Mammography X-ray System



MX - 600 Service Manual Document No.: RD0701001 Issued Date: 30-January-2001 Revision: 1.0 The original version of MX-600 is being written in English



RD0309001



Table of Contents



Section 1.



Specification Manual



Section 2.



Technical Document Manual



Section 3



Calibration Manual



Section 4



Maintenance Manual



Section 5



Schematic Manual



Series



Mammography X-ray System



1



Service Manual



Section 1. Specification Manual of the MX-600



Diagnostic Mammography X-ray System



RADIEQ CO., LTD.



MX-600 1



Mammography X-ray System



Series



1



Service Manual



1. System Power Requirements



………………...………………………………………..



3



2. Environment ………...………………………………………………...............................



3



3. X-ray Tube(Insert)



3



………...…………………………………………………………….



4. X-ray Tube(Housing)



……………...…………………………………………………….



4



5. Bucky Device Assembly ………...……………………………………………………….



4



6. X-ray Source Assembly ………...……………………….……………………………….



4



7. x-ray Generator



4



RADIEQ CO., LTD.



………...……………………….……………………………………...



MX-600 2



Mammography X-ray System



Series



1



Service Manual



1. SYSTEM POWER REQUIREMENTS Input power :



3500W (6500VA) max



Output power :



2500W max



Output Voltage :



22~39kV



Output Current :



10~85mA



Frequency :



20KHz



Line Voltage :



220~240 ± 10%



IRMS(rest) :



1A



IRMS(exposure):



25A



IPEAK:



55A



50/60 Hz



2. ENVIRONMENT Operational :



10˚C~40˚C , 30~75% rH , 70~106kPa



Transport & Storge



-40˚C~70˚C , 10~85% rH , 50~106kPa



3. X-RAY TUBE(INSERT) Manufacturer :



VARIAN



Model :



RAD 85S



Target Materials :



Molybdenum



Nominal X-ray Tube Voltage :



22~39kV



Inherent Filtration :



0.63mm Be (IEC 60522)



Nominal Focal Spot Value :



0.1/0.3 mm (IEC 60613)



Anode Heat Storage :



300kHU



Anode Heat Cooling :



60 kHU / min



Filament Current



Large Focus : 2.0~4.0 A Small Focus : 2.0~4.0 A



Filament Voltage



Large Focus : 6.0~8.5 V Small Focus : 1.0~1.5



Serial NO. : RADIEQ CO., LTD.



MX-600 3



Series



Mammography X-ray System



1



Service Manual



4. X-RAY TUBE ASSEMBLY(HOUSING) Assembler :



VARIAN



Type of X-ray Tube :



Rotating Anode



Model :



B110



Target Angle :



16°



Maximum Housing cooling rate :



135HU



Housing Heat Storage Capacity :



500kHU



Weight :



22.5lbs (10.2kg)



Serial NO. :



5. Bucky Device Assembly Manufacturer:



Radieq Medical Systems



Model :



RB-600



Electric Energy Input :



24VDC



Grid :



Ratio 4:1 , Line 36 L/cm



6. X-ray Source Assembly Assembler :



Radieq Medical Systems



Position of The Focal Spot :



600mm From Sticking Surface of The Head cover



Serial No.



7. X-ray Generator Generator Type :



High Frequency Inverter



Input Power :



Single Phase 220VAC , 60Hz



Radiographic Ratings :



Large Focal Point 22-39kV / 1-600mA Small Focal Point 22-35kV / 1-100mA



Power Rating :



RADIEQ CO., LTD.



3.5kW



MX-600 4



Series



Mammography X-ray System



2



Service Manual



Section 2. Technical Document of the MX-600



Diagnostic Mammography X-ray System



RADIEQ CO., LTD.



MX-600 1



Series



Mammography X-ray System



2



Service Manual



Table of Contents



1. Single Load Ratings IEC 60613……....………………………………………………….



3



2. Focus and tolerance between the focus and datum axis……………………………...



5



3. The attenuation of X-ray beam..………………………………………………………….



6



4. The location of focus mark .………...…………………………………………………….



7



5. X-ray tube teat data …… ...……………………………………………………………… 6. The limit of X-ray tube temperature …………………………………………………….



9 10



RADIEQ CO., LTD.



MX-600 2



Series



Mammography X-ray System



2



Service Manual



1. Single Load Ratings IEC 60613



RADIEQ CO., LTD.



MX-600 3



Series



Mammography X-ray System



2



Service Manual



RADIEQ CO., LTD.



MX-600 4



Series



Mammography X-ray System



2



Service Manual



2. Focus and tolerance between the focus and datum axis.



RADIEQ CO., LTD.



MX-600 5



Series



Mammography X-ray System



2



Service Manual



3. The attenuation of X-ray beam. 1. Measurement at Non-attenuation



X-Ray



xxx R/min



2. Measurement at 0.3mmAl



O.3mmAl X-Ray xxx R/min



Measurement equipment :Radiation Detector



3. Measurement at Bucky(Grid + FRP)



Grid X-Ray xxx R/min



Bucky



RADIEQ CO., LTD.



MX-600 6



Series



Mammography X-ray System



2



Service Manual



4. Measurement data ( At 30kV, 20mAs, Large focus) Item 1st mR Non-attenuation 15.28 0.3mmAl 8.29 Bucky(with Grid) 8.94



2nd mR 15.45 8.30 8.83



3rd mR 15.33 8.30 8.96



Avr. mR 15.35 8.30 8.91



% 100 54 58



Standard Pass



- Blank -



RADIEQ CO., LTD.



MX-600 7



Series



Mammography X-ray System



2



Service Manual



4. The location of focus mark



RADIEQ CO., LTD.



MX-600 8



Series



Mammography X-ray System



2



Service Manual



5. X-ray tube Test data PURPOSE : To verify housing/unit integrity under simulated catastrophic conditions. The T/U will be ran at a power level that will bring the housing to thermal equilibrium @ 80°C for 10 minutes, following this the T/U power will be increased to obtain 100% heat storage for addition 2 min. < Over pressure test>



RADIEQ CO., LTD.



MX-600 9



Series



Mammography X-ray System



2



Service Manual



6. The limit of X-ray tube temperature. - Max. Temperature shutdown Test PURPOSE : To verify that safety devices on the unit are functioning properly during an over temperature / pressure condition. Temperature rating of thermal cutoff : 80°C Thermocouple location : Center of housing : The thermocouple is actually rated at 79.4°C ±3.9°C. The thermocouple is inside the housing and connected in series with the common leg of the SRATOR. See attached and data.



RADIEQ CO., LTD.



MX-600 10



Series



Mammography X-ray System



2



Service Manual



RADIEQ CO., LTD.



MX-600 11



Series



Mammography X-ray System



3



Service Manual



Section 3. Calibration Manual of the MX-600



Diagnostic Mammography X-ray System



RADIEQ CO., LTD.



MX-600 1



Series



Mammography X-ray System



3



Service Manual



Table of Contents



1. Calibration Overview



………………...………………………………………………….



3



2. Required Staff and Used Tools ………...………………………………………………...



5



3. Pre-Check list



………...………………………………………………………………….



6



4. Calibration and Adjustment ………...…………………………………………………….



7



Block Diagram ………...……………………………………………………………….



7



4.1. Cover Removal of Control Box ………...……………………………………….



8



4.2. Mode Switch of CPU Board



9



………...………………………………………...



4.3. Selection Method in Adjustment Mode



………...…………………………….



10



4.4. Selection Method in Adjustment Sub Mode 1.x ………...…………………….



10



4.5. Initial Data Transmission of CPU Board ………...……………………………..



11



4.6. Connection of Measuring Instrument ...………………………………………...



13



4.7. Setting Adjustment Mode ...……………………………………………………...



14



4.8. Adjustment of X-ray Tube Current ...……………………………………………



15



4.9. Adjustment of AEC ...……………………………………………………………..



17



4.10 Mode Setting ..................……………………………………………………….



23



4.11 Data Sheet ...…………..…………………………………………………………



24



. 5. Mechanical



Adjustment ...………………………………………………………………



28



5.1. Adjusting the Light Field ...………………………………………………………



28



5.2. Adjusting X-ray exposure field ...………………………………………………..



29



RADIEQ CO., LTD.



MX-600 2



Series



Mammography X-ray System



3



Service Manual



1. Calibration Overview All new equipment installations will require that M101-5401 MAIN CPU board be adjusted. If you are comfortable with your evaluation of kV/mA output findings, proceed to AEC & mAs procedures. ALL Radieq Co.,Ltd. EQUIPMENT IS PRE-CALIBRATED AT THE FACTORY. THE FOLLOWING PROCEDURES ARE PROVIDE ONLY IN EVENT THAT THE UNIT FAILS TO MEET SPECIFICATIONS. It is recommended that all electrical and mechanical adjustments be checked annually. When performing the following procedures, follow the sequence at which the procedure is given, as well as its content. It is suggested that you calibrate the equipment annually, first verify the performance to determine whether or not a complete calibration is required. Always begin by checking the kV/mA/mAs DIGITAL REGERENCE SIGNAL that were recorded on the "DATA SHEET". You can either spot check or verify all selected ranges for output comparison for what was recorded on the "DATA SHEET" on the initial installation or previous calibration. Always start with a new copy of the "DATA SHEET" and add it to the equipment records. In the event the CPU(IC1,Main CPU Board) is replaced or the Non-volatile RAM should fail, the information on the "DATA SHEET" will reduce the number of X-ray's required to re-calibrate the mammography unit. By following the procedures, we are actually reprogramming the information fund in the RAM memory. This allows us to fine tune the broad exposure information supplied to the buffer from Non-volatile RAM chip. The information in the Non-volatile RAM will be maintained even with power off by inner the battery back-up system. Should this system fail for whatever reason, the RAM memory will be lost and no exposure can take place. In this event, the programming must be re-supplied to the Non-volatile RAM from the ROM memory and the kV and mA calibration performed again. Having recorded all of the data on the "DATA SHEET" and within this section of the manual, you should be able to quickly restore the conditions within a very short time.



RADIEQ CO., LTD.



MX-600 3



Series



Mammography X-ray System



3



Service Manual



Caution:



Remember we are dealing with small focal spot size. Take your time to prevent tube over loading, and you will gain faster result in calibration. Allow 30 seconds between exposures.



RADIEQ CO., LTD.



MX-600 4



Series



Mammography X-ray System



3



Service Manual



2. Required Staff and Used Tools The following personnel and tools are required to Calibration MX-600, and they must be organized beforehand. (1) Tools and instruments required (2) One set of general purpose tools „ „ „



Screwdriver (small and medium size): Flat screwdriver for adjustments: M6 and M10 wrenches:



Each 1 1 Each 1



(3) Unpacking tools:



1 set



(4) AC voltmeters to measure the power supply voltage:



1



(5) One X-ray tube voltmeter set: 1 set „ X-ray tube voltmeter/mAs meter as Dynalizer III or equivalent „ X-ray tube voltmeter cable (6) X-ray tube current meter: 1 set „ X-ray tube voltmeter/mAs meter as Dynalizer III or equivalent (7) Acrylic phantoms for AEC adjustment „ 10mm thickness (120mm x 120mm):



8



(8) Density meter:



1



RADIEQ CO., LTD.



MX-600 5



Series



Mammography X-ray System



3



Service Manual



3. Pre-Check List Equipment S/N:



Date Installed : Caution:



Do not apply power without first checking that all boards are properly seated and all connectors and wires are connected exactly, damaged could occur to the unit.



Changing the Power Regulator (PSU1) supply voltages can result in a change in the calibration of the mammography unit. Record below the initial PWR supply voltage setting, if a PWR supply is replaced adjust it to the recorded value below to avoid extensive re-calibration. 1) Turn on the line breaker on the Control Box side. 2) Check that all power supply voltages present, do not change outputs if they are within the tolerance of 2%: +5VDC



+12VDC



-12VDC



+24VDC



3) Turn off the line breaker on the Control Box side.



RADIEQ CO., LTD.



MX-600 6



PSU1



RADIEQ CO., LTD.



ROTOR & IF BOARD



MAIN CONTROL BOARD



MAIN OP BOARD



PSU2



SUB OP1 BOARD



M_INV BOARD1



INVERTER BOARD



COM. MOTOR DRIVER BOARD



MOTOR CONTROL BOARD



SUB OP1 BOARD



MX-600 BLOCK DIAGRAM M_INV BOARD2



HFG



TUBE



Series



Mammography X-ray System



Service Manual



3



4. Calibration and Adjustment



Fig.1 MX-600 Block Diagram.



MX-600 7



Mammography X-ray System



Series



3



Service Manual



4.1. Cover Removal of Control Box Before starting calibration and adjustment, please remove sideboards of control box as shown in Fig. 2.



4



2



3 1



Fig.2 Cover Removal of Control Box. 1) 2) 3) 4)



At first loosen screws and remove the front cover 1. Next loosen screws and remove the side cover 2. Next loosen screws and remove the side cover 3. Finally loosen screws and remove the rear cover 4.



RADIEQ CO., LTD.



MX-600 8



Mammography X-ray System



Series



3



Service Manual



The Boards are showed in Fig. 3 as below.



MAIN CPU BOARD



I/F&ROTOR BOARD



C1



PSU1



T1



S2 (MODE S/W)



TB10



HIGHTENSION TRANSFORMER



INVERTER



HIGHTENSION TRANSFORMER



FUSE BOARD



TB2 INVERTER



TB1 Control Box [ FRONT ]



Control Box [ BACK ]



Fig. 3 Position of Boards



4.2 Mode Switch of CPU Board Each following mode is chosen by selection of the number of the MODE switch(S2) on Main CPU board. 0 - - - - - - - - - - - - - - - Normal mode(Operation mode) 1 - - - - - - - - - - - - - - - Adjustment mode (Calibration mode)



CE F012



3 RADIEQ CO., LTD.



Mode switch : S2



MX-600 9



Series



Mammography X-ray System



3



Service Manual



4.3 Selection Method in Adjustment Mode The contents of Display in the adjustment mode are displayed by 7 segment Light Emitting Diode on the left-hand side of kgf. The adjustment mode is chosen by Density switch “▼“ 0‐‐‐‐‐‐‐‐ Adjusting mode of mA



0.



kgf



1‐‐‐‐‐‐‐‐Adjusting mode of AEC



1.



kgf



2 3 4



‐‐‐‐‐‐‐ Setting for a various Mode



2.



kgf



‐‐‐‐‐‐‐Selection mode of the transmission data



3.



kgf



4.4 Selection Method in Adjustment Sub Mode 1.x The contents of Display in the adjustment mode are 7 segment LED on the right-hand side of Density. It is displayed numerically. The switch “▲” of Density chooses the adjustment sub mode. In the case of AEC adjustment: 0 - - - - - - - - - - - - - - Semi AEC adjustment 1-------------



RADIEQ CO., LTD.



Density Data adjustment



1. 0



kgf



1. 1



kgf



MX-600 10



Mammography X-ray System



Series



3



Service Manual



4.5 Initial Data Transmission of CPU Board Initial data other than individual data of the following clause is transmitted to Non-volatile RAM. (1)Disconnect the connection cable between Inverter and MAIN CPU board(M101-5401) at MAINCPU board side. (2)The MODE switch S2 of CPU board is set as No. 1. (3)Turn on the circuit breaker



kV



mAs



DENSITY



The display “—“ of kV blinks. NOTE:







“ is showed a blank mark.



4.5.1 (1)



CPU Board Individual Data Transmission Set up the adjustment mode Push Switch"▼" of DENSITY. Then a display is made into 3. by 7 segment LED on the left side of Density. 3.



0 kV



mAs



DENSITY



kgf mm DEG.



(2)



Push Switch “▲” “▼“ of kV, A display of 7 segment LED of kV is made into the following selection number. 8…………………………...CPU RAM to Non-volatile RAM 3.



8 kV



mAs



DENSITY



kgf mm DEG.



(3)



RADIEQ CO., LTD.



Pushing AEC switch is continued for 1 second. 7 segment LED of mAs blinks, about 10 seconds after, a buzzer sounds for 1 second and completion is told.



MX-600 11



Series



Mammography X-ray System



3



Service Manual



(4)



Push Switch “▲” “▼“ of kV, A display of 7 segment LED of kV is made into the following selection number. 9 ………………………….Non-volatile RAM to Buffer 3.



9 kV



mAs



DENSITY



kgf mm DEG.



(5)



Pushing AEC switch is continued for 1 second. 7 segment LED of mAs blinks, about 10 seconds after, a buzzer sounds for 1 second and completion is told.



NOTE: 4.5.1 clauses needs to be operated, only when the Non-volatile RAM for backup the case where MAIN CPU board is changed into a new , or CPU board is exchanged Here, since the data transmission from RAM / Buffer finished to Non-volatile RAM, Line Breaker is turned off, and the cable between an inverter and CPU board is inserted in a connector.



RADIEQ CO., LTD.



MX-600 12



Mammography X-ray System



Series



3



Service Manual



4.6 Connection of Measuring Instrument Connect tube voltmeter and mA or mAs meters as shown in Fig. 4. High-voltage transformer High-voltage cable for measurement



Control box



Unit high-voltage cable Anode



E



TB 2 MA+



X-ray tube



MA-



Cathode



mA meter



kV meter



or



mAs meter



Fig. 3 Connection for kV and mA/mAs meter (1) X-ray tube voltmeter Between a high-voltage generator and an X-ray tube, the breeders for X-ray high-voltage measurement (DYNALYZER, etc.) are connected using the high-voltage cable adapter for the mammography device. (2) X-ray tube current The jumper line between terminal MA+ to MA- of X-ray tube current measurement terminal stand TB2 is removed, and an mA meter(or mAs meter) is connected between them.



Caution:



If between terminal MA+ to MA- of the terminal stand TB2 become open, the unusual high voltage is impressed by the X-ray tube, and since it may be damaged, you be careful of connection of an ammeter enough.



RADIEQ CO., LTD.



MX-600 13



Mammography X-ray System



Series



3



Service Manual



4.7 Setting Adjustment Mode (1) The MODE switch S2 of CPU board is set as position 1. (2) Turn on Line breaker.



kV



mAs



DENSITY



Display becomes as it is shown in a figure. NOTE: The setting method of the MODE switch S2 unites a number with the portion with a cut of a knob.



Caution:



X-ray exposure of continuation is not made. Since there is a possibility that equipment may be damaged, please take the pause time of 30sec or more in X-ray exposure.



(3)Push On-(internal) switch of Control Panel . Switch “ I” ON(



)



The control circuit will be activating. Display becomes as it is shown in a figure.



Lamp LARGE F.S. lights up.



(4) The switch ” ▼ “ of DENSITY is pushed and the adjustment mode is chosen. Display becomes as it is shown in a figure. 0.



2 2 X X. X kV



mAs



DENSITY



kgf mm DEG.



RADIEQ CO., LTD.



MX-600 14



Series



Mammography X-ray System



3



Service Manual



4.8 Adjustment of X-ray Tube Current NOTE:



There is no adjustment of X-ray tube voltage.



An adjusting point is shown below. „ Large Focus:22kV, 25kV, 30kV, 35kV, 39kV . „ Small Focus:22kV, 25kV, 30kV, 35kV 4.8.1Adjustment of mA NOTE: Record this adjustment data to DATA SHEET which are indicated by 4.12.1 clause. These data is used as future adjustment and reference data at the time of maintenance. 1)



A setup in the adjustment mode Display becomes as it is shown in a figure. Lamp LARGE F.S. lights up.



0.



2 2 X. X X kV



mAs



kgf



DENSITY



2) The switch "▲" of mAs and "▼" are pushed and mA data is inputted. 0.



2 2 7 5. 8 kV



mAs



kgf



DENSITY



A switch "AEC" is pushed and data is saved.



3) An X-ray exposure switch is pushed and expose, X-ray tube voltage and current are measured.



The tube current is displayed on mAs display position during about 5 seconds after an X-ray exposure end. (At the time of the adjustment mode).



RADIEQ CO., LTD.



MX-600 15



Mammography X-ray System



Series



3



Service Manual



When measured value does not fulfill specified value, a switch "▲" or "▼" of mAs is pushed, mA setting value is changed, and it measures again. If standard value is satisfied, switch "AEC" is pushed and data is saved.



4) The setting value of kV is changed by switch "▲" or "▼" of kV, and measurement and adjustment of 3) are performed about each kV. LARGE Focus : ① ② ③ ④ ⑤



22KV Æ 70mA 25KV Æ 85mA 30KV Æ 72mA 35KV Æ 60mA 49KV Æ 60mA



5) A switch "HELP" is pushed and a focus is changed to a small focus.



It is made 22kV. Display becomes as it is shown in a figure. SMALL F.S. lamp turns on.



NOTE:If "HELP" switch is pushed again, it will change to a large focus. 6)Adjustment to 1 to 5 clause performed on the large focus is similarly performed with a small focus. 0.



2 2 X. X X kV



mAs



kgf



DENSITY



SMALL Focus : ① ② ③ ④



RADIEQ CO., LTD.



22KV Æ 13mA 25KV Æ 15mA 30KV Æ 12mA 35KV Æ 10mA



MX-600 16



Mammography X-ray System



Series



3



Service Manual



4.9 Adjustment of AEC The tube voltage set as adjustment is 24kV, 27kV, 30kV, 32kV and 35kV, and a radiograph subject thick is 20mm, 40 mm, 60mm, and 80mm. Adjustment of MAIN CPU board of the clause 4.9.1 is done at the time of factory shipment. When the time of exchange of MAIN CPU board or AEC Detector About density adjustment of the clause 4.9.2 and 4.9.3 when it is installed or changed to other combination with film and a screen, please carry out. 4.9.1 AEC Tilt Adjustment of CPU Board 1) The switch "▼" and "▲" of DENSITY is pushed and the adjustment mode is set to 1.1 Display becomes as it is shown in a figure.



2)



24



2. 5



kV



mAs



1. 1



kgf



DENSITY



Conditions are set up. Tube Voltage: 24 kV(Fixed) Radiography conditions: ① BUCKY device ② Compression paddle ③ 50mm acrylics



3)The switch "▼" or "▲" of mAs is pushed, and mAs display is set to 2.5. Display becomes as it is shown in a figure.



24



2. 5



kV



mAs



1. 1



kgf



DENSITY



4)A switch “AEC” is pushed and data is saved.



RADIEQ CO., LTD.



MX-600 17



Series



Mammography X-ray System



3



Service Manual



5 ) An X-ray exposure switch is pushed and X-rays is exposed, then Voltage/Time is measured at the test pin TP1 on CPU board.



6)The adjustment of the relation between the voltage waveform of TP1 and time is carried out by VR1(at Main CUP board) at approximately 2.5V/1sec as shown in the following figure. 2.5V



*Adjustment of VR1



1.0sec



(Time will be extended if it rotates to clockwise)



4.9.2 100mAs Adjustment of CPU Board 1) Same condition 4.9.1.(AEC Tilt Adjustment of CPU Board) Conditions are set up. Tube Voltage:24 kV(Fixed) Radiography conditions: ① BUCKY device ② Compression paddle ③ 50mm acrylics 2) The switch "▲" of DENSITY is pushed and the adjustment mode is set to 1.1.



3) The adjustment of the relation between the voltage waveform of mAs meter or Dynalizer III is carried out by VR2(at Main CPU board) at exactly 100mAs.



*Adjustment of VR2



RADIEQ CO., LTD.



(Time will be extended if it rotates to clockwise)



MX-600 18



Series



Mammography X-ray System



3



Service Manual



4.9.3 Density Adjustment in Bucky Device 1) Bucky device is attached in a radiography stand. Radiography conditions: ① BUCKY device ② Compression paddle ③ 20mm acrylics : 4pcs ④ Cassette(with a non-exposed film) 2)The switch "▲" of DENSITY is pushed and the adjustment mode is set to 1.0. 1.01 1.



2 4 0. 1 4 kV



mAs



kgf



DENSITY



3) The switch of “FILM & SCREEN” is pushed and "1" is chosen. “1” Factory Setting, “2,3” User Setting.



1: KORDARK FLIM(MIN R) + KORDARK CASSETTE(MIN R)



4) It is set as 24kV by the switch "▼" or "▲" of kV. (acrylics 20mm) It is set as density data by the switch "▼" or "▲" of mAs Display becomes as it is shown in a figure. 1.01 1.



2 4 0. 1 4 kV



5)



mAs



kgf



DENSITY



A switch "AEC" is pushed and data is saved.



6)An X-ray exposure switch is pushed and X-ray is exposed.



The exposed film is developed and density is measured. NOTE 1.30~ 1.40 of density is set when the unit released from the factory.



RADIEQ CO., LTD.



MX-600 19



Series



Mammography X-ray System



3



Service Manual



7) When density is darker, it is smaller in setting value, and setting value is enlarged when density is lighter. 8)The operation from 6) clause to 7) clause is repeated until density suits.



9)It is set as 27kV by the switch "▼" or "▲" of kV. (acrylics 20mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



2 7 0. 1 4 kV



mAs



kgf



DENSITY



10) The operation from 6) clause to 7) clause is repeated until density suits. 11) It exchanges for an acrylics 40mm board, and the tube voltage is set as 24kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



2 4 0. 2 4 kV



mAs



kgf



DENSITY



12)The operation from 6) clause to 7) clause is repeated until density suits. 13)It is set as 27kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



2 7 0. 2 8 kV



mAs



kgf



DENSITY



14)The operation from 6) clause to 7) clause is repeated until density suits. 15)



It is set as 32kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



3 2 0. 3 8 kV



mAs



kgf



DENSITY



16) The operation from 6) clause to 7) clause is repeated until density suits.



17)



It is set as 35kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



3 5 0. 5 0 kV



RADIEQ CO., LTD.



mAs



kgf



DENSITY



MX-600 20



Series



Mammography X-ray System



3



Service Manual



18) The operation from 6) clause to 7) clause is repeated until density suits. 19) It exchanges for an acrylics 60mm board, and the tube voltage is set as 27kV by the switch "▼" or "▲" of kV. It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



2 7 0. 4 8 kV



mAs



kgf



DENSITY



20) The operation from 6) clause to 7) clause is repeated until density suits. 21)It is set as 30kV by the switch "▼" or "▲" of kV. (acrylics 60mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



3 0 0. 6 0 kV



mAs



kgf



DENSITY



22) The operation from 6) clause to 7) clause is repeated until density suits. 23)It is set as 32kV by the switch "▼" or "▲" of kV. (acrylics 60mm) It is set as density data by the switch "▼" or "▲" of mAs



1. 10



3 2 0. 6 4 kV



mAs



kgf kgf



DENSITY



24) The operation from 6) clause to 7) clause is repeated until density suits. 25) It is set as 35kV by the switch "▼" or "▲" of kV. (acrylics 60mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



3 5 0. 7 4 kV



mAs



kgf



DENSITY



26) It exchanges for an acrylics 80mm board, and the tube voltage is set as 32kV by the switch "▼" or "▲" of kV. It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



3 2 0. 9 4 kV



mAs



kgf



DENSITY



27) The operation from 6) clause to 7) clause is repeated until density suits.



RADIEQ CO., LTD.



MX-600 21



Mammography X-ray System



Series



3



Service Manual



28)It is set as 35kV by the switch "▼" or "▲" of kV. (acrylics 80mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.



3 5 1. 0 2 kV



mAs



kgf



DENSITY



29) The operation from 6) clause to 7) clause is repeated until density suits.



RADIEQ CO., LTD.



MX-600 22



Series



Mammography X-ray System



3



Service Manual



4.10 Mode Setting (1) Set up the adjustment mode Push Switch"▼" of DENSITY. Then a display is made into 2. by 7 segment LED on the left side of Density. Display becomes as it is shown in a figure.



0



0



kV



mAs



2.



kgf



DENSITY



(2) Push Switch “▲” “▼“ of kV, select modes. And Push Switch “▲” “▼“ of mAs Select data. 0‐‐‐Focal spot 1‐‐‐Filter conversion 2‐‐‐Autofilmchanger 3‐‐‐Rh Filter 4‐‐‐REL conversion



1:Manual 0:Auto 1:Manual 0:Auto 1:Install 0: Not Install 1:Rh 0:Al 1:Manual 0:Auto



(3) A switch "AEC" is pushed and data is saved.



4.10.1 Fiducial loading condition



kV 22 25 30 35 36 37 38 39



RADIEQ CO., LTD.



Large focus mA mAs(max) 70 600 86 600 72 600 60 600 60 500 60 400 60 320 60 320



kV 22 25 30 35



Small focus mA mAs(max) 13 100 15 100 12 100 10 100



MX-600 23



Series



Mammography X-ray System



3



Service Manual



4.11 Data sheet 4.11.1



Adjustment and measured value of X-ray tube voltage and current



Performed date : Condition: Large Focus Set Value



Measured Value by mA adjusted the instruments Ref. Digital Code kV mA



Specification Deviation 4% of Set Value



KV



mA



KV



mA



22



70



21.2







22.8



67.2







72.8



25



85



24.0







26.0



81.6







88.4



30



72



28.8







31.2



69.1







74.9



35



60



33.6







36.4



57.6







62.4



39



60



37.4







40.5



57.6







62.4



Small Focus Set Value



Measured Value by mA adjusted the instruments Ref. Digital Code kV mA



Specification Deviation 4% of Set Value



KV



mA



22



13



21.2







22.8



12.5







13.5



25



15



24.0







26.0



14.4







15.6



30



12



28.8







31.2



11.5







12.5



35



10



33.6







36.4



9.6







10.4



RADIEQ CO., LTD.



KV



mA



MX-600 24



Mammography X-ray System



Series



3



Service Manual



4.11.2 mAs measurement table Performed date: Condition: Radiography method:



Set Value



Measured Value by the instruments L.F.



mAs



BUCKY + compression paddle + 50mm acrylic



22kV / 70mA



S.F.



mAs



22kV / 13mA mAs



Specification Deviation 4% of Set Value



1.00



0.90







1.10



1.30



1.17







1.43



1.60



1.44







1.76



2.00



1.80







2.20



3.20



2.88







3.52



4.00



3.60







4.40



5.00



4.50







5.50



6.40



5.76







7.04



8.00



7.20







8.80



10.0



9.00







11.00



20.0



18.00







22.00



50.0



45.00







55.00



100



90.00







110.00



540.00







660.00



600



RADIEQ CO., LTD.















MX-600 25



Mammography X-ray System



Series



3



Service Manual



4.11.3 The data sheet for mAs of AEC mode Performed date: Condition: Radiography method



KV



24



Thick. of acrylic (mm)



1 Ref. of mAs



Actual mAs



BUCKY (L.F.)



Magnification(S.F.)



2 Density



Ref. Actual of mAs mA



3 Density



Re f. of



Actual mAs



Density



20 40 20



27



40 60



30



40 60 40



32



60 80 40



35



60 80



RADIEQ CO., LTD.



MX-600 26



Mammography X-ray System



Series



3



Service Manual



4.11.4



The data sheet for Density of AEC mode



Performed date: Condition: Radiography



Density Position



%



-9



-45



-8



-40



-7



-35



-6



-30



-5



-25



-4



-20



-3



-15



-2



-10



-1



-5



0



0



1



+5



2



+10



3



+15



4



+20



5



25



6



+30



7



+35



8



+40



9



+45



RADIEQ CO., LTD.



27kV



BUCKY + Comp. Paddle + 40mm acrylic



Adjustment coefficient



Actual mAs



Density



Actual %



MX-600 27



Series



Mammography X-ray System



3



Service Manual



5. Mechanical Adjustment 5.1. Adjusting the Light Field 1) Do not attach any units to the mammography table (Bucky unit, etc). 2) Mount the 24 x 30 cm mask. 3) Turn on the manual switch for field lamp, then measure dimensions a, b, and c as Fig.



c



b



a Fig. 5



4)



Check the size of a, b, and c. If the following conditions are met, adjustment is not necessary. | a – b| ≤ 6mm, c = 230 ±6mm.



RADIEQ CO., LTD.



MX-600 28



Series



Mammography X-ray System



3



Service Manual



5.2.



Adjusting X-ray exposure field Preparation Prepare the cassette shown by Fig. 6. z The edge marker of an aluminum board with a thickness of 0.5mm and a width of 10mm is stuck on a cassette from edge at about 20mm place. z A marker is given with an aluminum board with a thickness of 0.5mm so that right and left of a film may be known.



„



about 20mm



Radiography table



Cassette



Marker Edge marker Fig. 6 Test cassette (a) (b) (c) (d)



Set the cassette as shown in Fig. 6. Mount the 24 x 30 mask Exposure X-ray to the film ( large focus, 30kV – 20mAs ) Develop the film, and measure any exposure field deviation.



Center of the film X-ray exposure field



c a



b



Fig. 7 If | a – b| ≤ 3mm and 1 ≤ c ≤ 4, adjustment is not required. RADIEQ CO., LTD.



MX-600 29



Series



Mammography X-ray System



3



Service Manual



If the conditions above are not satisfied, then proceed with (f) below. (e) Place the developed film on the cassette, then the cassette is set by the preposition. (f) Align the lamp exposure field with the X-ray exposure field shown on the film. (g) Using the adjusted lamp field as a guide, move the mask holder by loosening the four M3 nuts fixing it, so that the conditions above are met.



Four M3 nuts



Mask holder Fig. 8 (h) After the lamp exposure field has been adjusted to within the conditions above, tighten the four M3 nuts, exposure X-ray and check the results on the film. (i) After checking the developed film, make sure that all nuts are securely tightened, then remount the covers.



RADIEQ CO., LTD.



MX-600 30



Series



Mammography X-ray System



4



Service Manual



Section 4. Maintenance Manual of the MX-600



Diagnostic Mammography X-ray System



RADIEQ CO., LTD.



MX-600 - 1



Mammography X-ray System



Series



4



Service Manual



Table of Contents 1. Maintenance



……………………………………………………………………………..



1.1. Everyday Check and Inspection 1.1.1. Check Display



3



………………………………………………



3



……………….……………………………………………



3



1.1.2. Check Radiography Stand



…………………………..………….....……



5



1.1.3. Evaluation Film Quality ( Density ) ………………………………….……



6



1.2. Cleaning and Disinfection 1.2.1. Cleaning



…………………………………..……….…...……



7



…………………………………………..………………..…..…



7



1.2.2. Disinfection



……………………………………………….……….………



1.3. Periodical Inspection



…………………...……………………..……………



7 8



2.1. REPLACEMENT OF LAMP …………………………………………………….. 12 2.2. REPLACEMENT OF LAMP …………………………………………………….. 13 2.3. Messages and Corrective Actions ……………………………………………… 15 2.3.1 Error & Check Massages ……………………………………………….… 15



RADIEQ CO., LTD.



MX-600 - 2



Series



Mammography X-ray System



4



Service Manual



1.



Maintenance 1.1 Everyday Check and Inspection A user needs to check that all safe related equipment operate pertinently, before using equipment for diagnosis. Attention also needs to be paid to check of whether a display and a pilot lamp operate correctly. An X-ray exposure indicator is turned on only within while operating the X-ray radiography button. While not operating these buttons, when this display lights up, the X-ray equipment must be stopped immediately and must be connected with our company service part. 1.1.1 Check Display (1) Engage the circuit breaker located on the Control Box side. The mark “-“ in the kV display on the panel flashes JUDGEMENT



.



(2) Then push the ON (internal) switch located on the Control Panel The display of the last radiographic condition is lighted. JUDGEMENT



.



(3) Press the kV switch +, and kV value on the display increases. JUDGEMENT



.



(4) Press the kV switch - , and kV value on the display decreases. JUDGEMENT



.



(5) Press the mAs switch + , and mAs value on the display increases. JUDGEMENT



.



(6) Press the mAs switch - , and mAs value on the display decreases. JUDGEMENT



.



(7) Remove the BUKY DEVICE. The BUKY lamp should extinguish. Next attach the Magnification box. MAGNIFY and SMALL F.S. lamp should illuminate. . JUDGEMENT



RADIEQ CO., LTD.



MX-600 - 3



Series



Mammography X-ray System



4



Service Manual



(8) Depress the AEC switch. Each time the display will sequence between MANUAL and Full AEC, Semi AEC. . JUDGEMENT (9) In Full AEC mode the kV and the mAs display should be blanked. JUDGEMENT



.



(10) In Semi AEC mode the mAs display should be blanked. JUDGEMENT



.



(11) In AEC mode the DENSITY display should illuminate. JUDGEMENT



.



When depress the DENSITY switch, DENSITY display should be variable between +9 to –9. . JUDGEMENT (12) Depress the FILM SPEED switch. Each time the display will sequence between 1, 2, 3 to 1. . JUDGEMENT



RADIEQ CO., LTD.



MX-600 - 4



Series



Mammography X-ray System



4



Service Manual



1.1.2. Check Radiography Stand (1) Vertical and Rotation movement of C-arm C-arm rotation Switch



This switch can make C-arm’s rotation. :Counterclockwise



:Clockwise. C-arm Up / Down switch



Keep pressing to make C-arm’s continuous rotation :Up movement



:Down movement



JUDGEMENT



.



(2) Attachment and detachment of a pressure paddle Each pressure paddle of accessories can attach and detach easily among a pressure paddle support part. . JUDGEMENT (3) Vertical movement of a Compression Paddle • Where a pressure paddle is attached, turn a Manual compression knob by hand and carry out vertical movement smoothly. JUDGEMENT . • Where a pressure paddle is attached, operate a Compression foot switch and carry out vertical movement smoothly. JUDGEMENT .



RADIEQ CO., LTD.



MX-600 - 5



Series



Mammography X-ray System



4



Service Manual



1.1.3.Evaluation Film Quality ( Density ) The quality control test of a film development machine is performed, and after checking that the parameter of a film development machine is in control within the limits, the function of automatic exposure control (AEC) is checked by performing phantom photography and checking film density. The MX-600 is designed to provide consistent densities on film across the full spectrum of tissue densities、regardless of breast size. Normally adjusted to provide an optical density of 1.4 at AEC mode, it can be adjusted to any density specified. This can be done to suit particular tastes or to compensate for processing irregularities etc..



RADIEQ CO., LTD.



MX-600 - 6



Series



Mammography X-ray System



4



Service Manual



1.2. Cleaning and disinfection Caution:



When performing cleaning and disinfection, it is required to turn off a power supply input.



1.2.1 Cleaning The parts that contact a patient among the parts of equipment must be cleaned before each use. Please make cloth become wet with neutral detergent thinned with water, and wipe equipment. . Please do not use a cleanser containing abrasive and cleaning liquid containing organic solvents or solvents (for example, benzene, a stain-removal solvent).



Caution:



Please do not hang a spray on equipment! cleaning equipment cleaning liquid.



In any case, don’t permeate



1.2.2.Disinfection Immerse the compression paddle, the magnification kit and the Bucky devices into a cleaning disinfection solution of minimum 2.4% glutaldehyde (ex. CIDEX, Omnicide., etc.) in a separate container for 45 minutes in every 14 days. Make sure the surface is completely moistened. After that take it off and rinse it with the clean water.



Caution:



1. Switch off the device before disinfecting . Disconnect it from the mains. 2. Take care that no liquid gets into the equipment. 3. If you put into the disinfection fluid too long, the life time of the compression paddles, the magnification kit and the Bucky device are reduced. 4. Do not perform the disinfection with E.O. Gas. 5. Do not use the degenerated solvent.



RADIEQ CO., LTD.



MX-600 - 7



Series



Mammography X-ray System



4



Service Manual



1.3. Periodical Inspection In order to secure the safety of a patient, an operator, and a third person, equipment needs to be checked every 12 months and reliability and operatively need to be maintained. The parts which may cause danger in respect of safety if it exhausts need to undergo the inspection by the qualified person periodically every at least 12 months. Moreover, parts need to be exchanged if needed. When the local rule and regulation define still frequent check and maintenance, please follow the regulation. 1. Stand (Inspecting the C-arm support) Item



1



2



3



Contents



1) Remove all cable lubricant from the cable with organic solvent. 2) Make sure there are no broken strands at the Cable cable ends or sections inspection that contact pulleys. Apply new cable lubricant uniformly to sections which contact the pulley over the entire C-arm vertical stroke. Vertical Press movement switches movement on both sides of the C-arm inspection to check operation Press movement switches Vertical on both sides of the C-arm movement belt to check operation. Make sure that the belt inspection fixing nut is not loose.



RADIEQ CO., LTD.



Standard



No-broken strands



Insp. jig



Inter-val 6M 1Y



Processing



Replace the cable.







Must operate







Should operate smoothly. Activation force should be 5kgf or less.







Replace belt.



MX-600 - 8



Series



Mammography X-ray System



4



Service Manual



Item 4



5



Insp. jig



Interval 6M 1Y



Contents



Standard



Rotational movement inspection



Press movement switches on both side of the C-arm to check operation.



Must operate







Cable operation confirmation



Move C-arm up and down; make sure that the cable moves smoothly inside the stand.



Make sure that the cable does not interfere with support section vertical movement.







Processing



Adjust cable



2. C-arm unit Item



1



2



3



Rotational shaft bracket screw tightness check Screw tightness check for the aluminum casting section of the mammography table mount Screw tightness check for the aluminum frame of the tube mount



Standard



Remove the rotational shaft section cover, then make sure that the six M8 screws are tight.



Screws must be tight







Tighten



Remove the plastic cover, and make sure that the four M6 screws are tight.



The screw must be tight.







Tighten



Remove the plastic cover, and make sure that the four M6 screws are tight.



The screws must be tight.







Tighten



RADIEQ CO., LTD.



Insp. jig



Interval 6M 1Y



Contents



Processing



MX-600 - 9



Series



Mammography X-ray System



4



Service Manual



3. Inspection the compression unit Insp. jig



Interval 6M 1Y



Item



Contents



Standard



1



Chain tension and damage check



1) Because tension cannot be measured, check it by touch 2) Check for damage.



When the belt is pressed by hand. There should not be loosened too.







2



Positioning motor drive inspection



It checks that pressure power is below 5kgves.



Less than 5kgf







Processing



4. X-ray high voltage unit Item



Tube voltage, Tube current 1 and mAs check



Contents



Insp. jig



Interval 6M 1Y



kV: 24.0 to 26.0 Generate X-ray using the mA: conditions below, 81.6 to 88.4 checking tube current mAs: and mAs. 18.0 to 22.0 Large focus: 25kV, 20mAs Small focus: 25kV, 10mAs



RADIEQ CO., LTD.



10



Standard







kV: 24.0 to 26.0 mA: 14.4 to 15.6 mAs: 9.00 to 11.0



Processing



If it is not satisfied, perform adjustment Refer Service Manual



MX-600 -



Series



Mammography X-ray System



4



Service Manual



5. AEC (Film density check) Item



1



2



Film density check



Actual mAs display check



Contents



Insp. jig



Perform AEC radiography using a 40mm thick acrylic *40mm phantom using the tube Make sure that thickness voltage setting below and the density is acrylic with DENSITY at “0”, then within the phantom check density. reference density set at Conditions the time of * Tube voltage: 23kV *Density installation. * Using the film system meter and the screen. * Bucky radiography Connect a mAs meter according to the tube voltage/current measurement methods The difference given section 4.6, then between the perform AEC compression mAs meter radiography using the and the *mAs conditions below. Make operation sure that the operation meter panel mAs panel mAs display is display should accurate. be within ±10%. - 40mm thickness acrylic phantom - Tube voltage: 25kV - DENSITY: 0%



RADIEQ CO., LTD.



11



Standard



Interval 6M 1Y



Processing







If it is not satisfied, perform adjustment Refer Service Manual







If it is not satisfied, perform adjustment Refer Service Manual



MX-600 -



Series



Mammography X-ray System



4



Service Manual



2.1. REPLACEMENT OF LAMP Replacement of halogen lamp (1)Tool required Screw driver Glove to replace the lamp



1 sec 1pair



(2)Lamp specification



Model : HLWT-J / 12V, 50W Maker :



NARVA SPEZIALLAMPEN Main in GERMANY



(3)Replacement procedure ① Remove the cover plate.



② Change the halogen lamp. (Don’t touch directly glass)



③ Get the cover plate back to its original position. ④ Press the lamp switch and make sure that lamp is lit up.



CAUTION Be sure not to touch the lamp with your bear hands as it is very hot, use gloves to protect from heat.



RADIEQ CO., LTD.



12



MX-600 -



Series



Mammography X-ray System



4



Service Manual



2.2. Bucky Device Maintenance for cassette size (for Kodak MIN-R II and Standard cassette) If the cassette thickness is not about 19mm(Kodak MIN-R type) that is 14.5mm(Standard cassette type) then please replace stopper and attach guide rail. * If the guide rail installed in the bucky then that is standard cassette type(14.5mm). (See No.9 Picture) (Standard cassette thickness : about 14.5mm -Kodak MIN-R II , Standard cassette type , CR Mammo cassette) Exchange of Part in order to use a standard cassette



1. Exchange of Cassette Stopper



2. Removal of Old Stopper



3. Exchange to New Stopper



4. Exchange of Both side Stopper



5. Setting of a Cassette Guide Rail



RADIEQ CO., LTD.



13



MX-600 -



Series



Mammography X-ray System



4



Service Manual



6. Tightening Bolts at Two points of Front Guide Rail



8. Tightening Bolts at Two points of Rear Guide Rail



RADIEQ CO., LTD.



14



7. Setting of a Rear Cassette Guide Rail



9. Completion of exchanging parts



MX-600 -



Mammography X-ray System



Series



4



Service Manual



2.3 Messages and corrective actions 2.3.1. Error & Check messages This unit has self-diagnostic function. Check when INHIBIT lamp is on.







Code



Corrective action



C01 – 00



Cassette is not inserted in Insert cassette into bucky device. bucky device



C01 – 01



Exposed cassette is in bucky.



Insert cassette with new film.



C02 – 01



Sensor can not sense filter.



Filter might be stuck at somewhere out of sensing range. If operation is unavailable, call engineer.



C04 – 00



No beam limiting mask is set.



Insert beam limiting mask.



C04 – 01



Beam limiting mask is 18×24 Exchange cassette with same size one. type but cassette is not for it.



C04 – 02



Beam limiting mask is 24×30 Exchange cassette with same size one. type but cassette is not for it.



C06 – 00



BUCKY error



RADIEQ CO., LTD.



15



Meanings of error



Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.



MX-600 -



Series



Mammography X-ray System



4



Service Manual



Error code



Corrective action



C06 – 01



Inverter / alarm 1 error



Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.



C06 – 02



Inverter / alarm 2 error



Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.



C06 – 03



Inverter / filament error



Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.



C06 – 04



Turn off circuit breaker then ON again. X-ray tube anode rotation error Turn on “ I “ switch. / Tube over heat If error repeats, call engineer.



C08 – 00



Turn off circuit breaker then ON again. Main Board and Motor Turn on “ I “ switch. Controller connection error If error repeats, call engineer.



C08 – 01



Emergency switch pushed



Emergency switch turn off. Turn off circuit breaker then ON again. Turn on “ I “ switch.



C08 – 02



Motor not moving error



Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.



C09 – 00



X-ray exposure setting mAs data.



small



then



C09 – 01



X-ray exposure setting mAs data.



large



then



C09 – 02



X-ray exposure mistake. CHECK lamp on error.



RADIEQ CO., LTD.



16



Meanings of error



Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer. Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer. X-ray can not be made during check lamp is on. Press CLR switch to restart.



MX-600 -



Series



Mammography X-ray System



5



Service Manual



Section 5. Schematics of the MX-600



Diagnostic Mammography X-ray System



RADIEQ CO., LTD.



MX-600 1



Series



Mammography X-ray System



5



Service Manual



Table of Contents 1



BLOCK DIAGRAM



2



FUSE BOARD : M101-5403



………………………………………………..………



4



3



COMPRESS MOTOR DRIVE(HEADER BOARD) : M101-5101……….…………



5



4



INTERFACE & ROTOR BOARD : M101-5402……………………………......……



6



5



MAIN CONTROL BOARD : M101-5401………………………………...……...……



7



6



MAIN OP BOARD : M101-5501………………………………..……….…...…….…



8



7



MOTOR INVERTER BOARD : M101-5405..…………………………..……..…..…



9



8



MOTOR CONTROL BOARD : M101-5103………………………….………………



10



9



SIDE OP BOARD : M101-5201...………………………...……………..……………



11



10



HFG PART BLOCKDIAGRAM



12



RADIEQ CO., LTD.



…………………………………………………………..………



.………………………...……………..……………



3



MX-600 2



Series



Mammography X-ray System



5



Service Manual



1.



BLOCK DIAGRAM



RADIEQ CO., LTD.



MX-600 3



Series



Mammography X-ray System



5



Service Manual



2.



FUSE BOARD : M101-5403



RADIEQ CO., LTD.



MX-600 4



Series



Mammography X-ray System



5



Service Manual



3.



COMPRESS MOTOR DRIVE(HEADER BOARD) : M101-5101



RADIEQ CO., LTD.



MX-600 5



Series



Mammography X-ray System



5



Service Manual



4.



INTERFACE & ROTOR BOARD : M101-5402



RADIEQ CO., LTD.



MX-600 6



Series



Mammography X-ray System



5



Service Manual



5.



MAIN CONTROL BOARD : M101-5401



RADIEQ CO., LTD.



MX-600 7



Series



Mammography X-ray System



5



Service Manual



6.



MAIN OP BOARD : M101-5501



RADIEQ CO., LTD.



MX-600 8



Series



Mammography X-ray System



5



Service Manual



7.



MOTOR INVERTER BOARD : M101-5405



RADIEQ CO., LTD.



MX-600 9



Series



Mammography X-ray System



5



Service Manual



8.



MOTOR CONTROL BOARD : M101-5103



RADIEQ CO., LTD.



10



MX-600



Series



Mammography X-ray System



5



Service Manual



RADIEQ CO., LTD.



11



MX-600



Series



Mammography X-ray System



5



Service Manual



9.



SIDE OP BOARD : M101-5201



RADIEQ CO., LTD.



12



MX-600



Series



Mammography X-ray System



5



Service Manual



10.



HFG PART BLOCKDIAGRAM



RADIEQ CO., LTD.



13



MX-600



Series



Mammography X-ray System



5



Service Manual



RADIEQ CO., LTD.



14



MX-600