BeneFusion VP1&VP3 - Service Manual - V1.0 - EN [PDF]

  • 0 0 0
  • Suka dengan makalah ini dan mengunduhnya? Anda bisa menerbitkan file PDF Anda sendiri secara online secara gratis dalam beberapa menit saja! Sign Up
File loading please wait...
Citation preview

BeneFusion VP3/VP1 (This manual is also applicable to BeneFusion VP3 EX, VP1 EX, SK-410, SK-430 and Vet series Infusion Pump)



Infusion Pump



Service Manual



Intellectual Property Statement SHENZHEN SHENKE MEDICAL INSTRUMENT TECHNICAL DEVELOPMENT CO., LTD. (hereinafter called SK Medical) owns the intellectual property rights to this product and this manual. SK Medical is a member of Mindray Medical Group. © Copyright 2015-2016 Shenzhen Shenke Medical Instrument Technical Development Co., Ltd. All rights reserved. Release, amendment, reproduction, distribution, rental, adaptation, translation of this manual in any manner whatsoever without the written permission of SK Medical is strictly forbidden.



,



,



and



are registered



trademarks or trademarks owned by SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO.,LTD. (hereinafter called Mindray) in Cina and other countries. K Medical is authorized by Mindray to use above registered trademarks or trademarks.



Statement SK Medical possesses the final interpretation right for this manual. Contents of this manual are subject to changes without prior notice. Any modified content will appear in the newly-published version of this manual. SK Medical is responsible for the effects on safety, reliability and performance of this product, only if:  All installation operations, expansions, changes, modifications and repairs of this product are conducted by SK Medical authorized personnel.  All replacement parts used in the maintenance as well as the supporting accessories and consumables must be originals (genuine originals) of SK Medical or approved by SK Medical.  The electrical installation of the relevant room complies with the applicable national and local requirements.  The product is used in accordance with the instructions for use.



I



Preface Manual Purpose This Service Manual provides detailed introductions to the hardware composition, installation, dismantling, testing and troubleshooting of this product and its related accessories, so that the maintenance personnel can handle the common problems effectively. It does not provide in-depth information on the product structure and design principle. If you experience problems that cannot be solved, please contact our Ater-sale Service Department. The introductions to this product in this Service Manual are based on the most complete configuration, so some of them may not be applicable to the product you're maintaining. In case of any question, please contact our after-sale service department. Before product maintenance, please read this Service Manual carefully, and ensure that you have fully understood the content, so that you can maintain the product correctly, avoiding product damage or personal injury.



Intended Audience This Service Manual is applicable to the professional biomedical engineers responsible for the product maintenance, the authorized maintenance personnel or the after-sale service representatives.



ConventionsVersion Information The version number of this Service Manual may be updated without notice at any time due to change in the software or technical specification. The version information of this Service Manual is as follows:  Version No.: 1.0  Issue date: 2015-06



II



Contents 1 Safety .....................................................................................................................1-1 1.1 Safety Information ...............................................................................................1-1 1.1.1 Dangers ...................................................................................................1-2 1.1.2 WARNING ...............................................................................................1-2 1.1.3 CAUTION ................................................................................................1-3 1.1.4 NOTE.......................................................................................................1-4 1.2 Equipment Symbols ............................................................................................1-5 2 Design Principle ...................................................................................................2-1 2.1 Product overview .................................................................................................2-1 2.1.1 Structural Composition and Performance ...............................................2-1 2.2 Product appearance ............................................................................................2-1 2.2.1 Front View ...............................................................................................2-1 2.2.2 Front View with the Door Opened ...........................................................2-3 2.2.3 Rear View ................................................................................................2-4 2.3 Hardware Principle ..............................................................................................2-5 2.3.1 Control Board ..........................................................................................2-6 2.3.2 Key Board ................................................................................................2-9 2.3.3 Motor Detection Board ............................................................................2-9 3 Test and Maintenance ..........................................................................................3-1 3.1 Description...........................................................................................................3-1 3.1.1 Test report................................................................................................3-1 3.1.2 Recommended frequency .......................................................................3-2 3.2 Complete Machine's Appearance Inspection ......................................................3-2 3.3 Startup test ..........................................................................................................3-3 3.4 Accuracy Calibration............................................................................................3-3 3.5 Pressure Calibration ............................................................................................3-5 3.6 Operational test ...................................................................................................3-7 3.7 Electrical safety test ............................................................................................3-8 3.7.1 Housing Leakage Current Test................................................................3-9 3.7.2 Ground Leakage Current Test .................................................................3-9 3.7.3 Patient Leakage Current Test............................................................... 3-10 3.8 Cleaning and Disinfection................................................................................. 3-10 3.8.1 Cleaning ............................................................................................... 3-10 3.8.2 Disinfection ........................................................................................... 3-10 4 Troubleshooting ...................................................................................................4-1 4.1 Overview ..............................................................................................................4-1 1



4.2 Replacement of components ..............................................................................4-1 4.3 Check the Infusion Pump Status .........................................................................4-1 4.3.1 Operation Methods for AD Value Channels and the Corresponding Tests4-1 4.3.2 Fault Table ...............................................................................................4-3 5 Maintenance and Disassembly ...........................................................................5-1 5.1 Tool preparation ...................................................................................................5-1 5.2 Disassembly procedure .......................................................................................5-1 5.2.1 Separate the Front and Rear Housings ..................................................5-1 5.2.2 Disassembly of Front Housing Kit ...........................................................5-2 5.2.3 Disassembly of Door Kit ..........................................................................5-6 5.2.4 Disassembly of PCBA Kit ........................................................................5-8 5.2.5 Disassembly of Pump Kit ..................................................................... 5-10 5.2.6 Disassembly of Rear Housing Kit ......................................................... 5-11 6 Components..........................................................................................................6-1 6.1 Description...........................................................................................................6-1 6.2 Infusion Pump Host .............................................................................................6-2 6.2.1 Explosive View ........................................................................................6-2 6.2.2 Parts List..................................................................................................6-2 6.3 Door Assembly ....................................................................................................6-3 6.3.1 Explosive View ........................................................................................6-3 6.3.2 Parts List..................................................................................................6-3 6.4 Front Housing Assembly .....................................................................................6-4 6.4.1 Explosive View ........................................................................................6-4 6.4.2 Parts List..................................................................................................6-4 6.5 Rear Housing Assembly ......................................................................................6-5 6.5.1 Explosive View ........................................................................................6-5 6.5.2 Parts List..................................................................................................6-5 6.6 PCBA Assembly...................................................................................................6-6 6.6.1 Explosive View ........................................................................................6-6 6.6.2 Parts List..................................................................................................6-6 6.7 Pump Assembly ...................................................................................................6-7 6.7.1 Explosive View ........................................................................................6-7 6.7.2 Parts List..................................................................................................6-7 7 Upgrading ..............................................................................................................7-1 7.1 Tools ....................................................................................................................7-1 7.2 Software Upgrading .............................................................................................7-1 7.3 Software Burning Method ....................................................................................7-1



2



1 Safety 1.1 Safety Information The safety statements presented in this chapter refer to basic safety information that the operator must pay attention to and abide by when using the infusion pump.There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specific to particular operations.



Dangers 



Indicates an imminent hazard that, if not avoided, could result in death, serious injury or damage to product/property.



WARNING 



Indicates a potential hazard or unsafe practice that, if not avoided, could result in death, serious injury or damage to product/property.



CAUTION 



Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury, product malfunction or damage to product/property.



NOTE 



Provides application tips or other useful information to ensure that you get the most out of the product.



1-1



1.1.1 Dangers This Manual does not contain any information at the "Danger" level.



1.1.2 WARNING WARNING 



Device, cables and accessories must be inspected before use to guarantee their normal and safe operation.







This equipment can only be connected to the socket with ground protection. Please adopt a rechargeable battery instead of the socket as the power supply if the socket is not provided with a ground lead.







To prevent fire or explosion, do not operate the equipment in the presence of anesthetic, flammable or explosive materials.







Do not open the equipment casing as there is the impending danger of electric shock. Equipment maintenance and upgrades must be carried out by maintenance technicians whom are trained and licensed by the manufacturer. Moreover, the process must be done only after the AC power supply is disconnected. Maintenance carried out by individuals non-affiliated to the manufacturer or by non-licensed personnel may affect the safety, performance and function of the product.







When used with electrosurgery equipment, the safety of patients should be ensured.







The patient's clinical condition and the working condition of the infusion pump must be monitored carefully, the alarm volume and alarm levels need to be set according to the actual needs. Operation and performance relying solely on the auditory alarm system alone is not sufficient, and setting the alarm at a low volume may endanger the patient. If the alarm volume is less than the surroundings volume, which can further lead to operators identify alarm mistakenly.







Please carefully install the power line and cables with various accessories to prevent the patient from choking or suffocation caused by entanglement of the cables or by electrical disturbance.







The packaging materials must be disposed of in compliance with local laws and regulations or the hospital policy on waste management.They must be kept out of the reach of children.







Infusion tube knots, filter coagulation and occlusions arising from needle insertion can cause the pressure inside the infusion tube to rise during infusion. When this occurs, removing the occlusion can cause 1-2



excessive liquid drug to be infused into the patient, so appropriate measures should be taken. 



The pump should not be placed more than 100 cm above or below the level of the patient’s heart. The smaller the height difference between the pump, the more accurate the pressure test in the infusion cannula will be.







This equipment has to be used with infusion tubes recommended by manufacturers, SK Medical is not responsible for its accuracy and alarm function with non-recommended infusion tubes.







Do not touch the patient when connecting the peripheral equipment via the input/output signal ports to prevent patient leakage current from exceeding the requirements specified by the standard.



1.1.3 CAUTION CAUTION 



Use the accessories specified in this Operator’s Manual to guarantee the patient’s safety.







When this infusion pump and its accessories exceed their service life, they must be disposed of in accordance with local statutes or hospital regulations. If you have any queries, please contact your distributor or the manufacturer.







After installation of the infusion tube and before infusion, check for leakages. If any are found, they should be rectified as soon as possible.







User should adjust the fixing site of the infusion tube every 4 hours after infusion begins, in order to guarantee its accuracy. We suggest to replace the infusion tube after the infusion has been running for 24 hours.







Electromagnetic fields may affect equipment performance. This makes it necessary for other equipment used in the vicinity of the pump to meet EMC standards. Mobile phones, X ray and MRI equipment are all potential interference sources because of their high-intensity electromagnetic radiation.







Before the equipment is connected to the power supply, check that the voltage and frequency of the power supply match the specifications on the label or in this Operator’s Manual.







Please install and carry the equipment correctly to protect the equipment from damage from drops, impacts, violent shaking or other 1-3



external mechanical forces. 



Disposable accessories must be disposed of after use in accordance with the relevant hospital regulations.







Avoid direct sunshine, high temperatures and dampness.







Check the built-in battery before use to make sure it has sufficient power. Recharge the battery if necessary.



1.1.4 NOTE NOTE 



Install the equipment in a position where it can be easily accessed for inspection, operation and maintenance.







Keep this Operator’s Manual near to the equipment for future ease of reference.







The software of the equipment is developed according to the software development demands of IEC60601-1 standard, which can minimize the possibility of the risk caused by program error.







This Operator’s Manual describes the most complete functional configuration of the equipment. The product you are using may not have some of the settings or functions described herein.







Do not insert devices that are not specified by the manufacturer into the DB9 interfaces.







During infusion, the infusion pump can accurately control the rate, infusion volume and infusion time, and monitor the operation in real-time, to effectively prevent over currents, under currents and instances of backflow.







The device is not in touch with the drugs or patients directly. Thus, there is no need to process Biocompatibility test on it.



1-4



1.2 Equipment Symbols The equipment you purchased may not provide you with all the following symbols. NOTE! Refer to the accompanying document (This Manual)



ON/OFF



Alternating current power supply (AC)



Battery



Alarms



Silence



Clear/Back



Start



Bolus



Confirm



Stop



Menu



Move up/Increase



Move down/Decrease



Move left



Move right



Infusion tube



Selected drug



Caution



Lock



Wireless modules work in order



Wireless transceiver



DB9 interface



Night mode



Drop sensor interface



Protected against defibrillation CF applied parts



Date of manufacture



Manufacturer



Electronic equipment: dispose of separately to avoid polluting the environment



Serial number



The European Union Representative Office



CE mark



1-5



IP34



Protected against solid foreign objects with a diameter no less than 2.5mm and protected against spraying liquid water



Environmentally-friendly use periods of electronic products (20 years)



Package should be kept between 50–106 kPa during transport



Package should be kept between 10%–95% humidity during transport



Package should be kept between -20–60°C during transport



1-6



2 Design Principle 2.1 Product overview 2.1.1 Structural Composition and Performance The Infusion Pump primarily consists of a housing and built-in battery, and all components of the infusion pump are suitable for use in patient environment. The drop sensor and wireless module are optional. Optional functions of the software comprise Rate Mode, Time Mode, Body Weight Mode, Sequential Mode, Drop Rate Mode, Drug Library, History Record, and Anti-bolus Function. Since some parts and functions are optional, the Infusion Pump you purchased may not contain these additional parts and their relevant functions.



2.2 Product appearance 2.2.1 Front View



1. Alarm indicator lamp The alarm indicator lamp indicates different alarm levels in different colors and flash frequencies. 2-1



2. Display Used for displaying infusion parameters and relevant content. 3.  Under non-setting status, indicate to return to the previous menu or operation.  Under the setting status, indicate to clear the current set or cancel the edit. 4. Used for adjusting value, change lines and pages. 5. Used for confirming input operation and saving values. 6.  Under non-operation status, used for switching [Main Menu] interface and other interfaces.  Under operation status, press and hold this key to lock; in locked state, press and hold to unlock. 7. AC/DC indicator light  On: The pump is connected to an AC/DC power supply (including shutdown).  Off: The pump is not connected to an AC/DC power supply. 8. Battery indicator  Steady green indicates that the battery is charging.  Flashing indicates that the battery is providing power.  Light off indicates that there is no battery or the equipment is turned off and not connected to an AC power supply. 9.  Used for turning power on, entering in standby state and turning off operations.  When power off, press and hold (>3 s) the key. 10.  During infusion, press this key to enter the Bolus settings screen.  When the pump is stopped, press this key to enter the Purge prompt screen. 11.  For high level and mid-level alarms, press this key to silence for 2 minutes. The alarm will be cancelled automatically if a new alarm is triggered within 2 minutes.  For low-level alarms, press this key to cancel alarm. 12. After installing the infusion tube correctly and completing setting infusion parameters, press this key to start the infusion. 13. During infusion, press this key to stop infusion. Infusion stops caused by alarms, such as occlusion and so on, press this key to cancel the alarm. 14. Handle 15. Door 16. Door holder Pull it to open the door.



2-2



2.2.2 Front View with the Door Opened



1. Liquid check clip button 2. Liquid check clip 3. Infusion tube slot



2-3



2.2.3 Rear View



1. DB9 interface, which combines the following interface functions:  DC power input interface  RS232 interface  Nurse call interface 2. Alternating current power supply (AC) port Connected by three-core-type power cord and AC power source. 3. Drop sensor interface 4. Pole clamp 5. Product label



2-4



2.3 Hardware Principle Hardware board framework of BeneFusion VP3 and VP1 Infusion Pump is shown in the following figure, which includes control board, key board and motor detection board. Wi-Fi function is optional for key board. BeneFusion VP3 infusion pump is equiped with display of 240×128 dots; BeneFusion VP1 infusion pump is equiped with display of 132×32 dots. Motor detecting board Bubble detector assembly



15V/1.33A



Display



J17



J4



J3



Yellow /red alarm light



Wifi module (optional)



J5 Holzer sensor circuit



Key circuit



Key board



Display interface circuit



Buzzer



J1



DB9 J3



J9



J13



Stepper yr motor e tt a B yr te t a B



J8 Drop sensor interface Infrared interface (backup) Pressure sensor



J11 Control panel: Master CPU STM32F103RCT6 Sub CPU STM32F030R8T6



J7



J10 J2 J5



J4



J6



J12



Loudspeaker ACDC module



100-240VAC 50/60Hz



Hardware Board Framework of BeneFusion VP3 and VP1 Infusion Pump



2-5



2.3.1 Control Board Control board of BeneFusion VP3 and VP1 Infusion Pump consists of power, main control and monitoring sections.



锂电池 2 7.4V/1600mAh Lithium batteries 7.4V/1600mAh



ACDC 15V/1.33A



Charging circuit with lithium battery



Monitoring CPU



DCDC TPS61175 23.7V



Motor drive



Main control CPU



VBUS Switch control circuit



DCDC TPS5420 5V



Speaker drive LCD backlight



LDO SPX1117



System 3.3V



Standby alarm circuit



VREF ZR431



ADC calibration



Standby power ZSR500GTA-5V



Standby power SPX5205



Main control CPU



Monitoring CPU



External DC input 10-16V/2-1.25A



Power System Chart Key signal testing points: Tested by a multimeter, TP124 (ACDC_IN) is about 15 V. Tested by a multimeter, TP13 (EDC_IN)is about 10 V~16 V. Tested by a multimeter, TP125 (BAT2) is about 6.8V-8.3V (when dual battery configuration). Tested by a multimeter, TP123 (BAT1)is about 6.8V-8.3V. Tested by a multimeter, TP127 (VBUS) is about 6.4V-15.6V. Tested by a multimeter, TP126 (VPP) is about 6.4V-15.6V. Tested by a multimeter, TP3 (VCCB) is about 4.5V-5.2V. Tested by a multimeter, TP2 (VBB) is about 3.3V. Tested by a multimeter, TP4 (VCC) is about 5.0V. Test ed by a multimeter, TP1 (VDD) is about 3.3V. Tested by a multimeter, TP83 (VMM) is about 23.7V.



2-6



RTC power supply Standby power of Monitoring CPU



ON/OFF control



Dot matrix LCD 132*32/240*128 Multiplex IO



Bubble detection



Key Pressure detection Loudspeaker



Main control CPU STM32F103RC LQFP64



Motor detection



Alarm light Motor power control



Infrared communication (reserved)



Monitoring CPU communication



Multiplex USART



RS-232 communication



RTC Enable control



Wifi module (optional)



EEPROM



Flash



Nurse call



Main Control Module Chart Key signal testing points: TP82



UART signal sent by the main control CPU to the monitoring CPU



TP44



UART signal received by the main control CPU from the monitoring CPU



TP6



External serial UART signal sent by the main control CPU



TP43



External serial UART signal received by the main control CPU



TP41



SDA signal with which the main control CPU communicates with EEPROM



TP42



SCL signal with which the main control CPU communicates with EEPROM



TP51



CS signal with which the main control CPU communicates with Flash



TP50



SCLK signal with which the main control CPU communicates with Flash



TP49



MISO signal with which the main control CPU communicates with Flash



TP52



MOSI signal with which the main control CPU communicates with Flash



TP47



Reset signal of the main control CPU



TP9



Nurse call control signal



TP107 TP108



PWM modulation input signal



TP109



Voice chip enabling signal



2-7



Bubble detection



Buzzer Standby Alarm circuit



Door on/off detection



Alarm light



Pressure detection



Main control CPU communication



Monitoring CPU STM32F030R8 LQFP64



Drip detection Motor detection



Battery status indicator light



System circuit voltage detection Motor drive



Battery charge status detection



Monitoring Module Chart Key signal testing points: TP81



Reset signal of the monitoring CPU



TP84



U7 stepping signal controlled by the monitoring CPU



TP88



U7 enabling signal controlled by the monitoring CPU



TP87



U7 direction signal controlled by the monitoring CPU



TP85



U7 current size signal controlled by the monitoring CPU



TP25, TP26



Motor coupler detection signal



TP24



Bubble detection output signal



TP29



Ultrasonic emission enabling signal



TP36



Holze switch door detection output signal



TP98



Pressure detection signal



TP38



Drip enabling signal



TP23



Drip detection signal



2-8



2.3.2 Key Board Serial communication



Keys



Wifi module (optional)



From control board Red-yellow alarm light



LCD LCD backlight control circuit



Battery indicator light



AC indicator light



Hall sensor



Ultrasonic transceiver circuit



Gold capacitor circuit



Motor detecting board interface



Key Board Chart Key testing points TP50



VDD 3.3V



TP51



VCC 5.0V



TP32



LCD backlight power



TP53



Signal of battery indicator light



TP52



Signal of AC indicator light



TP24



Serial output signal of Wifi module



TP25



Serial input signal of Wifi module



TP56



Gold capacitor power supply voltage



TP54, TP55



Red-yellow alarm light drive signal



TP1



Holze switch door detection signal



TP34, TP35



Double coupling motor testing signal



TP4, TP5



Ultrasonic emission signal



TP8



Ultrasonic receiving signal



2.3.3 Motor Detection Board Motor speed detection adopts double coupling circuit, which detects direction and speed; a gap on the detection encoder is used for detecting the camshaft of pump body. TP1, TP2



Motor coupler output signal



2-9



3 Test and Maintenance 3.1 Description To ensure long-term stability of the infusion pump, maintenance personnel must carry out regular inspection, maintenance and testing. This chapter provides the basic testing methods for the infusion pump, and recommends the proper testing frequency and testing tools. Maintenance personnel should choose proper testing tools to carry out inspection and testing according to the actual needs. The tests and testing methods provided in this chapter are mainly used to verify whether the performance of the infusion pump can meet the specifications. If test results do not meet the specifications, it indicates that a functional module has failed, which must be repaired or replaced immediately. In case of any other questions, please contact our After-sale Service Department in a timely manner.



CAUTION 



All the tests must be performed by qualified maintenance personnel only.







Prior to test, maintenance personnel shall ensure the applicability of the testing tools and connecting cables, and familiarize themselves with the use of these tools.



3.1.1 Test report After the professional maintenance personnel have performed the tests, please record in the following test report. Test equipment Name



Equipment Model



Equipment No.



Testing point



Test result



Test record Serial No.



Test content



1 2 3 4 5 Test judgment Pass or No Pass:



Tester:



Test date:



3-1



3.1.2 Recommended frequency The following tasks must be completed by professional maintenance personnel approved by our company. The equipment must be cleaned and disinfected before testing or maintenance. Inspection/ Maintenance Items Complete machine's appearance inspection



Frequency When first installation, or after each maintenance.



Startup test



1. 2.



When first installation; After each maintenance, upgrading or main accessory replacement.



Calibration



1. 2. 3.



At least once every two years; When you use a new brand of infusion set; When you doubt that the amount of infused liquid is not accurate; After each maintenance and upgrading.



4. Operational test



1. 2.



At least once every year; After maintenance of the infusion pump or when the user suspects that the battery is the failure source.



Housing leakage current test



1.



Patient leakage current test



2.



After maintenance or replacement of the power module; After machine fall-off or serious collision.



Battery powered test



Electrical safety test



At least once every year, or when you doubt that the injection flow or the alarm pressure is abnormal.



Cleaning and disinfection



1. 2.



Recommendation: once every month. Thoroughly clean the surface of the infusion pump before and after long-time storage.



3.2 Complete Machine's Appearance Inspection 1) 2) 3) 4)



Clean housing, no scratch, assembled firmly and no abnormal sound when shaken. Good handling with the keys. Complete and correct labels. Complete standard configuration and firm socket installation.



3-2



3.3 Startup test 1. 2. 3.



4.



Press



, the system will initiate the self-test and the screen will display



the [System Self-test] interface: The system will give out a sound "di" —— indicating the self-testing of the loudspeaker to be successful. The color of the alarm indicator lamp will change from red to yellow, turn on and off orderly —— indicating the self-testing of the alarm lamp to be successful. The system will give out a sound of "di" —— indicating the self-testing of the buzzer to be successful.



NOTE 



During startup, self-test will be performed in the following sequence: Self-test of register, RAM, ROM, jump instruction and arithmetic operation when CPU is powered on; Self-test of the alarm light; Self-test of the buzzer; Self-test of the speaker; Self-test of the external EEPROM; Self-test of the external FLASH; Self-test of the board power supply; Self-test of the battery status.



3.4 Accuracy Calibration Accuracy of infusion pump requires no daily calibration. However, calibration is required when the infusion pump is used for the first time, replaced with a new brand of infusion set or when you doubt that the deviation of the injection flow is much larger. Prepare the following materials before the calibration:  One electronic scales (500g)  One 100ml beaker  One infusion set The steps for accuracy calibration are as follows: 1. Connect the infusion set to the infusion bottle (bag) and install the set onto the infusion pump. 2. Enter the [Accuracy Calibration] interface: Select [Main Menu]→[System Maintenance]→Input Factory Maintenance Password 6686→[Accuracy Calibration]. 3. Interface will display the prompt information as shown in the figure, select the brand of infusion set as prompt.



3-3



4.



Press and hold



to pugre, put the ending of the infusion set into beaker.



5.



Press



6.



Weigh the liqud in beaker after infusion is completed, input the actual volume to [1st Volume].



7.



Press



to start the automic calibration of the infusion pump.



to confirm data and enter to the next calibration interface, put the



ending of the infusion set into beaker. 8.



Press



to start the automic calibration of the infusion pump.



9.



Weigh the liqud in beaker after calibration is completed, input the actual volume to [2nd Volume]. Press



10. Press



, calibration is completed.



to return and back to the [Main Menu] interface.



3-4



NOTE 



During accuracy calibration, please use medical normal saline solution, glucose solution or grade III water for experiments.







Please remove the needle portion of the infusion set, and do not place the ending of the infusion set above the liquid level in the beaker.







During the installation of the infusion set, please place the infusion set flat to the appropriate position and do not bend or stretch it.







The pictures in this manual are explained with BeneFusion VP3 Infusion Pump as the blueprint, with slight difference from BeneFusion VP1 Infusion Pump in the product display interface and other contents.



3.5 Pressure Calibration The steps for pressure calibration are as follows: 1. Enter the [Pressure Calibration] interface: Select [Main Menu]→[System Maintenance]→Input Factory Maintenance Password 6686→[Pressure Calibration]. 2. Follow the prompts and select the brand of the infusion set, prompts as shown in the diagram will appear on the interface.



3. Connect the infusion set to the infusion bottle (bag) and install the set onto the infusion pump. Connect the infusion set to a pressure gauge. 4. Press



to start the first calibration. Once the pressure reading on the



manometer reaches 150mmHg, press



to start collecting pressure values.



5. After the reading on the manometer stabilizes, follow the prompt and input the pressure value. 3-5



6. Press



to start the second calibration. Once the pressure reading on the



manometer reaches 500mmHg, press



to start collecting pressure values.



7. After the reading on the manometer stabilizes, follow the prompt and input the pressure value. Press 8. Press



, you will be prompted with [Calib done].



to return and back to the [Main Menu] interface.



NOTE 



The pump cannot be switched off directly from the Pressure Calibration screen.







If you have doubts about the accuracy of occlusion alarm pressure (pressure value exceeding the alarm threshold), you will need to perform a pressure calibration.







Only pressure measuring tools recommended and approved by this Company can be used for the pressure calibration.



3-6



WARNING 



The Pressure Calibration screen has no occlusion alarm. To avoid risk of serious harm to patients, do not perform infusion in this screen.



3.6 Operational test The steps for conventional functions test are as follows: 1. When various keys are operated on, the infuion pump shall perform or react normally in accordance with the User's Manual; abnormal operation should not cause the machine to crash. 2. Operate the infuion pump in accordance with the User's Manual. Check whether the sound-light alarm at various levels is normal, whether the volume control is normal and whether the alarm muting is working. 3. If an alarm is triggered when the pump is operated in accordance with the User's Manual, check whether the operations and alarms are properly recorded in the History Record. 4. Install the infusion set, set the infusion parameters correctly, and then press to enter the running interface. 5. During infusion, the operation icon, rate, volume and pressure value shall be normal. The steps for battery powered test are as follows: 1. The power source shall be able to supply power when the AC power supply is disconnected for a short time. 2. The indicator light shall be able to produce normal prompts in the battery charge and discharge status when battery is not installed.



WARNING 



The battery of this equipment is non-removable. Battery replacement must be performed by the maintenance personnel authorized by our company. Installation of fuel cell or replacement of battery by the insufficiently trained personnel will result in danger (e.g. overheat, fire or explosion).



3-7



3.7 Electrical safety test WARNING 



To test the electrical safety of the infusion pump, the electrical safety test is designed to detect abnormal electrical hazards which, if not discovered, may cause injury to the patient or the operator.







Commercially available testing devices such as the safety analyzer may be used for electrical safety tests. The maintenance personnel shall ensure the applicability, functional integrity and safety of such devices and familiarize themselves with the use of these devices.







Electrical safety test shall follow the following standard: IEC/EN 60601-1.







If local regulations provide provisions otherwise, please perform the relevant electrical safety test in accordance with the provisions.







In the patient area, all devices that are connected to the mains supply as well as to medical equipment must comply with the IEC/EN 60601-1 standard, and must be tested for electrical safety in accordance with the test interval for monitoring devices.



Electrical safety test is used to detect hazards that may pose electrical safety threats to the patient, operator and maintenance personnel. Please perform the electrical safety test in normal environment (including temperature, humidity and atmospheric pressure). While the 601 safety analyzer is used as an example in the electrical safety test described in this chapter, different safety analyzers may be used in different regions. Please ensure the applicability of the electrical safety test you are going to perform. Device connection diagram is as follows:



Tested equipment



3-8



A: AC power supply (programmable power supply, regulating frequency) B: Isolation transformer on the leakage current test tool C: Safety tester Testing tools:



 



Safety analyzer Isolation transformer



3.7.1 Housing Leakage Current Test 1. Connect the 601 safety analyzer to a power supply of 264 VAC and 60 Hz. 2. Using the connection tooling of the application section, connect the application section of the tested device and connect the SUM end of the connection tooling of the application section to the RA end of the safety analyzer. 3. Connect the tested device, via the power line, to the auxiliary power output jack of the 601 safety analyzer. 4. Connect one end of the red test lead to the “Red input terminal” of the safety analyzer and clip the other end to the metal foil attached to the surface of the housing of the tested device. 5. Power on the 601 safety analyzer, and press "5-Enclosureleakage” on the panel of the 601 safety analyzer to enter the interface for the housing leakage current test. 6. Housing leakage current is no more than 100μA under normal condition and no more than 300μA under single-fault condition.



3.7.2 Ground Leakage Current Test 1. Connect the 601 safety analyzer to a power supply of 264 VAC and 60 Hz. 2. Connect the application section of the tested device to the RA end of the safety analyzer. 3. Connect the tested device, via the power line, to the auxiliary power output jack of the 601 safety analyzer. 4. Power on the 601 safety analyzer, and press "4-Earth leakage” on the panel of the 601 safety analyzer to enter the interface for the ground leakage current test. 5. Housing leakage current is no more than 500 μA under normal condition and no more than 1000 μA under single-fault condition.



3-9



3.7.3 Patient Leakage Current Test 1. Connect the 601 safety analyzer to a power supply of 264 VAC and 60 Hz. 2. Using the connection tooling of the application section, connect the application section of the tested device and connect the SUM end of the connection tooling of the application section to the RA end of the safety analyzer. 3. Connect the tested device, via the power line, to the auxiliary power output jack of the 601 safety analyzer. 4. Power on the 601 safety analyzer, and press "6-Patient leakage” on the panel of the 601 safety analyzer. 5. Continuously press the "APPLIED PART" key to select the AC and DC measurements; “DC” is shown following the limit value of direct current. 6. Patient leakage current is no more than 10 μA under normal condition and no more than 50 μA under single-fault condition. If the electrical safety test fails, please contact our after-sale service department.



3.8 Cleaning and Disinfection 3.8.1 Cleaning The pump should be cleaned regularly. If operating in dirty or sandy areas, cleaning should be more frequent. Before cleaning, please consult or refer to the hospital's specific regulations concerning medical device cleaning. The recommended detergents include: Hydrogen peroxide (3%). To clean your equipment, follow these rules: 1. Turn off the pump and disconnect the AC power source line. 2. Wipe the display screen after soft cotton balls absorb an appropriate amount of detergent. 3. Use a piece of soft cloth which absorbs a modest amount of cleaning agent to wipe the surface of the device. 4. When necessary, use a piece of cloth to wipe off any excess cleaning agents. 5. Place the pump in a cool and ventilated environment to dry.



3.8.2 Disinfection The operation of disinfection may cause certain damage to the infusion pump. You are recommended to disinfect only when it is necessary in your desired maintenance plan. Clean the equipment before disinfection. The recommended disinfectants include: glutaraldehyde-type 2% liquid disinfectant. 3-10



CAUTION 



Never use EtO or formaldehyde for disinfection.







Do not conduct high pressure or high temperature disinfection for the infusion pump and its accessories.



3-11



4 Troubleshooting 4.1 Overview In this chapter, faults of infusion pump are classified according to the faulty components and faulty phenomena. Please refer to the relevant Fault Table when fault examination, identification and troubleshooting in sequence. The recommended solutions given in this chapter should help you solve most of the equipment faults you will encounter but not all possible problems. In the case of a fault not covered in this chapter, please contact our After-sale Service Department.



4.2 Replacement of components You may replace the circuit board components and other major components or parts of this infusion pump. Once you have identified the faulty circuit board component, you may follow the steps described in Chapter 5 Maintenance and Disassembly to replace the circuit board component. Then you may check whether the fault has been eliminated or whether the infusion pump is able to pass the relevant tests. If the fault has been eliminated, which shows that the original circuit board is damaged, then please return the faulty circuit board component to our company for repair. If the fault persists, please reassemble the original circuit board for troubleshooting according to other possible causes. Please refer to Chapter 6 Components for more information about the components to be replaced.



4.3 Check the Infusion Pump Status 4.3.1 Operation Methods for AD Value Channels and the Corresponding Tests 1. Operation method for entering the AD value interface: System maintenance --enter the password --- sensor AD value.



4-1



      



         



Rotation speed - -set the motor rotation speed Direction - - set the motor direction (1--forward, 0--reverse) Current - -set the motor drive current gear (four gears, from small to large: 0, 1, 2 and 3) Steps - -rotated steps of stepper motor Drops- -drop senfor detected value (when drop sensor detects drip, automatically add - -) MSD- -motor coupler detected value (motor rotation, foure status: 0, 01, 10 and 11) DOOR- -opening and closing door detection (1—close the door, 0—open the door)



VCC -- 5V power voltage value (normal range: 4.5~5.5) VMM -- motor power voltage value (normal range: 21.3~26.1) VDD -- 3.3V power voltage value (normal range: 3.1~3.5) VDDB -- standby 3.3V power voltage value (normal range: 3.0~3.7) EDC -- external DC power voltage value (normal range: 10~17) BAT1 -- primary battery power voltage value (normal range: 5.8~8.4) BAT2 -- secondary battery power voltage value (normal range of dual-battery configuration: 5.8~8.4) ACDC -- AC power module output voltage value (normal range: 10~17) VCCB -- standby 5V power voltage value (normal range: 4.5~5.5) VGC -- gold capacitor voltage value (normal range: 0~5.2) 4-2



 



BUBBLE-- bubble detected value (AD value less than 620 when there is bubble, AD value more than 621 when pure water) PS—pressure detected value (normal range: 500-4000)



NOTE 



The pictures in this manual are explained with BeneFusion VP3 Infusion Pump as the blueprint, with slight difference from BeneFusion VP1 Infusion Pump in the product display interface and other contents.







The explanations in this manual are based on the highest configuration product, while there is no rotation speed, direction, current or other items concerning eneFusion VP1 Infusion Pump.



4.3.2 Fault Table During transportation, storage and use of infusion pump, there can be faults that may affect the normal use as a result of various factors (e.g. grid voltage instability, change in ambient temperature, drop and shock and component aging). In such cases, professional personnel with certain experience in electronic medical equipment maintenance shall carry out component level overhauling in accordance with the following fault classification. The so-called part level overhauling refers to the analysis, replacement, test-run and identification of the equipment's faulty component(s), such as power supply, main control section, monitoring section and/or LCD. The maintenance of components, also known as component level maintenance, must be carried out by the manufacturer's engineers with rich experience in maintenance, using specialized equipment, in a specific environment and under specific conditions. Complete machine fault Faulty phenomenon



Possible cause



Handling method



When external power supply is not connected, there is no display on startup, battery indicator light is not light and buzzer does not sound.



1. Battery is dead; 2. Key board and control board are inserted in poor contact; 3. Control board fault.



1. Replace with a new battery; 2. Insert the key board and control board in the correct location and direction; 3. Replace the control board.



When external power supply is not connected, there is no display on



1. Display fault; 2. Short circuit.



1. Handle as with display fault; 2. Further check the control board and other boards.



4-3



startup, battery indicator light is on and buzzer sounds. When external AC power supply is connected, AC indicator light is not on.



1. The AC power supply wires are in poor contact; 2. Key board and control board are inserted in poor contact; 3. Faulty ACDC module.



1. Repair or replace the AC power supply wires; 2. Insert the key board and control board in the correct location and direction; 3. Repair or replace the ACDC module.



Self-test 4 not passed; alarm



Control board fault.



Replace the control board.



Self-test 5 not passed; alarm



1. Control board fault; 2. Battery fall-off or damage.



1. Replace the control board; 2. Replace the battery.



Unable to charge properly



1. The battery interface board wires are in poor contact; 2. Battery fault; 3. Control board fault.



1. Repair or replace the battery interface board wires; 2. Replace the battery; 3. Replace the control board.



Failure of an operation or measurement function



Damage to the control board or corresponding components.



Further check the control board and corresponding components.



Normal startup power supply, black or white display screen; or black or white display screen during normal operation



1. Display screen fault; 2. Key board and control board are inserted in poor contact; 3. Key board fault; 4. Control board fault.



1. Replace the display screen; 2. Insert the key board and control board in the correct location and direction; 3. Replace the keyboard; 4. Replace the control board.



Error in the displayed text or picture



1. Error in the resource file burning; 2. Control board fault.



1. Burn the resource file again; 2. Replace the control board.



Operation and alarm faults Faulty phenomenon Key failure



Possible cause



Handling method



1. Key board and control board are inserted in poor contact;



4-4



1. Insert the key board and control board in the correct



2. Key board fault; 3. Control board fault.



location and direction; 2. Replace the key board; 3. Replace the control board.



Buzzer sound failure



1. Control board fault; 2. Key board and control board are inserted in poor contact; 3. Key board fault.



1. Replace the control board; 2. Insert the key board and control board in the correct location and direction; 3. Replace the key board.



Speaker sound failure or hoarse sound



1. Speaker fault; 2. Speaker wire interface is in poor contact; 3. Speaker alarm circuit fault.



1. Repair the wires or replace the speaker components; 2. Repair or replace the speaker interface; 3. Further check the faulty circuit.



Failure of the red alarm light or yellow alarm light



1. Key board and control board are inserted in poor contact; 2. Key board fault; 3. Control board fault.



1. Insert the key board and control board in the correct location and direction; 2. Replace the key board; 3. Replace the control board.



Monitoring fault Faulty phenomenon



Possible cause



Handling method



Pressure value unchanged



1. Pressure sensor fault; 2. Amplifier circuit fault; 3. Wires fault.



1. Replace the pressure sensor; 2. Replace the control board; 3. Replace the wires or insert the wires again.



Pressure value overrange



1. Pressure sensor fault; 2. Wires fault.



1. Replace the pressure sensor; 2. Replace the wires or insert the wires again.



Motor fault



1. The motor wires are in poor contact; 2. Motor detection board fault; 3. Stepper motor drive circuit fault; 4. Motor fault; 5. Motor detection board and key board wires are in poor contact.



1. Repair or replace the motor wires; 2. Replace the motor detection board; 3. Replace the control board; 4. Replace the motor; 5. Replace the wires or insert the wires again.



Bubble detection fault



1. Ultrasonic chip fault; 2. Ultrasonic wires fault; 3. Ultrasonic receiving and transmitting fault; 4. Control board and key board



1. Replace the ultrasonic chip; 2. Replace the wires or insert the wires again; 3. Replace the key board; 4. Insert the control board and



4-5



inserted, in poor contact. Opening and closing door detection fault



1. Hall sensor fault; 2. Control board and key board are inserted in poor contact.



4-6



key board again. 1. Replace the sensor; 2.



Insert the control board and key board again.



5 Maintenance and Disassembly NOTE 



During maintenance and disassembly, an anti-static wrist strap should be worn at all times.







This manual only provides the disassembly method and sequence, while the assembly method and sequence are reverse.







When assembly, note that the wire connection direction, the wiring path and the wire fixation location shall be consistent with those before disassembly.







When assembly, note that the seal ring and sponge pat shall be consistent with those before disassembly.



5.1 Tool preparation During disassembly and replacement of components, you may need the following tools: 1. Phillips screwdriver (size 107) 2. Flathead screwdriver 3. Diagonal pliers 4. Anti-static wrist strap



Disassembly procedure 5.1.1 Separate the Front and Rear Housings Note: An anti-static wrist strap should be worn at all times. 1. As shown in the figure: Unscrew and remove 2 units of M5x12 Phillips pan head screws on tube clamp kit with the Phillips screwdriver. Remove the clamp kit and unscrew 4 units of M3x8 Phillips pan head screws on the rear housing.



5-1



2. As shown in the figure: Separate the front and rear housings, and disconnect the wires.



5.1.2 Disassembly of Front Housing Kit Note: An anti-static wrist strap should be worn at all times. 1. Disassembly of pump kit As shown in the figure: For starter, unplug the motor wires. Then unscrew and remove 4 units of M3x8 Phillips pan head screws on pump kit with the Phillips screwdriver. Take out the pump and mind the transparent water-proof film. The transparent water-proof film should be lubricated with oil, and please mind the layout of motor wires so as to separate the wire with the rear housing.



5-2



2. Disassembly of pressure sensor As shown in the figure: First, disconnect the sensor wires and unscrew 2 units of M3x12 Phillips pan head screws on pressure sensor kit with the Phillips screwdriver. Then pull the sensor kit upwards and separate the kit with the housing. (Note, the pressure sensor kit cannot be disassembled, otherwise the accuracy may be affected.). Use the Phillip screwdriver to remove 2 units of M3x12 self-tapping screws and take out the pressure sensor gasket.



5-3



3. Disassembly of liquid stopper clamp As shown in the figure: Unscrew 2 units of M3x12 Phillips pan head screws and take out the clamp kit.



4. Disassembly of motor detection board As shown in the figure: Unscrew the M3x8 Phillips self-tapping screws and take out the motor detection board.



5-4



5. Disassembly of ultrasonic kit As shown in the figure: Unscrew the M3x8 Phillips self-tapping screws with the screwdriver. Insert the flathead screwdriver into the hole, press slightly and remove the ultrasonic kit.



6. Disassembly of door clip As shown in the figure: Unscrew a M3x8 Phillips self-tapping screws with gasket and take out the door clip.



5-5



5.1.3 Disassembly of Door Kit Note: An anti-static wrist strap should be worn at all times. 1. As shown in the figure: Unscrew the bus screw by using flathead screwdriver, and unplug the extension wire that connected with the ultrasonic kit and the keyboard. The top of bus screw should be sealed by glue so as to prevent the screw from loosing.



2. As shown in the figure: Use the flathead screwdriver to press the door kit button (at the bottom of machine), push the door kit outward and remove the kit.



5-6



3. As shown in the figure: Use the Phillips screwdriver to remove 4 units of M3x8 Phillips pan head screws.



4. As shown in the figure: Open the front housing and the internal structure is demonstrated.



5-7



5.1.4 Disassembly of PCBA Kit Note: An anti-static wrist strap should be worn at all times. 1. As shown in the figure: Unscrew 2 units of M3x8 Phillips pan head screws on PCBA control board by using flathead screwdriver, unplug the speaker wire, and take out the PCBA control board.



、 2. As shown in the figure: Unscrew 6 units of M3x8 Phillip self-tapping screw on keyboard PCBA and take it out slightly. The indicator should be well protected in order to avoid the loosing.



5-8



3. As shown in the figure: Loose the clip and take out the indicator.



5-9



5.1.5 Disassembly of Pump Kit Note: An anti-static wrist strap should be worn at all times. 1. As shown in the figure: Unscrew 2 units of M3x8 Phillips pan head screws with Phillips screwdriver and take out the motor.



2. As shown in the figure: Unscrew and remore 2 units of M3x8 Phillips pan head double washer screws with Phillips screwdriver, take out the motor damping piece and gear.



5-10



5.1.6 Disassembly of Rear Housing Kit Note: An anti-static wrist strap should be worn at all times. 1. As shown in the figure: Use the Phillips screwdriver to remove 2 units of M3x8 Phillips pan head screws, push the waterproof sheet, remove 2 units of M3x8 Phillips pan head screws, disconnect the AC power plug and remove the power module.



5-11



2. As shown in the figure: Use the diagonal pliers to cut the ribbon of tightening lithium barttery, and take off the battery.



5-12



6 Components 6.1 Description This chapter provides the explosive view of the main components of the infusion pump host and their material codes to help the maintenance personnel identify the components during disassembly and replacement of components. Structural system of the infusion pump host is indicated in the following figure:



Structural composition



6-1



6.2 Infusion Pump Host 6.2.1 Explosive View



Explosive View of Infusion Pump Host



6.2.2 Parts List Serial No.



Sub-part code



Sub-part description



Quantity



1



115-036533-00



VP3 door kit



1



2



049-000445-00



Foot pad, silica gel



4



3



\



Front housing



1



4



\



Pump body



1



5



\



VP3 rear housing kit



1



6



115-031552-00



VP1/VP3 pole clamp (advanced)



1



7



115-031551-00



VP1/VP3 pole clamp (standard)



1



Parts List of Infusion Pump Host



6-2



6.3 Door Assembly 6.3.1 Explosive View



Explosive View of Door Assembly



6.3.2 Parts List Serial No. 1



Sub-part code \



Sub-part description



Quantity



Sub-component of VP3 door assembly



1



2



115-039030-00



VP3 door ultrasound assembly



1



3



043-006453-00



Rear housing of VP3 door



1



4



041-021328-00



VP3 door pivot



1



Parts List of Door Assembly



6-3



6.4 Front Housing Assembly 6.4.1 Explosive View



Explosive View of Front Housing Assembly



6.4.2 Parts List Serial No.



Sub-part code



Sub-part description



1



043-006443-00



VP3 front housing



1



2



043-006589-00



VP1 rear housing



1



3



020-000022-00



Loudspeaker



1



4



\



PCBA assembly



1



5



009-005810-00



Electronic wire 4P-4P direct-current output line 100mm



1



6



115-036537-00



VP3 Liquid check clip assembly



1



7



009-006431-00



VP3 motor connecting line



1



8



115-036538-00



VP3 pump assembly



1



9



115-036536-00



VP3 pressure sensor assembly



1



10



043-006516-00



VP3 waterproof sheet



1



11



051-002273-00



VP3 motor detection board



1



12



115-039031-00



VP3 ultrasound assembly



1



Parts List of Front Housing Assembly 6-4



Quantity



6.5 Rear Housing Assembly 6.5.1 Explosive View



Explosive View of Rear Housing Assembly



6.5.2 Parts List Serial No.



Sub-part code



Sub-part description



Quantity



1



047-016081-00



Insulating sheet of VP3 power module



1



2



022-000247-00



Power module



1



3



030-000262-00



Cross recessed pan head flat-nose self-tapping screw GB845-1976 F-type ST3x6 plated with environmental-friendly color zinc



2



4



022-000246-00



Battery: 2600mAh



1



5



043-006456-00



VP3 rear housing



1



6



049-000499-00



AC socket sealing gasket



1



7



009-006435-00



AC power outlet line (30mm)



1



8



043-005933-00



AC socket waterproof cover



1



9



M04-000305---



Cross recessed pan head flat-nose self-tapping screw- type PT3X12 plated with environmental-friendly color zinc



2



10



048-005294-00



DB9 waterproof sheet



1



6-5



11



009-005808-00



Electronic wire, DB9 connceting line



1



12



115-039032-00



VP3 dripping assembly



1



Parts List of Rear Housing Assembly



6.6 PCBA Assembly 6.6.1 Explosive View



Explosive View of PCBA Assembly



6.6.2 Parts List Serial No.



Sub-part code



Sub-part description



Quantity



1



030-000226-00



Cross recessed countersunk head screws GB/T818-1976 -type M3x12 plated with environmental-friendly color zinc



2



051-002282-01



VP3 control board



1



3



051-002272-02



VP3 key board



1



4



021-000207-00



LCD STN 240*128 LED white backlight (VP3)



1



5



021-000206-00



LCD STN 132*32 LED white backlight (VP1)



1



Parts List of PCBA Assembly



6-6



2



6.7 Pump Assembly 6.7.1 Explosive View



Explosive View of Pump Assembly



6.7.2 Parts List Serial No.



Sub-part code



Sub-part description



Quantity



1



024-000764-00



stepping motor



1



2



034-000069-00



Motor cushion



1



3



030-000226-00



Cross recessed countersunk head screws GB/T818-1976 -type M3x12plated with environmental-friendly color zinc



4



4



\



Sub-component of pump assembly



1



5



043-004193-00



Motor gear



1



Parts List of Pump Assembly



6-7



7 Upgrading 7.1 Tools During upgrading, you may need the following tools:  Multi-functional serial line  PC  PC burning tool software



7.2 Software Upgrading This infusion pump supports software upgrading. The burning software can be directly run on a desktop PC or mobile PC. The complete machine software can be upgraded by connecting this infusion pump to the PC via multi-functional interface. (Upgrading of drug library and built-in brands of infusion set will be available later).



7.3 Software Burning Method 1. Insert the multifunction serial line to the multifunction interface on the host as shown in the figure. Multifunctiona interface



Multifunction serial line



7-1



2. The other end of the multifunction serial line is a standard DB9 serial port. Connect this end to the PC as shown in the figure.



3. Press



and



simultaneously, and the machine will enter the "Boot"



interface as shown in the figure.



4. Open the PC burning tool, click the "Administrator Login" to pop up the longin password dialog box, and enter the administrator password to log in.



"Upgrade" menu is now available.



7-2



5. Select "Upgrade" and an interface as shown in the figure appears. Click here to import software.



6. Select "Browse" on the "Upgrade" interface, and choose the software package to be burnt, as shown in the figure. Software package saving path



7. Click "Upgrade" at the upper right corner of the PC burning tool software and the software starts burning. 8. Upon completion of burning, the upgrade status displays "Succeeded". Now, the software burning is completed.



7-3