CompoMat G5 Service Manual Edition: 3/11.10 Part. no.: M67 599 1 0123



Table of Contents 1



Index



2



Important Information



3



2.1



How to use the Service Manual................................................................................................ 2-1



2.2



Significance of the warnings.................................................................................................... 2-1



2.3



Significance of the note ............................................................................................................ 2-2



2.4



Significance of the tip ............................................................................................................... 2-2



2.5



Technician's qualification......................................................................................................... 2-2



2.6



Instructions to be observed when servicing the system....................................................... 2-2



2.7 2.7.1 2.7.2 2.7.3 2.7.4 2.7.5



Warnings .................................................................................................................................... 2-3 Basic warnings ............................................................................................................................ 2-3 Warnings regarding electric hazards ........................................................................................... 2-4 Warnings regarding mechanical hazards .................................................................................... 2-5 Warnings regarding risk of burning.............................................................................................. 2-5 Warnings regarding biological hazards ....................................................................................... 2-6



2.8



Addresses .................................................................................................................................. 2-7



Installation 3.1



Preface........................................................................................................................................ 3-1



3.2



Important information on initial start-up ................................................................................. 3-1



3.3 3.3.1



Installing the optional scales ................................................................................................... 3-3 Installing the PLS scales ............................................................................................................. 3-4 3.3.1.1 Mounting the PLS scales ............................................................................................. 3-4 Installing the RCC scales ........................................................................................................... 3-5 3.3.2.1 Mounting the RCC scales............................................................................................. 3-6



3.3.2



4



3.4



Connector layout ....................................................................................................................... 3-8



3.5



Initial start-up........................................................................................................................... 3-10



3.6



Decommissioning / removal from service / recommissioning............................................ 3-19



Specifications



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



0-i



5



Service mode 5.1



Accessing service mode when the device is switched off .................................................... 5-1



5.2 5.2.1 5.2.2 5.2.3



Accessing service mode from standby ................................................................................... 5-1 Self-test........................................................................................................................................ 5-2 Entering the password ................................................................................................................. 5-2 Main menu (Service).................................................................................................................... 5-2 5.2.3.1 Entering the serial number ........................................................................................... 5-3 5.2.3.2 Entering an E-code....................................................................................................... 5-4 5.2.3.3 Update/Init ARM7 ......................................................................................................... 5-5 5.2.3.4 Update/Init Trinamic ..................................................................................................... 5-5 5.2.3.5 Calibrating ADC & Heads............................................................................................. 5-6 Special functions (extended menu) ............................................................................................. 5-8 Application service mode............................................................................................................. 5-8 Program selection........................................................................................................................ 5-9 5.2.6.1 Display and operation of a service program ............................................................... 5-10



5.2.4 5.2.5 5.2.6 5.3 5.3.1 5.3.2



Seal & weighing (program-independent)............................................................................... 5-11 Program-independent sealing.................................................................................................... 5-12 Program-independent weighing................................................................................................. 5-12



5.4 5.4.1



5.4.5 5.4.6



Calibration ................................................................................................................................ 5-13 Calibrating top press.................................................................................................................. 5-13 5.4.1.1 Top press (calibrating press force)............................................................................. 5-14 5.4.1.2 Top press (calibrating press stroke) ........................................................................... 5-16 Calibrating the upper press........................................................................................................ 5-16 5.4.2.1 Upper press (calibrating the press force) ................................................................... 5-17 5.4.2.2 Upper press (calibrating press stroke) ....................................................................... 5-19 Calibrating the lower press ........................................................................................................ 5-19 5.4.3.1 Lower press (calibrating the press force) ................................................................... 5-20 5.4.3.2 Lower press (calibrating the press stroke) ................................................................. 5-22 Calibrating the scales ................................................................................................................ 5-22 5.4.4.1 Calibrating the press scales ....................................................................................... 5-22 5.4.4.2 Calibrating the RCC scales ........................................................................................ 5-23 5.4.4.3 Calibrating the PLS scales ......................................................................................... 5-24 Calibrating detector A ................................................................................................................ 5-25 Calibrating ADC & Heads (see chapter 5.2.3.5 on page 5-6).................................................... 5-26



5.5 5.5.1 5.5.2 5.5.3



Cleaning the press area .......................................................................................................... 5-26 Cleaning the upper press........................................................................................................... 5-27 Cleaning the lower press ........................................................................................................... 5-28 Cleaning the slide ...................................................................................................................... 5-29



5.6 5.6.1 5.6.2 5.6.3 5.6.4 5.6.5



Setup......................................................................................................................................... 5-30 Language selection.................................................................................................................... 5-30 Process data.............................................................................................................................. 5-31 Software update......................................................................................................................... 5-31 Barcode test............................................................................................................................... 5-33 System information .................................................................................................................... 5-34



5.4.2 5.4.3 5.4.4



0-ii



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6



7



8



TSC / TMC / MA 6.1



Important information for the procedure ................................................................................ 6-1



6.2



Test report - annual maintenance ............................................................................................ 6-2



6.3



Checking the electrical safety ................................................................................................ 6-12



6.4 6.4.1 6.4.2



Explanations on the reports ................................................................................................... 6-14 Explanations on the reports "Initial start-up" and "Annual maintenance" .................................. 6-14 Explanations on electrical safety checks ................................................................................... 6-36



Error Messages 7.1



CompoMat G5 Error Codes ...................................................................................................... 7-1



7.2 7.2.1 7.2.2 7.2.3 7.2.4



Actuator, action and error codes ............................................................................................. 7-4 Example....................................................................................................................................... 7-4 Error codes .................................................................................................................................. 7-4 Actuator codes............................................................................................................................. 7-6 Action codes ................................................................................................................................ 7-8



Calibration / adjustment 8.1



9



Calibration (see chapter 5.4 on page 5-13) ............................................................................. 8-1



Servicing / repair 9.1



Description / warnings .............................................................................................................. 9-1



9.2



Electrical hazards ...................................................................................................................... 9-2



9.3



Mechanical hazards................................................................................................................... 9-3



9.4



Risk of burning .......................................................................................................................... 9-3



9.5



Biological hazards..................................................................................................................... 9-4



9.6 9.6.1 9.6.2 9.6.3 9.6.4 9.6.5 9.6.6 9.6.7 9.6.8 9.6.9 9.6.10 9.6.11 9.6.12 9.6.13 9.6.14 9.6.15



Replacements / Repairs ............................................................................................................ 9-5 Replacing the Application Board bought-in assembly (M68 137 1)............................................. 9-5 Replacing the Display G5 printed circuit board bought-in assembly (M68 115 1)....................... 9-7 Replacing the Trinamic Board bought-in assembly ..................................................................... 9-8 Changing the top press silicon mat G5 (M68 129 1) ................................................................... 9-9 Replacing the BG-CompoFlow Opener Top and door............................................................... 9-10 Replacing the BG-CompoFlow Opener RCC ............................................................................ 9-14 Testing the BG-CompoFlow Opener ......................................................................................... 9-16 Replacing the Heads ................................................................................................................. 9-18 Replacing Head 3 (PLS scales)................................................................................................. 9-22 Replacing the weighing cell RCC (M63 526 1).......................................................................... 9-24 Testing the weighing cell RCC .................................................................................................. 9-27 Replacing the weighing cell PLS (M63 526 1)........................................................................... 9-27 Testing the weighing cell PLS ................................................................................................... 9-30 Replacing the scales 3 (M68 109 1) .......................................................................................... 9-30 Testing scales 3......................................................................................................................... 9-33



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



0-iii



9.7 9.7.1 9.7.2 9.7.3 9.7.4



Tests after the replacement of components ......................................................................... 9-34 BG-CompoFlow Opener height measurement .......................................................................... 9-34 BG-CompoFlow Opener installation .......................................................................................... 9-35 BG-CompoFlow Opener opening .............................................................................................. 9-35 BG-CompoFlow Opener height assembly ................................................................................. 9-35



10 Passwords 10.0.0.1 Service password ....................................................................................................... 10-1 10.0.0.2 Default settings for passwords in CompoMaster Net G5 .......................................... 10-1 10.0.0.3 Defining individual passwords in CompoMaster Net G5........................................... 10-1



11 Functional Description 11.1



Description of the therapy ...................................................................................................... 11-1



11.2



Physical Description ............................................................................................................... 11-1



11.3 11.3.1 11.3.2 11.3.3 11.3.4 11.3.5 11.3.6



Functional description ............................................................................................................ 11-1 Application Board bought-in assembly (M68 137 1) .................................................................. 11-1 Display G5 printed circuit board bought-in assembly (M68 115 1) ............................................ 11-1 HF generator 24V bought-in assembly (M66 178 1).................................................................. 11-2 Relay box G5 (M68 107 1)......................................................................................................... 11-2 Sealing Head El.Mag. Drive assembly (M68 144 1) .................................................................. 11-2 Trinamic Board bought-in assembly (M68 138 1)...................................................................... 11-2



11.4



Block diagram .......................................................................................................................... 11-3



11.5 11.5.1 11.5.2 11.5.3 11.5.4



Component layouts ................................................................................................................. 11-4 Application Board bought-in assembly ...................................................................................... 11-4 Display G5 printed circuit board bought-in assembly ................................................................ 11-5 LP1361 ...................................................................................................................................... 11-5 Trinamic Board bought-in assembly .......................................................................................... 11-6



11.6



Flow diagram (Service Menu) ................................................................................................. 11-7



12 Spare parts catalog 12.1



0-iv



Tools and accessories .......................................................................................................... 12-21



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 1: Index



1



Index



A



R



Accessories and tools 12-21



Replacing Application Board bought-in assembly 9-5



B



Replacing Display G5 printed circuit board bought-in assembly 9-7



Biological hazards 9-4 Block diagram 11-3



Replacing the Trinamic Board bought-in assembly 9-8



C



Risk of burning 9-3



Calibration / adjustment 8-1 Checking the electrical safety 6-12



S Scales, installing 3-3



D



Sealing seam, test criteria 6-30



Default passwords 10-1



Service mode 5-1 Service password 10-1 Servicing / repair 9-1



E Electrical safety inspections (explanation) 6-36 Error messages 7-1



F Functional description 11-1 Fuses 6-36



Spare parts catalog 12-1 Specifications 4-1



T Test report - annual maintenance 6-2 Test report explanation 6-14 Tools and accessories 12-21 TSC / TMC / maintenance 6-1



H Hazards (electric) 9-2 Hazards (mechanical) 9-3 Head 3, replacing (PLS scales) 9-22



W Warning description 9-1



Heads, replacing 9-18



I Important information 2-1 Initial start-up report 3-10 Installation 3-1



P Passwords 10-1



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



1-1



Chapter 2: Important Information



2



Important Information



2.1



How to use the Service Manual



2.2



Identification



The document can be identified by the following information on the title page and on the labels, if any: – Edition of the technical document – Part number of the technical document



Page identification



The page identification 1-3, for example, refers to Chapter 1, page 3.



Editorial information



The editorial information 3/11.10, for example, refers to: 3rd Edition, November 2010.



Organization of the chapters



To facilitate the use of documents from Fresenius Kabi, the organization of the chapters has been standardized in all manuals. There may therefore be chapters within this document without any content. Chapters without content are identified.



Illustrations



The illustrations used in the documents (e.g. screens, photos, etc.) may differ from the original if this does not have any influence on the function.



Document changes



Changes to the technical document will be released as new editions or supplements. In general, this manual is subject to change without notice.



Reproduction



Reproduction, even of extracts, may only be performed with written authorization.



Significance of the warnings Warning Advises the operator that incorrect operation can result in personal injury. Caution Advises the operator that incorrect operation can cause damage to the equipment.



2-1



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 2: Important Information



2.3



Significance of the note Note: Information alerting the operator to the fact that failure to follow the steps as specified may result in the specific function not being executed correctly, not being executed at all, or not producing the desired effect.



2.4



Significance of the tip Tip Information providing useful tips for easy handling.



2.5



2.6



Technician's qualification Purpose



This technical document is intended for service technicians and can be used for first studies (to acquire a basic knowledge) and for reference purposes (for TSC, maintenance and repair). The study of this document, however, does not replace the training courses offered by the manufacturer.



Requirements



Knowledge of the current Operating Instructions of the respective device. Background experience in mechanics, electrical and medical engineering.



Instructions to be observed when servicing the system Authorized persons



Assembly, extensions, adjustments, modifications or repairs may only be undertaken by the manufacturer or any person authorized by the manufacturer.



Test equipment and accessories



The activities described in this technical document require the availability of the necessary technical measuring equipment and accessories.



Specifications



Observe the information on the specifications.



Precautions



Before turning power on, repair any visible damage. Prior to opening the device and when working on the open device, the following precautions have to be observed:



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



2-2



Chapter 2: Important Information



– Protect the components against ingress of fluids. – Do not touch live parts. – Disconnect and connect all jacks, connectors and components only when the device is turned off.



2.7



Warnings



2.7.1



Basic warnings



Caution The CompoMat G5 may only be opened by qualified and skilled staff. Caution Operators are wholly responsible for any processes that they create in CompoMaster Net G5 and which are executed by the CompoMat G5. This applies in particular to possible loss of blood and blood products by incorrect programming. See the User Manual for CompoMaster Net G5 for information about using the CompoMaster Net G5 to program the CompoMat G5. Caution Operators are wholly responsible for preparatory tasks, such as storage, centrifuging, transport to the CompoMat G5, suspending on the pins and snapping any possible rated break points, all of which can have a significant influence on the quality of separation of the blood components. For these reasons, Fresenius Kabi does not provide information in advance concerning the quality of results as regards blood components obtained using CompoMat G5 processes. Fresenius Kabi only claims that there is a relatively high reproducibility of results for processed blood components within a relatively short preparation time. Please refer to the relevant literature for specific quality results for blood components processed with the CompoMat G5 in daily practice at a blood bank.



2-3



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 2: Important Information



2.7.2



Warnings regarding electric hazards



Warning Risk of injury caused by electrical voltage. When connecting the CompoMat G5 to a mains supply, the local regulations that apply have to be observed. Warning Risk of injury caused by electrical voltage. Improper commissioning and use of the electrical equipment can result in injuries, such as burns and electrical shocks. Warning Risk of injury caused by electrical voltage. When using safety class I systems, the quality of the protective ground of the installation is of particular importance. It must be taken into consideration that in many countries regulations have been enacted by the national authorities. The rear panel of the CompoMat G5 is provided with a connection for a grounded power cable. Warning Risk of injury caused by electrical voltage. If the power cable needs to be replaced, use only the original power cable listed in the spare parts catalog. If additional equipment, which is not included in the accessories, is connected to the device, there will be a danger that the permissible leakage currents will be exceeded. Warning Risk of injury by burning The devices are not suitable for use with flammable mixtures or nitrous oxide. Caution Loss of the preparation when disconnecting the power plug or in the event of a power failure. If there is a loss of power while processing a program in the CompoMat G5, all sealing heads/clamps will open and the presses will stop. This may cause a loss of the preparation.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



2-4



Chapter 2: Important Information



2.7.3



Warnings regarding mechanical hazards



Warning Risk of injury caused by parts moving autonomously. To avoid injury, operators must ensure to keep a sufficiently large distance to all moving parts of the CompoMat G5 . Moving parts are: – – – – – –



Door upper press lower press slide Top press CF-Opener



Warning Risk of injury by crushing Keep hands clear of the inside of the CompoMat G5 during operation of the device.



2.7.4



Warnings regarding risk of burning



Warning Risk of burning by touching the sealing electrodes – Do not touch the sealing electrodes of Heads 1-6 during the sealing process. – Do not place any live parts into the Heads. – Maintain a sufficient distance from the sealing electrodes in the Heads.



2-5



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 2: Important Information



2.7.5



Warnings regarding biological hazards



Warning Risk of infection by leaking blood bags It is always possible that donated blood could be infected by germs of contagious diseases. The blood must therefore always be treated as being potentially infectious. – Wear surgical or similar gloves when processing blood products. – Immediately clean and disinfect work surfaces that are contaminated with blood or blood products. Local laws and regulations concerning the handling of potentially infectious material must be observed at all times. Warning Risk of infection by leaking tubes – The operator should not exert any force on tubing, which is currently being sealed.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



2-6



Chapter 2: Important Information



2.8



Addresses Please address any inquiries to: Manufacturer



Fresenius Kabi AG 61346 Bad Homburg v.d.H. Germany Phone: +49 (0)6172 / 608 - 0



Service Central Europe



Fresenius HemoCare GmbH Technischer Service Pfingstweide 53 D-61169 Friedberg Germany International Service Hotline (08:30 - 17:00) Tel: +49 (0)6172 / 608 8469 Fax: +49 (0)6172 / 608 8539 Email: [email protected]



Service International



Responsible Regional Organization (Technical Service)



Local Service



2-7



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



3



Installation



3.1



Preface Instructions for all technicians authorized to commission our systems We, as manufacturers, permanently aim at delivering systems of highest quality. To reach this aim, we need your support. Please commission our devices uniformly using the enclosed "initial start-up report" and enter the values determined in the columns provided. The following applies: Corrections are necessary only if the measured values are outside of the specified tolerances!



3.2



Important information on initial start-up For initial start-up only



This technical document is intended for initial start-up only.



Environmental conditions



Variations in temperature during transport may lead to condensation water developing on conducting parts. In the event of major variations in temperature, the device must be given enough time to adapt to the ambient temperature before start-up.



Tester's qualification



The initial start-up must be performed by the Technical Service of Fresenius Kabi or a person authorized by them. The initial start-up procedure may only be performed by persons qualified to properly perform the specified checks on the basis of their educational background and training, as well as knowledge and experience gained in practice. Furthermore, the persons performing the tests must not be bound by any directives when performing this activity.



Test equipment and accessories



The activities described in this technical document require the availability of the necessary technical measuring equipment and accessories.



Specifications



Observe the information on the specifications.



Precautions



Before turning power on, repair any visible damage. Prior to opening the device and when working on the open device, the following precautions have to be observed: – Protect the components against ingress of fluids. – Do not touch live parts. – Disconnect and connect all jacks, connectors and components only when the device is turned off.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-1



Chapter 3: Installation



TSC/TMC/MA intervals TSC



Technical safety check Note: There is no TSC obligation for the CompoMat G5.



TMC



Technical measurement check Note: There is no TMC obligation for the CompoMat G5 .



MA



Servicing In order to avoid personal injury and damage to property, an annual maintenance must be carried out. (see chapter 6 on page 6-1) Once the CompoMat G5 has been opened, the Electrical Safety Check must be carried out before recommissioning the device. (see chapter 6.3 on page 6-12)



3-2



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



3.3



Installing the optional scales Warning Risk of injury when assembling the scales and when tilting the CompoMat G5. Electric shock – Turn CompoMat G5 off and disconnect the power plug. Caution Risk of damaging the CompoMat G5 when assembling the scales. Overvoltage and/or short-circuit – Turn CompoMat G5 off and disconnect the power plug. Note: Improperly-mounted scales can lead to incorrect weighing results. – A visible air gap must be present between the CompoMat G5 housing and the scales housing. – The connection cable must not touch the scales housing. Note: To ensure correct weighing results, the scales must be calibrated after mounting. (see chapter 5.4 on page 5-13) CompoMat G5 Tilting the



Fresenius Kabi



CompoMat G5



SM-EN



To facilitate mounting, the CompoMat G5 can be placed on its back (see the following illustrations).



3/11.10



3-3



Chapter 3: Installation



3.3.1



Installing the PLS scales PLS scales kit



Unpack the PLS scales Kit (902 552 1) and check that it is complete and has not been damaged in transit: 1 x PLS scales 3 x Allen screws 3.3.1.1



Mounting the PLS scales



CompoMat G5 being positioned on its rear.



1



– Place the PLS scales flush with the housing and use the three Allen screws supplied to secure them to the mounting points (1). Laying the connection cables 1 2



3



– Connect the BNC plug (1) for Head 3 to the BNC port (2). – Secure the scales connecting cable using the cable clip (3).



3-4



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



Caution Damage to the scales connecting cable caused by incorrect routing. Crushing hazard – The scales connecting cable must be passed through the brackets (see following illustration).



4



– Connect the PLS scales to the Balance A (4) connector. – Calibrate the PLS scales (see chapter 5.4 on page 5-13). – Check calibration of the PLS scales. – Check Head 3.



3.3.2



Installing the RCC scales RCC scales kit



Unpack the RCC scales Kit (902 551 1). Check that it is complete and has not been damaged in transit. 1 x RCC scales 3 x Allen screws



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-5



Chapter 3: Installation



3.3.2.1



Mounting the RCC scales



CompoMat G5 being positioned on its rear.



1



– Place the RCC scales flush with the housing and secure it with the three Allen screws (1) provided. – Route the scales connecting cable (1) and secure it using the two cable clips (2) (see following illustration). Routing the connecting cable 1



2



Caution Damage to the scales connecting cable caused by incorrect routing. Crushing hazard – The scales connecting cable must be passed through the brackets (see following illustration).



3-6



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



5



– Connect the RCC scales to the Balance B (5) connector. – Calibrate the RCC scales (see chapter 5.4 on page 5-13). – Check calibration of the RCC scales. – CF-Opener Check RCC scales.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-7



Chapter 3: Installation



3.4



Connector layout Caution Damage to the device when connecting the scales. Short circuit and/or overvoltage – Scales must only be connected when the device is switched off. View: CompoMat G5 from below



4/5



3/6



2



1



The allocation of the numbers complies with the following schematic view. Schematic view (from below): To rear 5



6



4



3



2



1



To front



(1) Network connection (LAN) (2) Two equal-priority USB ports (3) Scanner port (4) Connector for PLS scales (5) Connector for RCC scales (6) RFID scanner (pre-allocated, see following note) Note: The RFID (6) Scanner interface is allocated for future applications. At the moment no RFID scanner has been approved for use.



3-8



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



View: CompoMat G5 tilted 90° to the rear, with front facing up



1



2



3/6



4/5



The allocation of the numbers complies with the following schematic view. To front 1



2



3



4



6



5



To rear



(1) Network connection (LAN) (2) Two equal-priority USB ports (3) Scanner port (4) Connector for PLS scales (5) Connector for RCC scales (6) RFID scanner (pre-allocated, see following note) Note: The RFID Scanner (6) interface is allocated for future applications. At the moment no RFID scanner has been approved for use.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-9



Chapter 3: Installation



3.5



Initial start-up Note: A detailed description of the individual items can be found under chapter Explanations of reports (see chapter 6.4 on page 6-14)



CompoMat G5



Initial start-up report



Technician's name:



Customer/customer no.:



Version:



Optional:



Device no.:



Service report no.:



RCC scales: Serial number: PLS scales: Serial number:



Equipment code:



No.



Description



1



Unpacking (only for initial start-up!)



1.1



Unpack the device.







1.2



Check that all parts of the device have been accounted for.







1.3



Device without visible shipping damage.







2



Visual inspection



2.1



Check the fuses on the power input panel jack. Melting point of the fuses (2x) T 10 A / 230/240 V ~AC.



2.2



Labels and inscriptions are present and legible.



2.3



Mechanical condition permits safe use.



2.4



No contamination or damage.



2.5



The power cable is not damaged.



3-10



Correct.



Fresenius Kabi



CompoMat G5



Meas. value



SM-EN



3/11.10



Chapter 3: Installation



No.



Description



Correct.



3



Device-related inspections



3.1



Calibration of ADC and the head sensors Main menu (Service) / Cal. ADC & Heads



Meas. value



Yes No



Calibration value Head 1: Desired value 10-50



________



Calibration value Head 2: Desired value 10-50



________



Calibration value Head 3: Desired value 10-50



________



Calibration value Head 4: Desired value 10-50



________



Calibration value Head 5: Desired value 10-50



________



Calibration value Head 6: Desired value 10-50



________



3.2



Calibrating the top press position Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Top Press



Yes No



3.3



Calibrating the upper press position Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Upper Press



Yes No



3.4



Calibrating the lower press position Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Lower Press



Yes No



3.5



Calibrating the top press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Top Press



Yes No



3.6



Calibrating the upper press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Upper Press



Yes No



3.7



Calibrating the lower press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Lower Press



Yes No



3.8



Calibrating the press scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Press Scale



Yes No



3.9



Calibrating the RCC scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / RCC Scale



Yes No



3.10



Calibrating the PLS scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / PLS Scale



Yes No



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-11



Chapter 3: Installation



No.



Description



Correct.



3.11



Calibrating the A1-A8 detectors Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Detector A



Yes No



Calibration value det. A1: Desired value 50-200



________



Calibration value det. A2: Desired value 50-200



________



Calibration value det. A3: Desired value 50-200



________



Calibration value det. A4: Desired value 50-200



________



Calibration value det. A5: Desired value 50-200



________



Calibration value det. A6: Desired value 50-200



________



Calibration value det. A7: Desired value 50-200



________



Calibration value det. A8: Desired value 50-200



________



3.12



Testing the barcode scanner, barcode test (see chapter 5.6.4 on page 5-33) Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Setup / Barcode test



Yes No



3.13



Testing the network functionality



Yes No



3.14



Testing the WLAN functionality (n.v. = not available)



Yes No n.v.



3.15



Door / Door locks Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. door



Yes No



3.17



Top press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. TP force



Yes No



Press stops at a target value of 500 N (±30 N) 3.18



_______N



Upper press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. UP force



Yes No



Press stops at a target value of 500 N (±30 N) 3.19



_______N



Lower press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. LP force



Yes No



Press stops at a target value of 500 N (±30 N)



3-12



Meas. value



_______N



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



No.



Description



3.20



Test of press scales / PLS scales / RCC scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. X Scale



3.21



Correct. Yes No



PLS scales unladen target value: 0 g (±3 g)



_______g



PLS scales loaded with 500 g target value: 500 g (±3 g)



_______g



RCC scales unladen target value: 0 g (±3 g)



_______g



RCC scales loaded with 500 g target value: 500 g (±3 g)



_______g



Press scales unladen target value: 0 g (±3 g)



_______g



Press scales loaded with 500 g target value: 500 g (±3 g)



_______g



Slide occlusion Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Slide occl. or Serv. Slide occl. no PLS



Yes No



Target value 0 g (±15 g) 3.22



Meas. value



_______g



Test of CF-Opener top press / door / RCC scales (see Re: 3.22 on page 6-24) Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. CompoFlow



Yes No



Empty CF-Opener top press ⇒ Error message empty CF-Opener Place test tool 3.6 mm in CF-Opener top press ⇒ Error message ERROR CODE 0x0401 Empty CF-Opener door ⇒ Error message empty CF-Opener Place test tool 3.6 mm in CF-Opener door ⇒ Error message ERROR CODE 0x0401 Empty CF-Opener RCC scales ⇒ Error message empty CF-Opener Place test tool 3.6 mm in CF-Opener RCC scales ⇒ Error message ERROR CODE 0x0401



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-13



Chapter 3: Installation



No.



Description



Correct.



3.23



A1 - A8 detectors Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Detector A



Meas. value



Yes No



Lower press stops / retracts when det. A1 is covered Lower press stops / retracts when det. A2 is covered Lower press stops / retracts when det. A3 is covered Lower press stops / retracts when det. A4 is covered Lower press stops / retracts when det. A5 is covered Lower press stops / retracts when det. A6 is covered Lower press stops / retracts when det. A7 is covered Lower press stops / retracts when det. A8 is covered 3.24



Opening / closing the Heads Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. O./C. heads or Serv. O./C. heads no PLS



Yes No



Close Head 1 Close Head 2 Close Head 3 (optional, only if PLS scales present) Close Head 4 Close Head 5 Close Head 6 Open Head 1 Open Head 2 Open Head 3 (optional, only if PLS scales present) Open Head 4 Open Head 5 Open Head 6 Close all Heads Open all Heads



3-14



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



No.



Description



Correct.



3.25



Testing the Heads Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Check heads or Serv. Check heads no PLS



Meas. value



Yes No



All Heads empty Tubing requested for all Heads. Head 1, tubing inserted Tubing requested for all remaining Heads Head 2, tubing inserted Tubing requested for all remaining Heads Head 3, tubing inserted (optional, only if PLS scales present) Tubing requested for all remaining Heads Head 4, tubing inserted Tubing requested for all remaining Heads Head 5, tubing inserted Tubing requested for all remaining Heads Head 6, tubing inserted Tubing requested for all remaining Heads All Heads with tubings inserted No error message 3.26



Sealing behavior of the Heads Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Seal heads or Serv. Seal heads no PLS



Yes No



Head 1: Sealing seam and separation at the sealing seam are correct Head 2: Sealing seam and separation at the sealing seam are correct Head 3 (optional, only if PLS scales present): Sealing seam and separation at the sealing seam are correct Head 4: Sealing seam and separation at the sealing seam are correct Head 5: Sealing seam and separation at the sealing seam are correct Head 6: Sealing seam and separation at the sealing seam are correct



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-15



Chapter 3: Installation



No.



Description



Correct.



3.27



Head sensors Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Head det. or Serv. Head det. no PLS



Meas. value



Yes No



Head 1: At 30% visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70% visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 2: At 30% visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 3: At 30% visual sensitivity (optional) Release at grey scale tone 4 (Tolerance: ±1) At 70% visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 4: At 30 % visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 5: At 30 % visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 6: At 30 % visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) 3.28



Behavior during bag press-out Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Empty bag



Yes No



Bag is emptied (from the lower press) Bag is emptied (from the upper press) Bag is emptied (from the top press) 3.30



Expanded test CF-Opener



Yes No



CF-Opener Top CF-Opener Door CF-Opener RCC



3-16



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



No.



Description



Correct.



3.31



System information Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / System information



Meas. value



Yes No



Serial number



________



E-Code



________



Release



________



4



Checking the electrical safety (see chapter 6.4.2 on page 6-36) For Germany: in accordance with DIN EN 62353. In other countries, observe the local regulations!



4.1



Visual inspection performed.



4.2



Protective earth resistance max. 0.3 Ω (with power cable)



4.3



Leakage current measurement (device leakage current) max. 500 µA



______ Ω



Differential current measurement acc. to DIN EN 62353, Figure 5



_____ µA



Direct measurement acc. to DIN EN 62353, Figure 4



_____ µA



or



4.4



Insulation resistance measurement > 2 MΩ



_____ MΩ



Test equipment used: Force measuring device (Type, serial number): ________________________________________________________________________________________ Protective earth resistance, leakage current, insulation resistance (Type, serial number): ________________________________________________________________________________________ Comments:



Date:



Fresenius Kabi



Signature:



CompoMat G5



SM-EN



3/11.10



Stamp:



3-17



Chapter 3: Installation



The device has been released for its intended use.



Yes No



Next check: Comments:



Date:



Signature:



Confirmation of the test



Stamp:



Test equipment used: Type and serial number of the test equipment used. Comments: Any irregularities which occurred during the test are documented in this section. Date, signature, stamp Performance of the test has to be confirmed with date, tester's signature and stamp.



Assessment of the test



The device has been released for its intended use. During the intended use of the device it must be ensured that the device does not present a hazard to patients, employees or other third parties. As part of overall evaluation, testers are obliged to give their official approval for the use of the device. If defects are identified, then the device owner must be informed of these without delay. Date of next inspection: The next inspection date has to be entered in the report. The intervals specified by the manufacturer have to be observed. Comments: Any irregularities which occurred during the assessment are documented in this section. Date, signature, stamp: Assessment of initial start-up must be confirmed by the tester including the date, tester's signature and stamp.



3-18



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 3: Installation



3.6



Decommissioning / removal from service / recommissioning Note: For information regarding this chapter see: CompoMat G5 Operating Instructions



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



3-19



Chapter 4: Specifications



4



Specifications Note: For information regarding this chapter see: CompoMat G5 Operating Instructions



4-1



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



5



Service mode Tip The display has a yellow background when service mode is active.



Tip To help you find your way around the Service Menu, a flow diagram is available for this menu (see chapter 11.6 on page 11-7).



5.1



Accessing service mode when the device is switched off Accessories



– USB stick service G5 (M67 482 1)



Procedure



1. Insert the USB stick service G5. 2. Turn on the CompoMat G5 via the main switch. Service mode is activated after approx. 2 min. The two-part self-test is then executed. (see chapter 5.2.1 on page 5-2)



5.2



Accessing service mode from standby Accessories



– USB stick service G5 (M67 482 1)



Procedure



1. Insert the USB stick service G5. (see chapter 3.4 on page 3-8) 2. Wait at least 10 s (USB stick is initialized). 3. Press the menu key



(On/Off).



The two-part self-test is then executed. (see chapter 5.2.1 on page 5-2)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-1



Chapter 5: Service mode



5.2.1



Self-test The following components are tested: – – – – – – – –



5.2.2



System files Network IO communication Slide CF-Opener Top CF-Opener Door CF-Opener RCC Head



– – – – – – –



PLS Scale RCC scales Press Scale Upper Press Lower Press Top Press Door



Entering the password



Note: Passwords (see Passwords on page 10-1) Once the self-test has been passed, the screen message Password request will be displayed (shown here on the left). The password is entered as follows: – Select the input position using the function keys.



(yellow) and



(green)



– Use the (Up) and (Down) menu keys to select the required numerical value for the highlighted input position. The password is “1462”. – Confirm the input using the menu key



5.2.3



.



Main menu (Service) Once the password has been entered, the screen message Main menu (Service) will be displayed (shown here on the left). The following menu options are available: – – – – – – –



5-2



Set serial number Set E-Code Set MAC-Address Update/Init ARM7 Update/Init Trinamic Cal. ADC & Heads Special functions



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



5.2.3.1



Entering the serial number



Note: The serial number must be entered again after the replacement of the Display G5 printed circuit board bought-in assembly. Requirements



The menu Main menu (Service) must be selected. (see chapter 5.1 on page 5-1) to (see chapter 5.2.3 on page 5-2) – Use the (Up) and (Down) menu keys to select the menu option Set serial number. – Confirm the selection using the menu key



.



The screen message Set serial number will be displayed (shown here on the left). The serial number is entered as follows: – Select the input position using the function keys.



(yellow) and



(green)



– Use the (Up) and (Down) menu keys to select the required numerical value for the highlighted input position. – Confirm the input using the menu key



.



The screen message Advice will be displayed (shown here on the left). – Accept the message using the



(green) function key.



The CompoMat G5 will be rebooted. A screen message will be displayed, requesting password entry. (see chapter 5.2.2 on page 5-2)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-3



Chapter 5: Service mode



5.2.3.2



Entering an E-code



Note: The E-code must be entered again after the replacement of the Display G5 printed circuit board bought-in assembly. Requirements



The menu Main menu (Service) must be selected. (see chapter 5.1 on page 5-1) to (see chapter 5.2.3 on page 5-2) – Use the (Up) and option Set E-Code.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



.



The screen message Set E-Code will be displayed (shown here on the left). – Note the E-code from the rear of the CompoMat G5 and enter as follows: – Select the input position using the function keys.



(yellow) and



(green)



– Use the (Up) and (Down) menu keys to select the required numerical value for the highlighted input position. – Confirm the input using the menu key



.



The screen message Advice will be displayed (shown here on the left). – Accept the message using the



(green) function key.



You will be returned to the Main menu (Service) screen.



5-4



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



5.2.3.3



Update/Init ARM7



Note: The firmware for the Application Board bought-in assembly (M68 137 1) is installed automatically during installation of the G5 procedure software (see Software update on page 5-31). Manual installation normally is not required. Note: After the update/init on the Application Board bought-in assembly (M68 137 1) the CompoMat G5 must be fully calibrated (see chapter 6.2 on page 6-2). The calibration process is started automatically.



Requirements



The menu Main menu (Service) must be selected. (see chapter 5.1 on page 5-1) to (see chapter 5.2.3 on page 5-2)



Accessories



The subsequent calibration requires the use of the following accessories: – Calibrated force measuring device incl. pressure sensor (M67 499 1) – Tool for force measurement G5 (M67 961 1) – Synthetic block G5 (M67 851 1) – Use the (Up) and (Down) menu keys to select the menu option Update/init ARM7. – Confirm the selection using the menu key



.



The update is carried out without safety prompt. The following screen messages will be displayed in sequence: – – – – –



5.2.3.4



Wait... Reset IO board... Wait... Erase IO software... Wait... Install IO software... Restart... Please wait... Selftest...



Update/Init Trinamic



Note: The procedures for the Trinamic Board bought-in assembly (M68 138 1) are installed automatically during installation of the G5 procedure software (see Software update on page 5-31). Manual installation of the Trinamic board normally is not required.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-5



Chapter 5: Service mode



Requirements



The menu Main menu (Service) must be selected. (see chapter 5.1 on page 5-1) to (see chapter 5.2.3 on page 5-2) – Use the (Up) and (Down) menu keys to select the menu option Update/init Trinamic. – Confirm the selection using the menu key



.



The Application Board bought-in assembly (M68 138 1) update is carried out automatically without safety prompt.



The screen message Init motor control will be displayed (shown here on the left). The CompoMat G5 will be rebooted once the progress bar is complete.



Note: All presses must be re-calibrated following an update/init on the Trinamic Board bought-in assembly (M68 138 1). The calibration process is started automatically.



5.2.3.5



Calibrating ADC & Heads Requirements



5-6



The menu Main menu (Service) must be selected. (see chapter 5.1 on page 5-1) to (see chapter 5.2.3 on page 5-2)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



– Use the (Up) and (Down) menu keys to select the menu option Cal. ADC & Heads . – Confirm the selection using the menu key



.



The screen message Calibrate ADC & Head Detectors... will be displayed (shown here on the left).



After successful calibration, the screen message Advice will be displayed (shown here on the left). – The calibration data displayed must be documented in the report. (i.e. values for Heads 1 to 6) – Confirm the end of calibration using the



(green) function key.



The CompoMat G5 will be rebooted.



Note: The calibration values must be in the range between 10 and 50.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-7



Chapter 5: Service mode



5.2.4



Special functions (extended menu) Requirements



The menu Main menu (Service) must be selected. (see chapter 5.1 on page 5-1) to (see chapter 5.2.3 on page 5-2) – Use the (Up) and Special functions.



(Down) menu keys to select the menu option



– Confirm the selection using the menu key



.



The screen message Special functions will be displayed (shown here on the left). The Special functions menu expands the Main menu (Service) to include the following options: – – – – – –



5.2.5



Appl. Service mode Download files Unittest screenshots Unittest network stress Create Production Stick Production state



Application service mode Requirements



The extended menu Special functions must be selected in the Main menu (Service). (see chapter 5.2.3 on page 5-2) – Use the (Up) and (Down) menu keys to select the menu option Appl. Service-Mode. – Confirm the selection using the menu key



5-8



Fresenius Kabi



.



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



The screen message Ready to use will be displayed (shown here on the left). The program selected (here, 30 Demo no heads) may differ from the actual program shown. – Press the



(green) function key.



The program displayed will be executed. Or – Press the



(yellow) function key.



The screen message Main menu will be displayed (shown here on the left). The following menu options are available: – – – – – –



Special features in service mode



Programm selection Seal & weighing Calibrate Cleanup Setup System information



– All passwords are deactivated – The language is English throughout and cannot be modified. – Process data are deactivated



5.2.6



Program selection Requirements



The sub-menu Appl. Service-Mode has been selected in the extended Special functions menu. (see chapter 5.2.5 on page 5-8) – Use the (Up) and (Down) menu keys to select the menu option Program selection. – Confirm the selection using the menu key



Special features for service menu "Program selection"



Fresenius Kabi



CompoMat G5



SM-EN



.



In the Program selection service menu, a fixed set of service programs are available that are not provided by CompoMaster Net G5.



3/11.10



5-9



Chapter 5: Service mode



The screen message Program selection will be displayed (shown here on the left). – Use the program.



(Up) and



(Down) menu keys to select the desired



– Confirm the selection using the menu key



.



The screen message Ready to use will be displayed (shown here on the left). – Press the



(green) function key.



The selected program (here, 30 Demo no heads) will be executed. Or – Press the



(yellow) function key.



The display switches back to the Main menu



5.2.6.1



Display and operation of a service program The program step currently running is shown in field 1. 1



– To stop the service program: Press the – To interrupt the service program: Press the



5-10



Fresenius Kabi



(blue) function key. (yellow) function key.



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



– The corresponding parameters are shown in field 1 during execution of the program step.



1



5.3



Seal & weighing (program-independent) Requirements



The sub-menu Appl. Service-Mode has been selected in the extended Special functions menu. (see chapter 5.2.5 on page 5-8) – Use the (Up) and (Down) menu keys to select the menu option Seal & weighing. – Confirm the selection using the menu key



.



The screen message Seal & weighing will be displayed (shown here on the left).



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-11



Chapter 5: Service mode



5.3.1



Program-independent sealing



Note: Only Head 2 can be used for program-independent sealing. Tip The number of sealing processes is not limited. – Insert the tubing into sealing head 2. – Press the



(green) function key.



The inserted tubing will be sealed. – Press the



(blue) function key.



The display switches back to the Main menu



5.3.2



Program-independent weighing



Note: Only the RCC scales can be used for program-independent weighing. Tip The number of weighing processes is not limited. – Tare the RCC scales using the



(yellow) function key.



The weighing process can be carried out. – Press the



(blue) function key.



The display switches back to the Main menu



5-12



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



5.4



Calibration Warning Risk of injury from actuators moving automatically. Crushing hazard – Do not reach into the press area when the presses are active. Requirements



The sub-menu Appl. Service-Mode has been selected in the extended Special functions menu. (see chapter 5.2.5 on page 5-8) – Use the (Up) and option Calibrate.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



.



The screen message Calibrate will be displayed (shown here on the left). The following actuators can be calibrated: – – – – – – –



5.4.1



Top press upper press lower press Press scales RCC scales PLS scales Detector A



Calibrating top press Requirements



Fresenius Kabi



CompoMat G5



The menu Calibrate must be selected in the Main menu. (see chapter 5.4 on page 5-13)



SM-EN



3/11.10



5-13



Chapter 5: Service mode



– Use the actuator.



(Up) and



(Down) menu keys to select the desired



– Confirm the selection using the menu key



.



The screen message Calibrate Top Press will be displayed (shown here on the left). Two calibration functions are available: – Calibrate press force ( – Calibrate press stroke (



5.4.1.1



Top press (calibrating press force) Requirements



The menu Calibrate Top Press must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Calibrated force measuring device incl. pressure sensor (M67 499 1) – Use the



5-14



(yellow) function key) (green) function key)



(yellow) function key to select Calibrate force.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



The screen message Calibrate Top Press will be displayed (shown here on the left). The CompoMat G5 automatically calculates the null force for the top press.



On completion of the null force adjustment, the screen message Calibrate Top Press will be displayed (shown here on the left).



– Place the calibrated force measuring device in or under the top press (see illustration to the left). – Proceed by pressing the



(green) function key.



The screen message Calibrate Top Press will be displayed (shown here on the left). Note the determined force [N] from the force measuring device and enter it as follows: – Select the input position using the function keys.



(yellow) and



(green)



– Use the (Up) and (Down) menu keys to select the required numerical value for the highlighted input position. – Confirm the input using the menu key



.



After successful calibration, the message Calibration finished successfully! will be displayed.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-15



Chapter 5: Service mode



5.4.1.2



Top press (calibrating press stroke) Requirements



The menu Calibrate Top Press must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13) – Use the



(green) function key to select Calibrate position.



The screen message Calibrate Top Press will be displayed (shown here on the left). The calibration process for the press stroke will be performed and completed automatically.



5.4.2



Calibrating the upper press Requirements



The menu Calibrate must be selected in the Main menu. (see chapter 5.4 on page 5-13) – Use the actuator.



(Up) and



(Down) menu keys to select the desired



– Confirm the selection using the menu key



5-16



Fresenius Kabi



.



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



The screen message Calibrate Upper Press will be displayed (shown here on the left). Two calibration functions are available: – Calibrate press force ( – Calibrate press stroke (



5.4.2.1



(yellow) function key) (green) function key)



Upper press (calibrating the press force) Requirements



The menu Calibrate Upper Press must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Calibrated force measuring device incl. pressure sensor (M67 499 1) – Use the



(yellow) function key to select Calibrate force.



The screen message Calibrate Upper Press will be displayed (shown here on the left). The CompoMat G5 automatically calculates the null force for the upper press.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-17



Chapter 5: Service mode



On completion of the null force adjustment, the screen message Calibrate Upper Press will be displayed (shown here on the left).



– Place the calibrated force measuring device into the upper press (see illustration to the left). – Proceed by pressing the



(green) function key.



The screen message Calibrate Upper Press will be displayed (shown here on the left). Note the determined force [N] from the force measuring device and enter it as follows: – Select the input position using the function keys.



(yellow) and



(green)



– Use the (Up) and (Down) menu keys to select the required numerical value for the highlighted input position. – Confirm the input using the menu key



.



After successful calibration, the message Calibration finished successfully! will be displayed.



5-18



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



5.4.2.2



Upper press (calibrating press stroke) Requirements



The menu Calibrate Upper Press must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13) – Use the



(green) function key to select Calibrate position.



The screen message Calibrate Upper Press will be displayed (shown here on the left). The calibration process for the press stroke will be performed and completed automatically.



5.4.3



Calibrating the lower press Requirements



The menu Calibrate must be selected in the Main menu. (see chapter 5.4 on page 5-13) – Use the actuator.



(Up) and



(Down) menu keys to select the desired



– Confirm the selection using the menu key



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



.



5-19



Chapter 5: Service mode



The screen message Calibrate Lower Press will be displayed (shown here on the left). Two calibration functions are available: – Calibrate press force ( – Calibrate press stroke (



5.4.3.1



(yellow) function key) (green) function key)



Lower press (calibrating the press force) Requirements



The menu Calibrate Lower Press must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Calibrated force measuring device incl. pressure sensor (M67 499 1) – Use the



(yellow) function key to select Calibrate force.



The screen message Calibrate Lower Press will be displayed (shown here on the left). The CompoMat G5 automatically calculates the null force for the lower press.



5-20



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



On completion of the null force adjustment, the screen message Calibrate Lower Press will be displayed (shown here on the left).



– Place the calibrated force measuring device into the lower press (see illustration to the left). – Proceed by pressing the



(green) function key.



The screen message Calibrate Lower Press will be displayed (shown here on the left). Note the determined force [N] from the force measuring device and enter it as follows: – Select the input position using the function keys.



(yellow) and



(green)



– Use the (Up) and (Down) menu keys to select the required numerical value for the highlighted input position. – Confirm the input using the menu key



.



After successful calibration, the message Calibration finished successfully! will be displayed.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-21



Chapter 5: Service mode



5.4.3.2



Lower press (calibrating the press stroke) Requirements



The menu Calibrate Lower Press must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13) – Use the



(green) function key to select Calibrate position.



The screen message Calibrate Lower Press will be displayed (shown here on the left). The calibration process for the press stroke will be performed and completed automatically.



5.4.4



Calibrating the scales



Note: When calibrating the scales, the following requirements must be met: – The CompoMat G5 must be set up so that it is free of vibrations. – Do not allow any external force to affect the scales.



5.4.4.1



5-22



Calibrating the press scales Requirements



The menu Calibrate Press Scale must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Calibration weight 500 g (M63 925 1)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



– Relieve the press scales. – Start the calibration process using the



(green) function key.



The device calculates the zero point for the press scales (approx. 3 s).



The screen message Calibrate Press Scale will be displayed (shown here on the left). – Hang the 500 g calibration weight on the suspension pins. – Wait until the weight has stopped moving (approx. 3 s). – Confirm the weight using the



(green) function key.



After successful calibration, the screen message shown on the left will be displayed. – Check the calibration of the press scales. (see Re: 3.20 on page 6-22)



5.4.4.2



Calibrating the RCC scales Requirements



The menu Calibrate RCC Scale must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Calibration weight 500 g (M63 925 1) Note: If the calibration process is aborted using the then the previous calibration is retained.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



(blue) function key,



5-23



Chapter 5: Service mode



– Relieve the RCC scales. – Start the calibration process using the



(green) function key.



The device calculates the zero point for the RCC scales (approx. 3 s).



The screen message Calibrate RCC Scale will be displayed (shown here on the left). – Place 500 g calibration weight in the scales tray. – Wait until the scales tray has stopped moving (approx. 3 s). – Confirm the weight using the



(green) function key.



After successful calibration, the screen message shown on the left will be displayed. – Check calibration of the RCC scales. (see Re: 3.20 on page 6-22)



5.4.4.3



Calibrating the PLS scales Requirements



The menu Calibrate PLS Scale must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Calibration weight 500 g (M63 925 1) Note: If the calibration process is aborted using the then the previous calibration is retained.



5-24



Fresenius Kabi



(blue) function key,



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



– Make sure there is nothing on the scale. – Start the calibration process using the



(green) function key.



The device calculates the zero point for the PLS scales. (approx. 3 s).



The screen message Calibrate PLS Scale will be displayed (shown here on the left). – Place 500 g calibration weight on sealing head 3. – Wait until the weight has stopped moving (approx. 3 s). – Confirm the weight using the



(green) function key.



After successful calibration, the screen message shown on the left will be displayed. – Check calibration of the PLS scales. (see Re: 3.20 on page 6-22)



5.4.5



Calibrating detector A Requirements



The menu Calibrate Detector A must be selected in the Calibrate menu. (see chapter 5.4 on page 5-13)



Accessories



– Synthetic block G5 (M67 851 1)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-25



Chapter 5: Service mode



– Hang the G5 synthetic block on the suspension pins. – Press the



(green) function key.



The calibration process will be performed automatically (multiple screen messages shown)



After successful calibration, the screen message shown to the left will be displayed, presenting the detector calibration data. – The calibration data displayed must be documented. (i.e. values for detectors A1 to A8) – Confirm with the



(green) function key.



The screen display switches back to the Calibrate menu.



Note: The calibration values must be in the range between 50 and 200.



5.4.6



Calibrating ADC & Heads (see chapter 5.2.3.5 on page 5-6)



5.5



Cleaning the press area Warning Risk of injury caused by parts moving autonomously. Crushing hazard Do not reach into the press area while the press or slide are moving! Requirements



5-26



The sub-menu Appl. Service-Mode has been selected in the extended Special functions menu. (see chapter 5.2.5 on page 5-8)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



– Use the (Up) and option Cleanup.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



.



The screen message Cleanup will be displayed (shown here on the left). The following options are shown: – upper press – lower press – slide



5.5.1



Cleaning the upper press Requirements



The menu Cleanup must be selected in the Main menu. (see chapter 5.4 on page 5-13) – Use the (Up) and option Upper Press.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



.



5-27



Chapter 5: Service mode



The screen message Cleanup Upper Press will be displayed (shown here on the left). – Press and hold the



(green) function key.



The upper press is extended. The upper press can now be cleaned. – Press and hold the



(yellow) function key.



The upper press is retracted. – Press the



(blue) function key.



The screen display switches back to the Cleanup menu. Note: You can only exit the Cleanup Upper Press menu once the upper press has been fully retracted.



5.5.2



Cleaning the lower press Requirements



The menu Cleanup must be selected in the Main menu. (see chapter 5.4 on page 5-13) – Use the (Up) and option Lower Press .



(Down) menu keys to select the menu



– Confirm the selection using the menu key



.



The screen message Cleanup Lower Press will be displayed (shown here on the left). – Press and hold the



(green) function key.



The lower press is extended. The lower press can now be cleaned. – Press and hold the



(yellow) function key.



The lower press is retracted. – Press the



(blue) function key.



The screen display switches back to the Cleanup menu.



5-28



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



Note: You can only exit the Cleanup Lower Press menu once the lower press has been fully retracted.



5.5.3



Cleaning the slide Requirements



The menu Cleanup must be selected in the Main menu. (see chapter 5.4 on page 5-13) – Use the (Up) and option Slide.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



.



The screen message Cleanup Slide will be displayed (shown here on the left). – Press and hold the



(green) function key.



The slide is extended. The slide can now be cleaned. – Press and hold the



(yellow) function key.



The slide is retracted. – Press the



(blue) function key.



The screen display switches back to the Cleanup menu. Note: You can only exit the Cleanup Slide menu once the slide has been fully retracted.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-29



Chapter 5: Service mode



5.6



Setup Requirements



The sub-menu Appl. Service-Mode has been selected in the extended Special functions menu. (see chapter 5.2.5 on page 5-8) – Use the (Up) and option Setup.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



.



The screen message Setup will be displayed (shown here on the left). The following options are shown: – – – –



5.6.1



Language selection Process data Software update Barcode test



Language selection



Note: Language selection only affects the language settings for the operator. Service mode is only available in English. Requirements



5-30



The menu Setup must be selected in the Main menu. (see chapter 5.6 on page 5-30)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



– Use the (Up) and (Down) menu keys to select the menu option Language selection. – Confirm the selection using the menu key



5.6.2



.



Process data Requirements



The menu Setup must be selected in the Main menu. (see chapter 5.6 on page 5-30) Note: Process data settings only affect the user mode. These settings have no effect in service mode. – Use the (Up) and option Process data.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



5.6.3



.



Software update Requirements



The menu Setup must be selected in the Main menu. (see chapter 5.6 on page 5-30)



Accessories



– Update CompoMat G5 VX.X.X USB stick (M67 484 1)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



5-31



Chapter 5: Service mode



– Use the (Up) and (Down) menu keys to select the menu option Software update. – Confirm the selection using the menu key



.



The screen message Software update will be displayed (shown here on the left). – Press the



(green) function key.



The update process is started if there is a valid update on the USB stick.



The screen message Software update will be displayed (shown here on the left).



5-32



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 5: Service mode



The following screen messages will be displayed in sequence:



Once the self-test has been successfully completed, the password will once again be requested for service mode (see chapter 5.2.2 on page 5-2)



5.6.4



Barcode test Requirements



The menu Setup must be selected in the Main menu. (see chapter 5.6 on page 5-30) – Use the (Up) and option Barcode test.



(Down) menu keys to select the menu



– Confirm the selection using the menu key



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



.



5-33



Chapter 5: Service mode



The screen message Barcode scan will be displayed (shown here on the left). – Scan in the test barcode shown below.



The barcode is shown in plain text. (in the screen shot: BC_TEST_VAL or KBC_TEST_VAL)



5.6.5



System information Requirements



The sub-menu Appl. Service-Mode has been selected in the extended Special functions menu. (see chapter 5.2.5 on page 5-8) – Use the (Up) and (Down) menu keys to select the menu option System information. – Confirm the selection using the menu key



.



The screen message System information will be displayed (shown here on the left). The following system information is made available: – – – – – – –



Netname: Network name Process counter: Number of processes run Serial-No.: Serial number Software: Software version Firmware: Firmware version E-Code: Equipment code Version: Complete version number



– Press the



(blue) function key.



The display switches back to the Main menu



5-34



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



6



TSC / TMC / MA



6.1



Important information for the procedure TSC



Technical safety check Note: There is no TSC obligation for the CompoMat G5.



TMC



Technical measurement check Note: There is no TMC obligation for the CompoMat G5 .



MA



Annual maintenance In order to avoid personal injury and damage to property, an annual maintenance must be carried out. (see chapter 6 on page 6-1) Once the CompoMat G5 has been opened, the Electrical Safety Check must be carried out before recommissioning the device. (see chapter 6.3 on page 6-12) The following applies to the annual maintenance: After successful completion of the annual maintenance, the CompoMat G5 must be identified with an inspection label. This label must, in a unique and traceable manner, specify the year of the next annual maintenance and the authority or person having performed the annual maintenance.



Tester's qualification



The checks must be performed by the Technical Service of Fresenius Kabi or an institution/person authorized by them. The checks may only be performed by persons qualified to properly perform the specified checks owing to their educational background, training, knowledge and experience. Furthermore, the persons performing the checks must not be bound by any directives when performing this activity.



Test equipment and accessories



The activities described in this technical document require the availability of the necessary technical measuring equipment and accessories.



Specifications



Observe the information on the specifications.



Precautions



Before turning power on, repair any visible damage. Prior to opening the device and when working on the open device, the following precautions have to be observed: – Protect the components against ingress of fluids.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-1



Chapter 6: TSC / TMC / MA



– Do not touch live parts. – Disconnect and connect all jacks, connectors and components only when the device is turned off.



6.2



Test report - annual maintenance Annual maintenance



Maintenance here relates to checking the sensors and actuators. Checks are carried out without opening the device. Warning Risk of injury caused by electrical voltage. Caused by changes to the cable connection, the way the cables are laid or due to an electrical part. Once the CompoMat G5 has been opened, the Electrical Safety Check must be carried out before recommissioning the device. (see chapter 6.3 on page 6-12) Note: A detailed description of the individual items can be found under chapter Explanations of reports (see chapter 6.4.1 on page 6-14)



6-2



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



CompoMat G5



Test report - annual maintenance



Technician's name:



Customer/customer no.:



Version:



Optional:



Device no.:



Service report no.:



RCC scales: Serial number: PLS scales: Serial number:



Equipment code:



No.



Description



2



Visual inspection



2.1



Check the fuses on the power input panel jack. Melting point of the fuses (2x) T 10 A / 230/240 V ~AC.



2.2



Labels and inscriptions are present and legible.



2.3



Mechanical condition permits safe use.



2.4



No contamination or damage.



2.5



The power cable is not damaged.



2.6



The complete housing is cleaned



2.7



Welding head 1-6 (no penetrated liquids)



2.8



Silicone mats from top press, lower press, upper press and door are not worn or damaged



2.9



Top press cleaned and with no damage



2.10



Door cleaned and with no damage



2.11



Touch pad undamaged, key contact points OK



2.12



Check the upper press block with the cleaning program



2.13



Check the lower press block with the cleaning program



2.14



Check the slide with the cleaning program



2.15



Check the allocation of the sealing heads



2.16



Connector plug and circuit from PLS scales and RCC scales undamaged



2.17



Network connection and barcode scanner connection undamaged



Fresenius Kabi



CompoMat G5



Correct.



SM-EN



3/11.10



Meas. value



6-3



Chapter 6: TSC / TMC / MA



No.



Description



Correct.



2.18



Device feet undamaged



2.19



Check that the suspension pins of the upper press are secure. Do not bend the suspension pins; replace them if necessary



3



Device-related inspections



3.1



Calibration of ADC and the head sensors Main menu (Service) / Cal. ADC & Heads



Yes No



Calibration value Head 1: Desired value 10-50



________



Calibration value Head 2: Desired value 10-50



________



Calibration value Head 3: Desired value 10-50



________



Calibration value Head 4: Desired value 10-50



________



Calibration value Head 5: Desired value 10-50



________



Calibration value Head 6: Desired value 10-50



________



3.2



Calibrating the top press position Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Top Press



Yes No



3.3



Calibrating the upper press position Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Upper Press



Yes No



3.4



Calibrating the lower press position Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Lower Press



Yes No



3.5



Calibrating the top press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Top Press



Yes No



3.6



Calibrating the upper press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Upper Press



Yes No



3.7



Calibrating the lower press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Lower Press



Yes No



3.8



Calibrating the press scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Press Scale



Yes No



3.9



Calibrating the RCC scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / RCC Scale



Yes No



3.10



Calibrating the PLS scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / PLS Scale



Yes No



6-4



Meas. value



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



No.



Description



Correct.



3.11



Calibrating the A1-A8 detectors Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Calibrate / Detector A



Yes No



Calibration value det. A1: Desired value 50-200



________



Calibration value det. A2: Desired value 50-200



________



Calibration value det. A3: Desired value 50-200



________



Calibration value det. A4: Desired value 50-200



________



Calibration value det. A5: Desired value 50-200



________



Calibration value det. A6: Desired value 50-200



________



Calibration value det. A7: Desired value 50-200



________



Calibration value det. A8: Desired value 50-200



________



3.12



Testing the barcode scanner, barcode test (see chapter 5.6.4 on page 5-33)



Yes No



3.13



Testing the network functionality



Yes No



3.14



Testing the WLAN functionality (n.v. = not available)



Yes No n.v.



3.15



Door / Door locks Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. door



Yes No



3.16



Closing characteristics of the door with a bag Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. door Door must close correctly



Yes No



3.17



Top press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. TP force



Yes No



Press stops at a target value of 500 N (±30 N) 3.18



Upper press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. UP force



_______N Yes No



Press stops at a target value of 500 N (±30 N) 3.19



Meas. value



Lower press force Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. LP force Press stops at a target value of 500 N (±30 N)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



_______N Yes No _______N



6-5



Chapter 6: TSC / TMC / MA



No.



Description



3.20



Test of press scales / PLS scales / RCC scales Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. X Scale



3.21



Correct. Yes No



PLS scales unladen target value: 0 g (±3 g)



_______g



PLS scales loaded with 500 g target value: 500 g (±3 g)



_______g



RCC scales unladen target value: 0 g (±3 g)



_______g



RCC scales loaded with 500 g target value: 500 g (±3 g)



_______g



Press scales unladen target value: 0 g (±3 g)



_______g



Press scales loaded with 500 g target value: 500 g (±3 g)



_______g



Slide occlusion Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Slide occl. or Serv. Slide occl. no PLS



Yes No



Target value 0 g (±15 g) 3.22



Meas. value



_______g



Test of CF-Opener top press / door / RCC scales (see Re: 3.22 on page 6-24) Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. CompoFlow



Yes No



Empty CF-Opener top press ⇒ Error message empty CF-Opener Place test tool 3.6 mm in CF-Opener top press ⇒ Error message ERROR CODE 0x0401 Empty CF-Opener door ⇒ Error message empty CF-Opener Place test tool 3.6 mm in CF-Opener door ⇒ Error message ERROR CODE 0x0401 Empty CF-Opener RCC scales ⇒ Error message empty CF-Opener Place test tool 3.6 mm in CF-Opener RCC scales ⇒ Error message ERROR CODE 0x0401



6-6



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



No.



Description



Correct.



3.23



A1 - A8 detectors Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Detector A



Meas. value



Yes No



Lower press stops / retracts when det. A1 is covered Lower press stops / retracts when det. A2 is covered Lower press stops / retracts when det. A3 is covered Lower press stops / retracts when det. A4 is covered Lower press stops / retracts when det. A5 is covered Lower press stops / retracts when det. A6 is covered Lower press stops / retracts when det. A7 is covered Lower press stops / retracts when det. A8 is covered 3.24



Opening / closing the heads Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. O./C. heads or Serv. O./C. heads no PLS



Yes No



Close Head 1 Close Head 2 Close Head 3 (optional, only if PLS scales present) Close Head 4 Close Head 5 Close Head 6 Open Head 1 Open Head 2 Open Head 3 (optional, only if PLS scales present) Open Head 4 Open Head 5 Open Head 6 Close all Heads Open all Heads



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-7



Chapter 6: TSC / TMC / MA



No.



Description



Correct.



3.25



Testing the Heads Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Check heads or Serv. Check heads no PLS



Meas. value



Yes No



All Heads empty Tubing requested for all Heads. Head 1, tubing inserted Tubing requested for all remaining Heads Head 2, tubing inserted Tubing requested for all remaining Heads Head 3, tubing inserted (optional, only if PLS scales present) Tubing requested for all remaining Heads Head 4, tubing inserted Tubing requested for all remaining Heads Head 5, tubing inserted Tubing requested for all remaining Heads Head 6, tubing inserted Tubing requested for all remaining Heads All Heads with tubings inserted No error message 3.26



Sealing behavior of the Heads Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Seal heads or Serv. Seal heads no PLS



Yes No



Head 1: Sealing seam and separation at the sealing seam are correct Head 2: Sealing seam and separation at the sealing seam are correct Head 3 (optional, only if PLS scales present): Sealing seam and separation at the sealing seam are correct Head 4: Sealing seam and separation at the sealing seam are correct Head 5: Sealing seam and separation at the sealing seam are correct Head 6: Sealing seam and separation at the sealing seam are correct



6-8



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



No.



Description



Correct.



3.27



Head sensors Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Head det. or Serv. Head det. no PLS



Meas. value



Yes No



Head 1: At 30% visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70% visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 2: At 30% visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 3: At 30% visual sensitivity (optional) Release at grey scale tone 4 (Tolerance: ±1) At 70% visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 4: At 30 % visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 5: At 30 % visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) Head 6: At 30 % visual sensitivity Release at grey scale tone 4 (Tolerance: ±1) At 70 % visual sensitivity Release at grey scale tone 2 (Tolerance: ±1) 3.28



Behavior during bag press-out Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / Program selection / Serv. Empty bag



Yes No



Bag is emptied (from the lower press) Bag is emptied (from the upper press) Bag is emptied (from the top press) 3.29



Stud screws for CF-Opener covers tightened hand-tight



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-9



Chapter 6: TSC / TMC / MA



No.



Description



3.30



Expanded test CF-Opener



Correct.



Meas. value



Yes No



CF-Opener Top CF-Opener Door CF-Opener RCC 3.31



System information Main menu (Service) / Special functions / Appl. Service mode / Ready to use / Menu / System information



Yes No



Serial number



________



E-Code



________



Release



________



Test equipment used: Force measuring device (Type, serial number): ________________________________________________________________________________________ Comments:



Date:



6-10



Signature:



Stamp:



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



The device has been released for its intended use.



Yes No



Next check: Comments:



Date:



Fresenius Kabi



Signature:



CompoMat G5



SM-EN



3/11.10



Stamp:



6-11



Chapter 6: TSC / TMC / MA



6.3



Checking the electrical safety Warning Danger of injury from electric voltage due to changes to cable connections, the way the cable is laid or due to an electrical part. Once the CompoMat G5 has been opened, the Electrical Safety Check must be carried out before recommissioning the device. (see chapter 6.3 on page 6-12) Note: A detailed description of the individual items can be found under chapter Explanations of reports (see chapter 6.4.2 on page 6-36)



No.



Description



Correct.



4



Checking the electrical safety (see chapter 6.4.2 on page 6-36) For Germany: in accordance with DIN EN 62353. In other countries, observe the local regulations!



4.1



Visual inspection performed.



4.2



Protective earth resistance max. 0.3 Ω (with power cable)



4.3



Leakage current measurement (device leakage current) max. 500 µA



Meas. value



______ Ω



Differential current measurement acc. to DIN EN 62353, Figure 5



_____ µA



Direct measurement acc. to DIN EN 62353, Figure 4



_____ µA



or



4.4



6-12



Insulation resistance measurement > 2 MΩ



_____ MΩ



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Test equipment used: Force measuring device (Type, serial number): ________________________________________________________________________________________ Protective earth resistance, leakage current, insulation resistance (Type, serial number): ________________________________________________________________________________________ Comments:



Date:



Signature:



The device has been released for its intended use.



Stamp:



Yes No



Next check: Comments:



Date:



Fresenius Kabi



Signature:



CompoMat G5



SM-EN



3/11.10



Stamp:



6-13



Chapter 6: TSC / TMC / MA



6.4



Explanations on the reports Note: The tolerances specified in the report must be observed. Corrections are only required if the measured value lies outside the specified tolerance.



6.4.1



Explanations on the reports "Initial start-up" and "Annual maintenance" Identification



Technician's name: Technician's first name and last name. Device type with option(s) / software: System name with possible options or extras. Service report number: Service operation number. Customer/customer no.: Final customer's number. Device no.: The serial number printed on the type label Inventory no.: Inventory number assigned to the device. Equipment code: The configuration code on the device (E-code xxx). Version: Complete version number.



6-14



Re: 1



Unpacking (only for initial start-up!)



Re: 1.1



Remove the CompoMat G5 from its transport packaging and save the packaging for possible future return shipments.



Re: 1.2



Check that all parts of the CompoMat G5 have been accounted for. Inform your contact at Fresenius immediately of any missing parts.



Re: 1.3



CompoMat G5 cycler without visible shipping damage. Check that the CompoMat G5 has no visible damage or is dirty in a way that could affect safety.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Re: 2



Visual inspection



Re: 2.1 Check the line fuses accessible from the outside. Melting point of the line fuses (2x) T 10 A / 230/240 V ~AC. To change the fuses, first disconnect the device from the power supply. The line fuses (2) are located behind a cover underneath the power socket (1). 1 2



Re: 2.2



The line fuses (2) can be accessed by pulling and folding out the cover. Fig.: Power connection (1) with line fuses (2)



Labels and inscriptions are present and legible: Labels on front / controls: – – – –



Label about Head numbering; Top press warning label; Press / door warning label; CF-Opener warning label;



Labels on rear: – – – – – –



Type label; Fuse label; Balance A label; Balance B label; Scanner label; Network label.



Re: 2.3



The mechanical condition permits further use.



Re: 2.4



There are no signs of damage or contamination.



Re: 2.5



Power cable not damaged.



Re: 3



Calibration and functional testing of the components, Control of the functional ability and the operational safety The following features of the device are checked for correct function when the device is used for its intended purpose.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-15



Chapter 6: TSC / TMC / MA



– – – –



Functionality of intended use Functionality of the existing warning devices Measuring function Parameter specifications and their tolerances



Calibration can be carried out in user mode or service mode. For each description, a note will inform you whether calibration/testing must be carried out in user mode or service mode. The user mode is normal operation. The CompoMat G5 starts up as usual with no help from applications in user mode. Service Mode (see chapter 5 on page 5-1) is reserved exclusively for development, production and service and offers a number of advanced functions. Re: 3.1



Calibration of ADC and the head sensors (see chapter 5.2.3.5 on page 5-6)



Re: 3.2 to 3.11



Calibration – – – – – – – – – –



Top press position Upper press position Lower press position Top press force Upper press force Lower press force Press scales RCC scales PLS scales Detectors A1 to A8



(see chapter 5.4 on page 5-13) Re: 3.12



Testing the barcode scanner (see chapter 5.6.4 on page 5-33)



Re: 3.13



Testing the network ethernet RJ45 Accessories: – 2x network cable (M67 810 1) – 1x LAN hub 5-port (M67 481 1) Preparation: – Connect CompoMat G5 to LAN hub assy. using a network cable – Connect CompoMaster Net G5 PC to LAN hub assy. using a network cable – Start CompoMat G5 in user mode – Start CompoMaster Net G5. Test: – Deactivate WLAN interface on the PC on which the CompoMaster Net G5 is installed. – Configure Ethernet interface on the PC on which the CompoMaster Net G5 is installed as follows:



6-16



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



IP address:



192.168.0.250



Netmask:



255.255.255.0



Gateway:



(No entry)



DNS:



(No entry)



– Select configured ethernet interface in CompoMaster Net G5 (See User Manual, CompoMaster Net G5, Configuration / CompoMat G5 Parameters) – Set up CompoMat G5 on CompoMaster Net G5 (Allocate serial number, IP address and network name). – Reboot CompoMaster Net G5. The CompoMaster Net G5 recognizes the connected CompoMat G5. The status bar on the top right of the screen should show the following: = CompoMaster Net G5 connected, process data activated. or = CompoMaster Net G5 connected, process data deactivated. Re: 3.14



Testing the network WLAN functionality Accessories: – WLAN stick (M67 478 1) – WLAN access point (M67 479 1) Preparation: Note: Ensure nothing is connected to the network port on the CompoMat G5. – Plug in the WLAN stick into a free USB port on the CompoMat G5. – Deactivate Ethernet interface on the PC on which the CompoMaster Net G5 is installed. Test: – Configure WLAN interface on the PC on which the CompoMaster Net G5 is installed as follows:



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-17



Chapter 6: TSC / TMC / MA



IP address:



192.168.0.250



Netmask:



255.255.255.0



Gateway:



(No entry)



DNS:



(No entry)



ESSID:



g5Net



Password:



CompoMasterNet



Encryption:



WPA/PSK



EncyptType:



TKIP



Mode:



Managed



– Start CompoMat G5 in user mode – Start CompoMaster Net G5. – Select configured WLAN interface in the CompoMaster Net G5. (See User Manual, CompoMaster Net G5, Configuration / CompoMat G5 Parameters) – Set up CompoMat G5 on CompoMaster Net G5 (Allocate serial number, IP address and network name). – Reboot CompoMaster Net G5. The CompoMaster Net G5 recognizes the connected CompoMat G5. The status bar on the top right of the screen should show the following: = CompoMaster Net G5 connected, process data activated. or = CompoMaster Net G5 connected, process data deactivated. Re: 3.15



Door / Door locks If the door is blocked, then the error message Error: Close Door should be generated. Following this, the correct closure of the door is tested. Accessories: – USB stick service G5 (M67 482 1) – Suitable implement for blocking the door (e.g. screwdriver handle) Caution Risk of damage to device surface by introducing sharp objects. Risk of scratching device surface. – Do not use any metallic or edged implements to block the door.



6-18



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Warning Risk of injury since door closes automatically. Crushing hazard – Do not reach into the movement area of the door with your hands. Preparation: – Service program: Select Serv. Door. (see chapter 5.2.6 on page 5-9) Test: – Prevent door closure by introducing a suitable object between the door lock and the door. – Press the



(green) function key ⇒ Door closes.



– An audible warning is sounded. – Door re-opens automatically. The screen message shown to the left will be displayed: Error / Error: Close Door – Press the



(green) function key.



The audible warning sound is switched off. The screen message Ready to use is displayed. – Remove blockage between lock case and door.



Check that the door closes properly:



– Press the



(green) function key again ⇒ Door closes.



Program step Wait for Start is displayed. – Manual opening of the door must not be possible. – No air gap must be visible between the door lock and door. – Press the



(green) function key.



The screen message Ready to use is displayed.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-19



Chapter 6: TSC / TMC / MA



Re 3.16



Closing characteristics of the door with a bag The ability of the door to close properly under pressure is tested by placing a filled dual-bag system inside. Accessories: – USB stick service G5 (M67 482 1) – Dual-bag system (M63 925 1) – Tube clamp Preparation: – Service program: Select Serv. Door. (see chapter 5.2.6 on page 5-9) Test: – Allow the entire liquid content of the dual-bag system to run into one bag. – Close the tubing with a tube clamp. – Hang up the filled bag in the press area so that the tube hangs below. – Place the bag suspension in the front notch of the suspension pins. – Insert the tube over the pulley into Head 6. Caution Damage to press scales with tubing under tension. Overloading the press scales – Do not pull the tube when placing it over the pulley and when inserting Head 6. – Press the



(green) function key.



The door closes and locks without an error message. – Press the



(green) function key.



The screen message Ready to use is displayed. Re: 3.17



Top press force The press force of the top press is measured by placing a force measuring device into the top press. Accessories: – USB stick service G5 (M67 482 1) – Tool for force measurement G5 (M67 228 1) – Calibrated force measuring device incl. pressure sensor (M67 449 1) Preparation: – Service program: Select Serv. TP force. (see chapter 5.2.6 on page 5-9)



6-20



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Test: – Insert the pressure sensor underneath into the G5 force measuring device (see arrow). Position the G5 force measuring device with the pressure sensor in the middle of the top press. – Press the



(green) function key.



The door closes, the top press is retracted. 1. Screen display: Top Press close until 500 N 2. Screen display: Toppress force 500 N – After 10 s, note the value on the force measuring device and enter this into the report. – Press the



(green) function key.



The screen message Ready to use is displayed. Re: 3.18



Upper press force The press force of the upper press is measured by placing a force measuring device into the upper press. Accessories: – USB stick service G5 (M67 482 1) – Tool for force measurement G5 (M67 228 1) – Calibrated force measuring device incl. pressure sensor (M67 499 1) Preparation: – Service program: Select Serv. UP force. (see chapter 5.2.6 on page 5-9) Test: – Insert the pressure sensor at the top into the G5 force measuring device (see arrow). – Position the G5 force measuring device with pressure sensor centrally on the door (see illustration). – Press the



(green) function key.



The door closes, the upper press is extended. 1. Screen display: Upper Press close until 500 N 2. Screen display: Upper Press force 500 N – After 10 s, note the value on the force measuring device and enter this into the report. – Press the



(green) function key.



The screen message Ready to use is displayed.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-21



Chapter 6: TSC / TMC / MA



Re: 3.19



Lower press force The press force of the lower press is measured by placing a force measuring device into the lower press. Accessories: – USB stick service G5 (M67 482 1) – Tool for force measurement G5 (M67 228 1) – Calibrated force measuring device incl. pressure sensor (M67 499 1) Preparation: – Service program: Select Serv. LP force. (see chapter 5.2.6 on page 5-9) Test: – Insert the pressure sensor underneath into the G5 force measuring device (see arrow). – Position the G5 force measuring device with pressure sensor centrally on the door (see illustration). – Press the



(green) function key.



The door closes, the lower press is extended. 1. Screen display: Lower Press close until 500 N 2. Screen display: Lower Press force 500 N – After 10 s, note the value on the force measuring device and enter this into the report. – Press the



(green) function key.



The screen message Ready to use is displayed. Re: 3.20



Test of press scales / PLS scales / RCC scales Accessories: – Calibration weight 500 g (M63 925 1) Preparation: – Service program: Select Serv. Press scale, Serv. PLS scale or Serv. RCC scale. (see chapter 5.2.6 on page 5-9)



Testing the PLS scales



– Press the



(green) function key.



The following screen message is displayed: Processing... Press START for weighing Scale: PLS-Scale Weight: XXX g – Enter the weight shown (XXX g) into the report – Load the PLS scales with 500 g calibration weight. – Press the



6-22



(green) function key.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



The weight determined is transferred to the CompoMaster Net G5 and appears in the PLS scales field. – Enter the weight displayed into the report Testing the RCC scales



The following screen message is displayed: Processing... Press START for weighing Scale: RCC-Scale Weight: XXX g – Load the RCC scales with 500 g calibration weight. – Enter the weight shown (XXX g) into the report – Press the



(green) function key.



The weight determined is transferred to the CompoMaster Net G5 and appears in the RCC scales field. – Enter the weight displayed into the report The door opens. Testing the press scales



The following screen message is displayed: Processing... Press START for weighing Scale: Press Scale Weight: XXX g – Enter the weight shown (XXX g) into the report – Load the press scales with 500 g calibration weight. – Press the



(green) function key.



The weight determined is transferred to the CompoMaster Net G5 and appears in the Press scales field. – Enter the weight displayed into the report The screen message Ready to use is displayed. Re: 3.21



Slide occlusion Accessories: – Dual-bag system (M63 924 1) Preparation: – Service program: Select Serv. Slide occl. or Serv. Slide occl. no PLS. (see chapter 5.2.6 on page 5-9) Test: – Allow the entire liquid content of the dual-bag system to run into one bag. – Hang the dual-bag system on the suspension pins in the press area of the upper press. – Place the empty bag in the PLS scales plasma deaeration unit and place the connection tubing into clamp 1. – Press the



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



(green) function key.



6-23



Chapter 6: TSC / TMC / MA



The door closes. The display then shows the following sequence of messages: – – – – – –



Both Presses out until 150 N Both Presses to 45.0 mm Both Presses out until 650 N Slide out Wait 60 s Press START for weighing



– Press the



(green) function key.



– In the Process Data window (Scales weights/PLS scale), note the first value = reference value. – Lower Press out for 100 s – Press START for weighing – Press the



(green) function key.



– In the Process Data window (Scales weights/PLS scale), note the second value = reference value. – Enter the difference between the reference value and measurement value in the report. Re: 3.22



Test CF-Opener Top, CF-Opener Door and CF-Opener RCC Accessories: – USB stick service G5 (M67 482 1) – Test tool 3.6 mm (661 578 1) Preparation: – Service program: Select Serv. CompoFlow or Serv. CompoFlow no RCC. (see chapter 5.2.6 on page 5-9)



Test CF-Opener Top



– Press the



(green) function key ⇒ Door closes.



Screen message Processing... CF-Opener Top is displayed. – Press the



(green) function key.



Screen message Error / Error: CF-Opener is empty! is displayed, audible signal is sounded. – Press the



(green) function key.



The audible warning sound is switched off. Screen message Pause... Open CF-Opener Top is displayed, – Place 3.6 mm test tool in the CF-Opener Top. – Press the



(green) function key.



The following screen message is displayed: Technical error! Actor: 0x03, Action: 0x17 ERROR-CODE: 0x0401 – Press the



6-24



(green) function key.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Screen message Pause... Open CF-Opener Top is displayed, – Press the



(yellow) function key.



Screen message Processing... CF-Opener Door is displayed. Test CF-Opener door



– Press the



(green) function key.



Screen message Processing... CF-Opener Door is displayed. – Press the



(green) function key.



Screen message Error / Error: CF-Opener is empty! is displayed, audible signal is sounded. – Press the



(green) function key.



The audible warning sound is switched off. Screen message Pause... Open CF-Opener Door is displayed, – Place test tool 3.6 mm in the CF-Opener door – Press the



(green) function key.



The following screen message is displayed: Technical error! Actor: 0x04, Action: 0x17 ERROR-CODE: 0x0401 – Press the



(green) function key.



Screen message Pause... Open CF-Opener Door is displayed. – Press the



(yellow) function key.



Screen message Processing... CF-Opener RCC is displayed. Test of CF-Opener RCC (optional)



– Press the



(green) function key.



Screen message Processing... CF-Opener RCC is displayed. – Press the



(green) function key.



Screen message Error / Error: CF-Opener is empty! is displayed, audible signal is sounded. – Press the



(green) function key.



The audible warning sound is switched off. Screen message Pause... Open CF-Opener RCC is displayed. – Place test tool 3.6 mm in the CF-Opener RCC. – Press the



(green) function key.



The following screen message is displayed: Technical error! Actor: 0x05, Action: 0x17 ERROR-CODE: 0x0401 – Press the



(green) function key.



Screen message Pause... Open CF-Opener RCC is displayed. – Press the



(yellow) function key.



The program is terminated, the door opens.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-25



Chapter 6: TSC / TMC / MA



Re: 3.23



Multi-detector field (detectors A1 to A8) Accessories: – USB stick service G5 (M67 482 1) – Black foam mat (661 545 1)



Preparation:



– Service program: Select Serv. Detector A. (see chapter 5.2.6 on page 5-9) Test: – Press the



(green) function key.



The door closes, the program starts. The screen message Lower Press out until Det A1 is displayed. The lower press is extended slowly. – Hold a black foam mat in front of the A1 detector field. The lower press is retracted. The screen message Lower Press out until Det A2 is displayed. – Hold a black foam mat in front of the A2 detector field. The lower press is retracted. The screen message Lower Press out until Det A3 is displayed. – Hold a black foam mat in front of the A3 detector field. The lower press is retracted. The screen message Lower Press out until Det A4 is displayed. – Hold a black foam mat in front of the A4 detector field. The lower press is retracted. The screen message Lower Press out until Det A5 is displayed. – Hold a black foam mat in front of the A5 detector field. The lower press is retracted. The screen message Lower Press out until Det A6 is displayed. – Hold a black foam mat in front of the A6 detector field. The lower press is retracted. The screen message Lower Press out until Det A7 is displayed. – Hold a black foam mat in front of the A7 detector field. The lower press is retracted. The screen message Lower Press out until Det A8 is displayed. – Hold a black foam mat in front of the A8 detector field. The lower press is retracted, the door opens. Re: 3.24



Opening / closing the Heads (without tubes) Accessories: – USB stick service G5 (M67 482 1)



6-26



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Preparation: – Service program: Select Serv. O./C. heads or Serv. O./C. heads no PLS . (see chapter 5.2.6 on page 5-9) Test: – Press the



(green) function key.



The door closes, the program starts. Screen message Processing... Close head 1 is displayed. – Press the



(green) function key.



Head 1 is closed. Screen message Processing... Close head 2 is displayed. – Press the



(green) function key.



Head 2 is closed. Optional: Screen message Processing... Close head 3 is displayed. – Press the



(green) function key.



Head 3 is closed. Screen message Processing... Close head 4 is displayed. – Press the



(green) function key.



Head 4 is closed. Screen message Processing... Close head 5 is displayed. – Press the



(green) function key.



Head 5 is closed. Screen message Processing... Close head 6 is displayed. – Press the



(green) function key.



Head 6 is closed. Screen message Processing... Open head 1 is displayed. – Press the



(green) function key.



Head 1 is opened. Screen message Processing... Open head 2 is displayed. – Press the



(green) function key.



Head 2 is opened. Optional: Screen message Processing... Open head 3 is displayed. – Press the



(green) function key.



Head 3 is opened. Screen message Processing... Open head 4 is displayed. – Press the



(green) function key.



Head 4 is opened. Screen message Processing... Open head 5 is displayed.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-27



Chapter 6: TSC / TMC / MA



– Press the



(green) function key.



Head 5 is opened. Screen message Processing... Open head 6 is displayed. – Press the



(green) function key.



Head 6 is opened. Screen message Processing... Close all heads is displayed. – Press the



(green) function key.



All Heads are closed. The screen message Processing... open all heads is displayed. All Heads are opened. Screen message Processing... Wait for START is displayed. – Press the



(green) function key.



The program is terminated, the door opens. Re: 3.25



Testing the Heads (with tubings) Accessories: – USB stick service G5 (M67 281 1) – Empty 3.1 x 4.2 mm PVC tubing or six short tubing sections (each approx. 10 cm) (M60 024 1) Preparation: – Service program: Select Serv. Check heads or Serv. Check heads no PLS . (see chapter 5.2.6 on page 5-9) Test: – Press the



(green) function key.



The door closes, the program starts. Screen message Processing... No tubes in heads is displayed. – Press the



(green) function key.



The screen message shown to the left will be displayed: Error / Error: Check Head Inspect 1 2 3 4 5 6 – Press the



(green) function key.



– Insert the tubing into Head 1. – Press the



(green) function key.



Screen message Error / Error: Check Head Inspect 2 3 4 5 6 is displayed. – Remove the tubing from Head 1 and insert it into Head 2. – Press the



(green) function key.



Screen message Error / Error: Check Head Inspect 1 3 4 5 6 is displayed.



6-28



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



– Remove the tubing from Head 2 and insert it into Head 3, if the PLC scales are installed. If not, insert the tubing into Head 4. – Press the



(green) function key.



Screen message Error / Error: Check Head Inspect 1 2 4 5 6 is displayed. – Remove the tubing from Head 3 and insert it into Head 4. – Press the



(green) function key.



Screen message Error / Error: Check Head Inspect 1 2 3 5 6 is displayed. – Remove the tubing from Head 4 and insert it into Head 5. – Press the



(green) function key.



Screen message Error / Error: Check Head Inspect 1 2 3 4 6 is displayed. – Remove the tubing from Head 5 and insert it into Head 6. – Press the



(green) function key.



Screen message Error / Error: Check Head Inspect 1 2 3 4 5 is displayed. – Insert a tubing into all Heads. – Press the



(green) function key.



The program is terminated without an error message, the door opens. Re: 3.26



Sealing behavior of the Heads Accessories: – USB stick service G5 (M67 482 1) – A 4.2 mm tube (approx. 50 cm long) filled with saline solution (0.9% NaCl) for making corresponding pieces of tubing each with a length of 80 mm – Vise with a jaw length of 50 mm – Caliper (M67 963 1) – Spring balance Preparation: – Service program: Select Serv. Seal heads or Serv. Seal heads no PLS . (see chapter 5.2.6 on page 5-9) Test: – Insert a tubing into all Heads. – Press the



(green) function key.



The door closes, the program starts. The tubings in all Heads are sealed. The program is terminated without any errors. – Remove the tubing from Head 1 and check it: Sealing seam successfully passes a visual inspection.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-29



Chapter 6: TSC / TMC / MA



Separation at the sealing seam is correct. – Remove the tubing from Head 2 and check it: Sealing seam successfully passes a visual inspection. Separation at the sealing seam is correct. – Remove the tubing from Head 3 and check it: Sealing seam successfully passes a visual inspection. Separation at the sealing seam is correct. – Remove the tubing from Head 4 and check it: Sealing seam successfully passes a visual inspection. Separation at the sealing seam is correct. – Remove the tubing from Head 5 and check it: Sealing seam successfully passes a visual inspection. Separation at the sealing seam is correct. – Remove the tubing from Head 6 and check it: Sealing seam successfully passes a visual inspection. Separation at the sealing seam is correct. Test criteria for sealing seam:



1. Visual inspection:



The sealing seam must have a symmetrical shape on both, its bottom and top sides, as well as to the left and right of the cut. There must be no sharp edges or leaks in the transitional area between tubing and sealing seam. 2. Tensile load:



1



2



3



1 Sealed length of tubing 2 Sealing seam 3 Spring balance – Allow the sealing seam to cool down for 2 min. – Pull the two segments of the line apart smoothly and uniformly, using the spring balance. The tensile loading capacity of the sealing seam should be between 10–50 N (1–5 kg).



6-30



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



3. Pressure test: Note: The pressure test only functions with a completely filled tubing (no air bubbles), as only then can a correspondingly high pressure build up in the tubing. Two balloons form during the pressure test. Clamp the section symmetrical between the clamping jaws of the vise. The tubing must project for 15 mm at both ends of the jaws (see following illustration).



With the bench vise tightened, no sealing seams should be opened and no saline solution should escape. Re: 3.27



Head sensors Accessories: – USB stick service G5 (M67 482 1) – Gray wedge 13 levels (679 514 1) Preparation: – Service program: Select Serv. Head det. or Serv. Head det. no PLS. (see chapter 5.2.6 on page 5-9) Test:



Head 1



– Press the



(green) function key.



The door closes. Screen message Processing... Grey wedge in Head 1 | 30 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



(green) function key.



6-31



Chapter 6: TSC / TMC / MA



Screen message Processing... Close Head 2 at signal / Head detector 1 | red | 30% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 2 closes. Screen message Processing... Grey wedge in head 1 | 70 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 2 at signal / Head detector 1 | red | 70% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 2 closes. Head 2



Screen message Processing... Grey wedge in head 2 | 30 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 3 at signal / Head detector 2 | red | 30% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 3 closes. Screen message Processing... Grey wedge in head 2 | 70 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 3 at signal / Head detector 2 | red | 70% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 3 closes. Head 3



Screen message Processing... Grey wedge in head 3 | 30 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 4 at signal / Head detector 3 | red | 30% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 4 closes. Screen message Processing... Grey wedge in head 3 | 70 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



6-32



(green) function key.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Screen message Processing... Close Head 4 at signal / Head detector 3 | red | 70% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 4 closes. Head 4



Screen message Processing... Grey wedge in head 4 | 30 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 5 at signal / Head detector 4 | red | 30% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 5 closes. Screen message Processing... Grey wedge in head 4 | 70 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 5 at signal / Head detector 4 | red | 70% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 5 closes. Head 5



Screen message Processing... Grey wedge in head 5 | 30 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 6 at signal / Head detector 5 | red | 30% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 6 closes. Screen message Processing... Grey wedge in head 5 | 70 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 6 at signal / Head detector 5 | red | 70% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 6 closes. Head 6



Screen message Processing... Grey wedge in head 6 | 30 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



(green) function key.



6-33



Chapter 6: TSC / TMC / MA



Screen message Processing... Close Head 1 at signal / Head detector 6 | red | 30% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 1 closes. Screen message Processing... Grey wedge in head 6 | 70 is displayed. – Place a 13-level step wedge, beginning at 0, into the Head. – Press the



(green) function key.



Screen message Processing... Close Head 1 at signal / Head detector 6 | red | 70% is displayed. – Increase the grey level by moving the step wedge. – Note the grey value at which Head 1 closes. Re: 3.28



Behavior during bag press-out Accessories: – USB stick service G5 (M67 482 1) – Dual-bag system (M63 924 1) Preparation: – Service program: Select Serv. Empty bag. (see chapter 5.2.6 on page 5-9) Test: – Attach dual-bag system. – Place the empty bag under the top press. – Place the connection tubing into Head 1. – Press the



(green) function key.



The door closes. Both presses are extended until the bag is touched. Both presses are retracted to 55 mm. The lower press is fully extended. The slide is extended. The upper press is fully extended. Head 1 closes. ⇒ The bag in the press must have been pressed out completely. Screen message Processing... Wait for START is displayed. – Press the



(green) function key.



Both presses and the slide are fully retracted. Head 1 opens. Top press closes. Head 1 closes. ⇒ The dual-bag system in the top press must have been pressed out completely.



6-34



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Screen message Processing... Wait for START is displayed. – Press the



(green) function key.



Head 1 opens. Top press opens. The program is terminated without an error message, the door opens. Re: 3.30



Extended test CF-Opener Top, CF-Opener Door and CFOpener RCC Note: A service program is not required for this test. Accessories: – 9 CompoFlows Preparation: – Disconnect CompoMat G5 from the customer’s network and reboot without the service stick. – Connect the service technician’s CompoMaster Net G5 using a crossover network cable. Test: – Program the following processes for CF-Opener Top, door and RCC.



– 9 CompoFlows are to be marked (TOP_01 ... TOP_03, DOOR_01 ... DOOR_03, RCC_01 ... RCC_03). – Open 3 CompoFlows using the related CF-Opener.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-35



Chapter 6: TSC / TMC / MA



The CompoFlow in the tubing is completely opened (illustration on the left).



Re: 3.31



System information (see chapter 5.6.5 on page 5-34)



6.4.2



Explanations on electrical safety checks Re: 4



Check of the electrical safety: – In Germany according to DIN EN 62353 (DIN VDE 0751-1; edition 08/2008) – In other countries, observe the local regulations!



Re: 4.1



Perform visual inspection. – Check the fuses accessible from the outside Melting point of the fuses (2x) T 10 A / 230/240 V AC. – Labels and inscriptions are present and legible. – The mechanical condition permits further safe use. – There are no signs of damage or contamination. – The power cable is not damaged.



Re: 4.2



Protective earth resistance max. 0.3 Ω (with power cable) Measurement point: – Screws for attaching the housing sheet (see arrow)



6-36



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 6: TSC / TMC / MA



Re: 4.3



Measurement of the leakage current (device leakage current) Differential current measurement acc. to DIN EN 62353, Figure 5



or Direct measurement acc. to DIN EN 62353, Figure 4



Basic conditions: – Measurement of the protective earth resistance performed. – When performing a direct measurement, the following precautions also must be observed: The device must be insulated when installed. All external connections must have been removed from the device. Additional conditions: If the value scaled to the nominal voltage is higher than 90% of the admissible alarm limit (= 450 µA), the last measured value or the first measured value must additionally be considered for the rating. If necessary, reduce the interval for the regular test.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



6-37



Chapter 6: TSC / TMC / MA



Re: 4.4



Insulation resistance measurement > 2 MΩ Measurement between the active parts and any conductive part that can be touched, including the protective earth conductor.



Confirmation of the test



Test equipment used: Type and serial number of the test equipment used. Comments: Any irregularities which occurred during the test are documented in this section. Date, signature, stamp Performance of the check must be confirmed with the date, the signature of the person performing the check and a stamp.



Assessment of the test



The device has been released for its intended use. During the intended use of the device it must be ensured that the device does not present a hazard to patients, employees or other third parties. As part of overall evaluation, testers are obliged to give their official approval for the use of the device. If defects are identified, then the device owner must be informed of these without delay. Date of next inspection: The next inspection date has to be entered in the report. The intervals specified by the manufacturer have to be observed. Comments: Any irregularities which occurred during the assessment are documented in this section. Date, signature, stamp: Assessment of the initial start-up has to be confirmed with date, tester's signature and stamp.



6-38



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 7: Error Messages



7



Error Messages



7.1



CompoMat G5 Error Codes



Error message



Possible cause



Action required



0x0001



Internal error XXX



Turn off/on device



0x0002



Internal error XXX



Turn off/on device



0x0003



– Indicates presence of a syntax error in a programmed procedure. – CompoMaster Net G5 version incompatible with device version.



Turn off/on device



0x0004



– Indicates presence of a syntax error in a programmed procedure. – CompoMaster Net G5 version incompatible with device version.



Turn off/on device



0x0005



– Indicates presence of blockage or defect involving the presses.



Turn off/on device



In relation to other actuators (e.g. CF-Opener, door, clamp, etc.): – Indicates presence of defect in actuator 0x0006



Internal error XXX



Turn off/on device



0x0007



Internal error XXX



Turn off/on device



0x0008



Internal error XXX



Turn off/on device



0x0009



Internal error XXX



Turn off/on device



0x000A



– Indicates presence of a syntax error in a programmed procedure. – CompoMaster Net G5 version incompatible with device version.



Turn off/on device



0x000B



Internal error XXX



Turn off/on device



0x000C



Internal error XXX



Turn off/on device



0x0100



Internal error XXX



Turn off/on device



0x0101



Internal error XXX



Turn off/on device



0x0200



Internal error; ADC not calibrated



ADC must be re-calibrated. If the error continues to occur, then this may indicate that the Application Board bought-in assembly is defective.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



7-1



Chapter 7: Error Messages



Error message



Possible cause



Action required



0x0201



Internal error; ADC not calibrated



ADC must be re-calibrated. If the error continues to occur, then this may indicate that the Application Board bought-in assembly is defective.



0x0202



Internal error; digit < 2300 or digit > 2700



Check scales, re-calibrate device as required



0x0203



Internal error; digit < 1350 or ADCValueSum > 1750



Check scales, re-calibrate device as required



0x0204



Internal error XXX



ADC must be re-calibrated. If the error continues to occur, then this may indicate that the Application Board bought-in assembly is defective.



0x0400



Internal error XXX



Turn off/on device



0x0401



CF-Opener is blocked.



Turn off/on device



0x0402



CompoFlow has not been opened



Turn off/on device



0x0404



CompoFlow has been opened



Turn off/on device



0x0500



Scales are not calibrated



Check scales, re-calibrate device as required



0x0501



Internal error; communication with scales disrupted



Check scales, re-calibrate device as required



0x0502



Internal error; calibration value not plausible



Check scales, re-calibrate device as required



0x0503



Internal error; calibration value not plausible



Check scales, re-calibrate device as required



0x0600



Internal error XXX



Check Heads



0x0601



Sealing defective



Check Heads, insert tubing



0x0602



Error during Head check



Check Heads, insert tubing



0x0604



Erroneous ADC value from head sensors



Check head sensor



0x0607



Erroneous ADC value from head sensors during calibration



Check head sensor



0x0900



Internal error XXX



Check Trinamic Board bought-in assembly



0x0901



Fault on the Trinamic Board bought-in assembly, Application Board bought-in assembly, or software error.



Turn off/on device



0x0902



Step motor out of step.



If this error persists, then the corresponding unit must be verified/replaced.



0x0903



Calibration error



Check presses



7-2



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 7: Error Messages



Error message



Possible cause



Action required



0x0905



Door not closing during press movement



Check door sensor and door motor



0x0B00



Not possible to calibrate photodetection board



Check PDB



0x8001



Internal error ADC timeout



Re-calibrate ADC.



0x8002



5V in the range < 4500 or > 5500 mV Possible defect in power adapter or the Application Board bought-in assembly.



Change Application Board bought-in assembly (M68 137 1)



0x8004



12 V within range 11000 13000 mV Possible defect in power adapter or the Application Board bought-in assembly.



Change Application Board bought-in assembly (M68 137 1)



0x8008



24 V within range 23000 25000 mV Possible defect in power adapter or the Application Board bought-in assembly.



Change Application Board bought-in assembly (M68 137 1)



0x8010



Internal error; I2C bus 0 defective



Turn off/on device



0x8020



Internal error; I2C bus 0 defective



Turn off/on device



0x8040



Internal error; I2C bus 1 defective



Turn off/on device



0x8080



Internal error; I2C bus 1 defective



Turn off/on device



0xFFFF 0x7001



Turn off/on device Red cells have been detected by a head detector The error appears in plain text on the display: Error Head sensor x alarm This is a monitoring function that has been programmed into the procedure.



0x7010



Fresenius Kabi



The error appears in plain text on the display: Error: No weight rising on PLS Scale! e.g. – Kinks in the tubing; – clamp closed; – breaker not opened. – etc.



CompoMat G5



SM-EN



3/11.10



Confirm with the key.



(green) function



Then assess whether the procedure should be aborted or should continue. – Correct the problem; – Confirm with the (green) function key.



7-3



Chapter 7: Error Messages



7.2



Actuator, action and error codes



7.2.1



Example In the example on the left the following error is indicated: Actuator: 0x1d = AS_Slide Action: 0x03 = AC_Close Error-Code: 0x0200 = ERR_ADC_NOT_CALIBRATED



7.2.2



Error codes



Significance



Error codes



ERR_UNKNOWN



0x0001



ERR_OUTOFMEMORY



0x0002



ERR_INVALID_ACTOR



0x0003



ERR_INVALID_ACTION



0x0004



ERR_TIMEOUT 0X0005



0x0005



ERR_RESOURCE_INUSE



0x0006



ERR_RESOURCE_STOPPED



0x0007



ERR_PERIPH_INUSE



0x0008



ERR_ARGUMENT_OUTOFRANGE



0x0009



ERR_INVALID_PARAMETER



0x000A



ERR_PROCESSES_STILLACTIVE



0x000B



ERR_HIGHPOWERDEVICE_INUSE



0x000C



ERR_EEPROM_WRITE



0x0100



ERR_EEPROM_READ



0x0101



ERR_ADC_NOT_CALIBRATED



0x0200



ERR_ADCOFFSET_NOT_CALIBRATED



0x0201



ERR_ADC_LOW_CALIBRATION



0x0202



7-4



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 7: Error Messages



Significance



Error codes



ERR_ADC_HIGH_CALIBRATION



0x0203



ERR_ADC_OFFSET_CALIBRATION



0x0204



ERR_DOOR_LOCK



0x0300



ERR_DOOR_UNLOCK



0x0301



ERR_DOOR_CURRENT



0x0302



ERR_DOOR_STOPPED



0x0303



ERR_DOOR_SENSOR



0x0304



ERR_BREAKER_NOTAVAILABLE



0x0400



ERR_BREAKER_CURRENT



0x0401



ERR_BREAKER_NOT_OPEN



0x0402



ERR_BREAKER_CLOSURE_NOT_OPEN



0x0403



ERR_BREAKER_TUBE_OPEN



0x0404



ERR_SCALE_NOT_CALIBRATED



0x0500



ERR_SCALE_OFFLINE



0x0501



ERR_SCALE_INVALID_CALIBRATION



0x0502



ERR_SCALE_CALIBRATE_DELTA



0x0503



ERR_SCALE_NOTSTABLE



0x0504



ERR_HEAD_NOT_AVAILABLE



0x0600



ERR_HEAD_SEAL_FAILED



0x0601



ERR_HEAD_CHECK_FAILED



0x0602



ERR_HEADDETECTOR_BLINKING



0x0603



ERR_HEADDETECTOR_CALCULATION



0x0604



ERR_HEADDETECTOR_INUSE



0x0605



ERR_HEAD_INUSE



0x0606



ERR_HEADDETECTOR_CALIBRATION



0x0607



ERR_COMPRESSOR_TIMEOUT



0x0700



ERR_COMPRESSOR_LOW



0x0701



ERR_COMPRESSOR_PERMANENT_TIMEOUT



0x0702



ERR_SLIDE_NOTAVAILABLE



0x0800



ERR_SLIDE_CURRENT



0x0801



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



7-5



Chapter 7: Error Messages



Significance



Error codes



ERR_PRESS_TRINAMICCOMMAND



0x0900



ERR_PRESS_NOTINPROCESSMODE



0x0901



ERR_PRESS_CALIBRATE_DELTA



0x0902



ERR_PRESS_CALIBRATE_FORCE



0x0903



ERR_PRESS_VERIFY_DELTA



0x0904



ERR_PRESS_DOORNOTCLOSED



0x0905



ERR_PDB_CALIBRATION



0x0B00



ERR_STATE_ADC_ERROR



0x8001



ERR_STATE_5V



0x8002



ERR_STATE_12V



0x8004



ERR_STATE_24V



0x8008



ERR_STATE_I2C0_SCL



0x8010



ERR_STATE_I2C0_SDA



0x8020



ERR_STATE_I2C1_SCL



0x8040



ERR_STATE_I2C1_SDA



0x8080



ERR_PROCESSING



0xFFFF



ARM9_ERR_RED_CELL



0x7001



ARM9_ERR_TIMEOUT_MUTE



0x7005



ARM9_ERR_PLS_MONITORING



0x7010



ARM9_ERR_RCC_MONITORING



0x7015



ARM9_ERR_DET_A_COVERED



0x7020



7.2.3



Actuator codes



Actuator codes



Significance



0x00



AS_Null=0



0x01



AS_All



0x02



AS_Both_Presses



0x03



AS_Breaker_Top



Breaker_0



0x04



AS_Breaker_Front



Breaker_1



7-6



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 7: Error Messages



Actuator codes



Significance



0x05



AS_Breaker_RCC



Breaker_2



0x06



AS_Det_0



Head 1



0x07



AS_Det_1



Head 2



0x08



AS_Det_2



Head 3



0x09



AS_Det_3



Head 4



0x0a



AS_Det_4



Head 5



0x0b



AS_Det_5



Head 6



0x0c



AS_Det_6



Head 7



0x0d



AS_Det_7



Head 8



0x0e



AS_Det_A0



Multidetection board 1



0x0f



AS_Det_A1



Multidetection board 2



0x10



AS_Det_A2



Multidetection board 3



0x11



AS_Det_A3



Multidetection board 4



0x12



AS_Det_A4



Multidetection board 5



0x13



AS_Det_A5



Multidetection board 6



0x14



AS_Det_A6



Multidetection board 7



0x15



AS_Det_A7



Multidetection board 8



0x16



AS_Door



0x17



AS_Head



0x18



AS_Lower_Press



0x19



AS_Process_Mode



0x1a



AS_Scale_Press



Scale_0



0x1b



AS_Scale_RCC



Scale_1



0x1c



AS_Scale_Plasma



Scale_2



0x1d



AS_Slide



0x1e



AS_Sync_Presses



0x1f



AS_Top_Press



0x20



AS_Upper_Press



0x21



AS_BARCODE



0x22



AS_RFID



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



7-7



Chapter 7: Error Messages



Actuator codes



Significance



0x23



AS_EEProm



0x24



AS_Production



0x25



AS_Deaerate



0x26



AS_Debug



0x27



AS_Trincamic



0x28



AS_Counter



0x29



AS_Breaker_3



Breaker_3 NOT AVAILABLE



0x2a



AS_Breaker_4



Breaker_4 NOT AVAILABLE



0x2b



AS_Breaker_5



Breaker_5 NOT AVAILABLE



7.2.4



Action codes



Action codes



Significance



0x00



AC_Null=0



0x01



AC_Calibrate



0x02



AC_Check



0x03



AC_Close



0x04



AC_Close_Det



0x05



AC_Close_Weight



0x06



AC_Close_Delta



0x07



AC_Close_Det_A



0x08



AC_Close_Force



0x09



AC_Close_Scale_0



0x0a



AC_Close_Scale_1



0x0b



AC_Close_Scale_2



0x0c



AC_Get_Cal



0x0d



AC_Get_State



0x0e



AC_Get_Timeout



0x0f



AC_Get_Val



0x10



AC_Home



7-8



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 7: Error Messages



Action codes



Significance



0x11



AC_Seal



0x12



AC_Set_Cal



0x13



AC_Set_Timeout



0x14



AC_Set_Val



0x15



AC_Stop



0x16



AC_To_Pos



0x17



AC_Open



0x18



AC_Open_Close



0x19



AC_Open_Delta



0x1a



AC_Open_Force



0x1b



AC_Enable



0x1c



AC_Disable



0x1d



AC_Set_Limit



0x1e



AC_Get_Limit



0x1f



AC_Reset



0x20



AC_Init



0x21



AC_Flash



0x22



AC_Verify



0x23



AC_Debug_0



0x24



AC_Debug_1



0x25



AC_Start



0x26



AC_Wait



0x27



AC_Wait_for_Start



0x28



AC_Display_Text



0x29



AC_Close_Time



0x2a



AC_Close_with_Check



0x2b



AC_Take_Tare



0x2c



AC_Close_Det_A_PLS



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



7-9



Chapter 8: Calibration / adjustment



8



Calibration / adjustment



8.1



Calibration (see chapter 5.4 on page 5-13)



8-1



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



9



Servicing / repair



9.1



Description / warnings Caution CompoMat G5 and CompoMaster Net G5 may only be operated by qualified and specialized personnel. Caution The CompoMat G5 may only be opened by qualified and skilled staff. Caution Operators are wholly responsible for any processes that they create in CompoMaster Net G5 and which are executed by the CompoMat G5. This applies in particular to possible loss of blood and blood products by incorrect programming. See the User Manual for CompoMaster Net G5 for information about using the CompoMaster Net G5 to program the CompoMat G5. Caution Operators are wholly responsible for preparatory tasks, such as storage, centrifuging, transport to the CompoMat G5, suspending on the pins and snapping any possible rated break points, all of which can have a significant influence on the quality of separation of the blood components. For these reasons, Fresenius Kabi does not provide information in advance concerning the quality of results as regards blood components obtained using CompoMat G5 processes. Fresenius Kabi only claims that there is a relatively high reproducibility of results for processed blood components within a relatively short preparation time. Please refer to the relevant literature for specific quality results for blood components processed with the CompoMat G5 in daily practice at a blood bank.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-1



Chapter 9: Servicing / repair



9.2



Electrical hazards Warning Risk of injury caused by electrical voltage. Electric shock – When connecting the CompoMat G5 to a mains supply, the local regulations that apply have to be observed. Warning Risk of injury caused by electrical voltage. Electric shock – Improper commissioning and use of the electrical equipment can result in injuries, such as burns and electrical shocks. Warning Risk of injury caused by electrical voltage. Electric shock – When using safety class I systems, the quality of the protective ground of the installation is of particular importance. It must be taken into consideration that in many countries regulations have been enacted by national authorities. The rear panel of the CompoMat G5 is provided with a connection for a grounded power cable. Warning Risk of injury caused by electrical voltage. Electric shock – Disconnect the power plug before opening the CompoMat G5. Warning Risk of injury caused by electrical voltage. Electric shock If the power cable needs to be replaced, use only the original power cable listed in the spare parts catalog. If additional equipment, which is not included in the accessories, is connected to the device, there will be a danger that the permissible leakage currents will be exceeded.



9-2



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



Caution Loss of the preparation when disconnecting the power plug or in the event of a power failure. – If there is a loss of power while processing a program in the CompoMat G5, all sealing heads/clamps will open and the presses will stop. This may cause a loss of the preparation.



9.3



Mechanical hazards Warning Risk of injury caused by parts moving autonomously. Crushing hazard – To avoid injury, operators must ensure to keep a sufficiently large distance to all moving parts of the CompoMat G5 . Moving parts are: – – – – – –



Door upper press lower press slide Top press BG-CompoFlow Opener



Warning Risk of injury caused by parts moving autonomously. Crushing hazard – Keep hands clear of the inside of the CompoMat G5 during operation of the device.



9.4



Risk of burning Warning Risk of burning by touching the sealing electrodes. – Do not touch the sealing electrodes of Heads 1-6 during the sealing process. – Do not place any live parts into the Heads. – Maintain a sufficient distance from the sealing electrodes in the Heads.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-3



Chapter 9: Servicing / repair



Warning Risk of injury by burning Skin injury – The devices are not suitable for use with flammable mixtures or nitrous oxide.



9.5



Biological hazards Warning Risk of infection by leaking blood bags. It is always possible that donated blood could be infected by germs of contagious diseases. It must always be treated as being potentially infectious. – Wear surgical or similar gloves when processing blood products. – Immediately clean and disinfect work surfaces that are contaminated with blood or blood products. The cleaning instructions must be observed! Observe your local laws and regulations concerning handling of potentially infectious material. Warning Risk of contamination by leaking tubes. Risk of infection – The operator should not exert any force on tubing, which is currently being sealed.



9-4



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



9.6



Replacements / Repairs



9.6.1



Replacing the Application Board bought-in assembly (M68 137 1) CompoMat G5 preparation



– Turn on the CompoMat G5 via the main switch. The device performs a self-test. – Wait until the top press and the door are open. – Switch off main switch on the CompoMat G5 so that the top press and the door remain open.



Measures required



– Remove PLS scales. – Loosen screws for the SUB-D connectors for the following connections: Balance A/B as well as Scanner twice.



– Loosen the four screws for the Application Board bought-in assembly as per the following illustration.



– To simplify subsequent assembly, mark all cables and connections. – Loosen all cables. – Loosen all four fastening screws for the Application Board bought-in assembly as per the following illustration.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-5



Chapter 9: Servicing / repair



– To simplify subsequent assembly, mark all cables and connections, in particular those for the Heads. – Loosen all cables. – Pull Application Board bought-in assembly to the side of the PLS scales out of the CompoMat G5 (see arrow).



The new Application Board bought-in assembly is fitted in the reverse order of removal. The following must be noted during this process: – On connection X90 head 1 starts on the right side. – Position 3 is not used.



9-6



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Apply Loctite 243 to the screws for the SUB-D connectors undone previously. – Then tighten screws to a torque of 0.8Nm. – Ensure the connections X2 and X31 are in the correct position. Settings, calibration and tests



– All calibrations and tests must be performed again as specified in the test report - annual maintenance (see chapter 6.2 on page 6-2). – Explanation of Test report - annual maintenance (see chapter 6.4.1 on page 6-14).



9.6.2



Replacing the Display G5 printed circuit board bought-in assembly (M68 115 1) Settings, calibration and tests



Once the Display printed circuit board bought-in assembly in the CompoMat G5 has been replaced, the following steps have to be taken: – Use the menu key



to switch off the CompoMat G5.



– Insert USB stick service G5 in a USB port in the CompoMat G5. Note: The software version of the service G5 USB stick must match the customer's software version! Entering the E-Code



– (see chapter 5.2.3.2 on page 5-4)



Entering the serial number



– (see chapter 5.2.3.1 on page 5-3) – The Service-Menu is displayed.Select Set E-Code. – Set E-Code is displayed.



– The display shows: Advice Calibration successful. Confirm with the (green) function key. – The Service Menu is displayed.Select Set Serial-Num.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-7



Chapter 9: Servicing / repair



– The display shows: Set Serial-Num Enter serial number – The display shows: Advice Calibration successful. Please restart. Confirm with the (green) function key. – The following message is displayed: Standby the CompoMat G5 enters standby mode. – Use the menu key



to switch on the CompoMat G5.



– All calibrations and tests must be performed again as specified in the test report - annual maintenance. (see chapter 6.2 on page 6-2) – Explanation of Test report - annual maintenance (see chapter 6.4.1 on page 6-14).



9.6.3



Replacing the Trinamic Board bought-in assembly Settings, calibration and tests



– All calibrations and tests must be performed again as specified in the test report - annual maintenance (see chapter 6.2 on page 6-2). – Explanations regarding test report - annual maintenance (see chapter 6.4.1 on page 6-14).



9-8



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



9.6.4



Changing the top press silicon mat G5 (M68 129 1) Removing the old silicon mat



– Lift the silicon mat by one corner and fold it back as far as the old adhesive will permit. – At this point, pull slowly and evenly in order to detach the adhesive.



Cleaning the top press contact area



– remove as much as possible of the residual adhesive without damaging the surface of the top press. – Clean the surface with an alcohol-based cleaning agent (e.g. Freka®-NOL). The cleaning instructions must be observed!



Preparing the replacement mat



– Use emery paper (grain min. 180) to lightly roughen the side of the replacement mat which is to be affixed with adhesive and then clean it again.



Applying the adhesive



– The silicon mat is to be glued to the top press contact area by applying silicon adhesive M67 879 1 (Elastosil E43, 90 ml tube) in 3 places. To do so, apply adhesive to the circular areas (diameter approx. 10 mm) as highlighted in the illustration and spread slightly. – Lay the silicon mat down flat and press down evenly, removing all noticeable ridges.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-9



Chapter 9: Servicing / repair



9.6.5



Replacing the BG-CompoFlow Opener Top and door CompoMat G5 preparation



– Use the main switch to turn off the CompoMat G5 and – disconnect the power cable



Opening the housing



To open the CompoMat G5, proceed as follows: – Loosen the two screws on the lower left and unscrew completely (see arrows in the following illustration).



– Loosen the two screws on the lower right and unscrew completely (see arrows in the following illustration).



– Open the cover to the rear, disconnect the protective earth cable and remove the cover.



Note: To facilitate removal of the BG-CompoFlow Opener, it is helpful to dismount the PLS scales.



9-10



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



Disassembling BGCompoFlow Opener Top – Use an Allen wrench size 2 (M67 978 1) to unscrew the stud screws (see arrows) on both sides of the assembly. – Remove assembly. .



Tip The stud screws must not be removed.



– Remove four socket screws. (See 4 arrows in both illustrations)



– Unscrew the grounding cable. – Remove the cable completely from the cable ducts. – Disconnect related connector X51 on the Application Board boughtin assembly. – Remove BG-CompoFlow Opener Top from the housing of the CompoMat G5.



Fitting BG-CompoFlow Opener Top – Place mounting tool CFO G5 (661 586 1) on the pins on the CFO Top (red part in the illustration). – Guide BG-CompoFlow Opener Top into the housing of the CompoMat G5. – Fit the grounding cable.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-11



Chapter 9: Servicing / repair



Observe the installation direction of the BG-CompoFlow Opener Top as follows: Description



Position of the long leg (CompoMat G5 viewed from the front)



BG-CompoFlow Opener Top



Left



– Secure the BG-CompoFlow Opener in position using the four pan head screws M4x8. The BG-CompoFlow Opener Top is automatically correctly positioned by the mounting tool. – Remove CFO G5 mounting tool. – Re-route cables through the cable ducts. – Connect related connector X51 on the Application Board bought-in assembly. – Fit assembly and use an Allen wrench size 2 to fasten the stud screws on both sides of the assembly. Note: The BG-CompoFlow Opener Top assembly must be flush with the housing. Disassembling BGCompoFlow Opener door – Use an Allen wrench size 2 (M67 978 1) to unscrew the stud screws (see arrows) on both sides of the assembly. – Remove assembly.



.



Tip The stud screws must not be removed.



9-12



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Remove four socket screws. (See 4 arrows in both illustrations)



– Unscrew the grounding cable. – Remove the cable completely from the cable ducts. – Disconnect related connector X52 on the Application Board boughtin assembly. – Remove BG-CompoFlow Opener door from the housing of the CompoMat G5.



Fitting BG-CompoFlow Opener door – Place mounting tool CFO G5 (661 586 1) on the pins for the CFO door (red part in the illustration). – Guide BG-CompoFlow Opener door into the housing of the CompoMat G5. – Fit the grounding cable. Observe the installation direction of the BG-CompoFlow Opener door as follows: Description



Position of the long leg (CompoMat G5 viewed from the front)



BG-CompoFlow Opener Door



Right



– Secure the BG-CompoFlow Opener door in position using the four pan head screws M4x10. The BG-CompoFlow Opener door is automatically correctly positioned by the mounting tool.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-13



Chapter 9: Servicing / repair



– Remove CFO G5 mounting tool. – Re-route cables through the cable ducts. – Connect related connector X52 on the Application Board bought-in assembly. – Fit assembly and use an Allen wrench size 2 to fasten the stud screws on both sides of the assembly. Note: The BG-CompoFlow Opener assembly must be flush with the housing. Close the housing



– Slide the cover back onto the CompoMat G5 from the rear. – Connect the grounding cable. – Close the cover completely and secure it with screws.



9.6.6



Replacing the BG-CompoFlow Opener RCC Disassembling BGCompoFlow Opener RCC – Use an Allen wrench size 2 (M67 978 1) to unscrew the stud screws (see arrows) on both sides of the assembly. – Remove assembly.



.



Tip The stud screws must not be removed.



– Unscrew both mounting pins (see arrows) using a size 7 socket. – Remove yellow RCC (M68 119 1). – Unscrew four oval head screws M3x4 from the protective cover and remove protective cover.



– Use a hot air gun to heat up two screws M6x16 coated with Precote 80 on the top of the RCC platform and unscrew.



9-14



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Unsolder cable from the weighing cell (M63 526 1) and remove the BG-CompoFlow Opener RCC with platform.



Fitting BG-CompoFlow Opener RCC



– Fasten RCC platform and the spacer plate 5.0 mm to the weighing cell using two screws M6x16 (M68 324 1) coated with Precote 80. – During this process align the RCC platform and the base plate so they are parallel with each other. – Lay cable behind the motor to the microswitch and fasten using the two fastening elements H2P 029. – Lay cable away from the weighing cell as shown in the illustration below.



– During this process ensure that the cable does not protrude and then fix the cable in this position.



RCC Scale



Solder points



– Solder strain gauge cable to the RCC cable as shown in the illustration below. +VE input (green )



-VE output (white)



1 2 3 4



1 4 2



RCC Endposition



9 10



470



14 13



+



RCC Motor



1



-VE input (black ) +VE output (red)



200



30



M -



– Connect cable to the motor. – Pull a piece of heat shrink tubing over the cable and ensure the cable is not touching the motor housing.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-15



Chapter 9: Servicing / repair



– Apply Loctite 243 to four oval head screws M3x4. – Attach protective cover to the platform and the bearing block using the screws.



– Screw both mounting pins into the RCC platform using a torque wrench (M683151+M683171+M683181) and a torque of 0.8 Nm. – Fit assembly and use an Allen wrench size 2 to fasten the stud screws on both sides of the assembly.



Note: The BG-CompoFlow Opener RCC assembly must be flush with the housing.



9.6.7



Testing the BG-CompoFlow Opener The following tests must be carried out: – Check both legs for easy movement by pressing them together. – Both legs must move easily. – The legs must open again automatically.



2)



– None of the two legs may have any axial clearance. – When the sealer is open, the pin of leg 2 must reliably trigger the microswitch. 1) Pin, leg 2 and microswitch 2) Axial clearance 1)



9-16



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



The position of the eccentric must be correct: – Fit the assembly tool CFO Top/door (661 591 1). The assembly tool must sit firmly on the drive holder CFO (Motor board RCC) and the eccentric firmly on the assembly tool. – Apply Loctite 243 to the threaded pin M4x6. – Press down the eccentric and the drive shaft for the DC motor – Tighten the threaded pin to 1.8 Nm and in this way fasten the eccentric. – Remove assembly tool CFO Top/door. 1) 3)



1) Eccentric 2) Assembly tool CFO Top/door 3) Threaded pin



2)



Adjusting the BGCompoFlow Opener If the sealer is stuck or jammed, an adjustment is required: – Loosen the stud screw. 1)



– Adjust the clearance by moving the shaft of the BG-CompoFlow Opener using parallel pliers or a similar tool. – Apply Loctite 243 to the stud screw and then re-tighten the screw. Both legs must move easily, but neither leg may have any axial clearance.



2)



1) Stud screw 2) CFO shaft



Adjusting microswitch If the microswitch is not triggered reliably, it can be adjusted as follows:



1)



– Loosen the pan head screws (1). – Position the FCO switch bracket (2) so that the microswitch is triggered reliably.



2)



1) Pan head screws



1)



Fresenius Kabi



CompoMat G5



2) Switch bracket



SM-EN



3/11.10



9-17



Chapter 9: Servicing / repair



Calibration and tests



After the replacement of a BG-CompoFlow Opener the following tests must be performed: – Calibrate RCC scales (see chapter 5.4.4.2 on page 5-23) – Calibrate PLS scales (see chapter 5.4.4.3 on page 5-24) – Press force for the upper press (see Re: 3.20 on page 6-22) – Test press scales / PLS scales / RCC scales (see Re: 3.20 on page 6-22) – BG-CompoFlow Opener height measurement (see chapter 9.7.1 on page 9-34) – BG-CompoFlow Opener Installation (see chapter 9.7.2 on page 9-35) – BG-CompoFlow Opener opening (see chapter 9.7.3 on page 9-35) – BG-CompoFlow Opener height assembly (see chapter 9.7.4 on page 9-35) – Expanded test BG-CompoFlow Opener for the replaced BGCompoFlow Opener (see Re: 3.30 on page 6-35) – Perform the electrical safety check (see chapter 6.4.2 on page 6-36).



9.6.8



Replacing the Heads CompoMat G5 preparation



The following description applies to Heads 1-6, but does not include Head 3 which is located on the PLS scales (see chapter 9.6.9 on page 9-22). Disconnect the CompoMat G5 from the power supply. – Use the main switch to turn the device off and – disconnect the power cable



Opening the housing



To open the CompoMat G5, proceed as follows: Tip To open the housing, the PLS scales and the RCC scales must not be removed! – Loosen the two screws on the lower left and unscrew completely (see arrows in the following illustration).



9-18



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Loosen the two screws on the lower right and unscrew completely (see arrows in the following illustration).



– Pull the cover to the rear to remove. – Disconnect the protective earth cable. – Remove the cover.



Dismantling the Head



All Heads have the same design. The following description refers to Head 5 and can be applied to all other Heads.



2



1



– Loosen grooved nut. Tip If the grooved nut (1) is very tight, use the service tool for manual screwing (2) to loosen it (see illustration).



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-19



Chapter 9: Servicing / repair



1



– Remove the head cover (1) (see illustration).



3



2



1



Remove the following plug connectors from Head 5: – BNC connector, RF (1) (already removed in the illustration) – Four-pole connector, (sensor 2) (2). – Two-pole connector, (magnet 2) (3).



– Lift and remove Head 5 (see illustration). Installing the Head



9-20



The installation is described on the example of Head 5 and is identical for all heads.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Assemble the individual parts of the head as shown in the illustrations.



– Insert Head 5 into the housing from above. Note: Observe the following items when installing: – The Head position is determined by the design. – Make sure that the head opening is correctly aligned. Caution Damage can occur to cables. Crushing hazard – Make sure not to damage the cables on the side of the Head when inserting and positioning the head.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-21



Chapter 9: Servicing / repair



Insert the following plug connectors into Head 5: – BNC connector (RF) – Four-pole connector (sensor 2) – Two-pole connector (magnet 2)



– Position Head 5 cover and manually tighten the grooved nut (see illustration). Testing Heads



9.6.9



After the replacement of a Head the following tests must be performed: – Calibrate Head sensors (see chapter 5.2.3.5 on page 5-6)



Replacing Head 3 (PLS scales) CompoMat G5 preparation



– Use the main switch to turn the CompoMat G5 off. – Disconnect the power cable.



Disassembling the PLS scales



Head 3 is part of the PLS scales. The PLS scales must be dismounted before dismounting Head 3: – Tilt the CompoMat G5 to the rear and place it onto its rear panel (see chapter 3.3 on page 3-3). – Loosen and remove PLS scalesplug (Back board: Balance B). – Remove the connection cable from the cable clips. – Disconnect the BNC connector for Head 3. – Loosen and unscrew the three screws on the PLS scales fixing bracket. –



9-22



Completely remove the PLS scales.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



Dismantling Head 3



To replace Head 3, proceed as follows:



1



– Loosen both screws on the upper device housing (1) of the PLS scales Note: Both screws are located on the rear of the PLS scales and are only accessible once the PLS scales have been unmounted. – Disconnect the BNC connector (RF). – Disconnect the four-pole connector (sensor 2). – Disconnect the two-pole connector (magnet 2). – Loosen the grooved nut on Head 3. Tip If the grooved nut is very tight, use the tool for manual screwing to loosen it (see illustration). – Unscrew and remove the head cover. – Remove Head 3. Installing Head 3



(see Installing the Head on page 9-20) – Mount the upper scales housing and secure it with two screws.



Fitting PLS scales



Procedure for mounting the PLS scales(see chapter 3.3 on page 3-3).



Connect the PLS scales



Procedure for connecting the PLS scales (see chapter 3.4 on page 3-8). Note: The PLS scales must be re-calibrated after mounting. (see chapter 5.4.4.3 on page 5-24) After calibrating the scales, the scales must be tested (see Re: 3.20 on page 6-22).



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-23



Chapter 9: Servicing / repair



Checking Head 3 on the PLS scales. Note: The integrated Head 3 on the PLS scales must be inspected for functionality after assembly of the PLS scales. (see Re: 3.24 on page 6-26) (see Re: 3.25 on page 6-28) (see Re: 3.26 on page 6-29) (see Re: 3.27 on page 6-31)



9.6.10 Replacing the weighing cell RCC (M63 526 1) Disassembling weighing cell RCC – Loosen RCC scales at connection B. – Use an Allen wrench size 2 (M67 978 1) to unscrew the stud screws (see arrows) on both sides of the assembly. – Remove assembly. .



Tip The stud screws must not be removed.



– Unscrew both mounting pins (see arrows) using a size 7 socket. – Remove yellow RCC (M68 119 1). – Unscrew four oval head screws M3x4 from the protective cover and remove protective cover.



– Use a hot air gun to heat up two screws M6x16 secured using Precote 80 on the top of the RCC platform and unscrew.



9-24



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Unsolder cable from the weighing cell (M63 526 1) and remove the BG-CompoFlow Opener RCC with platform. – Use a hot air gun to heat up two screws M6x16 coated with Precote 80 screwed into the weighing cell from below and unscrew. – Similarly, unscrew the two countersink head screws M6x16 with which the weighing cell (M63 526 1) is fastened to the RCC base plate. – Remove spacer plate 0.6 mm. – Unsolder cable for the weighing cell and remove the weighing cell. Fitting weighing cell RCC



– Apply Loctite 243 to four countersink head screws M6x16. – Fasten weighing cell (M63 526) and the spacer plate 0.6 mm on the cable side of the weighing side and fasten to the RCC base plate using the four countersink head screws. – Prior to finally tightening the screws from below, screw the weighing cell assembly tool (661 534 1) into the threaded bores on the weighing cell and align it. – After tightening the countersink head screws, remove the assembly tool again. – Fasten RCC platform and the spacer plate 5.0 mm to the weighing cell using two screws M6x16 (M68 324 1) coated with Precote 80.



– During this process align the RCC platform and the base plate so they are parallel with each other. – Screw two screws M6x16 (M68 324 1) coated with Precote 80 screwed into the weighing cell from below.



– Adjust distance between base plate and countersink head screw to 0.6 mm.



Warning There must not be any dust or other foreign bodies between the weighing cell and the base plate.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-25



Chapter 9: Servicing / repair



– Align four spacer bolts until there is a distance of 0.7 mm to the RCC platform. For this purpose place the RCC scales upright and secure to a table using a screw clamp.



– Tighten four nuts M6 and ensure that the distance is now 0.6 mm.



RCC Scale



Solder points



– Solder RCC cable to the weighing cell for the RCC scales as shown in the illustration below. +VE input (green )



-VE output (white)



1 2 3 4



1 4



RCC Endposition



2



9 10



470



14 13



30



200



+



RCC Motor



1



-VE input (black ) +VE output (red)



M -



– Protect cable with heat shrink tubing. – Lay cable away from the weighing cell as shown in the illustration below.



– During this process ensure that the cable does not protrude and fix the cable in this position. – Apply Loctite 243 to four oval head screws M3x4. – Attach protective cover to the platform and the bearing block using the screws.



9-26



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Screw both mounting pins into the RCC platform using a torque wrench (M683151+M683171+M683181) and a torque of 0.8 Nm. – Fit assembly and use an Allen wrench size 2 to fasten the stud screws on both sides of the assembly.



Note: The BG-CompoFlow Opener RCC assembly must be flush with the housing.



9.6.11 Testing the weighing cell RCC After the replacement of the weighing cell RCC the following tests must be performed: – Calibrate RCC scales (see chapter 5.4.4.2 on page 5-23) – Calibrate PLS scales (see chapter 5.4.4.3 on page 5-24) – Test press scales / PLS scales / RCC scales (see Re: 3.20 on page 6-22) – BG-CompoFlow Opener height measurement (see chapter 9.7.1 on page 9-34) – BG-CompoFlow Opener Installation (see chapter 9.7.2 on page 9-35) – BG-CompoFlow Opener height assembly (see chapter 9.7.4 on page 9-35) – Perform the electrical safety check (see chapter 6.4.2 on page 6-36).



9.6.12 Replacing the weighing cell PLS (M63 526 1) CompoMat G5 preparation



– Use the main switch to turn off the CompoMat G5. – disconnect the power cable



Disassembling weighing cell PLS



Fresenius Kabi



CompoMat G5



SM-EN



– Loosen BNC connection and PLS connector. – Unscrew three screws M5x16 from the PLS scales (1).



3/11.10



9-27



Chapter 9: Servicing / repair



1



– Unscrew two screws M4x10 and replace lower cover. – Use a hot air gun to heat up two screws M6x16 coated with Precote 80 screwed into the weighing cell from below and unscrew. – Similarly, use a hot air gun to heat the two screws M6x16 secured with Precote 80 that fasten the platform to the spacer plates 5.0 mm and unscrew. – Similarly, using a hot air gun heat the two countersink head screws M6x16 with which the weighing cell (M63 526 1) is fastened to the PLS base plate and unscrew. – Remove spacer plate 0.6 mm. – Unsolder cable for the weighing cell and remove the weighing cell.



Fitting weighing cell PLS



– Apply Loctite 243 to two countersink head screws M6x16. – Fasten weighing cell (M63 526) and the spacer plate 0.6 mm on the cable side of the weighing side and fasten to the base plate using the two countersink head screws. – Prior to finally tightening the screws from below, screw the weighing cell assembly tool (661 534 1) into the threaded bores on the weighing cell and align it.



9-28



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– After tightening the countersink head screws, remove the assembly tool again. – Fasten platform and the spacer plate 5.0 mm to the weighing cell using two screws M6x16 (M68 324 1) coated with Precote 80. – During this process align the platform and the base plate so they are parallel with each other. – Screw two screws M6x16 (M68 324 1) coated with Precote 80 screwed into the weighing cell from below.



– Adjust distance between base plate and countersink head screw to 0.6 mm.



Warning There must not be any dust or other foreign bodies between the weighing cell and the base plate. – Loosen screw for the cable holder on the weighing cell (M63 526 1) and lay the cable as shown in the illustration on the left. – Lay cable for the weighing cell through the hole in the base plate. – Solder PLS cable to the weighing cell for the PLS scales as shown in the illustration below.



Plasma load cell +VE input (green )



-VE output (white)



Plasma Sensor



1



11



2



12



1



5 6 7



1



Plasma Magnet



8



1 2 3 4



-VE input (black ) +VE output (red)



Warning The cable must not touch the lower cover under any circumstances! – Check whether the distance between spacer bolt and platform is 0.6 mm.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-29



Chapter 9: Servicing / repair



Warning There must not be any dust or other foreign bodies between the weighing cell and the base plate. – If not: Align four spacer bolts until there is a distance of 0.7 mm to the platform. For this purpose place the PLS scales upright and secure to a table using a screw clamp. – Tighten four nuts M6 and ensure that the distance is now 0.6 mm. – Fasten lower cover with two screws M4x10 using a torque wrench and a torque of 0.4 Nm. – Finally re-fit the PLS scales (see chapter 3.3.1 on page 3-4).



9.6.13 Testing the weighing cell PLS After the replacement of the weighing cell PLS the following tests must be performed: – Calibrate Head sensors (see chapter 5.2.3.5 on page 5-6) – Calibrate RCC scales (see chapter 5.4.4.2 on page 5-23) – Calibrate PLS scales (see chapter 5.4.4.3 on page 5-24) – Test press scales / PLS scales / RCC scales (see Re: 3.20 on page 6-22) – Open/close the Heads (see Re: 3.24 on page 6-26) – Test the Heads (see Re: 3.25 on page 6-28) – Head sensors (see Re: 3.27 on page 6-31) – Sealing behavior of the Heads (see Re: 3.26 on page 6-29) – Perform the electrical safety check (see chapter 6.4.2 on page 6-36).



9.6.14 Replacing the scales 3 (M68 109 1) CompoMat G5 preparation



– Use the main switch to turn off the CompoMat G5 and – disconnect the power cable



Opening the housing



9-30



To open the CompoMat G5, proceed as follows:



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– Loosen the two screws on the lower left and unscrew completely (see arrows in the following illustration).



– Loosen the two screws on the lower right and unscrew completely (see arrows in the following illustration).



– Open the cover to the rear, disconnect the protective earth cable and remove the cover.



Disassembling scales 3



– Loosen connector X56 on the Application Board bought-in assembly. – Remove the cable completely from the cable ducts. – Loosen screw M4x10 with which the grounding cable is fastened to the front press guide. – Ensure that all washers and locking washers are removed. – Unscrew four screw M4x20 and remove the silicon insert top press G5 (M68 117 1). – Unscrew two screw M4x20 and carefully remove the sensor unit top press G5 (M68 113 1). – Unscrew two screws M4x30.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-31



Chapter 9: Servicing / repair



– Carefully move holder back and forward using pliers (see arrows).



Caution During the next work step do not grasp the scales at the weighing cell under any circumstances. – Carefully pull out scales 3 and remove cable. Fitting scales 3



– Fit scales 3 (M68 109 1) in the block for the upper press and fasten using two screws M4x30. – Tighten the two screws alternately until they are tight.



– Fasten sensor unit upper press G5 (M68 113 1) with two screws M4x20 using a torque wrench and a torque of 0.8 Nm.



– Fit silicon insert upper press G5 (M68 117 1) and fasten with two screws M4x20 using a torque wrench and a torque of 0.2 Nm. The plastic may break at a higher torque. – Fasten grounding cable for the front press guide (M67 469 1) to the scales 3 for the upper press using a screw M4x10.



9-32



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



– During this process place a locking washer and spacer under the cable lug as well as a further spacer under the Allen screw.



Allen screw Spacer Cable lug Spacer Locking washer



– Fasten cable using a few cable ties as shown in the illustration below.



– Lay new cable through the cable ducts like the old one and connect to connector X56 on Application Board bought-in assembly. Close the housing



– Slide the cover back onto the CompoMat G5 from the rear. – Connect the grounding cable. – Close the cover completely and secure it with screws.



9.6.15 Testing scales 3 After the replacement of the scales 3 the following tests must be performed: – Calibrate press stroke for upper press (see chapter 5.4.2.2 on page 5-19) – Calibrate press force for upper press (see chapter 5.4.2.1 on page 5-17) – Calibrate press scales (see chapter 5.4.4.1 on page 5-22) – Calibrate RCC scales (see chapter 5.4.4.2 on page 5-23) – Calibrate PLS scales (see chapter 5.4.4.3 on page 5-24) – Calibrate detector A (see chapter 5.4.5 on page 5-25) – Upper press force (see Re: 3.18 on page 6-21) – Test press scales / PLS scales / RCC scales (see Re: 3.20 on page 6-22)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-33



Chapter 9: Servicing / repair



– Multi-detector field (detectors A1 to A8) (see Re: 3.23 on page 6-26) – Perform the electrical safety check (see chapter 6.4.2 on page 6-36).



9.7



Tests after the replacement of components



9.7.1



BG-CompoFlow Opener height measurement Accessories: – Test tool height CF-Opener G5 (661 570 1) – Allen key size 2 (661 526 1) Preparation: – Fully switch off CompoMat G5. – Remove assembly from the CF-Opener (M67 987 1). – Switch on the dial gauge (part of the test tool height BG-CompoFlow Opener G5). – Place the dial gauge on the adjusting piece and press the ORIGIN button. – Place the dial gauge on each CF-Opener in turn and make a note of the measured value. Measured values: CF-Opener



Meas. value



Desired value Top: 0 mm ± 0.7 mm



______ mm



Desired value door: 0 mm ± 0.7 mm



______ mm



Desired value RCC: 0 mm ± 0.7 mm



______ mm



Ensure that the assembly for the dial gauge is firmly seated on the housing during all measurements. Test result: CF-Opener



Test result



All CF-Opener within tolerance



9-34



Fresenius Kabi



Yes No



CompoMat G5



SM-EN



3/11.10



Chapter 9: Servicing / repair



9.7.2



BG-CompoFlow Opener installation Accessories: – Mounting tool CFO G5 (661 586 1) – Allen key size 2 (M67 978 1) Preparation: – Fully switch off CompoMat G5. – Fit mounting tool CFO G5 to each CF-Opener. Test result: Mounting tool can be simply inserted.



9.7.3



Test result



CF-Opener Top



Yes No



CF-Opener Door



Yes No



CF-Opener RCC



Yes No



BG-CompoFlow Opener opening This test is also performed during the annual maintenance (see Re: 3.22 on page 6-24). – After the replacement of a BG-CompoFlow Opener the test must, however, be undertaken using the test tool production 3.3 mm (661 580 1) and not using the test tool service 3.6 mm (661 578 1). – Also only the BG-CompoFlow Opener that has been replaced needs to be tested. Note: During this service work all actual limits for the CF-Opener are exceeded. For this reason, on the completion of the service work the CompoMat G5 must be rebooted. Without a reboot the correct function of the CF-Opener cannot be ensured.



9.7.4



BG-CompoFlow Opener height assembly Accessories: – Test tool height assembly CF-Opener G5 (661 576 1) – Digital vernier caliper, round depth gauge (M67 963 1) – Allen key size 2 (M67 978 1) Preparation: – Fully switch off CompoMat G5. – Fit assembly (M67 987 1) to the CF-Opener and fasten. – Place test tool to each assembly.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



9-35



Chapter 9: Servicing / repair



– Using the vernier caliper measure the distance between the surface of the test tool and the bottom of the CF lever through the 3 mm opening.



– Make a note of the absolute values found during the height measurement (see BG-CompoFlow Opener height measurement on page 9-34). – Determine the desired value as well as the related tolerance and compare with the measured value. CF-Opener Top desired value 35.1 mm ± (magnitude of the measured value CFOpener Top + 0.3 mm)



Meas. value



35.1 mm ± ______ mm



______ mm



CF-Opener door desired value 35.1 mm ± (magnitude of the measured value CFOpener door + 0.3 mm)



Meas. value



35.1 mm ± ______ mm



______ mm



CF-Opener RCC desired value 29.7 mm ± (magnitude of the measured value CFOpener RCC + 0.3 mm)



Meas. value



29.7 mm ± ______ mm



______ mm



Example CF-Opener Top: The measured value for the height measurement was -0.52 mm. It follows that: Desired value: 35.1 mm ± (0.52 mm + 0.3 mm) = 35.1 mm ± 0.82 mm Measured values between 34.28 mm and 35.92 mm are in order. Test result: CF-Opener



Test result



All CF-Opener within tolerance



9-36



Fresenius Kabi



Yes No



CompoMat G5



SM-EN



3/11.10



Chapter 10: Passwords



10 Passwords 10.0.0.1 Service password The service password works only in Service Mode with an inserted Service G5 USB stick. It cannot be changed by the operator.



Passwor d



Function



Description



1462



Service



Enables the Service mode



10.0.0.2 Default settings for passwords in CompoMaster Net G5 The passwords only apply for the User mode. By default the following passwords are assigned. Passwor d



Function



1000



Program selection



2000



Calibration



3000



Cleaning



4000



Setup



Tip For further information see: User Manual, CompoMaster Net G5



10.0.0.3 Defining individual passwords in CompoMaster Net G5



Tip For further information see: User Manual, CompoMaster Net G5



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



10-1



Chapter 11: Functional Description



11 Functional Description 11.1 Description of the therapy Note: For information regarding this chapter see: Operating Instructions, CompoMat G5



11.2 Physical Description Note: For information regarding this chapter see: Operating Instructions, CompoMat G5



11.3 Functional description 11.3.1 Application Board bought-in assembly (M68 137 1) The functional component group Application Board bought-in assembly is responsible for the entire control of the actuators as well as for monitoring sensors. The Application Board bought-in assembly controls communication with the functional component group Trinamic Board bought-in assembly. Communication with the functional component group Display G5 printed circuit board bought-in assembly is carried out via a USB connection.



11.3.2 Display G5 printed circuit board bought-in assembly (M68 115 1) The functional component group Display G5 printed circuit board bought-in assembly controls the external interfaces to superior systems (such as BIS - Blood Bank Information System) as well as the user guidance and communication with the internal sub-systems.



11-1



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 11: Functional Description



11.3.3 HF generator 24V bought-in assembly (M66 178 1) The functional components group HF generator 24V G5 is responsible for the generation of HF (high frequencies). It is controlled via the functional component group Application Board bought-in assembly.



11.3.4 Relay box G5 (M68 107 1) The functional components group relay box G5 is responsible for the channeling of HF (high frequencies). It is controlled via the functional component group Application Board bought-in assembly.



11.3.5 Sealing Head El.Mag. Drive assembly (M68 144 1) The functional group Sealing Head El.Mag. Drive assembly is responsible for clamping and sealing the tubes and for the detection of red cells. Control and evaluation of the data is done by the functional component group Application Board bought-in assembly.



11.3.6 Trinamic Board bought-in assembly (M68 138 1) The functional component group Trinamic Board bought-in assembly controls the stepper motors and measures the power required by them. It is controlled via the functional component group Application Board bought-in assembly.



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



11-2



Fresenius Kabi



6



M



CompoMat G5



M



door lock



5



OUT IN



4



B/ASM HFgenerator 24V (M66178) B/ASM relay box (M66407) OUT6



OUT5



OUT4



OUT3



OUT2



OUT1



3



author: M.Hock



date: 21.04.2009 scale



M



X48



X48



X48



X48



X48



slider



K6



K5



K4



K2



K1



2.* ASMCompoFlow Opener (M66810)



1* B/ASM-Board detector (M60841)



1



2



1:1



V1.4



version



sheet



drawing no.



block diagram G5



1



1 of 1



Fresenius Kabi Deutschland GmbH



2*



RF



door sensor



Boot USB



S M



X9 3



A



CON304



Mixer



D



RF



Ethernetconnection



CON302



M



lower press



1*



upper press



breaker1



2*



breaker0



X9 3



CON306



USB



X55 X46 X47 X30 X31



Mlp



S M D lp



S M D up



Mup



top press



RF



door motor



CON303



CON301



X10



X60 X57



DMS 1



X81



X50



B/ASMTrinamic board (M66960) Communication



S M D tp



Mtp



X9 3



door



GND



X2



X30 0



B/ASM-Application board (M66639)



GND/24V



DMS 2 Motor 0 DMS 0 Motor 2



X80



X56



X51



X52



RF



K/ASM display G5 board (M66826) CON LED



CON300



(M66969)



+24V



power supply



X30 1



X1



X53



X200



X1000



X11



X9 3



display G5 front foil (M66829)



display G5 (M66824)



N



G



L



Reserve out



Jtag



USB Hub



X43



1 x USB intern



2



RF



Ethernet 4000V Isolation



ac



RF X93



K3



X48



X21



X42



3



X9 3



B



C



ASM-Plasma scale (M66846)



X20



X54



X41



X12



2 x USB



X13



X44



scanner



X90.6 X90.5 X90.4 X90.3 X90.2



RFID



breaker2



X40



D



CF opener



4



X45



5



X90.1



ASM-RCC-Scale (M66860)



0



REV.



ASM-sealing head cpl (M67174)



11-3 Motor 1



6



A



B



C



D



Chapter 11: Functional Description



11.4 Block diagram



SM-EN



3/11.10



Chapter 11: Functional Description



11.5 Component layouts 11.5.1 Application Board bought-in assembly



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



11-4



Chapter 11: Functional Description



11.5.2 Display G5 printed circuit board bought-in assembly



11.5.3 LP1361



11-5



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 11: Functional Description



11.5.4 Trinamic Board bought-in assembly



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



11-6



Chapter 11: Functional Description



11.6 Flow diagram (Service Menu) Entering the password



Main menu (Service) Set Serial number Program selection



Set E-Code Set MAC-Adress



01



Update/Init ARM7



02



Update/Init Trinamic Cal. ADC & Heads Special functions



50



Special functions Appl. Service mode Download files



Main menu



Unittest screenshot Unittest Network stress Create Production-Stick Production state



Program selection Seal & weighing Calibrate Cleanup Setup Cleanup



Calibrate



System information Upper Press



Top Press



Lower Press



Upper Press



Slide



Lower Press



Setup Press Scale Language selection



RCC scales



Process data Software update



PLS Scale Process data



Language selection



Process data off



English



Process data on



German



Detector A



Barcode test



Italian ...



11-7



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



12 Spare parts catalog



Pos.



Part no.



Description



1



M68 109 1



Scale 3 f. Service



2



M68 110 1



Suspending pin f. Service (2x)



Pos.



Part no.



Description



1



M68 113 1



Sensor unit upper press G5 f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-1



Chapter 12: Spare parts catalog



12-2



Pos.



Part no.



Description



2



M68 105 1



UP.DETECTOR COVER DETECTION f. Service



3



M68 149 1



DETECT.BOARD A8-A2 COMPOMAT f. Service



Pos.



Part no.



Description



1



M67 991 1



textile EMI gask f. Service (Set)



2



M68 114 1



Display G5 f. Service



3



M68 150 1



ferrite core filter display f. Service 10x



4



M68 115 1



Display G5 board ARM9 f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Part no.



Description



1



M63 526 1



SCALES CELL COMPOGUARD



2



M67 986 1



Rear cover PLS + clamp collar f. Service



3



M68 123 1



Upper cover G5 f. Service



4



M68 122 1



Lower cover f. Service



5



M68 121 1



Terminal strip f. Service



6



M68 144 1



Sealing head cpl. G5 f. Service



7



M68 120 1



Strip sheet coated f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-3



Chapter 12: Spare parts catalog



12-4



Pos.



Part no.



Description



1



M68 138 1



Trinamic board f. Service



2



M68 137 1



Application board ARM7 LP1349 f. Service



Pos.



Part no.



Description



1



M66 178 1



EE/ HF-generator 24V G5 adjusted + tested



2



M68 107 1



Relay box G5 f. Service



3



M68 139 1



Powerbox Medline 450 f. Service



4



M67 997 1



ethernet isolator f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Part no.



Description



5



M67 988 1



BNC panel jack 10HC 049 f. Service (5x)



6



M67 989 1



D-SUB-mounting set UNC 4-40 f. Serv. 4x



7



M67 996 1



components clip V-1001 f. Serv. (Set 10)



8



M68 103 1



COVERING CAP F. CIRCUIT f. Service (10x)



9



M68 142 1



Glider PGB 40x20x8 f. Service (8x)



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-5



Chapter 12: Spare parts catalog



12-6



Pos.



Part no.



Description



1



M68 125 1



Step motor f. press with spindle f. Serv.



2



M68 124 1



double guide bearing + flange f. Serv. 4x



3



M68 146 1



Proximity sensor G5 f. Service



4



M68 106 1



SHOCK ABSORBER f. Service



5



M68 127 1



Guide bearing slide f. Service 2x



6



M68 126 1



Slide drive cpl. f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Part no.



Description



1



M67 994 1



wiper 16_22_04 f. Service (2x)



2



M68 125 1



Step motor f. press with spindle f. Serv.



3



M68 128 1



Pressing plate G5 incl. label f. Service



4



M68 131 1



Double guide sleeve top press (Serv. 2x)



5



M68 130 1



Force sensor top press f. Service



6



M68 129 1



Silicone mat toppress G5 f. Service



7



M67 992 1



Disk wiper f. Service (2x)



8



M68 104 1



CountersinkScr.M6X16+precote80 f. S. 20x



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-7



Chapter 12: Spare parts catalog



12-8



Pos.



Part no.



Description



1



M67 987 1



Fixture active breaker G5 f. Service



2



M68 119 1



covering RCC f. Service



3



M63 526 1



SCALES CELL COMPOGUARD



4



M68 132 1



CompoFlow Opener G5 RCC f. Service



Pos.



Part no.



Description



M68 145 1



Interface Detector f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Pos.



Fresenius Kabi



Part no.



Description



M67 987 1



Fixture active breaker G5 f. Service



Part no.



Description



M68 101 1



passive insert activebreaker G5 f. Serv.



CompoMat G5



SM-EN



3/11.10



12-9



Chapter 12: Spare parts catalog



Pos.



Pos.



12-10



Part no.



Description



M68 108 1



Force sensor upper+down press f. Service



Part no.



Description



M68 111 1



CompoFlow Opener TOP-Pr. f. Service



M68 112 1



COMPOFLOW Opener Upper-Pr. f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Pos.



Fresenius Kabi



Part no.



Description



M68 117 1



Silicone inlay upper press G5 f. Service



Part no.



Description



M68 118 1



Silicone inlay lower press G5 f. Service



CompoMat G5



SM-EN



3/11.10



12-11



Chapter 12: Spare parts catalog



12-12



Pos.



Part no.



Description



1



M68 144 1



Sealing head cpl. G5 f. Service



2



M68 298 1



top electrode f. sealing head G5 f. Serv



Pos.



Part no.



Description



1



M68 134 1



assyembly door G5 f. Service



2



M68 135 1



Door G5 plate f. Service



3



M67 993 1



plug_B-Plastic_SFL 18x0_8-2 f. Service 5x



4



M68 136 1



Silicone inlay for door G5 f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Pos.



Fresenius Kabi



Part no.



Description



661 524 1



B/ASM-packaging RCC scales



Part no.



Description



661 525 1



B/ASM-packaging plasma-Scale



CompoMat G5



SM-EN



3/11.10



12-13



Chapter 12: Spare parts catalog



Pos.



Pos.



12-14



Part no.



Description



M67 216 1



B/ASM-packaging CompoMat G5



Part no.



Description



M68 126 1



Slider drive cpl. f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Pos.



Pos.



Fresenius Kabi



Part no.



Description



M68 133 1



Door drive cpl. f. Service



Part no.



Description



M68 141 1



Lock G5 f. Service



CompoMat G5



SM-EN



3/11.10



12-15



Chapter 12: Spare parts catalog



12-16



Pos.



Part no.



Description



1



M67 990 1



support bypass pin



2



M67 999 1



profile seal f. Service



3



M68 116 1



display front foil G5 f. Service



Pos.



Part no.



Description



M68 143 1



Rear panel sheet G5 f. Service



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



1



2



3



Pos.



Part no.



Description



1



M68 299 1



IEC power connector G5 f. Service



2



M68 301 1



ON/OFF SWITCH 10A 250V f. Service



3



M68 302 1



COVERING CAP F. CIRCUIT BRAKER f. Service



1



2



3



Pos.



Part no.



Description



1



M66 514 1



cable RF sealhead K1/5/6



2



M67 446 1



cable RF sealhead K2/4



3



M67 447 1



cable RF switchbox - RF-generator



1



2



3



Pos.



Part no.



Description



1



M67 449 1



cable RF sealhead K3



2



M67 301 1



cable power supply ARM7-board X1



3



M67 306 1



cable power supply ARM9-board X2



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-17



Chapter 12: Spare parts catalog



1



Pos.



Part no.



Description



1



M67 310 1



cable RFID ARM7-board - Connector X21



2



M67 314 1



cable power supply RF-generator X30



3



M67 322 1



cable relaybox - ARM7-board X31



1



2



3



Pos.



Part no.



Description



1



M67 330 1



cable PLS ARM7-board - Connector X53



2



M67 342 1



cable door motor X55



3



M67 346 1



cable slider motor X57



1



12-18



3



2



2



3



Pos.



Part no.



Description



1



M67 354 1



cable door lock X47



2



M67 357 1



Cable CF-Opener Toppress X51



3



M67 363 1



Cable CF-Opener Upperpress X52



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



1



2



3



Pos.



Part no.



Description



1



M67 368 1



cable power supply Trinamic Board X80



2



M67 372 1



cable sealhead K1/2/4 X90.1/90.2/90.4



2



M67 383 1



cable sealhead K5 X90.5



2



M67 386 1



cable sealhead K6 X90.6



3



M67 392 1



cable sealhead sensor K1 X40



1



2



3



Pos.



Part no.



Description



1



M67 395 1



cable sealhead sensor K2 X41



2



M67 397 1



cable sealhead sensor K4 X43



3



M67 399 1



cable sealhead sensor K5 X44



1



2



3



Pos.



Part no.



Description



1



M67 402 1



cable sealhead sensor K6 X45



2



M67 404 1



cable communication X81Trinamic



3



M67 412 1



cable powerline powerinlet



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-19



Chapter 12: Spare parts catalog



1



2



3



Pos.



Part no.



Description



1



M67 434 1



cable powerline powerinlet blue



2



M67 463 1



cable sealhead K3 crossing cable



3



M67 464 1



cable photodetection board X60



Other spare parts not illustrated Pos.



12-20



Part no.



Description



644 066 1



GV FUSE 10 A/T (D5 X 20 MM) SHORT



M68 324 1



Hex. socket scr.M6x16+Precote80 f. S. 10x



M67 995 1



Oval head screw ISO7380 M5x8 f. Serv. 20x



M67 998 1



warning label CompoFlow f. S. (54x)



M68 102 1



Hex. socket scr.M5x16+Precote80 f. S. 30x



M68 176 1



sticker set sealing head G4/G5 1-8 A B



M68 140 1



IEC power conn. PF0033/20/63 f. Service



M67 432 1



cable USB communication



M67 433 1



cable Ethernet communication



M67 438 1



Power cord EU except UK



M67 448 1



cable RF PLS-scales



M67 520 1



FUSE 8 A/T (D5 X 20 MM)



M67 416 1



cable grounding power line



M67 420 1



cable grounding power supply



M67 424 1



cable powerline powersupply black



M67 428 1



cable grounding back panel



M67 439 1



cable powerline powersupply blue



M67 469 1



cable grounding Front press guidance



M67 471 1



cable grounding Toppress



M67 494 1



cable grounding Scale CF-Toppress



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Other spare parts not illustrated Pos.



Part no.



Description



M67 495 1



cable grounding CF Upperpress



M68 147 1



RCC-cable f. Service



M68 148 1



PLS-cable f. Service



12.1 Tools and accessories



1



2



3



4



5



6



7



8



9



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-21



Chapter 12: Spare parts catalog



10



13



11



14



12



15



16



17



18



19



20



21



12-22



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



22



23



24



25



26



27



28



29



30



31



32



33



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-23



Chapter 12: Spare parts catalog



34



35



36



37



38



39



40



41



42



43



44



45



46



12-24



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



Chapter 12: Spare parts catalog



Tools and accessories Pos.



Part no.



Description



1



M68 287 1



ASM-electrical test equipment for G5



2



M68 288 1



EE motor switch box G5



3



661 545 1



black foam mat G5



4



M67 228 1



Service tool power measurement G5



5



M63 925 1



CALIBRATION WEIGHT 500G F1 + CERTIFICATE



6



M67 851 1



EE/SYNTHETIC BLOCK G4+G5 calibrated



7



M63 924 1



TWIN-BAG SYSTEM COMPOMAT G4



8



M67 501 1



CLOSURE DEVICE IN HOUSING 4.2 STERILIZED



9



661 537 1



flare double-ended ring spanner 17x19



10



679 514 1



OPTICAL FILTER 13 LEVELS



11



M67 980 1



hexagon socket screwkey SW4 (Service)



12



M60 024 1



PVC-TUBE 3.1X4.2 MM COMPOSEAL



13



M67 499 1



Mobile force gauge EasyForce 2kN



14



M67 502 1



service tool manual screwing G5



15



M67 981 1



Tool for manual screwing f. Service



16



M67 979 1



double-end. ring span. deep offset 12x13



17



M67 978 1



hexagon socket screw key SW2 f. Service



18



M67 507 1



ferrit opener



19



661 534 1



Assembly tool load cell



20



M67 482 1



USB Stick Service CompoMat G5



21



661 570 1



testing tool height CompoFlowOPener G5



22



661 576 1



testing tool height fixture G5



23



661 586 1



Assembly-Tool CFO G5



24



M67 963 1



dig. calliper gauge, round depth gauge



25



661 535 1



talcum Pressol 10 588 50gr.



26



M64 006 1



glue - SICOMET 77 20G bottle



27



661 575 1



Torque Wrench 4-20 Nm



28



M67 977 1



Silicone ELASTOSIL E43 90ml Tube



29



M68 198 1



case I for G5 service tools CFO empty



30



M68 199 1



case II for G5 service tools empty



31



M67 985 1



testing tool single empty blood bag



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10



12-25



Chapter 12: Spare parts catalog



Tools and accessories



12-26



Pos.



Part no.



Description



32



661 578 1



Testing-Pin Holder for Service G5



33



661 580 1



Testing-Pin Holder for Production



34



M68 170 1



Syringe Grease POLYLUB GLY151 6 ml



35



661 592 1



Assembly tool CFO RCC



36



661 591 1



assembly tool CFO Top/Door



37



641 950 1



GLUE LOCTITE NO.243; 50ML



38



M60 698 1



LOCTITE GLU NO.270/50 ML SCREWSAFE



39



M62 029 1



REINIGER LOCTITE 7063 400 ML-SPRUEHDOSE



40



M67 961 1



Service tool power measurement slide G5



41



M67 484 1



Update CompoMat G5 V01.01.06 USB-Stick



42



M68 314 1



dynamometric screwdriver 10-80cNm



43



M68 315 1



dynamometric screwdriver 40-250cNm



44



M68 316 1



reversing blade 6mm with 1/4”-bit-drive



45



M68 317 1



reversing blade 6mm with external square



46



M68 318 1



hexagon socket wrench - SW7



Fresenius Kabi



CompoMat G5



SM-EN



3/11.10