URK OPERATION MANUAL Contents 1. Features 2. Notes for Using the Instrument 3. Description 3.1 Main Unit 3.2 Operation Buttons 4. Practicing through Model Eye 5. Measurement 5.1 Continuous Keratometry and Refractometry 5.2 Refractometry 5.3 Keratometry 5.4 Contact Lens Base Curve Measurement 5.5 Corneal diameter measurement



6. Other Measurement Modes 6.1 SETUP Mode



1. Features 1. Various Measurements Supported. Not only the usual refractometry and keratometry, but also corneal diameter and base curve of contact lens can be measured with this one instrument. Thus, measurements of eye and prescriptions for glasses and contact lenses can be made more efficiently.



2. Wide Dioptric Measurement Range Because the URK-700 covers a wide measurement range, from -25D to +22D, even an examinee with strong myopia can be measured.



3. More accurate Measurement.



The foggy method of the eye fixation target makes examinee’s eye comfortable and enables to get more accurate measurement data.



4. Easy Connection with other Equipments This instrument is designed to connect other equipments such as VCR or TV monitor.



2. Notes for Using the Instrument



1. Do not hit or drop the instrument. The instrument may be damaged when it receives a strong impact. The impact may cause damage the function of this instrument. Please do handle with care.



2. An exposure to the direct sunlight or too bright indoor lights may



influence on the result of accurate measurement.



3. To connect this with other relevant equipments, consult with the dealer about the way to make it.



4. A sudden heating of the room in cold areas will cause condensation on the protective glass in the monitor screen and on optical parts inside the instrument. In this case, just wait until condensation disappears before performing measurement.



5. Keep the objective glass of the examinee side clean. If it was stained, it may cause on ERROR or inaccurate measurements.



6. In case there is smoke, strange odor or noise on working disconnect the power supply and consult the dealer.



7. Don’t use organic solution such as alcohol, thinner, benzene, etc. to clean the surface of this instrument. It may damage the instrument.



8. When moving this URK-700, fix the stage by using clamping bolt and stage holding knob, always keep power off, and then lift the bottom of the unit with both hands.



9. If you leave URK-700 without using for certain period, disconnect the power supply and protect the unit with dust cover.



3. Description. 3.1 Main Unit.



LCD Monitor



Functional Key



Measurement button



Stage Clamping Bolt



Printer



Headrest



Measurement



Chinrest Stage Holding



EXT Video



RS-232 Port



Power IN LET



Fuse



Power Switch



3.2 Operation Panel



Title



Technical skill



MODE



To change measurement mode



CLEAR



To erase measured data on the screen



MENU



To change and set up the function



PRINT



To print the data of measurement



IOL



To measure the



patient’s eye with cataract or IOL(intra-ocular lens)



4. Practicing by Model Eye Practice measurement by using the Test Model Eye before doing the measurement of patient’s eye.



(1) Turn ON the power Turn ON the power switch on the lower left side. (2)



Set the model eye on Remove the chin rest paper and align the holes on the base of the model eye with the holes on the chin rest. And insert pins.



(3) Release stage lock Fully turn the bolt in the underneath cover in counter clockwise direction. And turn the stage holding bolt in counter clockwise direction to release the stage lock. (4) Adjust height of Test Model Eye. Adjust the height of the chin rest by rotating Height Adjustment knob so that The Test Mode Eye is aligned with the Height Adjustment mark on the headrest. (5) Enter K/R or REF Mode If neither K/R nor REF is displayed on the monitor press MODE button until either one is displayed. (6) Adjust the position and focus on the Model Eye. Looking at the monitor incline the operation joystick toward the Model Eye until a bright dot appears near the inner alignment mark. Place the bright dot in the center of inner alignment mark. Focus on the Model Eye by inclining the operation joystick forward and backward so that KERATO RING image is displayed clearly on the monitor, at the time adjusted the bar would be appeared above the inner alignment mark. (7) Measurement 



Manual measurement



A.



Adjust the position and focus on the Mode Eye like the procedure 6 .



B.



Press the measurement button. If measurement is failed with message like RETRY repeat the procedure A. and press the measurement button again.



C.



Check whether diopter value is measured or not. Diopter value is recorded in the bottom area of the Model Eye.







Automatic measurement



A.



To initialize program press MENU button.



B.



After user setting is displayed select AUTO START by cursor and measurement will start automatically. Then press EXIT button,



menu will be



saved and measurement mode appear. C.



When the bright dot enters the inner alignment mark and Model Eye is in focus properly measurement starts automatically.



5. Measurement 5.1 Continuous Keratometry and Refractometry ( K/R Mode) In the K/R Mode refractometry is performed after keratometry.



(1) Enter the K/R Mode A.



Press MODE button until “R/K” is displayed on the top of the display.



(2) Adjust height of examinee’s eye A.



Have the examinee sit and place his/her chin and forehead against the chin rest and forehead rest.



B.



Adjust the height of the chin by rotating the height adjustment knob so that the eye of the examinee is aligned with the height adjustment mark on the headrest.



(3) Perform Focusing A.



Slide the operation joystick to the left so that the right eye of the examinee is displayed on the monitor.



B.



Ask the examinee to look at the red balloon in the center of the scene (eye fixation target).



C.



Looking at the monitor, check that image of the KERATO RING is not obscured by the upper eyelid, if it is covering the ring, instruct the examinee to keep his/her eye opened wide until measurement ends.



D.



Looking at the monitor, incline the operation joystick to right or left, and turn the operation joystick so the pupil is centered with the inner alignment mark and focus on the KERATO RING, at the time adjusted the bar would be appeared above the inner alignment mark.



(4) Measurement A.



Press the Measurement button



B.



The newest measured result will be displayed on the monitor.



C.



According to the 6.1 SETUP MODE, you can select and change VD, CYL points.



(5) Repeated measurement A.



Measurement could be repeatedly performed if necessary.



B.



The newest result will be displayed on the monitor whenever measured.



C.



CAUTION: before starting the new measurement you have to press CLEAR button to remove previous data.



(6) Measure the other eye. A.



Slide the stage to the right side and measure the left eye.



B.



After both eyes are measured, pupil distance (PD) will be displayed on the monitor.



(7) Print A.



Press the PRINT button.



B.



The selected data in the 6.1SETUP Mode will be printed.



5.2 Refractometry (REF Mode) Only refractometry can be performed in the REF mode.



(1) Enter the mode. A. Press the MODE button until “REF ” appears on the top of display



(2) Adjust height of examinee’s eye A.



Have the examinee sit and place his/her chin and forehead against the chin rest and forehead rest.



B.



Adjust the height of the chin by rotating the height adjustment knob so that the eye of the examinee is aligned with the height adjustment mark on the face rest.



(3) Perform Focusing A.



Slide the operation joystick to the left so that the right eye of the examinee is displayed on the monitor.



B.



Ask the examinee to look at the red balloon in the center of the scene (eye fixation target).



C.



Looking at the monitor, check that image of the KERATO RING is not obscured by the upper eyelid, if it is covering the ring, instruct the examinee to keep his/her eye opened widely until measurement ends.



D.



Looking at the monitor, incline the operation joystick to right or left, and turn the operation joystick so the pupil is centered with the inner alignment mark and focus on the KERATO RING, at the time adjusted the bar would be appeared above the inner alignment mark.



(4) Measurement A.



Press the Measurement button



B.



The newest measured result will be displayed on the monitor.



C.



According to the 6.1 SETUP MODE, you can select and change VD, CYL points.



(5) Repeated measurement A.



Measurement could be repeatedly performed if necessary.



B.



The newest result will be displayed on the monitor whenever measured.



C.



CAUTION: before starting the new measurement you have to press CLEAR button to remove previous data.



(6) Measure the other eye. A.



Slide the stage to the right side and measure the left eye.



B.



After both eyes are measured, pupil distance (PD) will be displayed on the monitor.



(7) Print A.



Press the PRINT button.



B.



The selected data in the 6.1SETUP Mode will be printed.



5.3 Keratometry (KER Mode) Only the radius of curvature of the cornea can be measured in KER Mode.



(1) Enter the KER Mode. A.



Press the MODE button until “KER” is displayed in the top of the display



(2) Adjust height of examinee’s eye A.



Have the examinee sit and place his/her chin and forehead against the chin rest and forehead rest.



B.



Adjust the height of the chin by rotating the height adjustment knob so that the eye of the examinee is aligned with the height adjustment mark on the face rest.



(3) Perform Focusing A.



Slide the operation joystick to the left so that the right eye of the examinee is displayed on the monitor.



B.



Ask the examinee to look at the red balloon in the center of the scene (eye fixation target).



C.



Looking at the monitor, check that image of the KERATO RING is not obscured by the upper eyelid, if it is covering the ring, instruct the examinee to keep his/her eye opened widely until measurement ends.



D.



Looking at the monitor, incline the operation joystick to right or left, and turn the operation joystick so the pupil is concentric with the inner alignment mark and focus on the KERATO RING, at the time adjusted the bar would be appeared above the inner alignment mark.



(4) Measurement



A.



Press the Measurement button



B.



The newest measured result will be displayed on the monitor.



C.



According to the 6.1 SETUP Mode you can select and change VD, CYL, KERATO points.



(5) Repeated measurement A.



Measurement could be repeatedly performed if necessary.



B.



The newest result will be displayed on the monitor whenever measured.



C.



CAUTION: before starting the new measurement you have to press CLEAR button to remove previous data.



(6) Measure the other eye. A.



Slide the stage to the right side and measure the left eye.



B.



After both eyes are measured, pupil distance (PD) will be displayed on the monitor.



(7) Print A.



Press the PRINT button.



The selected data in the 6.1SETUP Mode will be printed.



5.4 Contact Lens Base Curve Measurement (CLBC mode) Base curve of hard contact lens can be measured in the CLBC mode.



(1) Enter the Mode. A.



Press the MODE button until “CLBC MODE” is displayed in the top of the display



(2) Attach Contact Lens. A.



Put water in the concave section of contact lens holder at backside of the TEST Model Eye and place the contact lens on the holder with the concave surface facing upward. Take care that the contact lens not to drop.



(3)



Attach the test Model Eye. A.



Remove the chin rest paper. Fix the Model Eye attached with contact lens using fixation pin.



(4)



Perform Focusing A.



Looking at the monitor, incline the operation joystick to right or left, and turn the operation joystick so the pupil is centered with the inner alignment red mark.



B.



Moving operation joystick straight and back fix up KERATO RING to be clear at the time adjusted the bar would be appeared above the inner alignment mark.



(5) Measurement A.



Press the Measurement button



B.



The newest measured result will be displayed on the monitor.



C.



According to the 6.1 SETUP Mode you can select and change VD, CYL, KERATO points.



(6) Repeated measurement A.



Measurement could be repeatedly performed if necessary.



B.



The newest result will be displayed on the monitor whenever measured.



C.



CAUTION: before starting the new measurement you have to press CLEAR button to remove previous data.



(7) Print A.



Press the PRINT button.



The selected data in the 6.1SETUP Mode will be printed. 5.5 Corneal diameter Measurement (SIZE Mode) Diameter of cornea can be measured in the SIZE mode. (1) Enter the A.



Mode



Press MODE button until “SIZE MODE” is displayed in the top of the display.



(2) Adjust height of examinee’s eye A.



Have the examinee sit and place his/her chin and forehead against the chin rest and forehead rest.



B.



Adjust the height of the chin by rotating the height adjustment knob so that the eye of the examinee is aligned with the height adjustment mark on the headrest.



(3) Perform Focusing A.



Slide the operation joystick to the left so that the right eye of the examinee is displayed on the monitor.



B.



Ask the examinee to look at the red balloon in the center of the scene ( eye fixation target ).



C.



Looking at the monitor, check that image of the KERATO RING is not obscured by the upper eyelid, if it is covering the ring, instruct the examinee to keep his/her eye opened widely until measurement ends.



D.



Looking at the monitor, incline the operation joystick to right or left, and turn the operation joystick so the pupil is centered with the inner alignment mark and focus on the KERATO RING.



(4) Measurement A.



Press the Measurement button



B.



To move the left/right bar by pressing the →/←button.



C.



The newest measured result will be displayed on the monitor.



D.



According to the 6.1 SETUP Mode you can select and change VD, CYL, KERATO points.



(5) Repeated measurement A.



Measurement could be repeatedly performed if necessary.



B.



The newest result will be displayed on the monitor whenever measured.



C.



CAUTION: before starting the new measurement you have to press CLEAR button to remove previous data.



(6) Measure the other eye. A.



Slide the stage to the right side and measure the left eye.



(7) Print A.



Press the PRINT button.



B.



The selected data in the 6.1SETUP Mode will be printed.



6. Other Measurement Mode 6.1 SETUP MODE



Change all the settings about measurements, printouts.



(1)



STEP To change measurement data units. Select 0.25/0.12 pressing button →/←



(2)



VD To select Vertex Distance (VD) Select 0/10/12/13.5/15 pressing button →/←



(3)



CYL To select Cylinder value and form Select -/+/+- pressing button →/←



(4)



START To select Manual or Auto measurement Select Manual/Auto pressing button →/←.



(5)



REPEAT To select repeated measurements. Select 1/3/5/7 pressing button →/←.



(6)



TARGET Target LED brightness



control (“1” is the brightest)



Select 1/2/3/4/5/6/7 pressing button →/←. (7)



MONITOR Monitor brightness control



(“1” is the brightest)



Select 1/2/3/4 pressing button →/←



(8)



PRINT Change printout forms. select ALL/ECONO/OFF pressing button →/←.



(9)



PRINT NO. Printout Numbering. Select ON/OFF/RESET pressing button →/←.



(10) DATA Press PRINT button after DATA ON to display the latest measurements of KER mode or REF mode. select ON/OFF pressing button →/←. (11) SAVE Anytime you don’t operate the instrument it will turn to the power saving mode. select OFF/3/5/10 pressing button →/←. (12) MESSAGE Enter a message to be printed out with the results of measurement using the internal printer in this window.. To display the user’s message select the one and press EDIT button. Using buttons ←/↑/↓ select message and press SET button. You can edit the message using →,←, DEL, BAK buttons . then press SET button again. Select END and SET button to complete message. (13) DATE FORM To setup order of displaying the date. select YMD/DMY/MDY pressing button →/←. (14) 24H MODE to change 12 hour/24 hour time change 12H/24H pressing button →/←. (15) DATA To setup year/month/day Select year/month/d ay pressing button →/←and use +/- button. (16) TIME To setup hour/minute/second Select



hour/minute/second pressing button →/←and use +/- button.



(17) CODE To initialize program on the occasion of occurring errors among used data. After pressing EDIT “5” Reload the factory data? will appear. If you answer “yes” program can be initialized.