4 0 33 MB
Ziehm Vista Ziehm Vista Plus Ziehm 8000 Ziehm Compact Service Manual
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Copyright Copyright © 2010 Ziehm Imaging GmbH All rights reserved. Transmission or reproduction of this document, exploitation and disclosure of its contents to third persons is not permitted without express written consent of the manufacturer. Infringements shall entitle to damage claims. Registered Trademarks This document may contain the names of registered trademarks or brands, the use of which by third persons for their purposes may infringe the rights of their respective owners.
Quality Standards This document was produced in accordance with a certified QM system as per DIN EN ISO 13485. It conforms to the requirements of Council Directive 93/42 EEC, Annex I, and other applicable norms. The information provided in this document may be updated at regular intervals. Subject to change without prior notice. Manufactured by: Ziehm Imaging GmbH Donaustr. 31 90451 Nuremberg (Germany) e-mail: [email protected] http://www.ziehm.com Rev. 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
With the Ziehm Vista, Ziehm Vista Plus, Ziehm 8000 and the Ziehm Compact, Ziehm Imaging GmbH provides an active medical device connected to the mains power supply. The above listed medical devices are non-contact devices. They do not require any contact with the patient to perform its intended use. They provide contactless energy in the form of Xrays.
Preface
Ziehm Imaging GmbH authorizes only trained and skilled personnel to operate this medical device. The system is intended for use by health care professionals such as physicians, orthopedic surgeons, surgeons, radiologists and technologists in hospitals, outpatient clinics and other clinical environments. Ziehm Imaging GmbH anticipates the system will be used on a nearly daily basis. Ziehm Imaging GmbH applications specialists and/or qualified site personnel provide on-site operator training in the proper use of the system. The medical equipment is intended for fluoroscopies in the field of surgery, e.g. in traumatology, orthopedics, neurosurgery, urology, cardiology. Third-party devices and components used in combination with this system must comply with the safety requirements according to IEC 60601-1 and/or IEC 60601-1-1 or furnish proof of an equivalent degree of safety.
Normal use (Ziehm 8000)
To ensure CE conformity, these components must have a CE approval in accordance with Council Directive 93/42/EEC. In addition, a declaration in compliance with Article 12 of the said directive must be provided. For components without CE approval, a conformity assessment procedure is obligatory. Proper and safe operation of the system requires adequate transportation, storage, assembly and installation as well as appropriate use and maintenance. The limit values indicated in the present document must not be exceeded; this applies also when putting the system into service. The system is not suitable for interventional procedures acc. to IEC 60601-2-43. The medical equipment is intended for fluoroscopies in the field of surgery, e.g. in traumatology, orthopedics, neurology, urology, cardiology. Third-party devices and components used in combination with this system must comply with the safety requirements according to IEC 60601-1 and/or IEC 60601-1-1 or furnish proof of an equivalent degree of safety.
Intended use (Ziehm Vista product family, Ziehm Compact)
To ensure CE conformity, these components must have a CE approval in accordance with Council Directive 93/42/EEC. In addition, a declaration in compliance with Article 12 of the said directive must be provided. For components without CE approval, a conformity assessment procedure is obligatory. Proper and safe operation of the system requires adequate transportation, storage, assembly and installation as well as appropriate use and maintenance. The limiting values indicated in the present document must not be exceeded; this applies also when putting the system into service.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3
Operation
The system may only be operated by personnel who has undergone radiological training.
Operation (U.S.A.)
In the U.S.A., Federal law restricts use of this device to trained personnel on the order of a physician.
Ziehm Vista product family
The Ziehm Vista product family comprises the following models: Ziehm Vista (not available in the USA) and Ziehm Vista Plus (available only in the USA). Unless stated otherwise, all information given here for the Ziehm Vista also refers to the Ziehm Vista Plus.
Ziehm 8000 product family
The Ziehm 8000 product family comprises the following models: Ziehm 8000, Ziehm Compact, Ziehm Compact Litho, Ziehm Exposcop 8000.
Ziehm Vista Workstation option
The PC Workstation is intended for viewing and postprocessing DICOM images with the “e-Film” program. In addition, any software can be installed and used on it under the sole responsibility of the user. The PC Workstation is not a functional part of the medical equipment and does not influence its functions in any way.
Authorized personnel
Only authorized personnel are allowed to assemble and repair the medical equipment described in this document. Authorized personnel are persons who have attended an appropriate training course provided by the manufacturer.
Contraindications to the use of X-rays
The exposure of humans to ionizing radiation must always be medically justified. Especially when used on pregnant women, adolescents, children, and pediatric patients, all procedures using ionizing radiation should be used with caution or be avoided altogether. However, the final decision lies with the attending physician or attending surgeon.
Exclusion of liability
The manufacturer accepts responsibility for the safety, reliability and performance of the system only if −
any installation, modification or repair work is carried out exclusively by persons authorized by the manufacturer;
−
the electrical installation of the site where the system is operated complies with the requirements of VDE 0107 or the corresponding national regulations of the country of installation;
−
only original spare parts or components that comply with Ziehm Imaging’s specifications are used;
−
the system is used in accordance with the Operating Instructions.
−
Ziehm Vista Workstation option: the PC hardware is an integrated part of the system and is not altered. The responsibility for the installed software lies with the corresponding software suppliers.
The warranty becomes invalid in case that any repair, modification or installation work is carried out by unauthorized personnel, or any seals on components are broken. No consequential damages will be accepted either. The equipment conforms to Class IIb according to Council Directive 93/42/EEC. This document has been written and reviewed originally in English.
4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
0
Classification Mobile Stand/ Monitor Cart
22
No
Chapter
Mobile Stand
Monitor Cart
1
Warnings and Important Information
⊕
⊕
2
General
⊕
⊕
3
Mobile Stand Connections
⊕
4
Power Supply Mobile Stand
⊕
5
Module 2
⊕
6
Module 3
⊕
7
Image Intensifier
⊕
8
X-ray Generator
⊕
9
Vision Track
⊕
10
Keyboards Mobile Stand and Monitor Cart
⊕
11
Monitor Cart General & Wiring
⊕
12
Video Documentation
⊕
13
Monitors
⊕
14
Power Supply Monitor Cart
⊕
15
Image Memory System
⊕
16
DICOM Configuration
⊕
17
DICOM Conformance Statement
⊕
18
Software Tools
⊕
⊕
19
Adjustment Instructions
⊕
⊕
20
Error Messages
⊕
⊕
21
Active Cooling system
⊕
22
Maintenance Schedule
⊕
⊕
23
Index
24
DIN A3 Drawings
⊕
⊕
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
⊕
0-1
0 Classification Mobile Stand/Monitor Cart
0-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Table of Contents 0
Classification Mobile Stand/Monitor Cart Table of Contents
0-1 i
1
Warnings and Important Information 1.1 Safety Instructions............................................................................................ 1.2 Warnings .......................................................................................................... 1.3 Required tests after service..............................................................................
1-1 1-1 1-2 1-2
2
General 2.1 Mobile Stand System Overview ....................................................................... 2.2 Detailed Overview with U501 ........................................................................... 2.3 Detailed Overview with Module 3 ..................................................................... 2.4 Overview Ziehm 8000 Mobile Stand ................................................................ 2.5 Block Diagram with U501 ................................................................................. 2.6 Block Diagram with Module 3........................................................................... 2.7 Power Distribution Overview with U501 ........................................................... 2.8 Power Distribution Overview with Module 3 ..................................................... 2.9 Board Overview with U501............................................................................... 2.10 Board Overview with Module 3......................................................................... 2.11 Board Numbers ................................................................................................ 2.12 Numbers of CAN Modules with U501............................................................... 2.13 Numbers of CAN Modules with Module 3 ........................................................ 2.14 CAN Bus Distribution in the C-Arm Stand with U501 ....................................... 2.15 CAN Bus Distribution in the C-Arm Stand with Module 3................................. 2.16 Video Distribution in the C-Arm Stand.............................................................. 2.17 Video Distribution in the C-Arm Stand.............................................................. 2.18 Regulation Loops ............................................................................................. 2.19 Technical Data ................................................................................................. 2.19.1 Ziehm Vista and Ziehm Vista Workstation ........................................ 2.19.2 Ziehm Vista Plus (USA) .................................................................... 2.19.3 Ziehm 8000, Ziehm Compact and Ziehm Compact with adapter for lithotripter procedures ..................................................... 2.19.4 General technical data ...................................................................... 2.19.4.1 Systems with a voltage rating of 100 V, 120 V, 200 V....... 2.19.4.2 Systems with a voltage rating of 220 V, 230 V, 240 V....... 2.19.5 Power plug specification ................................................................... 2.19.5.1 Systems with a voltage rating of 120 V.............................. 2.19.5.2 Systems with a voltage rating of 230 V.............................. 2.19.6 Laser positioning device ................................................................... 2.19.7 Dose meter ....................................................................................... 2.19.7.1 VacuDAP ........................................................................... 2.19.7.2 VacuDAP 2002 .................................................................. 2.19.7.3 VacuDAP duo .................................................................... 2.19.8 Air kerma........................................................................................... 2.19.9 Focal spot position ............................................................................ 2.19.10 Dimensions .......................................................................................
2-1 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 2-11 2-13 2-13 2-14 2-15 2-16 2-17 2-18 2-19 2-19 2-22
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-24 2-24 2-26 2-29 2-31 2-31 2-31 2-32 2-33 2-33 2-34 2-35 2-36 2-37 2-38
i
3
Mobile Stand Connections 3.1 XP0 Cable ........................................................................................................ 3.1.1 Connectors ........................................................................................ 3.1.2 Wiring ................................................................................................ 3.2 C-Arm Wiring Diagram ..................................................................................... 3.3 CAN Connector ................................................................................................ 3.4 Grounds Overview............................................................................................ 3.5 Hand Switch & CAN Bus Interface Wiring ........................................................ 3.6 Foot Switch Wiring............................................................................................ 3.7 Four-Pedal Foot Switch .................................................................................... 3.8 Module 1........................................................................................................... 3.8.1 Main connector.................................................................................. 3.8.2 U333 connectors ............................................................................... 3.9 Module 2........................................................................................................... 3.9.1 Connector X60 .................................................................................. 3.9.2 Connector X81 .................................................................................. 3.9.3 Connector X62 .................................................................................. 3.9.4 Connector X83 .................................................................................. 3.9.5 Connector X84 .................................................................................. 3.9.6 Connectors X70, X71, X72, X73 ....................................................... 3.9.7 Connector X74 .................................................................................. 3.10 Module 3 / U501 ............................................................................................... 3.10.1 U180 connector X17 / U501 connector X17...................................... 3.10.2 U180 connector X9 / U501 connector X9.......................................... 3.10.3 U180 connector X5 ........................................................................... 3.10.4 U180 connector X12 / U501 connector X12...................................... 3.10.5 U321 connectors X7, X3 / U501 connectors X7, X3 ......................... 3.11 Module 4........................................................................................................... 3.11.1 Link connector C-arm – i.i. housing................................................... 3.11.2 U316 connector X9 ........................................................................... 3.11.3 U316 connector X11 ......................................................................... 3.12 Module 5........................................................................................................... 3.12.1 Signals and voltages on U185 generator filter ..................................
3-1 3-1 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-9 3-10 3-16 3-16 3-16 3-17 3-17 3-17 3-18 3-18 3-19 3-19 3-19 3-19 3-20 3-20 3-21 3-21 3-22 3-22 3-23 3-23
4
Power Supply 4.1 Trafoblech View................................................................................................ 4.2 Module 1 Assembly .......................................................................................... 4.3 Block Diagram .................................................................................................. 4.4 U333 Board versions ........................................................................................ 4.5 Wiring Diagram with U333f............................................................................... 4.6 Wiring Diagram with U333g.............................................................................. 4.7 Functional Diagram with U333f ........................................................................ 4.8 Functional Diagram with U333g ....................................................................... 4.9 Circuit Description ............................................................................................ 4.9.1 Power-up circuit................................................................................. 4.9.2 Output switching circuit ..................................................................... 4.9.3 Lifting motor control circuit ................................................................ 4.10 U333 ................................................................................................................. 4.10.1 U333f................................................................................................. 4.10.1.1 U333f board view ............................................................... 4.10.1.2 U333f board layout............................................................. 4.10.1.3 U333f schematic power ON/OFF....................................... 4.10.1.4 U333f schematic lift motor control......................................
4-1 4-1 4-3 4-4 4-5 4-6 4-7 4-8 4-9 4-10 4-10 4-11 4-11 4-13 4-13 4-13 4-14 4-15 4-16
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4.10.1.5 U333f schematic power switching ..................................... 4.10.1.6 U333f schematic BG1 & BG2 ............................................ 4.10.2 U333g ............................................................................................... 4.10.2.1 U333g board view.............................................................. 4.10.2.2 U333g board layout ........................................................... 4.10.2.3 U333g schematic power ON/OFF...................................... 4.10.2.4 U333g schematic lift motor control .................................... 4.10.2.5 U333g schematic power switching .................................... 4.10.2.6 U333g schematic BG1 & BG2 ........................................... 4.10.3 U333 jumper settings ........................................................................ U80 (Horn)........................................................................................................ 4.11.1 U80 schematic .................................................................................. U468 (BG1) ...................................................................................................... 4.12.1 U468 layout ....................................................................................... 4.12.2 U468 schematic ................................................................................ U437 (VacuDAP Connection)........................................................................... 4.13.1 U437 schematic ................................................................................ 4.13.2 U437 function .................................................................................... U424 (BG2, Pump Switching) .......................................................................... 4.14.1 U424 schematic ................................................................................ 4.14.2 U424 function .................................................................................... Module 1 Fuses................................................................................................ 4.15.1 Transformer fuses ............................................................................. 4.15.2 Fuses on U333.................................................................................. 4.15.3 Fuses on chassis ..............................................................................
4-17 4-18 4-19 4-19 4-20 4-21 4-22 4-23 4-24 4-25 4-26 4-26 4-27 4-27 4-28 4-29 4-29 4-29 4-30 4-30 4-30 4-31 4-31 4-31 4-31
Module 2 5.1 Module 2 19" Rack Description ....................................................................... 5.1.1 General ............................................................................................. 5.1.2 Module view ...................................................................................... 5.2 Module 2 Overview........................................................................................... 5.3 Module 2 Inputs and Outputs ........................................................................... 5.4 Block Diagrams ................................................................................................ 5.4.1 U325 block diagram .......................................................................... 5.4.2 U325 functional description............................................................... 5.4.3 U326 block diagram .......................................................................... 5.4.4 U326 functional description............................................................... 5.4.5 U328 block diagram .......................................................................... 5.4.6 U328 functional description............................................................... 5.4.7 U334 block diagram .......................................................................... 5.4.8 U334 functional description............................................................... 5.5 U325 and U326 Front Panels........................................................................... 5.6 Exchange of boards ......................................................................................... 5.6.1 Exchange of U325 ............................................................................ 5.6.2 Exchange of U326 ............................................................................ 5.7 Hardware Radiation Logic ................................................................................ 5.8 U326 Schematic Diagrams............................................................................... 5.8.1 U326 schematics 1 ........................................................................... 5.8.2 U326 schematics 2 ........................................................................... 5.8.3 U326 schematics 3 ........................................................................... 5.8.4 U326 schematics 4 ........................................................................... 5.8.5 U325 schematics 1 ........................................................................... 5.8.6 U325 schematics 2 ...........................................................................
5-1 5-1 5-1 5-1 5-2 5-3 5-4 5-4 5-5 5-6 5-7 5-8 5-9 5-10 5-11 5-12 5-13 5-13 5-14 5-16 5-17 5-17 5-18 5-19 5-20 5-21 5-22
4.11 4.12 4.13 4.14 4.15
5
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
iii
5.8.7 U325 schematics 3............................................................................ 5.8.8 U325 schematics 4............................................................................ 5.8.9 U325 schematics 5............................................................................ 5.8.10 U325 schematics 6............................................................................ 5.8.11 U325 schematics 7............................................................................ U501 Video Processing Board ......................................................................... 5.9.1 U501 Block Diagram ......................................................................... 5.9.2 Connectors on the front cover ........................................................... 5.9.3 LEDs on the front cover ....................................................................
5-23 5-24 5-25 5-26 5-27 5-28 5-28 5-29 5-29
6
Module 3 6.1 Module 3 Memory Unit .................................................................................... 6.1.1 General ............................................................................................. 6.1.2 Module 3 view ................................................................................... 6.2 Module 3 Overview........................................................................................... 6.3 Block Diagrams ................................................................................................ 6.3.1 U180 block diagram .......................................................................... 6.3.2 U180 functional description ............................................................... 6.3.3 U321 block diagram .......................................................................... 6.3.4 U321 functional description ............................................................... 6.3.5 Module 3 wiring diagram ................................................................... 6.4 Exchange of Module 3......................................................................................
6-1 6-1 6-1 6-1 6-2 6-3 6-3 6-4 6-5 6-6 6-7 6-8
7
Image Intensifier 7.1 Parts Location................................................................................................... 7.2 Block Diagram .................................................................................................. 7.2.1 Functional description ....................................................................... 7.3 I.I. power supply overview ................................................................................ 7.3.1 General ............................................................................................. 7.3.2 Main parts and their functions ........................................................... 7.4 Hardware description........................................................................................ 7.4.1 PCBs ................................................................................................. 7.4.2 U316 LEDs ........................................................................................ 7.4.3 U316 connectors ............................................................................... 7.4.4 Connectors cascade ......................................................................... 7.4.4.1 Connectors cascade location............................................. 7.4.5 U316 schematics............................................................................... 7.4.5.1 U316 schematics part 1 ..................................................... 7.4.5.2 U316 schematics part 2 ..................................................... 7.5 Adjustments...................................................................................................... 7.5.1 Potentiometer location ...................................................................... 7.6 CCD Camera .................................................................................................... 7.6.1 General description ........................................................................... 7.6.2 Setting the CCD camera distance ..................................................... 7.6.3 Scattered light lens............................................................................ 7.6.4 Changing the camera ........................................................................ 7.6.4.1 Removing the camera........................................................ 7.6.5 Installing the CCD camera ................................................................ 7.6.6 Adjustment of the optical focus ......................................................... 7.6.7 Adjustment instruction for 23 cm and 30 cm tube ............................. 7.6.8 Size and resolution requirements...................................................... 7.6.8.1 Without patient equivalent filter.......................................... 7.6.8.2 With patient equivalent filter 25 mm Al...............................
7-1 7-1 7-2 7-3 7-4 7-4 7-4 7-5 7-5 7-5 7-6 7-6 7-6 7-7 7-7 7-8 7-9 7-9 7-9 7-10 7-10 7-11 7-12 7-12 7-13 7-14 7-15 7-16 7-16 7-16
5.9
iv
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8
Generator 8.1 Overview .......................................................................................................... 8.2 Block Diagrams ................................................................................................ 8.2.1 Generator block diagram................................................................... 8.2.2 Generator detailed block diagram ..................................................... 8.3 Interface Generator – C-Arm Stand ................................................................. 8.4 Block Diagram Generator – Module 1 / Module 2 ............................................ 8.5 General PCBs .................................................................................................. 8.6 Checklist for the Generator .............................................................................. 8.6.1 U185 test point location .................................................................... 8.6.2 U185 schematics .............................................................................. 8.6.3 U185 board layout............................................................................. 8.7 Measuring kV and mA ...................................................................................... 8.8 Laser (option) ................................................................................................... 8.8.1 Generator with laser.......................................................................... 8.8.2 Image intensifier with laser................................................................ 8.8.3 Laser block diagram.......................................................................... 8.8.4 U160 laser control board...................................................................
8-1 8-1 8-2 8-2 8-3 8-4 8-5 8-6 8-7 8-8 8-9 8-10 8-11 8-13 8-13 8-14 8-15 8-16
9
Virtual Collimator 9.1 Virtual Collimator View ..................................................................................... 9.2 Block Diagram .................................................................................................. 9.3 U517d............................................................................................................... 9.3.1 U517d location of connectors, switches and LEDs ........................... 9.3.2 Switches............................................................................................ 9.3.2.1 DIP switch factory setting .................................................. 9.3.3 LEDs ................................................................................................. 9.3.4 Connectors........................................................................................ 9.4 Changing the Collimator................................................................................... 9.4.1 Part exchange ................................................................................... 9.4.2 Mechanical centering ........................................................................ 9.4.3 Collimator size adjustment ................................................................ 9.4.4 Preview size adjustment ...................................................................
9-1 9-1 9-2 9-3 9-3 9-3 9-5 9-5 9-6 9-10 9-10 9-10 9-11 9-12
10
Keyboards on Mobile Stand and Monitor Cart 10.1 Keyboard of Mobile Stand ................................................................................ 10.1.1 MST keyboard view........................................................................... 10.1.2 U400 connections ............................................................................. 10.1.3 U400 block diagram .......................................................................... 10.2 Keyboard of Monitor Cart ................................................................................. 10.2.1 MC keyboard view............................................................................. 10.2.2 U370 connections .............................................................................
10-1 10-1 10-1 10-2 10-3 10-4 10-4 10-5
11
Monitor Cart General and Wiring 11.1 General............................................................................................................. 11.1.1 Monitor cart with 18.1" flatscreen monitors ....................................... 11.1.2 Monitor cart with 24’’ split screen monitor ......................................... 11.1.3 Monitor cart with 23’’ split screen monitor ......................................... 11.1.4 Monitor cart with CRT monitors......................................................... 11.2 Monitor Cart Wiring .......................................................................................... 11.2.1 Wiring of flatscreen monitors............................................................. 11.2.2 Wiring of flatscreen monitors and PC................................................ 11.2.3 Wiring of CRT monitors.....................................................................
11-1 11-1 11-1 11-2 11-3 11-4 11-5 11-5 11-6 11-7
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
v
11.2.4 Wiring of external X-ray indicator and remote (options) .................... 11-8 11.2.5 U385 wiring ....................................................................................... 11-9 11.3 Yellow RJ45 network cables............................................................................. 11-10 12
Video Documentation 12.1 UP 960.............................................................................................................. 12.1.1 Front view.......................................................................................... 12.1.2 Rear view .......................................................................................... 12.2 UP 980.............................................................................................................. 12.2.1 Front view.......................................................................................... 12.2.2 Rear view .......................................................................................... 12.2.3 UP 980 configuration......................................................................... 12.3 VCR .................................................................................................................. 12.3.1 VCR wiring standard ......................................................................... 12.3.2 U411 VCR interface ..........................................................................
12-1 12-1 12-1 12-2 12-3 12-3 12-4 12-5 12-6 12-6 12-7
13
Monitors 13.1 18.1" Flatscreen Monitors................................................................................. 13.1.1 Description ........................................................................................ 13.1.2 Technical data ................................................................................... 13.1.3 Switches ............................................................................................ 13.1.4 Connectors ........................................................................................ 13.1.5 On-screen display ............................................................................. 13.1.5.1 Locking and unlocking the on-screen display .................... 13.1.5.2 Restoring the factory settings ............................................
13-1 13-1 13-1 13-2 13-2 13-3 13-4 13-4 13-5
14
Power Supply 14.1 View of Power Supply....................................................................................... 14.2 General Description.......................................................................................... 14.3 Main Components of Power Module 10 ........................................................... 14.4 Module 10 Block Diagram ................................................................................ 14.5 Module 10 Power Distribution Wiring Diagram................................................. 14.5.1 Wiring with transformer 11090 (105 V – 240 V) ................................ 14.5.2 Wiring with transformer 11091 (200 V – 240 V) ................................ 14.6 Connectors ....................................................................................................... 14.6.1 Mains input ........................................................................................ 14.6.2 Insulated XP0 PCB outlet.................................................................. 14.6.3 Internal mains distribution ................................................................. 14.6.4 On/Off connections ........................................................................... 14.7 Fuses................................................................................................................ 14.8 U357 schematics .............................................................................................. 14.8.1 Power-on logic .................................................................................. 14.8.2 U357 schematic power distribution ................................................... 14.9 Module 10 Transformers .................................................................................. 14.9.1 Transformer 11090 105 V – 240 V .................................................... 14.9.2 Transformer 11091 200 V – 240 V .................................................... 14.9.3 Mains transformer thermo fuse .........................................................
14-1 14-1 14-2 14-2 14-4 14-5 14-5 14-6 14-7 14-7 14-7 14-7 14-8 14-9 14-11 14-11 14-12 14-13 14-13 14-14 14-15
15
Image Memory System V2 1/2k 15.1 Features ........................................................................................................... 15.2 Barebone Front view Step 1 ............................................................................. 15.3 Barebone Front view Step 2 ............................................................................. 15.4 Complete Front View Step 1.............................................................................
15-1 15-1 15-3 15-4 15-5
vi
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15.5 15.6 15.7 15.8 15.9 15.10 15.11
Complete Front View Step 2 ............................................................................ Side View ......................................................................................................... Barebone Inside View Step 1 ........................................................................... Barebone Inside View Step 2 ........................................................................... Complete Inside View Step 1 ........................................................................... Complete Inside View Step 2 ........................................................................... Image Memory Service Settings ...................................................................... 15.11.1 Overview ........................................................................................... 15.11.2 Service settings................................................................................. 15.11.3 Step windowing ................................................................................. 15.11.4 Filter factors ...................................................................................... 15.11.4.1Filters for anatomical programs ......................................... 15.11.4.2Filters for subtraction modes ............................................. 15.11.4.3Filters for subtraction modes with CO2 .............................. 15.11.5 Windowing settings for subtraction modes........................................ 15.11.6 DICOM settings................................................................................. 15.11.7 System settings................................................................................. 15.11.8 Monitor calibration............................................................................. 15.11.9 HEDIS data ....................................................................................... 15.11.10 Software update ................................................................................ 15.12 Displaying Additional Information .....................................................................
15-6 15-7 15-8 15-9 15-10 15-11 15-12 15-12 15-12 15-13 15-15 15-16 15-19 15-20 15-21 15-22 15-22 15-25 15-27 15-29 15-30
16
DICOM Configuration Manual 16.1 DICOM Overview ............................................................................................. 16.2 DICOM and Network Parameters..................................................................... 16.3 DICOM Configuration ....................................................................................... 16.3.1 Configuration of Storage and Storage Commitment ......................... 16.3.2 Configuration of Media Storage ........................................................ 16.3.3 Configuration of DICOM Query/Retrieve........................................... 16.3.4 Configuration of Worklist and MPPS................................................. 16.3.5 Configuration of the print servers ...................................................... 16.3.6 Test of the configuration with DICOM Verification (Echo).................
16-1 16-1 16-2 16-4 16-4 16-6 16-8 16-12 16-17 16-21
17
DICOM Conformance Statement
18
Software Tools 18.1 Exposcop Studio .............................................................................................. 18.1.1 Connection between notebook and C-arm........................................ 18.1.1.1 Location of the service connectors .................................... 18.1.2 Starting Exposcop Studio.................................................................. 18.1.3 Exposcop Studio user interface ........................................................ 18.1.3.1 Main window ...................................................................... 18.1.3.2 Virtual keyboard................................................................. 18.1.3.3 Status line .......................................................................... 18.1.3.4 Workspace......................................................................... 18.1.4 Menu bar of Exposcop Studio software ............................................ 18.1.4.1 Exposcop ........................................................................... 18.1.4.2 “Workspace” ...................................................................... 18.1.4.3 “Settings” ........................................................................... 18.1.5 “General” tab ..................................................................................... 18.1.6 “Adjustment” tab................................................................................ 18.1.7 “Generator” tab ................................................................................. 18.1.8 “Update” tab ......................................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
17-1 18-1 18-1 18-1 18-1 18-1 18-3 18-3 18-4 18-4 18-5 18-6 18-6 18-8 18-9 18-10 18-11 18-13 18-14
vii
18.1.9
19
viii
“Options” tab ..................................................................................... 18.1.9.1 Option bits.......................................................................... 18.1.9.2 Temperature mode setup................................................... 18.1.9.3 Hand and foot switch assignment ...................................... 18.1.10 “Startup Config” tab ........................................................................... 18.1.11 “kVmA Tab” tab ................................................................................. 18.1.12 “Errors” tab ........................................................................................ 18.1.13 “Times” tab ........................................................................................ 18.1.14 “DAP” tab .......................................................................................... 18.1.15 “CAN” tab .......................................................................................... 18.1.16 Installation directory .......................................................................... 18.1.17 Update procedure ............................................................................. 18.1.18 Offline mode ...................................................................................... 18.1.18.1Floppy disks for software update ....................................... 18.1.18.2Options - switch assignment .............................................. 18.1.19 Problem correction ............................................................................ 18.2 Integrated Service Functions............................................................................ 18.2.1 General ............................................................................................. 18.2.2 Entering and exiting the Service mode.............................................. 18.2.3 Important keys................................................................................... 18.2.4 Service programs .............................................................................. 18.2.4.1 Program 0 (software version)............................................. 18.2.4.2 Program 1 (iris collimator size for format 24 cm) ............... 18.2.4.3 Program 2 (iris collimator size for format 30 cm) ............... 18.2.4.4 Program 3 (iris collimator size for 40 cm format) ............... 18.2.4.5 Program 4 (iris collimator size for i.i. format) ..................... 18.2.4.6 Program 5 (iris collimator size for magnification 1) ............ 18.2.4.7 Program 6 (iris collimator size for magnification 2) ............ 18.2.4.8 Program 7 (startup key adjustment)................................... 18.2.4.9 Program 8 (program the startup key adjustment) .............. 18.2.4.10Program 9 (adjust filter codes without magnifier)............... 18.2.4.11Program 10 (adjust filter codes with magnifier 1)............... 18.2.4.12Program 11 (adjust filter codes with magnifier 2)............... 18.2.4.13Program 12 (disable or enable radiography mode) ........... 18.2.4.14Program 13 (disable or enable autotransfer) ..................... 18.2.4.15Program 14 (show fluoroscopy program kV/mA curve) ..... 18.2.4.16Program 15 (show fluoroscopy program table CCD) ......... 18.2.4.17Program 16 (select crosshair Compact) ............................
18-16 18-16 18-19 18-20 18-21 18-22 18-23 18-24 18-25 18-26 18-27 18-28 18-29 18-30 18-31 18-32 18-33 18-33 18-33 18-33 18-34 18-34 18-34 18-34 18-34 18-35 18-35 18-35 18-35 18-36 18-37 18-37 18-37 18-38 18-38 18-38 18-38 18-38
Adjustment Instructions 19.1 Dose & Video Adjustment................................................................................. 19.1.1 Dose Adjustment ............................................................................... 19.1.1.1 Dose adjustment parameters............................................. 19.1.1.2 Dose adjustment procedure............................................... 19.1.1.3 Special cases for dose adjustment .................................... 19.1.2 Video Adjustment .............................................................................. 19.1.2.1 Video adjustment parameters ............................................ 19.1.2.2 Video adjustment procedure .............................................. 19.1.3 Exposcop Studio Dialog for Dose Adjustment................................... 19.1.4 kV/mA Values for Ziehm Vista / Ziehm 8000..................................... 19.1.5 kV/mA Values for Ziehm Vista Plus................................................... 19.1.6 Dose Tables for Ziehm Vista / Ziehm 8000 ....................................... 19.1.7 Dose Tables for Ziehm Vista Plus .....................................................
19-1 19-1 19-1 19-1 19-1 19-2 19-3 19-3 19-3 19-4 19-6 19-7 19-8 19-9
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19.1.8 Ziehm Vista / Ziehm 8000 Organ Tables 31/23cm............................ 19.1.9 Ziehm Vista / Ziehm 8000 Organ Tables 17/15/11 cm...................... 19.1.10 Ziehm Vista Plus Organ Tables ........................................................ 19.2 Default Values .................................................................................................. 19.2.1 Measuring Circle Sizes ..................................................................... 19.2.2 Measuring Circle Sizes Vista Plus .................................................... 19.2.3 Monitor Cart Settings ........................................................................ 19.2.3.1 Step windowing levels ....................................................... 19.2.3.2 DSA windowing values ...................................................... 19.2.3.3 Filter factors non interlaced ............................................... 19.2.3.4 Filter factors interlaced ...................................................... 19.2.3.5 Filter factors Vista Plus ......................................................
19-10 19-11 19-12 19-13 19-13 19-13 19-14 19-14 19-14 19-15 19-15 19-16
20
Error Messages & Fault Finding 20.1 General............................................................................................................. 20.2 Startup phase ................................................................................................... 20.2.1 WARNINGS ...................................................................................... 20.2.2 ERRORS........................................................................................... 20.3 Runtime ............................................................................................................ 20.3.1 WARNINGS ...................................................................................... 20.3.2 ERRORS........................................................................................... 20.4 List of ERRORS and WARNINGS.................................................................... 20.4.1 Module 2 ........................................................................................... 20.4.2 Module 15 ......................................................................................... 20.4.3 Module 3 / U501................................................................................ 20.4.4 Module 11 ......................................................................................... 20.5 Fuse List........................................................................................................... 20.5.1 Module 1 ........................................................................................... 20.5.2 Module 3 ........................................................................................... 20.5.3 Module 10 ......................................................................................... 20.6 Fault finding...................................................................................................... 20.6.1 Power faults ...................................................................................... 20.6.2 Module 2 errors ................................................................................. 20.6.3 Module 15 errors (virtual collimator and dose meter)........................ 20.6.4 Module 3 errors ................................................................................. 20.6.5 U501 errors ....................................................................................... 20.6.6 Module 11 errors (image memory).................................................... 20.6.7 Module 6 errors (keyboard)............................................................... 20.6.8 Radiation errors ................................................................................
20-1 20-1 20-1 20-1 20-1 20-1 20-1 20-1 20-2 20-2 20-4 20-5 20-6 20-7 20-7 20-10 20-10 20-11 20-11 20-12 20-15 20-17 20-19 20-21 20-23 20-24
21
Active Cooling System 21.1 Active Cooling System ..................................................................................... 21.1.1 Initiation of the cooling system .......................................................... 21.1.1.1 General .............................................................................. 21.1.1.2 Starting the pump manually ............................................... 21.1.1.3 Filling the tank with coolant................................................ 21.1.1.4 Filling the pump with coolant ............................................. 21.1.1.5 Checking the circulation..................................................... 21.1.2 Draining the cooling system ..............................................................
21-1 21-1 21-1 21-3 21-5 21-6 21-7 21-8 21-9
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
ix
22
Maintenance Schedule 22.1 General requirements....................................................................................... 22.2 Electrics ............................................................................................................ 22.3 Mechanics ........................................................................................................ Index
x
22-1 22-1 22-2 22-4 I
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
1 Warnings and Important Information
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 1.1 1.2 1.3
Safety Instructions............................................................................................ Warnings .......................................................................................................... Required tests after service..............................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
1-1 1-2 1-2
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
1 Warnings and Important Information
1
Warnings and Important Information
1.1
Safety Instructions
This manual does not constitute a complete catalog of all safety measures necessary for the operation of the respective medical equipment, since special operating conditions may require further measures. However, it does contain instructions which must be observed in order to ensure the personal safety of operating staff and patients as well as to avoid damage to property. These instructions are highlighted as follows: DANGER Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
WARNING Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a hazardous situation which, if not avoided, may result in minor or moderate injury.
NOTICE Indicates a property damage message.
NOTE Notes are informative. Additional useful information and hints are provided for the operator here.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
1-1
1 Warnings and Important Information
1.2
Warnings WARNING The monitor cart power module contains live voltage even when the unit is switched of from the C-arm or monitor cart keyboard or the emergency switch is pushed in. Be sure to disconnect the unit from the mains supply when performing service work to the unit. WARNING The generator power supply in module 1 contains a large storage capacitor that filters the line voltage. The capacitor may contain high voltage even after the unit is disconnected from the mains supply.
WARNING Modules 1 (mobile stand power supply), 10 (monitor cart power supply) and 5 (X-ray generator) contain dangerous voltages. During operation the voltages exceed 250 VDC.
WARNING Parts of the C-arm may be contaminated infectiously. We strongly recommend to wear disposable vinyl gloves while doing service.
1.3
Required tests after service
The following tests have to be performed after every service: • Is the mechanics working properly? • Are all cables (P0, mains and swivel cable) in good condition? • Are the keyboards working properly? • Is the printout of the video printer okay (if printer is available)? • Is the system working fine? • Perform electrical safety check according to relevant national regulation of the country of installation Possible discrepancies must be recorded in the service report.
1-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 2.10 2.11 2.12 2.13 2.14 2.15 2.16 2.17 2.18 2.19
Mobile Stand System Overview ....................................................................... 2-1 Detailed Overview with U501 ........................................................................... 2-2 Detailed Overview with Module 3 ..................................................................... 2-3 Overview Ziehm 8000 Mobile Stand ................................................................ 2-4 Block Diagram with U501 ................................................................................. 2-5 Block Diagram with Module 3........................................................................... 2-6 Power Distribution Overview with U501 ........................................................... 2-7 Power Distribution Overview with Module 3 ..................................................... 2-8 Board Overview with U501............................................................................... 2-9 Board Overview with Module 3......................................................................... 2-10 Board Numbers ................................................................................................ 2-11 Numbers of CAN Modules with U501............................................................... 2-13 Numbers of CAN Modules with Module 3 ........................................................ 2-13 CAN Bus Distribution in the C-Arm Stand with U501 ....................................... 2-14 CAN Bus Distribution in the C-Arm Stand with Module 3................................. 2-15 Video Distribution in the C-Arm Stand.............................................................. 2-16 Video Distribution in the C-Arm Stand.............................................................. 2-17 Regulation Loops ............................................................................................. 2-18 Technical Data ................................................................................................. 2-19 2.19.1 Ziehm Vista and Ziehm Vista Workstation ........................................ 2-19 2.19.2 Ziehm Vista Plus (USA) .................................................................... 2-22 2.19.3Ziehm 8000, Ziehm Compact and Ziehm Compact with adapter for lithotripter procedures .................................................................. 2-24 2.19.4 General technical data ...................................................................... 2-24 2.19.4.1 Systems with a voltage rating of 100 V, 120 V, 200 V....... 2-26 2.19.4.2 Systems with a voltage rating of 220 V, 230 V, 240 V....... 2-29 2.19.5 Power plug specification ................................................................... 2-31 2.19.5.1 Systems with a voltage rating of 120 V.............................. 2-31 2.19.5.2 Systems with a voltage rating of 230 V.............................. 2-31 2.19.6 Laser positioning device ................................................................... 2-32 2.19.7 Dose meter ....................................................................................... 2-33 2.19.7.1 VacuDAP ........................................................................... 2-33 2.19.7.2 VacuDAP 2002 .................................................................. 2-34 2.19.7.3 VacuDAP duo .................................................................... 2-35 2.19.8 Air kerma........................................................................................... 2-36 2.19.9 Focal spot position ............................................................................ 2-37 2.19.10 Dimensions ....................................................................................... 2-38
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2
General
2.1
Mobile Stand System Overview
Fig. 2-1
Mobile stand system overview
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-1
2 General
2.2
Detailed Overview with U501
Fig. 2-2
Detailed Overview with U501
2-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.3
Detailed Overview with Module 3
Fig. 2-3
Detailed overview with Module 3
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-3
2 General
2.4
Overview Ziehm 8000 Mobile Stand
Fig. 2-4
Overview Ziehm 8000 Mobile Stand
Modules and parts inside the Ziehm 8000 mobile stand correlate to the drawings above.
2-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.5
Block Diagram with U501
Fig. 2-5
Mobile Stand block diagram with U501
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-5
2 General
2.6
Block Diagram with Module 3
Fig. 2-6
Mobile stand block diagram with Module 3
2-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.7
Power Distribution Overview with U501
Fig. 2-7
Power distribution overview with U501
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-7
2 General
2.8
Power Distribution Overview with Module 3
Fig. 2-8
Power distribution overview with Module3
2-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.9
Board Overview with U501
Fig. 2-9
Board overview with U501
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-9
2 General
2.10
Board Overview with Module 3
Fig. 2-10
Board overview with Module 3
2-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.11
Board Numbers
Module
Board
1
Description Power Supply
U333
Power supply control
U80
Horn
2
Mobile Stand Control U325
Mobile stand control
U326
Generator adaption
U327
Interface board
U328
Motherboard
U334
Power supply
U501
Dose and Video control (option)
3
Memory Unit (missing if U501 is equipped) U180
Memory unit control
U321
CCD processing
U306
Image memory (option)
U313
100Hz converter (option)
4
Image Intensifier U316
Image intensifier power supply control
U317
Transformer board
U318
High voltage cascade
U160
Laser control & S-distortion correction (option)
5
Generator U45
Generator interface board
U54
kV control
U43
mA control
U42
Pre-stage
U40
Power stage
U41
Motherboard
U185
Generator filter
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-11
2 General
Module
Board
6
Description Keyboard
U400 9
Keyboard mobile stand Monitor Options
E891
Controller inside the VM4402 monitor
E890
VM 4402 chassis board
E889
VM 4402 CRT board
10
Power Supply Monitor Cart U357
11
Power supply control Digital Imaging System general
U385
Interface to mobile stand Digital Imaging System V2 ½K Step 1
U359
Frame grabber
U816
Rotation/DSA/Display Board Digital Imaging System V2 ½K Step 2
U544
Frame grabber
U616
Rotation/DSA/Display Board
15
Vision Track U517
Gen. Collimator Control
U608
Gen. Collimator Control (Z-Servos)
MC
2-12
Monitor Cart U406
Coax adapter for external video socket
U533
DICOM connection wall outlet (optionally)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.12
Numbers of CAN Modules with U501 Module Name
PCB/Module Number
CAN Node Number
CPU Type
Module 2, CPU
U325
3
LUCY
Module 2, Dose and Video processing
U501
13
LUCY
Module 6, MS keyboard
U400
8
LUCY
Module 15, Virtual Collimator
U517c
10
HC08
Module 11, Image Memory
U385
6
LUCY
2.13
Numbers of CAN Modules with Module 3 Module Name
PCB/Module Number
CAN Node Number
CPU Type
Module 2, CPU
U325
3
LUCY
Module 3, Memory unit
U180
13
LUCY
Module 6, MS keyboard
U400
8
LUCY
Module 15, Virtual Collimator
U517c
10
HC08
Module 11, Image Memory
U385
6
LUCY
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-13
2 General
2.14
CAN Bus Distribution in the C-Arm Stand with U501
Fig. 2-11
CAN bus distribution in the C-arm stand with U501
2-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.15
CAN Bus Distribution in the C-Arm Stand with Module 3
Fig. 2-12
CAN bus distribution in the C-arm stand with Module 3
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-15
2 General
2.16
Video Distribution in the C-Arm Stand
Fig. 2-13
Video distribution in the C-arm stand
2-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.17
Video Distribution in the C-Arm Stand
Fig. 2-14
Video distribution in the C-arm stand
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-17
2 General
2.18
Regulation Loops
Fig. 2-15
Regulation loops
2-18
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Technical Data
2.19.1
Ziehm Vista and Ziehm Vista Workstation
System
2.19
Voltage / frequency rating
230 VAC ±10%, 50/60 Hz
Power supply fuse rating
C 16 A
Required residual current circuit breaker (RCD)
IN ≥ 16 A, IAN = 30 mA
Typical current consumption
8 A continuous, 15 A momentary
Internal fusing
15 A, fast (2 pcs.)
Maximum line impedance
≤ 0.6 Ω
Equipment protection classification
Protection Class I, Type B ( ), ordinary equipment, continuous operation
Radiation controlled area (with generator in lowermost position and C-arm vertical)
23/31 cm i.i.:
4m
15 cm i.i.:
3m
Power
Fluoroscopy:
40–110 kV / 0.2–6 mA
Direct radiography:
40–110 kV / 20 mA
Single pulse:
40–110 kV / 8 mA / 1 s
Operating frequency:
20 kHz
Fluoroscopy:
110 kV / 0.2–6 mA
Direct radiography:
110 kV / 20 mA / 0.1–4 s
Fluoroscopy:
660 W
Single pulse:
880 W
Direct radiography:
2200 W
Generator
Max. operating data
Max. power output
Nominal electric power
2000 W at 100 kV / 20 mA / 0.1 s
X-ray tube
Dual-focus stationary-anode tube
Focal spot nominal size
Fluoroscopy:
0.6 mm
Direct radiography:
1.5 mm
Total filtration
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
≥ 4 mm aluminum equivalent, including Cu ≥ 0.1 mm
2-19
Environmental conditions
PC Workstation (Ziehm Vista Workstation only)
Monitors
Image intensifier
2 General
2-20
Tube
Input screen:
Cesium iodide
Nominal sizes:
31 / 23 / (15) cm or 23 / 15 / (10) cm or 15 / 10 cm
Anti-scatter grid
Pb 8/40
Flatscreen monitors
Screen size:
46 cm (18.1")
Luminance:
600 Cd/m2
Resolution:
1280 × 1024 pixels
Refresh rate:
75 Hz
Screen size:
44 cm (17")
Bandwidth:
45 MHz
Resolution:
625 lines
Refresh rate:
75 Hz (non-interlaced)
CRT monitors
Compaq Evo D 500 or similar models
For technical data, see documentation of PC manufacturer Default settings:
Screen resolution: 1280 × 1024 pixels with 16 bit color depth No screen saver No standby mode
PC keyboard:
Integrated in monitor cart
DICOM interface:
RJ45
Software:
Operating system: Microsoft® Windows® 2000 Monitor driver: Pivot Pro (for rotating the image) DICOM Workstation: e-Film for image viewing and editing
During storage and transport
Temperature:
–5°C to +55°C
Relative air humidity:
20–70%
During operation
Temperature:
+10°C to +35°C
Relative air humidity:
20–70%
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Weight
Dimensions
C-arm
Source/image receptor distance:
97 cm
Vertical free space (generator/i.i.):
76 cm
C-arm depth:
68 cm
Orbital rotation:
115° / 135° a
Angulation:
±225°
Swiveling (panning):
±10°
Horizontal movement:
22 cm
Vertical movement:
43 cm / 63 cm a
with 23 cm i.i.
Width: 80 cm Depth: 230 cm Height: 214 cm
with 31 cm i.i.
Width: 80 cm Depth: 236 cm Height: 219 cm
Monitor cart
Ziehm Vista, Ziehm Vista Workstation
Width: 71 cm Depth: 69.5 cm Height: 169 cm
C-arm stand
Ziehm Vista
approx. 286 kg (23 cm i.i.)
Monitor cart
Ziehm Vista with flatscreen monitors:
approx. 177 kg
Ziehm Vista with CRT monitors:
approx. 196 kg
Ziehm Vista Workstation:
approx. 180 kg
C-arm stand
a. Option, not available for 31 cm i.i. systems
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-21
2 General
System
2.19.2
Ziehm Vista Plus (USA) Voltage / frequency rating
120 VAC ± 10%, 50/60 Hz
230 VAC ± 10%, 50/60 Hz
Power supply fuse rating
C 20 A
C 16 A
Required residual current circuit breaker (RCD)
IN ≥ 16 A, IAN = 30 mA
IN ≥ 16 A, IAN = 30 mA
Typical current consumption
10 A continuous, 20 A momentary
8 A continuous, 15 A momentary
Internal fusing
20 A, fast (2 pcs.)
15 A, fast (2 pcs.)
Maximum line impedance
≤ 0.3 Ω
≤ 0.6 Ω
Equipment protection classification
Protection Class I, Type B ( ), ordinary equipment, continuous operation
Radiation controlled area (with generator in lowermost position and C-arm vertical)
23/31 cm i.i.: 4 m
23/31 cm i.i.: 4 m
15 cm i.i.: 3 m
15 cm i.i.: 3 m
Fluoroscopy:
40–110 kV / 0.1–6 mA
40–110 kV / 0.2–6 mA
Direct radiography:
40–110 kV / 18–20 mA
40–110 kV / 20 mA
Operating frequency:
20 kHz
Power
Generator
Max. operating data Fluoroscopy:
110 kV / 0.2–6 mA
110 kV / 0.2–6 mA
Direct radiography:
110 kV / 8 mA
110 kV / 20 mA
Exposure time:
0.1–4 s
0.1–4 s
Fluoroscopy:
660 W
650 W
Direct radiography:
1980 W
2200 W
Single pulse:
880 W
880 W
Nominal electric power
2000 W at 100 kV / 20 mA / 0.1 s
X-ray tube
Dual-focus stationary-anode tube DF-151-R
Focal spot nominal size
Fluoroscopy: 0.5 or 0.6 mm (depending on the X-ray tube)
Max. power output
Direct radiography: 1.5 mm Total filtration
2-22
≥ 4 mm aluminum equivalent, including Cu ≥ 0.1 mm
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Image intensifier
2 General
Tube
Scintillator: Cesium iodide Nominal sizes: 23 / 15 cm
Anti-scatter grid
Pb 8/40
Environmental conditions
Monitors
Flatscreen monitors Screen size:
46 cm (18.1")
Luminance:
600 Cd/m2
Resolution:
1280 × 1024 pixels
Refresh rate:
75 Hz
During storage and transport
Temperature:
–5°C to +55°C
Relative air humidity:
20–70%
During operation
Temperature:
+10°C to +35°C
Relative air humidity:
20–70%
C-arm
Source/image receptor distance: 97 cm Vertical free space (generator/i.i.): 76 cm C-arm depth: 68 cm Orbital rotation: 115 / 135°
Weight
Dimensions
Angulation: ± 225° Swiveling (panning): ±10° Horizontal movement: 22 cm Vertical movement: 43 cm C-arm stand
Width: 80 cm Depth: 230 cm Height: 214 cm
Monitor cart
Width: 71 cm Depth: 69.5 cm Height: 169 cm
C-arm stand
Ziehm Vista Plus: approx. 286 kg
Monitor cart
Ziehm Vista Plus: approx. 177 kg
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-23
2 General
Ziehm 8000, Ziehm Compact and Ziehm Compact with adapter for lithotripter procedures
2.19.4
General technical data
Image intensifier
2.19.3
Tube Scintillator
Cesium iodide
Nominal sizes
23 / 15 / 10 cm
Anti-scatter grid
Pb 8/40
Monitors on monitor cart
24" flat-screen monitors Display measurements
520 mm × 325 mm (20.4” × 12.8”)
Screen size (diagonal)
610 mm (24")
Luminance
max. 400 Cd/m2
Refresh rate
85 Hz
Resolution in pixels
1920 × 1200
18.1" flat-screen monitors Display measurements
300 mm × 360 mm (11,45” × 14,28”)
Screen size (diagonal)
460 mm (18.1")
Luminance
max. 600 Cd/m2
Refresh rate
75 Hz
Resolution in pixels
1280 × 1024
Video standard output
Compact monitors
Flat-screen monitor Display measurements
300 mm × 360 mm (11,45” × 14,28”)
Screen size (diagonal)
460 mm (18.1")
Luminance
max. 600 Cd/m2
Refresh rate
75 Hz
Resolution in pixels
1280 × 1024
CCIR, 50 Hz refresh rate, like PAL, no color EIA 343, 60 Hz refresh rate, like NTSC, no color
Table 2-1
2-24
General technical data of Ziehm 8000, Ziehm Compact and Ziehm Compact with adapter for lithotripter procedures
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Environmental conditions
During storage/transport Temperature
–5°C to +55°C
Relative air humidity
20 % – 70 %
During operation Temperature
+13°C to +35°C
Relative air humidity
20 % – 70 %
C-arm Source/image receptor distance 970 mm Vertical free space (generator/i.i.) 760 mm Dimensions
C-arm depth
680 mm
Orbital rotation Ziehm 8000
23 cm image intensifier: 135° (+45° – -90°)
Ziehm Compact
23 cm image intensifier: 115° (+25° – -90°)
Ziehm Compact with adapter for lithotripter procedures
23 cm image intensifier: 115° (+25° – -90°)
Angulation
±225°
Swiveling (panning)
±10°
Horizontal movement
220 mm
Vertical movement
420 mm
Weight
C-arm stand Ziehm 8000
23 cm image intensifier: approx. 264 kg
Ziehm Compact
23 cm image intensifier: approx. 270 kg
Ziehm Compact with adapter for lithotripter procedures
23 cm image intensifier: approx. 280 kg
Monitor cart
Table 2-1
Ziehm Vista with 24" flat-screen monitor
approx. 150 kg
Ziehm Vista with 18.1" flat-screen monitor
approx. 153 kg
General technical data of Ziehm 8000, Ziehm Compact and Ziehm Compact with adapter for lithotripter procedures (cont.)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-25
2 General
2.19.4.1
Systems with a voltage rating of 100 V, 120 V, 200 V
Systems with a voltage / frequency rating of:
100 VAC ± 10%, 50/60 Hz
120 VAC ± 10%, 50/60 Hz
200 VAC ± 10%, 50/60 Hz
Power supply fuse rating
C 20 A or C 32 A
C 20 A
C 20 A or C 32 A
Required residual current circuit breaker (RCD)
IN ≥ 20 A,
IN ≥ 20 A,
IN ≥ 16 A,
IAN = 30 mA
IAN = 30 mA
IAN = 30 mA
14 A continuous, 22 A momentary
10 A continuous, 16 A momentary
460 VA (3,8 A)
460 VA (2,3 A)
(tripping characteristic C acc. to VDE 0641, Part 11; DIN EN 60898 + IEC 898)
Typical current consumption 14 A continuous, 22 A momentary
System
Power supply in standby mode 410 VA (4,3 A)
The values depend on the integrated documentation systems. Power input fuse 20 A slow
20 A slow
15 A slow
Maximum line impedance
≤ 0.3 Ω
≤ 0.3 Ω
≤ 0.6 Ω
Equipment protection classification
Protection Class I, Type B, ordinary equipment, continuous operation
Radiation controlled area (with generator in lowermost position and C-arm vertical) 15 cm image intensifier: 3 m 23 cm image intensifier: 4 m Table 2-2
2-26
Systems with a voltage rating of 100 V, 120 V, 200 V
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Systems with a voltage / frequency rating of:
100 VAC ± 10%, 50/60 Hz
120 VAC ± 10%, 50/60 Hz
200 VAC ± 10%, 50/60 Hz
Power Direct radiography 40–110 kV 20 mA min. 0,1 - 4 s Continuous fluoroscopy 40–110 kV 0.1–6 mA Single pulse fluoroscopy (snapshot) 40–110 kV 8 mA 1s Operating frequency
20 kHz
DC-HV ripple
40 kHz
Generator
Max. operating data Fluoroscopy 110 kV / 6 mA Direct radiography 110 kV / 20 mA Digital radiography 110 kV / 8 mA Max. power output Fluoroscopy 660 W (110 kV / 6 mA) Direct radiography 2200 W (110 kV / 20 mA) Digital radiography 880 W (110 kV / 8 mA) Nominal electric power 2000 W at 100 kV / 20 mA / 0.1 s Table 2-2
Systems with a voltage rating of 100 V, 120 V, 200 V (cont.)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-27
2 General
Systems with a voltage / frequency rating of:
100 VAC ± 10%, 50/60 Hz
120 VAC ± 10%, 50/60 Hz
200 VAC ± 10%, 50/60 Hz
X-ray tube Dual-focus stationary-anode tube Focal spot nominal size, in relation to reference axis Fluoroscopy: 0.5 acc. to IEC 336
Generator
Direct radiography: 1,5 acc. to IEC 336 Anode angle, in relation to reference axis 12° Anode material Tungsten Total filtration ≥ 3.9 mm Al, including Cu ≥ 0.1 mm Maximum X-ray tube loading factors for 1h; 3 mA at 110 kV Table 2-2
2-28
10800 mAs/h
Systems with a voltage rating of 100 V, 120 V, 200 V (cont.)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.19.4.2
Systems with a voltage rating of 220 V, 230 V, 240 V
Systems with a voltage / frequency rating of:
220 VAC ± 10%, 50/60 Hz
230 VAC ± 10%, 50/60 Hz
240 VAC ± 10%, 50/60 Hz
Power supply fuse rating
C 16 A (tripping characteristic C acc. to VDE 0641, Part 11; DIN EN 60898 + IEC 898)
Required residual current circuit breaker (RCD)
IN ≥ 16 A, IAN = 30 mA
Typical current consumption
10 A continuous, 16 A momentary
Power supply in standby mode
System
approx. 370 VA (approx. 1.6 A) The values depend on the integrated documentation systems. Power input fuse
15 A slow
Maximum line impedance
≤ 0.6 Ω
Equipment protection classification
Protection Class I, Type B, ordinary equipment, continuous operation
Radiation controlled area (with generator in lowermost position and C-arm vertical) 15 cm image intensifier: 3 m 23 cm image intensifier: 4 m Power Direct radiography 40–110 kV 20 mA, 0.1–4 s Generator
Fluoroscopy 40–110 kV 0.1–6 mA Digital radiography (snapshot) 40–110 kV 8 mA 1s Operating frequency
20 kHz
DC-HV ripple
40 kHz
Table 2-3
Systems with a voltage rating of 220 V, 230 V, 240 V
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-29
2 General
Systems with a voltage / frequency rating of:
220 VAC ± 10%, 50/60 Hz
230 VAC ± 10%, 50/60 Hz
240 VAC ± 10%, 50/60 Hz
Max. operating data Fluoroscopy 110 kV / 6 mA Direct radiography 110 kV / 20 mA Digital radiography 110 kV / 8 mA Max. power output Fluoroscopy 660 W (110 kV / 6 mA) Direct radiography 2200 W (110 kV / 20 mA)
Generator
Digital radiography 880 W (110 kV / 8 mA) Nominal electric power 2000 W at 100 kV / 20 mA / 0.1 s X-ray tube Dual-focus stationary-anode tube Focal spot nominal size, in relation to reference axis Fluoroscopy: 0.5 acc. to IEC 336 Direct radiography: 1,5 acc. to IEC 336 Anode angle, in relation to reference axis 12° Anode material Tungsten Total filtration ≥ 3.9 mm Al, including Cu ≥ 0.1 mm Maximum X-ray tube loading factors for 1h; 3 mA at 110 kV Table 2-3
2-30
10800 mAs/h
Systems with a voltage rating of 220 V, 230 V, 240 V (cont.)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.19.5
Power plug specification
2.19.5.1
Systems with a voltage rating of 120 V 2.5 mm2
Cables
three-wire supply voltage: two wires ground wire: one wire Phase
single-phase
Strain relief
required
Voltage
125 VAC
Dielectric strength
> 2000 V
Current-carrying capacity
15 A continuous (min.)
Flammability rating
acc. to UL 2305 OEM
Temperature
–40°C to +80°C
Mark of conformity
VDE or equivalent safety testing and national mark of conformity of country of installation (if required)
Degree of protection
IP 20 or higher
2.19.5.2
Systems with a voltage rating of 230 V
Cables
2.5 mm2 three-wire supply voltage: two wires ground wire: one wire
Phase
single-phase
Strain relief
required
Voltage
250 VAC
Dielectric strength
> 2000 V
Current-carrying capacity
16 A continuous (min.)
Flammability rating
acc. to UL 94 V0
Temperature
–40°C to +80°C
Mark of conformity
VDE or equivalent safety testing and national mark of conformity of country of installation (if required)
Degree of protection
IP 20 or higher
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-31
2 General
2.19.6
Laser positioning device
Laser Class
Class 2M acc. to IEC 60825-1:2001
Max. power output of continuous laser radiation, measured at the laser beam apertures
< 1 mW
Wavelength of the radiation
635 nm
2-32
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.19.7
Dose meter
2.19.7.1
VacuDAP
Additional absorption
2.8 mm Al
Sensitivity (75 kV; 2.7 mm Al HVL)
≥ 800 pC / mGy•cm2
Measuring range of DAP power
0.1–104 mGy • cm2/s
Voltage range
35–150 kV
Voltage correction
→ Fig. 2-16, p. 2-33
Aluminum equivalent
< 0.4 mm Voltage correction
Correction factor
Additional absorption: 2.8 mm Al filtration
Fig. 2-16
Voltage correction of VacuDAP dose meter
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-33
2 General
2.19.7.2
VacuDAP 2002
Additional absorption
3 mm Al
Sensitivity (75 kV; 2.7 mm Al HVL)
≥ 700 pC / mGy•cm2
Measuring range of DAP power
0.1–104 mGy • cm2/s
Voltage range
40–150 kV
Voltage correction
→ Fig. 2-17, p. 2-34
Aluminum equivalent
< 0.4 mm Voltage correction
Correction factor
Additional absorption: 3 mm Al filtration
Fig. 2-17
2-34
Voltage correction of VacuDAP 2002 dose meter
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.19.7.3
VacuDAP duo
Additional absorption
2.8 mm Al
Sensitivity (70 kV; 2.7 mm Al HVL)
≥ 1350 pC / mGy•cm2
Measuring range of DAP power
0.75•10-2 – 0.3•104 mG•cm2/s
Voltage range
40–110 kV
Voltage correction
→ Fig. 2-18, p. 2-35
Aluminum equivalent
0.5 mm Voltage correction
Correction factor
Additional absorption: 2.8 mm Al filtration Additional absorption: 2.8 mm + 1 mm Al filtration
X-ray tube voltage U [kV]
Fig. 2-18
Voltage correction of VacuDAP duo dose meter
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-35
2 General
2.19.8
Air kerma
In compliance with the requirements of international standards (e.g. 21 CFR 1020.32-d (1), IEC 606012-43), Ziehm fluoroscopy systems with automatic dose rate control are not operable at any combination of tube voltage and tube current which will result in an exposure rate in excess of 88 mGy/min at the point where the center of the useful beam enters the patient, referred to the interventional reference point.
Image intensifier input plane Lead blocker Dose probe Interventional reference point
Focal spot
Fig. 2-19
Interventional reference point
The air kerma that is displayed on the system’s user interface refers to the interventional reference point according to IEC 60601-2-43 and 21 CFR 1020.32-d.
Deviation in %
Air kerma deviation
Voltage in kV
Fig. 2-20
2-36
Typical deviation of dose as compared to an external calibrated measurement system
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
2.19.9
Focal spot position
7° 7°
Reference axis
Fig. 2-21
Focal spot position on systems with 23 cm i.i.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-37
2 General
2.19.10
Dimensions
Ziehm Vista C-arm stand with 23 cm i.i.
Fig. 2-22
Dimensions of Ziehm Vista C-arm stand on systems with 23 cm i.i.
Ziehm Vista C-arm stand with 31 cm i.i. 25°
Fig. 2-23
2-38
90°
225° 225° 10° 10°
Dimensions of Ziehm Vista C-arm stand on systems with 31 cm i.i.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Ziehm Vista monitor cart with flatscreen monitors
Fig. 2-24
Dimensions of Ziehm Vista monitor cart with flatscreen monitors
Ziehm Vista monitor cart with CRT monitors (n/a for USA)
Fig. 2-25
Dimensions of Ziehm Vista monitor cart with twin CRT monitors (n/a for USA)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-39
2 General
Ziehm Vista Workstation monitor cart
Fig. 2-26
Dimensions of a Ziehm Vista Workstation monitor cart
520 - 940
760
800
155
305
345
925
970
745
1710 - 2130
420
450
Ziehm 8000 C-arm stand with 23 cm i.i.
Fig. 2-27
2-40
Dimensions of Ziehm 8000 C-arm stand on systems with 23 cm i.i.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Tolerances of Ziehm 8000 C-arm movements
Fig. 2-28
Tolerances of Ziehm 8000 C-arm movements on systems with 23 cm i.i.
Ziehm 8000 C-arm stand with lower C-arm position
Fig. 2-29
Dimensions of Ziehm 8000 C-arm stand with lower C-arm position
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-41
2 General
Ziehm 8000 monitor cart with 23" flat-screen monitor
Fig. 2-30
2-42
Dimensions of a Ziehm 8000 monitor cart with 23" flat-screen monitor
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2 General
Ziehm 8000 monitor cart with 24" flat-screen monitor
Fig. 2-31
Dimensions of a Ziehm 8000 monitor cart with 24" flat-screen monitor
Ziehm 8000 monitor cart with 18.1" flat-screen monitors
Fig. 2-32
Dimensions of a Ziehm 8000 monitor cart with 18.1" flat-screen monitors
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
2-43
2 General
Fig. 2-33
1710 - 2130 520 - 940
760
155
305
345
925
970 745
45 130
420
450
Ziehm Compact with flat-screen monitor and 23 cm i.i.
Dimensions of a Ziehm Compact with flat-screen monitor and 23 cm i.i.
Ziehm Compact with adapter for lithotripter procedures and 23 cm i.i. 225°
225°
90°
0°
0°
155
267 1600 2060
2-44
535 - 765
760
212
345
925
680
400
Fig. 2-34
1725 - 1955
130
0
45
745
970
130
230
450
25°
420 700 800
Dimensions of Ziehm Compact with adapter for lithotripter procedures and 23 cm i.i.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 3.1
XP0 Cable ........................................................................................................ 3.1.1 Connectors........................................................................................ 3.1.2 Wiring ................................................................................................ 3.2 C-Arm Wiring Diagram ..................................................................................... 3.3 CAN Connector ................................................................................................ 3.4 Grounds Overview............................................................................................ 3.5 Hand Switch & CAN Bus Interface Wiring........................................................ 3.6 Foot Switch Wiring ........................................................................................... 3.7 Four-Pedal Foot Switch.................................................................................... 3.8 Module 1........................................................................................................... 3.8.1 Main connector ................................................................................. 3.8.2 U333 connectors ............................................................................... 3.9 Module 2........................................................................................................... 3.9.1 Connector X60 .................................................................................. 3.9.2 Connector X81 .................................................................................. 3.9.3 Connector X62 .................................................................................. 3.9.4 Connector X83 .................................................................................. 3.9.5 Connector X84 .................................................................................. 3.9.6 Connectors X70, X71, X72, X73 ....................................................... 3.9.7 Connector X74 .................................................................................. 3.10 Module 3 / U501 ............................................................................................... 3.10.1 U180 connector X17 / U501 connector X17...................................... 3.10.2 U180 connector X9 / U501 connector X9.......................................... 3.10.3 U180 connector X5 ........................................................................... 3.10.4 U180 connector X12 / U501 connector X12...................................... 3.10.5 U321 connectors X7, X3 / U501 connectors X7, X3 ......................... 3.11 Module 4........................................................................................................... 3.11.1 Link connector C-arm – i.i. housing................................................... 3.11.2 U316 connector X9 ........................................................................... 3.11.3 U316 connector X11 ......................................................................... 3.12 Module 5........................................................................................................... 3.12.1 Signals and voltages on U185 generator filter ..................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-1 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-9 3-10 3-16 3-16 3-16 3-17 3-17 3-17 3-18 3-18 3-19 3-19 3-19 3-19 3-20 3-20 3-21 3-21 3-22 3-22 3-23 3-23
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3
Mobile Stand Connections
3.1
XP0 Cable
3.1.1
Connectors
Fig. 3-1
XP0 cable connector assignment
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-1
3 Mobile Stand Connections
3.1.2
Wiring
Fig. 3-2
XP0 cable wiring mobile stand
3-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.2
C-Arm Wiring Diagram
Fig. 3-3
C-arm wiring diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-3
3 Mobile Stand Connections
3.3
CAN Connector
Fig. 3-4
CAN connector
CAN connectors are: • X70, X71, X72, X73 on U327 of module 2 • X74 on U327 of module 2 • X10, X13 on U333 of module 1 • X7 on U180 of module 2 • X1 on U400 of module 6 NOTE X74 is only to connect the monitor cart via the XP0 connector. On this connector Pin 2 is CAN sense (→ Ch. 3.9.7, p. 3-18).
3-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.4
Grounds Overview
Fig. 3-5
Grounds overview
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-5
3 Mobile Stand Connections
3.5
Hand Switch & CAN Bus Interface Wiring
Fig. 3-6
Hand switch & CAN bus interface wiring
X5 is the hand switch socket of the C-arm. It is also used to connect a notebook for service purposes.
3-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.6
Foot Switch Wiring
Fig. 3-7
Foot switch wiring
X1 is the foot switch socket of the C-arm.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-7
3 Mobile Stand Connections
3.7
Four-Pedal Foot Switch
Fig. 3-8
Four-pedal foot switch NOTE Systems equipped with a four-pedal foot switch do not support a hand switch. The hand switch socket is only used as a CAN bus interface for service.
3-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.8
Module 1
3.8.1
Main connector
d
Name
b
z
Name
CANH
2
CAN Reserve
2
CAN-L
CANL
2
4
+5 V for CAN
CAN+5V
4
4
/CAN fault
6
GND for CAN
CANGND
6
6
/CAN radiation
8
+18 V for Horn
+18VB
8
GND +18 V for horn
GND18VB
8
+24 V stabilized
10
+18 V
+18VB
10
GND +18 V
GND18VB
10
GND for +24 V stabilized
12
/Pump
OPUMPE
12
GND for /pump
PUMGND
12
14
+24 V from rectifier
+24V~
14
GND +24 V from rectifier
GND +24 VAC
14
/power off
16
+24 V from rectifier
+24V~
16
GND +24 V from rectifier
GND +24 VAC
16
/power on
18
/power switching 2
POWER2
18
horn
18
GND for power on / power off
20
/power switching 1
POWER1
20
dose rate serial
KX_SIGN
20
GND for DAP meter
22
GND for Analog Aux.
GNDD
22
dose rate /Test1
KX_TST1
22
dose rate /Test2
24
Analog Aux. input
ANA_AUX
24
dose rate /Reset
KX_RES
24
+18 V for DAP meter
26
AC. E5 for +5 V Video
E5
26
AC. E6 for +5 V Video
E6
26
AC supply voltage 1 for rotation (Compact)
28
AC. E4 for ±5 V dig.
E4
28
28
AC supply voltage 2 for rotation (Compact)
30
AC. E3 for ±5 V dig.
E3
30
30
AC. E8 for ±5 V Video
32
AC. E1 for +5 V CAN
E1
32
32
AC. E7 for ±5 V Video
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
CAN-H
Name
AC. E2 for +5 V CAN
E2
3-9
3 Mobile Stand Connections
3.8.2
U333 connectors
X1: Mains input connector STV4 connector Part Number: 21388 Pin 1:
230 V L1
Pin 2:
110 V L1
Pin 3:
110 V N
Pin 4:
230 V N
X2: To TR1 STV4 connector Part Number: 21388 Pin 1:
230 V
Pin 2:
110 V
Pin 3:
0V
Pin 4:
Not used
X5: To Compact unit MSTB2,5/3 connector Part Number: 21368 Pin 1:
SL
Pin 2:
0V
Pin 3:
230 V
X6 and X8 RM6006 connector Part Number: 23097 X6:
To softstart resistor
X8:
To softstart resistor
3-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
X7: To TR 2 STV4 connector Part Number: 21388 Pin 1:
230 V from Tr1
Pin 2:
0 V from TR1
Pin 3:
To F1 of TR2
Pin 4:
PE to chassis
X9: TR1 generator supply STV4 connector Part Number: 21388 Pin 1:
From GL2~
Pin 2:
From TR1 90 V
Pin 3:
From TR1 136 V
Pin 4:
From TR1 195 V
X10: CAN to keyboard 6410-8 molex connector Part Number: 21212 Pin 1:
CAN-shield
Pin 2:
CAN +5 V
br
Pin 3:
CAN-GND
wh
Pin 4:
not used
Pin 5:
CAN-H
pi
Pin 6:
CAN-L
gr
Pin 7:
CAN-STÖR
yl
Pin 8:
CAN-STR
gn
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-11
3 Mobile Stand Connections
X11: From main connector MLSS100-07 Panduit connector Part Number: 21354 Pin 1:
NETZEIN
z16
yl
Pin 2:
NETZAUS
z14
gn
Pin 3:
EAGND
z18
bk
Pin 4:
not used
Pin 5:
/PWR2
d18
yl
Pin 6:
/PWR1
d20
gn
Pin 7:
GND24V
b14
bk
X12: From GL4 MLSS156-04 Panduit connector Part Number: 21313 Pin 1:
not used
Pin 2:
not used
Pin 3:
From GL4 +
Pin 4:
From GL4 -
X13: CAN from main connector 6410-8 molex connector Part Number: 21212 Pin 1:
CAN-shield
Pin 2:
CAN +5 V
d4
br
Pin 3:
CAN-GND
d6
wh
Pin 4:
not used
Pin 5:
CAN-H
b2
pi
Pin 6:
CAN-L
d2
gr
Pin 7:
CAN-STÖR
z4
yl
Pin 8:
CAN-STR
z6
gn
3-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
X15 and X16 RM6006 connector Part Number: 23097 X15:
To Generator -
X16:
To Generator +
X17: To keyboard 2 609-1027 flat cable connector Part Number: 21090 Pin 1:
Mains on
Pin 2:
Mains off
Pin 3:
EAGND
Pin 4:
Mains LED
Pin 5:
Lift +
Pin 6:
Lift
Pin 7:
Lift
Pin 8:
Lift -
Pin 9:
+18 V GND
Pin 10:
+18 V
X18: Limit switch lift motor MLSS100-04 Panduit connector Part Number: 21329 Pin 1:
lift -
Pin 2:
down
Pin 3:
lift +
Pin 4:
up
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-13
3 Mobile Stand Connections
X14: To electrolytic capacitor STV4 connector Part Number: 21388 Pin 1:
From GL2 -
Pin 2:
to electrolytic capacitor -
Pin 3:
From GL2 +
Pin 4:
to electrolytic capacitor +
X19: Via main connector to image processing MLSS156-02 Panduit connector Part Number: 21326 Pin 1:
+18 V
Pin 2:
0V
X20: to water pump MLSS156-02 Panduit connector Part Number: 21326 Pin 1:
approx +30 V if pump is switched on
Pin 2:
0V
X21 and X22 RM6006 connector Part Number: 23097 X21:
To lift motor
rd
X22:
To lift motor
gn
X23: Lift motor supply STV4 connector Part Number: 21388 Pin 1:
from GL1 +
bk
Pin 2:
to R02
br
Pin 3:
from GL1 -
bk
Pin 4:
to R01
br
3-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
X24: From GL3 MLSS156-04 Panduit connector Part Number: 21313 Pin 1:
From GL3 +
Pin 2:
From GL3 -
Pin 3:
to main connector d16
Pin 4:
to main connector b16
X26: To transformer thermo switch MLSS100-02 Panduit connector Part Number: 21335 Pin 1:
From TR1
Pin 2:
To NEAGND
X27: From GL1 (Only U333g and above) MLSS156-02 Panduit connector Part Number: 21326 Pin 1:
From GL1 ~
Pin 2:
From GL1 ~
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-15
3 Mobile Stand Connections
3.9
Module 2
3.9.1
Connector X60
Power supply for module 003 Pin
signal
description
level active
level not active
reference
1
GND18VB
ground for +18 V
3
+18VB
+18 V
DC +18 V
Pin 1
4
+18VB
+18 V
DC +18 V
Pin 1
2
3.9.2
Connector X81
Connector to generator Pin
signal
description
level active
level not active
reference
1
KVMASSE
ground for ma. actual value
DC 0.7 V
DC 0.7 V
GNDGEN X63 Pin 3
2
KVSOLL
kV nominal value
DC 8.5 to 13 V
DC 8.5 to 13 V
GNDGEN
3
STÖRUNG
fault line from generator
DC 0.5 V
DC 5 V
GNDGEN
4
STRAHL
radiation on
DC 0.5 V
DC 15 V
GNDGEN
5
MAIST
ma. actual value
6
VORB
radiation preparation
DC 0.5 V
DC 15 V
GNDGEN
7
0.5MA
overheat
DC 0.5 V
DC 5 V
GNDGEN
8
+5V_IRIS
+5 V for iris diaphragm position sensor
DC 5 V
DC 5 V
GND_IRIS
9
GND_IRIS
ground for iris diaphragm position sensor
10
IRIS_IST
iris diaphragm position sensor
DC 0 V – 5 V
DC 0 V – 5 V
GND_IRIS
11
IRIS_ZU
close iris diaphragm
DC 0.5 V
DC 15 V
GNDGEN
12
IRIS_AUF
open iris diaphragm
DC 0.5 V
DC 15 V
GNDGEN
13
SCHL_ZU
close slot diaphragm
DC 0.5 V
DC 15 V
GNDGEN
14
SCHL_RE
open slot diaphragm
DC 0.5 V
DC 15 V
GNDGEN
15
SCHL_LI
rotate slot diaphragm left
DC 0.5 V
DC 15 V
GNDGEN
16
SCHL_AUF
rotate slot diaphragm right
DC 0.5 V
DC 15 V
GNDGEN
3-16
GNDGEN
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.9.3
Connector X62
Power supply for generator Pin
signal
description
level active
1
UH
filament voltage
DC 9 V – 20 V
Pin 3
2
+15VGEN
+15 V generator supply
+15 V
Pin 3
3
GNDGEN
ground
3.9.4
level not active
reference
Connector X83
Foot switch 1 and 2 Pin
signal
description
level active
level not active
reference
1
FUSS_1
foot switch 1
DC 1.75 V
DC 3.2 V if connected
Pin 2
2
ISOGND
ground
3
FUSS_2
foot switch 2
DC 1.75 V
DC 3.2 V if connected
Pin 4
4
ISOGND
ground
3.9.5
Connector X84
Hand switch 1 and 2 Pin
signal
description
level active
level not active
reference
1
HAND_1
hand switch 1
DC 1.75 V
DC 3.2 V if connected
Pin 2
2
ISOGND
ground
3
HAND_2
hand switch 2
DC 1.75 V
DC 3.2 V if connected
Pin 4
4
ISOGND
ground
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-17
3 Mobile Stand Connections
3.9.6
Connectors X70, X71, X72, X73
CAN bus Pin
signal
description
1
C_SCHIRM
shield to ground
2
CAN+5V
+5 V
3
CANGND
ground
4
CANRESV
reserve
not used
5
CANH
CAN bus data H
pulses
Pin 3
6
CANL
CAN bus data L
pulses
Pin 3
7
CANSTÖR
fault
TTL Low
TTL High
Pin 3
8
CANSTR
radiation
TTL Low
TTL High
Pin 3
level active
level not active
reference
DC +1 V Monitor cart connected
DC +5 V Monitor cart not connected
Pin 3
not used
3.9.7
level active
level not active
DC +5 V
reference
Pin 3
not used
Connector X74
CAN bus to P0 cable Pin
signal
description
1
C_SCHIRM
shield to ground
2
CAN sense
C-arm / monitor cart detection
3
CANGND
ground
4
CANRESV
reserve
not used
5
CANH
CAN bus data H
pulses
Pin 3
6
CANL
CAN bus data L
pulses
Pin 3
7
CANSTÖR
fault
TTL Low
TTL High
Pin 3
8
CANSTR
radiation
TTL Low
TTL High
Pin 3
3-18
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.10
Module 3 / U501
Signals on Module 3 are identical to signals on U501. Either Module 3 or U501 is equipped.
3.10.1
U180 connector X17 / U501 connector X17
Supply image intensifier power block Pin
signal
description
level active
level not active
reference
1
GND
ground
2
Magnifier1
switch to magnifier 1
DC 0 V
DC 10.7 V
Pin 1
3
H Sync
horizontal sync
5 Vss
4
Magnifier2
switch to magnifier 2
DC 0 V
4
18V
18 V supply voltage
DC 18 V
3.10.2
Pin 1 DC 10.7 V
Pin 1 Pin 1
U180 connector X9 / U501 connector X9
Control CCD iris Pin
signal
1
+ iris motor
description
level active
level not active
reference Pin 2
open and close CCD iris 2
- iris motor
3
GND
ground
4
iris potentiometer
sliding contact
5
+5V
supply for iris potentiometer
3.10.3
Pin 1
DC 0 V – 5 V DC 5 V
Pin 3 Pin 3
U180 connector X5
Supply memory unit Pin
signal
description
1
GND18V
ground
2
+18V
18 V supply voltage
DC 18 V
Pin 1
3
+18V
18 V supply voltage
DC 18 V
Pin 1
4
level active
level not active
reference
not used
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-19
3 Mobile Stand Connections
3.10.4
U180 connector X12 / U501 connector X12
AGC out Pin
3.10.5
signal
description
AGC out
AGC control for CCD chip
level active
level not active
reference
U321 connectors X7, X3 / U501 connectors X7, X3
Video Pin
signal
description
X7
Video in
video input
X3
Video out
video output
3-20
level active
level not active
reference
1Vss
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.11
Module 4
3.11.1
Link connector C-arm – i.i. housing
Fig. 3-9
Connection between II Housing and C-Profile
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-21
3 Mobile Stand Connections
3.11.2
U316 connector X9
Supply CCD camera Pin
signal
description
1
GND
ground
2
H Sync
horizontal sync
3
GND
ground
4
12V
supply voltage CCD camera
5
GND
ground
6
12V
supply voltage CCD camera
3.11.3
level active
level not active
reference
DC 12 V
Pin 1
DC 12 V
Pin 1
U316 connector X11
Control CCD iris Pin
signal
1
+ iris motor
description
level active
level not active
reference Pin 2
open and close CCD iris 2
- iris motor
3
GND
ground
4
iris potentiometer
sliding contact of iris potentiometer
5
+5V
supply for iris potentiometer
3-22
Pin 1
DC 0 V – 5 V DC 5 V
Pin 3 Pin 3
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3 Mobile Stand Connections
3.12
Module 5
3.12.1
Signals and voltages on U185 generator filter
Fig. 3-10
Signals and voltages on U185 generator filter
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
3-23
3 Mobile Stand Connections
3-24
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply Module 1
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9
4.10
4.11 4.12 4.13 4.14 4.15
Trafoblech View................................................................................................ Module 1 Assembly .......................................................................................... Block Diagram .................................................................................................. U333 Board versions........................................................................................ Wiring Diagram with U333f............................................................................... Wiring Diagram with U333g.............................................................................. Functional Diagram with U333f ........................................................................ Functional Diagram with U333g ....................................................................... Circuit Description ............................................................................................ 4.9.1 Power-up circuit ................................................................................ 4.9.2 Output switching circuit ..................................................................... 4.9.3 Lifting motor control circuit ................................................................ U333................................................................................................................. 4.10.1 U333f ................................................................................................ 4.10.1.1 U333f board view............................................................... 4.10.1.2 U333f board layout ............................................................ 4.10.1.3 U333f schematic power ON/OFF....................................... 4.10.1.4 U333f schematic lift motor control ..................................... 4.10.1.5 U333f schematic power switching ..................................... 4.10.1.6 U333f schematic BG1 & BG2 ............................................ 4.10.2 U333g ............................................................................................... 4.10.2.1 U333g board view.............................................................. 4.10.2.2 U333g board layout ........................................................... 4.10.2.3 U333g schematic power ON/OFF...................................... 4.10.2.4 U333g schematic lift motor control .................................... 4.10.2.5 U333g schematic power switching .................................... 4.10.2.6 U333g schematic BG1 & BG2 ........................................... 4.10.3 U333 jumper settings ........................................................................ U80 (Horn)........................................................................................................ 4.11.1 U80 schematic .................................................................................. U468 (BG1) ...................................................................................................... 4.12.1 U468 layout ....................................................................................... 4.12.2 U468 schematic ................................................................................ U437 (VacuDAP Connection)........................................................................... 4.13.1 U437 schematic ................................................................................ 4.13.2 U437 function .................................................................................... U424 (BG2, Pump Switching) .......................................................................... 4.14.1 U424 schematic ................................................................................ 4.14.2 U424 function .................................................................................... Module 1 Fuses................................................................................................ 4.15.1 Transformer fuses ............................................................................. 4.15.2 Fuses on U333.................................................................................. 4.15.3 Fuses on chassis ..............................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-1 4-3 4-4 4-5 4-6 4-7 4-8 4-9 4-10 4-10 4-11 4-11 4-13 4-13 4-13 4-14 4-15 4-16 4-17 4-18 4-19 4-19 4-20 4-21 4-22 4-23 4-24 4-25 4-26 4-26 4-27 4-27 4-28 4-29 4-29 4-29 4-30 4-30 4-30 4-31 4-31 4-31 4-31
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4
Power Supply
4.1
Trafoblech View
Fig. 4-1
View of Trafoblech
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-1
4 Power Supply
WARNING The generator power supply in module 1 contains a large storage capacitor that filters the line voltage. The capacitor may contain high voltage even after the unit has been disconnected from the mains supply.
WARNING Module 1 (power supply) contains dangerous voltages. During operation the voltages exceed 250V DC.
4-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.2
Module 1 Assembly
Fig. 4-2
Module 1 assembly
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-3
4 Power Supply
4.3
Block Diagram
Fig. 4-3
Module 1 block diagram
4-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.4
U333 Board versions
In the following; two different versions of the U333 board are described. In the board version "f" both the lift motor control and the lift motor power are connected to the same rectifier "GL1" In the board version "g" only the lift motor power is connected to the output of rectifier "GL1". The lift motor control is supplied via two additional diodes "D16" and "D18" connected to the input of the rectifier "GL1" by an additional cable. (→ Ch. 4.8, p. 4-9) This cable ia available as part No.29265. NOTE If an older board U333 "f" or below has to be replaced by a U333 "g" board, the cable between "GL1" input and U333g connector "X27" has to be installed also.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-5
4 Power Supply
4.5
Wiring Diagram with U333f
Fig. 4-4
Module 1 wiring diagram with U333f
4-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.6
Wiring Diagram with U333g
Fig. 4-5
Module 1 wiring diagram with U333g
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-7
4 Power Supply
4.7
Functional Diagram with U333f
Fig. 4-6
Module 1 functional diagram with U333f
4-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.8
Functional Diagram with U333g
Fig. 4-7
Module 1 functional diagram with U333g
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-9
4 Power Supply
4.9
Circuit Description
4.9.1
Power-up circuit
Module 001 is supplied with mains voltage (230 VAC or 110 VAC) via terminal X1 (→ Fig. 4-10, p. 4-15). Transformer TR1 is constantly powered via mains fuses F1, F2, F3 and F4 (→ Fig. 4-10, p. 4-15) and back-up fuse F7. It thus keeps capacitor C1 constantly charged via secondary fuse F8 and diode D7. Module 001 is switched on by connecting the NETZEIN and EAGND signals (→ Fig. 4-10, p. 4-15), either via plug X17 (→ Fig. 4-11, p. 4-16) on the FST keyboard or via plug X11 (→ Fig. 4-11, p. 4-16) on the MW keyboard. Module 001 is switched off by connecting the NETZAUS and EAGND signals (→ Fig. 4-10, p. 4-15), either via plug X17 (→ Fig. 4-11, p. 4-16) on the FST keyboard or via plug X11 (→ Fig. 4-11, p. 4-16) on the MW keyboard. After switching on, relay K4 (→ Fig. 4-10, p. 4-15) picks up and latches with diode D1. The Power On LED (on the FST keyboard) is powered via resistor R5. Relay K1 (cut-off relay with 3 mm contact gap) picks up, and transformer TR1 (on the transformer panel) is switched to the power supply via fuse F5 and soft start resistor RL (on the transformer panel). Once electrolytic capacitor C2 has charged via the secondary winding (24 V, 3 A) (protected by fuse F3 on TR1) of transformer TR1 and rectifier GL3 (both on the transformer panel), and connected via plug X24 (→ Fig. 4-12, p. 4-17) and resistor R7, relay K2 short-circuits with the closing relay RL (contacts → Fig. 4-10, p. 4-15). Relay K3 (→ Fig. 4-12, p. 4-17) ensures that capacitors C and C2 are always discharged safely via resistors R6 and R3 during power off. The secondary winding (24 V, 5 A) (protected by fuse F2 on TR1) of transformer TR1 powers the lifting motor switching circuit via rectifier GL1 (both on the transformer panel). The secondary winding (24 V, 3 A) of transformer TR1 supplies module U326 (generator adaptation) in the 19" rack via rectifier GL3 (both on the transformer panel), plug X24 (→ Fig. 4-12, p. 4-17) and the central plug (on the transformer panel). The secondary winding (230 VAC) of transformer TR1 (on the transformer panel) supplies transformer TR2 (protected by fuse F1 on TR2 on the transformer panel). It also supplies plugs X3, X4 and X5 (→ Fig. 4-10, p. 4-15) via plug X7 (→ Fig. 4-10, p. 4-15) and fuse F6. The “Compact unit” section is connected to plug X5. The four 9 V secondary voltages (protected by fuses F5 – F8 on TR2) from transformer TR2 are supplied to module U334 (power unit for Module 2 via the central plug. The 24 V, 0.5 A secondary voltage (protected by fuse F9 on TR2) is used to rotate the image of the monitors (in the "Compact unit" section), while the 24 V, 4 A secondary voltage (protected by fuse F4 on TR2) powers modules BG1(18 V) and BG2 (coolant pump) via rectifier GL4 (on the transformer panel) and plug X12 (→ Fig. 4-13, p. 4-18). The 18 VDC voltage generated by module BG1 supplies Module 3, Module 4, the VacuDAP dose meter electronics and the signal horn (U80) via plug X19 and the central plug. The 18 VDC voltage also supplies the mobile stand keyboard via plug X17 (→ Fig. 4-11, p. 4-16).
4-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
The screening between the primary winding and the secondary windings is connected to PE potential on all three transformers. (TR1 and TR2 on the transformer panel and TR1 on U333). During power off, the excitation voltage is disconnected from relay K4 (→ Fig. 4-10, p. 4-15). Relay K4 then drops out, followed by cut-off relay K1 (→ Fig. 4-10, p. 4-15). Diodes D2 and D3 are used to isolate the on/off circuits in the mobile C-arm stand and monitor cart.
4.9.2
Output switching circuit
Secondary winding 1 of transformer TR1 (on the transformer panel) supplies the operating voltage for generating high voltage for the generator. The three tapping points 90 V/120 V or 136 V/ 195 V are linked to plug X9 (→ Fig. 4-12, p. 4-17). The /POWER1 signal at plug X11 (→ Fig. 4-11, p. 4-16) activates relay K5 (→ Fig. 4-12, p. 4-17), while the /POWER2 signal at plug X11 (→ Fig. 4-11, p. 4-16) activates relay K6 (→ Fig. 4-12, p. 4-17). The selected voltage is supplied to capacitor C (on the transformer panel) via rectifier GL2 (on the transformer panel) and plug X14 (→ Fig. 4-12, p. 4-17). Relay K7 is activated via transistor T1 when the /CANSTÖR signal at plug X10 and X13 (→ Fig. 4-12, p. 4-17) is “HIGH” (default). The capacitor voltage is then connected through to the generator via plug X15 and X16. The CAN signals are routed from the central plug (on the transformer panel), via plug X10 and X13 (→ Fig. 4-12, p. 4-17) to the FST keyboard. Resistor R6 is used to quickly discharge capacitor C (on the transformer panel) when the device is switched off.
4.9.3
Lifting motor control circuit
This part of the circuit is supplied by secondary winding 2 of transformer TR1 (on the transformer panel). This voltage is routed from X23 (→ Fig. 4-11, p. 4-16) via GL1 (on the transformer panel). Resistors R01 and R02 (on the transformer panel) are also connected to this plug, and serve to limit the lifting motor operating and braking currents. The lifting motor is connected to plugs X21 and X22 (→ Fig. 4-11, p. 4-16) and is short-circuited via the relay contacts of K8 and K9 when set to Off. The voltage regulator IC5 is supplied via diode D16, resistor R28 and Zener diode D17. It generates the supply voltage for the electronic circuit. The “Lifting motor up” and “Lifting motor down” keys on the FST keyboard are linked to PCB U333 via plug X17. The limit switches connected to plug X18 limit the traveling range of the lifting motor. When they reach their end position, they open the contact, and the lifting motor cannot travel any further in this direction. The Schmitt triggers IC1D and IC1E are linked to AND gates IC2B and IC2C and prevent the motor running if the “Lifting motor up” and “Lifting motor down” keys are pressed at the same time.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-11
4 Power Supply
Starting the lifting motor: The travel command (jump to “HIGH”) is routed to resistor R20 and R21 and to AND gate IC2B or IC2C via resistor R19 or R22. Capacitors C12 and C13 debounce the control keys. When the travel command is reversed by IC3C (or IC3F), capacitor C11 quickly discharges via diode D14 and resistor R24 (or C3 via D13 and R8). The Schmitt trigger IC3A (or IC1B) activates transistor T3 (or T2), and relay K9 (or K8) picks up, thus selecting the direction of travel. The travel command changes to a jump to “LOW” at the output from AND gate IC2A. Capacitor C14 is slowly discharged via resistor R17. The Schmitt trigger IC3F then activates transistor T4 and relay K10 picks up. The lifting motor is powered with voltage and moves down. Stopping the lifting motor: The stop command (issued by releasing the “Lifting motor up” or “Lifting motor down” key) discharges capacitor C12 or C13. Capacitor C14 is also quickly discharged via diode D11 and resistor R18. Relay K10 thus drops out immediately and disconnects the lifting motor from the supply voltage. Capacitor C11 or C3, on the other hand, is slowly charged via resistor R23 or R9. Relay K8 or K9 drops out after a delay, and the lifting motor is short-circuited once more. Monoflop IC4A responds to the falling edge at pin 1. The monoflop issues the “LOW” signal for approximately 200 ms (set using resistor R26 and capacitor C17). This prevents the lifting motor from starting up again after a stop command for the duration of the set time constant.
4-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10
U333
4.10.1
U333f
4.10.1.1
U333f board view
Fig. 4-8
U333f board view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-13
4 Power Supply
4.10.1.2
U333f board layout
Fig. 4-9
U333f board layout
4-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10.1.3
U333f schematic power ON/OFF
Fig. 4-10
U333f schematic Power ON/OFF
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-15
4 Power Supply
4.10.1.4
U333f schematic lift motor control
Fig. 4-11
U333f schematic lift motor control
4-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10.1.5
U333f schematic power switching
Fig. 4-12
U333f schematic power switching
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-17
4 Power Supply
4.10.1.6
U333f schematic BG1 & BG2
Fig. 4-13
U333f schematic BG1 & BG2
4-18
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10.2
U333g
4.10.2.1
U333g board view
Fig. 4-14
U333g board view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-19
4 Power Supply
4.10.2.2
U333g board layout
Fig. 4-15
U333g board layout
4-20
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10.2.3
U333g schematic power ON/OFF
Fig. 4-16
U333g schematic Power ON/OFF
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-21
4 Power Supply
4.10.2.4
U333g schematic lift motor control
Fig. 4-17
U333g schematic lift motor control
4-22
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10.2.5
U333g schematic power switching
Fig. 4-18
U333g schematic power switching
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-23
4 Power Supply
4.10.2.6
U333g schematic BG1 & BG2
Fig. 4-19
U333g schematic BG1 & BG2
4-24
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.10.3
U333 jumper settings
Jumpers are factory wired. Jumper
Function
Default Setting
BR1
bridge in circuit of /NETZ EIN
closed
BR2
bridge in circuit of /NETZ-EIN between X11 and X17
closed
BR3
connects CAN shield to CAN +5 V
open
BR4
connects CAN shield to CAN GND
closed
BR5
input 240 VAC for TR1
open
BR6
input 230 VAC for TR1
closed
BR7
Connects GL4+ to BG2
closed if BG2 is equipped
BR8
Connects GL4- to BG2
closed if BG2 is equipped
BR1, BR2 BR5, BR6
BR3, BR4
BR7 BR8
Fig. 4-20
U333 jumper location
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-25
4 Power Supply
4.11
U80 (Horn)
This part is supplied via X1 with 18 VDC. As soon as the signal at X1 gets “HIGH”, the transistor T2 gets conductive. Then the oscillator formed by buzzer H1, transistor T1, resistor R1 and condenser C1 is activated and the horn sounds. NOTE For service, the connector X1 can be removed temporarily in order to turn off the alarm sound.
4.11.1
U80 schematic
Fig. 4-21
U80 schematic
4-26
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.12
U468 (BG1)
4.12.1
U468 layout
Fig. 4-22
U468 layout
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-27
4 Power Supply
4.12.2
U468 schematic
Fig. 4-23
U468 schematic
4-28
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.13
U437 (VacuDAP Connection)
4.13.1
U437 schematic
Fig. 4-24
U437 schematic
4.13.2
U437 function
The U437 board is used in order to connect the VacuDAP meter electronics to the connecting cable located in the module 1. Additionally the board provides a calibration function. By setting the jumper to the service setting, the DAP meter can be adjusted using the S1 - and S2 + keys.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-29
4 Power Supply
4.14
U424 (BG2, Pump Switching)
4.14.1
U424 schematic
Fig. 4-25
U424 schematic
4.14.2
U424 function
The U424 board is optionally equipped in units with active cooling. It switches the pump supply voltage on and off controlled by a signal coming from Module 2 via the main connector.
4-30
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4 Power Supply
4.15
Module 1 Fuses
4.15.1
Transformer fuses
Fuse
Value
Transformer TR1
Part No.
Function
11120
F2
10 A slow (5×20mm)
17007
Supply lifting motor
F3
10 A slow (5×20mm)
17007
Supply U326 and U333 control relay
Transformer TR2
11081
F1
12 A slow (5×20mm)
17006
Main fuse TR2
F4
4 A slow (5×20mm)
17021
Supply BG1 (18 V) in U333
F5
4 A slow (5×20mm)
17021
Supply for ± 5 V Digital
F6
1 A slow (5×20mm)
17019
Supply for +5 V ISO (CAN)
F7
1 A slow (5×20mm)
17019
Supply for +5 V Video
F8
1 A slow (5×20mm)
17019
Supply for -5 V Video
F9
1 A slow (5×20mm)
17019
Supply for monitor rotation (Compact)
4.15.2
Fuses on U333
Fuse
Value
Part No.
Function
F1
15 A medium (6.3×32mm)
17020
Mains input fuse 230 V
F4
15 A medium (6.3×32mm)
17020
Mains input fuse 230 V
F5
1 A slow (5×20mm)
17019
RL resistor
F6
3.15 A slow (5×20mm)
17056
230 V out for Compact unit
F7
100 mA slow (5×20mm)
17056
Primary transformer 1
F8
250 mA slow (5×20mm)
17001
Secondary transformer 1
F1 on U424 (BG2)
2.5 A slow (5×20mm)
17035
Water pump
4.15.3
Fuses on chassis
The following fuses are located behind the U333 board. Fuse
Value
Part No.
Function
F2
20 A medium (6.3×32mm)
17044
Mains input fuse 120 V
F3
20 A medium (6.3×32mm)
17044
Mains input fuse 120 V
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
4-31
4 Power Supply
4-32
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 5.1 5.2 5.3 5.4
5.5 5.6 5.7 5.8
5.9
Module 2 19" Rack Description ....................................................................... 5.1.1 General ............................................................................................. 5.1.2 Module view ...................................................................................... Module 2 Overview........................................................................................... Module 2 Inputs and Outputs ........................................................................... Block Diagrams ................................................................................................ 5.4.1 U325 block diagram .......................................................................... 5.4.2 U325 functional description............................................................... 5.4.3 U326 block diagram .......................................................................... 5.4.4 U326 functional description............................................................... 5.4.5 U328 block diagram .......................................................................... 5.4.6 U328 functional description............................................................... 5.4.7 U334 block diagram .......................................................................... 5.4.8 U334 functional description............................................................... U325 and U326 Front Panels........................................................................... Exchange of boards ......................................................................................... 5.6.1 Exchange of U325 ............................................................................ 5.6.2 Exchange of U326 ............................................................................ Hardware Radiation Logic ................................................................................ U326 Schematic Diagrams............................................................................... 5.8.1 U326 schematics 1 ........................................................................... 5.8.2 U326 schematics 2 ........................................................................... 5.8.3 U326 schematics 3 ........................................................................... 5.8.4 U326 schematics 4 ........................................................................... 5.8.5 U325 schematics 1 ........................................................................... 5.8.6 U325 schematics 2 ........................................................................... 5.8.7 U325 schematics 3 ........................................................................... 5.8.8 U325 schematics 4 ........................................................................... 5.8.9 U325 schematics 5 ........................................................................... 5.8.10 U325 schematics 6 ........................................................................... 5.8.11 U325 schematics 7 ........................................................................... U501 Video Processing Board ......................................................................... 5.9.1 U501 Block Diagram ......................................................................... 5.9.2 Connectors on the front cover........................................................... 5.9.3 LEDs on the front cover ....................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-1 5-1 5-1 5-2 5-3 5-4 5-4 5-5 5-6 5-7 5-8 5-9 5-10 5-11 5-12 5-13 5-13 5-14 5-16 5-17 5-17 5-18 5-19 5-20 5-21 5-22 5-23 5-24 5-25 5-26 5-27 5-28 5-28 5-29 5-29
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5
Module 2
5.1
Module 2 19" Rack Description
5.1.1
General
The 19" rack combines several functions: − C-arm control: U325 − Generator adaption: U326 − Video & Dose control: U501 (optionally)
5.1.2
Module view
Fig. 5-1
Module 2 view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-1
5 Module 2
5.2
Module 2 Overview
Fig. 5-1
Module 2 overview
5-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.3
Module 2 Inputs and Outputs
Fig. 5-2
Module 2 inputs and outputs
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-3
5 Module 2
5.4
Block Diagrams
5.4.1
U325 block diagram
Fig. 5-3
U325 block diagram
5-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.4.2
U325 functional description
The slide-in board U325 is the central control unit of the C-arm. All machine data are stored on this board. Machine data include information about: − options − start up configuration − diaphragm settings in case of standard collimator − kV - mA characteristics − hand and foot switch configuration. The board receives data via the CAN bus from module 3, calculates the kV and mA nominal values and forwards them as data words via the internal LUCY bus to the U326 board. Also the information for diaphragm setting and laser switching are transmitted that way. The hand and foot switches are monitored for short circuits. The real time clock counts up the radiation time and is used to create an error log. The safety logic prevents from unintended radiation. There is a detailed description in → Ch. 5.7, p. 5-16 “Hardware Radiation Logic” The monitor cart detection controls the CAN bus termination logic on board U334. The power switching controls the relays K5 and K6 on the board U333.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-5
5 Module 2
5.4.3
U326 block diagram
Fig. 5-4
U326 block diagram
5-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.4.4
U326 functional description
The slide-in board U326 is the control board for the generator. Here the data words coming from the U325 board are converted to analog values and discrete switching signals: − 12 bit words are converted to mA and kV analog control values − Slot and Iris collimator controls are decoded and sent to the collimators via drivers. (only used for standard collimator) − Laser switch command is decoded and sent to the generator − radiography mode command is decoded and sent to the generator. The mA nominal value is compared to the actual value measured in the generator. The difference controls a switching regulator, which generates the filament voltage from 24 V input voltage. A temperature switching signal (0.5 mA) is received from the generator. This signal is encoded on the U326 board and sent to the U325 board in order to control the system depending on the generator temperature. The fault line coming from the generator is received. This signal is monitored by the fault LED on the U326 board and is connected to the CAN fault line by an optocoupler. Also 15 V and ±10 V for the board supply are generated. The voltages are monitored on the front of the board.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-7
5 Module 2
5.4.5
U328 block diagram
Fig. 5-5
U328 block diagram
5-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.4.6
U328 functional description
Board U328 is the motherboard of the 19" rack. It links the boards U325, U326, U327, and U334. It provides a connection to module 1. The optocouplers isolate module 1 from module 2. The serial number of the system is permanently stored in the GAL chip. The number can be read using the Exposcop Studio (→ Ch. 18.1.5, p. 18-10).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-9
5 Module 2
5.4.7
U334 block diagram
Fig. 5-6
U334 block diagram
5-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.4.8
U334 functional description
Board U334 is the power supply for module 2. It is supplied by transformer 2 with AC voltages. The voltages are rectified and stabilized. Green LEDs on the front indicate the voltages. The CAN bus termination logic is also located on the U334 board.
Fig. 5-7
U334 front view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-11
5 Module 2
5.5
U325 and U326 Front Panels
Fig. 5-8
U325 and U326 front panels
5-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.6
Exchange of boards
5.6.1
Exchange of U325
After exchange the U325 board has to be programmed. There are two steps of programming. First, if the warning E35 appears, the software versions of the board and the rest of the system have to be adapted. Then the machine data have to be restored. It is recommended to save the machine data before you remove the original board, if communication to the board is possible. If the data cannot be read out and no archived data are available, use default data and set options manually. If there is a warning E32 after exchange, the RAM test pattern must be set. If there is a warning E34 after exchange, the Clock must be set. → Ch. 18.1.4.1, p. 18-6 and → Ch. 18, p. 18-1 Software Tools U325 to be changed
stored machine data avaliable
no
comunication to U325 possible
no
Use default data and set options manually
yes yes
save machine data change U325
Start C-Arm
E 35?
yes
adapt software versions
no
restore machine data
restart
E 32?
yes
Set RAM test pattern
yes
Set Clock
no
E 34?
no
restart
finished
Fig. 5-9
U325 exchange procedure
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-13
5 Module 2
5.6.2
Exchange of U326
After exchange the U326 board has to be adjusted according to the following instructions:
1 General: C-arm has to run properly before adjustments can be made. The potentiometers and their labels at the front panel (U326): − P5 mA offset − P4 pre-heating − P3 mA DAC offset − P2 exposure − P1 kV offset Necessary test equipment: Test cable to generator (to check the actual generator values) Test cable to the front connector of U326 (for the signal names refer to the U326 schematic) PC with USBCAN or PARCAN test set.
2 Adjust filament voltage without exposure • Set fluoroscopy mode, connect test cable to the front connector of U326 Connect voltage meter to: + with UH - with GND Adjust potentiometer P4 to 12.0 V (P4 = pre-heating)
3 Adjust mA offset with exposure • Set fluoroscopy mode, set kV manual and 40 kV Connect ampere meter instead of the mA bridge in the generator. Adjust potentiometer P5 to 0.2 mA (P5 = mA Offset) • Measure the voltage UH with exposure and keep the voltage value in mind. Adjust pre-heating to memorized value with potentiometer P4 without exposure.
4 Adjust kV with exposure • Set fluoroscopy program extremities, set kV manual and 80 kV Connect voltage meter to + with kV measuring point “f” – with generator ground Adjust potentiometer P1 to 8.0 V (1 V equals 10 kV) Check voltages at 40, 60, 100, 110 kV
5 Adjust mA with exposure, top value Set spine / pelvis fluoroscopy program (program 2), set kV manual and 65 kV (mA display has to indicate 6.0 mA) Connect ampere-meter instead of the mA bridge in the generator. Adjust potentiometer P3 to 5.95 mA
6 Adjust mA with exposure, bottom value • Set spine / pelvis fluoroscopy program (program 2), set kV manual and 40 kV Connect ampere-meter instead of the mA bridge in the generator. Adjust potentiometer P5 to 0.2 mA !! Repeat the two steps above, because these steps interact with each other !!
5-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
7 Current rise time in active radiography mode • Set radiography mode, set radiography exposure time to 0.1sec • NOTE: the exposure starts approximately 1 sec. after pressing the hand switch (pre-heating time) • Connect the mAs meter instead of the mA bridge in the generator. Adjust potentiometer P2 to 2.0 mAs
8 Adjust radiography current with exposure • Set radiography mode, 80 kV and 4 sec. radiography time. • NOTE: the exposure starts approximately 1 sec. after pressing the hand switch (pre-heating time) Connect ampere meter instead of the mA bridge in the generator. Adjust the current to 20.0 mA which means you have to enter the appropriate number in the field “D/A Converter Radiography” of the Exposcop Studio software. The default value is 20 (20 mA). Allowed value range is from 15 (15 mA) to 28 (28 mA). ! If the value is out of range there is a hardware failure ! For adjustment use the “DAC value” entry of the above menu. Allowed value range here is from 614 to 1147.
9 Adjust radiography mAs with exposure • Set radiography mode, 80 kV and radiography exposure time to 0.5 s • Do not forget: the exposure starts approximately 1 s after pressing the hand switch Connect mAs meter instead of the mA bridge in the generator. Adjust potentiometer P2 to 10.0 mAs.
10 Check pulse fluoroscopy • mA values identical to fluoroscopy.
11 Check snapshot mA • This value has to be set to 8 mA. This is done using the Exposcop Studio software. • Section Generator, D/A converter snapshot.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-15
5 Module 2
5.7
Hardware Radiation Logic
Fig. 5-10
Hardware radiation logic
5-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.8
U326 Schematic Diagrams
5.8.1
U326 schematics 1
Fig. 5-11
U326 schematics 1
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-17
5 Module 2
5.8.2
U326 schematics 2
Fig. 5-12
U326 schematics 2
5-18
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.8.3
U326 schematics 3
Fig. 5-13
U326 schematics 3
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-19
5 Module 2
5.8.4
U326 schematics 4
Fig. 5-14
U326 schematics 4
5-20
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.8.5
U325 schematics 1
Fig. 5-15
U325 schematics 1
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-21
5 Module 2
5.8.6
U325 schematics 2
Fig. 5-16
U325 schematics 2
5-22
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.8.7
U325 schematics 3
Fig. 5-17
U325 schematics 3
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-23
5 Module 2
5.8.8
U325 schematics 4
Fig. 5-18
U325 schematics 4
5-24
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.8.9
U325 schematics 5
Fig. 5-19
U325 schematics 5
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-25
5 Module 2
5.8.10
U325 schematics 6
Fig. 5-20
U325 schematics 6
5-26
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.8.11
U325 schematics 7
Fig. 5-21
U325 schematics 7
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-27
Fig. 5-22
5-28
U306
U180
Blockdiagram U501 (former U180 U321 U306)
Module 2
Image Memory
R. Frank R. Frank .
Drawn: Release: Changed:
20. Sept. 2007 File Name : Page :
.
. of
.
18V= input
8000/Vista Family
Unit:
outputs to monitors
Power Supply
to I.I.
to CCD camera
AGC to CCD camera
BSB_U501_02.cdr
+18V
01. Feb. 2006 Drawing No.:
-9V
-5V
+5V
+15V
DAC
VIDEO
magn. buffers
Iris control
DAC
to ADC
U501 Block Diagram
ADC
Lattice
LUCY U300
ADC
+
output
Weitergabe sowie Vervielfaeltigung dieser Unterlage, Verwertung und Mitteilung ihres Inhalts nicht gestattet. Zuwiderhandlungen verpflichten zu Schadenersatz.
5.9.1
VIDEO
test
I/O
DAC
measurement circles
U501
Alle Rechte vorbehalten. Technische Aenderungen vorbehalten
U501 Video Processing Board
from above
CAN
Sync
log
All rights reserved. Subject to change without prior notice.
5.9
U321
CCD camera input
Transmission as well as reproduction of this document, exploitation or disclosure of its contents is not permitted. Infringements shall entitle to damage claims.
5 Module 2
U501 Block Diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5 Module 2
5.9.2
Connectors on the front cover
X7
Video input from CCD camera
(SMB)
X12
AGC output to CCD camera
(SMB)
X9
Iris interface
(5-pin Panduit)
X17
I.I. interface
(5-pin Panduit)
X3
Video output from U321
(SMB)
X2
Video output from U321, filtered
(SMB)
X90
Video output from U306
(SMB)
X100
Video output from U306
(SMB)
5.9.3
LEDs on the front cover
LED 45
ERROR
(rd)
LED 35A
+15 V
(gn)
LED 40A
+5 V
(gn)
LED 41A
+3.3 V
(gn)
LED 40B
-5 V
(gn)
LED 41B
-9 V
(gn)
LED 35B
-15 V
(gn)
CAUTION Dose and video adjustments have to be performed if a Module 3 has been replaced by a U501.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
5-29
5 Module 2
5-30
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6 Module 3
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 6.1 6.2 6.3
6.4
Module 3 Memory Unit .................................................................................... 6.1.1 General ............................................................................................. 6.1.2 Module 3 view ................................................................................... Module 3 Overview........................................................................................... Block Diagrams ................................................................................................ 6.3.1 U180 block diagram .......................................................................... 6.3.2 U180 functional description............................................................... 6.3.3 U321 block diagram .......................................................................... 6.3.4 U321 functional description............................................................... 6.3.5 Module 3 wiring diagram ................................................................... Exchange of Module 3......................................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6-1 6-1 6-1 6-2 6-3 6-3 6-4 6-5 6-6 6-7 6-8
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6 Module 3
6
Module 3
6.1
Module 3 Memory Unit
6.1.1
General
Module 3 provides the following functions: − Dose regulation − Video regulation − Image memory in Compact units: U306 (optionally) − 100 Hz converter in Compact units with CRT monitor: U313 (optionally) NOTE Module 3 will be missing if system is equipped with a U501 board.
6.1.2
Module 3 view
Fig. 6-1
Module 3 view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6-1
6 Module 3
6.2
Module 3 Overview
Fig. 6-2
Module 3 overview
6-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6 Module 3
6.3
Block Diagrams
6.3.1
U180 block diagram
Fig. 6-3
U180 block diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6-3
6 Module 3
6.3.2
U180 functional description
There is a U300 processor board (LUCY) on the U180 board. Thus module 3 is part of the CAN bus. On the U300 board the anatomical program data are stored. Anatomical program data contain values for: − dose and video adjustment − diameter of the gauging circle − regulation mode. Values are stored independently for every anatomical mode and magnifier size. The U180 transmits values for Iris and AGC to the CCD camera. The iris position is fed back to the analog inputs by a potentiometer. At the analog multiplexer there are also values gained from the video signal in order to do dose and video regulation. The average or peak value is compared to an internal value. The difference is transmitted by the CAN bus to the module 2 to regulate the necessary dose value. The black and white values are compared to values stored in the organ table. The difference controls the black clamping and white clipping (video adjustment, → Ch. 19.1.2, p. 19-3).
6-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6 Module 3
6.3.3
U321 block diagram
Fig. 6-4
U321 block diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6-5
6 Module 3
6.3.4
U321 functional description
There is a digital circle generator on the U321 board. This generates the gauging and blanking circles. Within those areas, the video signal is measured and the result is sent to the LUCY of the U180. The signal stages “curve” and “white level regulator” are controlled by the LUCY of the U180 board. In the signal mixer the video signal is overlaid by the blanking circle. Sync signals are separated from the video signal in order to synchronize the circle generator and additional hardware for Compact units. (Image memory, 100 Hz converter)
6-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6 Module 3
6.3.5
Module 3 wiring diagram
Fig. 6-5
Module 3 wiring diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6-7
6 Module 3
6.4
Exchange of Module 3
After exchange the module 3 has to be programmed. There are two steps of programming. First, if warning E35 appears, the software versions of the board and the rest of the system have to be adapted. Then the organ data have to be restored. It is recommended to save the organ data before you remove the original module 3, if communication to the board is possible. If the data cannot be read out and no archived data are available, use default data and adjust dose manually (→ Ch. 18.1.8, p. 18-14).
Fig. 6-6
6-8
Module 3 exchange procedure
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6 Module 3 To change the memory unit, the 4 nuts have to be removed → Fig. 6-7, p. 6-9. Now the module can be opened. Remove connectors X5, X7, X9, X17 on the board U180 as well as the coax connectors X7, X3 on U321and X12 on U 180 → Fig. 6-8, p. 6-9. The unit is fixed by two screws → Fig. 6-9, p. 6-9.
Nuts to be removed
Fig. 6-7
Fixation of Module 3 cover
Fig. 6-8
Module 3 connections
Fixation screws
Fig. 6-9
Module 3 fixation
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
6-9
6 Module 3
6-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier Module 4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 7.1 7.2 7.3
7.4
7.5 7.6
Parts Location .................................................................................................. Block Diagram .................................................................................................. 7.2.1 Functional description ....................................................................... I.I. power supply overview ................................................................................ 7.3.1 General ............................................................................................. 7.3.2 Main parts and their functions ........................................................... Hardware description ....................................................................................... 7.4.1 PCBs ................................................................................................. 7.4.2 U316 LEDs........................................................................................ 7.4.3 U316 connectors ............................................................................... 7.4.4 Connectors cascade ......................................................................... 7.4.4.1 Connectors cascade location............................................. 7.4.5 U316 schematics .............................................................................. 7.4.5.1 U316 schematics part 1 ..................................................... 7.4.5.2 U316 schematics part 2 ..................................................... Adjustments...................................................................................................... 7.5.1 Potentiometer location ...................................................................... CCD Camera.................................................................................................... 7.6.1 General description ........................................................................... 7.6.2 Setting the CCD camera distance..................................................... 7.6.3 Scattered light lens ........................................................................... 7.6.4 Changing the camera........................................................................ 7.6.4.1 Removing the camera........................................................ 7.6.5 Installing the CCD camera ................................................................ 7.6.6 Adjustment of the optical focus ......................................................... 7.6.7 Adjustment instruction for 23 cm and 30 cm tube ............................. 7.6.8 Size and resolution requirements...................................................... 7.6.8.1 Without patient equivalent filter.......................................... 7.6.8.2 With patient equivalent filter 25 mm Al ..............................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-1 7-2 7-3 7-4 7-4 7-4 7-5 7-5 7-5 7-6 7-6 7-6 7-7 7-7 7-8 7-9 7-9 7-9 7-10 7-10 7-11 7-12 7-12 7-13 7-14 7-15 7-16 7-16 7-16
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7
Image Intensifier
7.1
Parts Location
Counterweight CCD camera
II-Power supply
Fig. 7-1
Image intensifier parts location
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-1
7 Image Intensifier
7.2
Block Diagram
Fig. 7-1
Image intensifier block diagram
7-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.2.1
Functional description
The image intensifier can be divided into three electrical units: − The CCD sensor forms together with the iris collimator and the lens the CCD camera. − The image intensifier tube. − The intensifier power supply consisting of the boards U316, U317 and U318. The latter supplies the intensifier tube and the CCD camera. The U316 is supplied with 18 V. From this 18 V the supply voltages for the CCD camera, the grid voltages and the anode voltage are generated. The voltages for G2 (focus) and G3 (image size) are separately adjustable for every fluoroscopy format. The grid voltages are switched by signals coming from module 3. The anode voltage is factory pre-adjusted to 30 kV and must not be changed. (Tube might be damaged) The voltage for G1 is given by the tube manufacturer according to the tube type. It makes up the edge focusing.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-3
7 Image Intensifier
7.3
I.I. power supply overview
Fig. 7-2
Image intensifier power supply
7.3.1
General
The i.i. power supply supplies the necessary voltages for the different image intensifiers. The U316 also supplies power for the CCD camera, switchable between 6 V and 12 V.
7.3.2
Main parts and their functions
− Controller U316
Controls the magnification modes of the i.i.
− Power inverters U317
High voltages
− Cascade U318
Multiplies the voltages
7-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.4
Hardware description
7.4.1
PCBs
PCBs U316:
controller board
U317:
high voltage transformers
U318:
high voltage cascade
The i.i. power supply is divided into several boards to build a very compact module.
7.4.2
U316 LEDs
Status LEDs for Service Requirements LED 10 yellow:
H-Sync
LED 11 yellow:
indicates Magnifier 1
LED 12 yellow:
indicates Magnifier 2 D12 D11 Magnifier
D10 H-Sync
Fig. 7-3
U316 LEDs
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-5
7 Image Intensifier
7.4.3
U316 connectors
X11 to CCD Iris
X12 to Module3/U501
X9 to CCD Camera
Fig. 7-4
U316 connectors
7.4.4
Connectors cascade
Connectors
Range
Function
Anode
30 kV factory set
Anode voltage
Grid 1
approx. 100 V – 300 V
Edge focus (depends on tube type)
Grid 2
approx. 0.5 kV – 2 kV
Focus adjustment
Grid 3
approx. 2 kV – 12 kV
Size adjustment
Grid 4
approx. 2.5 kV
Not adjustable
Ground
7.4.4.1
Connectors cascade location
Fig. 7-5
Connectors cascade location
7-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.4.5
U316 schematics
7.4.5.1
U316 schematics part 1
Fig. 7-6
U316 schematics 1
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-7
7 Image Intensifier
7.4.5.2
U316 schematics part 2
Fig. 7-7
U316 schematics 2
7-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.5
Adjustments
7.5.1
Potentiometer location P1 Anode P2, P3, P4 Size M2, M1, M0 P5, P6, P7 focus M2, M1, M0
Fig. 7-8
U316 potentiometer location
7.6
CCD Camera
Fig. 7-9
CCD camera without cover
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-9
7 Image Intensifier
7.6.1
General description
The CCD sensor is a solid-state imaging device that produces a black and white image. Light entering wells in the sensor are converted to electrons, which are read out from the sensor, amplified and converted to a digital video signal. The video signal is transmitted via an analog video coax cable to the image processing system.
7.6.2
Setting the CCD camera distance
The camera distance is set by using different mounting rings. The camera will be shipped without a mounting ring. For installing the camera, the ring has to be taken from the camera that has been installed before. II Tube
D [mm]
Scattered light lens assembled
17 cm HP
16.1
Yes
23 cm HP
13.6
Yes
23 cm HX
20.4
No
30 cm HX
23.5
No
I
Mounting ring
Fig. 7-10
7-10
Mounting ring for distance adjustment
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.6.3
Scattered light lens
Before the camera is mounted onto the image intensifier, check whether the scattered light lens is required or not.
Fig. 7-11
Scattered light lens location
Fig. 7-12
Scattered light lens removed
Check the type of the tube: For a HX tube type, the lens has to be removed. For a HP tube type the lens has to be installed.
Fig. 7-13
II Tube Types: HP type (left), HX Type (right)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-11
7 Image Intensifier
7.6.4
Changing the camera
7.6.4.1
Removing the camera
Assembly and disassembly of the camera has to be done only at power down. • Remove cover of the intensifier housing. • In order to remove the camera, the led weight has to be removed first. Thereto the 3 nuts and the plate have to be removed (→ Fig. 7-14, p. 7-12). • Then the connectors X9 and X11 on the U316 have to be removed and the coax connectors have to be separated (→ Fig. 7-15, p. 7-12). • To remove the camera, three Allen screws have to be removed. Do not loosen the clamps. NOTE For the 12" tube, a 7/64" Allen key is required. For all other tubes, a 3 mm Allen key is required.
Remove those screws Clamps
Remove nuts Remove weight
Fig. 7-14
Camera removal
X9 Camera connector
X11 Iris connector Coax connectors AGC and video Fig. 7-15
7-12
Camera connections
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.6.5
Installing the CCD camera
Fig. 7-16
Mounting of the CCD camera
The camera is put onto the output window and is fixed by the mounting ring using 3 allen screws. If the camera has not been removed from the mounting ring, the assembly is done in reversed order. Please check optical focus afterwards. If the camera has to be changed, it has to be mounted and centered to the ring that has been assembled to the unit. After installing, the optical focus has to be set. NOTE For the 12" tube, a 7/64" Allen key is required. For all other tubes, a 3 mm Allen key is required.
NOTE The connections have to face towards the C profile, in order to provide correct image orientation.
CAUTION Make sure that you have selected the correct supply voltage at the i.i. power supply, U316, J1 (→ Ch. , p. 7-8).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-13
7 Image Intensifier
7.6.6
Adjustment of the optical focus
• Switch off the unit. • To focus the camera lens, the camera cover has to be removed. • For that, remove the intensifier power supply (only the screws, not the connectors) and loosen the grub screws at the camera housing (→ Fig. 7-17, p. 7-14). • Switch on. • Center a line pair phantom to the input window. The line pairs should be 45° to the monitor line structure and 90° to the scattered radiation grid • Loosen the locking screw. Set to magnifier 2 format and adjust to maximum resolution. Use a 2 mm nib (Part No. 91118) for focus adjustment. • Switch to full field size and check edge focus. Tighten locking screw hand screwed. • Put on camera cover and lock it by the grub screws. • Assemble i.i. supply, led weight and i.i. cover (incl. earth cable)
WARNING Focus adjustment is done while screening. Take care for radiation protection!
Locking screw for focusing ring
Grub screw for camera cover Focusing ring with 2mm holes Fig. 7-17
7-14
CCD camera focus adjustment
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7 Image Intensifier
7.6.7
Adjustment instruction for 23 cm and 30 cm tube
Fig. 7-18
I.I. power supply adjustment CAUTION Target values are specified according to non interlaced systems. Refer to tables → Ch. 7.6.8, p. 7-16 for corresponding values of interlaced systems.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
7-15
7 Image Intensifier
7.6.8
Size and resolution requirements
7.6.8.1
Without patient equivalent filter
Extremities mode, automatic dose II Type
15 cm
23 cm
30 cm
N
M1
N
M1
M2
N
M1
M2
Size [cm]
14.0
10.0
20.1
15.1
11.1
27.1
20.1
15.1
Resolution [lp/mm] Non Interlaced
2.0
2.5
1.6
2.0
2.5
1.2
1.6
2.0
Resolution [lp/mm] Interlaced
1.7
1.9
1.4
1.8
2.0
1.0
1.4
1.8
7.6.8.2
With patient equivalent filter 25 mm Al
Extremities mode, automatic dose II Type
15 cm
23 cm
30 cm
N
M1
N
M1
M2
N
M1
M2
Size [cm]
14.0
10.0
20.1
15.1
11.1
27.1
20.1
15.1
Resolution [lp/mm] Non Interlaced
1.8
2.2
1.4
1.8
2.2
1.0
1.4
1.8
Resolution [lp/mm] Interlaced
1.5
1.7
1.2
1.6
2.8
0.8
1.2
1.6
NOTE Size and resolution adjustments of the i.i. tube are interacting. Do the adjustment reciprocally.
NOTE Interlaced: ZIEHM 8000, ZIEHM Compact Non Interlaced: ZIEHM Vista To get information about interlaced / non interlaced, display image system information → Ch. 15.12, p. 15-30.
7-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator Module 5
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 8.1 8.2
8.3 8.4 8.5 8.6
8.7 8.8
Overview .......................................................................................................... Block Diagrams ................................................................................................ 8.2.1 Generator block diagram................................................................... 8.2.2 Generator detailed block diagram ..................................................... Interface Generator – C-Arm Stand ................................................................. Block Diagram Generator – Module 1 / Module 2 ............................................ General PCBs .................................................................................................. Checklist for the Generator .............................................................................. 8.6.1 U185 test point location .................................................................... 8.6.2 U185 schematics .............................................................................. 8.6.3 U185 board layout............................................................................. Measuring kV and mA ...................................................................................... Laser (option) ................................................................................................... 8.8.1 Generator with laser.......................................................................... 8.8.2 Image intensifier with laser................................................................ 8.8.3 Laser block diagram.......................................................................... 8.8.4 U160 laser control board...................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-1 8-2 8-2 8-3 8-4 8-5 8-6 8-7 8-8 8-9 8-10 8-11 8-13 8-13 8-14 8-15 8-16
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8
Generator
8.1
Overview
Fig. 8-1
Generator assembly
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-1
8 Generator
8.2
Block Diagrams
8.2.1
Generator block diagram
Fig. 8-2
Generator block diagram
8-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.2.2
Generator detailed block diagram
Fig. 8-3
Generator detailed block diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-3
8 Generator
8.3
Interface Generator – C-Arm Stand
Fig. 8-4
Interface generator – C-arm stand
8-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.4
Block Diagram Generator – Module 1 / Module 2
Fig. 8-5
Block diagram generator – Module 1 / Module 2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-5
8 Generator
8.5
General PCBs
PCBs summary U40
Heating current control Current control Heating amplifier kV amplifier
U41
Motherboard
U42
3 stage kV amplification with push-pull controlling Pre-amplifier
U43
Fault and filament control mA measuring range mA control
U45
Slot and iris collimator control Temperature control
U54
8-6
kV control
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.6
Checklist for the Generator
The following steps show the signals required for the generator to work properly: Measurements are made on the U185 Filter frame. 1. U1 (final stage supply) must be in the range of approx. 130 to 280 VDC depending on the operating mode. (Generator power in W). This voltage has to be present in standby also. Check the voltages at each of the three stages. (see table below) Screen once before measuring. If E26 is present, the voltage is switched off by module1 for safety reasons. The test points are: plus minus
X11 Pin 2 X11 Pin 4
2. U2 (supply voltage) must be 15 VDC. The test points are: plus minus
X12 Pin 2 X12 Pin 1
3. UH (filament voltage) must be in a range of approx. 9.5 to 11 VDC in standby and approx. 9.5 to 18 VDC while screening depending on selected mA value. If the voltage gets below 7.5 VDC, the generator pulls the fault line (E26) If the voltage gets above 22 VDC for a certain time, the filament regulator on the U326 switches off and radiation will be interrupted. The test points are: plus minus
X11 Pin 3 X12 Pin 1
4. kV reference voltage must be in the range of 8.1 to 12 VDC depending on the selected kV value. Voltage has to be also present during standby. 8.1 VDC equals to 40 kV and 12 VDC equals to 110 kV; 55 mV is 1 kV step. The test points are: plus minus
BG2 Pin 4 (kV ref.) BG2 Pin 2 (kV ground)
5. Radiation command input is active if BG2 Pin 8 is about 0 V; inactive level is 15 V. The test points are: plus minus
BG2 Pin 8 (radiation) X12 Pin 1
Mode
AC Voltage at transformer
DC Voltage measured
at 40 kV
90 VRMS
127 V
120 VRMS
169 V
136 VRMS
192 V
195 VRMS
275 V
at 80 kV (depending on transformer type) Snapshot
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-7
8 Generator
8.6.1
U185 test point location
Fig. 8-6
U185 test point location
WARNING Module 5 (generator) contains dangerous voltages. During operation the voltages exceed 250 VDC.
WARNING Measurements partly have to be done while screening. Take care for radiation protection.
8-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.6.2
U185 schematics
Fig. 8-7
U185 schematics
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-9
8 Generator
8.6.3
U185 board layout
Fig. 8-8
U185 board layout
8-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.7
Measuring kV and mA
In order to measure mA, the tank connector is used. Before the mA meter is connected the jumper has to be removed.
Fig. 8-9
Tank connector with jumper (left) and without jumper (right)
Fig. 8-10
Test cable connected to tank connector CAUTION After measuring the jumper has to be set again.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-11
8 Generator
In order to measure kV the test points “f” and “GND” on the U54 board are used.
Test points “f” and “GND”
Fig. 8-11
kV and mA measuring points
The reading of the voltmeter is the actual kV divided by 10,000. For example 67 kV is measured as 6.7 V.
Fig. 8-12
Test cable connected to “f” and “GND” on U54 WARNING Module 5 (generator) contains dangerous voltages. During operation the voltages exceed 250 VDC.
WARNING Measurements have to be done while screening. Take care for radiation protection.
8-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.8
Laser (option)
8.8.1
Generator with laser
Fig. 8-13
Generator with laser
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-13
8 Generator
8.8.2
Image intensifier with laser
Fig. 8-14
Image intensifier with laser
8-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8 Generator
8.8.3
Laser block diagram
Fig. 8-15
Laser block diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
8-15
8 Generator
8.8.4
U160 laser control board
Fig. 8-16
U160 laser control board
8-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Vision Track Module 15
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 9.1 9.2 9.3
9.4
Virtual Collimator View ..................................................................................... Block Diagram .................................................................................................. U517d............................................................................................................... 9.3.1 U517d location of connectors, switches and LEDs ........................... 9.3.2 Switches............................................................................................ 9.3.2.1 DIP switch factory setting .................................................. 9.3.3 LEDs ................................................................................................. 9.3.4 Connectors........................................................................................ Changing the Collimator................................................................................... 9.4.1 Part exchange ................................................................................... 9.4.2 Mechanical centering ........................................................................ 9.4.3 Collimator size adjustment ................................................................ 9.4.4 Preview size adjustment ...................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-1 9-2 9-3 9-3 9-3 9-5 9-5 9-6 9-10 9-10 9-10 9-11 9-12
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
9
Virtual Collimator
9.1
Virtual Collimator View
Fig. 9-1
Virtual collimator
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-1
9 Virtual Collimator
9.2
Block Diagram
Fig. 9-2
Virtual collimator block diagram
9-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
9.3
U517d
9.3.1
U517d location of connectors, switches and LEDs
Fig. 9-3
U517d connectors, switches and LEDs
9.3.2
Switches
DIP Switch 2 Switch 1
Standalone mode for production requirements on == standalone active
Switch 2
Size Bit 1
Switch 3
Size Bit 2
Switch 4
Size Bit 3
Switch 5
not used
Switch 6
not used
switch 7
not used
switch 8
VacuDAP duo installed
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-3
9 Virtual Collimator
DIP switch 1 module number virtual collimator Switch 1
CAN bus address OFF
Switch 2
CAN bus address ON
Switch 3
CAN bus address OFF
Switch 4
CAN bus address ON
Switch 5
CAN bus address OFF
Switch 6
CAN bus address OFF
Switch 7
CAN bus address OFF
Switch 8
CAN bus address OFF
Switches must be set like indicated.
9-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
9.3.2.1
DIP switch factory setting
Fig. 9-4
U517 S1 and S2 factory settings
9.3.3
LEDs
Status LEDs for Service Requirements LED 1 red
flashing: shows startup phase off: shows system is running
LED 5 red
future use
LED 6 green
system is running
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-5
9 Virtual Collimator
9.3.4
Connectors
X1: Programming Interface Connector 10 pin Panduit cable connector Pin 1
PTC0
Pin 2
PTC1
Pin 3
PTC3
Pin 4
IRQ
Pin 5
RST
Pin 6
WDO
Pin 7
PTA0
Pin 8
OSC
Pin 9
+5 V output
Pin 10
ground
X2: RS232 Debug Connector 3 pin Panduit cable connector Pin 1
ground
Pin 2
TxD
Pin 3
RxD
X3: CAN Bus Connector J12 6 pin flat ribbon cable connector Pin 1
CAN +5 V
Pin 2
not used
Pin 3
CAN_RxD
Pin 4
CAN_TxD
Pin 5
not used
Pin 6
ground
9-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
X4: Rotation Motor Connector 5 pin molex cable connector Pin 1
+5 V
Pin 2
ground
Pin 3
servo control pulse
Pin 4
feedback servo position
Pin 5
servo return (there or not)
X5: Iris Motor Connector 5 pin molex cable connector Pin 1
+5 V
Pin 2
ground
Pin 3
servo control pulse
Pin 4
feedback servo position
Pin 5
servo return (there or not)
X6: Slot Motor Connector 5 pin molex cable connector Pin 1
+5 V
Pin 2
ground
Pin 3
servo control pulse
Pin 4
feedback servo position
Pin 5
servo return (there or not)
X8: Rotation Potentiometer 3 pin Panduit cable connector Pin 1
+5 V
Pin 2
feedback voltage
Pin 3
ground
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-7
9 Virtual Collimator
X9: Laser Diode 1 and 2 Connector 3 pin Panduit cable connector Pin 1
Laser -
Pin 2
Laser +
Pin 3
not connected
X10: DAP Connector 8 pin molex cable connector Pin 1
count input
Pin 2
reset output
Pin 3
calibrate output +
Pin 4
calibrate output -
Pin 5
ground
Pin 6
unused
Pin 7
unused
Pin 8
+UB 18 V
X11: Power Input 2 pin Panduit cable connector Pin 1
ground
Pin 2
+14 V
X12: VacuDAP duo Connector 4 pin Panduit cable connector Pin 1
+15 V
Pin 2
ground
Pin 3
RS 485 A
Pin 4
RS 485 B
9-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
X13: Wellhöfer Connector RJ45 connector Pin 1
RS 485 A
Pin 2
unused
Pin 3
+15 V
Pin 4
ground
Pin 5
RS 485 B
Pin 6
RS 485 Z
Pin 7
unused
Pin 8
RS 485 Y
Pin 9
ground
Pin 10
ground
X14: Laser Diode Flat Detector 3 pin Panduit cable connector Pin 1
Laser -
Pin 2
Laser +
Pin 3
not connected
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-9
9 Virtual Collimator
9.4
Changing the Collimator
9.4.1
Part exchange
• Remove connectors X7, X9, X10, X11, X12, X13. • Remove the 4 mounting screws. With power switched off, the collimator can be rotated in order to access the screws. • Fit the DAP-meter to the new collimator • Mount the new collimator with the 4 screws. Take care about the pre-collimator. Do not tighten the screws, because centering is necessary. • Connect X7, X9, X10, X11, X12, X13. Power up the unit.
9.4.2
Mechanical centering
• Press briefly both buttons of each function of the collimator (iris, slot, rotation) to get the starting position. • Close the iris collimator to a diameter of approx. 10 cm. • Screen in order to get an image of the collimator. • Press briefly one of the collimator buttons to display the preview. • Adjust the center of the collimator. • Repeat the two steps above until collimator is centered to preview. • If the image of the collimator is in the center and you can see a hexagon, the adjustment was successful. • Tighten the collimator mounting screws. • Press the both buttons of each collimator control (iris, slot, rotation) to get the starting position.
9-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
9.4.3
Collimator size adjustment
• Go to the Generator tab in the Exposcop Studio. • Set the collimator size for every image format with parameters “Collimator values”: “Normal”, “Magnifier1” and “Magnifier2”.
Fig. 9-5
Collimator size adjustment
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-11
9 Virtual Collimator
9.4.4
Preview size adjustment
To match the virtual collimator to the iris collimator, do the following: • Press the Close Iris Collimator key until the collimator is closed as far as possible. • Initiate radiation for visual control and terminate it again. • The menu system on the reference screen disappears. • Press the MENU key. • The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the System settings menu item. • The virtual collimator appears on the live screen. • Press the up arrow or down arrow key until the cursor is on the Collimator offset menu item. • Press the left arrow or right arrow key until the actual and the virtual collimator coincide. • Press key F1 – Save. • Press the open iris collimator key until the edges of the collimator are just visible on the live screen. • Initiate radiation for visual control and terminate it again. • The menu system on the reference screen disappears. • Press the MENU key. • The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the System settings menu item. • The virtual collimator appears on the live screen again. • Press the up arrow or down arrow key until the cursor is on the Collimator factor menu item. • Press the left arrow or right arrow key until the actual and the virtual collimator coincide. • Press key F1 – Save. • Repeat the collimator offset and collimator factor adjustment alternately until the deviations are minimal.
9-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9 Virtual Collimator
Fig. 9-6
Iris collimator settings
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
9-13
9 Virtual Collimator
9-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10 Keyboards Module 6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 10.1 Keyboard of Mobile Stand ................................................................................ 10.1.1 MST keyboard view........................................................................... 10.1.2 U400 connections ............................................................................. 10.1.3 U400 block diagram .......................................................................... 10.2 Keyboard of Monitor Cart ................................................................................. 10.2.1 MC keyboard view............................................................................. 10.2.2 U370 connections .............................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10-1 10-1 10-2 10-3 10-4 10-4 10-5
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10 Keyboards on Mobile Stand and Monitor Cart
10 Keyboards on Mobile Stand and Monitor Cart 10.1
Keyboard of Mobile Stand
10.1.1
MST keyboard view
Fig. 10-1
MST keyboard view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10-1
10 Keyboards on Mobile Stand and Monitor Cart
10.1.2
U400 connections
Fig. 10-2
U400 connections
10-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10 Keyboards on Mobile Stand and Monitor Cart
10.1.3
U400 block diagram
Fig. 10-3
U400 block diagram
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10-3
10 Keyboards on Mobile Stand and Monitor Cart
10.2
Keyboard of Monitor Cart
10.2.1
MC keyboard view
Fig. 10-4
MC keyboard view
10-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10 Keyboards on Mobile Stand and Monitor Cart
10.2.2
U370 connections
Fig. 10-5
U370 connections
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
10-5
10 Keyboards on Mobile Stand and Monitor Cart
10-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 11.1 General............................................................................................................. 11.1.1 Monitor cart with 18.1" flatscreen monitors ....................................... 11.1.2 Monitor cart with 24’’ split screen monitor ......................................... 11.1.3 Monitor cart with 23’’ split screen monitor ......................................... 11.1.4 Monitor cart with CRT monitors......................................................... 11.2 Monitor Cart Wiring .......................................................................................... 11.2.1 Wiring of flatscreen monitors............................................................. 11.2.2 Wiring of flatscreen monitors and PC................................................ 11.2.3 Wiring of CRT monitors..................................................................... 11.2.4 Wiring of external X-ray indicator and remote (options).................... 11.2.5 U385 wiring ....................................................................................... 11.3 Yellow RJ45 network cables ............................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11-1 11-1 11-2 11-3 11-4 11-5 11-5 11-6 11-7 11-8 11-9 11-10
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11 Monitor Cart
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11 Monitor Cart General and Wiring
11 Monitor Cart General and Wiring 11.1
General
11.1.1
Monitor cart with 18.1" flatscreen monitors
Fig. 11-1
System overview of monitor cart with 18.1" flatscreen monitors
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11-1
11 Monitor Cart General and Wiring
11.1.2
Monitor cart with 24’’ split screen monitor
Fig. 11-2
System overview of monitor cart with 24’’ split screen monitor
11-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11 Monitor Cart General and Wiring
11.1.3
Monitor cart with 23’’ split screen monitor
Fig. 11-3
System overview of monitor cart with 23’’ split screen monitor
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11-3
11 Monitor Cart General and Wiring
11.1.4
Monitor cart with CRT monitors
Fig. 11-4
System overview of monitor cart with CRT monitors
11-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11 Monitor Cart General and Wiring
11.2
Monitor Cart Wiring
11.2.1
Wiring of flatscreen monitors
Fig. 11-5
Wiring of flatscreen monitors
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11-5
11 Monitor Cart General and Wiring
11.2.2
Wiring of flatscreen monitors and PC
Fig. 11-6
Wiring of flatscreen monitors and PC
11-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11 Monitor Cart General and Wiring
11.2.3
Wiring of CRT monitors
Fig. 11-7
Wiring of CRT monitors
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11-7
11 Monitor Cart General and Wiring
11.2.4
Wiring of external X-ray indicator and remote (options)
Fig. 11-8
Wiring of external X-ray indicator and remote (options)
11-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11 Monitor Cart General and Wiring
11.2.5
U385 wiring
Fig. 11-9
U385 wiring
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
11-9
11 Monitor Cart General and Wiring
11.3
Yellow RJ45 network cables
In Monitor Carts shipped from october 2008 all network connections are made with yellow RJ45 cables. This is done in order to distinguish the network cables from the CAN-Bus cables.
Fig. 11-10
11-10
Yellow network cable at image memory
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12 Video Documentation Module 7
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 12.1 UP 960 ............................................................................................................. 12.1.1 Front view ......................................................................................... 12.1.2 Rear view .......................................................................................... 12.2 UP 980 ............................................................................................................. 12.2.1 Front view ......................................................................................... 12.2.2 Rear view .......................................................................................... 12.2.3 UP 980 configuration ........................................................................ 12.3 VCR.................................................................................................................. 12.3.1 VCR wiring standard ......................................................................... 12.3.2 U411 VCR interface ..........................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12-1 12-1 12-2 12-3 12-3 12-4 12-5 12-6 12-6 12-7
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12 Video Documentation
12 Video Documentation 12.1
UP 960
12.1.1
Front view
Fig. 12-1
UP 960 front view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12-1
12 Video Documentation
12.1.2
Rear view
Fig. 12-2
UP 960 rear view
12-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12 Video Documentation
12.2
UP 980
12.2.1
Front view
Fig. 12-3
UP 980 front view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12-3
12 Video Documentation
12.2.2
Rear view
Fig. 12-4
UP 980 rear view
12-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12 Video Documentation
12.2.3
UP 980 configuration
• Press the menu button. The door panel opens and the sub panel appears. • Display the desired main menu by pressing the Õ or Ö button. • Display the desired sub menu by pressing the × or Ø button. • Select the desired selection or change the value by pressing the Õ or Ö button. Menu Picture
Menu Layout
Menu Printer
Submenu
Value 2
For thermal paper UPP-210SE/210HD
Value 3
For thermal film UPT-210BL
Interlaced
Non Interlaced
Aspect adjust
Value -0.5
Value -0.1
H-Shift
Value 170
Value 165
H-Size
Value 1225
Value 1157
V-Shift
Value 25
Value 25
V-Size
Value 585
Value 585
AUTO-SCAN
OFF
75 Ohm
ON
If no equipment is connected to the video out of the printer
Pixel Density
High
AGC
ON
Save
Value1
Paper Type
Submenu
Submenu
• Save the settings. Menu Printer
Submenu
• Afterwards press EXEC button. All other values remain factory preset.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12-5
12 Video Documentation
12.3
VCR
12.3.1
VCR wiring standard
Fig. 12-5
VCR wiring standard
12-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12 Video Documentation
12.3.2
U411 VCR interface
Fig. 12-6
U411 VCR interface
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
12-7
12 Video Documentation
12-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13 Monitors Module 9
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 13.1 18.1" Flatscreen Monitors................................................................................. 13.1.1 Description ........................................................................................ 13.1.2 Technical data................................................................................... 13.1.3 Switches............................................................................................ 13.1.4 Connectors........................................................................................ 13.1.5 On-screen display ............................................................................. 13.1.5.1 Locking and unlocking the on-screen display .................... 13.1.5.2 Restoring the factory settings ............................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13-1 13-1 13-2 13-2 13-3 13-4 13-4 13-5
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13 Monitors
13 Monitors 13.1
18.1" Flatscreen Monitors
13.1.1
Description
The 18.1" flatscreen monitor is a monochrome, high-resolution LC display. All settings and adjustments are made using an integrated setup menu.
Fig. 13-1
18.1" flatscreen monitor
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13-1
13 Monitors
13.1.2
Technical data
Viewable Diagonal (inches)
18.1
Brightness (cd/m², typical)
700 (without anti-glare filter)
Native Resolution
1280 × 1024
Dot Pitch (mm)
0.28
Vertical Viewing Angle
170o
Horizontal Viewing Angle
170o
Contrast Ratio (typical)
400:1
VGA Input signal level at 75 Ohm
0.7 Vp-p
BNC Input signal level
1 Vp-p
Power (nominal)
42 W
Monitor Weight
6.35 kg
Operating Temperature
0 oC to 40 oC
Storage Temperature
-20 oC to 60 oC
Relative Humidity (non condensing)
20% – 85%
13.1.3
Switches
There is a power switch on the rear side of every monitor.
Fig. 13-2
13-2
Monitor power switches
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13 Monitors
13.1.4
Connectors
+12 V in
DVI / VGA in
Video in
RS232 interface
Fig. 13-3
Monitor connectors
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13-3
13 Monitors
13.1.5
On-screen display
13.1.5.1
Locking and unlocking the on-screen display
The monitors are set up and adjusted by an on-screen display. The functions are controlled by an integrated keypad.
Fig. 13-4
Monitor keypad
The on-screen display is activated by pressing the “Menu” key. On the “Setup” tab, you can lock the menu.
Fig. 13-5
“Setup” tab
If this function is activated, the menu will disappear. Instead of the menu, a message “MENU LOCKED” is displayed for a short time. Now the access to the on-screen display is denied. You will get this message anytime you press a key on the keypad.
Fig. 13-6
13-4
Menu locked
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13 Monitors
To unlock the menu, you have to press the “MENU” and “SCROLL” keys until the message “MENU UNLOCKED” appears.
Fig. 13-7
Menu unlocked NOTE You can use the “MENU” and “SCROLL” key combination also to lock the onscreen display directly. This function will toggle between “MENU LOCKED” and “MENU UNLOCKED”.
13.1.5.2
Restoring the factory settings
On the “Defaults” tab of the on-screen menu, you can restore the factory settings. The factory settings are optimized for the video output of the Vista image memory system.
Fig. 13-8
“Defaults” tab
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
13-5
13 Monitors
13-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply Module 10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 14.1 14.2 14.3 14.4 14.5 14.6
14.7 14.8 14.9
View of Power Supply ...................................................................................... General Description.......................................................................................... Main Components of Power Module 10 ........................................................... Module 10 Block Diagram ................................................................................ Module 10 Power Distribution Wiring Diagram................................................. 14.5.1 Wiring with transformer 11090 (105 V – 240 V) ................................ 14.5.2 Wiring with transformer 11091 (200 V – 240 V) ................................ Connectors ....................................................................................................... 14.6.1 Mains input........................................................................................ 14.6.2 Insulated XP0 PCB outlet ................................................................. 14.6.3 Internal mains distribution ................................................................. 14.6.4 On/Off connections ........................................................................... Fuses................................................................................................................ U357 schematics.............................................................................................. 14.8.1 Power-on logic .................................................................................. 14.8.2 U357 schematic power distribution ................................................... Module 10 Transformers .................................................................................. 14.9.1 Transformer 11090 105 V – 240 V.................................................... 14.9.2 Transformer 11091 200 V – 240 V.................................................... 14.9.3 Mains transformer thermo fuse .........................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-1 14-2 14-2 14-4 14-5 14-5 14-6 14-7 14-7 14-7 14-7 14-8 14-9 14-11 14-11 14-12 14-13 14-13 14-14 14-15
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
14 Power Supply 14.1
View of Power Supply
Fig. 14-1
Monitor cart power supply WARNING Module 10 (monitor cart power supply) contains dangerous voltages.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-1
14 Power Supply
14.2
General Description
The monitor cart module 10 power supply is the main power module for the Ziehm Vista C-arm and provides − power isolation from mains input − internal power distribution of 230 VAC to the monitor cart and through the XP0 for the mobile stand − inrush current limiter − on/off circuitry − emergency stop and key switch lock for power
14.3
Main Components of Power Module 10
Isolation transformer TR1 provides: − isolation from the mains line power supply − input voltage taps for matching the input voltage to maintain 230 VAC output − the main supply voltage for the complete C-arm U357 power distribution board provides: − power on/off control − distribution of power − U357 allows an easy method for matching the input line voltage to the facility power − matching TR1 input to mains line voltage − the ability to set the correct voltage is accomplished by following the power selection table in the diagram below − X12 to X21 terminals for matching input line to TR1 − output 230 VAC for system
14-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
Matching the line input to TR1: Matching the line input to facility is accomplished by performing a simple measurement of the wall outlet connector with a digital volt meter. Monitor the voltage for a period of time long enough to detect any unusual dips in the mains line (10 to 20 minutes). NOTE Those facilities demonstrating large fluctuations in the mains line voltage may require a more detailed measurement test. Placing a lines voltage monitor and monitor the line voltage for 12 to 24 hours. This is an important issue and should not be overlooked. Once the line voltage is determined to be stable and within the range of the selection table the service technician must follow the tap setting and jumper placement as required in the Voltage Selection table for the U357 board.
230 VAC power distribution U357: − the output of the TR1 transformer (X10 and X11) supplies 230 VAC to the U357 board − distributes 230 VAC to auxiliary equipment by means of the terminal connectors X24 to X30 − supplies power to devices such as video monitors, printers, image system, etc. − distributes 230 VAC power to the mobile stand by means of the power connector terminals X22 and X23
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-3
14 Power Supply
14.4
Module 10 Block Diagram
Fig. 14-2
Module 10 block diagram
14-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
14.5
Module 10 Power Distribution Wiring Diagram
14.5.1
Wiring with transformer 11090 (105 V – 240 V)
Fig. 14-3
Module 10 wiring with transformer 11090 set to 115 V Select Voltage Transformer 11090
105 VAC
110 VAC
115 VAC
120 VAC
210 VAC
220 VAC
230 VAC
240 VAC
Connections with AWG 14 (Blue) A-H
A-F
A-D
A-B
A-H
A-F
A-D
A-B
H-I
F-G
D-E
B-C
K-I
K-G
K-E
K-C
K-L
K-L
K-L
K-L
F3 Mains Fuses F1 = F2 = 20 A medium
F2
F1
Mains Fuses F1 = F2 = 15 A medium
CAUTION Only one of the fuses F1, F2, F3 must be inserted!
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-5
14 Power Supply
14.5.2
Wiring with transformer 11091 (200 V – 240 V)
Fig. 14-4
Module 10 wiring with transformer 11091 set to 115 V Select Voltage Transformer 11091 200 VAC
230 VAC
240 VAC
A-K
A-I
A-H
Connections with AWG 14 (Blue)
F2
F1
Mains Fuses F1 = F2 = 15 A medium
CAUTION Only one of the fuses F1, F2, F3 must be inserted!
14-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
14.6
Connectors
The module 10 has many connectors and voltage tap settings. The following will provide information on the name and function of the listed connector.
14.6.1
Mains input
X1: To P0 connector / Earth 1:
Earth
Provides connection to the earth or chassis ground for safety X2: Mains Input I / Line 1:
Line
X3: Mains Input I / Neutral 1:
14.6.2
Neutral
Insulated XP0 PCB outlet
X22: XP0 connection screw connection 1:
XP0 connector / Internal mains to mobile stand
X23: XP0 connection screw connection 1:
14.6.3
XP0 connector / Internal mains to mobile stand
Internal mains distribution
X24 - X30: Cart internal mains distribution 230 V~ 3 pin Phoenix MSTBV 2,5/3-G connector Order number Phoenix: 17 53 45 3 Part Number: 21368 Pin 1:
Protective earth
Pin 2:
230 VAC insulated mains I / Neutral
Pin 3:
230 VAC insulated mains I / Line
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-7
14 Power Supply
14.6.4
On/Off connections
X9: XP0 On/Off 3 pin Panduit MLSS 100 connector Part Number: 21364 Pin1:
OFF / active low
Pin 2:
ON / active low
Pin 3:
Ground
X8: Keyboard On/Off Flat Ribbon Connector 10 pin flat ribbon connector Part Number: 21090 Pin 1:
Ground
Pin 2:
Ground
Pin 3:
LED output
Pin 4:
LED output
Pin 5:
n. c.
Pin 6:
n. c.
Pin 7:
ON / active low
Pin 8:
ON / active low
Pin 9:
OFF / active low
Pin 10:
OFF / active low
X31: Thermo Switch Main Transformer 2 pin Panduit MLSS 100 connector Part Number: 21335 Pin 1:
Thermo switch
Pin 2:
Thermo switch
Main power off
14-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
X4: Additional Key Switch 2 pin Panduit MLSS 100 connector Part Number: 21335 Pin 1:
key switch
Pin 2:
key switch
Main power off Must be shorted if no key switch is installed
14.7
Fuses
Fuse F1 Mains Fuse 210-220-230-240 VAC system Mains Fuse: 15 A medium / Part Number: 17020 105-110-115-120 VAC system Mains Fuse: 20 A medium / Part Number: 17044 Fuse F2 Mains Fuse 210-220-230-240 VAC system Mains Fuse: 15 A medium / Part Number: 17020 105-110-115-120 VAC system Mains Fuse: 20 A medium / Part Number: 17044 Fuse F1 240 VAC system Supply voltage: 100 mA slow / Part Number: 17010 Not installed for 230 VAC; 115 VAC; 110 VAC; 105 VAC Fuse F2 210-220-230 VAC system Supply voltage: 100 mA slow / Part Number: 17010 Not installed for 240 VAC; 115 VAC; 110 VAC; 105 VAC
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-9
14 Power Supply
Fuse F3 105-110-115-120 VAC system Supply voltage: 100 mA slow / Part Number: 17010 Not installed for 240 VAC; 230 VAC Fuse F4 Supply voltage: 315 mA slow Installed for all mains input voltages
14-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
14.8
U357 schematics
14.8.1
Power-on logic
Fig. 14-5
U357 schematic power-on logic
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-11
14 Power Supply
14.8.2
U357 schematic power distribution
Fig. 14-6
U357 schematic power distribution
14-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
14.9
Module 10 Transformers
14.9.1
Transformer 11090 105 V – 240 V
Fig. 14-7
Transformer 11090 105 V – 240 V
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-13
14 Power Supply
14.9.2
Transformer 11091 200 V – 240 V
Fig. 14-8
Transformer 11091 200 V – 240 V
14-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14 Power Supply
14.9.3
Mains transformer thermo fuse
Protective Device Thermo sensor Rating: 120 °C UL recognized Not connected to primary and/or secondary circuit
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
14-15
14 Power Supply
14-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System Module 11
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 15.1 15.2 15.3 15.4 15.5 15.6 15.7 15.8 15.9 15.10 15.11
Features ........................................................................................................... Barebone Front view Step 1............................................................................. Barebone Front view Step 2............................................................................. Complete Front View Step 1 ............................................................................ Complete Front View Step 2 ............................................................................ Side View ......................................................................................................... Barebone Inside View Step 1 ........................................................................... Barebone Inside View Step 2 ........................................................................... Complete Inside View Step 1 ........................................................................... Complete Inside View Step 2 ........................................................................... Image Memory Service Settings ...................................................................... 15.11.1 Overview ........................................................................................... 15.11.2 Service settings................................................................................. 15.11.3 Step windowing ................................................................................. 15.11.4 Filter factors ...................................................................................... 15.11.4.1Filters for anatomical programs ......................................... 15.11.4.2Filters for subtraction modes ............................................. 15.11.4.3Filters for subtraction modes with CO2 ...................................... 15.11.5 Windowing settings for subtraction modes........................................ 15.11.6 DICOM settings................................................................................. 15.11.7 System settings................................................................................. 15.11.8 Monitor calibration............................................................................. 15.11.9 HEDIS data ....................................................................................... 15.11.10 Software update ................................................................................ 15.12 Displaying Additional Information .....................................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-1 15-3 15-4 15-5 15-6 15-7 15-8 15-9 15-10 15-11 15-12 15-12 15-12 15-13 15-15 15-16 15-19 15-20
15-21 15-22 15-22 15-25 15-27 15-29 15-30
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15 Image Memory System V2 1/2k 15.1
Features
Mains
230 VAC
Mains Cycles
50/60 Hz
Rating
300 W
CPU
VIA 750 MHz, 128 MB RAM
Hard disk
8 GB partition for system, Ziehm file system for images
Number of Images
3 /150 / 1000 / 5000
Patient Folder System
Mosaic view with 16 thumbnail images and search function
Video Input/Digitizing
25 Hz non interlaced, 8 bit / 50Hz interlaced, 8 bit
Scan Rate
15 MHz
Pixels
576 × 576 × 8 bit
Alu
16 bit
Filtering
Stack filter: 1....16 Recursive Filter: 0....3 Real-time edge filter: 0....4
Subtraction
DSA, MSA, RSA, Auto-MSA, Auto-RSA Phase 1; with pixel shift and landmarking
Post Processing
Zoom, rotation and image reversal Measuring function, inversion, Windowing
Cine
1 / 2 / 4 / 8 / 12.5 frames per second Length: 50, 100, 150, 200, 250 images
Monitor out
75 Hz non interlaced (type detection by dongle / IIC interface) types: − 17" CRT − 18.1" TFT − 23" 16:9 split screen
Brightness Sensor via IIC Bus
Automatic brightness control by ambient light
VIDEO OUT external
50 Hz CCIR
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-1
15 Image Memory System V2 1/2k
Interfaces
CAN, USB, RS232, SCSI, Ethernet, Floppy, IIC; Centronics (VGA + keyboard for service)
Floppy
3.5"; 1.44 MB; DICOM images or TIF images
CD-R drive
for backup, TIF images, DICOM images
MO drive
3.5"; 640 MB for backup, TIF images or DICOM images
USB drive
for backup, TIF images or DICOM images.
DICOM
RJ45 or LWL Storage / Query / Retrieve / Print / Worklist / MPPS / Storage commitment / Media Storage
Paper Printer
UP960, UP980
VCR
Remote Interface implemented NOTE The availability of features depends on the options installed.
15-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.2
Barebone Front view Step 1
Fig. 15-1
Module 11 barebone front view Step 1
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-3
15 Image Memory System V2 1/2k
15.3
Barebone Front view Step 2
Fig. 15-2
Barebone Front view Step 2
15-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.4
Complete Front View Step 1
Fig. 15-3
Module 11 complete front view Step 1
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-5
15 Image Memory System V2 1/2k
15.5
Complete Front View Step 2
Fig. 15-4
Module 11 complete front view Step 2
15-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.6
Side View
Fig. 15-5
Module 11 side view
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-7
15 Image Memory System V2 1/2k
15.7
Barebone Inside View Step 1
Fig. 15-6
Module 11 barebone inside view Step 1
15-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.8
Barebone Inside View Step 2
Fig. 15-7
Module 11 barebone inside view Step 2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-9
15 Image Memory System V2 1/2k
15.9
Complete Inside View Step 1
Fig. 15-8
Module 11 complete inside view Step 1
15-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.10
Complete Inside View Step 2
Fig. 15-9
Module 11 complete inside view Step 2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-11
15 Image Memory System V2 1/2k
15.11
Image Memory Service Settings
15.11.1
Overview
General A large number of parameters can be preset for the Ziehm Vista. This is done in a configuration menu. How to use menus To open the Configuration Menu, you use the MENU key on the monitor cart. To switch between the different menu items, do the following: • Press the up arrow or down arrow key until the cursor is on the desired menu item. To select a setting option, do the following: • Press the left arrow or right arrow key until the desired option appears, or • Enter the desired value using the alphanumeric keypad on the control panel.
15.11.2
Service settings
Function The Service Settings menu is used for modifying different system parameters which affect both the quality of the fluoroscopic image and the user interface. Password The Service Settings menu is password-protected. If you are an authorized person, you can obtain the password from your local Ziehm dealer.
15-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
Fig. 15-10
Service Settings menu
15.11.3
Step windowing
Function The Step Windowing menu is used for presetting 4 fixed steps for the contrast window level (L) and width (W). During fluoroscopy, you can select these steps with the contrast keys on the Carm stand. NOTE For the subtraction modes, no step windowing is available.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-13
15 Image Memory System V2 1/2k
Fig. 15-11
Step Windowing menu
Parameter range The available parameter range (0–255) is displayed as soon as the menu item is selected with the down arrow or up arrow key. The desired values must be entered with the alphanumeric keys. NOTE Default values are listed in the adjustment chapter.→ Ch. 19.2.3.1, p. 19-14
Activating the step windowing mode The Windowing menu item is used for setting either standard windowing mode or step windowing mode for the contrast keys on the C-arm stand. − Windowing: Off The contrast keys on the C-arm stand work in standard windowing mode. − Windowing: On The contrast keys on the C-arm stand work in step windowing mode. Default setting The default windowing steps which are activated after power-up can be programmed for each individual anatomical program under Configuration Menu → Service Settings → Filter Factors → Organ 0 / Organ 1 / Organ 2 / Soft (→ Ch. 15.11.4.1, p. 15-16).
15-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.11.4
Filter factors
Function The Filter Factors submenu is used for setting different filter factors for each anatomical program and each subtraction mode in separate submenus.
Fig. 15-12
Filter Factors menu
Menu item
Anatomical program
Organ 0 = extremities & children Organ 1 = head, spine & pelvis Organ 2 = thorax & abdomen
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-15
15 Image Memory System V2 1/2k
15.11.4.1 Filters for anatomical programs Filter types For each of the anatomical programs Organ 0, Organ 1, Organ 2 and Soft, the values for the following filters can be programmed: − Recursive filter for continuous fluoroscopy (independent settings for normal view, magnification 1 and magnification 2) − Stack filter for continuous fluoroscopy − Edge filter − Default windowing step after power-up − Stack filter for continuous pulse fluoroscopy − Stack filter for snapshot
Fig. 15-13
15-16
Filter Factors for Organ Key 0 menu
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
Parameter values The following parameter values can be set: Filter type
Fluoroscopy mode
Available parameters
Recursive filter
Continuous fluoroscopy
Level 0, 1, 2, 3
Stack filter
Continuous fluoroscopy
1, 2, 4, 8, 16 images
Edge filter
Continuous fluoroscopy, continuous pulse fluoroscopy, snapshot
Level 0, 1, 2, 3, 4
Windowing step
Continuous fluoroscopy, continuous pulse fluoroscopy, snapshot
Level -2, -1, 0, +1, +2
Continuous pulse
Continuous pulse fluoroscopy (= stack filter)
1, 2, 4, 8, 16 images
Snapshot
Snapshot (= stack filter)
1, 2, 4, 8, 16 images
Recursive filter The recursive filter adds a certain number of images (which are determined by the preset level) during fluoroscopy. Each newly-acquired image is superimposed by the result of the previous addition with a certain weighting factor. The higher the number of images, the greater the noise suppression, but also the greater motion blurring. The values can be set independently for normal view, magnification 1 and magnification 2. The values are displayed in the lower left corner of the live screen in the following form: N=0, N=1, N=2, N=3. The preset values can be adjusted during operation with the F2 and F3 keys on the C-arm stand control panel. Stack filter The stack filter generates and adds the preset number of images after radiation has been terminated. The higher the preset number of images, the greater the noise suppression, but also the greater motion blurring if the patient moves during image generation.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-17
15 Image Memory System V2 1/2k
Edge filter The edge filter is used for setting a greater or lesser degree of enhancement (sharpening) of edges within the image. 4 levels of edge enhancement are available: Level
Meaning
0
No edge enhancement (original fluoroscopic image)
1
Slight edge enhancement
2
Medium edge enhancement
3
Strong edge enhancement
Increasing the level of edge enhancement will also cause any noise in the image to become more apparent, though. Windowing step This value determines which windowing step is automatically set by the system for the corresponding anatomical program after power-up, provided that the step windowing mode has been activated (→ Ch. 15.11.3, p. 15-13). You can adjust these default settings later during operation using the contrast keys on the C-arm stand. Continuous pulse This setting defines how many images are used for the stack filter in continuous pulse fluoroscopy mode (overriding the value specified under Stack filter). Snapshot This setting defines how many images are used for the stack filter in snapshot mode (overriding the value specified under Stack filter). NOTE Default values are listed in the adjustment chapter.→ Ch. 19.2.3.3, p. 19-15
15-18
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.11.4.2 Filters for subtraction modes Filter types For each of the subtraction modes DSA, RSA and MSA, the values for the following filters can be programmed: − Recursive filter (independent settings for normal view, magnification 1 and magnification 2) − Stack filter − Edge filter
Fig. 15-14
Filter Factors for DSA Key menu
Parameter values The following parameter values can be set: Filter type
Fluoroscopy mode
Available parameters
Recursive filter
Continuous fluoroscopy
Level 0, 1, 2, 3
Stack filter
Continuous fluoroscopy
1, 2, 4, 8, 16 images
Edge filter
Continuous fluoroscopy, continuous pulse fluoroscopy, snapshot
Level 0, 1, 2, 3, 4
NOTE Default values are listed in the adjustment chapter.→ Ch. 19.2.3.3, p. 19-15
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-19
15 Image Memory System V2 1/2k
15.11.4.3 Filters for subtraction modes with CO2 Filter types If the system is equipped with the CO2 option, the values for the following filters can be selected for each of the subtraction modes DSA, RSA and MSA in combination with CO2 negative contrast medium: − Recursive filter (independent settings for normal view, magnification 1 and magnification 2) − Stack filter − Edge filter
Fig. 15-15
Filter Factors for CO2 DSA menu
Parameter values The following parameter values can be set: Filter type
Fluoroscopy mode
Available parameters
Recursive filter
Continuous fluoroscopy
Level 0, 1, 2, 3
Stack filter
Continuous fluoroscopy
1, 2, 4, 8, 16 images
Edge filter
Continuous fluoroscopy, continuous pulse fluoroscopy, snapshot
Level 0, 1, 2, 3, 4
NOTE Default values are listed in the adjustment chapter.→ Ch. 19.2.3.3, p. 19-15
15-20
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.11.5
Windowing settings for subtraction modes
Function This menu is used for setting default values for the width (W) and level (L) of the contrast window. This is done separately for DSA, MSA and RSA. If the system is equipped with the CO2 option, the default windowing values for each individual subtraction mode in combination with CO2 can be set in addition.
Fig. 15-16
Windowing Settings for Subtraction Modes menu
Parameter range The available parameter range (0–255) is displayed as soon as the menu item is selected with the down arrow or up arrow key. The desired values must be entered with the alphanumeric keys. Function key F3 – CO2 moves the cursor to the column where the default windowing values for the subtraction modes with CO2 negative contrast medium are set. Function key F3 – Normal returns the cursor to the column where the default windowing values for the subtraction modes with positive contrast medium are set. NOTE The default values are listed in the adjustment chapter (→ Ch. 19.2.3.3, p. 19-15).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-21
15 Image Memory System V2 1/2k
15.11.6
DICOM settings
→ Ch. 16
15.11.7
System settings
In the System Settings menu, you can set or modify the following system parameters with the cursor keys: − Vario 3D function − DSA mask − Ambient light sensor − Histogram − Dose area product − File check − Collimators − Language
Fig. 15-17
System Settings menu
Vario 3D function On a Ziehm Vista, the Vario 3D function must always be set to Off.
15-22
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
DSA mask If the option Always is set, a new mask image is acquired during each DSA. NOTICE You must always set the Always option under DSA mask.
Ambient light sensor gain Under Bri sensor gain you define to which extent the system adjusts the screen backlight brightness to the ambient light conditions. You can select a value between 1 (weak gain) and 20 (strong gain). Ambient light sensor limit value Under Bri sensor limit you specify a maximum value for the screen backlight brightness. You can select a value between 0 and 32. Ambient light sensor speed Under Bri sensor speed you define how long it takes to adjust the screen backlight brightness to changes in ambient light conditions. You can select a value between 10 s and 60 s. Histogram Here you define whether the histogram is displayed or not. NOTE Histogram function will be disabled automatically after system has been turned off.
Dose area product Here you define whether the dose area product is displayed on the live screen or not. File check Here you define whether the system checks the internal image data list for corrupt files during power-up. NOTICE You must always set the On option under File check. Collimator offset and collimator factor → Ch. 9.4.4, p. 9-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-23
15 Image Memory System V2 1/2k
Language You can choose between the following languages for the user interface: − DEUTSCH − ENGLISH − FRANÇAISE − ITALIANO − ESPAÑOL − SVENSKA − POLSKI − CZECH To apply the settings, do the following: • Press key F1 – Save. • Switch off the system. • Switch the system back on. • The new settings are valid now. To discard system settings that have not been applied yet, do the following: • Press key F2 – Cancel, or • Quit the Service Settings menu without applying the settings.
15-24
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.11.8
Monitor calibration
This menu only appears in units equipped with CRT monitors. In the Monitor Calibration menu, you set the vertical and horizontal scaling and position of the screen display using a test image. There are 6 different test images available. Furthermore, the monitor’s serial number is automatically displayed in the Monitor Calibration menu.
Fig. 15-18
Monitor Calibration menu
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-25
15 Image Memory System V2 1/2k
To calibrate a monitor, do the following: • In the Monitor Calibration menu, press the up arrow or down arrow key until the cursor is on the Select monitor menu item. • Press the left arrow or right arrow key until the desired screen (Left or Right) is displayed. • The Monitor Calibration menu appears on the selected screen. • Press the down arrow key. • The Test image menu item is selected. • Press the left arrow or right arrow key until the desired test image (test image 1–6) is selected. • The selected test image is displayed on the other screen. • Press the down arrow key. • The Hor. size menu item is selected. • Press the left arrow or right arrow key until the image has the desired horizontal size. • Press the down arrow key. • The Hor. position menu item is selected. • Press the left arrow or right arrow key until the image is in the desired horizontal position. • Press the down arrow key. • The Vert. size menu item is selected. • Press the left arrow or right arrow key until the image has the desired vertical size. • Press the down arrow key. • The Vert. position menu item is selected. • Press the left arrow or right arrow key until the image is in the desired vertical position. • Press key F1 – Save to apply the settings.
15-26
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.11.9
HEDIS data
To upload the image memory data of the system to a USB storage device, do the following: • In the Service Settings menu, press the up arrow or down arrow key until the cursor is on the HEDIS data menu item. • Press the Enter key. • The keys F1 – Export and F2 – Import appear on the screen.
Fig. 15-19
Service Settings menu, HEDIS data menu item
• Plug the USB storage device to which you want to upload the image memory data into the USB port. • Press key F1 – Export. • The image memory data is uploaded to the USB storage device and saved there. • Press the ESC key.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-27
15 Image Memory System V2 1/2k
To download image memory data from a USB storage device to the system, do the following: • In the Service Settings menu, press the up arrow or down arrow key until the cursor is on the HEDIS data menu item. • Press the Enter key. • The keys F1 – Export and F2 – Import appear on the screen.
Fig. 15-20
Service Settings menu, HEDIS data menu item
• Plug the USB storage device with the image memory data you want to download to the system into the USB port. • Press key F2 – Import. • The image memory data is downloaded from the USB storage device to the system and saved there. • Press the ESC key.
15-28
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
15.11.10 Software update To perform a software update, do the following: • In the Service Settings menu, press the up arrow or down arrow key until the cursor is on the Software update menu item. • Plug the USB storage device that contains the update package into the USB port. • Press the Enter key. The following confirmation prompt appears: Do you really want to update? Y/N
Fig. 15-21
Service Settings menu, Software update menu item
• Press the Y key on the alphanumeric keypad. The software update is installed. Once the complete software update package has been installed, the following message is displayed: Software update is completed Please switch the system OFF and ON • Switch the system off and back on.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-29
15 Image Memory System V2 1/2k
15.12
Displaying Additional Information
To display additional information about the image memory do the following: • Press the menu key. • Select “Basic Settings”. • Press Enter. • Press the “Right Arrow” and “Image Transfer” keys simultaneously.
Fig. 15-22
“Right Arrow” and “Image Transfer” keys
Additional information will appear in the “Basic Settings” menu.
Fig. 15-23
15-30
“Basic Settings” menu with additional information
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15 Image Memory System V2 1/2k
Information contains: − First row − Version and release date of image memory software − Serial number of image memory − Serial number of C-arm (only if C-arm is connected) − Second row − Firmware versions of image memory hardware − Third row − Information about enabled options − Fourth row − Information about enabled options − Storage capacity of image memory − Fifth row − Information about video system (interlaced / non interlaced) − Information about enabled options
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
15-31
15 Image Memory System V2 1/2k
15-32
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 16.1 DICOM Overview ............................................................................................. 16.2 DICOM and Network Parameters..................................................................... 16.3 DICOM Configuration ....................................................................................... 16.3.1 Configuration of Storage and Storage Commitment ......................... 16.3.2 Configuration of Media Storage ........................................................ 16.3.3 Configuration of DICOM Query/Retrieve........................................... 16.3.4 Configuration of Worklist and MPPS................................................. 16.3.5 Configuration of the print servers ...................................................... 16.3.6 Test of the configuration with DICOM Verification (Echo).................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-1 16-2 16-4 16-4 16-6 16-8 16-12 16-17 16-21
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
16 DICOM Configuration Manual 16.1
DICOM Overview
DICOM means Digital Imaging and Communications in Medicine. This worldwide standard was developed by ACR and NEMA for the exchange of information between digital imaging computer systems in medical environments. The Ziehm C-arm can be supplied with 5 DICOM options: 1 With the DICOM Storage option you can send x-ray documentation to a DICOM computer network. With DICOM Storage Commitment you can get a confirmation from the archive that an image is saved. 2 With the DICOM Media Storage option you can save x-ray documentation to a 650 MB formatted MO disk or DVD/CD. 3 The DICOM Query/Retrieve option is used to query patient, study, series and image data from a computer network. 4 The DICOM Worklist option is used to query a worklist job for today from a computer network. MPPS can be used to send worklist jobs into the status “in progress” and finally “completed”. 5 The DICOM Basic Grayscale Print option is used to print x-ray images to a computer network. The basic document of the DICOM application is the “Conformance Statement”, which is obligation for every “DICOM participant”. With this document you can decide whether the communication between Ziehm C-arm (SCU = Service Class User) and server (SCP = Service Class Provider) will be successful. The DICOM transmission needs the network parameters described in the following section:
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-1
16 DICOM Configuration Manual
16.2
DICOM and Network Parameters
To enable the system to establish a connection for a DICOM data transfer, you must specify the following network parameters: − IP addr. DICOM server (IP = Internet Protocol) Each host on the network is assigned an IP address consisting of a 32-bit number, e.g. 192.168.127.080 (192.168.127 = group address, .080 = individual host) IP address ranges (examples, other configurations are possible): − Class A network: 10.0.0.0 to 10.255.255.255 − Class B network: 128.16.0.0 to 172.31.255.255 − Class C network: 192.168.0.0 to 192.168.255.255 − IP subnet mask 255.255.255.0 (subnet mask, combined with IP address yields group address, see above) IP address ranges (examples, other configurations are possible): − Class A network: 255.0.0.0 − Class B network: 255.255.0.0 − Class C network: 255.255.255.0 − Host name DICOM server Name of the network host, e.g. sun-server, with unique IP address assignment − AE title DICOM server (Application Entity = name of the application) Name of the DICOM application on the server, e.g. INOARCHIVE for DICOM Storage
16-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual − Port addr. DICOM server At this address, the server “listens” for requests from Ziehm NetPort. − IP address gateway If the server and Ziehm NetPort belong to different groups, a gateway must be specified, e.g. 192.168.127.254. − IP address ZiehmNetPort (IP address of Ziehm Vista) − AE title ZiehmNetPort (AE title of Ziehm NetPort, freely configurable) The DICOM Verification function enables you to check whether all parameters have been entered correctly (for error messages, → Ch. 16.3.6, p. 16-21. The default port address for DICOM is set to 104 and cannot be reconfigured. Entering special characters In some lines of the network parameter input screen, you may enter certain special characters. Function keys F4 – < and F5 – > appear in this case. A list of the available special characters is displayed above these two function keys. As long as you hold down one of these two function keys, the available special keys are displayed one after the other in the order of the list. With F4 – < you can scroll forward in the list. With F5 – > you can scroll backward in the list.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-3
16 DICOM Configuration Manual
16.3
DICOM Configuration
16.3.1
Configuration of Storage and Storage Commitment
Storage information model When configuring the storage server, you specify not only the general network parameter (→ Ch. 16.2, p. 16-2), but also the storage information model (SOP class = Service Object Pair). You can choose between the following options: − XA = X-ray Angiography − CR = Computed Radiography − SC = Secondary Capture The original images furnished by Ziehm NetPort are of type XA. Since some servers do not support the XA SOP class, CR and SC are provided as alternatives. Image attributes Under Image attributes included you define which filter algorithms from the image memory are applied to the DICOM image prior to transfer. By default, all attributes are set to Off, and the raw image is transferred. To configure the storage server, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Storage server menu item. The DICOM Settings Storage Server menu appears.
16-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Fig. 16-1
DICOM Settings Storage Server menu
• Enter the network parameters. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Activate or deactivate the desired image attributes. • Press key F6 – Verify. The system checks the network connection between SCU (Service Class User = Ziehm NetPort) and SCP (Service Class Provider = server) by means of a “ping”, the DICOM settings (e.g. AE title) and common SOP classes (“association negotiation”). It sends an ECHO request to the SCP and waits for an ECHO response. If all network parameters have been entered correctly, the message Verification successful is displayed (for error messages, → Ch. 16.3.6, p. 16-21). If your system includes the Storage Commitment function, the following message is displayed: Do you want to activate NetCommit? Y/N • Press the Y key if you want to activate the Storage Commitment function, or • Press the N key if you do not want to activate the Storage Commitment function. • Press key F1 – Save.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-5
16 DICOM Configuration Manual
16.3.2
Configuration of Media Storage
Storage information model When configuring the storage server, you specify not only the general network parameters (→ Ch. 16.2, p. 16-2), but also the storage information model (SOP class = Service Object Pair). You can choose between the following options: − XA = X-ray Angiography − CR = Computed Radiography − SC = Secondary Capture The original images furnished by Ziehm NetPort are of type XA. Since some servers do not support the XA SOP class, CR and SC are provided as alternatives. Image attributes Under Image attributes included you define which filter algorithms from the image memory are applied to the DICOM image prior to saving. By default, all attributes are set to Off, and the raw image is saved. DICOM graphics format The DICOM format must have been set as graphics format for writing images to USB storage device or CD. To configure Media Storage, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Media storage menu item. The DICOM Settings Media Storage menu appears.
16-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Fig. 16-2
DICOM Settings Media Storage menu
• Enter the AE title and the SOP class. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Activate or deactivate the desired image attributes. • Press key F1 – Save.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-7
16 DICOM Configuration Manual
16.3.3
Configuration of DICOM Query/Retrieve
Query model When configuring the query/retrieve server, you specify not only the general network parameters (→ Ch. 16.2, p. 16-2), but also the query model: − PATIENT ROOT: Patient-level data − STUDY ROOT: Study-level data NOTE If several patients with the same patient ID are stored in the PACS, you will get the patient with the respective ID who is sent by the PACS first. Network parameters for Retrieve To enable Ziehm NetPort to connect to the Query server for the Retrieve function, you must enter its AE title, IP address as well as port 104. Use the same IP address and AE title as in the DICOM Settings Query Server menu (→ Fig. 16-3, p. 16-9). To configure the query server, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Query server menu item. The DICOM Settings Query Server menu appears.
16-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Fig. 16-3
DICOM Settings Query Server menu
• Enter the network parameters. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Press key F6 – Verify. The system checks the network connection between SCU (Service Class User = Ziehm NetPort) and SCP (Service Class Provider = server) by means of a “ping”, the DICOM settings (e.g. AE title) and common SOP classes (“association negotiation”). It sends an ECHO request to the SCP and waits for an ECHO response. If all network parameters have been entered correctly, the message Verification successful is displayed (for error messages, → Ch. 16.3.6, p. 16-21).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-9
16 DICOM Configuration Manual
Retrieve server The retrieve server settings are entered in the DICOM Settings Retrieve Server X menu. If any requested images cannot be found on the query server, Ziehm NetPort will automatically select the appropriate retrieve server using the retrieve AE title (which is provided by the query server). To configure a retrieve server, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Query server menu item. The DICOM Settings Query Server menu appears.
Fig. 16-4
DICOM Settings Query Server menu
• Place the cursor on the IP addr. DICOM server menu item. • Press key F4 – < or F5 – >. The DICOM Settings Retrieve Server menu for the selected retrieve server (1 or 2) is displayed.
16-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Fig. 16-5
DICOM Settings Retrieve Server menu (here: Retrieve Server 1)
• Enter the network parameters. The Gateway, IP address Ziehm NetPort and AE title Ziehm NetPort settings must be identical to the corresponding query server settings. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Press key F6 – Verify. The system checks the network connection between SCU and SCP by means of a “ping”, the DICOM settings (e.g. AE title) and common SOP classes (“association negotiation”). It sends an ECHO request to the SCP and waits for an ECHO response. If all network parameters have been entered correctly, the message Verification successful is displayed (for error messages, → Ch. 16.3.6, p. 16-21).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-11
16 DICOM Configuration Manual
16.3.4
Configuration of Worklist and MPPS
Search arguments for Worklist Server When configuring the worklist server, you specify not only the general network parameters (→ Ch. 16.2, p. 16-2), but also one or several search arguments: − Modality: You can limit the Worklist to images of the XA or CR SOP class (→ p. 16-4) − Station AE title: You can limit the Worklist to jobs that are assigned to the AE title ZiehmNetPort specified (= YES; NO = no limitation) − Time span: The default setting for the time span is 0 – 24 o’clock of the current day. You may set a shorter period of time (full hours only). − Location: You can limit the Worklist to jobs that are related to the location of the respective Ziehm NetPort, e.g. OR 1. Operator’s name In addition, you must enter the name of the radiological technologist under Operator’s Name. To configure the worklist server, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Worklist server menu item. The DICOM Settings Worklist Server menu appears.
16-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Fig. 16-6
DICOM Settings Worklist Server menu
• Enter the network parameters. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Enter the desired search arguments. • Enter the name of the radiological technologist under Operator’s Name. • Press key F6 – Verify. The system checks the network connection between SCU and SCP by means of a “ping”, the DICOM settings (e.g. AE title) and common SOP classes (“association negotiation”). It sends an ECHO request to the SCP and waits for an ECHO response. If all network parameters have been entered correctly, the message Verification successful is displayed (for error messages, → Ch. 16.3.6, p. 16-21). • Press key F1 – Save.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-13
16 DICOM Configuration Manual
MPPS Server When configuring the MPPS server, you specify only the general network parameters (→ Ch. 16.2, p. 16-2). The search arguments are irrelevant for the MPPS server. To configure the MPPS server, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Worklist server menu item. The DICOM Settings Worklist Server menu appears.
Fig. 16-7
DICOM Settings Worklist Server menu
• Place the cursor on the IP addr. DICOM server menu item. • Press key F5 – >. The DICOM Settings MPPS Server menu appears.
16-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Fig. 16-8
DICOM Settings MPPS Server menu
• Enter the network parameters. • Enter the name of the radiological technologist under Operator’s Name. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Press key F6 – Verify. The system checks the network connection between SCU and SCP by means of a “ping”, the DICOM settings (e.g. AE title) and common SOP classes (“association negotiation”). It sends an ECHO request to the SCP and waits for an ECHO response. If all network parameters have been entered correctly, the message Verification successful is displayed (for error messages, → Ch. 16.3.6, p. 16-21). If the connection has been successfully established, the following message appears: Do you want to activate MPPS? Y/N • Press the Y key if you want to activate the MPPS function, or • Press the N key if you do not want to activate the MPPS function. If the system was unable to establish a connection, the MPPS function is automatically deactivated. • Press key F1 – Save.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-15
16 DICOM Configuration Manual
Returning to the Worklist server configuration To return from the DICOM Settings MPPS Server menu to the DICOM Settings Worklist Server menu, do the following: • In the DICOM Settings MPPS Server menu, place the cursor on the IP addr. DICOM server menu item. • Press key F4 – until the desired special character appears. • Specify the print configuration parameters. • Press key F1 – Save.
16-18
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
Configuring print server 2 To configure print server 2, do the following: • Press the MENU key. The Configuration Menu appears. • Select the Service settings menu item. • Type the password and press the Enter key. • Select the DICOM settings menu item. • Type the password and press the Enter key. • Select the Print server menu item. The DICOM Settings Print Server 1 menu appears.
Fig. 16-10
DICOM Settings Print Server 1 menu
• Place the cursor on the IP addr. DICOM server menu item. • Press key F5 – >. The DICOM Settings Print Server 2 menu appears.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-19
16 DICOM Configuration Manual
Fig. 16-11
DICOM Settings Print Server 2 menu
• Enter the network parameters. When the cursor is in certain lines of the input screen, additional function keys for entering lower case letters and special characters are displayed. • To enter lower case letters, press key F3 – abc... • To enter a special character, press and hold down key F4 – < or F5 – > until the desired special character appears. • Specify the print configuration parameters. • Press key F1 – Save. Returning to the print server 1 configuration To return from the DICOM Settings Print Server 2 menu to the DICOM Settings Print Server 1 menu, do the following: • In the DICOM Settings Print Server 2 menu, place the cursor on the IP addr. DICOM server menu item. • Press key F4 – 20 s))
dicom\pctcp\pctcpsrv.drv file is incompatible
Network unexpectedly shutdown [v11]
Server does not send an ECHO response, but association negotiation is okay
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Application on the server is not running
16-21
16 DICOM Configuration Manual
Error message
Cause(s)
Message service unacceptable for this association [v8]
Wrong mergecom.app
Timed out waiting on network partner [v11]
Server does not recognize ECHO request or STANDARD_ECHO service
STANDARD_ECHO is missing
It may be necessary to restart the server (e.g. HIPAX server) Network partner sent an invalid message [v11]
Server does not recognize ECHO request or STANDARD_ECHO service It may be necessary to restart the server (e.g. HIPAX server)
16-22
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16 DICOM Configuration Manual
DICOM Retrieve The following error messages frequently occur at user level: Response from server
Cause(s)
0xa801
Ziehm NetPort is not properly connected to the server: wrong network settings
0xC000, 0xC001
Ziehm NetPort is not properly connected to the server: wrong DICOM settings, or image data is not stored on the query server
DICOM hardware faults Fault
Remedy
Image memory module fault
Replace image memory module
Defective cables
Replace cables
Visible damage (kinks, pinches, abrasion, cuts) No connection between two correctly functioning components (e.g. hub – hub), simple continuity test can be performed with a multimeter Defective socket(s)
Replace FOC sockets on the monitor cart
Visible damage
If the defective sockets are within the hospital network: Plug the monitor cart into another socket for checking
No connection between correctly functioning components (e.g. hub – cable – socket – server) Cable connection interrupted
Establish cable connection
Cable is not plugged in, cable is defective, socket is defective FOC version: RX and TX cables are swapped
Swap the RX and TX cables
The LINK LED on port 2 of the fiber optic converter is not lit.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
16-23
16 DICOM Configuration Manual
16-24
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
17 DICOM Conformance Statement
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
17 DICOM Conformance Statement
17 DICOM Conformance Statement For more information on DICOM, please refer to our DICOM Conformance Statement.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
17-1
17 DICOM Conformance Statement
17-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 18.1 Exposcop Studio .............................................................................................. 18.1.1 Connection between notebook and C-arm........................................ 18.1.1.1 Location of the service connectors .................................. 18.1.2 Starting Exposcop Studio.................................................................. 18.1.3 Exposcop Studio user interface ........................................................ 18.1.3.1 Main window .................................................................... 18.1.3.2 Virtual keyboard............................................................... 18.1.3.3 Status line ........................................................................ 18.1.3.4 Workspace....................................................................... 18.1.4 Menu bar of Exposcop Studio software ............................................ 18.1.4.1 Exposcop ......................................................................... 18.1.4.2 “Workspace” .................................................................... 18.1.4.3 “Settings” ......................................................................... 18.1.5 “General” tab ..................................................................................... 18.1.6 “Adjustment” tab................................................................................ 18.1.7 “Generator” tab ................................................................................. 18.1.8 “Update” tab ...................................................................................... 18.1.9 “Options” tab ..................................................................................... 18.1.9.1 Option bits........................................................................ 18.1.9.2 Temperature mode setup ................................................ 18.1.9.3 Hand and foot switch assignment.................................... 18.1.10 “Startup Config” tab........................................................................... 18.1.11 “kVmA Tab” tab ................................................................................. 18.1.12 “Errors” tab ........................................................................................ 18.1.13 “Times” tab ........................................................................................ 18.1.14 “DAP” tab .......................................................................................... 18.1.15 “CAN” tab .......................................................................................... 18.1.16 Installation directory .......................................................................... 18.1.17 Update procedure ............................................................................. 18.1.18 Offline mode...................................................................................... 18.1.18.1 Floppy disks for software update ..................................... 18.1.18.2 Options - switch assignment............................................ 18.1.19 Problem correction ............................................................................ 18.2 Integrated Service Functions............................................................................ 18.2.1 General ............................................................................................. 18.2.2 Entering and exiting the Service mode ............................................. 18.2.3 Important keys .................................................................................. 18.2.4 Service programs .............................................................................. 18.2.4.1 Program 0 (software version) .......................................... 18.2.4.2 Program 1 (iris collimator size for format 24 cm) ............. 18.2.4.3 Program 2 (iris collimator size for format 30 cm) ............. 18.2.4.4 Program 3 (iris collimator size for 40 cm format) ............. 18.2.4.5 Program 4 (iris collimator size for i.i. format) ................... 18.2.4.6 Program 5 (iris collimator size for magnification 1).......... 18.2.4.7 Program 6 (iris collimator size for magnification 2).......... 18.2.4.8 Program 7 (startup key adjustment) ................................ 18.2.4.9 Program 8 (program the startup key adjustment) ............
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-1 18-1 18-1 18-1 18-3 18-3 18-4 18-4 18-5 18-6 18-6 18-8 18-9 18-10 18-11 18-13 18-14 18-16 18-16 18-19 18-20 18-21 18-22 18-23 18-24 18-25 18-26 18-27 18-28 18-29 18-30 18-31 18-32 18-33 18-33 18-33 18-33 18-34 18-34 18-34 18-34 18-34 18-35 18-35 18-35 18-35 18-36
i
18.2.4.10 18.2.4.11 18.2.4.12 18.2.4.13 18.2.4.14 18.2.4.15 18.2.4.16 18.2.4.17
ii
Program 9 (adjust filter codes without magnifier)............. Program 10 (adjust filter codes with magnifier 1)............. Program 11 (adjust filter codes with magnifier 2)............. Program 12 (disable or enable radiography mode) ......... Program 13 (disable or enable autotransfer) ................... Program 14 (show fluoroscopy program kV/mA curve) ... Program 15 (show fluoroscopy program table CCD) ....... Program 16 (select crosshair Compact) ..........................
18-37 18-37 18-37 18-38 18-38 18-38 18-38 18-38
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18 Software Tools 18.1
Exposcop Studio
This program is a tool with which you can perform various service procedures on the mobile X-ray systems of the Ziehm 7000, 8000, Compact and Vista series. This requires, however, that you have acquired the necessary skill and knowledge by attending a training at Ziehm Imaging. If this is not the case, you should not try to change the settings of your Ziehm C-arm with this program, but instead request a trained service technician from Ziehm Imaging.
18.1.1
Connection between notebook and C-arm
Communication between notebook and C-arm is established via a CAN interface. There are two possibilities to connect a notebook to the C-arm: • Via a USB-CAN adapter • Via a PARCAN adapter (not available as a new part anymore) The connection must be configured in the Exposcop Studio “Interface Selection” dialog box.
18.1.1.1
Location of the service connectors
The CAN adapter is connected to the hand switch socket.
18.1.2
Starting Exposcop Studio
In order to start the Exposcop Studio, you will find the Exposcop Studio under Start menu > Programs > Ziehm Imaging > Exposcop Studio. It is recommended to create a shortcut on the desktop. When you start the Exposcop Studio software, the GUI (Graphical User Interface) possibly may not start. If you get a window “Connecting the Exposcop”, followed by a message box “Interface”, the USB-CAN or PARCAN adapter cannot be found.
Fig. 18-1
“Interface” message box
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-1
18 Software Tools
Check your cable connection. If the cable is connected correctly, press the “Change” button.
Fig. 18-2
“Interface Selection” dialog box
Select either “PARCAN-Adapter” or “USB-CAN Adapter”. Select the LPT port your notebook uses. The other options refer to cards for stationary computers and are not used for field service. If everything is configured correctly, after pressing “OK” the GUI will start. If no dose measuring system is connected to the computer, enter value -1 in the Dosemeter Port field. After pressing OK, another window may appear, if the adapter cannot communicate with the Carm:
Fig. 18-3
“Exposcop” dialog box
For details on how to solve this problem, → Ch. 18.1.19, p. 18-32, Problem correction
18-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.3
Exposcop Studio user interface
18.1.3.1
Main window
The main window displays a virtual keyboard for the Ziehm C-arm and below that a status line and a workspace. You can select different tabs with the tab labels at the bottom of the workspace for the various service tasks. Please note that you must select an image intensifier size before you can access another tab. Since the dose for the protocol and various default settings depend on this value, an adjustment is only possible after you have made a selection here.
Fig. 18-4
Graphical user interface of Exposcop Studio
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-3
18 Software Tools
18.1.3.2
Virtual keyboard
With the virtual keyboard you can control functions of the Ziehm C-arm that need to be changed frequently when working with the Studio Software. Additionally it shows the current state of the Ziehm C-arm. You can press the keys with the mouse or use the following shortcuts: PC key
Ziehm C-arm key
F2
Thorax program
F3
Pelvis program
F4
Extremities program
F5
DSA
F6
RSA
F7
MSA
F8
Soft
F9
Metal
F10
Magnifier
F11
Half dose
Ctrl+F2
Fluoroscopy
Ctrl+F3
Pulsed fluoroscopy
Ctrl+F4
Snapshot
Ctrl+R
Reset Ziehm C-arm
18.1.3.3
Status line
In the status line you can see information about the selected organ program in plain text. Additionally the number of the used organ table and kV/mA table is displayed. In the organ table the characteristics for AGC, CCD iris and the video adjustment data are stored. The kV/mA table defines the current for a given voltage and so the intensity of X-ray radiation. Also the actual value of the CCD iris is shown.
18-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.3.4
Workspace
The workspace has a number of tabs for the various service procedures you can do with Studio Software. To switch between them, press the corresponding tab label at the bottom. You can also press Ctrl+Tab or Ctrl+Shift+Tab to switch to the next or previous tab. The following tabs are available: − General shows serial numbers, software versions and the error status. − Adjustment allows changing the adjustment data for the selected organ program. − Generator is where you can change the generator settings. − Update allows you to update the software in the Ziehm C-arm. − Options is the tab where you can change the option settings. − Under Startup Config you can change the way the Ziehm C-arm starts up after power-on. − Under kVmA Tab you can view the kV/mA table for the current organ program. − Error shows the error log and allows download and erase the error log − Times provides different timer settings − DAP: dose meter setup − CAN: performs a ping to all CAN modules
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-5
18 Software Tools
18.1.4
Menu bar of Exposcop Studio software
18.1.4.1
Exposcop
Fig. 18-5
Menu “Exposcop”
The “Exposcop” menu contains three items: − “Program module” − “Write RAM signature” − “Exit Exposcop Studio”
18-6
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.4.1.1 Program module... The “Program module...” command opens a window which allows to program single modules manually without machine data conversion and transmission. This is required if a module has lost its software and is not displayed on the “Update” tab anymore.
Fig. 18-6
“Program single module” dialog box
The module name and the required software version (corresponding to a global release) have to be selected. Now programming can be started. After programming, machine data for the module has to be transmitted. Available modules are: − U325
Module 2
− U180
Module 3
− U181
Module 6 Keyboard Ziehm 7000
− U400
Module 6 Keyboard Ziehm 8000 and Vista
− U353
CAN interface Ziehm 7000 Monitor Cart USA
− U385
Module 11 Image Memory
− U501
Video & Dose control
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-7
18 Software Tools
18.1.4.1.2 “Write RAM signature” This function has to be used after battery replacement on the U325 board in order to clear the Warning “E32”.
18.1.4.1.3 “Exit Exposcop Studio” This will terminate the software.
18.1.4.2
“Workspace”
Allows to switch between the different workspaces.
Fig. 18-7
18-8
“Workspace” menu
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.4.3
“Settings”
Fig. 18-8
“Settings” menu
The “Settings” menu contains three items: − “Protocol dose adjustment” This function creates a protocol of the adjustment values and the nominal dose values. − “Protocol Directory” Sets up the directory for the protocol file − “Machine Data directory” The “Machine data directory” command opens a browser window which allows you to set the machine data directory of the Exposcop Studio. This directory is required for storing the downloaded machine data of the unit. − “Interface” Opens the “Interface Selection” dialog box as described above (→ Ch. 18-2, p. 18-2).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-9
18 Software Tools
18.1.5
“General” tab
Fig. 18-9
“General” tab
On the “General” tab, you receive information about: − C-arm type − Serial number of the mobile stand (from GAL on U328 of module 2) − Serial number of the image memory (from GAL on U385) − Global SW version − SW versions and release date and of every CAN bus module − Organ key assignment The global software version displayed is the version of the U325. The software is consistent if all modules except U517 (virtual collimator) show an identical release date. To activate the other tabs of the Exposcop Studio, the image intensifier size has to be selected from the drop-down box. This information is required for creating the protocol file.
18-10
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.6
“Adjustment” tab
Fig. 18-10
“Adjustment” tab
On the “Adjustment” tab you can do the dose and video adjustment for the current organ program. Which organ program is the current one is shown in the status line. You can also see there which organ table number and which kV/mA table are used for this organ program. In the “fluoroscopy program to adjust” combo box, you specify which organ program will be adjusted. The following options are available: current
The current program of the Ziehm C-arm is adjusted. This is the default setting.
Thorax
The “Thorax” organ program is put on the corresponding organ key.
Pelvis
The “Pelvis” organ program is put on the corresponding organ key. The appropriate kV/mA table is transmitted into the Ziehm C-arm.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-11
18 Software Tools
Spine
The “Spine” organ program is put on the “Pelvis” organ key. The appropriate kV/mA table is transmitted into the Ziehm C-arm.To put the original program on the “Pelvis” organ key, select “Pelvis” in the combo box again.
CO2
The Ziehm C-arm is prepared for the adjustment of the CO2 program. You can load default settings for the corresponding organ and kV/mA table. The CO2 program is put to the “Thorax” key for adjustment. Switch back to the normal settings after adjustment by selecting “Thorax” in the combo box.
DSA
The Ziehm C-arm is prepared for the adjustment of the DSA program. You can load default settings for the corresponding organ and kV/mA table. The DSA program is put to the Thorax key for adjustment. Switch back to the normal settings after adjustment by selecting “Thorax” in the combo box.
In the area for dose adjustment you can adjust the AGC values and the settings for the CCD iris. You can use the slider, the +/- keys or the edit field. Please note that any entries in the edit field become effective only after you move away the cursor (e.g. with the tab key). The Ziehm C-arm uses different values for these parameters depending on the state of the functions “Magnifier” and “Half Dose”. The entered values are always used for the current state which is displayed in the status line. Additionally you can select the settings for the measurement area black, white and the control mode in the combo boxes. In the area video adjustment, you set the parameters for white value, black value and the second black value (when the “METAL” key is pressed). As above, you can also use slider, +/- keys or edit field to make your settings. Any changes that you make on this tab are sent to the Ziehm C-arm and become effective immediately. To store the data in the permanent memory (EEPROM) press the “Store” key. Otherwise the data will be overwritten after the next change of the organ program. When you press the “Store Protocol” button, an entry in the adjustment protocol is created, if you have chosen to use a protocol with the “Settings\Protocol adjustment” menu item. Please note that you cannot change some values in the standard edition of the Studio Software. The corresponding fields are dimmed.
18-12
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.7
“Generator” tab
On this tab, you can change the generator settings. Limit values and parameters for the iris, settings for the D/A converter and time for the signal horn can be set.
Fig. 18-11
"Generator" tab
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-13
18 Software Tools
18.1.8
“Update” tab
Fig. 18-12
“Update” tab
The “Update” tab is used to update the CAN module software, to download machine data to a storage medium and to restore machine data. − “New software version”: Combo box for selecting the software version to be programmed. − “Module”: Checked boxes indicate the modules that will be programmed. Modules which require an update are determined by comparing the release dates on the “General” tab. If release dates are not the same for all modules (except U517), the warning E35 will be displayed. For a complete update, all modules have to be checked. Only modules detected in the bootup phase will be displayed. For programming missing modules use “Program module...”. − “Current Version” indicates the currently installed module version. The information will disappear when programming of the respective module has started. − “Read data” downloads the machine data to the selected machine data directory. − “Select data file” allows to select and restore machine data from a storage medium (HD, FD, CD, ....). − “Program” will start the automatic programming procedure. − “Verify program” will compare the installed software to the software files of the Exposcop Studio. − “Write data” will restore machine data files to the C-arm. − “Progress” indicates the programming progress. The upper bar indicates the complete procedure, the lower bar indicates the current step.
18-14
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools − “Current data files” indicates the machine data folder which is currently used. The picture shows the proceeding of the software update.
Fig. 18-13
Update procedure
After a software update the machine parameters have to be reloaded into the Ziehm C-arm. Therefore it is required to specify the desired machine data files before the update. These files will be administered according to the following rules: − When you press the “Select data file” button, you specify a set of machine data files (.mad and .org) that you want to use. A temporary copy will be made of these files. − When you press the “Read data” button, the machine data will be read out of the Ziehm Carm. The directory where they will be stored is specified under the “Settings\Machine data directory ...” menu item. The file names will be created from the serial number of the Ziehm C-arm and the extensions .mad (for machine data) and .org (for organ data). Once this data is known in the Studio Software, the buttons “Program” and “Write data” are enabled. With the “Program” button you start the programming procedure. This will perform all the necessary steps for the software update: − The machine data are converted to fit the desired software version. This means that for new machine data default values will be given and the current machine data will be transformed in the format of the new version. Should this conversion be not possible, the programming will be cancelled in this place before data in the Ziehm C-arm will be changed. − The software will be written to the EEPROM memories of the Ziehm C-arm modules. Only those modules that are checked in the list of modules on the right, will be programmed. However you should always program all modules to avoid compatibility problems. − The Ziehm C-arm will be reset, so that the modules are synchronized. Because the machine data are not yet loaded by this time, various error messages might be displayed which can be ignored. − The machine data that had been converted in the first step are transmitted back to the Ziehm C-arm. Depending on the old and new software version another reset of the Ziehm C-arm might be necessary. Eventually perform this reset by switching the device off and on. Please note that the programming procedure can take more than 30 minutes. Cancelling the programming usually leaves the Ziehm C-arm not functional so that you have to repeat the programming before you can put it into operation.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-15
18 Software Tools
NOTICE Laptop must not run on battery while programming.
18.1.9
“Options” tab
Fig. 18-14
“Options” tab
On the “Options” tab, you can set the device options of the Ziehm C-arm. Device options are the option bits for releasing or modification of detail functions and the hand and foot switch assignment.
18.1.9.1
Option bits
With the option bits certain functions of the Ziehm C-arm can be enabled or modified. Currently bit groups 0, 1, 2 and 3 are in use. With a click on one of the radio buttons you can select the corresponding group. With a mouse click in the check box you can set or clear the corresponding option bit. The changes become effective immediately. The options have the following meaning.
18-16
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
Bit Group “0” Laser on
When the Ziehm C-arm is supplied with a laser light it can be enabled with this bit.
Dose measurement on
When the Ziehm C-arm is supplied with a dose meter it can be enabled with this bit.
X-ray + Alarm
When this bit is set the Ziehm C-arm will always sound during radiation.
Disable Radiography
When this bit is set the Ziehm C-arm will disable Radiography at all.
Enable Magnify 2
The magnification 2 can be enabled with this bit.
X-ray time without radiography time
When this bit is set the Ziehm C-arm will not add the Radiography time to the Fluoroscopy time.
Do not check Compact lamp
When the Ziehm C-arm is a no Compact unit, the lamp check has to be disabled with this bit.
No monitor rotation at power-on
When this bit is set, the image will be displayed in standard (12 o’clock) position after power-up. Otherwise the last rotation setting is preserved.Only for 7000 series.
Bit Group “1” No dose measurement at radiography
When this bit is set, the dose in the Radiography mode will not be added to the total dose metering in the display on the keyboard.
Auto transfer Compact
When this bit is set, a two-image memory system (if installed!) automatically transfers the first image to the second storage position at every beginning of fluoroscopy.
Foot switch for radiography/ snapshot
When this bit is set, you can activate Radiography or Snapshot with the foot switch.
Disable iris collimator
When this bit is set the iris collimator is totally disabled and movements are not monitored (no Errors E21, E22, E23).
Overtemperature MODE 1
→ Ch. 18.1.9.2, p. 18-19, Temperature mode setup
Disable Magnify 1
When this bit is set, the magnification 1 is disabled
Overtemperature MODE 2
→ Ch. 18.1.9.2, p. 18-19, Temperature mode setup
Enable METAL 2
Not used, should not be checked
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-17
18 Software Tools
Bit Group “2” Snapshot: mA × 2
In single shot mode the machine does not use the fixed 8.0mA, but the current from the last fluoroscopy mode is being doubled.
Black level fixed
The black level is not controlled according to the video signal
VCR pause On/Off
The VCR is started and stopped synchronous to radiation.
SOFT toggle
Pressing the SOFT key during SOFT mode switches the Ziehm C-arm back to the last organ program. If this bit is not set, SOFT mode can be left only by pressing an organ key.
New collimator
The virtual collimator (With U517) is enabled with this bit. Standard collimator functions are disabled.
Disable position check
The position check of the new collimator is disabled.
New DAP (VacuDAP 2002)
VacuDAP 2002 installed in the system
Crosshair Compact
Assigns crosshair function to F key (created in Mod 3)
Bit Group “3” Crosshair Compact on at power-up
Crosshair will automatically appear after bootup.
Stepwise image rotation Compact
Enables 90 degree stepwise image rotation on Compact units equipped with flatscreen monitors. (Calculated in Mod. 3)
Compact with Flatscreen
Has to be checked for Compact units equipped with flatscreen monitors. This affects the image orientation.
VacuDAP duo / DAP
Has to be checked for DAP measurement.
VacuDAP duo / Kerma
Has to be checked for air kerma measurement.
Block radiation on DAP overflow
Disables next radiation release if measuring range is exceeded. Press “>0 “Machine data directory”. 4 Select the software version you want to install. 5 All Modules which are marked automatically. 6 Press “Read data” 7 Press “Program” 8 An automatic programming procedure is started. • All selected modules are programmed and verified. • The system is reset automatically • Machine data is transferred. • The system is reset automatically. • A message box “Programming and machine data transfer from the current files successful” appears. • Press “OK”. 9 Programming is finished.
18-28
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.18
Offline mode
In offline mode you create floppy disks for updating the Ziehm C-arm or for assigning the foot switches without a direct CAN bus connection to the device. Because the Studio Software cannot read the data from the device, you first need to provide some information about the Ziehm C-arm for which you create the floppy disks.
Fig. 18-22
Offline mode serial number dialog box
Enter the type of the Ziehm C-arm (Ziehm 7000 or Ziehm 8000 and Ziehm Vista) and the serial number of the mobile stand. In the following dialog box you can choose if you want to create floppy disks for a software update or for changing the foot switch assignment. (Floppy disks for changing the foot switch assignment can be created only for the Ziehm 8000 and Ziehm Vista.)
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-29
18 Software Tools
18.1.18.1 Floppy disks for software update NOTE Programming only will work on a PC with Microsoft® Windows® 98 or older using the PARCAN adapter. To prepare floppy disks for the software update, select the desired software version. If there is any data in your machine data directory for this machine, you can start immediately with the creation of the floppy disks, otherwise the “Create” button is disabled (dimmed) and you must select the data files first. During the creation the software will generate two subdirectories called “Old” and “New” in your directory for the disk images. “New” contains the files for the update. Copy these files (e.g. using the Windows® Explorer) to a floppy disk. To perform the update, connect a PC via a PARCAN adapter to the Ziehm C-arm, insert this floppy disk into the PC and start the update procedure with the command a:\prog. During the update the machine data which have been fitted to the software version will be transmitted, too. In the “Old” subdirectory, you will find a disk image to undo the update process. Execute the undo exactly like the update, i.e. copy the files on to a floppy disk and start prog.bat from this floppy disk on the PC which is connected to the Ziehm C-arm via PARCAN.
Fig. 18-23
18-30
Offline tool for update
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.1.18.2 Options - switch assignment For the Ziehm 8000 and Ziehm Vista you can create a floppy disk set to change the hand and foot switch assignment. Since these floppy disks are used with the HEDIS, you have to enter its serial number. The setting of the assignments done the same way as in the online mode. During the creation a subdirectory called “Switch” will be generated in the directory for disk images. Copy the files from this directory to a floppy disk (e.g. using the Windows® Explorer). You can transfer the switch assignment into the Ziehm C-arm via the floppy disk drive of the monitor cart.
Fig. 18-24
Offline tool for switch programming
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-31
18 Software Tools
18.1.19
Problem correction
With the selected interface no connection to the Ziehm C-arm can be established. Possible causes of this problem are: The Ziehm C-arm is not switched on.
Switch on the Ziehm C-arm and wait until the kV and mA values are displayed on the keyboard of the mobile stand. Then press “Repeat”.
The cables to the Ziehm C-arm are defect, not plugged in or loose.
Check all the cable connections. Eventually replace the cables with new ones. Then press “Repeat”.
An external PARCAN-Adapter has no power supply.
Check the power LED on the PARCAN.
PPARCAN Adapter requires a reset
Please note that the external PARCAN-Adapter has to be eventually reset after a failed connection attempt. To do this unplug the PARCAN power connection, wait for a few seconds and plug it in again. Then press “Repeat”.
CAUTION Radiation might be released during reset of the PARCAN adapter!
If the Studio Software cannot connect the Ziehm C-arm even though it succeeded to initialize the interface, several causes are possible. Possible causes of this problem are: The cables to the Ziehm C-arm are defect, not plugged in or loose.
Check all the cable connections. Eventually replace the cables with new ones. Then press “Repeat”.
The Ziehm C-arm has not finished its startup.
Wait until the kV and mA values are displayed on the keyboard of the mobile stand and try again.
The module “Main Controller” in the Ziehm C-arm does not respond.
The module 2 (U325) has no system program. Select the option “Module Programming” in the dialog box that is being displayed, choose U325 and the required software version. After successful programming you have to turn the Ziehm C-arm off and on again and load the machine data anew.
The Ziehm C-arms breaks the programming procedure after a short time with the message “Programming failed!”.
The reason is in most cases the configuration of the parallel port. This interfaces operates with modern printers in ECP or EPP mode. This does not work with a PARCAN adapter, however. Please set the parallel port back to “Standard Mode” if you get this message. This is done after power-up of your PC in CMOS setup. For details on that refer to the manual from your PC manufacturer.
18-32
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.2
Integrated Service Functions
18.2.1
General
Several service functions are integrated in the Ziehm Vista. These functions can be used without having a service PC.
18.2.2
Entering and exiting the Service mode
1 Switch to fluoro mode. 1 Switch off the laser if the laser is on. 2 Press the “laser” and “snapshot” keys for approx. 2 seconds. 3 press the “fluoroscopy” key until the laser control LED switches off. NOTE Press the sequence of the above described key combination within 3 sec.
The service mode is now enabled. The display shows on the dose display “P 00“. To select the service program use the radiography time up and down keys, for security programs additionally the slot collimator close key. To set the values use the iris open and close keys. To save the new values press the store key (store picture to hard disk). To exit the service mode press the radiography on/off key.
18.2.3
Important keys
laser + snapshot, fluoroscopy
enter service mode
radiography key
exit service mode
radiography time up and down
change service program
radiography time up/down + slot collimator close
SECURITY service program
iris open and close
adjust value
store key (memory)
save changes
NOTE Certain functions are not available with CAN module software 13.06 and above. With CAN module software 13.08/14.03 and above, these functions are not available anymore
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-33
18 Software Tools
18.2.4
Service programs
18.2.4.1
Program 0 (software version)
Shows the installed CAN module software version.
18.2.4.2
Program 1 (iris collimator size for format 24 cm)
(SECURITY) not available with SW13.06 and above Adjust the radiography iris format for 24 cm. Move the iris to the correct size with iris open and close key, then press the store button. The actual value of the size is shown on the radiation time display.
18.2.4.3
Program 2 (iris collimator size for format 30 cm)
(SECURITY) Not available with SW13.06 and above Adjust the radiography iris format for 30 cm. Move the iris to the correct size with iris open and close key, then press the store button. The actual value of the size is shown on the radiation time display. A value less than 20 cancels this function that means the 30 cm format key is disabled.
18.2.4.4
Program 3 (iris collimator size for 40 cm format)
(SECURITY) Not available with SW13.06 and above Adjust the radiography iris format for 40 cm. Move the iris to the correct size with iris open and close key, then press the store button. The actual value of the size is shown on the radiation time display. A value less than 20 cancels this function that means the 40 cm format key is disabled.
18-34
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.2.4.5
Program 4 (iris collimator size for i.i. format)
(SECURITY) Not available with SW13.06 and above Adjust the fluoroscopy iris format for the image intensifier. Move the iris to the correct size with iris open and close key then press the store button. The actual value of the size is shown on the radiation time display.
18.2.4.6
Program 5 (iris collimator size for magnification 1)
(SECURITY) Not available with SW13.06 and above Adjust the fluoroscopy iris format for the image intensifier with magnifier 1 on. (The magnifier will switch on automatically if this program is entered.) Move the iris to the correct size with iris open and close key then press the store button. The actual value of the size is shown on the radiation time display.
18.2.4.7
Program 6 (iris collimator size for magnification 2)
(SECURITY) Not available with SW13.06 and above Adjust the fluoroscopy iris format for the image intensifier with magnifier 2 on. (The magnifier will switch on automatically if this program is entered.) Move the iris to the correct size with iris open and close key then press the store button. The actual value of the size is shown on the radiation time display.
18.2.4.8
Program 7 (startup key adjustment)
Selection of a user defined startup keyboard configuration, i.e. which keys are preselected (prepushed) when the machine is turned on. With iris open and close keys select one of 6 possible configurations (value 0 to 5) Value 0 is the default factory setting.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-35
18 Software Tools
18.2.4.9
Program 8 (program the startup key adjustment)
Before entering this service menu, set the keys which have to come up after the machine is turned on. E.g. press 1/2 Dose a fluoroscopy program of your choice and image reverse left / right, then enter the service mode and step to this program (program 8). Select a program place between 1 and 5. Press the store key. Now enter program 7 and select the proper value you set before. You can store up to 5 different key definitions. The following setups can be programmed: − Values: − kV-Value − radiography time − Keys: − Extremities & Children fluoroscopy program − Pelvis fluoroscopy program − fluoroscopy program thorax − fluoroscopy program soft − Manual Mode function − METAL function − 1/2 Dose function − Magnification function − Vertical image reversal (up/down) function − Horizontal image reversal (left/right) function
NOTE For programming modes snapshot, pulsed fluoroscopy and radiography, use the Exposcop Studio (→ Ch. 18.1.10, p. 18-21).
18-36
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18 Software Tools
18.2.4.10 Program 9 (adjust filter codes without magnifier) Not available with SW13.06 and above Ziehm Compact only Before entering this service menu, set the proper fluoroscopy menu you want to edit. Then enter this menu. To adjust the filter code (0..3, 0 means no filtering), use the iris open and close key to select the filter value. Then press the store key. You can check the result with radiation on already now. If you do not like the result, change the filter code until you are satisfied and press the store key. Now you can leave the program.
18.2.4.11 Program 10 (adjust filter codes with magnifier 1) Not available with SW13.06 and above Ziehm Compact only Before entering this service menu set the proper fluoroscopy menu you want to edit. Do not switch the magnifier on. Then enter this menu. To adjust the filter code (0..3, 0 means no filtering) use the iris open and close key to select the filter value. Then press the store key. You can check the result with radiation on already now. If you don’t like the result change the filter code until you are satisfied and press the store key. Now you can leave the program. This program can be used directly after program 9 (without leaving the service menu).
18.2.4.12 Program 11 (adjust filter codes with magnifier 2) Not available with SW13.06 and above Ziehm Compact only Before entering this service menu set the proper fluoroscopy menu you want to edit. Do not switch the magnifier on. Then enter this menu. To adjust the filter code (0..3, 0 means no filtering) use the iris open and close key to select the filter value. Then press the store key. You can check the result with radiation on already now. If you don’t like the result change the filter code until you are satisfied and press the store key. Now you can leave the program. This program can be used directly after program 9 or 10 (without leaving the service menu).
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
18-37
18 Software Tools
18.2.4.13 Program 12 (disable or enable radiography mode) Not available with SW13.06 and above Use the iris open and close key to enable or disable the radiography mode.
18.2.4.14 Program 13 (disable or enable autotransfer) Not available with SW13.06 and above Ziehm Compact only Use the iris open and close key to enable or disable the autotransfer function. Available only with Compact version and a 2-image memory system.
18.2.4.15 Program 14 (show fluoroscopy program kV/mA curve) Before entering this service menu, set the proper fluoroscopy menu you want to check. By using the service mode counter P14, you can see the number for the kV/mA curve used in the fluoroscopy program chosen such as (standard, pelvis, thorax, heart, soft) or in the subtraction mode (DSA, MSA, RSA). The number cannot be changed or stored.
18.2.4.16 Program 15 (show fluoroscopy program table CCD) Before entering this service menu, set the proper fluoroscopy menu you want to check. By using the service mode counter P14, you can see the number for the fluoroscopy program table CCD chosen such as (standard, pelvis, thorax, heart, soft) or in the subtraction mode (DSA, MSA, RSA). The number cannot be changed or stored.
18.2.4.17 Program 16 (select crosshair Compact) Ziehm Compact only Enables the crosshair on a Compact unit (SW 13.03 and above). Use the “open iris” and “close iris” keys to enable or disable the crosshair function. The Crosshair function will be assigned to the “F” key.
18-38
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19 Adjustment Instructions
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
Contents 19.1 Dose & Video Adjustment ................................................................................ 19.1.1 Dose Adjustment............................................................................... 19.1.1.1 Dose adjustment parameters............................................. 19.1.1.2 Dose adjustment procedure............................................... 19.1.1.3 Special cases for dose adjustment .................................... 19.1.2 Video Adjustment .............................................................................. 19.1.2.1 Video adjustment parameters............................................ 19.1.2.2 Video adjustment procedure.............................................. 19.1.3 Exposcop Studio Dialog for Dose Adjustment .................................. 19.1.4 kV/mA Values for Ziehm Vista / Ziehm 8000 .................................... 19.1.5 kV/mA Values for Ziehm Vista Plus .................................................. 19.1.6 Dose Tables for Ziehm Vista / Ziehm 8000....................................... 19.1.7 Dose Tables for Ziehm Vista Plus..................................................... 19.1.8 Ziehm Vista / Ziehm 8000 Organ Tables 31/23cm............................ 19.1.9 Ziehm Vista / Ziehm 8000 Organ Tables 17/15/11 cm...................... 19.1.10 Ziehm Vista Plus Organ Tables ........................................................ 19.2 Default Values .................................................................................................. 19.2.1 Measuring Circle Sizes ..................................................................... 19.2.2 Measuring Circle Sizes Vista Plus .................................................... 19.2.3 Monitor Cart Settings ........................................................................ 19.2.3.1 Step windowing levels ....................................................... 19.2.3.2 DSA windowing values ...................................................... 19.2.3.3 Filter factors non interlaced ............................................... 19.2.3.4 Filter factors interlaced ...................................................... 19.2.3.5 Filter factors Vista Plus ......................................................
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19-1 19-1 19-1 19-1 19-2 19-3 19-3 19-3 19-4 19-6 19-7 19-8 19-9 19-10 19-11 19-12 19-13 19-13 19-13 19-14 19-14 19-14 19-15 19-15 19-16
i
ii
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19 Adjustment Instructions 19.1
Dose & Video Adjustment
Dose adjustment has to be performed prior to the video adjustment. The adjustment of dose is performed with 25 mm Al + 1.5 copper put onto the generator. The dose is measured in front of the image intensifier grid. The dose values have to be taken from the dose table. You can adjust dose with the following parameters:
19.1.1
Dose Adjustment
19.1.1.1
Dose adjustment parameters
− Iris collimator − This value influences the actual Iris CCD collimator. A smaller value leads to a higher dose and vice versa. − AGC − This value influences the internal gain of the CCD camera −
Increasing the value will increase the dose.
19.1.1.2
Dose adjustment procedure
• Set to Full Dose • Set iris collimator and “AGC full dose” to values given in the Organ Table. • Adjust to full dose value by setting the iris collimator in the specified range • If dose level is not adjustable then change “AGC full Dose” value. • Set to Half dose • Adjust to half dose value by changing “AGC half dose” value
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19-1
19 Adjustment Instructions
19.1.1.3
Special cases for dose adjustment
− Half Dose: Dose has to be adjusted at full dose first. By activating half dose, the iris collimator value is blocked. Half dose value adjustment is only done by changing “AGC half dose” value − Snapshot: Snapshot is a modifier to each application program and is adjusted via “Iris collimator snapshot”. Measure the accumulated dose for one shot. Do not measure the first shot, make sure that regulation has found its target before. Dose has to be below 2uG. − DSA / MSA / RSA: Assign the DSA function to the thorax key to do the adjustments without the DSA functionality of the Image Memory. → Ch. 18.1.6, p. 18-11 ff. After Adjustment assign the prior function to the thorax key again.
19-2
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19 Adjustment Instructions
19.1.2
Video Adjustment
19.1.2.1
Video adjustment parameters
− Black level: − influences the black level of the video signal. − increasing the value will increase the black level. − White level: − influences the white level of the video signal. − increasing the value will increase the white level. − 2. Black level: − sets up a second black level if metal key is selected. − increasing the value will increase the second black level.
19.1.2.2
Video adjustment procedure
• Put the grey step wedge into the optical path so that both angles of the grey step wedge (black) just contact the upper mask circle: Position
II Active area
Type and distance of the step wedge
31 cm: (N with 31 cm II)
large, about 30 cm to II
23 cm: (N with 23cm II) (M1 with 31cm II)
large, about 10 cm to II
15 cm: (M1 with 23cm II) positioning at N
large, about 10 cm to II
(M2 with 31cm II) positioning at M1 Fig. 19-1
Grey Step Wedge
11 cm: (M2 with 23cm II) positioning at M1
small, about 10 cm to II
• Use histogram function in order to measure the video signal. The histogram can be activated with the histogram function that appears in the menu Service > System Settings • Adjust the Black level as described in the video adjustment table. • White level is adjusted in all organ programs to full range without clipping. Exception: In Extremities and Pelvis/Spine mode it is set like indicated in the organ table. • Adjust video Metal level. Therefore take a proper reference image without metal in each organ program. After getting metal in and pressing the metal switch there should be no difference
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19-3
19 Adjustment Instructions
within the bone grayscale values of the image with and without metal.
CAUTION Before making any adjustments not described here, please contact our Service Department.
19.1.3
Exposcop Studio Dialog for Dose Adjustment
Fig. 19-2
Exposcop Studio dialog for dose adjustment
19-4
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19 Adjustment Instructions
Name
Description
Dose adjustment AGC full/half dose
This value is setting the dose with respect to the analog gain in CCD camera. Increasing the value will increase the dose. The setting (full / half dose) offered depends on the selected mode.
Iris collimator
This value is setting the dose with respect to the CCD iris. Increasing the value will decrease the dose. This setting is only done in full dose mode, it cannot be changed at half dose.
Iris collimator (snapshot)
This value is setting the snapshot dose with respect to the CCD iris. Increasing the value will decrease the dose. Snapshot only works at full dose, no half dose setting is required.
mA Factor Half Dose
Sets up the factor for reduced mA in half dose mode. kV/mA curve will be modified in half dose mode if factor is different from 100.
Control mode
Sets up average or peak value measurement for dose regulation
Measurement area black
Measurement area for video black level regulation
Measurement area white
Measurement area for dose regulation
Blanking
Size of the blanking circle generated in the module 3
Video Gamma
Sets up the gamma correction value.
Black level
0–255: This is to adjust the black level within the image based on the black measurement area.
White level
0–255: This is to adjust the white level within the image based on the whole image.
Black Limit
Limits the range of the black level regulator. Set to 255 for full range
White Limit
Limits the range of the white level regulator. Set to 0 for full range
Filter
Sets up the filter factor for Compact units.
Shading correction off
Switches off the vignette correction.
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19-5
19-6
+F
Key
2
3
0
5
4
4
Spine
Heart
Soft
DSA, RSA, MSA
CO2 DSA, RSA, MSA
Organ code
Extremities
Program
2
3
4
0
1
5
kV/mA Table kV
0.2
0.2
0.2
0.2
0.2
0.2
40
0.4
42
0.7
45
kV/mA Values for Ziehm Vista / Ziehm 8000
According to test instruction release B0.1
19.1.4
0.4
50
0.7
55
6.0
56
58
0.1
0.4
3.0
60
63
6.0
65
70
3.8
4.0
80
6.0
90
6.0
100
5.6
4.7
4.7
3.1
5.5
3.2
110
19 Adjustment Instructions
mA
Ziehm Vista P_26363/CD_28115 - 09/03/2010
Key
2
3
1
5
4
Pelvic
Heart
LPD
DSA, RSA, MSA
Organ code
Extremities
Program
Ziehm Vista P_26363/CD_28115 - 09/03/2010
3
4
2
1
5
kV/mA Table kV
0.2
0.2
0.2
0.2
0.2
40
42
kV/mA Values for Ziehm Vista Plus
According to test instruction release B
19.1.5
45
50
1.0
55
56
58
6.0
1.0
60
6.0
6.0
63
65
70
4.5
4.6
80
6.5
83
100
5.4
5.4
5.0
5.4
5.4
110
19 Adjustment Instructions
mA
19-7
19 Adjustment Instructions
19.1.6
Dose Tables for Ziehm Vista / Ziehm 8000
According to test instruction release B0.1 All values in µGy/s (allowed tolerance ± 10%) Caution max. allowed limit value is 0.6 µGy/s for AVR and 1.2 µGy/s for PVR 1/2 Dose
1/1 Dose
regulation mode
measured
calculated value for dose label (if required)
measured
calculated value for dose label (if required)
31cm
AVR
0.18
0.18
0.36
0.36
23cm
AVR
0.22
0.22
0.43
0.43
17 / 15cm
AVR
0.22
0.22
0.43
0.43
10cm
PVR
0.55
0.28
1.10
0.55
AVR
0.18
0.18
0.36
0.36
Organ Program
Extremities
Spine 31cm 23cm
AVR
0.20
0.20
0.40
0.40
17 / 15cm
AVR
0.20
0.20
0.40
0.40
10cm
PVR
0.40
0.20
0.80
0.40
31cm
AVR
0.23
0.23
0.45
0.45
23cm
AVR
0.28
0.28
0.56
0.56
17 / 15cm
AVR
0.28
0.28
0.56
0.56
10cm
PVR
0.55
0.28
1.10
0.55
31cm
AVR
0.18
0.18
0.36
0.36
23cm
AVR
0.18
0.18
0.36
0.36
17 / 15cm
AVR
0.18
0.18
0.36
0.36
10cm
PVR
0.55
0.28
1.10
0.55
31cm
AVR
-
-
0.36
0.36
23cm
AVR
-
-
0.54
0.54
17 / 15cm
AVR
-
-
0.54
0.54
10cm
PVR
-
-
1.10
0.55
31cm
AVR
-
-
0.6
0.6
23cm
AVR
-
-
0.6
0.6
17 / 15cm
AVR
-
-
0.6
0.6
10cm
PVR
-
-
1.30
0.65
Heart
Soft
DSA, MSA, RSA
CO2
According to service information 5/16 - 21.09.05 Snapshot is adjusted to 1.6 µGy in all modes and image formats
19-8
Ziehm Vista / Ziehm 8000 P_26363/CD_28115 - 09/03/2010
19 Adjustment Instructions
19.1.7
Dose Tables for Ziehm Vista Plus
According to test instruction release B
Organ Program
regulation mode
1/1 Dose
1/2 Dose
max Dose
measured [µGy/s]
measured [µGy/s]
[mGy/min]
Extremities 23cm
AVR
0.65 (0.82)
0.35 (0.32)