Operation Manual For MEC1000 h22302-13 [PDF]

  • 0 0 0
  • Suka dengan makalah ini dan mengunduhnya? Anda bisa menerbitkan file PDF Anda sendiri secara online secara gratis dalam beberapa menit saja! Sign Up
File loading please wait...
Citation preview

MEC-1000



Portable Multi-parameter Patient Monitor



Operation Manual



Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaption and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.



,



,



,



and



are the registered



trademarks or trademarks owned by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners. Contents of this manual are subject to changes without prior notice.



For this Operation Manual, the issued Date is July 2005 (Version: 1.3). © 2002-2005 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.



I



Manufacturer’s Responsibility All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this manual. Mindray is responsible for safety, reliability and performance of this product only in the condition that: all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel; and, the electrical installation of the relevant room complies with the applicable national and local requirements; and, this product is operated under strict observance of this manual.



Warranty This warranty is exclusive and is in lieu of all other warrantyies, expressed or implied, including warranties of merchantability or fitness for any particular purpose.



Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel. This warranty shall not extend to any Mindray product which has been subjected to misuse, negligence or accident; or any Mindray product from which Mindray's original serial number tag or product identification markings have been altered or removed; or any product of any other manufacturer.



II



Return Policy In the event that it becomes necessary to return a unit to Mindray, follow the instructions below.



1.



Obtain a return authorization.



Contact the Mindray Service Department and obtain a Mindray Customer Service Authorization Number. The Mindray Customer Service Authorization Number must appear on the outside of the shipping container. Return shipments will not be accepted if the Mindray Customer Service Authorization Number is not clearly visible. Please provide the model number, serial number, and a brief description of the reason for return.



2.



Freight policy



The customer is responsible for freight charges when this product is shipped to Mindray for service (including any relevant customs fees or other freight related charges).



3.



Return address



Please send the part(s) or equipment to the address offered by Customer Service Department.



III



Contact Information Manufacturer:



Shenzhen Mindray Bio-Medical Electronics Co., Ltd.



Address:



Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen 518057 P.R. China



Tel:



+86 755 26522479



+86 755 26582888



Fax:



+86 755 26582500



+86 755 26582501



Website:



www.mindray.com.cn



EC Representative:



Shanghai International Holding Corp. GmbH (Europe)



Address:



Eiffestraße 80, 20537 Hamburg Germany



Tel:



0049-40-2513175



Fax:



0049-40-255726



IV



Equipment Symbols This symbol means 'BE CAREFUL '. Refer to the manual. This symbol indicates that the instrument is IEC 60601-1 Type CF equipment. The unit displaying this symbol contains an F-Type isolated (floating) patient applied part providing a high degree of protection against shock, and is suitable for use during defibrillation.



Equipotential grounding system. Protective earth ground.



Power On/Off This mark means that this device is fully in conformance with the Council Directive Concerning Medical Devices 93/42/EEC. The number adjacent to the CE marking (0123) is the number of the EU-notified body that certified meeting the requirements of Annex II of the Directive. The following definition of the WEEE label applies to EU member states only. This symbol indicates that this product should not be treated as household waste. By ensuring that this product is disposed of correctly, you will help prevent bringing potential negative consequences to the environment and human health. For more detailed information with regard to returning and recycling this product, please consult the distributor from whom you purchased it. * For system products, this label may be attached to the main unit only. NOTE: Points to be noted. CAUTION: Points to be noted to avoid damage to the equipment. WARNING: Points to be noted to avoid injury to the patient and the operator.



V



FOR YOUR NOTES



VI



Contents Chapter 1



Introduction........................................................................................................ 1-1



1.1 General Information .................................................................................................. 1-2 1.2 Screen Display .......................................................................................................... 1-4 1.3 Button Functions ....................................................................................................... 1-7 1.4 Interfaces ................................................................................................................ 1-10 1.5 Built-in battery ......................................................................................................... 1-12 Chapter 2



Getting Started ................................................................................................. 2-1



2.1 Unpacking and Checking .......................................................................................... 2-1 2.2 Connect the Power Cables ....................................................................................... 2-1 2.3 Power On the Monitor ............................................................................................... 2-2 2.4 Connect Sensors/Probes .......................................................................................... 2-2 2.5 Check the Recorder .................................................................................................. 2-2 Chapter 3



System Menu ................................................................................................... 3-1



3.1 Patient Setup ............................................................................................................. 3-2 3.2 Default Setup............................................................................................................. 3-4 3.3 System Setup ............................................................................................................ 3-5 3.4 Selection Setup ....................................................................................................... 3-10 3.5 Monitor Version ....................................................................................................... 3-11 3.6 Drug Calculation...................................................................................................... 3-12 3.7 Maitenance .............................................................................................................. 3-12 3.8 DEMO Function....................................................................................................... 3-15 Chapter 4



Face Select ...................................................................................................... 4-1



4.1 Standard Screen ....................................................................................................... 4-1 4.1 Viewbed Screen ........................................................................................................ 4-2 Chapter 5



Alarm................................................................................................................ 5-1



5.1 Alarm Modes ............................................................................................................. 5-1 5.2 Alarm Self-test........................................................................................................... 5-4 5.3 Alarm Cause.............................................................................................................. 5-4 5.4 SILENCE/CLOSE/PAUSE ........................................................................................ 5-5 5.5 Parameter Alarm ....................................................................................................... 5-6 5.6 When an Alarm Occurs ............................................................................................. 5-6 Chapter 6



Freeze ............................................................................................................... 6-1



6.1 General Description .................................................................................................. 6-1 6.2 Entering and Exiting Frozen Mode............................................................................ 6-1 6.3 Frozen Menu ............................................................................................................. 6-2 6.4 Recording Frozen Waveform .................................................................................... 6-2 Chapter 7



Recording .......................................................................................................... 7-1



7.1 General Description .................................................................................................. 7-1 7.2 Recording Type ......................................................................................................... 7-1 7.3 Start Recording ......................................................................................................... 7-4 1



Contents



7.4 Recorder Operations and Status Information ........................................................... 7-5 Chapter 8



Trend and Event .............................................................................................. 8-1



8.1 Trend Graph .............................................................................................................. 8-1 8.2 Trend Table................................................................................................................ 8-3 8.3 NIBP Recall ............................................................................................................... 8-5 8.4 Alarm Event Recall .................................................................................................... 8-5 Chapter 9



Drug Calculation and Titration Table................................................................ 9-1



9.1 Drug Calculation........................................................................................................ 9-1 9.2 Titration Table ............................................................................................................ 8-4 Chapter 10



Patient Safety................................................................................................. 10-1



Chapter 11



Maintenance / Cleaning ................................................................................. 11-1



11.1 System Check ....................................................................................................... 11-1 11.2 General Cleaning .................................................................................................. 11-1 11.3 Cleaning Agents .................................................................................................... 11-2 11.4 Sterilization ............................................................................................................ 11-3 11.5 Disinfection............................................................................................................ 11-3 Chapter 12



ECG/RESP Monitoring ................................................................................ 12-1



12.1 What Is ECG Monitoring ....................................................................................... 12-1 12.2 Precautions ........................................................................................................... 12-1 12.3 Monitoring Procedure............................................................................................ 12-2 12.4 ECG Screen Hot Keys .......................................................................................... 12-5 12.5 ECG Menu ............................................................................................................ 12-6 12.6 ECG Alarm Information and Prompt ................................................................... 12-10 12.7 ST Segment Monitoring (optional) ...................................................................... 12-11 12.8 Arr. Monitoring (optional)..................................................................................... 12-15 12.9 Measuring RESP................................................................................................. 12-20 12.10 Maintenance and Cleaning ............................................................................... 12-24 Chapter 13



SpO2 Monitoring ............................................................................................ 13-1



13.1 What is SpO2 Monitoring....................................................................................... 13-1 13.2 Precautions ........................................................................................................... 13-2 13.3 Monitoring Procedure............................................................................................ 13-3 13.4 Limitations for Measurement................................................................................. 13-5 13.5 SpO2 Menu............................................................................................................ 13-6 13.6 Alarm Description and Prompt .............................................................................. 13-8 13.7 Maintenance and Cleaning ................................................................................... 13-9 Chapter 14



NIBP Monitoring............................................................................................. 14-1



14.1 Introduction ........................................................................................................... 14-1 14.2 NIBP Monitoring .................................................................................................... 14-1 14.3 NIBP SETUP menu............................................................................................... 14-6 14.4 NIBP Alarm Message .......................................................................................... 14-11 14.5 Maintenance and Cleaning ................................................................................. 14-13 Chapter 15



TEMP Monitoring ........................................................................................... 15-1



15.1 TEMP Monitoring .................................................................................................. 15-1 15.2 TEMP SETUP Menu ............................................................................................. 15-2



2



Contents



15.3 TEMP Alarm message .......................................................................................... 15-3 15.4 Maintenance and Cleaning ................................................................................... 15-3 Chapter16



Accessories and Ordering Information .......................................................... 16-1



16.1 ECG Accessories .................................................................................................. 16-1 16.2 SpO2 Accessories ................................................................................................. 16-2 16.3 NIBP Accessories.................................................................................................. 16-2 16.4 TEMP Accessories ................................................................................................ 16-3



Appendix A EC Declaration of Conformity......................................................................A-1 Appendix B Product Specification ..................................................................................B-1 Appendix C EMC............................................................................................................C-1 Appendix D System Alarm Prompt .................................................................................D-1



3



Contents



FOR YOUR NTOES



4



Chapter 1 Introduction For an overview of the monitor, refer to 1.1 General Information. For various messages displayed on the screen, refer to 1.2 Screen Display. For basic operations, please refer to 1.3 Button Functions. For allocation of interface sockets, please refer to 1.4 Interfaces. For battery recharging procedure, refer to 1.5 Built-in Battery. For safety precautions of the monitor, refer to Chapter 10 Patient Safety.



Warning The Monitor is intended for use by qualified clinical Personnel. Its operation is to be performed by a physician or appropriate medical staff under the direction of a physician. The monitor can monitor single patient once. Opening the monitor housing presents a risk of hazard due to electrical shock. All servicing and future upgrades to this equipment must be carried out by personnel trained and authorized by Mindray. Possible explosion hazard if used in the presence of flammable anesthetics. You must verify if the device and accessories can function safely and normally before use. You must customize the alarm setups according to individual patient situation and make sure that alarm sound can be activated when alarm occurs. Do not use cellular phone in the vicinity of this device. High level electromagnetic radiation emitted from such devices may greatly affect the monitor performance. Do not touch the patient, table, or the device during defibrillation. Devices connected to the monitor shall form an equipotential system (protectively earthed). When used with Electro-surgery equipment, you (doctor or nurse) must give top priority to the patient safety. Dispose of the packaging material, observing the applicable waste control regulations 1-1



Introduction



and keeping it out of children’s reach.



Note This equipment complies with the standard CISPR11 (EN55011) class A. The software was developed per IEC601-1-4. The possibility of hazards arising from errors in the software program is minimized. Caution At the end of its service life, the product described in this manual, as well as its accessories, must be disposed of in compliance with the guidelines regulation the disposal of such products. If you have questions concerning disposal of the product, please contact Mindray Customer Service.



1.1



General Information



Environmental Specifications Temperature Working



0 ~ 40 °C



Transport and Storage -20 ~ 60 °C Humidity Working



15% ~ 95%, noncondensing



Transport and Storage 10% ~ 95%, noncondensing Altitude Working



-500 ~ 4,600m



Transport and Storage -500 ~ 13,100m Power Supply 100 ~ 240V, AC, 50/60Hz Pmax=110VA FUSE T 1.6A Contraindications: None



Note The system may not meet its performance specifications if stored or used outside the specified temperature and humidity range.



General Instruction The Monitor is applicable for bedside monitoring of Adult, Pediatric and Neonate. The Monitor is intended to monitor vital signals such as ECG, Respiratory Rate, SpO2, NIBP and TEMP. It integrates parameter measurement, display and recorder in one device, featured by compactness, lightweight and portability. Its large and high-resolution display can clearly display 4 waveforms and all parameter information. 1-2



Introduction



Front Panel ③







⑤ ①



④ Figure 1-1



Patient Monitor



① POWER Switch: On the bottom left quarter of the front panel. ② CHARGE Indicator: A LED above the POWER Switch. ③ ALARM Indicator: A LED on the upper side of the front panel. The ALARM indicator flashes or lights when an alarm occurs. The visible LEDs are CLASS 1 LED PRODUCT according with EN 60825-1 A11 Oct 1996. ④ Control Panel: Refer to 1.3 Button Functions for details. ⑤ Rotary Knob: Refer to 1.3 Button Functions for details. The sockets of the sensors are at the left side. The recorder socket is at the right side. Other sockets and power plug-in are at the rear panel.



Monitored Parameters Heart Rate (HR) ECG



1-channel ECG waveform Arrhythmia and S-T segment analysis, Pace analysis



RESP



SPO2



Respiratory Rate (RR) Respiration Waveform Oxygen Saturation (SpO2), Pulse Rate (PR) SpO2 Plethysmogram



NIBP



Systolic Pressure (NS), Diastolic Pressure (ND), Mean Pressure (NM)



TEMP



Temperature DATA



The monitor has additional functions including visual & audible alarm, trend data storage and output, NIBP data storage and review, alarm event identification, and drug calculation. 1-3



Introduction



1.2



Screen Display



The display of the monitor is a color TFT, Which can display the collected patient parameters, waveforms, alarm information as well as bed number, time and monitor status, etc. As shown in Figure 1-2, the main screen is divided into three areas: ①











① Figure 1-2



Main Screen



① Message Area; ② Waveform Area; ③ Parameter Area.



① Message Area The Message Area is at top and bottom part of the screen, displaying the current status of both the monitor and the patient. The top part displays “patient information”, “monitoring status” and “alarm message” from left to right. The “monitoring status” and “alarm message” appear or disappear together with the status being reported. Patient information include: BED NO



Bed numbers of all patients under monitoring



Patient type



Three options: Adult, Pediatric, Neonate



“2000-01-08”



Current date: “year - month - day”



“00:49:30”



Current time: “hour : minute : second”



Patient name



This item will display blank if the operator does not input patient name



Patient sex



Male of Female



Monitoring status reporting the current status of the monitor or sensor/probe, which always appears to the right side of the system time. When this information appears, it will cover patient sex and name. 1-4



Introduction



Alarm message icon for alarm PAUSE. Press “SILENCE” button once (less than 1 second) to suspend all alarm sounds for the pause duration. Press the button again to terminate the PAUSE status. The duration for PAUSE status can be set for 1 minute, 2 minutes, or 3 minutes.



icon for alarm SILENCE. Press and hold the “SILENCE” button for more than 1 second to manually mute the alarm sound and this icon will be displayed. The SILENCE status terminates when you press “SILENCE” button again or new alarm occurs.



icon for Alarm Volume Off. It appears indicating that you have turned the alarm sound OFF. To terminate this function, you must turn the Alarm Volume ON.



Note If the



symbol appears, the system will no longer give an audible alarm sound.



You must be very careful in using this function.



② Waveform/Menu Area This area can display four waveforms under normal operation mode. From the top to the bottom, there are 2 ECG waveforms (one-channel cascade ECG waveform), SpO2 Plethysmogram waveform, and RESP waveform. You can choose the waveforms for display. Refer to 3.3.6 Tracing Waveforms Selection for details. The gain, channel as well as filter way of the ECG waveform are also displayed above the ECG waveform. A 1mv scale bar is displayed to the right side of the ECG waveform for reference. The name of other waveform is displayed at the upper left of the waveform. When the functional menu is accessed during normal monitoring operation, the menu always occupies a fixed position in the middle part of the waveform area. Therefore, part of the waveform can’t be viewed. After exiting the menu, the monitor will return to the original display. You may select the refresh rate for the waveforms. The method to adjust the refresh rate for each waveform is discussed in the setup description for each parameter.



③ Parameter Area Parameters are displayed on the right side of the waveform area, and their positions basically 1-5



Introduction



correspond to the waveform. The measured parameters displayed in this area include: ①HR or PR ②ST Anlysis ③PVC ④NIBP



⑤SPO2 ⑥RESP ⑦TEMP



Figure 1-3



Main Screen



ECG: ⎯ ① Heart Rate ( Unit: bpm) ⎯ ② ST-segment analysis result (Unit: mV) ⎯ ③ Arrhythmia (PVCs) events (Unit: event/min) NIBP: — ④ Systolic, Mean, Diastolic (From left to right) (Unit: mmHg or kPa) SpO2: ⎯ ⑤ SpO2 (Unit: %) RESP: ⎯ ⑥ Respiration Rate (Unit: breath/min) TEMP: ⎯ ⑦ Temperature (Unit:℃ or ℉) The system refreshes each parameter value once per second except NIBP, which is refreshed after each measurement.



Alarm indicator and status During normal monitoring, the alarm indicator is not on. When an alarm occurs, the alarm indicator lights on or flashes. The color of the indicator corresponds to the alarm priority level. Refer to Chapter 5 Alarm for detailed information. For the details of the alarm and alarm message information for the individual measured parameters, refer to the chapter for each parameter.



1-6



Introduction



Warning Always verify the self-check function of the audible and visual (LED) alarms when the power is turned on.



1.3



Button Functions



The operation of the monitor is performed using a few buttons and a Rotary Knob on the control panel, located at the lower right of the monitor.















Figure 1-4



















Buttons and Knob



① MAIN The screen will return to the main screen when this button is pressed. ② FREEZE Press this button and the monitor will switch to the FREEZE mode. Press this button again to exit the FREEZE mode. In FREEZE mode, the frozen waveform can be printed out. For detailed information, refer to Chapter 6 Freeze. ③ SILENCE Press this button to access the Alarm Pause status. All the alarm sounds are muted. However, other sounds such as heart beat, button sound and pulse sound will still exist. At this time, the “ALARM PAUSE xx s” appears in the parameter alarm information area. Press this button again or when the pause time runs out, the system terminates the alarm pause status and returns to the normal monitoring status, then the alarm sound resumes and the prompt “ALARM PAUSE xx s” disappears from the screen.



Note When the monitor exits the Alarm Pause status, whether the alarm sound is restored depends on whether the alarm exists. However, the technical alarms such as lead off will not be restored. That means the user may also close the technical alarms such as lead off through accessing and then exiting the Alarm Pause status. Press this button for more than 1 second to access the Silence status. The system will mute all kinds of sound such as heart beat, pulse and button sound and all the current alarm sound will be muted. The symbol



appears in the information area. Press the



SILENCE once again to restore all kinds of sound and the



1-7



disappears.



Introduction



In the Alarm Silence status, if there is new alarm, the system will terminate the Silent status and restore the normal alarming status. At this time, the system gives alarm according to the setup alarm sound. For the detailed rules, refer to related chapter: Alarm. ④ RECORD Press this button to start a real time recording. The recording time is set in REC TIME of RECORD SETUP submenu. Press during recording to stop the recording. For detailed information, refer to Chapter 7 Recording. ⑤ START Press this button to inflate the cuff to start a blood pressure measurement. When measuring, press to cancel the measurement and deflate the cuff. ⑥ MENU Press this button to display the SYSTEM MENU, in which the user may set up system information and perform review operation. For detailed information, refer to Chapter 3 System Menu and Chapter 8 Trend and Event. ⑦ Rotary Knob Use the rotary knob to select menu items and modify the monitor setup. It can be rotated clockwise or counter-clockwise, and pressed to make a desired selection, similar to other buttons.



How to use the Rotary Knob The rectangular mark on the screen that moves with the rotation of the knob is called a “cursor”. Operation can be performed at any position at which the cursor is located. When the cursor is in the waveform area, you may immediately modify the current setup by selecting one of the following hot keys: ECG lead name ECG gain ECG filter When the cursor is in the parameter area, you may open the setup menu of the corresponding parameter to review or change the menu items. ECG menu NIBP menu SpO2 menu RESP menu TEMP menu Highlight any item that you want to operate and press the knob, the system will fulfill one of the following three activities: A menu pops up, or the current menu is replaced by a new one; The solid frame becomes a dotted one, indicating that rotating the knob can change its content;



1-8



Introduction



The system can immediately execute an operation.



Basic Operations To display desired waveform: Press the MENU button to access the SYSTEM MENU and then select the [TRACE SETUP] item. The TRACE SETUP menu will appear, in which you can select the desired waveform. Refer to 3.3.6 Tracing Waveforms Selection for details. To adjust waveform sweep speed: ECG: access the ECG SETUP menu and select the [SWEEP] item; PLETH: access the SpO2 SETUP menu and select the [SWEEP] item; RESP: access the RESP SETUP menu and select the [SWEEP] item. To change Alarm limits: ECG: access the ECG SETUP menu and select the [ALM HI] or [ALM LO] item. ST: access the ECG SETUP menu and select the [ST ANALYSIS], then select the [ALM HI] or [ALM LO] item. PVCs: access the ECG SETUP menu and select the [ARR ANALYSIS], then select the [ALM HI] or [ALM LO] item. SpO2: access the SPO2 SETUP menu and select the [SpO2 ALM HI] or [SPO2 ALM LO] item. PR: access the SPO2 menu and select the [PR HI] or [PR ALM LO] item. NIBP Systolic: access the NIBP SETUP menu and select the [SYS ALM HI] or [SYS ALM LO] item. NIBP MEAN: access the NIBP SETUP menu and select the [MEAN ALM HI] or [MEAN ALM LO] item. NIBP Diastolic: access the NIBP SETUP menu and select the [DIA ALM HI] or [DIA ALM LO] item. RESP: access the RESP SETUP menu and select the [ALM HI] or [ALM LO] item. TEMP: access the TEMP SETUP menu and select the [ALM HI] or [ALM LO] item. To record real-time waveform: Push RECORD button to start the real-time recording process. Refer to Chapter 7 Recording for details. To adjust volume: Alarm volume: Press the MENU button to access the SYSTEM MENU \ SYSTEM SETUP \ ALARM SETUP and select the [ALARM VOL] item. Key volume: Press the MENU button to access the SYSTEM MENU \ SELECTION and select the [KEY VOL] item. Beat volume: Access the ECG SETUP\OTHER SETUPS and select the [BEAT VOL]. You can select 0~10. 1-9



Introduction



Pulse volume: Access the SPO2 SETUP Menu and select the [PR SOUND]. To set up system date and time: Press the MENU button to access the SYSTEM MENU \ SYSTEM SETUP and select the [TIME SETUP]. To restore default settings: Refer to 3.2 Default Setup for details.



1.4



Interfaces



For your convenience, we put different interfaces on different positions of the monitor. Recorder is on the right side of the monitor.



③ ① ④











Figure 1-5 Right Panel



Figure 1-6 Left Panel



Patient cable and sensor/probe sockets are on the left side of the monitor. As shown in Figure 1-6, there are: ①



Socket for ECG cable







Socket for NIBP probe







Socket for TEMP probe







Socket for SpO2 sensor



This symbol means “BE CAREFUL". Refer to the accompanying document (this manual).



This symbol indicates that the instrument is IEC60601-1 Type CF equipment. The unit displaying this symbol contains an F-Type isolated (floating) patient applied part providing a 1-10



Introduction



high degree of protection against shock, and is suitable for use during defibrillation. Other symbols in the monitor are explained in Chapter 10 Patient Safety.



① ③











Figure 1-7 Back Side On the rear panel are the following connectors, shown in Figure 1-7. ① Network Interface: Standard RJ45 Socket. ② Equipotential grounding terminal for connection with the hospital’s grounding system. ③ Fuse: T 1.6A ④ AC Power Input Connector: 100~240 (VAC), 50/60 (Hz). Network interface is a multi-functional interface, which support 100 BASE-TX Ethernet communication channel, and Mindray Hyper III and CMS network protocol, as well as online software upgrading function by connecting with a PC.



Warning Network interface is can be used to upgrade the system, and is only executed by Mindray authorized personnel Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC standards (e.g. IEC 60950 for data processing equipment and IEC 60601-1 for medical equipment). Furthermore all configurations shall comply with the valid version of the system standard IEC 60601-1-1. Everybody who connects additional equipment to the signal input part or signal output part configures a medical system, and is therefore responsible that the system complies with the requirements of the valid version of the system standard IEC 60601-1-1. If in doubt, consult Mindray Customer Service.



1-11



Introduction



1.5



Built-in Battery



The Monitor is equipped with rechargeable batteries. The battery in the Monitor will automatically recharge when AC power is applied until it is full. A symbol “



” is displayed on



the lower left quarter of the screen to indicate the battery status. The solid part represents the relative electric energy of the battery. This symbol will be covered when some information appears. And, if no battery is installed, battery state will be displayed as “



” under a cross to



indicate that no battery is available. There are two battery slots inside the monitor, each can hold a battery, and one battery can support the monitor working. Battery can be installed into and pulled off from battery slot. Under connectors to patient cables there are battery slots with cover. See Figure 1-8 Battery Slot Cover.



Warning Don’t pull off battery when the monitor is working. When operating on battery, the monitor will sound an alarm and shut off automatically when the energy is low. When the battery is depleted, the monitor will sound continuous high-level alarm tone and display “BATTERY TOO LOW” in the Message Area. Connect the monitor to AC power at this moment can recharge the battery while operating. If the monitor continues to operate from the battery, the monitor will shut off automatically about 5 minutes after alarm.







Figure 1-8 Battery Slot Cover Warning After the patient monitor has been placed unused for a long time, the battery must be recharged before use again. When not using the patient monitor for a long time without charging the battery, the battery capacity will decrease. Connect the patient monitor to appropriate AC mains to charge the battery. Mindray recommends replace the sealed Lead-Acid battery once per year. Dispose or recycle the battery and other parts of the patient monitor by following local government regulations.



1-12



Chapter 2 Getting Started







Unpacking and Checking







Connect the Power Cable







Power On the Monitor







Connect Sensors/Probes







Check the Recorder



Note To ensure the monitor works properly, please read “Chapter 10 Patient Safety” and follow the steps indicated before using this monitor.



2.1 Unpacking and Checking Open the package and remove the monitor and accessories carefully. Save the packing material for possible future transportation or storage. Check the components by comparing the packing list. ■



Check for any mechanical damage.







Check all the cables and accessories for damage.



If there is any problem, contact Mindray Customer Service immediately.



2.2 Connect the Power Cable Following these steps to connect the AC power cable: Make sure the AC power supply complies with following specification: 100~240 VAC, 50/60 Hz. Use the power cable supplied with the monitor. Plug the power cable into the “AC Power Input connector” of the monitor. Connect the other end of the power cable into a 3-prong, grounded hospital grade power outlet. Connect the equipotential earth line if necessary. Refer to Chapter 10 Patient Safety for details.



Note Connect the power cable to a 3-prong, hospital grade outlet only.



2-1



Getting started



Note Mindray does not provide MULTIPLE PORTABLE SOCKET-OUTLETS. IF use it, please do not place it on the floor. Mindray advises that every one monitor uses one MULTIPLE PORTABLE SOCKET-OUTLETS.



2.3 Power On the Monitor Press POWER Switch to power on the monitor. A beep “Du…” will be heard and the alarm indicator will flash once in different color. After about 10 seconds, the system will self-test successfully and enter main monitoring screen. You can perform normal operations now. During self-test, the system can display machine version number.



Note If the system detects any fatal error during self-test, it will give alarm.



Note Check all usable functions in order to ensure the normal function of the monitor.



Warning If any sign of damage is detected, or the system displays some error messages, do not use it on any patient. Contact biomedical engineer in the hospital or Mindray Customer Service Center immediately.



Note You must turn on the monitor at least 1 minute after turning off the monitor.



2.4 Connect Sensors/Probes Connect the necessary patient sensors/probes to the monitor and the patient.



Note To know how to correctly connect sensors/probes, refer to Chapter 12-15.



2.5 Check the Recorder Check if the recorder on the right side of the monitor has paper. If there is no paper, refer to Chapter 7 Recording for details.



2-2



Chapter 3 System Menu







Patient Setup







Default Setup







System Setup







Selection Setup







Monitor Version







Drug Calculation







Demo Function







Maintenance



The Monitor features flexibility in monitoring configurations. You can customize the monitoring presentation, waveform sweep speed, sound volume, and output interface. Press the “MENU” button on the front panel to display the “SYSTEM MENU” as shown in Figure 3-1.



Figure 3-1



SYSTEM MENU



In this chapter, the submenus will be described one by one, except the “TREND GRAPH”, “TREND TABLE”, “NIBP RECALL” and “ALARM RECALL”, which will be described in Chapter 8 Trend and Event.



3-1



System Menu



3.1 Patient Setup Note To clear current patient data, refer to “New Patient” for details. Select [PATIENT SETUP] from the “SYSTEM MENU” to display the following menu. This is done by moving the cursor to [PATIENT SETUP] and pressing on the knob.



Figure 3-2



PATIENT SETUP



1. Input the patient information by pressing the knob and the cursor will jump to select the alphanumeric keypad located at the bottom of the menu. 2. Once a character is highlighted with the cursor, press the knob to enter that character. 3. Repeat the process until all the characters are input. 4. Select OK and press the knob. The cursor will return to the starting position. DEPT.



Department in which the patient receives treatment.



PAT NO



Patient Identification Number.



BED NO



Patient bed number (Range: 1~100)



DOCTOR



Name of the doctor.



NAME



Patient name (Valid characters: A~Z, 0~9 and space bar; Maximum of 12 characters)



SEX



Patient gender ("F" for Female, "M" for Male)



3-2



System Menu



PAT TYPE



Patient type (ADU, PED, and NEO for Adult, Pediatric, and Neonate respectively)



ADMIT



Hospitalization starting date (format: year\month\ day)



BIRTH



Patient date of birth (format: year\month\day)



HT. (cm/inch )



Patient height (Rotating the knob will increase or decrease the height in 0.5 cm or inch each time.) When select the HT. unit in the measured parameter area, the unit of measurement must correspond with the unit you select here.



WT. (kg/Ib)



Patient weight (Rotating the knob will increase or decrease the weight in 0.5 kg or pound each time.) When select the WT. unit in the measured parameter area, the unit of measurement must correspond with the unit you select here.



BLOOD



Patient blood type (Select A, B, O, AB, or N. "N" represents unknown blood type)



NEW PATIENT



Admission of new patient



When you select [NEW PATIENT], you access the “CONFIRM TO UPDATE PATIENT” menu as shown below. This allows you to delete previous patient data and begin monitoring a new patient.



Figure 3-3



Confirm To Update Patient Menu



1. Select [YES] to delete all information of the patient that was being monitored and exit the menu. 2. Select [NO] to exit the menu and the system will keep the information of the patient that was being monitored.



Note If you select [YES], the system will delete all information of previously monitored patient, or the patient being currently monitored.



3-3



System Menu



3.2



Default Setup



Note After selecting any item in this sub-menu, the selected item will replace the current setup of the system and accordingly become the system default configuration.



Figure 3-4



DEFAULT Menu



In this menu, you can select both the factory default and the user-defined default. You can also save the current system configuration as the user-defined default configuration. At this time, the system will automatically save all the setups in the parameter menus as the user-defined default configuration according to the patient type. After selecting any item in the DEFAULT menu and exiting that menu, the “CONFIRM DEFAULT CONFIG” menu will be displayed. You can select [YES] to confirm your selection or [NO] to give up your selection.



Figure 3-5



CONFIRM DEFAULT CONFIG



Warning All configurations will be replaced by “default configurations” if selecting [YES].



3-4



System Menu



3.3 System Setup Select the [system setup] item from the “SYSTEM MENU”.



Figure 3-6 System setup In the [SYSTEM SETUP] menu, you can configure the following items.



3.3.1 Face Select Select “FACE SELECT” item in “SYSTEM SETUP” menu to access “FACE SELECT” dialog box as shown below, in which two selections are available: STANDARD SCREEN and VIEWBED SCREEN. Only one selection can be chosen for each time.



Figure 3-7 FACE SELECT For details about STANDARD SCREEN and VIEWBED SCREEN, please refer to Chapter 4 Face Select.



3-5



System Menu



3.3.2 Alarm setup The monitor provides ten levels of alarm volume. You can select any of them depending on your specific requirement. 1. Select the [ALARM SETUP] item in the “SYSTEM SETUP” menu. You can select the alarm volume, and other alarm information. For detailed information, refer to Chapter 5 Alarm.



2



Figure 3-8 Alarm Setup 2. By moving cursor to [ALARM VOL] and rotating the knob, you can change the alarm volume. There are 10 options: “1~10”. “1” means the minimum volume, and “10” means the maximum volume.



3.3.3 Time Setup Select the [TIME SETUP] from the “SYSTEM SETUP” menu. The menu as shown below will be displayed. The system time is in the format of year, month, day, hour, minute and second. Use the cursor to highlight the item that you want to modify and rotate the knob to select time. Then select [EXIT]. Note You shall set up the system time upon turning on the monitor (if you need to set up the system time); otherwise, when you review the content with time information, the system may not display the correct time.



3-6



System Menu



Figure 3-9



System Time Setup



3.3.4 Recorder setup Select [RECORD] from the “SYSTEM SETUP” menu to display the “RECORD” menu, you can perform the following setup:



Figure 3-10



Record Setup



REC WAVE1; REC WAVE2 You can select up to two waveforms for recording. The waveforms that can be selected include: ECG



If no ECG waveform is displayed on the screen, you cannot select this item.



3-7



System Menu



SPO2



SpO2 Plethysmogram. (If no SpO2 Plethysmogram is displayed , you cannot select this item.)



RESP



RESP waveform (If no RESP waveform is displayed, you cannot select this item).



OFF



Do not select this waveform.



RT REC TIME: There are two options, CONTINUAL and 8s. “CONTINUAL”: When the recorder is started by pressing the “RECORD” button on the monitor panel, the recorder will continuously print out the waveform or parameter until this button is pressed again. “8s”: When the “RECORD” button is pressed, the recorder will run for 8 seconds and stops automatically. TIMING REC TIME: This is used to select a time interval between automatic recordings. Ten selections are available: “OFF, 10min, 20min, 30min, 40min, 50min, 1hour, 2hours, 3hours and 4hours”. The monitor will start recording at the selected time interval, record for 8 seconds, and automatically shut off.



Note RT REC TIME takes priority over TIMING REC TIME OFF. REC RATE: You can select either 25.0, or 50.0 mm/s. REC GRID: You can select ON to print a grid on the paper, or OFF for no grid. CLEAR REC TASK:



This is used to clear a recording event that has been waiting for



recording.



Note If two same waveforms are selected, the system will automatically change one of the waveform to a different one.



3-8



System Menu



3.3.5 Module Setup Select [MODULE SETUP] from the “SYSTEM SETUP” menu.



Figure 3-11



Module Setup



Select the parameters to be monitored. This will avoid displaying information for parameters that are not being monitored. If a module is not selected, its corresponding parameters and waveform will not be displayed.



3.3.6 Tracing Waveforms Selection Select the [TRACE SETUP] from the “SYSTEM SETUP” menu.



Figure 3-12



Tracing Waveforms Selection



You can choose the waveform to be displayed in the screen. If a parameter in this menu is not selected, its corresponding waveform will not be displayed. However, its corresponding measured parameters will still be displayed.



3-9



System Menu



3.3.7 Event Setup The monitor enables you to mark four different events. 1. Select [MARK EVENT] from the “SYSTEM SETUP”.



Figure 3-13



MARK EVENT Menu



2. Use the knob select event A, B, C or D. The “@” symbol will appear in the frame of the event being selected. If you make an incorrect selection, press the knob on the event again to unmark the selection. 3. Select [EXIT] to exit the menu and the mark function will be effective. The event marking function helps to classify the records into different categories, such as those having influence on patients versus those having influence on parameter monitoring. Example: Dose taking, injections, or therapy. A mark will be displayed on the trend graph/table indicating the time the mark was initiated in relation to the event it represents.



3.4 Selection Setup Select [SELECTION] from the “SYSTEM MENU”.



2



Figure 3-14 Selection Setup



3-10



System Menu



Key Volume: Select [KEY VOL] from the “SELECTION” menu. Rotate the knob to select the level of volume. There are 11 options: “0~10”. “0” means the key volume is turned off, and “10” means the maximum volume.



Help Function: The system incorporates On-line Help for the menu operations. You can choose any help information as needed. 1. Select [SELECTION] from the “SYSTEM MENU” 2. Highlight [HELP] and rotate the knob to select “ON” or “OFF”. When it is “ON”, you can browse the on-line help information. When it is “OFF”, the system will turn off the on-line help function.



Alarm Limits: The system can display the alarm limits. 1. Select [SELECTION] from “SYSTEM MENU”. 2. You can select [ALM LIMIT], and set to “ON” or “OFF”. When it is “ON”, the alarm limit will be displayed in the parameter area. When it is “OFF”, the alarm limit will not be displayed in the parameter area.



3.5 Monitor Version Select [VERSION] from the “SYSTEM MENU” to see the software version of the monitor.



Figure 3-15



Monitor Version



Due to continuing product innovation, the version information of your product may differ from that displayed in this menu. Select [DEVICE CONFIG LIST] to view the configuration of the monitor.



3-11



System Menu



Figure 3-16



Device Configuration List



3.6 Drug Calculation You can use the drug calculation and titration table function of the monitor to calculate the concentration of 15 kinds of drugs. Refer to Chapter 9 Drug Calculation and Titration Table for detailed information.



3.7 Maintenance 1. Select [MAINTAIN] from the “SYSTEM MENU”. 2. Enter the password into the frame under the “USER KEY”. 3. Select confirm to enter the “USER MAINTAIN” menu shown in Figure 3-18.



Figure 3-17



Enter Maintain Password 3-12



System Menu



Warning It is forbidden to execute factory maintenance function, which is only available for the service engineers of MINDRAY company.



Figure 3-18



User Maintain



In the “USER MAINTAIN” menu, you can set up following items: [LANGUAGE]: You can select the screen language you need. [LEAD NAMING]: You can select “AHA” or “EURO”. To know the difference between two styles, refer to Chapter 12 ECG/RESP Monitoring. [ALM SOUND]: You can set the alarm volume to “ON” or “OFF”.



Warning When the alarm volume is set to “OFF”, you will not hear the alarm sound if new alarm occurs. Therefore, you must be very careful in using this selection.



Warning If setting the alarm volume to “OFF” when the system is in Silence or Pause status, the system will automatically discharge Silence or Pause status.



3-13



System Menu



Warning If you select “Silence” or “Pause” when the alarm volume is set to “OFF”, the system will restore the alarm volume before the alarm volume is set to “OFF” and enter Silence or Pause status.



Note After the alarm volume is set to OFF, a



symbol will appear in the Technical Alarm



Area.



Note Setting Alarm Volume to “OFF” is valid only when the monitor is turned on for this time. After turning on the monitor next time, this setup will restore its value of the previous time when the system is turned on. [ALM PAUSE TIME]: You can set up the duration of Alarm Pause status. Three options are available, 1 minute, 2 minutes and 3 minutes. [TEMP SENSOR]: You can choose either “YSI” or “CY-F1”. “CY-F1” is a kind of TEMP probe made in China. [NET TYPE]: You can choose “HYPER III” or “CMS”. [LOCAL NET NO]: It refers to the net No. COLOR SELF-DEFINE This is used to define the color of the waveform displayed on the screen. Five colors can be chosen from: green, cyan, red, yellow and white.



Figure 3-19 Color Self-define



3-14



System Menu



3.8 DEMO Function Warning The purpose of the demonstration display is to demonstrate the performance of the monitor, and for training purpose only. In clinical application, this function is forbidden because the DEMO display can mislead the medical staff to treat the DEMO waveform and parameter as the actual data of the patient. 1. Select [DEMO] from the “SYSTEM MENU” to display the “ENTER DEMO PASSWORD”. 2. After entering the password, the system enters DEMO status.



Figure 3-20 Input Demo Key



3-15



System Menu



FOR YOUR NOTES



3-16



Chapter 4



Face Select



In the “SYSTEM MENU”, select [FACE SELECT] from the “SYSTEM SETUP” menu to display the figure below. There are two options in this dialog, which are “STANDARD SCREEN”, and “VIEWBED SCREEN”. Only one item can be selected at one time.



Figure 4-1 FACE SELECT



4.1



Standard Screen



Select the “STANDARD SCREEN” to enter the Standard Screen. The Standard Screen is the basic operating screen of the monitor.



Figure 4-2 STANDARD SCREEN 4-1



Face Select



4.2



Viewbed Screen



If another monitor is connected on the same LAN of this monitor, you can use this monitor to view any measured waveform and information about all measured parameters from another monitor. Enter Viewbed Screen Select the “VIEWBED SCREEN” option in the “FACE SELECT” menu. Viewbed Screen window occupies the space of the bottom four waveforms.



Figure 4-3



VIEWBED SCREEN



Hot key of Viewbed There are two hot keys in the Viewbed Screen: Select Bed Number and Select Waveform. The hot key of Select Bed Number displays the bed numbers and patient names of other monitors currently connected on the LAN. You can select a monitor to be monitored according to the patient name and bed number. If at this time no other monitors are connected on the same LAN of this monitor, the hot key of Bed Number will therefore display “N/A”. After you use this hot key to select a monitor to be viewed, the system will toggle to the display of the selected monitor for your view. The selected waveform is one of those listed in the hot key of Select Waveform. The hot key of Select Waveform is used to select a waveform generated by the monitor being viewed. If the hot key of Select Waveform displays “N/A”, it indicates that the bedside monitor being viewed has no waveforms. You can use this hot key to select and therefore view different waveforms of the monitor being viewed.



4-2



Face Select



Alarm indicator of Viewbed On the upper right side of the Viewbed Screen, there is an Alarm Indicator used to tell the alarm status of the monitor being viewed. The activity of this alarm indicator is identical with that of the alarm lamp on the panel of the monitor being viewed. That is to say, if the monitor being viewed occurs medium/low level alarm, this alarm indicator illuminates yellow; if it occurs high level alarm, this alarm indicator illuminates red. If the monitor being viewed has no alarm or the alarm is screened, the icon for this alarm indicator will not be displayed. Parameter area of Viewbed Screen Under the hot key of Select Bed Number is the Parameter area, in which parameters of all monitors being viewed are displayed. Waveform area of Viewbed Screen Under the hot key of Select Waveform is the Waveform area. The Sweep manner (refreshing or scrolling) of the waveform is identical with that of this monitor. The feature description of the displayed waveform is given above the waveform. Sweep speed is also identical with that set up for the same waveform on this monitor. Technical Information area Technical Information area is to the right of patient name in Viewbed Screen. This area displays related technical information to Viewbed, such as due to network failure or network too busy, Viewbed is disabled. Close Viewbed Screen In the FACE SELECT menu, select options of other operating screens to close the Viewbed Screen. Rules for automatically selecting monitor to be viewed and waveform When you turn on the monitor or enter Viewbed Screen, the system will automatically select a networked bedside monitor and a waveform of this monitor for you to view. If the monitor being currently viewed is disconnected, the viewed monitor will automatically close, clear displays of all alarms, parameters and waveforms. However in this situation, the Viewbed Screen still displays. If you want to view another monitor, you must select again through using hot keys.



If a measure module of the viewed monitor is disconnected or closed, its corresponding waveform will disappear and the waveform in the waveform area will not be refreshed. Instead this waveform area will display empty. At this time, if you want to view other waveforms of this monitor, you need to select again.



4-3



Face Select



FOR YOUR NOTES



4-4



Chapter 5



Alarm







Alarm Modes







Alarm Self-test







Alarm Cause







SILENCE/CLOSE/PAUSE







Parameter Alarm







When an Alarm Occurs



Warning When the monitor is powered on, the system will test whether the audio and visual alarm function is working normally or not. Upon turning on the monitor, a tone “DO-” will be heard. At the same time the alarm indicator will flash once in different color. If the audio and visual alarm function is not normal, do not use the monitor on the patient. Please contact Mindray Customer Service immediately.



5.1 Alarm Modes 5.1.1. Alarm Level The monitor classifies alarms into three categories. 1. Physiological alarm: Physiological alarm is on the upper right part of the screen. It refers to those alarms triggered by the patient’s physiological condition, which could be considered dangerous to his or her life. These are alarms such as heart rate (HR) exceeding alarm limit (parameter alarms). 2. Technical alarm: Technical alarm is to the left side of the physiological alarm. It refers to a system failure that can make certain monitoring processes technically impossible, or make the monitoring result unreliable. A technical alarm is also called a System Error Message. 3. General alarm: General alarms belong to those situations that can not be categorized into the other two types, but still need to have some attention given to them. Each type of alarm, either technical or physiological, has its own priority level. 1. High Level Alarms: A high level alarm indicates the patient’s life is in danger or the monitor has a serious technical problem. the most serious alarm.



5-1



This is the highest level of alarm and is



Alarm



2. Medium Level Alarms: A medium level alarm means a serious warning. 3. Low Level Alarms: A low level alarm is a general warning. For high priority alarms, when they occur, the monitor will give a prompt in a more alert way. Some alarms levels can be changed via the software in the monitor. Alarm levels of all technical alarms and general alarms as well as some physiological alarms are pre-set in the system and cannot be changed.



5.1.2. Alarm Modes When an alarm occurs, the monitor will notify the user in at least three ways. 1. Audio Sound: A beep or tone is generated. Usually this consists of a series of short beeps, or tones followed by a pause, and repeated. 2. Visual prompts: A flashing or constant alarm lamp, or flashing indication of the measured parameter is generated at the same time as the audio sound. 3. Screen Messages: The description is displayed on the screen. Note The type of presentation for each alarm is dependant on the alarm level.



Screen Messages For technical alarm, when an alarm occurs, the description will be displayed in the information area without asterisk signal. For a measured parameter, when its value exceeds the alarm limits, a physiological alarm will be triggered. The description with asterisk signal will appear on the screen. 1.



Triple “***” means high-level alarm.



2.



Dual “**” means medium-level alarm



3.



Single “*” means low-level alarm.



Visual Prompts When the measured value of a monitored parameter exceeds the alarm limit, the monitor indicates the alarm by making the monitored parameter value flash every 1 second. This is in addition to the alarm indications described above. If, at this time, the upper and lower alarm limits for the parameter are displayed, they will flash at the same 1 second intervals. The alarm indicator uses the following visual presentation to indicate high, medium, or low level alarms. High level: Alarm indicator flashes red with high frequency. Medium level: Alarm indicator flashes yellow with low frequency. Low level: Alarm indicator lights on yellow.



5-2



Alarm



Audio Sound The monitor uses the following audible presentations to indicate high, medium or low level alarms. Alarm level



Audio prompt



High



“DO-DO-DO------DO-DO, DO-DO-DO------DO-DO” once every 8 seconds.



Medium



“DO-DO-DO” once every 24 seconds.



Low



“DO-” once every 24 seconds.



Note When alarms of different levels occur at the same time, the monitor gives the sound of the highest level.



5.1.3. Alarm Setup The setup for alarms is done by using the alarm setup menu. 1. Select [ALARM SETUP] from the “SYSTEM SETUP” menu. 2. Select [ALM SEL], you can set the information for “COMMON ALM SETUP” and settings for each parameter.



2



Figure 4-1



ALARM SETUP



COMMON ALM SETUP Select [COMMON ALM SETUP] to display following setup items for all parameters. ALARM VOL: Alarm volume. There are ten options: 1~10. “1”means the minimum volume, and “10” means the maximum volume. ALM REC TIME: It refers to the recording duration and has three options: 8s, 16s, 32s. The system can output the waveform of 4s, 8s or 16s prior to and after the alarm occurs. PARA ALM TYPE: Select LATCH or UNLATCH. LATCH refers to the situation where once an alarm occurs, the monitor will stay in alarm until intervention by the operator (press SILENCE on the panel). UNLATCH refers to the situation where the monitor will stop the alarm once the condition 5-3



Alarm



causing the alarm is returned to a state that no longer violates the alarm conditions. This is done automatically by the monitor. Alarm setup of each parameter 1.



From the “ALARM SETUP” menu, select the [ALM SEL].



2.



You can set the alarm information for the monitored parameters. These are HR, ST, PVC, SPO2, NIBP, RESP and TEMP.



Example: To set the alarm information for HR 1.



Select [HR ALM SETUP] from the pull-down list of [ALM SEL] to displays the “ALARM SETUP” menu for heart rate.



2.



Five items are available for HR alarm settings. HR ALM: ON of OFF ALM LEV: Alarm level, high, medium or low ALM REC: Alarm recording ON or OFF ALM HI: Upper limit for the HR alarm ALM LO: Lower limit for HR alarm



3.



Use the rotary knob to select each item, and then press the knob and a pull-down list appears for you to make the desired selection.



The method for setting the alarm information for other parameters is the same as for HR.



5.2 Alarm Self-test When the monitor is turned on, the system will test the audio and visual alarm function. Each time when the power is turned on, the system will give “DO-” alarm sound, and the alarm indicator on the display will flash once in different color. If the system does not give “DO-“ sound and the alarm does not flash, do not use the monitor on patients, and notify Mindray Customer Service.



5.3 Alarm Cause Alarms occur when: 1.



Physiological measured parameter alarm limit is violated.



2.



An alarm for system error (technical alarm) occurs.



3.



A general alert occurs.



A. Physiological alarms When the measured value of a physiological parameter exceeds the alarm limit and the alarm is set to “ON”, an alarm will be activated. No alarm will be activated if the alarm is set to “OFF”. 5-4



Alarm



B. Technical alarms When a system error occurs, the system initiates an alarm immediately and proceeds to implement a corresponding remedy, such as stop displaying corresponding parameter and waveform and clear the final displayed value to avoid misleading the treatment. If more than one error occurs, they will be displayed in order. C. General alert In some circumstances, general alerts will behave similar to a monitored parameter alarm. But in a normal sense, they are not regarded as patient health related items. For example, if patient probe/sensor is not connected when the monitored is turned on, the system will give technical alarm such as SENSOR OFF.



5.4



SILENCE/CLOSE/PAUSE SILENCE/CLOSE



Press the SILENCE button on the panel for more than 1 second, the system will shut off all sounds. Press the SILENCE button again, the system can exit the SILENCE status and restore the PAUSE status and accordingly suspend the alarm as per the previously defined time duration. Press the SILENCE button for the third time, the system will exit the PAUSE status and restore the normal alarm status by giving the alarm sound again. When the system is in the SILENCE status, any new alarm will terminate the SILENCE status and make the system restore the normal alarm status.



Note When the



symbol appears indicating the alarm sound is shut off and accordingly the



system will not give alarm sound. Therefore, you must be very careful in using this function. There are two methods to terminate this status. One is to set the alarm volume to “ON” in the MAINTAIN menu. The other method is to Press the SILENCE button shortly to make the ; Press the SILENCE button again and the system will restore the normal symbol become alarm status again. PAUSE Press the SILENCE button on the panel shortly, the system will shut off all alarm sound and visual prompt as well as description of physiological alarm, and enter the PAUSE status. The countdown of PAUSE status is displayed in the Physiological Alarm area, in which area the symbol is also displayed. The time duration of the PAUSE status can be set to 1min, 2min or 3min. You can select in the [ALM PAUSE TIME] item in the “SYSTEM MENU\MAINTAIN”. After Pressing the SILENCE button again, the system will restore the normal status. Besides, 5-5



Alarm



the occurrence of any new technical alarm will also terminate the PAUSE status and let the system restore the normal status. The



symbol disappears, too.



Note After the system goes back to the normal status, the existence of alarm depends on whether the alarm condition is complied with. After Pressing the SILENCE button, the system will temporarily shut off the alarm sound for LEAD OFF/SENSOR OFF alarm.



5.5 Parameter Alarm The settings for parameter alarms are located in the parameter individual menus. In the menu for a specific parameter, you can check and set the alarm limit or alarm status. The settings for the parameters are isolated from each other.



When a parameter alarm is set to OFF, a “ ” icon is displayed near the parameter. You can set the alarm to ON/OFF for each parameter. For the parameters whose alarm is set to ON, an alarm will be triggered when at least one of the settings is violated. The following actions take place: 1.



An alarm message is displayed on the screen as described in 5.1.2 Alarm Modes.



2.



If you have set up the alarm volume, the system can give alarm sound as per the pre-set alarm level and volume. When alarms of more than one parameter occur at the same time, the system will give the alarm sound in the highest alarm level.



3.



The alarm indicator flashes.



4.



If you have set up the alarm recording time in the “ALARM SETUP” menu to 8s, 16s or 32s, the system will store the waveform of 4s, 8s or 16s prior to and after the alarm;



5.



If the alarm recording is set to ON, the system will automatically activate the recorder to start alarm recording. For detailed information, refer to Chapter 7 Recording.



5.6 When an Alarm Occurs Note When an alarm occurs, you shall always check the patient's situation first. The alarm message appears at the top right of the screen. appropriately, depending on the cause of the alarm.



5-6



Identify the alarm and act



Alarm



1.



Check the patient's condition.



2.



Identify the alarming parameter or the alarm type.



3.



Identify the cause for the alarm.



4.



If necessary, silence the alarm sound.



5.



If the alarm status terminates, you need to check if the cause for the alarm has been removed.



You can find the alarm information and prompt information for parameters in their corresponding chapters of this manual.



5-7



Alarm



FOR YOUR NOTES



5-8



Chapter 6



Freeze



General Description Entering and Exiting Frozen Mode Frozen Menu Recording Frozen Waveform



6.1



General Description When monitoring a patient, you may freeze the waveforms of interest so you can view them closely. If necessary, you may also use recorder to print out a frozen waveform. The Freeze function of this monitor has following features: Freeze status can be activated on any operating screen; At the same time of entering the Freeze status, the system exits all other operating menus and freezes all waveforms in the waveform area of the basic screen. However, the system can still refresh the parameters normally. You can review and record the frozen waveforms.



6.2



Entering and Exiting Frozen Mode



Entering Frozen Mode 1. In the normal mode, press the “FREEZE” button on the front panel and the menu being displayed (if available) will disappear. 2. The monitor will enter the Frozen mode. At the same time, the “FROZEN” menu will pop up. 3. In the Freeze mode, all waveforms are frozen.



Exiting Frozen Mode When the monitor is in frozen mode, performing any of the following operations will allow the monitor to exit the frozen mode: 1. Select [EXIT] on the “FROZEN” menu. 2. Press the “FREEZE” button on the front panel again. 3. Press the “MAIN” button on the front panel. After exiting the Frozen mode, the system displays real-time waveforms again. 6-1



Freeze



6.3



FROZEN Menu



Press the “FREEZE” button on the front panel, the FROZEN menu will appear on the bottom part of the screen and the system enter the frozen mode at the same time.



Figure 6-1 WAVE 1:



FROZEN menu



used to select the first frozen waveform to record. The pull-down list



of this item gives you the names of all frozen waveforms displayed on the screen. WAVE 2:



used to select the second frozen waveform to record. The pull-down



list of this item gives you the names of all waveforms displayed on the screen. REC: select REC and the system will begin recording the frozen waveforms selected in “WAVE 1” and ”WAVE 2”. EXIT: select EXIT and the system will close the FROZEN menu and exit the frozen mode. Note Pressing the “FREEZE” button repeatedly may result in discontinuous waveforms on the screen.



6.4



Recording Frozen Waveform



In the Frozen mode, you may record the frozen waveforms. A maximum of 2 waveforms can be recorded simultaneously. On the FROZEN menu, the pull-down lists for “WAVE 1” and “WAVE 2” give you all names of the frozen waveforms on the screen. You may select two. Select the “REC” option on the FROZEN menu to record the frozen waveforms. If one of the two selected waveforms is not available, only the measured parameters and the other waveform are recorded. If both selected waveforms are not available, only the measured parameters are recorded. For recording frozen waveforms, you can only record the waveforms displayed at the time the monitor enters the freeze status. The recording time length is the same as the length of the waveform displayed on the screen.



6-2



Freeze



While recording frozen waveforms, the monitor is still in the Frozen mode. After completion of recording you may select another waveform to be recorded. Select the “REC” option again to record new selected waveforms. If the recorder is not installed, selecting the “REC” option will display the message “Recorder does not exist” in the information bar. For detailed information about recording, refer to Chapter 7 Recording.



6-3



Freeze



FOR YOUR NOTES



6-4



Chapter 7



Recording



General Description Recording Type Start Recording Recording Operating and Status Information



7.1



General Description



The monitor uses a thermal dot matrices recorder. The record paper is 48mm wide.



Performance of the Recorder Waveforms are recorded at a speed of 25 or 50 mm/s. Up to two waveforms can be recorded. The recording can have a printed grid, or be on plain paper. English or Chinese language can be printed. The recording time and waveforms are selectable. Time intervals for automatic recording can be selected. The waveform is the same as for real time recording. The alarm recording waveform is automatically selected by the monitor.



7.2



Recording Type



The monitor provides several recording methods: Continuous real-time recording 8 second real-time recording Auto 8 second recording Parameter alarm recording Frozen waveform recording Trend graph/table recording ARR review recording Alarm review recording NIBP review recording Drug calculation and titration table recording Monitor information recording



7-1



Recording



Real-time Recording Real-time recording starts as you press the RECORD button. The system automatically select waveforms (usually the first two waveforms displayed on the screen) for continuous real-time recording and continuous 8 second recording. Or you can specify the waveforms in the menu. Refer to Chapter 3.4.4 Recorder Setup for details. In the “RECORD” menu, you can choose the function to record two waveforms at the same time or to record only one waveform by closing the other one. If you start recording when two waveforms are closed, the recorder only outputs parameters.



Note When the system is executing a recording task, the system can start executing next alarm recording task only when the current one is finished.



Auto recording The monitor starts 8-second real-time recording automatically with the interval set up in the [TIMING REC TIME OFF] item of the “RECORDER ” menu. Refer to Chapter 3.4.4 Recorder Setup for details.



Alarm Recording Parameter Alarm Conditions for alarm recording: The alarm recording switch must be set to ON and there is parameter alarm. Recording length is 4, 8, or 16 seconds prior to and after the alarm (totally 8, 16 or 32 seconds) (which can be selected in the SYSTEM MENU). The monitor can output all parameter values when the alarm is happening. Output two waveforms as per following rules: 1) If more than one alarm switch is set to ON and more than one alarm is triggered simultaneously, the monitor will output the waveform of the parameter with the highest alarm level. If the alarms are of the same level, the monitor will output the waveform of the latest alarm. 2) If other parameters have alarms when the monitor is recording a parameter alarm, the monitor will record other alarms after it finishes the current recording process. 3) If many alarms occur at the same time, the monitor will first store part of them and then record them out later in turn. ST Segment Alarm The monitor can record ECG waveforms of 4, 8, or 16 seconds prior to and after the ST alarm (totally 8, 16, or 32 seconds) and all parameter values when the alarm is happening. Arrhythmia Alarm The monitor can record waveforms of 4 seconds prior to and after the alarm (totally 8 seconds) 7-2



Recording



and all parameter values when the alarm is happening.



Frozen Waveform Recording In the Freeze status, the monitor can output the specified screen waveforms. As the result, you can keep the abnormal waveforms for further study.



Trend Graph / Table Recording The monitor can output the trend graph/table in the current trend review window.



Arrhythmia Review Recording The monitor can output the Arrhythmia alarm event in the current ARR RECALL window.



Alarm Review Recording The monitor can output the parameter alarm event in the current ALARM RECALL window.



NIBP Review Recording The monitor can output NIBP review event in the current NIBP RECALL window.



Monitor Information The monitor can output messages in the current STATUS window.



Titration Table The monitor can output the messages in the current TITRATION window.



Notes on Recording Record types: Continuous real-time recording 8 second real-time recording Auto 8 second recording Parameter alarm recording Frozen waveform recording Trend graph/table recording ARR review recording Alarm review recording NIBP review recording Drug calculation and titration table recording Monitor information recording Alarm parameters, alarm time and freeze time. Patient bed number, name, sex, height, weight, date of birth, admission date. Parameter name and value Recording time Waveform name Waveform scale (only for ECG waveform) ECG lead, scale, filter mode, (if having ECG waveforms, it will be printed out within the first 7-3



Recording



second or when changing the lead, gain and filter mode during real-time recording.) Date and time Company name



7.3



Start Recording



You can use following ways to start the recording process: Continuous real-time recording Press the “RECORD” button to start/stop recording. 8 seconds real-time recording



Press the “RECORD” button to start recording. The system will automatically stop recording in 8 seconds.



Auto recording



Start recording automatically as per the interval selected in the [TIMING REC TIME OFF] item in the “RECORD” menu. The system will automatically stop recording in 8 seconds.



Alarm recording



When the alarm recording switch is set to ON, the system will automatically start recording when alarm is happening



Frozen waveform recording



After accessing the FREEZE menu, you can use knob to select two waveforms for output and then select [REC] to output waveforms.



If two waveforms are both closed, the monitor only outputs parameter values in the FREEZE status. Trend graph recording



Select [REC] in the “TREND GRAPH” menu, the monitor will output the trend graph of the parameters being currently displayed.



Trend table recording



Select [REC] in the “TREND TABLE” menu, the monitor will output the trend table of the parameters being currently displayed.



Arrhythmia review recording



Select [REC] in the “ARR WAVE RECALL” menu, the monitor will output the ARR waveform being currently displayed and its relevant parameter value.



Alarm review recording



Select [REC] in the “ALARM RECALL CONDITION” sub-menu of the “SYSTEM MENU”, the monitor will output the waveforms of the alarming parameter and relevant parameter value.



NIBP review recording



Select [REC] in the “NIBP RECALL” sub-menu of the “SYSTEM MENU”, the monitor will output the NIBP value in the current window. 7-4



Recording



Monitor information recording



Select [REC] in the “STATUS” sub-menu, the monitor will output status information of the monitor.



Titration table recording



Select [REC] in the “TITRATION” sub-menu of the “DRUG CALCULATION” menu, the monitor will output the information in the current titration table.



Note You can Press the RECORD button on the panel to stop the current recording process. You can stop recording process and clear the stored alarm recording tasks by selecting the [CLEAR REC TASK] item in the “RECORD” sub-menu of the “SYSTEM MENU”.



7.4



Recorder Operating and Status Information



Record Paper Requirement You must use the thermal-sensitive record paper in accordance with requirement; otherwise the recorder may not function normally, the recording quality may be poor, and the thermo-sensitive printhead may be damaged. Proper operation ■



When the recorder is working, the record paper goes out steadily. Do not pull the paper outward with force; otherwise the recorder may be damaged.







Do not use the recorder before loading the paper.



Paper Out When the monitor displays the "RECORDER OUT OF PAPER" message in the Information Area, the recorder cannot be started. At this time, you should load the record paper in accordance with the requirement.



Procedures to load record paper ■



Open the recorder catch;







Lift up the bail on the left axis;







Load a new roll of record paper into the paper cassette with the printing side facing the printhead;







When you can see the paper from the other side, pull it outside. You must load the paper at the proper position with tidy margin;







Pull the bail back to the original position;







Feed the paper out from the recorder outlet;







Close the recorder catch.



7-5



Recording



Note Be careful when loading record paper. Avoid damaging the printhead. Unless when loading paper or shooting troubles, do not leave the recorder catch open. Removing Paper Jam When the recorder functions abnormally or sounds strange, open the recorder catch to check for paper jam. You shall obey the following steps to remove the paper jam. ■



Cut off the record paper from the feeding edge;







Lift up the bail on the left axis of the recorder;







Pull the paper out from below;







Re-load the paper.



Recorder Status Message (Technical Alarms) Message



Cause



Alarm Level



Remedy



RECORDER HEAD HOT



The thermal terminal is too hot.



low



Stop using the recorder



REC HEAD IN WRONG POS.



The printhead is not in recording place.



low



Press down the bail on the left axis of the recorder.



RECORDER OUT OF PAPER



Record paper runs out.



low



Load a new roll of record paper.



low



Re-load paper.



low



Reset the recorder.



RECORDER PAPER JAM RECORDER COMM ERR



Recording continuously for more than 30m Operating status error



RECORDER INIT ERR



Error occurs during initialization



low



Shut down and re-start the monitor



TOO MANY REC TASKS



Too many alarm events take place simultaneously.



low



Close alarm recording switch



RECORDER INITIALIZING



The recorder is in initialization process.



low



Wait for the initialization process to end



REC NOT AVAILABLE



Working status of the recorder is incorrect.



low



Shut down and re-start the monitor



RECORDER COMM ERR



Serial port communication error.



low



Shut down and re-start the monitor.



low



Wait for the recording process to end



RECORDER BUSY



In recording status



If after shutdown and re-start, error still exists, contact MINDRAY Customer Service Department.



7-6



Chapter 8



Trend and Event



The monitor can store 72-hour trend data of all parameters, 400 NIBP measurement results and 60 alarm events. This chapter tells how to view the data stored by the system.



8.1



Trend Graph







For the latest 1-hour trend graph, display the value of the data every 1 or 5 seconds;







For the latest 72-hour trend, display the value of the data every 1, 5 or 10 minutes;



Select the [TREND GRAPH] item in the “SYSTEM MENU” to call up the following menu:



Figure 8-1



TREND GRAPH window



In the trend graph, the y-axis refers to the measured value and x-axis time. “ ” is the cursor of the trend graph, the parameter value of the position pointed by the cursor is displayed below the trend graph and the corresponding time is displayed above the trend graph. Except for NIBP trend, the system displays other trends in continuous curves. In the NIBP trend graph, "▼" indicates systolic value, "▲" the diastolic value, and "*" the mean value.



8-1



Trend and Event



Select to display trend graph of a specific parameter: Move cursor to highlight the [PARA] item and turn the knob to view its content and press the knob when the desired parameter appears. Then the system will display the trend graph of the selected parameter.



Select 1-hour or 72-hour trend graph: Select the [RES] item, you can select 1s or 5s to view 1-hour trend or 1min, 5min or 10min to view 72-hour trend.



View other trend curves: When "



" appears on the right part of the screen, Select the [L-RIGHT] item and turn the



knob clockwise to view later trend curve. When "



" appears on the left part of the screen,



Select the same item and turn the knob counterclockwise to view earlier trend curve.



Change the display scale You can change the proportion of y-axis by Selecting [ZOOM] item. The proportion of the trend curve will change accordingly. The value beyond maximum value will be represented by the maximum value.



Obtain trend data of a specific time on the current trend graph The time at which the cursor points will change as the knob is turned. Parameter of the time is " appears on the right part of the screen, the trend



displayed below the x-axis. When "



graph will automatically page down for you to view later trend curve as the cursor moves here. When "



" appears on the left part of the screen, the trend graph will automatically page up



for you to view earlier trend curve as the cursor moves here.



Output trend curve Press [REC] button to print out the trend curve of the current selected parameter.



Mark event If an event is marked A, B, C, or D, the corresponding event type will be displayed on the axis time of the trend graph, such as



,



,



or



.



Operation example View the NIBP trend graph of the latest 1 hour: ■ Select the MENU button in the lower right corner of the screen; ■ Select the [TREND GRAPH] item in the SYSTEM MENU; ■ Select parameter: Select the [PARA] item and turn the knob until NIBP appears; ■ Select 1S or 5S in the [RES.] item; ■ Select the [L-RIGHT] item and turn the knob, observe the changes of the trend graph time and the trend curve; ■ Stop at the trend time section to be observed. If the y-axis proportion is improper such as 8-2



Trend and Event



some trend values exceed the maximum value in the current y-axis, you can Select the [ZOOM] item to adjust; ■ To know the measured value of a certain time, you can Select the [CURSOR] item and move the cursor to the time, accordingly time will appear above and the measured value below the curve.; ■ To output the trend graph, you can Select [REC] and then the recorder will output the NIBP trend in the current review window; ■



8.2



Select [EXIT] to exit trend graph window.



Trend Table



■ The system can display the latest 72-hour trend table data with the resolution of 1min, 5min, 10min, 30min, or 60 minutes. Select the [TREND TABLE] item in the SYSTEM MENU to call up the following menu:



Figure 8-2



TREND TABLE window



Time corresponding to each group of trend data is display in the leftmost column with date in brackets. Events that are once marked are listed below the event. They correspond to the time of marked event. Trend data of each parameter is divided into 6 groups. HR, PVCs ST RR 8-3



Trend and Event



TEMP SPO2, PR NIBP (S/M/D) NIBP trend data is displayed in a special way. Except the measured NIBP value, the system also displays measurement time under the corresponding value. If there are many NIBP values in a time section, the system can only display one group and also a “*” symbol on the “MORE” to indicate that there are two or more than two measured results.



Choose trend table of different resolutions Use the cursor to highlight the [RES] item and turn the knob to change its content so as to change the time interval of the trend data.



View later or earlier trend curve When " " appears on the upper part of the screen, select the [UP-DOWN] item and turn the knob clockwise to view later trend data. When " " appears on the lower part of the screen, select the same item and turn the knob counterclockwise to view earlier trend data.



Obtain trend data of different parameter Select the [L-RIGHT] to select one of the 6 parameter groups. When there is ">" to the right of the rightmost parameter, it indicates the following page is available. When there is "5).



User-sele ctable



The RR interval is less than 600ms. VT>2



Without pacemaker



3 < the number of cluster PVCs < 5



12-19



VT>2



User-sele ctable



ECG/RESP Monitoring



COUPLET



Without pacemaker



BIGEMINY



Without pacemaker



TRIGEMINY



Without pacemaker



R ON T



Without pacemaker



PVC



Without pacemaker



TACHY



All patients



BRADY



BEAT MISS



PNP



PNC



2 consecutive PVCs VENT BIGEMINY Vent Trigeminy



COUPLET



User-sele ctable



BRGEMINY



User-sele ctable



TRIGEMINY



User-sele ctable



A type of single PVC under the condition that HR