NOTIFIED BODY: TUV PRODUCT SERVICE GMBH EC CERTIFICATE NO.: G1 08 06 46736 003



POX-100BT High Powered Mobile X-ray System Product Manual (Operation & 0DQXDO)



Product Manual: POX-100BT



TABLE OF CONTENTS



PART I. OPERATION MANUAL 1. General Description



---------------------------------------------------------------------



3



2. Notice for Safe Use



--------------------------------------------------------------------



5



3. Operating Environment



---------------------------------------------------------------------



6



4. Radiation Safety



---------------------------------------------------------------------



7



5. Composition



---------------------------------------------------------------------



8



6. Specification



---------------------------------------------------------------------



9



7. Name of Each Part



---------------------------------------------------------------------



16



8. Operation



---------------------------------------------------------------------



19



9. How to Operate



---------------------------------------------------------------------



21



10. Remote Control Function



---------------------------------------------------------------------



25



1. Notice



--------------------------------------------------------------------



29



2. Regular Maintenance



---------------------------------------------------------------------



30



3. Block Diagram



---------------------------------------------------------------------



31



PART II. SERVICE MANUAL



4. Calibration Procedures



---------------------------------------------------------------------



32



5 Error Code and Resolution



---------------------------------------------------------------------



37



6. Radiation Dose



---------------------------------------------------------------------



39



7. Error Code and Resolution



---------------------------------------------------------------------



40



8 Schematics



---------------------------------------------------------------------



42



2/47



Product Manual: POX-100BT



PART 1: OPERATION MANUAL 1. Gener al Description 1) Outline Thank you for purchasing POX-1 00BT, high powered mobile x-ray syste m. POX-100BT is the most powerful high frequency x-ray machine. Our mobile systems are ripple free high frequency type with 40~50% stronger power output efficiency compared with conventional x-ray systems. Therefore, it can be called as the perfect concept of radiography system providing easy operation and mobility for indoor and outdoor radiography. 2) Unique Features -High frequency type x-ray generator, high efficiency -Maximum output: 110kV / 100mA -750 APR memories -Remote control software -Equipped digital reverse display function -Electrical requirement: 220VAC, 50/60HZ 3) Product Manual This consolidated manual consists of operating and service instruction. Product m anual provides POX-100BT operating instructions for x-ray practitioner and other medical personnel trained in the use of x-ray equipment. Service manual provides installation in struction, calibration and routine maintenance for field service staffs. This manual is not designed to replace or substitute for certified training in the radiological or medical field. By reading this manual, the operator should be qualified in the use of this equipment. This X-ray system is only to be used by licensed operators in diagnostic, operative and post-operative procedures. 4) Notice For Safety This manual provides guideline for safe use and o peration of the medical radiographic system. The operator should be trained in the use of this equipment and refer to this manual in conjunction with that training. It is owner ’s responsibility to ensure that the system is operated by properly trained and qualified personnel who have obtained credentials from local, state, and federal authorities where required



3/47



Product Manual: POX-100BT



only. Potential ha zards exist in the use of medical electronic devi ces and X-ray systems. All users and operators must understand the safety and emergency procedures, and operating instructions written in this manual Poskom certifies that each system complies with the Radiation Control for Health Safety Act in T itle 21, Cha pter I, subchapter J of the Code of Federal Reg ulation in the USA and EN 6060 1, genera l requirements for radiation protection in diagnostic X-ray equipment for IEC 601-1 -3 in European Nations. 1.EN60601-1:1990+A1:1993+A2:1995+A13:1996 Medical electrical equipment— Part1: General requirements for safety and essential performance. 2.EN60601-2:2001/A1:2006 Medical electrical equipment—Part1-2:General requirements for safety-collateral standard: Electromagnetic compatibilityRequirements and tests. 3.EN60601-2-32:1994 Medical electrical equipment—Part2:Particular requirements for the safety of associated equipment of X-ray equipment. 4.EN60601-1-3:1994 Medical electrical equipment—Part1:General requirements for safety—3.collateral standard: General requirements for radiation protection in Diagnostic X-ray equipment. 5.EN6061-2-7:1998 Medical electrical equipment—Part2-7:Particular requirements for the safety of high-voltage generators of diagnostic X-ray generators. 6.EN60601-2-28:1993 Medical electrical equipment—Part2:Particular requirements For the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis. ACOMA'S assume no responsibility or liability for after-sales operating and safety practices, nor can it be responsible for personal injury or damage resulting from misuse of its x-ray systems. Please address questions and comments regarding safety issue to ACOMA'S service organization and/or each distributor in your region. 5) Cautions and Warnings Possible hazards are associated with unsafe operation of x-ray equipments. These hazards are described on the following warning or caution signs with the safety precautions which should be taken. CAUTION The X-ray ra diation could be dangerous to patients and operators unless following safety regulations are strictly observed.



4/47



Product Manual: POX-100BT



2. Notice for Safe Use 1) Equipment Direction User’s safety, easy operation and product reliability are our major concern in designing this system of generators. However, it is advisable t o observe the following instructions and guidelines for your additional safety in operating this equipment. (1) This product should be operated only by or under the supervision of legally qualified persons. (2) POX-100BT i s de signed f or the radiographic u ses and n ot for fl uoroscopic o r other associated applications. (3) POX-100BT should be used for the diagnosis, not for the therapy. (4) POX-100BT is specified as ClassⅠtype B equipment under the IEC 601-1, 2 standards. CAUTION This equipment is not protective against ingress of water or liquid.



(5) Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask help of ACOMA'S or its authorized dealers for their service. (6) This system has been calibrated for optimal operation. (7) If you h ave any troubles with the equipm ent, turn off power immediately, and notify ACOMA'S or its authorized dealers of their service and assistance. (8) If you want connecting a device made by other manufacturers to this equipment, ask ACOMA'S or its authorized dealers for their instructions. 2) Direction on Hazardous waste This equipment and accessories are to be disposed of safely after the life span of the mand national regulation must be observed. - This equipment and accessories are to be disposed according to the instructions given by national regulations or ask a specialized waste service company to dispose of it. 3) Direction on Electromagnetic Wave “This e quipment has been tested and found to comply wit h the limits for medical d evices in IEC 60601-1-2:1994. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. This equipment generates and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other de vices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which ca n be determined by t urning



5/47



Product Manual: POX-100BT



the equipment off and on, t he user is encouraged to try to correct the interference by one or more of the following measures: - Reorient or relocate the receiving device. - Increase the separation between the equipment. - Connect the equipment into an outlet o n a circuit di fferent from that to which the other device(s) are connected. - Consult the manufacturer or field service technician for help.” 3. Working Environment 1) Maintenance Environment For proper operation and maintenance, please avoid following places. -Where the equipment is exposed to water vapor. -Where the equipment is exposed to direct sunlight. -Where the equipment is exposed to heavy dust. -Where the equipment is exposed to high humidity. -Where there is insufficient ventilation. -Where the equipment is exposed to salty atmosphere. -Where the equipment is exposed to chemicals or gas. 2) Operating Environment Keep safe distance from strong vibration and maintain good environment Operational Environment Temperature range



10°C ~ 40°C ( 50℉ ~ 104℉)



Humidity range



30% ~ 75%RH



Optimal Working Environment Temperature range



17°C ~ 23°C ( 63℉ ~ 73℉)



Humidity range



40% ~ 60%RH



For safe sto rage a nd transport ation, please keep proper temperature, humidity and other conditions Storage and Transportation Temperature range



-25°C ~ +60°C (-1 3℉ ~ +140℉)



Humidity range



10% ~ 95%RH



6/47



Product Manual: POX-100BT



4. Radiation Safety CAUTION The ionized radiation could be dangerous to patients and operators unless following safety regulations are strictly observed



-Users and operators are requested to wear appropriate protecting gears and clothes. -Users and operators are requested to be leave safe distance from radiant sources and secondary radiation zones. -Users and operators are requested to eliminate all unnecessary objects near the exposure zones. -Users and operators are requested to leave at least 20cm away from the target. -Users and operators are requested to use minimum kV and mAs values for experimental use. -Users and operators are requested to be careful not to exceed the limited radiograms in the exposure zone.



7/47



Product Manual: POX-100BT



5. Composition POX-100BT is a radiological device for medical us es. It should be operated by qualified practitioners for radiographic diagnosis only. Users should comply with safety and health regulations concerning the ionizing protection and the electrical/ mechanical safety of medical devices. POX-100BT consists of followin g parts an d devi ces: Foll owing items should be firmly fixed and connected for regular functioning. 1) Operation Console 2) Power Factor Controller Circuit & Inverter 3) HV Tank including X-ray Tube 4) Collimator 5) Han d Switch Supplement 1) Operation & Service Manual 2) Power Plug Hand Switch Remote Control (Picture 1)



1st Stage = Prep Switch 2nd Stage = Exposure Switch



Hand Switch Holder



All of remote control functions are operated by depressing the first stage of the hand switch (ready).



8/47



Product Manual: POX-100BT



6. Product Specifications 1) Electrical Features Model



POX-100BT



Input Power



Battery



Output Power



5.0kW (Max.)



Generator Type



Inverter



Switching Frequency



75kHz Tube Voltage



40kV~50kV



51kV~60kV



61kV~70kV Radiography



71kV~80kV



81kV~100kV



101~110kV



Current



mAs



20mA



0.1mAs



40mA



0.2~0.3mAs



100mA



0.4~10mAs



80mA



12~32mAs



60mA



36~80 mAs



50mA



100mAs



20mA



0.1mAs



40mA



0.2~0.3mAs



80mA



0.4~8.0mAs



50mA



9.0~80mAs



40mA



100mAs



20mA



0.1mAs



40mA



0.2~0.3mAs



50mA



0.4~8.0mAs



40mA



9.0~80mAs



30mA



100mAs



20mA



0.1mAs



40mA



0.2~0.3mAs



50mA



0.4~8.0mAs



40mA



9.0~21mAs



30mA



25~64mAs



20mA



0.1mAs



40mA



0.2~8.0mAs



30mA



9.0~30mAs



20mA



32~80mAs



20mA



0.1mAs



40mA



0.2~5.0mAs



30mA



6.4~12mAs



20mA



18~80mAs



Max. kV Deviation



±7 %



Max. mAs Deviation



±10 %



9/47



Product Manual: POX-100BT



Posko



Max. sec. Deviation



m Co., Ltd.



±10 %



Display



7-Segment & LED Model



Battery



ES24-12



Battery Voltage



12VDC /24A X 4EA



Capacity



5.0AH / 20°C



Tube Model



OX/110-15



Anode heat storage X-ray Tube



30kJ



Inherent Filtration



0.5mm Al



Target angle



15°



Focal Sport



0.6mm × 1.8mm



Total Filtration



3.0 mm Al



Battery Charger Input



110 / 220VAC, 50/60Hz



Battery Input



54VDC / 0.5A



Collimat



Type



Double slit type, Manual operation



or



Min. X-ray Field Size



5cm X 5cm @ 1m



Max. X-ray Field Size



35cm X 35cm @ 65cm



Timer



Push button illuminator with 30 sec timer



Lamp



24VDC 150W Halogen



FAN



24VDC / 0.07A



Inherent Filtration



0.5mm Al



2) Mechanical Features Section



SPCIFICATION



Output 5.0 Generator



㎾ Pos



kom



POX-100BT PFC 6kW(380-400VDC)



Poskom



Power Board



POX-100BT PWR



Poskom



HV Tank



POX-100BT TANK



Poskom



kV & mA & mAs/sec digital display -Thickness(for anim al) / Den sity(for human) Console(OP)



digit al display



Poskom



-Prep. & Exposure LED -Collimator lamp Switch -APR Store Switch Lamp: 12V100 Watt, halogen lamp, Collimator (R72)



Timer: 30sec X-ray Field Size: 43cm * 4 3cm @ 1m SID



10/47



Ralco



Product Manual: POX-100BT



Type: Fixed anode Focal spot size: 0.6mm / 1.8mm X-RAY TUBE



Heat Storage: 30kJ



(OX/110-15)



Cooling rate: 250W Inherent Filtration: 0.5mm Al Target angel: 15°



3) Specification Dimension (1) Main Body



11/47



NAGO



Product Manual: POX-100BT



4) Tube Specifications (1) General Data Operating Tube Voltage: OX/110-15 (CEI) Tube kV range: 40 ~ 110kV Focal spot size: 0.6mm x 1.8mm D. Heat Storage: 30kJ E. Cooling rate: 250W F. Inherent Filtration: 0.5mm Al G. Target angle: 15° (2) Filament volt curve A. Small Focus



B. Large Focus



12/47



Product Manual: POX-100BT



(3) Max. Output Curve A. Small Focus



B. Large Focus



Anod e Heat Curve



3) Dimension



13/47



Product Manual: POX-100BT



(4) Bipolar Thermal Curve



4) Name of Major Parts 7.1 Name of Parts



(5) Dimension



14/47



Product Manual: POX-100BT



7. Name of Each Part 1) Main Console



17



1 2



16 15



3



14



4



13



5



12



11



10 6



9



7



8



[1]



Exposure LED



[10]



Exposure Switch



[2]



Ready LED



[11]



Density Up/Down Switch



[3]



kV Up/Down Switch



[12]



sec/mAs Mode Select Switch



[4]



APR Size Select Switch



[13]



Collimator Lamp Switch



[5]



APR Mode Select Switch



[14]



sec/mAs Select Switch



[6]



APR Store Switch



[15]



Battery Leftover Display



[7]



APR Region Select Switch



[16]



Error LED



[8]



APR Direction Select Switch



[17]



Wait LED



[9]



Ready



Switch



-



15/47



-



Product Manual: POX-100BT



2) Sub Console



1 2 3



4



8



7



5



6



[1]



Wait LED



[5]



kV Up/Down Switch



[2]



Exposure LED



[6]



mAs Up/Down Switch



[3]



Ready LED



[7]



Collimator Lamp Switch



[4]



kV Display



[8]



mAs Display



16/47



Product Manual: POX-100BT



Product Labels: Human (POX-100BT_H)



Warning: This X-ray unit maybe dangerous to patient and operator unless safe exposure factors and operation instruction are observed



DIAGNOSTIC X-RAY UNIT Model: POX-100BT Serial No: DATE: Input Rating: 110/220V~, 50/60Hz, 5kVA Rating: X-ray Tube Voltage: 40-110kV X-ray Tube Current: 20-100mA Total filtration: Min 3.0 mmAl (Tube: 0.5 mmAl, Collimator: 0.5 mmAl, Al Filt: 2.0mmAl) X-ray Tube : OX/110-15 Serial No: Focal spot: 0.6mm x 1.8mm ACOMA'S R oom 405, Unitech Ville Venture Town, 1141-2, Baekseok-dong, I lsandong-gu, Goyang-si Gyeonggi-do 411-722, Korea Take adequa te P recautions to p revent the possib ility of a ny pe rsons carelessly exposing themselves or others to radiation.



HIGH VOLTAGE TANK MODEL: HV-100HF Serial No: DATE: Tube Model: OX/110-15 Tube Serial No: Focal spot: 0.6mm x 1.8mm Inherent Filtration: 0.5mmAl Rating: MAX. 110kV, 40mA, 50kV, 100mA ACOMA'S Room 405, Unitech Ville Venture Town, 1141-2, Baekseok-dong, Ilsandong-gu, Goyang-si Gyeonggi-do 411-722, Korea Warning: Electric shock hazard do not remove cover Reserve Servicing to qualified personnel.



BEAM LIMITING DEVICE (COLLIMATOR) MODEL: MCM-100BT Serial No: Input Rating: 24VDC, 150W Min. X-ray field size: Below 5m x 5m at SID 100cm Max. X-ray field size: Below 35cm x 35cm at SID 65cm Max. Tube Voltage: 110kV, Inherent filtration: 0.5 mmAl ACOMA'S Room 405, Unitech Ville Venture Town 1141-2, Baekseok-dong, Ilsandong-gu, Goyang-si Gyeonggi-do 411-722, Korea



17/47



Product Manual: POX-100BT



8. Operation 1) Prep aration A. Wear a lead apron for x-ray exposures. B. Leave at least 2 meters away from the unit or use extended hand switch cable in order to keep safe distance. C. Apply pro per beam size a nd techni cal values to produce the best radiographic results while minimizing exposure dose. D. Be careful about the decimal point on the mAs display in order to acquire intended mAs values. E. For the indoor radiography for the p atient in t he bed, ask visito rs to leave the room for a wh ile and keep safe distance from the patient. F. Pay attention to proper management of the equipment and careful maintenance. G. Cumulative radiation dose should not exceed the recommended amount. Ask the guide and advice of experts for the requirement of additional protective devices at frequent use of high kV and mAs. WARNING Please check if the voltage and current of the power supply observe the specifications written on t he system labels, attached t o backside of machine body. The levels of voltage should move within ±10% of the standard values.



CAUTION Please check up all the line connections before the operation and x-ray exposures of machine.



2) Collimator Check-up A. Place the loaded cassette on the patient’s backside. B. Arrange the SID (Focal Spot to Image Receptor Distance) using a scale located on the side of the collimator. C. Turn on the collimator lamp and laser pointer by pushing collimator/laser on switch. D. Adjust the size of the x-ray beam in accordance with the film size using the x-ray field adjustment knobs. Turn off the collimator and laser pointer after 30 seconds on the timer.



18/47



Product Manual: POX-100BT



3) Preheat (1) Overview All portable x-ray units require preheat procedure to protect the x-ray tube from an instant surge of current and to secure long and safe operation of the system. Preheat of x-ray tube is required under following situation. A. Initial use or installation. B. Non-use of the equipment over 1 month at low temperature C. Temperature of x-ray tube is below 0℃ (32℉). (Cold tube) (2) Method to Preheat Following preheat procedure is recommended. A. Install the equipment at a safe place for radiation. B. In case outdoor temperature is below 0℃ (32℉), move the equipment indoor and warm it up. C. In order to preheat filament of x-ray tube, exposure x-rays in 15 second time intervals. First: 3 exposures at 50 kV @ 0.2 mAs Second: 3 exposures at 80 kV @ 5.0 mAs Third: 3 exposures at 110 kV @ 5.0 mAs CAUTION Beep Sound will occur during x-ray exposu res. Do n ot press exposure buttons at wait mode after x-ray exposure is made.



19/47



Product Manual: POX-100BT



9. Operation Mode 1) General Mode A. Open the carrying case. B. Connect Power supply cable, communication cable and hand switch as following. Power Supply Cable



Communication Cable



[OP Console]



[Console Side]



Hand Switch Connection



20/47



[Body Side]



Product Manual: POX-100BT



※ CAUTION※ Please use designated electronic parts (Power plug and Hand switch) and be careful not to apply other components. ※ CAUTION※ and



on



power switch represent ON and OFF,



respectively



C. Turn on the power on switch and initial conditions are displayed as follows. (1) Main Console



(2) Sub Console



21/47



Product Manual: POX-100BT



D. Select kV settings by kV up/down switch. The speed of increasing or decreasing values can be faster by keep pressing the buttons. E. Select sec & mAs values by sec & mAs up/down switch. The sec segment displays the real value of sec without separate table. F. After setting the radiography conditions, make the x-ray at preposition by pressing the 1 st stage of hand switch. G. Press the 2nd step of hand switch and make exposure as soon as the ready LED is lighted on. The exposure LED is lighted on during exposures. H. The wait LED is lighted off for 0.5 seconds after the x-ray exposures. I.



Radiography conditions (kV & mAs) can be operated by sub console as well as the light on/off of collimator lamp.



2) APR Mode A. Select APR Switch. B. Select the size of patient by APR Size Select Switch. C. Select the subject region by APR region select switch. D. Select the radiography direction by APR direction select switch. E. Select wanted values by density up/down switch. F. APR Storage 1) Users can save 30 (Body Size 2 X Region 5 X Direction 2) data for APR. 2) If you press Store Switch after changing kV/mA/Sec, the changed condition is saved.



※ NOTICE※ APR data can not be saved when OVER LED is lighted on or Density is not Zero. In this case, change radiography condition (kV, mA, Sec) and density and resave.



※ NOTICE※ Under APR mode, the last saved technique values are automatically selected and display and saving of data is available only at x-ray exposures.



※ NOTICE※ As explained at previous pages, the exposure switch should be operated by 2 stages. It, however , instant exposure is available when the filament board is heated enough.



22/47



Product Manual: POX-100BT



23/47



Product Manual: POX-100BT



PART II. SERVICE MANUAL



1. NOTICE 1) If the output calibration is needed after periodical inspection or repair, please observe following rules and methods. CAUTION The ionized radiation could be dangerous to the operator u nless f ollowing safety measures are strictly observed.



2) This manual is designed to ensure correct use and operation o f POX-100BT. Please read all the lines thoroughly before using this equipment. 3) Incorrect use and operation exceeding described conditions in this manual may cause d amage of the machine and shorten i ts life. Particular attention must be paid to all the warnings, cautions and notes incorporated here in. 4) This equipment should be used only by the legally qualified persons and practitioners. 5) Regarding the AC cable and software, only the one developed and supplied by ACOMA'S should be used. 6) No responsibility is taken by ACOMA'S for any infringement of patients or other right s of the third parties which many results from the use of this manual. 7) Always keep the manual at hand for your reference.



24/47



Product Manual: POX-100BT



2. Regular Maintenance 1) Overview Portable x-ray equipment should be inspected regularly according to following instruction. . 2) Periodical Inspection 1) Every 6 Month A. Check all visual displays: display readout windows (FND), LEDs and other indication. B. Check all functions: Operation switches, collimator, laser pointer and programme. C. Check all status of calibration according to section 4 of service manual. D. Test the thermal switch of x-ray tube inside the equipment. Remove the connection to tube thermal swit ch and check the display of error message. Test normal operation of x-ray exposure as well. E.



Execute additional test required by laws and official regulation.



2) Every 12 Month A. Check damages on the outer case: Operation console including buttons. B. Check cable connections to main body: hand switch cable, collimator cable and other connections. C. Open the top cover of equipment and check visible damage: missing or loosened ground connection, D. Open the system cover and check the v isible damage: missing or loose groun d connections, oil leakage, damaged cables and etc.



WARNING This equipment should be handled and operated by the legally qualified persons and practitioners only. WARNING Wait at least 5 minutes after turn off in order to discharge capacitors.



25/47



PLUG A C MALE



INLET



2



SW1 CARLING S/W



1



ON 보조 접점



1



1



SW2



2



2



HANDLE S/W



1 2 3 4 5 6 7 8 9 10 11 12 13 14 15



1 2 3 4 5



CN11



53258-05



CN1



CN12



CN2 53258-03 3 2 1



F1 100A



BT4 12V24AH



BT3 12V24AH



BT2 12V24AH



BT1 12V24AH



NO.



C ON TEN TS



1 2 3 4 5 6



CN21



35312-06



TX RX EXP GND GND +15V



CON1 53258-02



D a t e



100 mAS LAMP



Rev . By



TB1



58.5V RTN



Checked By



Approv ed By



TB2



58.5V+



DRV CON5 53258-03



HT CONTROL & REMOTE OP



110 kV



2009-08-14



Date



Designed By



Date:



Size A3



Title



Friday , August 14, 2009



Document Number POX-100BT-S001



POXBILE-100BT BLOCK DIAGRAM



LAMP



Sheet



1



COLLIMATOR



HV TANK



of



1



Rev B



3. Block Diagram



A



B



C



D



E



Battery



623PCB4-B2



4 3 2 1



OP BOARD



PHONEJACK 6 P4C *2



Batt Charger



BODY



D-SUB



4 3 2 1



623PCB4-B



1 1 2 1 2 1 2



26/47 2



HAND SWITCH



1 2



HEAD



1 2 3



OP



Product Manual: POX-100BT



Product Manual: POX-100BT



4. Calibration procedures ▣ Preparation Following instrument and tools are required. a. Tools: Philips screwdriver No.1 & 2, Straight slot screwdriver b. Testers: DVM, Digital Oscilloscope (Tektronics TDS 3032 or other device with higher calibers. 1) Power Supply -The power switch should be at off condition. -Connect power cables to power consent. -Power on the switch. 2) Calibration method is composed of three procedures. -Setting and adjusting tube kV. -Setting and adjusting tube mA -Adjusting filament preheat. A. Setting and adjusting tube kV (Pre shipment adjustment) -Turn off the power switch. -Connect Oscilloscope to HT Controller inside machine. -Chanel 1: connect the probe of oscilloscope to TP11 (IP F/B) and GND (TP16, 17 or 18). -Chanel 2: connect the probe of oscilloscope to TP10 (EP F/B) and GND (TP16. 17 or 18). Chanel 1: mA, 1V = 10mA Chanel 2: kV, 1V = 25kV -Turn on the power switch. -Set the conditions a s 4 0kV and 0. 1mAs (2 0mA) and observe mA waveforms at Channel 1 of Oscilloscope. -If the value at channel 1 is not 2V (20mA), adjust the mA output with VR1 of HT board. And observe if the waveform of Channel 2 is 40kV (1.625V). -Set the conditions as 1 00kV and 0.2 mAs (4 0mA) and observe mA waveforms at Channel 1 of Oscilloscope. -If the value at channel 1 is not 4 V (40mA), adjust the mA output with VR1 of HT board. And observe if the waveform of Channel 2 is 100kV (4V). ※ CAUTION ※ As the kV value of channel 2 is automatically adjusted by standard parameters, Additional calibration is not required.



27/47



Product Manual: POX-100BT



B. Setting and adjusting tube mA -Turn off the power switch. -PCB HT board is located under control panel. -All DIP swit ches o n PCB HT b oard a re at of f con ditions. (Note: DIP switch i s lo cated i n this pl ace normally. -Chanel 1: connect the probe of oscilloscope to TP11 (IP F/B) and GND (TP16, 17 or 18). -Chanel 2: connect the probe of oscilloscope to TP10 (EP F/B) and GND (TP16. 17 or 18). Chanel 1: mA, 1V = 10mA Chanel 2: kV, 1V = 25kV -Turn on the power switch. -Set the conditions as 4 0kV and 0. 2mAs (1 0mA) and observe mA waveforms at Channel 1 of Oscilloscope. -If the value at channel 1 is not 4V (40mA), adjust the mA output under calibration mode. -If you press “kV Down”, “APR S”, “APR L” switch for 2 seconds, it is enter into calibration mode.



Calibration mode is displayed as below. “C00” on the kV segment is kV calibration code and “20” on the mA segment is mA calibration value. Refer to content of each menu at table 1.



28/47



Product Manual: POX-100BT



The kV display segment of FND is displayed as CXX.



Calibration Code



Calibration Data



Press mA up switch (Refer to the OP sheet) and it displays as follows. C01 is calibration code of tube current (40mA at 40-50kV) and 40 is calibration data of tube current. .(40mA from 40 to 50kV) -Select mA with mAs up/down switch (Refer to the OP sheet) -Press the ‘Store Key’ longer than 1 second after application (Refer to the OP sheet) -Exposure x-rays after saving the data -If the measu red data is not 4V(40 mA), reset t he cali bration dat a. Set the mA value s as thi s method. -Table 1 represents calibration code and data of each tube current vales at 5kW. C. Filament preheat calibration -If the preheat calibration data or VR1 are set in wrong way, over current or low current can occurat X-ray tube. -There are two methods about preheat calibration. Calibration data can be changed by tube conditions. -VR1 can be applied to overall tube current. Filament preheat calibration dat a can be applied to each tube current. -Turn on the power switch. -Under normal mode, press [‘S’+‘L’+‘kV_Down’] Key in case of human and [‘Bird’+‘Dog’+‘kV_Down’] Key in case of veterinary over 2 seconds.



29/47



Product Manual: POX-100BT



⑦ kV Up/Down 스위치를 이용하여 관전류 조정모드에서 필라멘트 예열 조정모드로 변경한다.



P00 is filament preheat calibration code (preheat calibration code and data of 20mA) and 28 is calibration data. -Apply the data by mAs up/down switch. (Refer to the OP sheet) -Press the ‘Store’ key longer than 1 second (Refer to the OP sheet) -Reset the calibration data if any over-shoot or under-shoot is observed. Press kV up switch. (Refer to the OP sheet) and it displays as below. -P01 is filament preheat calibration code and 30 is the calibration data (preheat calibration code and data of 40mA) -Apply the data by mAs up/down switch. (Refer to the OP sheet) -Press the ‘Store’ Key longer than 1 second after application (Refer to the OP sheet) -Exposure x-rays after saving the data. -Reset the calibration data if any over-shoot or under-shoot is observed. -Table 2 represents calibration code and data of each tube current vales at 3kW. -Escape from calibration mode by press [‘M’=‘Cat’] Key one time. [Human]







[Veterinary]



Notice※



Calibration data can be changed by tube conditions.







Notice※



If you press [‘S’=‘Cat’] Key consecutively, it repeats P -> C -> L _ Mode. [ P: Preheat, C: mA for 5kW, L: mA for 3kW ]



30/47



Product Manual: POX-100BT



Posko



m Co., Ltd.



5. Calibration data [Table 1. mA calibration data for 5kW] AC Power



Code



Data



Tube kV



Tube mA



mAs 220V C00



220V C01



40



40kV ~ 50kV



40mA



0.2 ~ 0.3



220V C02



100



40kV ~ 50kV



100mA



0.4 ~ 10



220V C03



80



40kV ~ 50kV



80mA



12 ~ 32



220V C04



60



40kV ~ 50kV



60mA



36 ~ 200



220V C05



20



51kV ~ 60kV



20mA



0.1



220V C06



40



51kV ~ 60kV



40mA



0.2 ~ 0.3



220V C07



80



51kV ~ 60kV



80mA



0.4 ~ 8.0



220V C08



50



51kV ~ 60kV



50mA



9.0 ~ 100



220V C09



40



51kV ~ 60kV



40mA



125 ~ 200



220V C10



20



61kV ~ 70kV



20mA



0.1



220V C11



40



61kV ~ 70kV



40m



0.2 ~ 0.3



220V C12



50



61kV ~ 70kV



50mA



0.4 ~ 8.0



220V C13



40



61kV ~ 70kV



40mA



9.0 ~ 80



220V C14



30



61kV ~ 70kV



30mA



100 ~ 150



220V C15



20



71kV ~ 80kV



20mA



0.1



220V C16



40



71kV ~ 80kV



40mA



0.2 ~ 0.3



220V C17



50



71kV ~ 80kV



50mA



0.4 ~ 8.0



220V C18



40



71kV ~ 80kV



40mA



9.0 ~ 21



220V C19



30



71kV ~ 80kV



30mA



25 ~ 64



220V C20



20



81kV ~ 100kV



20mA



0.1



220V C21



40



81kV ~ 100kV



40mA



0.2 ~ 8.0



220V C22



30



81kV ~ 100kV



30mA



9.0 ~ 30



220V C23



20



81kV ~ 100kV



20mA



32 ~ 80



220V C24



20



101kV ~ 110kV



20mA



0.1



220V C25



40



101kV ~ 110kV



40mA



0.2 ~ 5.0



220V C26



30



101kV ~ 110kV



30mA



6.4 ~ 12



20



101kV ~ 110kV



20mA



15 ~ 64



220V



C27



[Table 2. mA Calibration Data for 2.3kW] AC Power



Code



Data



Tube kV



31/47



Tube mA



mAs



Product Manual: POX-100BT 220V L00



20



40kV ~ 50kV



20mA



0.1



220V L01



40



40kV ~ 50kV



40mA



0.2 ~ 0.3



220V L02



60



40kV ~ 50kV



60mA



0.4 ~ 200



220V L03



20



40kV ~ 50kV



20mA



0.1



220V L04



40



40kV ~ 50kV



40mA



0.2 ~ 0.3



220V L05



50



51kV ~ 60kV



50mA



0.4 ~ 100



220V L06



40



51kV ~ 60kV



40mA



125 ~ 200



220V L07



20



61kV ~ 70kV



20mA



0.1



220V L08



40



61kV ~ 70kV



40mA



0.2 ~ 80



220V L09



30



61kV ~ 70kV



30mA



100 ~ 150



220V L10



20



71kV ~ 80kV



20mA



0.1



30



71kV ~ 80kV



30mA



0.2 ~ 64



220V L12



20



81kV ~ 100kV



20mA



0.1



220V L13



30



81kV ~ 100kV



30mA



0.2 ~ 30



220V L14



20



81kV ~ 100kV



20mA



32 ~ 80



220V L15



20



101kV ~ 110kV



20mA



0.1 ~ 64



220V L1



1



[Table 3. Filament preheat calibration code and data] AC Power



Code



Data



220V P00



28 20mA



220V P01



30 30mA



220V P02



31 40mA



220V P03



31 50mA



220V P04



33 60mA



Tube Current



220V



P05 35



80mA



220V



P06 38



100mA



※ No tice ※ Do not use the machine in case of frequent errors. Please make contact to ACOMA"S service team or authorized dealers for trouble shootings.



32/47



Product Manual: POX-100BT



6. Radiation dose data [Table 4. Radiation dose data]



29.204.3 T



able: Measured LEAKAGE RADIATION in the LOADING STATE Conditions of LOADING



Points of



Mode of



measurement



operation



s



CURRENT,



Allowed limit of



LEAKAGE



X-RAY TUBE



LEAKAGE



RADIATION



VOLTAGE, (kV)



RADIATION



(mGy/h)



(mA) or



CURRENT TIME



P



PRODUCT, (mAs)



(mGy/h)



Front -



20 mAs



110



1,0



0,24



Rear -



20 mAs



110



1,0



0,05



Left -



20 mAs



110



1,0



0,75



Right -



20 mAs



110



1,0



0,74



Top -



20 mAs



110



1,0



0,55



Bottom -



20 mAs



110



1,0



0,71



Front-left -



20 mAs



110



1,0



0,54



Front-right -



20 mAs



110



1,0



0,19



Rear-left -



20 mAs



110



1,0



0,03



Rear-right -



20 mAs



110



1,0



0,02



Supplementary information: - Model: POX-100BT, Serial No.: PTH100-0809-001 - Test Instrument: RADCAL, 10x5-180/1015



33/47



Product Manual: POX-100BT



7. Error code and resolution 1) Error at stand-by Without hand switch op eration, if hardware is over th an EP=10kV, IP=5mA, FIL=0.1V F/B Check, it is in error. Error Code



Contents



Measures



FILAMENT FEEDBACK OVER



Calibration or adjust VR.



Err. 11



kV FEEDBACK OVER



Calibration or adjust VR.



Err. 12



mA FEEDBACK OVER



Calibration or adjust VR.



Err. 10



2) Error at ready With ready switch on, check the filament conditions and if hardware is over than FIL=0.35V F/B Check, it is in error. Error Code Err. 14



Contents



Measures Calibration or adjust VR.



FILAMENT FEEDBACK LOW ERROR



3) Error at Exposures EP, lP Low Error: Lower than minimum Value (30kV). EP,IP Over Error: Higher than maximum Value (120kV). EP,IP State Low Error: Lower than output kV (-10kV) EP,IP State Over Error: Higher than output kV (+10kV) Error Code



Contents



Measures



Err. 15



kV FEEDBACK LOW DETECT ERROR AT EXPOSURE



Calibration or adjust VR.



Err. 16



mA FEEDBACK LOW DETECT ERROR AT EXPOSURE



Calibration or adjust VR.



Err. 17



kV FEEDBACK OVER DETECT ERROR AT EXPOSURE



Calibration or adjust VR.



Err. 18



mA FEEDBACK OVER DETECT ERROR AT EXPOSURE



Calibration or adjust VR.



-10kV kV FEEDBACK LOW DETECT ERROR AT



Calibration or adjust VR.



Err. 19 Err. 20 Err. 22 Err. 23



EXPOSURE +10kV kV FEEDBACK LOW DETECT ERROR AT



Calibration or adjust VR.



EXPOSURE -10kV mA FEEDBACK LOW DETECT ERROR AT



Calibration or adjust VR.



EXPOSURE +10kV mA FEEDBACK LOW DETECT ERROR AT



34/47



Calibration or adjust VR.



Product Manual: POX-100BT



EXPOSURE 4) Inter_Lock Error Error Code Err. 21



Contents



Measures



Tank Thermal S/W Check



35/47



PRODUCT MANUAL: POX-100BT



8. Schematics 1) Battery Charger



PRODUCT MANUAL: POX-100BT



2) DRV Board



2



PRODUCT MANUAL: POX-100BT



3) HT Controller



3



PRODUCT MANUAL: POX-100BT



4) HT Controller



4



PRODUCT MANUAL: POX-100BT



5) HT Controller



5



PRODUCT MANUAL: POX-100BT



6) HT Controller



6



PRODUCT MANUAL: POX-100BT



7) Console Board



7



PRODUCT MANUAL: POX-100BT



8) Console Board



8



PRODUCT MANUAL: POX-100BT



9) Console Board



9



PRODUCT MANUAL: POX-100BT



10) Console Board



10



PRODUCT MANUAL: POX-100BT



11) Console Board



11