EPX-SERIES Diagnostic X-ray System



Service Manual



Ecotron Co., Ltd Made by Korea 4F&5F, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul 08511, Korea Obelis S.A Bd. General Wahis 53, 1030 Brussels, Belgium / E-mail:[email protected] Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03



D/N: ECO-SM-EPX-001 Rev. G



EPX-SERIES SERVICE MANUAL



(This page intentionally left blank)



Rev. F



2/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



CONTENTS 1. CALIBRATION ________________________________________________________ 8 1.1 EPX-F1600 ____________________________________________________________________________________ 8 1.1.1 LINE POWER SWITCH ___________________________________________________________________ 8 1.1.2 CALIBRATION PROCEDURE ____________________________________________________________ 8 1.2 EPX-F2400 __________________________________________________________________________________ 12 1.2.1 LINE POWER SWITCH _________________________________________________________________ 12 1.2.2 CALIBRATION PROCEDURE __________________________________________________________ 12 1.3 EPX-F2800 __________________________________________________________________________________ 16 1.3.1 LINE POWER SWITCH _________________________________________________________________ 16 1.3.2 CALIBRATION PROCEDURE __________________________________________________________ 16 1.4 EPX-F3200 __________________________________________________________________________________ 20 1.4.1 LINE POWER SWITCH _________________________________________________________________ 20 1.4.2 CALIBRATION PROCEDURE __________________________________________________________ 20 1.5 EPX-F4000 __________________________________________________________________________________ 24 1.5.1 LINE POWER SWITCH _________________________________________________________________ 24 1.5.2 CALIBRATION PROCEDURE __________________________________________________________ 24 1.6 EPX-F5000 __________________________________________________________________________________ 28 1.6.1 LINE POWER SWITCH _________________________________________________________________ 28 1.6.2 CALIBRATION PROCEDURE __________________________________________________________ 28



2. DIP SWITCH SETTING _________________________________________________ 32 3. HARD-WIRED HAND SWITCH CONTROL _________________________________ 33 3.1 HAND SWITCH INTRODUCTION___________________________________________________________ 33 3.2 EXPOSURE PROCESS _____________________________________________________________________ 33



4. COLLIMATOR ________________________________________________________ 34



Rev. F



3/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



5. TROUBLE SHOOTING _________________________________________________ 36 5.1 EPX-F1600 / F2400 / F2800 / F3200 / F4000 / F5000 ERROR CODE TABLE ______________ 36 5.2 ACTION TO CORRECT THE ERROR OF EPX-SERIES ____________________________________ 37



6. CALIBRATION _______________________________________________________ 38 6.1 EPX-F1600 __________________________________________________________________________________ 38 6.1.1 VOLTAGE CHECK ______________________________________________________________________ 38 6.1.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 39 6.1.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 41 6.1.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 42 6.1.5 mA CALIBRATION DATA CHECK _____________________________________________________ 43 6.1.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 44 6.2 EPX-F2400 __________________________________________________________________________________ 46 6.2.1 VOLTAGE CHECK ______________________________________________________________________ 46 6.2.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 47 6.2.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 49 6.2.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 50 6.2.5 mA CALIBRATION DATA CHECK _____________________________________________________ 51 6.2.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 52 6.3 EPX-F2800 __________________________________________________________________________________ 54 6.3.1 VOLTAGE CHECK ______________________________________________________________________ 54 6.3.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 55 6.3.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 57 6.3.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 58 6.3.5 mA CALIBRATION DATA CHECK _____________________________________________________ 59 6.3.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 60



Rev. F



4/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.4 EPX-F3200 __________________________________________________________________________________ 62 6.4.1 VOLTAGE CHECK ______________________________________________________________________ 62 6.4.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 63 6.4.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 65 6.4.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 66 6.4.5 mA CALIBRATION DATA CHECK _____________________________________________________ 67 6.4.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 68 6.5 EPX-F4000 __________________________________________________________________________________ 70 6.5.1 VOLTAGE CHECK ______________________________________________________________________ 70 6.5.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 71 6.5.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 73 6.5.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 74 6.5.5 mA CALIBRATION DATA CHECK _____________________________________________________ 75 6.5.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 77 6.6 EPX-F5000 __________________________________________________________________________________ 79 6.6.1 VOLTAGE CHECK ______________________________________________________________________ 79 6.6.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 80 6.6.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 82 6.6.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 83 6.6.5 mA CALIBRATION DATA CHECK _____________________________________________________ 84 6.6.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 86



Rev. F



5/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



REVSON HISTORY Revision Number



Date



Description



A



SEP 11, 2013



First Edition



B



SEP 30, 2014



European Authorized Representative Change



C



JAN 12, 2015



Manual Updates



D



MAR 23, 2015



Manual Updates



E



JUN 11, 2015



Manual Updates



F



AUG 23, 2018



Manual Update



G



June 17, 2019



Modify misprinted contents



ADVISORY SYMBOLS The following advisory symbols are used throughout this manual. Their application and meaning are described below.



WARNING Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death or radiation exposure.



CAUTION Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of equipment.



NOTE Note symbol used to indicate important information need to proper use and right operation of equipment.



NOTE KEEP THIS SERVICE MANUAL WITH THE EQUIPMENT AT ALL TIMES, AND REVIEW THE IMPORTANT INFORMATION WHENEVER REQUIRED. THE STATISTICS AND SPECIFICATIONS OF THIS UNIT AND MANUAL CAN BE MODIFIED WITH OR WITHOUT NOTIFICATION FOR THE IMPROVEMENT OF PERFORMANCE AND SAFETY.



Rev. F



6/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



SERVICE CHAPTER This manual is designed to ensure correct use and operation of EPX model diagnostic x-ray units. Please read all the lines thoroughly before you use the equipment.



Incorrect use and operation exceeding described conditions in this manual may cause damage of the equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. This equipment should be used only by the legally qualified persons and practitioners.



Regarding the AC cable and software, only the ones developed and supplied by Ecotron Co., Ltd should be used.



Only trained maintenance staff may remove the covers of the x-ray system.



Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask the help of Ecotron or its authorized dealer for the service.



Information for repair will be provided by Ecotron. (Circuit diagram, Component part list, descriptions, Calibration instructions, or other information)



No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties which many results from the use of this manual.



Always keep the manual at hand for your reference.



NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following rules and methods described in this manual.



CAUTION



THE IONIZING RADIATION IS DANGEROUS FOR THE OPERATOR IF THE FOLLOWING SAFETY MEASURES ARE NOT STRICTLY OBSERVED.



NOTE ASSEMBLY INFORMATION: NO ASSEMBLY IS REQUIRED.



Rev. F



7/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1. CALIBRATION 1.1 EPX-F1600 1.1.1 LINE POWER SWITCH - Line switch should be at off position. - Connect power line cord to the power outlet. - Turn on the line switch. - The FND on the control panel should illuminate.



1.1.2 CALIBRATION PROCEDURE 1.1.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40 KV and 5 mAs (30mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.



j. Set the kV and the mAs at 90 KV and 5 mAs (16 mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 4.5 V, adjust KV by turning VR3 on the OP board.



Rev. F



8/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.1.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40 kV and 5 mAs (30 mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.



Then the display should indicate as follows,



Where, C00~C09 are calibration code of tube current at AC110V model and C10~C19 are calibration code of tube current at AC220V model. C00 or C10 is calibration code of tube current (30mA at 40kV~50kV) at AC110V or AC220V and 30.0 is calibration data of tube current (30mA at 40kV~50kV).



Rev. F



9/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec (see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,



Where, C01 or C11 is calibration code of tube current (25mA at 40kV~50kV) and 25.0 is calibration data of tube current (25mA at 40kV~50kV).



i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec (see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 5V, set the calibration data again.



Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,



Rev. F



10/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Where, P00~P07 are calibration code of filament preheat current at AC110V, P08~P15 are calibration code of filament preheat current at AC220V, P00 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA), and 1.50 or 1.40 is Filament preheat calibration data (Preheat calibration data of 30mA).



g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,



Where, P01 or P09 is Filament preheat calibration code (Preheat calibration code of 25mA) and 1.45 or 1.35 is Filament preheat calibration data (Preheat calibration data of 25mA).



m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.



Rev. F



11/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.2 EPX-F2400 1.2.1 LINE POWER SWITCH - Line switch should be at off position. - Connect power line cord to the power outlet. - Turn on the line switch. - The FND on the control panel should illuminate.



1.2.2 CALIBRATION PROCEDURE 1.2.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40 KV and 5 mAs (40mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.



j. Set the kV and the mAs at 100 KV and 5 mAs (20 mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5 V, adjust KV by turning VR3 on the OP board.



Rev. F



12/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.2.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40 kV and 5 mAs (40 mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.



Then the display should indicate as follows,



Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration code of tube current at AC220V model. C00 or C08 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0 is calibration data of tube current (40mA at 40kV~60kV).



Rev. F



13/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec (see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,



Where, C01 or C09 is calibration code of tube current (35mA at 40kV~60kV) and 35.0 is calibration data of tube current (35mA at 40kV~60kV).



i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec (see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 5.25V, set the calibration data again.



Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition.



a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,



Rev. F



14/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Where, P00~P04 are calibration code of filament preheat current at AC110V, P05~P09 are calibration code of filament preheat current at AC220V, P00 or P05 is Filament preheat calibration code (Preheat calibration code of 40mA), and 0.70 or 0.72 is Filament preheat calibration data (Preheat calibration data of 40mA).



g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec (see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,



Where, P01 or P06 is Filament preheat calibration code (Preheat calibration code of 35mA) and 0.67 or 0.69 is Filament preheat calibration data (Preheat calibration data of 35mA).



m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec (see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.



Rev. F



15/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.3 EPX-F2800 1.3.1 LINE POWER SWITCH - Line switch should be at off position. - Connect power line cord to the power outlet. - Turn on the line switch. - The FND on the control panel should illuminate.



1.3.2 CALIBRATION PROCEDURE 1.3.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs (40mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.



j. Set the kV and the mAs at 120 KV and 5 mAs (16 mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 6 V, adjust KV by turning VR3 on the OP board.



Rev. F



16/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.3.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40 kV and 5 mAs (40 mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.



Then the display should indicate as follows,



Where, C00~C15 are calibration code of tube current at AC110V model and C16~C31 are calibration code of tube current at AC220V model. C00 or C16 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0 is calibration data of tube current (40mA at 40kV~60kV).



Rev. F



17/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec (see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,



Where, C01 or C17 is calibration code of tube current (30mA at 61kV~70kV) and 30.0 is calibration data of tube current (30mA at 61kV~70kV).



i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec (see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 5.15V, set the calibration data again.



Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition.



a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,



Rev. F



18/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



where, P00~P06 are calibration code of filament preheat current at AC110V, P07~P13 are calibration code of filament preheat current at AC220V, P00 or P07 is Filament preheat calibration code (Preheat calibration code of 35mA), and 0.70 or 0.70 is Filament preheat calibration data (Preheat calibration data of 35mA).



g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,



Where, P01 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA) and 0.67 or 0.67 is Filament preheat calibration data (Preheat calibration data of 30mA).



m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.



Rev. F



19/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.4 EPX-F3200 1.4.1 LINE POWER SWITCH - Line switch should be at off position. - Connect power line cord to the power outlet. - Turn on the line switch. - The FND on the control panel should illuminate.



1.4.2 CALIBRATION PROCEDURE 1.4.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs (60mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.



j. Set the kV and the mAs at 100KV and 5mAs(30mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5V, adjust KV by turning VR3 on the OP board.



Rev. F



20/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.4.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40 kV and 5 mAs (60 mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.



Then the display should indicate as follows,



Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration code of tube current at AC220V model. C00 or C08 is calibration code of tube current (60mA at 40kV~50kV) at AC110V and 60.0 is calibration data of tube current (60mA at 40kV~50kV).



Rev. F



21/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec (see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6.0V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,



Where, C01 or C09 is calibration code of tube current (50mA at 51kV~60kV) and 50.0 is calibration data of tube current (50mA at 51kV~60kV).



i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec (see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 5.0V, set the calibration data again.



Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition.



a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,



Rev. F



22/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Where, P00~P05 are calibration code of filament preheat current at AC110V, P06~P11 are calibration code of filament preheat current at AC220V, P00 or P06 is Filament preheat calibration code (Preheat calibration code of 60mA), and 1.65 or 1.65 is Filament preheat calibration data (Preheat calibration data of 60mA).



g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,



Where, P01 or P07 is Filament preheat calibration code (Preheat calibration code of 50mA) and 1.60 or 1.60 is Filament preheat calibration data (Preheat calibration data of 50mA).



m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.



Rev. F



23/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.5 EPX-F4000 1.5.1 LINE POWER SWITCH - Line switch should be at off position. - Connect power line cord to the power outlet. - Turn on the line switch. - The FND on the control panel should illuminate.



1.5.2 CALIBRATION PROCEDURE 1.5.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs (80mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2.5V, adjust KV by turning VR3 on the OP board.



j. Set the kV and the mAs at 110KV and 5mAs (25mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.



Rev. F



24/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.5.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 50 kV and 5 mAs (80 mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.44V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.



Then the display should indicate as follows,



Where, C00~C14 are calibration code of tube current at AC110V model and C15~C29 are calibration code of tube current at AC220V model. C00 or C15 is calibration code of tube current (80mA at 50kV) at 80.0 is calibration data of tube current (80mA at 50kV).



Rev. F



25/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 4V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,



Where, C01 or C16 is calibration code of tube current (70mA at 50kV) and 70.0 is calibration data of tube current (70mA at 50kV).



i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec (see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 3.0V, set the calibration data again.



Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition.



a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,



Rev. F



26/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Where, P00~P09 are calibration code of filament preheat current at AC110V, P10~P19 are calibration code of filament preheat current at AC220V, P00 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA), and 2.50 or 2.50 is Filament preheat calibration data (Preheat calibration data of 80mA).



g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec (see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.



j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,



Where, P01 or P11 is Filament preheat calibration code (Preheat calibration code of 70mA) and 2.20 or 2.20 is Filament preheat calibration data (Preheat calibration data of 70mA).



m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.



Rev. F



27/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.6 EPX-F5000 1.6.1 LINE POWER SWITCH - Line switch should be at off position. - Connect power line cord to the power outlet. - Turn on the line switch. - The FND on the control panel should illuminate.



1.6.2 CALIBRATION PROCEDURE 1.6.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 50KV and 5mAs (100mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2.0V, adjust KV by turning VR3 on the OP board.



j. Set the kV and the mAs at 110KV and 5mAs (40mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.



Rev. F



28/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



1.6.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 50 kV and 5 mAs (100 mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.



Then the display should indicate as follows,



Where, C00~C20 are calibration code of tube current at AC110V model and C21~C41 are calibration code of tube current at AC220V model. C00 or C21 is calibration code of tube current (100mA at 50kV) at A0.0(=100) is calibration data of tube current (100mA at 50kV).



Rev. F



29/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec (see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,



Where, C01 or C22 is calibration code of tube current (80mA at 50kV) and 80.0 is calibration data of tube current (80mA at 50kV).



i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.08V, set the calibration data again.



Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition.



a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,



Rev. F



30/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Where, P00~P08 are calibration code of filament preheat current at AC110V, P09~P15 are calibration code of filament preheat current at AC220V, P00 or P09 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.40 or 2.40 is Filament preheat calibration data (Preheat calibration data of 100mA).



g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,



Where, P01 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.25 or 2.25 is Filament preheat calibration data (Preheat calibration data of 80mA).



m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec (see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.



Rev. F



31/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



2. DIP SWITCH SETTING



DIP Switches are used for following setting.



SW #1: ERROR Disable (ON) / Enable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to OFF position.



Rev. F



32/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



3. HARD-WIRED HAND SWITCH CONTROL 3.1 HAND SWITCH INTRODUCTION The Exposure Hand Switch consists of a two step hand switch. The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage. The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.



3.2 EXPOSURE PROCESS There are two ways to use the hand switch to get the exposure. a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or Ready” status on, then press Switch #2 to get the “Exposure”.



b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1 second. This action will result in “Prep or Ready” status and “Exposure” at the same time.



Rev. F



33/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



4. COLLIMATOR a. Halogen Lamp * 1 b. Laser pointer * 2 c.



Horizontal and vertical lighting range control



I. Block Diagram



II. Layout a. External Layout



Rev. F



34/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



b. Internal Layout



c. Knob Layout



Rev. F



35/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



5. TROUBLE SHOOTING If you have any error message when you are using portable x-ray, please check below service check list



5.1 EPX-F1600 / F2400 / F2800 / F3200 / F4000 / F5000 ERROR CODE TABLE Error Code



DESCRIPTION



INDICATION



2



THERMAL ERROR (INTERLOCK)



Excess from 60℃



7



READY SW ERR



Ready Switch Malfunction



8



EXPOSURE SW ERROR



Exposure Switch Malfunction



10



STANDBY FIL FB ERROR



Excess from 200mA (10% of Max. Current, 2A)



11



STANDBY EP FB ERROR



Excess from 12kV (10% of Max. Voltage, 120kV)



12



STANDBY IP FB ERROR



Excess from 3.5mA (10% of Max. Current, 35mA)



14



RUN FIL FB ERROR



Out of Range from Reference ±10%



15



RUN EP FB or AC POWER LOW ERROR



Out of Range from Reference -10% or AC Power Line poor



16



RUN IP FB HIGH ERROR



Out of Range from Reference +10%



17



RUN EP FB HIGH ERROR



Out of Range from Reference +10%



18



RUN IP FB LOW ERROR



Out of Range from Reference -10%



28



PULSE BY PULSE ERROR



FET Board Fault (Hardware)



29



FET-OTP ERROR



FET board overheat



30



KV_OVP ERROR



kV Over Voltage (Hardware)



31



FIL_OC ERROR



FIL Over Current (Hardware)



32



IP_OCP ERROR



mA Over Current (Hardware)



Rev. F



36/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



5.2 ACTION TO CORRECT THE ERROR OF EPX-SERIES Error Codes 7



Check points and Action



Ready SW ERR a. Replace the hand switch



8



Exposure SW ERR



10



Standby Fil FB Error



11



Standby EP FB Error



12



Standby IP FB Error



14



Run Fil FB Error



15



Run EP FB low ERR



16



Run IP FB High ERR



a. If error repeats at the condition of 40kV 0.4mAs EXP, replace the OP board



a. If error repeats at the condition of 40kV 0.4mAs EXP, check the followings by sequence. 1. Check TP in the power input of the OP board 2. Check the Reference voltage value 17



Run EP FB High ERR



18



Run IP FB low ERR



28



Pulse by Pulse ERR



30



KV_OVP ERR



31



FIL_OC ERR FIL



3. If the reference voltage is normal, replace the FET board 4. If replacing FET board does not solve the problem, the replace the HV tank.



a. If error pops up right after the unit power is turned on, then replace the OP board. b. If error repeats at the condition of 40kV 0.4mAs EXP, 1. Replace the FET board



32



IP_OCP ERR



2. If replacing FET board does not solve the problem, the replace the HV tank.



Warning Codes 2



29



Thermal ERR excess



Wait until the temperature inside the monotank comes down to normal



from 60°C



(Once the temperature drop to normal, the Warning Message will disappear)



FET-OTP FET board



Wait until the temperature inside the FET board comes down to normal



overheat



(Once the temperature drop to normal, the Warning Message will disappear)



Rev. F



37/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6. CALIBRATION 6.1 EPX-F1600 Please fill in the blank.



Kv / mAs condition (When error is occurred)



Product S/N



Error code



※ Open upper cover of portable x-ray below picture.



6.1.1 VOLTAGE CHECK 1) Turn on AC switch



VDC Range



TP26



TP24



TP25



TP6



4.8V~5.0V



11.5V~12.5V



-11.5V~-12.5V



+3V~3.6V



Pass



Fail



Pass



Fail



Pass



Fail



Pass



Fail



Pass/Fail Measured Voltage If it is fail case



Rev. F



38/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.1.2 kV, mA WAVEFORM CHECK 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.



4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



40msec/div



Setting value



EP_FB



IP_FB



Ref voltage 40kV



2V



5.1V



Pass/Fail



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV, mA waveforms are fail, please attach your kV, mA waveform



Rev. F



39/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL 5) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 80kV Pass/Fail



EP_FB



IP_FB



4V



3.06V



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform 6) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 90kV Pass/Fail



Pass



EP_FB



IP_FB



4.5V



2.72V Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms



Rev. F



are fail, please attach your kV, mA waveform



40/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.1.3 WAVEFORM OF POOR AC POWER CONDITION If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.



Rev. F



41/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.1.4 kV, mA PWM FREQUENCY CHECK 1) Turn off AC swtich 2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture. 3) Turn on AC switch.



4) kV / mA PWM frequency CH 1: kV



2V/div



CH



CH 2 : mA



2V/div



CH 1 mA PWM



CH 2 kV PWM



80kHz



140kHz



Frequency Pass/Fail



4μsec/div



Pass



Fail



Pass



Fail



Measured frequency



Rev. F



42/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.1.5 mA CALIBRATION DATA CHECK



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



1) Turn on AC switch.



Up/Down switches increase and decrease C00~C09 (AC110V), C10~C19 (AC220V). Please check Input data. Input data and Tube current should be the same with below table.



Rev. F



43/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Tube Current Parameter



Tube Voltage



mAs



Tube Current



INPUT DATA



AC110V



AC220V



KV



MIN



MAX



mA



mA



C00



C10



40~50 (1)



0.4



20



30.0



30.0



C01



C11



40~50 (2)



25



32



25.0



25.0



C02



C12



51~60 (1)



0.4



20



25.0



25.0



C03



C13



51~60 (2)



25



32



20.0



20.0



C04



C14



61~70 (1)



0.4



20



20.0



20.0



C05



C15



61~70 (2)



25



32



15.0



15.0



C06



C16



71~80 (1)



0.4



20



18



18



C07



C17



71~80 (2)



25



32



14



14



C08



C18



81~90 (1)



0.4



20



16



16



C09



C19



81~90 (2)



25



32



12



12



4) If mA calibration is done, Please dip switch 4 set to OFF position.



6.1.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK 1) Turn off AC switch 2) Dip switch on op board



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



Rev. F



44/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.



Up/Down switches increase and decrease P00~P07 (AC110V), P08~P15 (AC220V) Please check Input data. Input data and Tube current should be the same.



Fill in the blank below table because P_DATA is different for each product.



Model



1.6kW



Current(mA)



P_Mode (110V)



P_Mode (220V)



30



P00



P08



25



P01



P09



20



P02



P10



18



P03



P11



16



P04



P12



15



P05



P13



14



P06



P14



12



P07



P15



P_DATA



5) If Preheat calibration is done, Please dip switch 4 set to OFF position.



Rev. F



45/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.2 EPX-F2400 Please fill in the blank.



Kv / mAs condition (When error is occurred)



Product S/N



Error code



※ Open upper cover of portable x-ray below picture.



6.2.1 VOLTAGE CHECK 1) Turn on AC switch



VDC Range



TP26



TP24



TP25



TP6



4.8V~5.0V



11.5V~12.5V



-11.5V~-12.5V



+3V~3.6V



Pass



Fail



Pass



Fail



Pass



Fail



Pass



Fail



Pass/Fail Measured Voltage If it is fail case



Rev. F



46/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.2.2 kV, mA WAVEFORM CHECK 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.



4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



40msec/div



Setting value



EP_FB



IP_FB



Ref voltage 40kV



2V



6V



Pass/Fail



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV, mA waveforms are fail, please attach your kV, mA waveform



Rev. F



47/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL 5) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 80kV Pass/Fail



EP_FB



IP_FB



4V



4.5V



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform 6) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 100 kV Pass/Fail



Pass



EP_FB



IP_FB



5V



3V Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms



Rev. F



are fail, please attach your kV, mA waveform



48/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.2.3 WAVEFORM OF POOR AC POWER CONDITION If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.



Rev. F



49/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.2.4 kV, mA PWM FREQUENCY CHECK 1) Turn off AC swtich 2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture. 3) Turn on AC switch.



4) kV / mA PWM frequency CH 1: kV



2V/div



CH



CH 2 : mA



2V/div



CH 1 mA PWM



CH 2 kV PWM



80kHz



140kHz



Frequency Pass/Fail



4μsec/div



Pass



Fail



Pass



Fail



Measured frequency



Rev. F



50/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.2.5 mA CALIBRATION DATA CHECK



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



1) Turn on AC switch.



Up/Down switches increase and decrease C00~C07 (AC110V), C08~C15 (AC220V). Please check Input data. Input data and Tube current should be the same with below table.



Rev. F



51/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Tube Current Parameter



Tube Voltage



mAs



Tube Current



INPUT DATA



AC110V



AC220V



KV



MIN



MAX



mA



mA



C00



C08



40~60(1)



0.4



64



40.0



40.0



C01



C09



40~60(2)



80



100



35.0



30.0



C02



C10



61~70(1)



0.4



100



30.0



25.0



C03



C11



71~80(1)



0.4



32



30.0



30.0



C04



C12



71~80(2)



40



80



25.0



25.0



C05



C13



81~90(1)



0.4



40



25.0



25.0



C06



C14



81~90(2)



50



80



20.0



20.0



C07



C15



91~100(1)



0.4



80



20.0



20.0



4) If mA calibration is done, Please dip switch 4 set to OFF position.



6.2.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK 1) Turn off AC switch 2) Dip switch on op board



DIP Switches are used for following setting: SW #1: ERROR Enable (ON )/ Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



Rev. F



52/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.



Up/Down switches increase and decrease P00~P04 (AC110V), P05~P09 (AC220V) Please check Input data. Input data and Tube current should be the same.



Fill in the blank below table because P_DATA is different for each product.



Model



2.4 kW



Current(mA)



P_Mode (110V)



P_Mode (220V)



40



P00



P05



35



P01



P06



30



P02



P07



25



P03



P08



20



P04



P09



P_DATA



5) If Preheat calibration is done, Please dip switch 4 set to OFF position.



Rev. F



53/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.3 EPX-F2800 Please fill in the blank.



Kv / mAs condition (When error is occurred)



Product S/N



Error code



※ Open upper cover of portable x-ray below picture.



6.3.1 VOLTAGE CHECK 1) Turn on AC switch



VDC Range



TP26



TP24



TP25



TP6



4.8V~5.0V



11.5V~12.5V



-11.5V~-12.5V



+3V~3.6V



Pass



Fail



Pass



Fail



Pass



Fail



Pass



Fail



Pass/Fail Measured Voltage If it is fail case



Rev. F



54/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.3.2 kV, mA WAVEFORM CHECK 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.



4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



40msec/div



Setting value



EP_FB



IP_FB



Ref voltage 40kV



2V



6V



Pass/Fail



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV, mA waveforms are fail, please attach your kV, mA waveform



Rev. F



55/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL 5) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 80kV Pass/Fail



EP_FB



IP_FB



4V



5.25V



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform 6) 120kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 120kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 120 kV Pass/Fail



Pass



EP_FB



IP_FB



6V



2.4V Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms



Rev. F



are fail, please attach your kV, mA waveform



56/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.3.3 WAVEFORM OF POOR AC POWER CONDITION If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.



Rev. F



57/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.3.4 kV, mA PWM FREQUENCY CHECK 1) Turn off AC swtich 2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture. 3) Turn on AC switch.



4) kV / mA PWM frequency CH 1: kV



2V/div



CH



CH 2 : mA



2V/div



CH 1 mA PWM



CH 2 kV PWM



80kHz



140kHz



Frequency Pass/Fail



4μsec/div



Pass



Fail



Pass



Fail



Measured frequency



Rev. F



58/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.3.5 mA CALIBRATION DATA CHECK



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



1) Turn on AC switch.



Up/Down switches increase and decrease C00~C15 (AC110V), C16~C31 (AC220V). Please check Input data. Input data and Tube current should be the same with below table.



Rev. F



59/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Tube Current Parameter



Tube Voltage



mAs



Tube Current



INPUT DATA



AC110V



AC220V



KV



MIN



MAX



mA



mA



C00



C16



40~60



0.4



100



40.0



40.0



C01



C17



61~70(1)



0.4



64



30.0



30.0



C02



C18



61~70(2)



80



100



25.0



25.0



C03



C19



71~80(1)



0.4



10



35.0



35.0



C04



C20



71~80(2)



13



32



30.0



30.0



C05



C21



71~80(3)



40



80



25.0



25.0



C06



C22



81~90(1)



0.4



10



30.0



30.0



C07



C23



81~90(2)



13



40



25.0



25.0



C08



C24



81~90(3)



50



80



20.0



20.0



C09



C25



91~100(1)



0.4



13



25.0



25.0



C10



C26



91~100(2)



16



50



20.0



20.0



C11



C27



91~100(3)



64



80



16.0



16.0



C12



C28



101~110(1)



0.4



32



18.0



18.0



C13



C29



101~110(2)



40



64



16.0



16.0



C14



C30



111~120(1)



0.4



20



16.0



16.0



C15



C31



111~120(2)



25



50



14.0



14.0



4) If mA calibration is done, Please dip switch 4 set to OFF position.



6.3.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK 1) Turn off AC switch 2) Dip switch on op board



Rev. F



60/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL DIP Switches are used for following setting: SW #1: ERROR Enable (ON )/ Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.



Up/Down switches increase and decrease P00~P06 (AC110V), P07~P13 (AC220V) Please check Input data. Input data and Tube current should be the same. Fill in the blank below table because P_DATA is different for each product. Model



2.8 kW



Current(mA)



P_Mode (110V)



P_Mode (220V)



40



P00



P08



35



P01



P09



30



P02



P10



25



P03



P11



20



P04



P12



18



P05



P13



16



P06



P14



14



P07



P15



P_DATA



5) If Preheat calibration is done, Please dip switch 4 set to OFF position.



Rev. F



61/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.4 EPX-F3200 Please fill in the blank.



Kv / mAs condition (When error is occurred)



Product S/N



Error code



※ Open upper cover of portable x-ray below picture.



6.4.1 VOLTAGE CHECK 1) Turn on AC switch



VDC Range



TP26



TP24



TP25



TP6



4.8V~5.0V



11.5V~12.5V



-11.5V~-12.5V



+3V~3.6V



Pass



Fail



Pass



Fail



Pass



Fail



Pass



Fail



Pass/Fail Measured Voltage If it is fail case



Rev. F



62/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.4.2 kV, mA WAVEFORM CHECK 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.



4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



40msec/div



Setting value



EP_FB



IP_FB



Ref voltage 40kV



2V



6V



Pass/Fail



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV, mA waveforms are fail, please attach your kV, mA waveform



Rev. F



63/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL 5) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 80kV Pass/Fail



EP_FB



IP_FB



4V



4V



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform 6) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 100 kV Pass/Fail



Pass



EP_FB



IP_FB



5V



3V Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms



Rev. F



are fail, please attach your kV, mA waveform



64/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.4.3 WAVEFORM OF POOR AC POWER CONDITION If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.



Rev. F



65/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.4.4 kV, mA PWM FREQUENCY CHECK 1) Turn off AC swtich 2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture. 3) Turn on AC switch.



4) kV / mA PWM frequency CH 1: kV



2V/div



CH



CH 2 : mA



2V/div



CH 1 mA PWM



CH 2 kV PWM



80kHz



140kHz



Frequency Pass/Fail



4μsec/div



Pass



Fail



Pass



Fail



Measured frequency



Rev. F



66/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.4.5 mA CALIBRATION DATA CHECK



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



1) Turn on AC switch.



Up/Down switches increase and decrease C00~C07 (AC110V), C08~C15 (AC220V). Please check Input data. Input data and Tube current should be the same with below table.



Rev. F



67/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Tube Current Parameter



Tube Voltage



mAs



Tube Current



INPUT DATA



AC110V



AC220V



KV



MIN



MAX



mA



mA



C00



C08



40~50



0.4



100



60.0



60.0



C01



C09



51~60



0.4



80



50.0



50.0



C02



C10



61~70



0.4



100



40.0



40.0



C03



C11



71~80(1)



0.4



32



40.0



40.0



C04



C12



71~80(2)



40



80



35.0



35.0



C05



C13



81~90



0.4



80



30.0



30.0



C06



C14



91~100(1)



0.4



50



30.0



30.0



C07



C15



91~100(2)



64



80



25.0



25.0



4) If mA calibration is done, Please dip switch 4 set to OFF position.



6.4.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK 1) Turn off AC switch 2) Dip switch on op board



DIP Switches are used for following setting: SW #1: ERROR Enable (ON )/ Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



Rev. F



68/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.



Up/Down switches increase and decrease P00~P05 (AC110V), P06~P11 (AC220V) Please check Input data. Input data and Tube current should be the same.



Fill in the blank below table because P_DATA is different for each product.



Model



3.2 kW



Current(mA)



P_Mode (110V)



P_Mode (220V)



60



P00



P06



50



P01



P07



40



P02



P08



35



P03



P09



30



P04



P10



25



P05



P11



P_DATA



5) If Preheat calibration is done, Please dip switch 4 set to OFF position.



Rev. F



69/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.5 EPX-F4000 Please fill in the blank.



Kv / mAs condition (When error is occurred)



Product S/N



Error code



※ Open upper cover of portable x-ray below picture.



6.5.1 VOLTAGE CHECK 1) Turn on AC switch



VDC Range



TP26



TP24



TP25



TP6



4.8V~5.0V



11.5V~12.5V



-11.5V~-12.5V



+3V~3.6V



Pass



Fail



Pass



Fail



Pass



Fail



Pass



Fail



Pass/Fail Measured Voltage If it is fail case



Rev. F



70/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.5.2 kV, mA WAVEFORM CHECK 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.



4) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



40msec/div



Setting value



EP_FB



IP_FB



Ref voltage 50kV



2.5V



5.44V



Pass/Fail



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV, mA waveforms are fail, please attach your kV, mA waveform



Rev. F



71/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL 5) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 80kV Pass/Fail



EP_FB



IP_FB



4V



3.4V



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform 6) 110kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 110kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 110 kV Pass/Fail



Pass



EP_FB



IP_FB



5.5V



2.04V Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms



Rev. F



are fail, please attach your kV, mA waveform



72/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.5.3 WAVEFORM OF POOR AC POWER CONDITION If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.



Rev. F



73/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.5.4 kV, mA PWM FREQUENCY CHECK 1) Turn off AC swtich 2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture. 3) Turn on AC switch.



4) kV / mA PWM frequency CH 1: kV



2V/div



CH



CH 2 : mA



2V/div



CH 1 mA PWM



CH 2 kV PWM



80kHz



140kHz



Frequency Pass/Fail



4μsec/div



Pass



Fail



Pass



Fail



Measured frequency



Rev. F



74/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.5.5 mA CALIBRATION DATA CHECK



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



1) Turn on AC switch.



Up/Down switches increase and decrease C00~C14 (AC110V), C15~C29 (AC220V). Please check Input data. Input data and Tube current should be the same with below table.



Rev. F



75/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Tube Current Parameter



Tube Voltage



mAs



Tube Current



INPUT DATA



AC 110V



AC 220V



KV



MIN



MAX



mA



mA



C00



C15



50 (1)



0.4



16



80.0



80.0



C01



C16



50 (2)



20



32



70.0



70.0



C02



C17



50 (3)



40



100



60.0



60.0



C03



C18



51~60 (1)



0.4



32



60.0



60.0



C04



C19



51~60 (2)



40



100



50.0



50.0



C05



C20



61~70 (1)



0.4



32



50.0



50.0



C06



C21



61~70 (2)



40



100



40.0



40.0



C07



C22



71~80 (1)



0.4



10



45.0



45.0



C08



C23



71~80 (2)



13



80



40.0



40.0



C09



C24



81~90 (1)



0.4



32



35.0



35.0



C10



C25



81~90 (2)



40



80



30.0



30.0



C11



C26



91~100 (1)



0.4



32



30.0



30.0



C12



C27



91~100 (2)



40



80



25.0



25.0



C13



C28



101~110 (1)



0.4



32



25.0



25.0



C14



C29



101~110 (2)



40



80



20.0



20.0



4) If mA calibration is done, Please dip switch 4 set to OFF position.



Rev. F



76/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.5.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK 1) Turn off AC switch 2) Dip switch on op board



DIP Switches are used for following setting: SW #1: ERROR Enable (ON )/ Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.



Up/Down switches increase and decrease P00~P09 (AC110V), P10~P19 (AC220V) Please check Input data. Input data and Tube current should be the same.



Rev. F



77/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Fill in the blank below table because P_DATA is different for each product.



Model



Current(mA)



P_Mode(110V)



P_Mode(220V)



80



P00



P10



70



P01



P11



60



P02



P12



50



P03



P13



45



P04



P14



40



P05



P15



35



P06



P16



30



P07



P17



25



P08



P18



20



P09



P19



P_DATA



4.0KW



5) If Preheat calibration is done, Please dip switch 4 set to OFF position.



Rev. F



78/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.6 EPX-F5000 Please fill in the blank.



Kv / mAs condition (When error is occurred)



Product S/N



Error code



※ Open upper cover of portable x-ray below picture.



6.6.1 VOLTAGE CHECK 1) Turn on AC switch



VDC Range



TP26



TP24



TP25



TP6



4.8V~5.0V



11.5V~12.5V



-11.5V~-12.5V



+3V~3.6V



Pass



Fail



Pass



Fail



Pass



Fail



Pass



Fail



Pass/Fail Measured Voltage If it is fail case



Rev. F



79/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.6.2 kV, mA WAVEFORM CHECK 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.



4) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



40msec/div



Setting value



EP_FB



IP_FB



Ref voltage 40kV



2V



5.1V



Pass/Fail



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV, mA waveforms are fail, please attach your kV, mA waveform



Rev. F



80/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL 5) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 80kV Pass/Fail



EP_FB



IP_FB



4V



3.06V



Pass



Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform 6) 110kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 110kV/5mAs CH 1: kV



2V/div



CH 2: mA



2V/div



100msec/div



Setting value Ref voltage 110 kV Pass/Fail



Pass



EP_FB



IP_FB



5.5V



1.53V Fail



Pass



Fail



Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV, mA waveform



Rev. F



81/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.6.3 WAVEFORM OF POOR AC POWER CONDITION If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.



Rev. F



82/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.6.4 kV, mA PWM FREQUENCY CHECK 1) Turn off AC swtich 2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture. 3) Turn on AC switch.



4) kV / mA PWM frequency CH 1: kV



2V/div



CH



CH 2 : mA



2V/div



CH 1 mA PWM



CH 2 kV PWM



80kHz



140kHz



Frequency Pass/Fail



4μsec/div



Pass



Fail



Pass



Fail



Measured frequency



Rev. F



83/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.6.5 mA CALIBRATION DATA CHECK



DIP Switches are used for following setting: SW #1: ERROR Enable (ON) / Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



1) Turn on AC switch.



Up/Down switches increase and decrease C00~C20 (AC110V), C21~C41 (AC220V). Please check Input data. Input data and Tube current should be the same with below table.



Rev. F



84/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Tube Current Parameter



Tube Voltage



mAs



Tube Current



INPUT DATA



AC110V



AC220V



KV



MIN



MAX



mA



mA



C00



C21



50(1)



0.4



10



100



100



C01



C22



50(2)



13



32



80



80



C02



C23



50(3)



40



100



70



70



C03



C24



51~60(1)



0.4



10



80



80



C04



C25



51~60(2)



13



32



70



70



C05



C26



51~60(3)



40



100



60



60



C06



C27



61~70(1)



0.4



10



70



70



C07



C28



61~70(2)



13



32



60



60



C08



C29



61~70(3)



40



100



50



50



C09



C30



71~80(1)



0.4



10



60



60



C10



C31



71~80(2)



13



32



50



50



C11



C32



71~80(3)



40



100



40



40



C12



C33



81~90(1)



0.4



10



50



50



C13



C34



81~90(2)



13



32



40



40



C14



C35



81~90(3)



40



100



30



30



C15



C36



91~100(1)



0.4



10



45



45



C16



C37



91~100(2)



13



32



40



40



C17



C38



91~100(3)



40



100



30



30



C18



C39



101~110(1)



0.4



10



40



40



C19



C40



101~110 (2)



13



32



30



30



C20



C41



101~110 (3)



40



100



20



20



4) If mA calibration is done, Please dip switch 4 set to OFF position.



Rev. F



85/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



6.6.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK 1) Turn off AC switch 2) Dip switch on op board



DIP Switches are used for following setting: SW #1: ERROR Enable (ON )/ Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON) / Normal mode (Off)



For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,



3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.



Up/Down switches increase and decrease P00~P08 (AC110V), P09~P17 (AC220V) Please check Input data. Input data and Tube current should be the same.



Rev. F



86/87



ECOTRON Co., Ltd.



EPX-SERIES SERVICE MANUAL



Fill in the blank below table because P_DATA is different for each product.



Model



5kW



Current(mA)



P_Mode(110V)



P_Mode(220V)



100



P00



P09



80



P01



P10



70



P02



P11



60



P03



P12



50



P04



P13



45



P05



P14



40



P06



P15



30



P07



P16



20



P08



P17



P_DATA(A)



5) If Preheat calibration is done, Please dip switch 4 set to OFF position.



Rev. F



87/87