Service Manual V8800 Ventilator [PDF]

  • 0 0 0
  • Suka dengan makalah ini dan mengunduhnya? Anda bisa menerbitkan file PDF Anda sendiri secara online secara gratis dalam beberapa menit saja! Sign Up
File loading please wait...
Citation preview

 



 



Service Manual‐ V8800 Ventilator 



V8800 Critical Care Ventilator Service Manual



  P a g e  | 1 



 



 



P a g e  | 2 



 



Service Manual‐ V8800 Critical Care Ventilator 



 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



Contents 



1  Safety Information ..................................................................... 6  2  Notices ......................................................................................... 9  2.1  2.2  2.3  2.4  2.5  2.6  2.7 



Warranty Notice ........................................................................................................ 9  Copyright Notice ....................................................................................................... 9  EMC Notice ................................................................................................................ 9  MRI Notice ............................................................................................................... 10  Intended Use Notice ................................................................................................ 10  IEC Classification ...................................................................................................... 11  Declaration of Conformity Notice ........................................................................... 11 



3  Equipment Symbols ................................................................. 12  4  Introduction ............................................................................. 15  4.1  4.2 



General description ................................................................................................. 15  Abbreviations and Definitions ................................................................................. 16 



5  Theory of Operation ................................................................. 18  5.1 



Pneumatic System ................................................................................................... 18  5.1.1  Pneumatic Theory ........................................................................................... 18  5.1.2  Main Pneumatic Assemblies Overview ........................................................... 19  5.2  Electrical system ...................................................................................................... 26  5.2.1  Host inside View 1 ........................................................................................ 26  5.2.2  Host inside View 2 ........................................................................................ 28  5.2.3  Expiratory Valve heating Module view ....................................................... 29  5.2.4  Rear panel view ............................................................................................ 30  5.2.5  Display inside view .......................................................................................... 32  5.2.6  Monitor back plastic shell ............................................................................ 33  5.2.7  Power Management Supply board Components Overview ............................ 34  5.2.8  Ventilation Unit Motherboard Components    Overview ................................ 35  5.2.9  Monitor Control Board Components Overview .............................................. 36  5.2.10  Monitor interface board Components Overview ............................................ 37  5.2.11  Stepper motor driver board Components Overview ...................................... 38  5.2.12  BDU Isolate Board Components Overview ...................................................... 39  5.2.13  Touch Screen control board ................................................................ 40  5.2.14  LCD Backlight inverter board .............................................................. 40  5.2.15  Single Board Computer Overview ................................................................... 41  5.2.16  System wiring diagram .................................................................................... 43  5.2.17  Main board ..................................................................................................... 50  5.2.18  Power supply board......................................................................................... 53  5.2.19  UIC board......................................................................................................... 55  5.3  Software System ...................................................................................................... 57  P a g e  | 3 



 



 



5.3.1  5.3.2  5.3.3 



Service Manual‐ V8800 Critical Care Ventilator 



Software Units and Their Relationship ............................................................ 57  Structure of the Software System ................................................................... 58  Hardware and Software Interface ................................................................... 59 



6  Disassembling and assembling ............................................... 59  6.1  6.2 



General .................................................................................................................... 59  Disassembling of the main assemblies .................................................................... 61  6.2.1  Disassembling of the display from the host ................................................. 61  6.2.2  Disassembling of the host from the trolley .................................................. 62  6.2.3  Disassembling of the air compressor from the trolley .................................... 63  6.2.4  Disassembling of the humidification assembly from the host ........................ 65  6.3  Disassembling of maintenance spare parts............................................................. 66  6.3.1  Disassembling of the gas supply inlet filter ..................................................... 66  6.3.2  Disassembling of oxygen sensor ...................................................................... 67  6.3.3  Disassembling of W15 cable ............................................................................ 68  6.3.4  Disassembling of pressure sensor board ......................................................... 69  6.3.5  Disassembling of W14 cable ............................................................................ 70  6.3.6  Disassembling of Stepper motor driving board ............................................... 72  6.3.7  Disassembling of TSI mass flow sensor ........................................................... 73  6.3.8  Disassembling of the W4 cable ....................................................................... 74  6.3.9  Disassembling of step motor ........................................................................... 74  6.3.10  Disassembling of PEEP valve ........................................................................... 76  6.3.11  Disassembling of diaphragm and scaleboard .................................................. 76  6.3.12  Disassembling of W24 cable ............................................................................ 79  6.3.13  Disassembling of W26 cable ............................................................................ 79  6.3.14  Disassembling of power supply AC/DC module .............................................. 80  6.3.15  Disassembling of components in the display .................................................. 81  6.3.16  Disassembling of electric panel assembly ....................................................... 82  6.3.17  Disassembling of BDU assembly...................................................................... 83 



7  User maintenance ....................................................................... 84  7.1.1  Basic guidelines ............................................................................................... 85  7.1.2  Disinfection and Sterilization ........................................................................... 85  7.2  Regular Maintenance .............................................................................................. 87  7.2.1  Maintenance Principles ................................................................................... 87  7.2.2  User Maintenance ........................................................................................... 87  7.2.3  Service life of product/accessories .................................................................. 88  7.3  Maintenance in Operation and Transportation ....................................................... 89  7.3.1  Transportation ................................................................................................. 89  7.3.2  Storage ............................................................................................................ 90  7.4  Battery maintenance ............................................................................................... 90  7.4.1  Battery specification ........................................................................................ 90  7.4.2  Precautions ...................................................................................................... 91  7.4.3  Battery storage ................................................................................................ 91  P a g e  | 4 



VL9004 Rev 1 



 



 



 



7.5  7.6  7.7  7.8  7.9 



Service Manual‐ V8800 Ventilator 



7.4.4  Battery replacement ....................................................................................... 91  Disposal ................................................................................................................... 92  Manufacturing techniques and materials: .............................................................. 92  Free Obligations: ..................................................................................................... 93  Security, reliability and operating condition: .......................................................... 93  Return ...................................................................................................................... 93 



8  Pre‐Use Test and calibration ........................................................ 94  8.1.1  Technical Test .................................................................................................. 94  8.1.2  AC/Battery Test ................................................................................................ 94  8.1.3  Gas Supply Test ................................................................................................ 94  8.1.4  Oxygen Sensor Test ......................................................................................... 94  8.1.5  Leak Test .......................................................................................................... 95  8.1.6  Flow Sensor Test .............................................................................................. 95  8.1.7  Pressure Sensor Test ........................................................................................ 95  8.1.8  Safety Valve Test .............................................................................................. 95  8.1.9  Breath Circuit Test ........................................................................................... 95  8.1.10  etCO2 Sensor Test ........................................................................................... 95  8.2  Calibration ............................................................................................................... 96  8.2.1  Pressure Sensor ............................................................................................... 96  8.2.2  Flow Sensor ..................................................................................................... 96  8.2.3  O2 Sensor ........................................................................................................ 96  8.2.4  CO2 Sensor ...................................................................................................... 96  8.2.5  Inspiratory Valve .............................................................................................. 96  8.2.6  Expiratory Valve ............................................................................................... 97  8.2.7  Touch Screen ................................................................................................... 97 



9  Troubleshooting .......................................................................... 97  9.1 



Technical Error ......................................................................................................... 97  9.1.1  Technical Test Error Code ................................................................................ 97  9.1.2  Other Errors ................................................................................................... 100  9.1.3  Service Tools .................................................................................................. 101 



10 



Appendix A Contact & Ordering Information ........................ 102 



11 



Appendix B    Diagrams and Schematics ............................... 103 



12 



Appendix C Specifications .................................................... 104 



13 



Appendix D spare parts ........................................................ 107 



 



P a g e  | 5 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



1 Safety Information  The operator of this ventilator must recognize their full responsibility for choosing appropriate ventilation settings to ensure proper ventilation and patient safety. The responsibility for the selection of the appropriate level of patient monitoring depends solely on the equipment operator. All the monitoring information is for reference only; it should not be used as the sole basis for therapeutic or diagnostic decisions. Whenever a patient is connected to the ventilator, constant attention by qualified medical personnel is required in order to provide immediate corrective action in case of a malfunction and alarm occurrence. Terms This manual uses three special indicators to convey information of a specific nature. They include:



WARNING:



Identify conditions or practices that could result in serious adverse reactions or potential safety hazards.



CAUTION:



Identify conditions or practices that could result in damage to the ventilator or other equipment



NOTE:



Identify supplemental information to help you better understand how the ventilator works.



P a g e  | 6 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



Warnings, cautions, and notes Please take the time to familiarize yourself with the following safety considerations, special handling requirements, and regulations that govern the use of the V8800ventilator. To ensure proper servicing and avoid the possibility of physical injury, only qualified personnel should attempt to service or make authorized modifications to the ventilator. The user of this product shall have sole responsibility for any ventilator malfunction due to operation or maintenance performed by anyone not trained by Oricare. WARNING!  Back‐up ventilation with an independent manual ventilation device    If a fault is detected in the V8800ventilator so that its life-support functions are no longer assured, start ventilation using an independent ventilation device (resuscitation bag) without delay, if necessary with PEEP and / or increased inspiratory O2 concentration. To avoid an electrical shock hazard while servicing the ventilator, be sure to remove all power to the ventilator by disconnecting the power source and turning off all ventilator power switches. To avoid a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (e.g., flammable anesthetics and/or heaters) away from V8800ventilator and oxygen hoses. Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials such as grease or oils. Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen. In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen supply, facility power, and backup power source. Follow your hospital infection control guidelines for handling infectious material. Oricare recognizes that cleaning, sterilization, sanitation, and disinfection practices vary widely among health care institutions. It is not possible for Oricare to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of cleaning, sterilization, and other practices carried out in the patient care setting. P a g e  | 7 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



Patients on life-support equipment should be appropriately monitored by competent medical personnel and suitable monitoring devices. V8800 ventilator is not intended to be a comprehensive monitoring device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment. For a thorough understanding of ventilator operations, be sure to thoroughly read this manual before attempting to operate the system. Before activating any part of the ventilator, be sure to check the equipment for proper operation and, if appropriate, run PUT (pre-use test) as described in this manual. Do not use sharp objects to make selections on the LCD touch screen or panel. Check the ventilator periodically as outlined in this manual; do not use if defective. Immediately replace parts that are broken, missing, obviously worn, distorted, or contaminated. An alternative source of ventilation should always be available when using V8800ventilator, such as manual respiratory equipment. Do not connect items which are not specified as part of the system. Do not place containers of liquids (such as humidifier water reservoirs) on top of or above ventilator. Liquids getting into the ventilator can cause equipment malfunction with the risk of patient injury. Do not use in conjunction with magnetic resonance imaging (MRI). Equipment malfunction with the risk of patient injury may result.



P a g e  | 8 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



2 Notices  2.1 Warranty Notice  Do not make any service repairs on this equipment during the states warranty period. Any unauthorized work immediately voids the warranty. Oricare will not be liable for any repairs attempted by the owner. Any such attempted repairs other than specified no warranty repairs void the warranty.  



2.2 Copyright Notice    Copyright Notice: Inc. This work is protected the copyright law and is the sole property of the Company. No part of this document permitted to be copied or otherwise reproduced, translated into other languages, or stored in any electronic information retrieval system, without the prior written consent of the Company.



2.3 EMC Notice  This equipment can radiate radio frequency energy ,as well can be interfered by electromagnetic If not installed and used in accordance with the instructions



mentioned



in the manual, electromagnetic interference may result. The equipment has been tested and found to comply with the limits set forth in EN60601-1-2 for Medical Products. These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments described in this manual. The ventilator has been tested to conform to the following specifications: MIL-STD-461D:



1993,



MIL-STD-462D:1993,



1000-4-2:1994, IEC 1000-4-3:1994, P a g e  | 9 



EN



55011:2007+A2:2007,



IEC



 



 



Service Manual‐ V8800 Critical Care Ventilator 



IEC 1000‐4‐4:1994, IEC 1000‐4‐5:1994, QUASI‐STATIC:1993  This ventilator is also designed and manufactured to comply with the safety requirements of IEC 60601-1, IEC 60601‐2‐12, CAN/CSA C22.2 No. 601.1‐M90, and UL 60601‐1. 



2.4 MRI Notice  This equipment contains electromagnetic components whose operation can be affected by intense electromagnetic fields. Do not operate the ventilator in a MRI environment or in the vicinity of high-frequency surgical diathermy equipment, defibrillators, or short-wave therapy equipment. Electromagnetic interference could disrupt the operation of the ventilator.



2.5 Intended Use Notice  The V8800 Critical Care Ventilator is intended for patients ranging from pediatric to adult, and for use in a wide variety of clinical conditions. Specifically, the V8800 Critical Care Ventilator is applicable for adult and pediatric patients weighing at least 5 kg for adult and 3.5 Kg for child (11 lbs.), who require the following types of ventilatory support: Positive Pressure Ventilation, delivered invasively (by ET or Tracheotomy tube) or non-invasively (by mask) via Assist/Control, SIMV, CPAP and other modes of ventilation. The V8800Critical Care Ventilator is intended for use in hospital and hospital-type facilities. It may be used during intra-hospital transport provided that electrical power and compressed gas are supplied.



WARNING: The V8800 is not designed for use in an MRI environment. Do not use the V8800 near an MRI machine or injury or equipment damage could result.



WARNING: The user of the Ventilator must be professional and trained.



P a g e  | 10 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



2.6 IEC Classification  Type of Equipment: Medical Equipment, Class 1 type B Adult/Pediatric Lung Ventilator  



2.7 Declaration of Conformity Notice  This medical equipment complies with the Medical Device Directive, 93/42/EEC, and the following Technical Standards, to which Conformity is declared: EN60601‐1                                                                                                    EN60601‐1‐2  ISO 9001, EN 46001                                                   



P a g e  | 11 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



3 Equipment Symbols    Warnings and



Cautions indicate all the possible dangers in case of violation



of the stipulations in this manual. Refer to and follow them.



Warnings Indicates potential hazards to operators or patients.



Cautions Indicates potential damage to equipment. Instead of illustrations, other symbols may also be utilized. Not all of them may necessarily appear in the equipment and manual. The symbols include: Instead of illustrations, symbols may be utilized. Not all of these symbols may necessarily appear on the equipment or in this User manual. The symbols include: 



On (Power)



Protection Class Type B



Attention: Please read Off (Power)



documents provided with the equipment



Attention: Please refer to Product Specifications and



Dangerous voltage



Relevant Standards



Protective earth ground



loudspeaker



EQUIPOTENTIAL Address of production connection



Lock



P a g e  | 12 



Date of production



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



Unlock



Serial Number



Inspiratory hold



Expiratory hold



Nebulization



Manual inspiration



Intelligent increase of Waveform freeze oxygen



P a g e  | 13 



Standby



int.



Internal Battery



AC power



ext.



Extended Battery



USB device



Already online



Prompt message



Pressure trigger



Flow trigger



Manual trigger



Adult



Child



NIV modes



Invasive modes



Main Menu



Alarm Silence Key



 



 



Protection Class Type BF



P a g e  | 14 



VL9004 Rev 1 



Service Manual‐ V8800 Critical Care Ventilator 



Disposal of Waste



 



 



 



Service Manual‐ V8800 Ventilator 



4 Introduction  4.1 General description  The V8800 Critical Care Ventilator System consists of two required Components: a Main Control Unit and a Graphical User Interface (GUI). Optional Components available for the V8800 Critical Care Ventilator system are: Ventilator Cart, Air Compressor, Drawer & Extended Battery Backup Assembly, Gas Cylinder Transport Cart, Patient Circuit Positioning Arm, and Hose Assemblies for O2 and Air, CO2 monitoring subsystem.







Main Control Unit 







Humidifier   







Air Compressor 







connector 







Nebulizer connector







Y‐piece 







Patient Circuit Positioning Arm







Mask 







Test Lung 



11



Patient Circuit 



12



Water trap 



10  Nebulizer tube 



P a g e  | 15 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



4.2 Abbreviations and Definitions  CPAP 



Continuous Positive Airway Pressure (Set value, hereinafter “S”) 







Breath rate (frequency) in bpm, i.e. ventilation times per minute (S) 



fspont 



ftotal 



Patient’s  spontaneous  respiratory  frequency  (Measured  value,  hereinafter “M”)  Total  breath  rate,  i.e.  the  sum  of  breath  rate  f  and  spontaneous  breath  rate fspont (M) 



O2 



Inspiratory O2 concentration (S & M) 



I : E 



The ratio of Inspiration to Expiration (M) 



MV 



Expiratory minute volume (M) 



MVspont 



Spontaneously breathed minute volume (M) 



MVleak 



Leakage minute volume (M) 



Paw 



Patient airway pressure (M) 



PEEP 



Positive  End‐Expiratory  Pressure,  which  can  improve  the  patient’s  oxygenation (S & M) 



PEEPi 



Intrinsic Positive End‐Expiratory Pressure (M) 



Pinsp 



Upper pressure level in PCV mode (S) 



Pmean 



Mean  airway  pressure.  This  value  is  updated  at  the  end  of  the  last  respiratory cycle, hence, is a continuous average (M) 



Ppeak 



Airway pressure peak value during one ventilatory cycle (M) 



Pplat 



End‐inspiratory airway pressure (M) 



Pmin 



Minimum airway pressure (M) 



P a g e  | 16 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



Psens 



Pressure sensitivity (S) 



Psupp 



Pressure support (S) 



Phigh 



Upper pressure level in BIVENT and APRV (S) 



Plow 



Lower pressure level in BIVENT and APRV (S) 



T Imax 



Maximum inspiratory time (S) 



T insp 



Inspiratory Time (S) 



T pause 



 



Inspiratory Pause Time, to increase the inspiratory time to improve the  patient’s oxygenation (S)  Trigger by flow rate (S) 



VT 



Tidal volume of mechanical ventilation (S) 



VTE 



Expiratory tidal volume (M) 



VTI 



Inspiratory tidal volume (M) 



Esens 



Expiratory trigger sensitivity (S) 



ETCO2 



End‐expiratory CO2 concentration (M) 



WOB 



Work of breathing (M) 



R*C 



Time constant (M) 



Leak% 



Leakage percentage (M) 



Cdyn 



Dynamic compliance (M) 



Cstatic 



Static compliance (M) 



Rinsp 



Inspiratory resistance (M) 



Rexp 



Expiratory resistance (M) 



Elastic 



Elastic resistance (M) 



 



P a g e  | 17 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



5 Theory of Operation  5.1 Pneumatic System    5.1.1 Pneumatic Theory 



  P a g e  | 18 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



                         Figure 6‐1 Pneumatic Theory 



5.1.2 Main Pneumatic Assemblies Overview  5 6 







7  8 



3







1                                        Figure 6‐2 Main Airway Assemblies sketch view    1



Gas buffer module



2



Supply gas module



3



Pressure Regulator module



4



Inspiratory module



5



Electronic board module



6



Oxygen mixer module



7



PEEP valve



8



Expiratory module



  The oxygen mixer module in this cell is different from the oxygen module in the airway theory.  The  oxygen  mixer  module  in  the  airway  theory  includes  the  pressure  regulator,  the  gas  buffer  module and the oxygen mixer module in this cell. 



    P a g e  | 19 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



    6.1.2.1 Pneumatic Connecting Diagram 



  Figure 6‐3    Pneumatic Connecting Diagram     



PU tube¢10 



 



φ3x1.5 silica gel tube 



 



Vapor lock 



 



φ2x1.5 silica gel tube 



 



Three‐ entry connecting 



1,2 



Gas buffer module 



3,6 



Inspiratory module 



4,7 



Supply air module 







Oxygen mixer module 







Expiratory module 







PEEP valve 



10 



Circuit module 



     



P a g e  | 20 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



    6.1.2.2 Supply Gas Module Overview 



5  6  4  3  2      



 7



9







8



                           10       Figure 6‐4 Supply Gas Module      item   



Description  Gas flow direction 







Oxygen connector, connecting with oxygen supplies. 







Air connector, connecting with external air supplies or air compressor. 







Inlet filter, collecting any condensed water from the air or oxygen gas 







Oxygen sensor, monitoring the oxygen density of mixed gas. 







Pressure sensor, monitoring the pressure of inlet air or oxygen gas. 







Solenoid  valves (the  W15  cable),  carrying  out  self‐inspection  in  the  start‐up  of  the  ventilator. 







Connector, connecting with gas drain tube 







Connector, connecting with the inspiratory module 







Gas drain connectors 



10 



Drain filter 



 



P a g e  | 21 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



    6.1.2.3 Pressure Regulator Module Overview 







    



Air 



                     Oxygen gas    



3











Figure 6‐4 Pressure Regulator Module    item   



 



Description  Gas flow direction 







Precision air pressure regulator, regulating gas pressure to 0.16MPa. 







Inlet connector, connecting with the gas flow export of the supply air module. 







Precision Air Relay, regulating gas pressure to 0.16MPa. 







Export connector, connecting with the gas flow inlet of the oxygen mixer module. 



  6.1.2.3 Oxygen Mixer Module Overview 



3



4 5



1        2    Figure 6‐5 Oxygen Mixer Module   



P a g e  | 22 



 







VL9004 Rev 1



6



 



 



item   



Service Manual‐ V8800 Ventilator 



Description  Gas flow direction 







Gas flow inlet, connecting with gas flow export of the pressure regulator module 







Export connector, connecting with the connector of the gas buffer module 







Step motor driving board, controlling mixed gas oxygen density.   







Step motor 







Solenoid  Valves  (the  W14  cable),  controlling  whether  affording  the  mixed  gas  to  nebulizer and supplying gas pressure to PEEP valve. 







Connector, connecting with nebulizer pipe. 







Connector, connecting with PEEP valve 



  6.1.2.4 Gas Buffer Module  4 



3  2 



  







 



5



        Figure 6‐6 Gas Buffer Module    item 



Description 







Connector, connecting with gas drain tube in the supply air module. 







Solenoid  Valves  (the  W13  cable),  opening  the  connector  for  item  1  to  exhaust  when  the pressure in the gas cylinder gets to the appointed value. 







Safety  valve,  exhausting  gas  when  the  pressure  in  the  gas  cylinder  gets  to  the  appointed va 







Pressure sensor, monitoring the pressure of the gas cylinder. 







Connector, connecting with inspiratory module and oxygen mixer module 



P a g e  | 23 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



    6.1.2.5 Inspiratory Module Overview 



6 4



5



8



7







9  2    Figure 6‐7 Inspiratory Module  1  item 



Description 



 



Gas flow direction 







Connector, connecting with the export connector in the supply gas module. 







Connector, connecting with export in the gas buffer module. 







Proportional valve(the W4 cable), adjusting the gas volume. 







Flow sensor 1, monitoring the volume of the gas 







Connector, connecting with the pressure sensor of circuit module 







Low connector of flow circuit board 







High connector of flow circuit board 







Inspiratory connector, connecting with silica gel tube for patient. 







Connector, connecting with the corresponding tubes of the oxygen mixer module and  the PEEP valve module. 



   



P a g e  | 24 



 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



6.1.2.6 Expiratory Module Overview   



5



4  3  2



     







                    



  Figure 6‐8 Expiratory module 



6   Figure 6‐9    Expiratory valve core component  item   



Description  Movement  direction,  the  component  such  as  figure6‐9  can  take  off  when  turning  the  rod  along the direction.   







Connector, connecting with high connector of flow circuit board in the inspiratory module 







Connector, connecting with export in the PEEP valve module. 







Connector, connecting with high connector of flow circuit board in the inspiratory module 







Nebulizer tube, connecting with the connector in the oxygen mixer module. 







The  rod,  turned  right  along  the  direction  and  then  it  can  come  true  to  take  off  the  component such as figure 6‐9 







The cup collecting water, 



  6.1.2.7 PEEP Valve Module 



1 P a g e  | 25 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



4  2



3                                     Figure    6‐10 PEEP Valve    item 



Description 







Valve bracket 







Pressure control valve. 







Connector, connecting with the export in the oxygen mixer module. 







Connector, connecting with the export in the expiratory module. 



         



5.2 Electrical system      5.2.1 Host inside View 1   



P a g e  | 26 



VL9004 Rev 1



18 



 



17   



Service Manual‐ V8800 Ventilator 



16 







14  15 



13 







12 







1 1







10  9 



5  6 



8 1.Display connector  2.Fan 



3.Three Solenoid Valves 4.BDU Isolate board 5.pressure sensor board 6.One Solenoid Valve 7.Stepper motor Position Sensor 8.Stepper motor 9.Peep Valve 10. Stepper motor driver board 11. Two Solenoid Valves 12. Expiratory flow sensor



P a g e  | 27 



7



 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



13. Airway safety pressure monitoring switch 14. Inspiratory flow sensor 15. Inspiratory Valve 16. Ventilation Unit Motherboard 17. Two pressure sensor board (one for O2 Supply pressure, one for Air Supply pressure,) 18. Power Switch



 



5.2.2 Host inside View 2 



12 



11 



1  10  9



8



   







3 4



    P a g e  | 28 



VL9004 Rev 1



5







7



 



 



Service Manual‐ V8800 Ventilator 



1. Safety pressure monitoring switch     



2. AC‐DC Switching power supply module      3. Battery charging board (Internal Battery)    4. Battery charging board (Extended battery)      5. Power Supply board      6. AC Power filter      7. AC power input circuit breaker    8. Internal Battery    9. pressure sensor board(O2 supply)    10. pressure sensor board(Air supply)    11. BDU board    12. Ventilation Unit Motherboard     



5.2.3 Expiratory Valve heating Module view  



P a g e  | 29 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



 







1. Expiratory  Valve  heating  Module(The  above  is  temperature  sensor,  below is Heating resistor)   



5.2.4 Rear panel view  



P a g e  | 30 



VL9004 Rev 1



 



 







Service Manual‐ V8800 Ventilator 



11 







10 



3  4  9















   







1.Fan  2.Monitor Connector  3. Power Switch  4.O2 sensor  5. AC power input circuit breaker  6. Equipotential terminal      7. AC power input socket  8. Standard Battery circuit breaker  9. Nurse Call Connector    P a g e  | 31 



 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



10.RS232 Connector(Functional reserve)    11.CO2    Module Connector   



5.2.5 Display inside view   



12 



11  10  







  7



 



8



7



   1 



 



 







1. Knob board 



3



4



5



2. Knob LED & Mute Key Board  3. Monitor cable  4. Display connector   



P a g e  | 32 



VL9004 Rev 1



6



 



 



Service Manual‐ V8800 Ventilator 



5. USB connector  6. NET connector  7. Monitor interface board  8. Industrial control board  9. Alarm LED board  10. Touch Screen Control Board  11. Monitor Control Board    12. LCD Backlight inverter board      5.2.6



Monitor back plastic shell 



   







1. Loud Speaker  P a g e  | 33 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



5.2.7 Power  Management  Supply  board  Components  Overview  9



 







8







7







6



 



1. 2. 3. 4. 5. 6. 7. 8. 9.



    4  5   P5 Connector                      Expiratory Valve heating connection  J7  Connector                    Power  indicator  LED  board(Installed  on  the  front  plastic  panel)  connection  J3 Connector                    Software debugger connection  J10,J11 Connector            Battery charging board connection(Standard Battery)  J12,J13 Connector            Battery charging board connection(Extended battery)  J6 Connector                      Power input  (AC‐DC , Batteries)connection  J9 Connector                      Power switch connection  J8 Connector                      Fan connection  J2 Connector                    Ventilation Unit Motherboard connection   



 



P a g e  | 34 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



5.2.8 Ventilation  Unit  Motherboard  Components    Overview   



19 16 



1  2 



15 14







13 4  5 



12



6  7 



11







10



   



18  9



  1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11.



17 



U11                                        Expiratory pressure sampling tube connection  U15                                         pressure sampling tube connection  JP4 Connector                      Inspiratory flow sensor(pin 5‐pin 10) connection  JP6 Connector                      Inspiratory flow sensor(pin 1‐pin 4) connection  JP2 Connector                      Expiratory flow sensor connection  JP12 Connector                    Inspiratory Valve connection  JP13 Connector                    Peep Controller connection  JP16 Connector                    air way safety pressure monitor switch connection  JP10 Connector                    Power Management Supply    board connection  JP1Connector                        O2 sensor connection  JP11  Connector                    Three  pressure  sensor  boards  (O2  supply,  Air  supply  ,          mixture Tank)connection 



P a g e  | 35 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



12. JP9 Connector                Step motor control board(J1 Connector) connection  13. JP14 Connector                six solenoids valves connection  14. JP18 Connector                BDU isolate board connection  15 . JP19 Connector                display monitor    connection  16. Ventilation Unit Motherboard  17. Ventilation Unit daughter board  18. Ventilation Unit daughter board Software debugger connection   



 



5.2.9 Monitor Control Board Components Overview   







3







1



5  11  6 



    7 



9







  1. 2. 3. 4. 5. 6. 7. 8. 9.



P4 Connector            Alarm LED board connection  P1 Connector            LCD Backlight inverter board connection  P6 Connector            knob LED & mute key board connection  P2 Connector            knob board connection  P8 Connector            monitor cable power connection  P9 Connector            monitor cable RS232 Signal connection    P7 Connector          Louder speaker connection  P5 Connector            Single Board Computer CN35    AUDIO‐OUT connection  P12 Connector          Single Board Computer CN38    Power Input connection 



P a g e  | 36 



VL9004 Rev 1



10 



 



 



Service Manual‐ V8800 Ventilator 



10.   P11 Connector        Single Board Computer CN30    COM1/2/3/4Connection  11. P10 Connector          Software debugger connection     



5.2.10   Monitor interface board Components Overview  1 



2



3



 



   







5







    1. 2. 3. 4. 5. 6.



J3 Connector            Single Board Computer CN27    CRT/VGA/LCD connection  J6 Connector            Single Board Computer CN15    USB1/2 connection  J2 Connector            Single Board Computer CN26 LAN connection  J4 Connector            VGA    interface  J5,J7 Connector        USB      interface  J1 Connector            NET    interface   



P a g e  | 37 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



5.2.11 Stepper  motor  driver  board  Components  Overview 







2







  1. J1 Connector            Ventilation Unit Motherboard (J9 Connector) connection  2. J3 Connector            Stepper motor home position sensor connection  3. J2 Connector            Stepper motor connection                           



P a g e  | 38 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



5.2.12 BDU Isolate Board Components Overview   







     







2



1. P1 Connector          Ventilation Unit Motherboard (J18 Connector) connection  2. P2Connector            nurse call connection  3. P3 Connector          CO2 module(upper) & reserve RS 232(lower)                       



P a g e  | 39 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



5.2.13 Touch Screen control board











  1.    JP1 Connector Touch Screen connection  2. J1 Connector      Single Board Computer CN17 USB 3/4 connection     



5.2.14 LCD Backlight inverter board



    1       2     1.    CN1 Connector              Power Input  2.    CN2, CN3 Connector      LCD Backlight lamp     



P a g e  | 40 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



5.2.15   Single Board Computer Overview 



 



  P a g e  | 41 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



  Connector Table:  CN4    TFT LCD I/F  CN5    Inverter Power  CN6 SMBus CN11 IDE CN15 USB1/2 CN17 USB 3/4 CN18 KB/MS CN19 Print Port CN20 RS422/485 CN23 ISA -5V & -12V Input CN26 LAN CN27 CRT/VGA/LCD CN30 COM1/2/3/4 CN32 Reset / Buzzer pin header CN34 AUDIO-IN CN35 AUDIO-OUT CN36 GPIO1 CN37 GPIO2 CN38 Power Input



P a g e  | 42 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



CN39 COM2 Setting CN40 DDR RAM Socket CN41 CF Socket CN42 LVDS LCD I/F CN43 Inverter Power (LVDS) 



 



5.2.16   System wiring diagram      Line BOM size and diagram:  Line size 



BOM size 



Figure 



W1 



122004898    W2 



122004899    W3 



122004900    W4 



122004901    W5 



122004902 



 



P a g e  | 43 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



W6 



122004903 



  W7 



122004904 



  W8 



122004905 



  W9 



122004906 



W10 



122004907 



W11 



122004908 



W12 



122004909 



P a g e  | 44 



VL9004 Rev 1 



 



 



 



W13 



122004910 



W14 



122004911 



W15 



122004912 



W16 



122004913 



W17 



122004914 



W18 



122004915 



W19 



122004916 



P a g e  | 45 



Service Manual‐ V8800 Ventilator 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



W20 



122004917 



W21 



122004918 



W22 



122004919 



  W23 



122004920 



  W24 



122004921 



  W25 



122004922 



 



P a g e  | 46 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



W26 



122004923 



  W27 



122004924 



  W28 



122004925 



  W29 



122004926 



  W30 



122004927 



  W31 



122004928 



 



P a g e  | 47 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



W32   



122004929 



  W33   



122004930 



  W34   



122004931 



  W35 



122004932 



  W36 



122004933 



  W37 



122004934 



 



P a g e  | 48 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



W38 



122004935 



  W39 



122004936 



  W40 



122004937 



W41 



122004938 



W42 



122004939 



W43 



122005313 



    P a g e  | 49 



 



 



 



5.2.17



Service Manual‐ V8800 Critical Care Ventilator 



  Main board   



    Connection and  Leading  function  function  feet 



Connection  and function 



Le ed in g  fe et 



JP4  TSI_FLOW_SEN SOR 



+12V 



function 







TSI_SPI_MOSI 







TSI_SPI_SCLK 







TSI_SPI_SYNC 



JP18  1  BDU  ISOLATE  2  BOARD  3 







TSI_SPI_MISO 







GND 







NC 







NURSE CALL 







GND 







+3.3V 







CO2 MOUDLE RXD 







CO2 MOUDLE TXD 







BDU_TXD2 



GND  +12V 



10  BDU_RXD2  JP6  TSI_FLOW_SEN SOR 



P a g e  | 50 







TSI_FLOW(参数 范 ) 







TSI_TEMP 







+5V 



JP19  DISPLAY  MONITOR 



VL9004 Rev 1 







GND 







+24V 







GND   



 



 



Service Manual‐ V8800 Ventilator 







GND 







+24V 



JP2  EXP_FLOW_SE NSOR 







EXP_FLOW 







GND 







GND 







+24V 







+5V 







PS_RXD 



JP12  INS_VALVE 







INS VALE ‐ 







PS_TXD 







BDU_RXD2 



10  BDU_TXD2  11  BDU_RXD1  12  BDU_TXD1 



JP13  EXP_VALVE 







NC 







1, 2, 3, 4 



+12V 



NC 



5, 6, 7, 8 



GND 







INS VALE + 







+3.3V 







EXP VALE + 



10  GND 







NC 



JP9  STEP_MOTOR _BOARD 



11  MOTOR_STEP  12  MOTOR_DIR 



3   



EXP VALE ‐   



13  MOTOR_HOME  14  MOTOR_SLP  15  GND  16  MOTOR_POS1  17  MOTOR_POS2  18  GND  19  SCL  20  SDA 



JP16  1  SAFTY  AIR  PRESSURE  SENSOR 



SAFTY  PRESSURE  SIGNAL 



AIR  JP14  6 solenoids 







Nebulizer solenoid Valve ‐ 







Nebulizer solenoid Valve + 







Relief solenoid valve ‐ 







Relief solenoid Valve + 







Peep solenoid Valve ‐ 







peep solenoid Valve + 







O2    air    choice  solenoid  Valve ‐ 







O2    air    choice  solenoid  Valve + 







21% 100% solenoid Valve ‐ 



10  21% 100% solenoid Valve +  11  O2 sensor check solenoid Valve  ‐  P a g e  | 51 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



12  O2 sensor check solenoid Valve  +  2 



GND 



  JP1    O2    SENSOR     







O2    SENSOR + 







  O2    SENSOR‐ 



  Caution:check the tab of circuit board connection end       



P a g e  | 52 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



5.2.18 Power supply board 



  Power supply board layout    Connection and  Leading  function  function  feet 



Connection  and function 



Le ed in g  fe et 



function 



P5  Exp_valve_heat ing 



P8  FAN 







GND 







FAN POWER 



3   



SPEED SIGNAL 







SWITCH1 







SWITCH1 







SWITCH2 







SWITCH2 







EXTEND BATT POWER IN 







GND 







EXTEND BATT PLUG IN SIGNAL 



P7  POWER  INDICATION  LED BOARD 







GND 







HEATING  POWER 







GND 







DS18B20_DAT 







+3.3V 







AC_POWER_LED 







CHANGING_LED 







GND 



S3  POWER  SWITCH 



P6 



P a g e  | 53 



 



   



P a g e  | 54 



 



Service Manual‐ V8800 Critical Care Ventilator 







GND 







+24V POWER IN   







STANDARD  BATT  PLUG  IN  SIGNAL 







STANDARD BATT POWER IN 







GND 



 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



5.2.19 UIC board   



  Mouth  function 



and  Leading  function  feet  1,2,3   



+24V   



4,5,6,7 



GND   



P9







PC_TXD3 



RS232







PC_RXD3 







PC_TXD2 







PC_RXD2 







PC_TXD1 



M2







P7  SPEAKER   



P8 POWER INPUT



PS-COM3, BDU-COM1,



Mouth  function 



and  Leading  function  feet  1 



+12 







RED_ALARM_LED 







YELLO_ALARM_LED 







+12V 







GND 







BLACK  ON/OFF 



PC_RXD1 







BRIGHTNESS_CONTR OL 







SPEAKER + 







+5V 







SPEAKER ‐ 



CO2_module-CO



P4  ALARM_LED_BO ARD 



P1  BRIGHTNESS_C ONTROL_BOAR D 



LIGHT 



  P5  Audio input   







Audio input_R 







GND 







Audio input_L 



P6  1  MUTE  &  KNOB  2  LED_BOARD  3  4 



P a g e  | 55 



KEY  KEY  KNOB LED CONTROL  +12V 



 



 



P12  1  SCB POWER SUPPLY  2 



P11    RS232 



+5V  +5V 



Service Manual‐ V8800 Critical Care Ventilator 



P2  KNOB_BOARD 







 







KNOB_KEY 







GND 







TSB1 







GND 







TSA1 







GND 







+5V 







GND 







+12V 







PC_RXD1 







PC_TXD1 







GND 



13 



PC_RXD2 



14 



PC_TXD2 



19 



GND 



23 



PC_RXD3 



24 



PC_TXD3 



29 



GND 



33 



PC_RXD3 



34 



PC_TXD3 



39 



GND 



 



       



P a g e  | 56 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



5.3 Software System  5.3.1 Software Units and Their Relationship  User  Interface  Software  Unit  bears  the  responsibility  of  user  interface  of  the  whole  software system. It includes operation interface for the users, graphic interface for showing  information of the device, communication with other units, etc.    Breathing Delivery Software Unit bears the responsibility of ventilation control.    User  Interface  Connection  Software  Unit  bears  the  responsibility  of  controlling  knob,  hard key, alarm indicator and alarm speaker.    Power Supply Software Unit bears the responsibility of power management.  The relationship between these four units is as following: 



  Fig 1 Software Units of NVP System 



In  above  figure,  the  orientation  of  arrow  points  to  information  is  flowing  to  where.  Furthermore,  a  bidirectional  arrow  means  communication  between  SW  units  is  bi‐  directional.   



P a g e  | 57 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



5.3.2 Structure of the Software System 



  Fig 2 Structure of Software Units   



P a g e  | 58 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



5.3.3 Hardware and Software Interface 



  Fig 3 Interface between Software and Hardware System     



6 Disassembling and assembling  6.1 General    The  ventilator  is  mainly  formed  of  the  following  components  and  some  accessories.  Disassembling of the main units and maintenance spare parts is described in this chapter. If  not stated otherwise, the assembling procedure is the reverse of the described disassembling  procedure.   



P a g e  | 59 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



                                                          Fig 7‐1      Whole Shape    1 



Display 







Trolley assembly 







Ventilator host 







Arm assembly 







Air Compressor 







Arm assembly 



P a g e  | 60 



VL9004 Rev 1 



  



 



 



 



Service Manual‐ V8800 Ventilator 



6.2 Disassembling of the main assemblies    6.2.1 Disassembling of the display from the host   



                                      Fig 7‐2        To separate the display from the host:  a. Loose the screws (No.1)  b. Drag out the spring lock(No.2)  c. Pull out the display along the direction; 



P a g e  | 61 



 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.2.2 Disassembling of the host from the trolley 



                                    Fig 7‐3       



 



  Fig 7‐4                                                                                    Fig 7‐5       



 



  To separate the host machine from the trolley  a. Drag out drawer(No.3),as fig.7‐4 shows;  b. Screw off knurling bolts(No.2), as fig.7‐5 shows;  c. Lift up the host machine(No.1);  Note: when installing the host machine to the trolley, please take the plastic padding to aim  at the holes,such as central line in Fig 7‐3. 



P a g e  | 62 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



6.2.3 Disassembling  of  the  air  compressor  from  the  trolley 



   Fig 7‐6  To separate the air compressor from the trolley (No.2)  a. Screw off four bolts(No.3);  b. Lift up the air compressor(No.1)    1) Disassembling of the arm assembly from the host 



P a g e  | 63 



 



 



                                    Fig 7‐7 



Service Manual‐ V8800 Critical Care Ventilator 



  



To separate arm assembly from the host  a. Loosen the knob(No.1);  b. Open up the clasp(No.2) and take off arm assembly from handrail (No.3) of host;     



P a g e  | 64 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



6.2.4 Disassembling of the humidification assembly  from the host   



     



 



                                  Fig 7‐8    To separate the humidification assembly from the host  a. Loosen the knob(No.1);  b. Move the clasp(No. 2) along handrail(No.3) and take off it from the host                                                        Humidification (No.4) can been freely take off from the humidification assembly along the  guide‐rail.   



P a g e  | 65 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.3   Disassembling of maintenance spare parts  6.3.1 Disassembling of the gas supply inlet filter 



                                    Fig 7‐9  To separate the gas supply inlet filter from the host:  a. Screw off four bolts(No.1)  b. Remove the Inlet valve body (No.2);  c. Drag out the gas supply inlet filter (No.3). 



P a g e  | 66 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



6.3.2 Disassembling of oxygen sensor 



  Fig 7‐10    To separate the gas supply inlet filter from the host:  a. Pull out oxygen sensor cover(No.1)  b. Screw off the oxygen sensor(No.2) 



P a g e  | 67 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.3.3 Disassembling of W15 cable 



       Fig 7‐11  To removal the W15 cable from the host:  a. Screw off the four screw(No.1) ;  b. Lift up covered component(No.2);  c. Disconnect the cable connector;  d. Screw off the six screw(No.3) on the solenoid valve;    e. Pull out the W15 cable(No.4) 



P a g e  | 68 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



6.3.4 Disassembling of pressure sensor board 



                                Fig 7‐12  To removal the two pressure sensor board from the host:  a. Screw off the four screw(No.1) ;  b. Move the side panel(No.2) to a side;  c. Disconnect the cable connector;  d. Screw off the four screw(No.3)  f. Pull out the two pressure sensor board(No.4)   



P a g e  | 69 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.3.5 Disassembling of W14 cable 



   



P a g e  | 70 



       



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



                  Fig 7‐13                                                            Fig 7‐14  To removal the W14 cable from the host:  a. Screw off the four screw, No.2 in fig7‐13 ;  b. Move the covered component to a side, No.1 in fig 7‐13;  c. Disconnect the cable connector;  d. Screw off the four screw (No.3 in fig 7‐14);  e. Take out of the W14 cable (No.4 in fig 7‐14) ; 



P a g e  | 71 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.3.6 Disassembling of Stepper motor driving board 



                          Fig 7‐15  To remove Stepper motor driving board from the host:  a. Screw off the four screw, No.2 in fig7‐13 ;  b. Move the covered component to a side, No.1 in fig 7‐13;  c. Disconnect the cable connector;  d. Screw off the two screw, No.3 in fig7‐15;  e. Take out of step motor driving board, No.4 in fig7‐15 



P a g e  | 72 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



6.3.7 Disassembling of TSI mass flow sensor   



                                 Fig 7‐16  To removal TSI mass flow sensor from the host:  a. Screw off the four screw, No.2 in fig7‐13;  b. Move the covered component to a side, No.1 in fig 7‐13;  c. Disconnect the cable connector;  d. Screw off the two screw, No.3 in fig7‐16;  e. Move the proportional valves component(No.4 in fig7‐16) to the left side;  f. Drag out the TSI mass flow sensor (No.5 in fig 7‐16). 



P a g e  | 73 



  



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.3.8 Disassembling of the W4 cable   



  Fig 7‐17  To removal the W4 cable from the host:  a. Screw off the four screw, No. 2 in fig7‐13;  b. Move the covered component(No.1 in fig 7‐13) to a side;  c. Disconnect the cable connector;  d. Screw off the two screw, No.3 in fig7‐17;  e. Lift up the proportional valve, No.4 in fig7‐17.   



6.3.9 Disassembling of step motor     



P a g e  | 74 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



                        Fig 7‐18    a. b. c. d.



To remove the step motor from the host:  Screw off the four screw, No.2 in fig7‐13;  Move the covered component to a side, No.1 in fig 7‐13;  Screw off the four screw, No.1 in fig7‐18;  Drag out the step motor, No. 2 in fig7‐18.   



P a g e  | 75 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



6.3.10 Disassembling of PEEP valve 



                      Fig 7‐19  To removal the PEEP valve from the host:  a. Screw off the four screw, No.2 in fig7‐13 ;  b. Move the covered component(No.1 in fig 7‐13) to a side;  c. Pull out the plastic tube on the joints of PEEP valve(No.1 in fig 7‐19);  d. Disconnect the cable connector;  e. Screw off the four screw, No. 2 in fig7‐19;  f. Take off the PEEP valve 



6.3.11 Disassembling of diaphragm and scaleboard   



P a g e  | 76 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



                                      Fig 7‐20  The  diaphragm  and  the  scaleboard  appear  in  the  two  components,  expiratory  module  assembly and inspiratory module assembly.  To removal the diaphragm and the scaleboard from inspiratory module assembly:  a. Screw off the four screw, No.2 in fig7‐13;  b. Move the covered component (No.1 in fig 7‐13) to a side,    c. Pull out the plastic tube on the joints of safety valve;  d. Screw off the safety valve cover, No. 1 in fig7‐20    e. Take off the diaphragm and the scaleboard, No. 2 and 3 in fig7‐20 



P a g e  | 77 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



                          Fig 7‐21                                                                                              Fig 7‐22  a. b. c. d.



To removal the diaphragm and the scaleboard from expiratory module assembly:  Press the button(No.1 in fig7‐21);  Pull out the expiratory valve core component (No.2 in fig 7‐21);  Screw off the safety valve cover(No.1 in fig7‐22)  Take off the diaphragm(No. 2 in fig 7‐22) and the scaleboard(No. 3 in fig7‐22) 



P a g e  | 78 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



6.3.12 Disassembling of W24 cable 



                      Fig 7‐23  To removal the W24 cable:  a. Grasp the drawer handle(No.1 in fig 7‐23) and pull out battery box assembly;  b. Disconnect the cable connector;  c. Open up the spring slices, No.2 in fig7‐23;  d. Take off the W24 cable, No. 3 in fig7‐23 



6.3.13 Disassembling of W26 cable 



   



P a g e  | 79 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



            Fig 7‐24                                                                Fig 7‐25  To removal the W26 cable:  a. Pull out drawer, No.1 in fig7‐24 ;  b. Disconnect the cable connector;  c. Pull out the W26 cable(No.2 in fig 7‐25) along the lead‐rail(No.3 in fig 7‐25) 



6.3.14 Disassembling of power supply AC/DC module 



                  Fig 7‐26  To removal the power supply module:  a. Screw  off  the  four  screws(No.1  in  fig7‐26),  and  take  down  the  side  panel(No.2  in  fig  7‐26) ;  b. Screw  off  the  screws(No.3  in  fig7‐26),  disconnect  the  correlative  connectors,  and  take  down the electronic board assembly (No.4 in fig7‐26);  c. Take  out  the  power  supply  module(No.5  in  fig  7‐26)  from  host  when  screwing  off  the  P a g e  | 80 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



four screws on the host pedestal; 



6.3.15 Disassembling of components in the display  Two screws 



Four screws 



   



  



 



              Fig 7‐27                                                                                  Fig 7‐28  a. Screw off the four screws in fig 7‐27 and take off the fixed board;  b. Screw off the two screws in fig 7‐28; 



                                      Fig 7‐29                                                      Fig 7‐30  c. Pry open the back‐shell of display with the tool. You can use the tool to prize the red circle  signed places such as fig 7‐29.  d. Open the back‐shell slowly and disassemble the two screws in the fig7‐30, move away the  back‐shell.    Screws     PCB mask      e.  Screw  off  the  eight  screws,  and  then  move  away the PCB mask, such as fig 7‐30  f.  Seeing  fig7‐31,  the  components  in  the  display  can be easy to be taken away.                      Fig 7‐31            P a g e  | 81 



 



 



          











Service Manual‐ V8800 Critical Care Ventilator 











12  11 







10  9 



                                      Fig 7‐32 











 







      1. Alarm light panel      2. Display Control Panel        3. W36 cable    4. PC/104 single board computer suite, industrial grade CF card          5. Speaker Component  6. 1G memory card        7. Display interface panel      8. Touch screen control    9. Aperture and mute Panel        10. W31 cable      11. Inverter TPI‐02‐0606      12. Front shell component 



6.3.16 Disassembling of electric panel assembly  1  2 











                              Fig 7‐33  P a g e  | 82 



VL9004 Rev 1



 



 



Service Manual‐ V8800 Ventilator 



1.  Main  Control  board      2.  Main  Control  core  board        3.  Power  Board            4.  Charger Board    a. Take down the side panel, such as fig 7‐26;    b. The components on the electric panel assembly appear such as fig 7‐33, which can            be easy to be taken away; 



6.3.17 Disassembling of BDU assembly  1 







                            Fig 7‐34                                                          Fig 7‐35  a. Take down the side panel, such as fig 7‐34;  b. Screw off the two screws (No.1 in fig7‐35);  c. Take away the BDU assembly(No.1 in fig7‐35); 



P a g e  | 83 



 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



7 User maintenance  WARNING: Oricare recognizes that cleaning, disinfection, and sterilization practices vary widely among medical institutions. It is not possible to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of cleaning, sterilization, and other practices carried out in the patient care setting.



WARNING: Use a cleaning and disinfection schedule that conforms to your institution’s disinfection and risk-management policies. • Refer to the material safety data as applicable. • Refer to the operation and maintenance manuals of all disinfection equipment. • Do not inhale fumes that may result from any disinfection process.



WARNING: Movable and removable parts may clamp or even crush your hand. Use caution when moving or replacing system components.



CAUTION: The disposal of environmentally harmful devices (such as batteries and LCD display) must be in accordance with local regulations.     WARNING: Use a cleaning and disinfection schedule that conforms to your institution’s disinfection and risk-management policies. • Refer to the material safety data as applicable. • Refer to the operation and maintenance manuals of all disinfection equipment. • Do not inhale fumes that may result from any disinfection process. WARNING: Do not use talc, zinc stearate, calcium carbonate, corn starch or similar material to prevent sticking of the bellows, as these materials may enter the patient’s lungs or airway, causing irritation or injury. CAUTION:



To prevent system damage:



• Refer to the literature supplied by the manufacturer of the cleaning agent. • Never use organic, halogenated or petroleum-based solvents, anesthetic, glass cleaning agents, acetone or other irritant agents. • Never use abrasive agents (i.e. steel wool or silver polish) to clean components. P a g e  | 84 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



• Keep all liquids away from electronic components. • Prevent liquid from entering the equipment. • All cleaning solutions used must have a pH between 7.0 and 10.5. CAUTION: Never immerse the oxygen sensor or its connector in any type of liquid. • Dispose of the oxygen sensor per the manufacturer’s specification. CAUTION: Do not wash the inner surface of the oxygen sensor. CAUTION: Prior to use after cleaning or disinfecting, power up the system as described in section 6 and follow the on-screen Pre-Use test prompts to perform the Leak Test and Circuit Compliance Test. 



7.1.1 Basic guidelines    Discard one time use components after using. Don’t use hard brushes or other sharp tools in  cleaning to avoid damage to parts.  Clean components with warm water and light detergent.  Dry the components after rinsing with water.  Check each component when cleaning and replace damaged ones.  After a component is replaced, test the ventilator first prior to putting the ventilator into  service.  CAUTION:  Follow  the  instructions  of  detergent  manufacturer.  Concentrated  detergent  may  hurt  some  components.  Detergent  residue  may  cause  spots  and  cracks,  especially  in  high temperature disinfection. 



7.1.2 Disinfection and Sterilization  Discard one time use components after using. Using an FDA cleared Gluteraldehyde  disinfection solution; follow the manufacturer’s instructions for high level disinfection and  rinsing of the part while adhering to facility policies and procedures. Ensure that the part has  been cleaned as described before disinfecting.    Alternative method to disinfection solution  is steam autoclaving. High level disinfection of the part is accomplished, through steam  autoclaving up to a maximum temperature of 134 ºC (273 ºF). When using an autoclave, 



P a g e  | 85 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



follow the manufacturer’s instructions for high level disinfection.  Table 8‐1 Guidelines for cleaning and disinfection  Item



How Often



Procedure



Note



External



After every



Wipe clean with a damp cloth and



Be careful not to let



surface of



patient



water-soluble mild soap or with one



disinfectant into the



of the chemicals listed below or its



machine or power



equivalent. Apply soap or chemical



supply plug.



ventilator (including



 



LCD)



to damp cloth and directly to the equipment. If necessary use damp



 



 



cloth and water to rinse off chemical residue. • Mild dishwashing detergent • Isopropyl alcohol (70% solution) • Window cleaning solution (with isopropyl alcohol and ammonia) • Ammonia (15% solution) • Bleach (10% solution) • Hydrogen peroxide (3% solution)   



Expiratory



After each



Remove Expiratory module from



Reinstall the



module



patient/



ventilator and remove diaphragm



module after



weekly



and water trap. Clean the shell,



cleaning and



diaphragm and water trap



drying. Pay



separately. Using an approved



attention to assure



Gluteraldehyde disinfection solution,



connections are



follow the manufacturer’s instruction



gas tight.



for high level disinfection and rinsing of the diaphragm, water trap and shell. Note: The shell and water trap can also be high level disinfected through steam autoclaving up to a maximum temperature of 134°C (273°F). Using an autoclave, follow the manufacturer’s instruction for high level disinfection of the shell and water trap while adhering to facility procedures. Note: autoclaving should not be used for high level disinfection of the diaphragm.



P a g e  | 86 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



7.2 Regular Maintenance  7.2.1 Maintenance Principles  Don’t  use  a  faulty  machine.  Ask  an  authorized  agent  of  our  company  to  carry  out  all  necessary  maintenance  tasks.  Test  the  machine  after  maintenance  for  normal  operation.  Every parameter should meet requirement in specification.  In order to ensure the reliability of the machine, all maintenance and repair work should be  carried  out  by  an  authorized  agent  of  our  company.  In  case  such  person  is  unavailable,  a  qualified person with similar maintenance experience is acceptable for the work.  CAUTION:  Person  with  no  sufficient  experiences  is  strictly  prohibited  to  perform  maintenance. 



Use products of our company to replace damaged ones and test, ensuring all specifications  met.  Contact  local  service  agent  of  our  company  in  case  support  is  needed.  In  all  cases,  maintenance fee is the current component price plus reasonable labor cost, except for those  within guarantee period. 



7.2.2 User Maintenance  Minimum maintenance 



Task 



interval  Daily 



Drain water in gas supply inlet filter, check  liquid 



in 



expiratory 



module 



water 



trap(collected  liquid  volume  cannot  be  more than half of the bottle)  1‐3 month(s) 



P a g e  | 87 



Clean air filter 



 



 



Every year 



Service Manual‐ V8800 Critical Care Ventilator 



Calibrate flow and pressure sensors;  Calibrate  inspiratory  valve  and  expiratory  valve (if necessary) 



Every year or after calibration    Replace the O2 sensor(actual life depends  on temperature and O2 concentration)  When cleaning and installing 



Check  components  and  replace  or  repair  when necessary. 



Notes related to User Maintenance:  To change the Oxygen sensor, remove the Oxygen sensor cover on the rear of the machine,  replace the installed Oxygen sensor with a new Oxygen sensor supplied by Oricare. Use care  to assure connections and installation are complete. Replace Oxygen sensor cover.  Note: V8800Ventilator performs an O2 sensor calibration automatically, every 24 hours, no  operator intervention is needed.    WARNING:  If  the  V8800  Ventilator  will  not  be  in  use  for  a  period  of  more  than  6  months, the internal batteries must be disconnected or removed to prevent possible damage  to the equipment or risk to users or service personnel. 



  WARNING:The  ventilator  must  not  use,  nor  be  connected  to,  any  anti‐static  or  electrically conductive hoses and tubing.   



7.2.3 Service life of product/accessories    Caution:The  service  life  of  the  following  items  is  based  on  normal  operating  conditions. 



P a g e  | 88 



Mask, patient pipeline 



One‐off 



Power supply cable, gas supply pipe 



8 years 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



Machine 



8 years 



Battery 



1 year 



 



7.3 Maintenance in Operation and Transportation  The  location  of  machine  should  be  proper  so  that  it  cannot  obstruct  or  be  disturbed  by  medical care personnel. Fix power supply cable well to avoid failure. Use caution not to touch  accidental keys on the panel, which may make tidal volume setting wrong.  During transportation of the ventilator with or without a patient connected, make sure that  the following conditions are fulfilled;  Gas  cylinders  are  connected  with  a  sufficient  amount  of  gas,  the  Battery  module  is  functioning. Follow the hospital guidelines.  Use  the  handles  on  the  Mobile  Cart.  Transport  the  bed  and  the  ventilator  slowly,  and  watch the patient connection carefully to see that no pulling or other movement occurs.  Be careful not to tip the Mobile Cart when crossing an obstacle such as a doorstep   



7.3.1 Transportation  Use care when moving machine within hospital or clinical environment.      WARNING: If Control Unit of Ventilator is dropped or damaged during transportation,  equipment failure could result in patient injury.    WARNING:  Tip  over  hazard  –  use  care  when  moving  Ventilator  mounted  to  Cart  as  device could tip over leading to injury or damage to the equipment and possible subsequent  patient injury.  User can carry packaged machine while riding vehicle, plane and train. Impact, severe shock  and  damp  should  be  avoided  during  transportation  (other  conditions  should  accord  with  P a g e  | 89 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



purchase contract), with ambient temperature ‐20°C‐+60°C and relative humidity not more  than  95%.  In  case  transportation  conditions  don’t  agree,  put  the  machine  in  normal  operating environment at least 24h before using. 



7.3.2 Storage    CAUTION: Do not put ventilator into the shock environment.    CAUTION: Do not lay the heavy on the top.  The  machine  should  be  stored  in  a  room  with  temperature  ‐20°C  ‐+60°C  and  relative  humidity 5 ‐ 95% non‐condensating, with ventilation and no corrosive gas.      WARNING:  If  the  storage  environment  conditions  don’t  agree,  put  the  machine  in  normal operating environment at least 24h before using.    CAUTION: The device should be stored at the room that is drafty and no corrosion gas  exists.    CAUTION:  When  the  storage  conditions  are  beyond  the  requirements  of  operational  environment, and the storage state is transferred into operation state,  the  product  only  can  be  used  after  being  stored  in  environment  for  over 24 hours. 



7.4 Battery maintenance  7.4.1 Battery specification  Internal battery module  -



DC12V, 2.3AH, lead‐acid battery, 2 in series 



-



Typical charge time: 4h 



-



Typical discharge time: 1h 



Extended battery module  -



DC12V, 5.0AH, lead‐acid battery, 2 in series 



P a g e  | 90 



VL9004 Rev 1 



 



 



 



-



Typical charge time: 8h 



-



Typical discharge time: 2h 



Service Manual‐ V8800 Ventilator 



7.4.2 Precautions  Charge: When operating with AC power supply on, the system will maintain the battery  automatically. Charge time is less than 4h.  Discharge:  Extended  battery  module  (if  installed)  will  supply  power  prior  to  the  internal battery module when machine is operating on battery.  In case of low battery condition, an alarm message “Low battery” will appear, notifying  the user to restore AC power supply to charge batteries, otherwise the batteries will be  depleted  and  alarm  “Limited  Battery  Capacity”  will  be  displayed,  and  eventually  the  system will be shut down (for safety reason, manual power‐on is required to start the  machines again after an automatic shutdown). 



7.4.3 Battery storage  In case the battery is to be stored for a long time, charge it fully prior to storage.    In case storage time exceeds 3 months, charge battery every 3 months.  High humidity and high temperature environments should be avoided for storage.  If battery is damaged due to improper maintenance, replace promptly, otherwise liquid  leakage may corrupt the machine. Contact the manufacturer when replacing battery. 



7.4.4 Battery replacement  Same  model  battery  with  CE  certification  is  suggested.  Make  sure  AC  power  supply  is  disconnected before replacing.    CAUTION: An authorized Oricare services representative can replace battery. If not to  use  the  battery  for  long‐time,  please  contact  Oricare  service  representatives  to  disconnect  P a g e  | 91 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



battery. The waste battery should be disposed in accordance with the local policies.    CAUTION: When ‘battery low’ alarm occurs, charging should be done immediately. Or  else, the Shangrila590 Ventilator System will shut off in several minutes automatically. 



7.5 Disposal    Correct Disposal of Batteries and O2 Sensors   



WARNING: Treatment of batteries and O2-sensor capsules: • Do not throw into fire! Risk of explosion. • Do not force open! Danger of bodily injury. • Follow all local regulations with respect to environmental protection when disposing of batteries and O2-sensor capsules.



This product must not be disposed of with your other waste. Instead, it is your responsibility  to dispose of your waste equipment by handing it over to a designated collection point for  the recycling of waste electrical and electronic equipment, or by returning it to Medical  Illumination International, Inc for reprocessing. The separate collection and recycling of your  waste equipment at the time of disposal will help to conserve natural resources and ensure  that it is recycled in a manner that protects human health and the environment. For more  information about where you can drop off your waste equipment for recycling, please  contact your local city office, your waste disposal service, or your product distributor or  retailer.   



7.6 Manufacturing techniques and materials:    For a period of one year from the date of original delivery, the components and assemblies of  this product are warranted to be free from defects manufacturing techniques and materials,  provided that the same is properly operated under the conditions of normal use and regular  maintenance. The warranty period for other parts is three months. Expendable parts are not  included.  Oricare’s  obligation  under  the  above  warranties  is  limited  to  repairing  free  of  charge.   



P a g e  | 92 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



7.7 Free Obligations:    • Oricare’s obligation under the above warranties does not include the freight and other  fees;    •  Oricare  is  not  responsible  for  any  direct,  indirect  or  final  product  broken  and  delay  which  result  from  improper  use,  alteration  by  using  unapproved  assemblies  and  maintenance by anyone other than Oricare;    • This warranty does not apply to the followings:    Improper use;    Machines without maintenance or machines broken;    The label of Oricare original serial number or mark is removed or replaced;    Other manufacturers’ products.   



7.8 Security, reliability and operating condition:    Oricare is not responsible for the security; reliability and operating condition of this product  in case that:    • The assemblies are disassembled, extended and readjusted    • This product is not operated correctly in accordance with the manual instruction. The  power  supply  used  or  operating  environment  does  not  follow  the  requirements  in  this  manual.   



7.9 Return    Follow the steps in case that the product needs to be returned to Oricare:    1、Obtain the rights of return    Contact with the customer service of Oricare by informing them the number and type of the  product. The number is marked on the surface of the product. Return is unacceptable if the  number  cannot  be  identified.  Enclose  a  statement  of  the  number,  type  and  the  reason  of  return as well.    2、Transportation charges    Transportation  and  insurance  charges  must  be  prepaid  by  the  user  for  transporting  the  product to Oricare for repairing.   



  NOTE: Each Oricare product has a serial number, such as V8800 xx xx xxx. This serial  number should be referenced in all correspondence. Where: V8800 = machine model number xx xx xxx: equipment number  1st two ‘xx’ digits – last two digits of current year  2nd two ‘xx’ digits indicate month product was manufactured



 



rd



Z=O,Y=1,X=2,W=3,V=4,U=5,T=6,S=7,R=8,K=9.   



3 group of ‘xxx’ digits indicates the number unit shipping in a current month.



P a g e  | 93 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



Product serial number appears at nameplate and on equipment packaging.    Manufacturer 



Oricare Inc 



Address 



1900 AM Drive  Quakertown, PA 18951 USA 



EC Representative 



Obelis SA 



Address 



Bd  General  Wahis  53,  103D  Brussels,  Belgium 



Tel. 



+(32) 2732‐59 54 



Fax 



+(32) 2732‐60 03 



Web site 



www.oricaremed.com 



Email 



[email protected] 



8 Pre‐Use Test and calibration  8.1.1 Technical Test  Possible reason and recommended action if failed:  See the Technical Test Error Code below. 



8.1.2 AC/Battery Test  Possible reason and recommended action if failed:  Make sure the internal battery is installed and charged.  Make sure the AC supply is OK.   



8.1.3 Gas Supply Test  Possible reason and recommended action if failed:  1. Make sure the two gas supplies are connected, and the pressure is in the spec 



8.1.4 Oxygen Sensor Test  Possible reason and recommended action if failed:  1. Perform 21% and 100% oxygen sensor test at first, and try again.  2. Make sure the two gas supplies are connected, and the pressure is in the spec.  3. Make sure the oxygen sensor is valid and connected to the main box.  4. Make sure the oxygen auto calibration valve is connected right and no leak. 



P a g e  | 94 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



8.1.5 Leak Test  Possible reason and recommended action if failed:  1. Calibrate Inspiratory valve at first, and try again;  2. Make sure machine and connected tube have no leak. 



8.1.6 Flow Sensor Test  Possible reason and recommended action if failed:  1. Calibrate Inspiratory valve at first, and try again;  2. Make sure machine and connected tube have no leak;  3. Make sure inspiratory flow sensor and expiratory flow sensor work fine. 



8.1.7 Pressure Sensor Test  Possible reason and recommended action if failed:  1. Calibrate Inspiratory valve at first, and try again;  2. Make sure machine and connected tube have no leak;  3. Make sure inspiratory pressure sensor and expiratory pressure sensor work fine. 



8.1.8 Safety Valve Test  Possible reason and recommended action if failed:  1. Calibrate Inspiratory valve at first, and try again;  2. Make sure safety valve works fine;  3. Make sure machine and connected tube have no leak;  4. Make sure expiratory pressure sensor works fine. 



8.1.9 Breath Circuit Test  Possible reason and recommended action if failed:  1. Calibrate Inspiratory valve and Expiratory valve at first, and try again.  2. Make sure machine and connected tube have no leak;  3. Make sure expiratory pressure sensor works fine. 



8.1.10   etCO2 Sensor Test  Possible reason and recommended action if failed:  1. Take off the CO2 sensor from the breath circuit, and put it in the air.  2. Make sure the CO2 sensor has warmed up. The LED indicator in the sensor turns green.  P a g e  | 95 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



3. Make sure the adapter is installed into the sensor. 



8.2 Calibration  8.2.1 Pressure Sensor  Possible reason and recommended action if failed:  1. Make sure inspiratory pressure sensor and expiratory pressure sensor work fine;  2. Make sure no gas flow pass inspiratory pressure sensor and expiratory pressure sensor. 



8.2.2 Flow Sensor  Possible reason and recommended action if failed:  1. Make sure inspiratory flow sensor and expiratory flow sensor work fine;  3. Make sure no gas flow pass inspiratory flow sensor and expiratory flow sensor. 



8.2.3 O2 Sensor  Possible reason and recommended action if failed:  1. Calibrate Inspiratory valve and Expiratory valve at first, and try again;  2. Make sure the two gas supplies are connected, and the pressure is in the spec;  3. Make sure the oxygen sensor is valid and connected to the main box;  4. Make sure the mixer works fine;    5. Make sure regulators wok fine. 



8.2.4 CO2 Sensor  Possible reason and recommended action if failed:  See Section 9.2.10. 



8.2.5 Inspiratory Valve  Possible reason and recommended action if failed:  1. Make sure the two gas supplies are connected and work normal;  2. Make sure machine and connected tube have no leak;  3. Make sure inspiratory flow sensor and expiratory flow sensor work fine.  4. Make sure inspiratory valve works fine;  P a g e  | 96 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



8.2.6 Expiratory Valve  Possible reason and recommended action if failed:  1. Make sure the two gas supplies are connected and work normal;  2. Make sure machine and connected tube and lung have no leak;  3. Make sure inspiratory pressure sensor and expiratory pressure sensor work fine;  4. Make sure inspiratory valve and expiratory valve work fine; 



8.2.7 Touch Screen  Possible reason and recommended action if failed:    1. Trying the fewest test points calibration (4 points).  2. Make sure press the center of each test point (the red region).  3. Make sure the plastic frame of screen doesn’t constrict the touching area of touch screen.     



9 Troubleshooting  9.1 Technical Error  9.1.1 Technical Test Error Code  The  error  code  is  made  up  by  8  digits  hexadecimal  numbers,  which  are  four  byte  using  binary  system.  The  highest  two  bytes  are  internal  AD  test  result  flag,  the  lowest  byte  are  external AD test result flag. The meaning of each bit is shown in the below table.    External  AD##(0xF FFF00xx) 



Test Point 



Bit 1 



Zero  point  of  expiratory  1. Remove  the  test  lung  and  breath  circuits,  flow rate  perform the test again.  2. Don’s  start  PUT  too  quickly  after  quit  from  ventilation.  3. Calibrate  the  pressure,  flow  rate  sensor,  Insp.  valve and Exp valve.  4. Proportional  solenoid  valve  can’t  close  completely.  Replace  proportional  solenoid  valve. 



P a g e  | 97 



Possible Reason and Recommended Action 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



5. Replace expiratory flow rate sensor.  Bit 2 



Zero  point  of  inspiratory  1. Remove  the  test  lung,  connect  Insp.  port  and  flow rate  Exp. port directly with a tube.    2. Don’s  start  PUT  too  quickly  after  quit  from  ventilation.  3. Calibrate  the  pressure,  flow  rate  sensor,  Insp.  valve and Exp valve.  4. Proportional  solenoid  valve  can’t  close  completely.  Replace  proportional  solenoid  valve.  5. Replace mother board. 



Bit 3 



Inspiratory temperature 



1. 2. 3.



4.



Check  the  cable  between  the  mother  board  to the temperature sensor  The  interface  IC  on  the  mother  board  is  broken. Replace mother board.  The  input  port  of  external  AD  chip  on  the  mother  board  is  broken.  Replace  mother  board.  Replace temperature sensor. 



Bit 4 



‐‐ 



Bit 5 



Zero  point  of  inspiratory  1. Remove  the  test  lung,  connect  Insp.  port  and  pressure  Exp. port directly with a tube.  2. Don’s  start  PUT  too  quickly  after  quit  from  ventilation.  3. Calibrate  the  pressure,  flow  rate  sensor,  Insp.  valve and Exp valve.  4. Proportional  solenoid  valve  can’t  close  completely.  Replace  proportional  solenoid  valve.    5. The  amplifier  on  the  mother  board  is  broken.  Replace mother board.  6. The  input  port  of  external  AD  chip  on  the  mother  board  is  broken.  Replace  mother  board.  7. The sensor is broken. Replace mother board. 



Bit 6 



Zero  point  of  expiratory  1. Remove  the  test  lung,  connect  Insp.  port  and  pressure  Exp. port directly with a tube.  2. Don’s  start  PUT  too  quickly  after  quit  from  ventilation.  3. Calibrate  the  pressure,  flow  rate  sensor,  Insp.  valve and Exp valve.  4. Proportional  solenoid  valve  can’t  close  completely.  Replace  proportional  solenoid  valve. 



P a g e  | 98 



‐‐ 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



5. The  amplifier  on  the  mother  board  is  broken.  Replace mother board.  6. The  input  port  of  external  AD  chip  on  the  mother  board  is  broken.  Replace  mother  board.  7. The sensor is broken. Replace mother board.  Bit 7 



Barometric pressure 



  1. Proportional  solenoid  valve  can’t  close  completely.  Replace  proportional  solenoid  valve.    2. The  amplifier  on  the  mother  board  is  broken.  Replace mother board.  3. The  input  port  of  external  AD  chip  on  the  mother  board  is  broken.  Replace  mother  board.  The sensor is broken. Replace mother board. 



Bit 8 



‐‐ 



‐‐ 



Internal  AD(0xxxxx 00FF) 



­­ 



­­ 



Bit 1 



‐‐ 



‐‐ 



Bit 2 



‐‐ 



‐‐ 



Bit 3 



‐‐ 



‐‐ 



Bit 4 



‐‐ 



‐‐ 



Bit 5  Bit 8 



Inspiratory  valve  AD/DA  1. Poor  connection  between  the  mother  board  sample  and  daughter  board.  Re‐plug  the  daughter  board.  2. The  daughter  board  is  broken.  Replace  daughter board. 



Bit 6  Bit 7 



Expiratory  valve  AD/DA  1. Poor  connection  between  the  mother  board  sample  and  daughter  board.  Re‐plug  the  daughter  board.  2. The  driver  chip  of  expiratory  valve  on  the  mother  board  is  broken.  Replace  mother  board.  3. The  daughter  board  is  broken.  Replace  daughter board. 



Bit 9 



‐‐ 



‐‐ 



Bit 10 



‐‐ 



‐‐ 



Bit 11 



‐‐ 



‐‐ 



Bit 12 



‐‐ 



‐‐ 



Bit 13 



‐‐ 



‐‐ 



Bit 14 



‐‐ 



‐‐ 



Bit 15 



‐‐ 



‐‐ 



P a g e  | 99 



 



 



Bit 16 



‐‐ 



Service Manual‐ V8800 Critical Care Ventilator 



‐‐ 



1. Error Codes in the Log  Error  Code 



Test Point 



Possible Reason and Recommended Action 



1000 



PS board 7.5V DC 



The 7.5 V out is out of range. Replace PS board. 



PS board LDO3.3V DC 



The  3.3V  LDO  output  is  out  of  range.  Replace  PS  board. 



PS board 12V DC 



The 12V output is out of range. Replace PS board. 



PS board Analog 3.3V DC 



  The 3.3V output is out of range. Replace PS board. 



1001  1002  1004 



1013 



Temperature  sensor  in    the  PS  board  1. Sensor failed. Replace W17.  (Malfunction)  2. PS board failed. Replace PS board. 



2000 



12V DC on UIC board 



The 12V out is out of range. Replace PS board. 



2001 



5V DC on UIC board 



The 5V out is out of range. Replace PS board. 



2003 



Alarm  Speaker  in  the  UIC  Check  if  the  alarm  speaker  in  the  screen  is  board short  disconnected or short. 



2004 



Alarm  Speaker  in  the  UIC  board open  Data check 



3000 



Data  files  are  broken  in  the  embedded  computer  of UI. Configurations, previous patient’s data, logs,  trends  data  may  be  lost,  and  the  manufactory  default settings will be applied.  Check  these  data  and  re‐configure  them  if  necessary. 



Communication  failure  Check if the MCU of BDU, PS and UIC are working.  between BDU to UI  Make  sure  cables  linking  to  the  BDU,  PS  and  UIC  Communication  failure  are OK.  3005: Replace the daughter board  between PS board to UI  Communication  failure  3006: Replace the PS board  between UIC board to UI  3007: Replace the UIC board 



3005  3006  3007 



9.1.2 Other Errors  Alarm  1) Communication  Check if the UI are working.  sound  failure  between  UI  to  Make sure the cable linking to the PS board is OK.  from  PS board  Replace power supply board.  buzzer  in  2) No  AC  and  battery  the  PS  connected  board     



P a g e  | 100 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



9.1.3   Service Tools  Service tools are listed underside:    Sequence  number   



Name 



Quantity 



Remark 



1   



Metric hex driver set   



One set   



 



2   



#2cross screwdriver   



One entries   



Staff width is about 5mm 



3   



Multi‐meter   



One block   



 



4   



V8800 checkout frock and  its attachments for ESD  (Electro‐Static Discharge)  protection) 



One desk   



If anyone of the electric  machine, electric  machine driver and the  main board is replaced   



5   



Breath pipeline cover bag 



One set   



 



6   



Simulated lung   



One entries   



 



   



P a g e  | 101 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



10 Appendix  A  Contact  &  Ordering  Information  How to Call for Service      This service manual provides procedures for testing and maintaining the Ventilator V8800. It  is not intended to be a complete maintenance document; therefore, it contains no  disassembly or reassembly instructions.  Refer any repairs or adjustments that exceed the scope of this manual to the Service Center  of Oricare Inc., by calling 800‐810‐8333.  This manual contains proprietary information. It is intended for use only by individuals  qualified in the installation and maintenance of the Oricare anesthetic machine. Receipt,  purchase, or possession of this document in no way confers or transfers any other rights for  the use of this information. Disclosure or reproduction of the enclosed, without the written  permission of Oricare Inc. is prohibited.          This manual is intended for use only by technicians who have successfully completed  Oricare, Inc. training on this product.  Oricare, Inc. believes the information herein is accurate but accepts no responsibility for  errors, omissions, or misrepresentation. 



P a g e  | 102 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



11 Appendix B    Diagrams and Schematics   



    Pneumatic diagram          P a g e  | 103 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



12 Appendix C Specifications  System  Operating Conditions  Operating Temperature Range: +10 to +40 C    Relative Humidity: 15 to 95% non‐condensing    Atmospheric Pressure: 500 to 1060 hPa  Non‐operating Conditions    Storage Temperature Range: ‐20 to +60 C    Storage Relative Humidity: 5 to 95% non‐condensing    Storage Atmospheric Pressure: 500 to 1060 hPa  Power Supply    Power Supply, Automatic Range Selection    Battery Backup, Internal (Standard)    Battery Backup, Extended (Optional)  Maximum Power Consumption    100‐120 +/‐ 10%,    220‐240 +/‐ 10% VAC 50/60 Hz    Two rechargeable battery modules, 12 V, 2.3 Ah each. Recharge time approximately 4 hours.  Battery backup time of 60 minutes minimum with only standard internal battery.    Two rechargeable battery modules 12 V, 5  Ah each. Recharge time approximately 12 hours  for  both  the  Internal  and  Extended  batteries  together.  Battery  backup  time  of  3  hours  minimum for both the internal and external batteries.    At 110‐120 VAC: 2A, 150VA (Ventilator only); 15 A (Ventilator plus Auxiliary Outlets)    At 220‐240 VAC: 1A, 150VA (Ventilator only); 7.5 A (Ventilator plus Auxiliary Outlets)      CAUTION: The Power Supply should meet the above specifications. 



Ventilator  General Dimensions  User Interface: 406  wide x  70  deep  x 289  high (mm)  Ventilation Delivery  Unit: 460 wide  x  504  deep x 669 high (mm) System with Cart (optional): 774 wide x 624 deep x 1410 high (mm)    Weight : Total: 60 kg    (Approximate) User Interface: 3 kg    Ventilation Delivery Unit: 27 kg    Cart: 30 kg    Trigger: Flow and Pressure    Method    Maximum 110 cm H2O    Limited pressure    Maximum 100 cm H2O, ensured by high pressure limit    Working pressure 90 cm H2O, pressure ventilation    Sub‐atmospheric pressure:In the expiratory phase (‐40 cm H2O)    P a g e  | 104 



VL9004 Rev 1 



 



 



 



Service Manual‐ V8800 Ventilator 



Gas Supply    Supplied gases must be free of water, oil and particles    Inlet Gas Pressure : O2, Air: 200 kPa to 650 kPa (29 – 94 PSI)    Connection Standards Available    DISS, NIST    Patient System Connectors    Male 22 mm Conical Fittings in accordance with ISO 5356‐1    User Interface    Attaches to the cart, a rail, or other mounting system.  Standard Conditions Specifications  Error ranges in this document assume the following standard conditions:    Ambient pressure: 101.3 kPa (1 atmosphere)    Room temperature: 20 C    Dry gases in patient system    Inlet pressure: 345 kPa (50 PSI)  Inspiratory Channel  Pressure Drop : Maximum 5 cm H2O at a flow of 60 liters/min without CO2 airway adapter.  Compliance : Maximum 2 ml/cm H2O (with Fisher & Paykel MR 810 humidifier and    reusable silicone adult patient circuit)    Gas Delivery System    Microprocessor controlled valves    Gas Delivery Device    Flow Range:    1 to 180 liters/min. (Adult)    0 to 60 liters/min. (Child)    Maximum Pressure Setting: 90 cm H2O    NIV Max leakage compensation level    Adult: 30 liters/min    Child: 15 liters/min  Expiratory Channel  Pressure Drop (Resistance)    Maximum 5 cm H2O at a flow of 60 liters/min without CO2 airway adapter.    Compliance    Maximum 2 ml/cm H2O (with Fisher & Paykel MR 810 humidifier and    reusable silicone adult patient circuit)    PEEP Regulation    Microprocessor controlled valve    PEEP Setting Range : 0 – 50 cm H2O    Expiratory Flow Measurements    Range: 0 – 180 liters/min.  Monitoring  Expiratory Minute Volume    Range: 0 – 60 liters/min.    Accuracy: +/‐ 0.5 LPM or +/‐ 8% of measured value (whichever is greater)   



P a g e  | 105 



 



 



Service Manual‐ V8800 Critical Care Ventilator 



Resolution: 0.1 liters > 1 liter/min, 0.001 liters