17 0 4 MB
Service Manual
Preface
pHOx Service Manual
Stat Profile® pHOx® Service Manual Part Number and Ordering Information The Stat Profile® pHOx® Service Manual (PN 24301) can be ordered from Nova Biomedical Order Services. Write or call: Nova Biomedical 200 Prospect Street Waltham, MA 02454-9141 U.S.A. Tel:
1-800-458-5813
FAX: (781) 893-6998 (in the U.S.A.) or (781) 899-0417 (outside the U.S.A.)
Technical Assistance (USA) For technical assistance, call Nova Biomedical Technical Services at: Tel:
1-800-545-NOVA or (781) 894-0800
FAX: (781) 894-0585
Trademarks Stat Profile® is a registered trademark of Nova Biomedical. pHOx® is a registered trademark of Nova Biomedical.
Copyright Printed in the U.S.A. Copyright 2005, Nova Biomedical, Waltham, MA 02454.
i
Preface
ii
pHOx Service Manual
Stat Profile pHOx Service Manual
Contents 1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1.1
1.2 1.3
The Stat Profile pHOx Analyzer ............................................................................1-1 1.1.1 Environmental ............................................................................................1-1 1.1.2 Energy consumption ..................................................................................1-1 1.1.3 Physical Dimensions ..................................................................................1-1 1.1.4 Regulatory Compliance .............................................................................1-1 1.1.5 Location - Stable stationary surface. .........................................................1-2 1.1.6 Operational.................................................................................................1-2 1.1.7 General .......................................................................................................1-2 Cautions and Hazards ............................................................................................1-3 Required tools .......................................................................................................1-4
2.0 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 2.1
2.2
Mechanical Assemblies..........................................................................................2-1 2.1.1 Pump ..........................................................................................................2-2 2.1.2 Sampler Assembly ....................................................................................2-2 2.1.3 Rotary Valve Assembly .............................................................................2-2 2.1.4 Sensor Module Assembly .........................................................................2-3 Electronic Assemblies. ...........................................................................................2-3 2.2.1 Power Entry Module .................................................................................2-3 2.2.2 Power Supply ............................................................................................2-3 2.2.3 Digital Control Board ...............................................................................2-4 2.2.4 Analog Board .............................................................................................2-6 2.2.5 SO2 Control Board.....................................................................................2-6 2.2.6 Electrode Interface Boards .........................................................................2-7 2.2.7 Door Assembly Electronics........................................................................2-7
PN 24301 Rev. B 2005/11
TOC-1
Table of Contents 3.0 Replacement Procedures . . . . . . . . . . . . . . . . . . . . 3-1 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 3.10 3.11 3.12 3.13 3.14 3.15 3.16 3.17 3.18 3.19
Cover Removal.......................................................................................................3-1 Tubing Harness .....................................................................................................3-2 Power Supply Removal ..........................................................................................3-4 Sampler Assembly .................................................................................................3-5 Rotary Valve Assembly ........................................................................................3-10 Valve Assembly ....................................................................................................3-11 Rotary Valve Motor ..............................................................................................3-12 Pump Assembly ...................................................................................................3-13 Digital Control Board ..........................................................................................3-14 Analog Board ......................................................................................................3-15 Printer Assembly .................................................................................................3-16 Sensor Module Light Source ...............................................................................3-17 Door Removal /Reinstallation Procedure ...........................................................3-18 Display Assembly ................................................................................................3-20 Distribution board ...............................................................................................3-21 Keypad Assembly ...............................................................................................3-21 Sensor Module Assembly ....................................................................................3-22 Electrode Interface Boards ...................................................................................3-23 Pinch Valve Assemblies .......................................................................................3-24
4.0 Troubleshooting Information . . . . . . . . . . . . . . . . . . 1-1 4.1
4.2
4.3
TOC-2
Service Screens ......................................................................................................1-3 4.1.1 System Test Screen ....................................................................................1-4 4.1.1.1 Operator Flow Test .........................................................................1-4 4.1.1.2 Service Flow test ............................................................................1-5 4.1.2 Analog Input Screen ..................................................................................1-6 4.1.3 Sensor Subsystem .....................................................................................1-7 4.1.4 Printer Menu ..............................................................................................1-8 4.1.5 Communication Tests .................................................................................1-9 Flow Problems .....................................................................................................1-10 4.2.1 Isolating a Plug in the W/R to Waste Tubing ...........................................1-11 4.2.2 Isolating a flow problem in the Sampler/Sensor module .........................1-11 4.2.3 Isolating a flow problem between the Fluid Pack and the Fluid Fountain1-12 Status Codes .......................................................................................................4-13 4.3.1 Analyte Troubleshooting ( Series code 1-6 and Digit codes 1-4) ............4-18 4.3.2 Single Channel Failures ......................................................................4-20 4.3.3 Temperature Problems .............................................................................4-21 4.3.4 Air Detector (ADT ) Problems ................................................................4-21 Operator Flow Test ...............................................................................................4-26 Sensor Module Back Flush Procedure .................................................................4-27 PN 24301 Rev. B 2005/11
Stat Profile pHOx Service Manual A
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 A.1 A.2 A.3 A.4 A.5 A.6
S
Quick Reference Guide .........................................................................................A-1 General Maintenance/QC......................................................................................A-8 A.2.1 Other Analyzer Maintenance ..................................................................A-12 pHOx Electrode/Standards..................................................................................A-13 Electronics and Mechanicals Overview ..............................................................A-13 Shutdown Procedure ...........................................................................................A-14 Start Up Procedure ..............................................................................................A-14 Installation Check List ........................................................................................A-15 Preventive Maintenance Check List....................................................................A-17 Explosion Parts List ............................................................................................A-19
Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1 #10-1087A S-2 #10-1071B #10-1072B #10-1073B #10-1074B #10-1075A #10-1076B #10-1077B #10-1078B #10-1079B #10-1080B #10-1081B #10-1082B #10-1083B #10-1084B #10-1085B #10-1086A
Wiring Doucument ...................................................................................... S-1 Wireing Charts ............................................................................................. S-2 24292-01 pHOx Digital Board Processor and Memory ............................. S-3 24292-02 pHOx Digital Board Comm. Printer, Display Touch Memory ... S-4 24292-03 pHOx Digital Board Motor & Valve Drivers .............................. S-5 24292 pHOx Digital Board Assembly Drawing .................................... S-6 21642 pHOx SO2 Board ........................................................................ S-7 24727 pHOx Distribition Board ............................................................ S-8 24291 Analog Board - Cortex Assembly Drawing .............................. S-9 24292-01 New Gen. Sensor Board ........................................................... S-10 24292-02 New Gen. Sensor Board ........................................................... S-11 24292-03 New Gen. Sensor Board ........................................................... S-12 24292-04 New Gen. Sensor Board ........................................................... S-13 23897 Pump Assembly ........................................................................ S-14 23899 Sampler Assembly .................................................................... S-15 23898 Rotary Valve Drive Assembly ................................................... S-16 24614 Printer Interface IF3001 Seiko ................................................. S-17 24720 Power Supply Universal input .................................................. S-18
PN 24301 Rev. B 2005/11
TOC-3
Table of Contents
TOC-4
PN 24301 Rev. B 2005/11
1 Introduction
This manual contains information and procedures used in servicing the Stat Profile pHOx Analyzer. The intent is to supplement the information provided in the Stat Profile pHOx Instructions for Use Manual (PN 37865) and the Stat Profile pHOx Advanced User Applications Manual (PN 35235). This manual does not supersede any current performance specification claims, expendable maintenance procedures, or warranty criteria as outlined in the reference manual.
1.1
The Stat Profile pHOx Analyzer
1.1.1 Environmental Ambient Temperature: Ambient Humidity:
15° - 32° C ( 59° - 86° F) 0% - 95% Non-condensing
1.1.2 Energy consumption Air Conditioning Load: Peak 44 BTU/Hour Power Consumption: ~130 watts peak Power requirements: 100-120; 230-240 VAC, 50/60 Hz • 2 Amp Time Delay (SB 2A or T2A) at 100-120VAC line • 1 Amp Time Delay (T1A) at 220-240 VAC line
1.1.3 Physical Dimensions Weight: Dimensions:
8.1 Kg (18 lbs) 38 cm high (15 in) x 31 cm (12 in) wide x 38 cm (15 in) depth
1.1.4 Regulatory Compliance The analyzer is tested and acceptable to attach the CSA, GS (Europe), JIS (Japan), and CE (self declaration) safety marks, and compliant with EN1010-1, EN55011, and IEC 801-2.
1-1
1. Intro.
1.0 Introduction
pHOx Service Manual 1.1.5 Location - Stable stationary surface. • • •
8 cm ( 3 in) clearance to a wall or adjacent instrumentation Not intended for use on vibrating (helicopter, etc.) surface Placement on a cart is acceptable if the cart is stationary during operation.
1.1.6 Operational The analyzer is designed to be on at all times. It may be turned off without special procedure or consideration for up to 1 hour with no detrimental effect on the sensors or fluidics. The tubing should be flushed and the pump tubing relaxed prior to extended shutdown periods. Any shutdown period will require a 2-point calibration and verification by control material upon restarting operation.
1.1.7 General The analyzer utilizes an Intel 386EX processor, PCMCIA ROM programming. There are three RS-232 connectors, utilizing ASTM protocol, and one Bar Code reader connection. The display is a backlit liquid crystal (LCD) with type that is CCFT backlit with 320 x 240 pixel resolution. The internal optional printer is a 20 column thermal international font /166 dot graphics printer with ~5 cm (2 in) wide paper. NOTE:
1-2
The circuit boards use fine pitch surface mount technology. Field repair of these circuit boards is not possible. DO NOT attempt to repair any components on the board.
1 Introduction 1. Intro.
1.2
Cautions and Hazards There are NO user serviceable assemblies inside the analyzer. Only a trained, authorized service representative should remove the cover of this analyzer. WARNING: Removal of the top cover allows access to power supply voltages. Care should be taken to avoid electrical shock. Remove the power cord prior to accessing any internal assemblies. WARNING: Blood samples and blood products are potential sources of hepatitis and other infectious agents. Handle all blood products and flow path components (waste-line, capillary adapter, probe, sensor module, etc.) with care. Gloves and protective clothing are recommended. NOTE:
This International Caution Label appears on the rear of the pHOx Analyzer and means refer to the manual.
It is the responsibility of the service representative to decontaminate any assembly or analyzer being returned to Nova for repair or warranty claim. Decontamination of external surface of the flow path may be accomplished by a wipe down with cleaning agent or 10% bleach solution. The Internal surfaces of the flow path and tubing should always be considered contaminated. Prior to removal from the analyzer, ensure that the flow path and tubing are emptied. WARNING: Internal surfaces may have sharp edges. Care should be taken to avoid cuts and scrapes when accessing internal assemblies. Used tubing, biosensors, electrodes, reagents, controls, etc. may be disposed by normal laboratory waste procedures.
1-3
pHOx Service Manual 1.3
Required tools The analyzer is designed to require no unique or special tools. Flow path cleaning wires/kits are provided in the accessory kit. Avoid metal wires or other material that may scratch the internal surfaces of the flow path. • Digital Volt Meter (2 decimal accuracy) • Phillips® screwdriver (crosshead) preferably one with 15 cm (6 in) shank • Spring clip expander wrench or comparable means of releasing tension of retention clips • Pliers - needle nose or small snub nose
1-4
2 Product Description 2.0 Product Description
2.1
Mechanical Assemblies The analyzer is designed to calibrate using calibration standards and reference solution contained in a reagent pack. The tubing automatically connects with the pHOx when the pack is placed into the analyzer. Likewise, a space is provided for a plug-in control pack. The content and serial number of the pack is coded in a “Dallas Semiconductor Add-Only Memory,” which is read by a One-Wire MicroLAN® system. The analyzer has 4 major mechanical assemblies to move calibrator or sample in front of the biosensors.
Figure 2-1 Flow Diagram
2-1
2. Descrip.
The pHOx analyzer accepts syringe or capillary samples through a sample probe. The operator selects the sample mode via the keypad. The probe moves from its home position in the analyzer to an extended 45° position for syringe sampling. If the capillary sample is selected, the probe moves to a horizontal position that is back inside the capillary adapter.
pHOx Service Manual 2.1.1 Pump The analyzer uses one pump (PN 21492) for 2 functions. The pump aspirates either sample or calibration standards from the sample probe, through the sensor module/reference electrode, through a solenoid actuated pinch valve, and delivers to the waste bottle of the reagent pack. The second function of the pump is to aspirate the reference solution from the reagent pack and deliver it to the reference electrode through a solenoid actuated pinch valve.
2.1.2 Sampler Assembly The sampler assembly (PN 21504) utilizes a probe that rotates to 1 of 3 possible positions. The HOME position is vertical with the tip of the probe seated against a "Fluid Fountain” assembly that is the common port of the rotary valve. The syringe position is the probe tip extended out in front of the analyzer at approximately a 45° angle. The capillary position is horizontal, but the tip is totally inside the capillary adapter. This allows the capillary to be used without a separate adapter. (This position is also used to change the capillary adapter.) The capillary adapter is pushed onto the end of the sampler. Extending the probe outward from the capillary position ensures that the probe will be seated inside of the adapter. The sampler will exert enough force to push the adapter off of the sampler if it is not aligned correctly.
2.1.3 Rotary Valve Assembly The rotary valve assembly (PN 22215) has one common port connected to the sampler’s fluid fountain. Each calibration standard and quality control standard is connected to a port on the rotary valve. The rotor positions a central port connecting one of the fluids to the outlet port. The outlet port is connected to the Fluid Fountain by the S-Line. Starting the pump aspirates the selected fluid into the flow path. The rotary valve's manifold also has positions that allow air to enter the S-line. Refer to the replacement procedures for actual positions.
2-2
2 Product Description 2.1.4 Sensor Module Assembly
2.2
Electronic Assemblies.
2.2.1 Power Entry Module The power entry module (PN 22818) provides the interconnect to wall power. The fuses are found in the accessory kit for the operating voltage (120 VAC PN 22814, 236 VAC PN 22817). The power entry module connects to the power supply through TB1.
2.2.2 Power Supply The power supply (PN 21602) provides all of the power requirements for the analyzer. The analog and logic voltage used in the pHOx are + 5VDC. The stepper motors are driven by +24 VDC. This module has no repairable parts inside. The power supply connects to the digital control board by TB2.
2-3
2. Descrip.
The Sensor Module Assembly contains the measuring sensors, flow cell, and preheater. The measuring sensors clip into place upon insertion. The sensors have contacts instead of cabled connectors. When mounted, the sensor module forces the sensor contacts to align with the matching contacts on the left/right interconnect boards. The module also contains the light sources and detector for the SO2% channel and 2 air detectors. Please refer to the drawing section and Appendix A for individual part numbers.
pHOx Service Manual 2.2.3 Digital Control Board There are 2 Digital Control Boards: all units with monochromatic display use PN 24292, and all units with a color display use PN 40797. The main controlling board of the analyzer uses an Intel 386X processor with the following integrated peripheral functions: • Clock and Power Management unit • Watchdog Timer Circuit • Asynchronous Serial I/O unit • Synchronous Serial I/O unit • Parallel I/O unit • Refresh control unit • JTAG- Complaint Test Logic Unit The digital control board provides one 72 pin socket for a 4 MB SIMM module of RAM. This volatile memory stores the calibration, slope, and millivolt data. Major circuit identification: U25 is the main clock generator; it is crystal controlled 14.318 MHz and a phaselocked-loop frequency generator. This generates a 50.11 MHz CPU clock, 14.318 MHz for VGA video timing, and 1.84 MHz for Baud rate generation. U35 is the real time clock containing a battery which will retain the date/time, setup options, daily QC (up to 144 points), Error log (last 96 errors FIFO), and sample counter data. The battery has an expected 10 year life. U37 PEROM stores the monthly (256 data points) and YTD QC data, QC setup information, and recognition of the QC fluid pack presence. PCMCIA contains the operational program and boot program. This card must be in place for the unit to function. U43 is the video control chip. U26 controls the printer output and the I/O ports Comm 1 and Comm 2. U36 is the DUART that controls the Comm 3 I/O port. U16 provides the serial interface for the (RMS) interface and fluid pack memories. The mechanical devices are controlled by U18. The Keypad and “home” sensors for the rotary valve and sampler are ported through U18. The stepper motors are 4 phase stepper motors. U18 controls the speed and direction each phase is powered. U3 and U4 provide the power for each winding of the pump motor. U5 and U6 power each phase winding of the sampler motor. U7 and U8 power each phase of the rotary valve motor. U18 also provides the logic signal to operate the other mechanical devices. Typically, the devices are connected to either 12 VDC or 24 VDC and the logic provides a ground path through a transistor. The waste valve solenoid (12 VDC) grounds through Q3, the reference solenoid (12 VDC) through Q4, the sensor module lamp (24 VDC) through Q6, and the preheater (24 VDC) through Q2. The 24 VDC valve boost circuit is on the board but not used on the pHOx analyzer. 2-4
2 Product Description Power is received from the power supply through the J7 connector on the digital control board. Power Supply Connector TB2 5 Ground
6 Ground
7 +5 VDC
8 +5 VDC
Digital Board Connector J7 Pin 1 2 3 Ground +24 VDC. +5 VDC
4 +5 VDC
5 6 7 8 Ground +12 VDC. Ground -12 VDC
Sensing Wire Color Black Violet
White
Red
Black
Orange
Green
Yellow
Individual circuits are protected from an over current condition by the use of Positive Temperature Coefficient Resistors (PTC). These resistors act as a direct short if the current is below a set current. If the current increases, the resistors heat up and open, similar to a fuse. When the device cools off, the connection is remade, allowing the circuit to function. The following lists these devices: R22 - Solenoid common boost circuit (not used on pHOx) R24 - Heater power circuit R29 - Printer motor power R43 - Print Head heater power R44 - +24 VDC R45 - +5 VDC R71 - Bar Code Reader (+5 VDC out) R249 - Distribution board power (This board is mounted in the door.) The LED indicators on this board are mostly unused. All will light at power up and then go off. The exception is D8 - 100 Hz Clock. D9 - 386 Strobe will appear to be always on. D7 is your program fail indicator and will come on with a software reset. Connectors: J1 Bar Code Reader J2 COMM 1 J3 COMM 2 J4 COMM 3 J5 JTAG J6 PCMCIA J7 Power Supply J8 Mechanical Assemblies J9 Printer Connector J10 Analog Board J11 - Display/Keypad (Motors/heater/lamp/solenoids and door)
2-5
2. Descrip.
1 2 3 4 +12 VDC. -12 VDC +24 VDC. Ground
pHOx Service Manual 2.2.4 Analog Board There are 2 Analog Boards: pHOx, pHOx Basic, BiopHOx use PN 24291 and pHOx Plus, pHOx Plus L, and pHOx Plus C use PN 33390. U34 is the microprocessor that controls the signal acquisition and digital conversion. The signals are connected through several connectors. • Left Sensor Interface board J6 - PCO2, pH, Reference, Air Detector (ADT) 4 • Right Sensor Interface board J7 - PO2, Na+ • Sensor module J4 -SO2% detector, Preheater temp, ADT2, ADT3/Hct • Mechanical assemblies J3 - ADT1 Barometric pressure sensor U9 is on the analog board. The Analog to Digital Conversion (ADC) is performed by U31. The signals are multiplexed through U26 and U32. Preheater temperature control circuits are on the analog board. +24 VDC are ported through J3 to the SO2%/sensor module board (J4). The heater power wiring is protected on the digital control board by a PTC Resistor; the Heater has an in line thermostat (50°C) in the sensor module. The thermistor has 18.25 Kohms resistance at 37°C. The signal will change at a rate of 22.5 millivolts/0.1°C. There are 4 ADTs in the system. ADT1 holds the sample probe (40 uL setting). ADT2 at the inlet to the sensor module (70 uL). ADT3 is at the input to the sensors and is also used for the Hct. ADT4 is in the reference electrode housing. Logic and analog power is supplied to the analog board from the digital control board. Logic voltage (+5 VDC) enters J1 pin 14. The analog +5 VDC enter J1 pin 3 and -5 VDC J1 pin 4. The analog board provides power to the SO2 board through +5 VDC J4 pin 5 and -5 VDC J4 pin 6.
2.2.5 SO2 Control Board SO2 Control Board is an integral part of the Sensor module. It controls the 2 LED light sources that glow at 880 nM and 660 nM, respectively.
2-6
2 Product Description 2.2.6 Electrode Interface Boards The electrode interface boards are found only on pHOx and pHOx Basic.
The Electrode Interface Boards have 2 part numbers: PN 20641 (Left) and PN 20674 (Right). The electrodes are wireless. The electrodes have point contacts which engage the Interface board contacts when inserted into the sensor module. The left interface board, connecting to the analog board through J6, has 3 sets of contacts: one set for the reference electrode/ADT4, one set for Na+, and one set for PCO2 sensors. The reference electrode/ADT4 has 3 individual point contacts while the Na+ and PCO2 have 2 individual point contacts. The right interface board has 2 sets of contacts: one for the pH and one for the PO2 sensors; each consisting of 2 individual point contacts. This board is connected through J7 to the analog board.
2.2.7 Door Assembly Electronics The door assembly electronics include the Display Assembly, the Distribution Board, and the Keypad Assembly. The digital control board sends/receives signals to the door via the flat cable J104 on the distribution board. NOTE:
See Section 3.13 for part numbers that are model specific.
The keypad, which is glued onto the door, connects to the distribution board via J102 and has a ground cable clamped to the conductive surface on the inside of the door. The Analyze key also connects to the distribution board through this connector. The distribution board also provides the signal path to the display assembly and power to the display back lighting lamp. The display is a liquid crystal, graphics display. It has a 320 x 240 dot matrix (1/4 VGA) yielding a 122 x 92 mm (4.99 x 1.33 inch) viewing area. This translates to a 40 x 20 character display using a standard character set. The display is backlit by cold cathode fluorescent tube (CCFT). An 8 ohm voice coil loudspeaker is mounted on the distribution board, producing all associated sounds.
2-7
2. Descrip.
NOTE:
pHOx Service Manual
2-8
3 Replacement Procedures 3.0 Replacement Procedures CAUTION: Assemblies may have loose hardware mounting. Ensure no loose hardware is left in the unit before restoring power to the analyzer.
Cover Removal 1. 2.
3.
4.
Turn the unit OFF, disconnect the line cord, and have the back of the unit facing you. R e m o ve t h e 4 Screws (2) rear-panel-cornermounting screws. Lift the printer cover and remove the 2 cover mounting screws. Slightly spread the bottom sides and slide the cover toward the back. This Mount will disengage the Screws (4) internal clips at the front of the cover. Lift the cover off the unit.Take care not to damage any cables or wiring during removal. Figure 3-1 Cover Removal
3-1
DWG #10-1029A
3.1
3. Replace.
WARNING: Removal of the analyzer cover will allow access to line voltage wiring. Power should be turned off and the power cord disconnected before any assembly replacement procedure is performed.
pHOx Service Manual Tubing Harness 1. 2.
Remove the analyzer cover per Section 3.1. Remove the digital control shield. There are 4 mounting screws: One at the top front One at the bottom front Two at the bottom rear (above and below the PCMCIA card)
NOTE:
Take care not to lose the spacers on the screws.
NOTE:
It is optional to remove the digital control board and mount plate when accessing the reference and waste line. To remove the digital control board and mount plate, follow these steps:
a.
Disconnect the electrical connectors (J7, J8, J9, and J10) from the top of the board.
b.
Disconnect J11 connector at the bottom front of the board.
c.
The digital control board and mount plate come off as one assembly. There are 3 mounting screws: two on the rear panel and one at the top front on the printer box. (Take care not to damage the board.)
d.
Slightly lift the board and mount plate upward and out from the side of the analyzer.
e.
Remove the ground strap on the back of the board.
Digital Control Board Mount Plate
Mount Screws (3) (4)
Digital Control Board
DWG #10-1030A
3.2
Figure 3-2 Removing the Digital Control Board/Mount Plate 3.
3-2
Remove the R and W-lines from their front panel fitting. (Looking at the front panel, R is on the left.) If you have not removed the digital control board, it may be easier to remove the panel fitting nut on the front panel and move the fitting to the new tubing.
3 Replacement Procedures 4.
Carefully remove the tubing from the rotary valve including the center port to fluid fountain. Take care not to damage the valve ports. Push the tubing end off the port to reduce the risk of damaging the valve. 3. Replace.
G DW
0 0-1 #1
A 40
NOTE:
Figure 3-3 Removing Tubing from Rotary Valve
6. 7.
8. 9.
NOTE:
The tubing bracket is held in place on each end by a screw. Two screws hold the bracket in alignment; access them from the bottom of the analyzer. Remove both screws. Lift the bracket and tubing set, removing it from the sampler side of the analyzer. Slide the new assembly into the analyzer and reconnect the W and Front Panel W and R Fittings R-tubing to the front panel ports. WR Ensure that they go to the correct ports and brackets. Reinstall the 2 mount screws at the bottom of the analyzer and the 2 end mount screws. Place the reagent tubing and S-line W onto their ports on the rotary valve. R Take care that the metal ports are not damaged. DWG #10-1039A
5.
The A line connects at the lower left (7 O’clock position). Figure 3-4 Reconnect W and R-lines
10. If the control board and mount plate were removed, replace them. 11. Reinstall the cover and restore the unit to operation.
3-3
pHOx Service Manual 3.3
Power Supply Removal The Power Supply (PN 24720) is removed as follows: WARNING: Unplug the Analyzer before proceeding. 1. 2. 3. NOTE: 4.
Remove the analyzer cover per Section 3.1. Disconnect the power harness connectors: P1 and P3 (top of the power supply assembly). Remove the 4 power supply mounting screws from the rear analyzer panel. Support the power supply before removing the mounting screw. Carefully lower the supply. Rotate it to access the line input module wiring (TB1) on the bottom of the power supply.
P3
P1
Use Center Holes to Mount
Rear Panel Power Supply Mount Screws (4)
DWG #10-1038B
Power Supply
Power Plug On/Off Switch Figure 3-5 Power Supply Removal 5. 6.
3-4
Disconnect TB1 and remove the power supply. Position the new power supply with the lower side toward the analyzer. Plug in TB1.
3 Replacement Procedures NOTE:
NOTE:
NOTE:
9.
The holes closest to the middle of the supply bracket are the ones used to mount the supply.
Brown
Connect P1 and P3. The wires leave the top of P1. The white wire of P3 goes to the digital display board P7.
Black
Green & Yellow
TB1
Guide
Power Supply
Reinstall the unit cover and power cord.
Figure 3-6 Power Supply Position
3.4
Sampler Assembly The Sampler Assembly (PN 23899) is removed as follows: 1. Remove the probe and the air detector. a. From the Menu screen, select Flowpath/Probe Maintenance and press Enter. b. Press Move Probe (soft key). Open the door. c. Remove the capillary adapter from the front of the probe by gently pulling it off. d. Disconnect the air detector's sample line from the sensor module. e. Disconnect the 2-prong cable. f. Push the air detector down and pull the air detector with the probe out of the sampler assembly. WARNING: Unplug the Analyzer before proceeding. 2. 3. NOTE:
Remove the analyzer cover per Section 3.1 Cut the tie wraps on the rear of this assembly. Care should be taken to avoid cutting wires.
3-5
3. Replace.
8.
Position the new power supply to align with the mounting holes in the back panel.
DWG #10-1037A
7.
TB1 male end has 3 wires. The female end has 5 sockets. The Green/Yellow wire is closest to you.
pHOx Service Manual 4.
5. STOP:
Disconnect the following connectors: J5 Motor cable connector J10 Home detector P6 door switch cable Sensor module light cable Ground wire Remove the S-Line from the fluid fountain assembly. There are washers and nuts that are NOT captive. Do not lose any hardware into the unit. Flow Cell Light Source
Mount Screw
Mount Screws
Ground
Power Supply
S-Line
J5, J10
DWG #10-1032B
CN3
Figure 3-7 Sampler Assembly
3-6
3 Replacement Procedures 6.
DWG #10-1034B
NOTE:
Remove sampler subassemblies. a. Motor assembly 1. From the Gear set side of the assembly, remove the flywheel (solid metal disk) by loosening the set screws on its side. 2. On the motor side remove the 2 mount screws. 3. Remove the motor (some Mount Screws tie wraps may need cutting). Install the new motor by reversing these steps. Use new tie wraps as required. b. Home detector (PN 21504) 1. Remove the sensor module light and bracket. Touching the light glass will shorten use life.
DWG #10-1035A
8.
Figure 3-9 Removal of Motor Assembly 3-7
3. Replace.
7.
Remove the 4 mounting screws. Mount • One at the sensor modScrew ule light bracket to front panel • One at the lower front of bracket • Two at the rear of the sampler bracket, they Mount Screws also mount the rotary valve. If NOT replacing the entire assembly, skip to Step 8, otherwise: a. Install the sensor module light bracket onto the new sampler asMount Screw sembly. b. Position the new assembly onto the unit. Install the mounting screws and tie wraps. Figure 3-8 Removal of Sampler Assembly
pHOx Service Manual 2. 3. 4.
NOTE:
Remove the 2 mounting screws from the home detector. Install the new detector by reversing these steps. Use new tie wraps as required to prevent fouling of the wires. After restoring power, ensure the capillary adapter seals against the Fluid fountain in the Home position. If not, adjust the Home flag on the top gear. Ensure the home flag enters the detector without obstruction.
DWG #10-1045A
C. Gear Set (PN 22634) 1. The gears are held in position by compression springs. These can be removed by spreading the open end and lifting. 2. Remove the old gears noting the position and orientation of Gear each gear. The noise Compression Spring damper is behind the middle gear. Spread Compression 3. After installing the new Spring to loosen gears, ensure that the Post gears run freely. Figure 3-10 Removal of the Compression Spring
3-8
3 Replacement Procedures
DWG #10-1041A
Figure 3-11 Fluid Fountain Assembly
3-9
3. Replace.
D. Fluid Fountain Assembly (PN 24620) 1. Rotate the gears so the capillary adapter lifts off the fluid fountain assembly. 2 . Remove the S-Line from the bottom of the spike assembly. 3. Loosen the mounting nut (under the Capillary sampler bracket). Adapter 4. Slide the fluid fountain assembly off the sampler. 5. Move the washer and mount nut to the new assembly. Fluid 6. Slide the new assemFountain bly onto the sampler. 7. Rotate the sampler Sampler gears until the capillary presses down on the fluid fountain Washer assembly. Nut NOTE: The fluid fountain should be centered and should move S-Line downward between 0.06 to 0.08 inches (1.5 to 2.0 cm).
pHOx Service Manual Rotary Valve Assembly The Rotary Valve Assembly (PN 23898) is removed as follows. 1. Remove the cover per Section 3.1. 2. Optional: Perform this step to enhance accessibility. Remove the sampler assembly per Section 3.4. 3. Remove the 7 tubes and SLine from the rotary valve. CAUTION: Care should be taken not to scratch the metal tubes of the valve. 4. 5. 6. NOTE:
Mount Screws (2)
Disconnect the motor connector J2 and the optical detector connector J9. Loosen the 2 mounting screws at the bottom of the assembly. Slide the assembly toward the front and lift. The front of the valve bracket is hook shaped. Slide the assembly toward the back of the unit to clear the sampler bracket. G DW
#1
0-1
0 04
A
7.
Flow Cell Light Source
DWG #10-1036B
3.5
Figure 3-12 Removing Rotary Valve Assembly 8. NOTE:
3-10
If replacing the entire assembly, install the new rotary valve assembly by following Steps 1 thru 7 in reverse order. Ensure the red, nonmetallic washers for the sampler mount screws stay in place during the remounting process. If only repairing the rotary valve assembly, refer to Section 3.6. If repairing the rotary valve motor, refer to Section 3.7.
3 Replacement Procedures Valve Assembly The Valve Assembly (Valve Repair Kit PN 24626) is removed as follows: NOTE: The rotary valve assembly must be removed first before proceeding. (If not already removed, see Section 3.5.) 1. Remove “C” clamp on the valve shaft and remove the drive gear. CAUTION: Do not lose the shaft pin when removing the drive gear. With the drive gear removed, access and remove the 2 optical-detector mount screws. Remove the optical detector. Remove the 4 screws from the compression spring washer. Remove the washer and compression spring. Slide the rotor and tube manifold out of the valve body.
3. 4. 5. 6. NOTE:
Before proceeding, note the alignment pin in the valve body that aligns with the slot in the manifold.
NOTE:
Before installing the new parts, lubricate with a small drop of sealing oil on the mating surfaces.
7.
Slide the new rotor into the valve body. The large flange side goes in first. Align with the drive pins.
B C
A
D
S
J9 DWG #10-1033A
E
G F
Figure 3-13 Valve Assembly Removal
3-11
3. Replace.
2.
DWG #10-1019A
3.6
pHOx Service Manual 8. 9. NOTE:
Align the manifold with the pin and slide it into place. Reinstall the compression spring and washer Ensure the drive gear goes all the way over the shaft pin.
10. Install the “C” clamp on the drive gear shaft. 11. Install the valve by reversing Steps 1 - 10 above. 12. Install the rotary valve assembly. See Section 3.5 and reverse Steps 1 thru 7.
3.7
Rotary Valve Motor WARNING: Unplug the power from the analyzer. The Rotary Valve Motor (PN 24627) is removed as follows: NOTE: 1. 2. 3. 4. 5.
The rotary valve assembly must be removed first before proceeding. (If not already removed, see Section 3.5.) Remove the drive gear/opto detector as in Steps 1 and 2 in Section 3.6. Remove the 4 mounting screws from the motor. Remove the fly wheel from the motor shaft byFlow loosening the Source Allen (hex) Cell Light screw. Mount the new motor and opto detector. Reinstall the rotary valve assembly. See Section 3.5 and reverse Steps 1 thru 7.
Access Holes in Gear
Optical Detector DWG #10-1026B
DWG #10-1027A
Optical Detector on Reverse Side
Mounting Screws
Figure 3-14 Optical Detector
3-12
3 Replacement Procedures Pump Assembly The Pump Assembly (PN 23897) is removed as follows: 1. Remove the analyzer cover per Section 3.1. 2. Remove the digital control board shield. There are 4 mounting screws: One at the top front One at the bottom front Two at the bottom rear (above and below the PCMCIA card)
3.
Take care not to lose the spacers on the screws. Remove the digital control board and mount plate. a. Disconnect the electric connectors (J7, J8, J9, and J10) from the top of the board. b. Disconnect J11 connector at the bottom front of the board. c. The digital control board and mount plate come off as one assembly.There are 4 mounting plate screws: two WR on the rear panel and two under the printer box.
NOTE:
Take care not to damage the board. d.
Slightly lift the mount plate and the board upward and out the side of the analyzer.
Pump Assembly W R
Figure 3-15 Pump Assembly 4. 5. 6. 7.
Disconnect the motor connector J1. Remove the pump cap (on the front side of the pHOx) by pulling it off. Remove the 4 corner mounting screws then remove the pump. Install the new pump by reversing the above steps.
3-13
3. Replace.
NOTE:
DWG #10-1048A
3.8
pHOx Service Manual 3.9
Digital Control Board WARNING: Unplug the power from the analyzer. The Digital Control Board is removed as follows:
1. 2.
Units manufactured after June 1, 2005, use PN 40797; Units manufactured through June 1, 2005, use PN 24292.
Remove the analyzer cover per Section 3.1. Remove the digital control board shield. There are 4 mounting screws: -One at the top front, -One at the bottom front -Two at the bottom rear (above and below the PCMCIA card)
NOTE:
The shield is not present on units manufactured after June 1, 2005.
Mount Screws (4)
Shield
DWG #10-1047A
NOTE:
Figure 3-16 Digital Board Removal
3.
4.
Take care not to lose the spacers on the screws.
At the top of the board, disconnect the electrical connectors: • J 7 - Power Supply • J 8 - Mechanical devices harness • J 9 - Printer (Pull side tabs outward to release the cable.) • J10 - Analog Board At the front lower corner disconnect J11 – Display/ PCMCIA keyboard (pull side tabs Release outward to release cable). Button
Mount Screws
Digital Control Board
Figure 3-17 Install PCPMIC Card
3-14
Shield Stand Offs
DWG #10-1046B
NOTE:
3 Replacement Procedures 5. 6.
7.
3.10 Analog Board The Analog Board is removed as follows:
1. 2. 3.
Basic/BiopHOx, pHOx/pHOx use PN 24291; pHOx Plus, Plus L,Plus C use PN 33390; all jumpers will be in the 3 to 4 position. To access the top/rear mounting screw, use a long shank (15 cm or 6 inch) Phillips® head (cross head) screwdriver. Remove the analyzer cover per Section 3.1. Disconnect the sampler ground strap. Disconnect the cable connectors from the Analog board Sensor Interface board connectors: J6 - 3 sensor board J7 - 2 sensor board J4 - Sensor module/SO2 cable J3 - RMS/Fluid Deck (Pull side tabs outward to release cable.) J1 - Digital control board
J4 J6 J7
J3
Analog Board Mount Screws
J1
DWG #10-1049B
NOTE:
Figure 3-18 Analog Board Removal 3-15
3. Replace.
8. 9. 10. 11. 12. 13. 14. 15.
Remove the PCMCIA card (PN 22805) Push the button below the PCMCIA card (bottom rear corner) to release the card. Remove the 2 mounting screws and 2 shield standoffs: One screw is at the top rear corner One screw is between J8 and J9 connectors One shield standoff is at the top front and the other is at the bottom front Remove the 8 COM connectors screw standoffs at the rear of the pHOx by unscrewing 2 standoffs from each COM port. Remove the board. Position the new board on the mount plate. Install the 2 mounting screws and the 2 shield standoffs. Install the 8 COM connector screw standoffs. Reinstall the PCMCIA card into its slot. Reconnect all the cable connectors. Reinstall the digital board shield (4 screws). Reinstall the cover and restore the analyzer to operation.
pHOx Service Manual 4. 5.
Remove the 3 mounting screws: one in each corner of the board. You may need to place a screw driver through the holes in the power supply to access the screw in the upper back corner of the board. Install the analog board by reversing the above steps.
3.11 Printer Assembly WARNING: Unplug the power from the analyzer.
DWG #10-1050A
The Printer Assembly (PN 24615) is removed as follows: 1. Remove the analyzer cover per Section 3.1. 2. Remove the digital control board per Section 3.7. 3. Open the front door; remove the 2 screws on top of the printer cover plate; remove the printer cover plate by pushing the plate from the inside outward; then remove the paper and roller bar. 4. Remove the 3 print Printer Cover head mounting screws: two at the top of the print head 2 Screws and one at the lower right. 5. The flat ribbon cable from the print head sits in a connector. Slide the top of the connector upward to release the cable. The connector stays on the board. Tilt the print head from the assembly. Remove the paper advance motor cable that is connected into the printer control board. Figure 3-19 Printer Cover Removal 6. NOTE:
3-16
The Printer Control Board (PN 24614) is mounted to the paper box by 2 screws that fit into the slots. Loosen the screws and slide out the board. Save cable that goes from the printer control board to J9 of the digital control board.
3 Replacement Procedures 7. 8. NOTE:
When installing the new assembly, install the printer control board first. Do not tighten the 2 slotted screws. This allows access when connecting the print head cable. Attach the paper advance motor cable to the printer control board. Install the print head. Connect the flat ribbon cable. The easiest way to plug in the print head cable is using one finger on each side. Tighten the 3 print head screws. The print head cable is clamped in place. Lift up on the white connector to release the flat cable for easy removal or replacement.
9.
Use 2 Fingers to Connect Printer Cable to Printer Board
DWG #1
0-1051A
3. Replace.
After connecting the print head, tighten the printer control board mounting screws inside the paper box. 10. Attach the cable from the printer control board (key faces up, blue or black wire faces towards the front of unit) to the digital control board. 11. Reinstall the digital control board, shield, printer cover, analyzer cover, and paper. Restore unit to operation.
Figure 3-20 Attaching the Printer Cable
3.12 Sensor Module Light Source Flowcell Light Source Mount Screws
DWG #10-1052A
The Sensor Module Light Source (PN 24717) is removed as follows: 1. Remove the analyzer cover per Section 3.1. 2. Remove the 2 screws of the light source mounting bracket. 3. The light bulb is removed by pushing it into the socket, rotating it counterclockwise a 1/4 turn, then releasing it.
Figure 3-21 Removing Mounting Screws CAUTION: Do not touch the glass surface of the new light. 4. 5. 6. 7.
Place the glass end of the new light bulb into the tubing provided. Insert the light bulb into the socket. Push it in and rotate it clockwise a 1/4 turn. Then pull off the tubing. Reinstall the bracket. Reinstall the covers and restore the unit to operation. 3-17
pHOx Service Manual
Push and rotate 1/4 turn counterclockwise to unscrew light.
DWG #10-1053B
Push plastic tube over lamp . Do not touch glass of new light bulb.
Push and rotate 1/4 turn to clockwise to install new bulb into socket Figure 3-22 Tubing for Light
3.13 Door Removal /Reinstallation Procedure
CAUTION: Until free of the hinge and cables, support the door to prevent stress to other cables/ assemblies.
105 4B
Use a pencil to 3 Mounting trace the position Screws of the hinge onto the door mounting plate.
10-
STOP:
Turn the unit OFF and disconnect the line cord.
DW G#
1.
Figure 3-23 Door Removal
3-18
3 Replacement Procedures
NOTE:
4.
Remove the 3 hinge screws from the door plate. Remove the 5 plate-mounting screws: one in each corner, plus one at right side. Care should be taken not to strip the threads.
Keypad
Display Shield
5 Plate Mounting Screws Door
Distribution Board
Figure 3-24 Door Removal Continued 5.
Reinstall the Door by reversing the above steps. Use the pencil marks made on the door plate to realign the door to the hinge. Ensure the door opens and closes easily and aligns with the chassis. Analyzer pHOx pHOx Basic/ BiopHOx pHOx Plus pHOx Plus L pHOx Plus C
PN of Door 21493 33102 30817 34082 35350
3-19
3. Replace.
Remove the plate and disconnect the flat cable (J101). This frees the door. Place it face down on a protective surface to prevent scratching.
DWG #101055B
2. 3.
pHOx Service Manual 3.14 Display Assembly The Display Assembly is removed as follows: Prior to June 1, 2005 Display Assy. 24716 Door Distribution Board 24727 Digital Control Board 24292 Door Interface Cable 21983 1. 2. 3. 4.
June 1, 2005 forward 40910 40795 40797 40495
Remove the analyzer door assembly per Section 3.13. Disconnect the cable from the display board (CN2). Disconnect the light source, J103, on the distribution board. Remove the 4 corner mounting screws.
CAUTION: When reinstalling the mount screws, DO NOT overtighten. Assure the door plastic is not stripped. The screws will tighten in about one turn clockwise. 5. 6.
Position the new display and install the 4 mounting screws. Reconnect the cables CN2 and J103. Reinstall the door per Section 3.13. CN2
Keypad J104 DWG #10-1065A
J101
J102
J102 Light Source J103 Keypad Clip Ground Strip
Figure 3-25 Display Removal
3-20
3 Replacement Procedures 3.15 Distribution board
CAUTION: When reinstalling the mount screws, DO NOT overtighten. Assure the door plastic is not stripped. The screws will tighten in about one turn clockwise. 5. 6.
Reconnect cables J104, J103, J102, and J101. Reinstall the door per Section 3.13.
The Keypad Assembly (PN 24625) is removed as follows: 1. Remove the analyzer door per Section 3.13. 2. Disconnect the cable connector J102 from the distribution board. 3. Remove the clip holding the keypad ground strip to the door. Save this for reuse. 4. The Keypad is held by adhesive to the front panel. Peel the old assembly from the front of the door. 5. Clean the surface of any debris. 6. Remove the protective backing from the new keypad. 7. Slide the cable/ground strap through the hole in the door. Align and press new keypad in place. 8. Fold the ground strap over the ledge and reinstall the clip. Keypad Clip 9. Plug in J102 and reinstall the door per Ground Section 3.13. Strip
DWG #10-1056A
3.16 Keypad Assembly
Figure 3-26 Keypad Removal
3-21
3. Replace.
The Distribution Board (PN 24727) is removed as follows: 1. Remove the analyzer door per section 3.13. 2. Disconnect the cable connectors J104 and J103 coming from the display board, J101 coming from the digital control board (should already be disconnected), and J102 coming from the Keypad. (Refer to Figure 3-25.) 3. Remove the 4 corner mounting screws. 4. Position the new distribution board (J101 will be on the hinge side of the door) and remount.
pHOx Service Manual 3.17 Sensor Module Assembly WARNING: Unplug the power from the analyzer.
NOTE:
The preheater and SO2 component are all part of this assembly.
R
DWG #10-1059A
The Sensor Module Assembly is removed as follows: 1. Remove the sensors (electrodes) and sample probe/ADT1 assemblies. 2. Remove the reference electrode assembly. 3. Rotate the retaining clamp on the top and the bottom of the assembly. 4. Pull the assembly toward you. 5. Remount by reversing these steps.
W
Figure 3-27 Removing the Sensor Module Sensor Module Part Numbers Analyzer pHOx pHOx Basic/BiopHOx pHOx Plus pHOx Plus L pHOx Plus C
3-22
Part Number 21494 34653 27807 34738 37343
3 Replacement Procedures 3.18 Electrode Interface Boards This section pertains to the pHOx, pHOx Basic, and BiopHOx Analyzers only. WARNING: Unplug the power from the analyzer.
NOTE:
3 Mount Screws to Remove Electrode Interface Board Bracket
DW G#
10-1
057
A
The top left corner screw was removed with sampler removal.
Figure 3-28 Removing the 3 Mount Screws 6.
Pull the electrode interface board out thru the inside of the pHOx.
3-23
3. Replace.
The Electrode Interface Boards (PN 24719 {left} and PN 24718 {right}) are removed as follows: 1. Remove the analyzer's cover per Section 3.13. 2. Remove the sampler assembly per Section 3.4. 3. Remove the sensor module assembly per Section 3.17. 4. Remove the spring loaded sensor module retaining clips (front of pHOx). 5. Remove the 3 remaining mount screws.
pHOx Service Manual 7.
Each interface board is held to the plate by 2 screws. Ensure that the 3 connector board is on left, and the 2 connector board is on the right when looking at it from the front.
3 Connector Board
DWG #10-1058A
2 Connector Board
Figure 3-29 Correct position of Interface Board 8. 9.
Disconnect the affected board from the analog board: 3 connector board to J6, 2 connector board to J7. Install new assembly by reversing the above steps.
3.19 Pinch Valve Assemblies
0-1059A
#10 Screw
DWG #1
The Pinch Valve Assemblies (PN 16928) are interchangeable. The assembly plugs into the front panel and is held in by one screw.
#10 Screw
Figure 3-30 Pinch Valve Assemblies
3-24
4 Troubleshooting 4.0 Troubleshooting Information This section contains suggestions for individual error codes, cycle information to help identify the cause of an error condition, and hints on basic troubleshooting. 1.
Troubleshooting is recognizing the variance from normal operation. The calibration cycle can be used as the sequence of events where the cycle response and results are known. Fluid Function all 4 ADTs, Hct, PCO2, Na+, K+, iCa, Cl-, Gluc, Lact PCO2 pH, Na+, K+, Cl-, Gluc, Lact, (iCa on pHOx +L) all ADTs, pH, Hct, (iCa on pHOx+C)
4. Trouble.
Start cycle Std. A Std. B Std. C Std. D
PO2 uses Room Air (A-air position of the rotary valve). SO2/Hb utilizes external calibrators. (These are model specific.) 2.
Sensor Methodology ISE sensors - Na+, K+, iCa, Cl-, pH, and PCO2 ISE requiring a reference signal - Na+, K+, iCa, Cl-, pH ISE with its own internal reference - PCO2 Reflected Impedance - Air Detectors 1,2,4, 3/Hct Amperometric - PO2, Glucose, Lactate Optical Reflectance - SO2%
3.
Calibrator Functions other than calibration. PO2 Std. B is used to flush the system. PO2 is monitored on all fluids in the Calibrator pack. If the PO2 millivolt reading varies from that of room air, a 72 Delta mV error is generated indicating a contaminated calibrator pack. Std. D has surfactant and is used to clean the flowpath between pump cycles.
4-1
pHOx Service Manual 4.
Dependencies (Refer to the quick reference guide in the appendix.) Hct Calibrated Na+ required
NOTE: The pHOx Na+ is a 30 day expendable electrode. SO2% Hb PCO2
Calibrated Hct and PO2 required Calibrated Hct and SO2% required Calibrated pH required
Calculated values will not be reported if a needed test is suppressed. P50 is reported only if PO2 is between 30 mmHg (4 kPa) and 75 mmHg (10 kPa). 5.
Password protection - One password for Set Up, QC, and Service menus. Should the password be lost, enter the following number when asked for the password. Date Code “Back-door” Password = (((month x 31) + day) x 137) + (year x 16)) + 5987 2 digits 2 digits last two digits Example: If today is 02 January 2003 the “back-door” password would be: Password = (((01 x 31) + 02) x 137) + (03 x 16)) + 5987 = (33 x 137) + 48 + 5987 = 4521 + 48 + 5987 = 10556 Change the password to a known value and write it down.
4-2
4 Troubleshooting 4.1
Service Screens Pressing the “MENU” soft key from the “Ready Screen” displays an “Operation Menu.” Note that the soft key labels change. The “QC” label changes to “Service” (second key from the left). Pressing this key will display the Service Menu. If the home key is pressed, the system returns to the “Ready Screen.”
Manual Device Control
Service Menu Use
to select. Press
to change / enter.
System T est
Active analog input signal (1 - 25)
Analog Input - Real Time mV
Calibration data
Sensor Subsystem
Printer enable, Error log, System configuration print
Printer Menu
Error log (Last 100 FIFO)
I/O Signals / Loop Back Test
Error Log Comm unications Tests
Home
Exit
4-3
4. Trouble.
Manual Device Control + Flow test
pHOx Service Manual 4.1.1 System Test Screen System Test Rev: 123.456 Measured Parameters/ Current signal (millivolt)
Use to select. Press to select / enter. pH: +047.43 LED1: +314.3 Na: +147.4
AD3: +314.3
PCO2: -082.12 LED2: +314.3
AD4: +314.3
PO 2: -051.20
Device Names (column 1&3) / Current state of the device (column 2&4)
11-23-96 10:54 am
AD1: +314.3
HCT: +007.43 AD2: +314.3
Rot. Valve:
STD A
Sampler:
Pump:
CTRL 1
SO 2% LEDs:
Waste valve: Ref. valve: Home
Open Closed Exit
Temp:
Home On
Air Osc:
On
X level:
Hct
Operator Exit Flow Test
37.0
Service Flow Test Test Flow
Manual Control of a mechanical device: • Move the cursor over the device /Press “ENTER.” • The Option selections will appear. (If there are only 2 options, the device will toggle its state when the “ENTER” key is pressed, i.e., solenoid valves.) • Move the cursor over the option wanted /Press “ENTER.” NOTE:
To successfully move fluids, a knowledge of the complete flow path is required.
4.1.1.1 Operator Flow Test Pressing the soft key under the Operator Flow Test turns on the pump and raises the sample probe. Use the guide in this Section titled Flow Problems in Conjunction with the Operator Flow Test.
4-4
4 Troubleshooting 4.1.1.2 Service Flow test
4-5
4. Trouble.
Pressing the soft key under the Service Flow Test brings up a screen that displays the results of this test. On this screen, pressing the soft key under Run Test causes the unit to pump each fluid twice: once with the Reference valve shut and once with the Reference valve Open. The expectation is that the flow time will increase with the decrease in vacuum caused by pumping reference solution into the waste line junction in the reference electrode. Ideally the 2 pump windings are equal and the flow time will double with the reference solenoid valve open. The pumping times will yield 2 pieces of information: 1. The individual flow times appear at the end of pumping that fluid. 2. All the fluids will pump at the same rate. Usually Standard D is the one standard that varies in flow rate from the other fluids. A good rule of thumb for the flow rates is pHOx Basic/BiopHOx and pHOx is in the 1000 – 1200 range. pHOx Plus/C is in the 1200 – 1300 range. pHOx Plus L pumps in the 1300 – 1400 range.
pHOx Service Manual At the end of the Test, the Average flow time for Reference Valve Closed and Reference Valve Open is displayed. A ratio of the average times is presented. The ratio should fall between 0.5 – 0.65. • Ratios approaching 1 indicate the loss of reference flow. • Ratios approaching 0.0 indicate an obstruction or air leak in the system before the flowpath reaches the Reference electrode. FLOW TEST RESULTS
CAL A 1180 CAL B 2010 CAL C CAL D 0.58 QC 1 QC 2
1190
AVG FLOW 1
1195
AVG FLOW 2
1190 1170
RATIO
Average Flow time Reference Valve Shut Average Flow time Reference Valve Open Flow 1/Flow 2 (nominal 0.5 – 0.65)
1195 1193
QC 3
1195 Print
EXIT
Run Test
4.1.2 Analog Input Screen Selecting this option, displays the 25 possible input channels. The display is active. Chnl Signal Unit Chnl Signal Unit PO2 pHOx+, pHOx+ C 1 Analog GND ALL 15 + 2 K pHOx+,+L,+C 15 Lac pHOx+L pHOx+,+L 16 Glu pHOx+,+L,+C 3 Cl-/Ca++ 3 ClpHOx+C 17 PO2 pHOx+L ++ 4 Ca pHOx+C 18 PO2 pHOx+,+C 5 pH pHOx 18 Lac pHOX+ L pHOx+,+L,+C 20 Temp 0.00=37C 5 Na+ pHOx 22 LED1 ALL 6 Na+ 6 pH pHOx+,+L,+C 23 LED2 ALL 7 PCO2 All 24 ISE Ref ALL PCO2,Ref ALL 8 Barr.Press. All 25 9 +5V Ref All 13 HCT pHOx, pHOx+C 14 PO2 pHOx, pHOx+C 10 Voltage Monitor Sums +24 and +5 4-6
4 Troubleshooting 4.1.3 Sensor Subsystem This set of screens shows the calibration data for each analyte measured by the pHOx. Use the “NEXT” soft key to page to the analyte of interest. Last 2-point calibration data Millivolts for each standard Slope calculated Last 1 point data (Millivolts should be similar to the same Standard in 2 point cal)
pH Sensor Callibration & Analysis Data mV
mV
STDA
+ 43.1
STDA
STDB
+ 43.1
SAMPLE
STDC
+ 43.1
CONC
STDD
+ 43.1
SLOPE
+ 43.1 + 43.1 97.4
- 7.1
Next Page
Exit
Print
Sample millivolts and concentration - Unknown and therefore not relevant to troubleshooting Useful to the clinician for any possible dilution SO2 LED 1
SO2 sensors - Each LED has its own screen.
Displays similar data to the other Measured parameters, plus
The “dark” (off) and “Light” (full) Signals are shown
Callibration & Analysis Data STD1
mV + 379.1
SAMPLE
STD2
+ 379.1
CONC
DARK
+ 379.1
LIGHT
+ 379.1
SLOPE
+11.4
Home
Print
Exit
AIR + 117.0
FLUID
THRESHOLD
AD1
+ 54.0
+ 99.0
mV
+ 379.1
Next Page
Air Detector Subsystems Air Detectors
Displays the millivolts for Air and Liquid. (Air column is actually Standard D.) Shows the AIR/Liquid threshold (set by the analyzer) millivolt level NOTE:
Numbers shown on this screen are not real values.
Callibration Data
AD2
+ 117.0
+ 54.0
+ 99.0
AD3
+ 117.0
+ 54.0
+ 99.0
AD4
+ 117.0
+ 54.0
+ 99.0
Exit
Next Page
Home
Print
4-7
4. Trouble.
Home
pHOx Service Manual 4.1.4 Printer Menu This menu is straight forward. Move the cursor to the desired action and press “ENTER”
Printer Menu Use
to select. Press
to change / enter.
Enable/Disable internal printer
Printer: Enabled
Prints last errors
Print System Error Log Print System Dump
Prints setup configuration and millivolt data
Character Set Test
Test Print Home
4-8
Cancel Printout
Exit
4 Troubleshooting 4.1.5 Communication Tests This screen allows you to transmit out a particular port for testing. Likewise using a test plug you can perform a Loopback test for each I/O function. Communications Tests Use to select. Press
to change / enter. Loop Back Tests Send test characters.
4. Trouble.
Home
Exit
Loopback Tests Use
to select. Press
to change / enter.
COOX
1. Select Port and install loopback plug.
ASTM
2. Press to begin testing.
Remote
Tests Completed: 0
Bar Wand
RXD
DTR
RTS
TXD DSR DCD CTS RI Errors
Home
0
0
0
0
0
Exit
Loopback Tests Use
to select. Press
to change / enter.
COOX
1. Select Port and install loopback plug.
ASTM
2. Press to begin testing.
Remote
Tests Completed: 0
Bar Wand
RXD
DTR
RTS
TXD DSR DCD CTS RI Errors
Home
0
0
0
0
0
Exit
Pressing the “HOME” key returns to the READY screen, or EXIT to go back one screen.
4-9
pHOx Service Manual 4.2
Flow Problems NOTE:
Each of the following troubleshooting procedures has an associated flow diagram that may be found at the end of this chapter.
General procedures when isolating flow problems: NOTE:
Refer to the FLOW TEST decision tree. Do not be confused with the Flow Test Cycle that only shows the flow times.
Unit calibrates, but the sample will not aspirate. This usually means you have an air leak into the system. The sample, being more viscous than the aqueous standards, causes more vacuum to be created by the roller pump. If a membrane is leaking, this increase in vacuum will pull the aqueous fill solution from the electrode, easier than pulling sample from the probe. Likewise, a poor connection at the interconnect tubing or between the electrode sealing washer and the sensor module will allow air to enter easier than pulling sample up the flow path. • Check the connections between probe and sensor module, sensor module, and reference electrode. • Check the electrodes for the loss of internal fill solution. • Check the flow cell window during a pump cycle, looking for bubbles entering the flow path or pump and waste tubing. A No flow or obstructed flow cause can be isolated by performing the FLOW TEST. The fluidic system can be isolated into three large sections. This will limit the area of your search. The 3 sections are Fluid Pack to Fluid Fountain, Sampler/Sensor module, and W/R to Waste tubing. Use the SYSTEM TEST screens to control the mechanical devices. 1. 2. 3. 4. 5. NOTE:
4-10
Extend the sampler to the syringe position. Open the waste line solenoid. Turn on the pump and attempt to aspirate water then air from a sample cup. If water is aspirated from the cup, the sample probe, sensor module, external tubing, and waste line are functioning. The problem exists between the Fluid Fountain and Fluid Pack. If water is NOT aspirated from the cup, remove the tubing from the outlet port of the Reference electrode. Place it in a cup of water. This assumes the waste solenoid has functioned and is not pinching the tubing.
4 Troubleshooting 6. 7.
If step 5 fails to aspirate from the cup, the problem lies between the W/R tubing and the waste bottle of the Fluid Pack. If Step 5 does aspirate water, the problem is in the Sampler/ Sensor module.
4.2.1 Isolating a Plug in the W/R to Waste Tubing NOTE:
1.
3. 4. 5.
Disconnect the waste tubing from the reference electrode and place in a cup of water. If water is NOT aspirated, there is a plug. Lift the waste tubing out of the waste solenoid. If flow starts, the valve is dirty. Clean the dirt from the waste solenoid or replace it. If flow does NOT start, remove the waste line from the front panel connection. If water flows from this tube there is a plug in the internal waste tubing or waste bag of the fluid pack. Turn off the pump. Connect a syringe of water to the waste port on the front panel. Attempt to inject water through the line to clear the plug. If there is no flow, remove the Fluid Pack to ensure the problem is not a poorly seated or damaged waste connection. Should the problem continue, replace the internal tubing set.
4.2.2 Isolating a flow problem in the Sampler/Sensor module 1. 2. 3. 4.
NOTE:
You noted flow when aspirating from the waste tubing in step 2 of Section 4.2.3. Raise the Sampler to the Syringe position. The Pump should be OFF for this test. Connect a syringe of water to the reference electrode outlet port. Attempt to inject the water. If water flows freely, air is entering the system preventing aspiration. a. Inspect each electrode to assure the internal fill solution level is not dropping. b. If yes, replace the electrode or membrane as required. c. Inspect the seal between the reference electrode and sensor module (interconnect tubing) . If the flow path starts to empty while the syringe is still attached, only the water below the leak will leave the flow path out of the tip of the probe.
4-11
4. Trouble.
2.
If the reference flow is obstructed, the sample may not stop in the sensor module due to increased vacuum. Flow time decreases due to increased flow rate.
pHOx Service Manual d.
5.
Inspect the sensor sealing washers for distortion or misalignment. Replace as necessary. e. The seal between the air detector and the sample probe may be tested by attaching the syringe to the sample probe, blocking the air detector outlet tube. Apply pressure to the syringe and look for water leaking out. If water does not flow freely, there is a blockage. a. Loosen the reference electrode. Water should pass through or the electrode is plugged. Flushing with cleaning agent or hot water may free the blockage. b. If the reference is clear, reinstall it and remove the bottom electrode. c. Apply a moderate pressure with the syringe. d. If water does not enter the sensor module of the bottom electrode the blockage is above it. e. Removing one electrode at a time. Look for water entering the sensor module. If no water enters the module as the electrode is removed, the plug is above it. f. If water did enter the bottom sensor module in step C above, the plug is in the Sample Probe/Air detector. Remove these assemblies and flush through them or replace them.
4.2.3 Isolating a flow problem between the Fluid Pack and the Fluid Fountain If you have successfully aspirated a sample, but the analyzer will not calibrate or aspirate on-board controls, proceed as follows. 1. Position the sampler in the syringe position and inspect the capillary adapter for signs of wear. Replace if necessary. 2. Lower the sampler to the Home position; connect a syringe to the reference outlet port. 3. Remove the analyzer cover. 4. Disconnect the S-Line (center tube) from the rotary valve. 5. Inject the water and look for flow from the S-Line. 6. Look for a leak between the fluid fountain and the capillary adapter. Realign if necessary. 7. If there is no flow from the S-Line, remove the fluid fountain/S-Line and flush with hot water or cleaning agent, or replace the assembly. 8. If there is free flow through the S-Line, the problem is in the rotary valve, tubing harness, or connection to the fluid pack. 9. Reconnect the S-Line and remove the fluid pack. 10. Move the rotary valve to all the possible positions, observing which ports do not allow flow. 11. If there is no flow in ALL positions, remove the valve and clean or replace it.
4-12
4 Troubleshooting 12. Remove the tubing for each position that does not flow freely. 13. If water can be flushed through the individual lines, remove and clean or replace the valve. 14. If water cannot be flushed through the affected lines, replace the tubing harness.
4.3
Status Codes Error codes come in 2-digit groups. The tens (first) number, called the “series code,” identifies the affected component. Series 1 through 6 are sensors, 0 = pH 1 = PCO2 2 = PO2 3 = SO2 % 4 = Hb 5 = Hct 6 = Na+
Series Digit code -1 -2 -3 -4 -5 -6 -7
1-6
7
8
Slope No Std.A No QC1 Instability No Std.B No QC2 Overload No Std.C No QC3 Drift No Std.D Dependency Sample flow Air flow
pH Status Codes Code Status Screen Text 01 pH Slope 02 pH Instability 03 pH Overload 04 pH Drift 96 Hardware 97 Software
9
10
Barometer Not assigned Printer Flow Cell Temp Communication Digital Board Analog Board (ADX) Software
Error Log Text pH Slope pH Instability pH Overload pH Drift pH ADC pH Math
4-13
4. Trouble.
Series 7 = flow, 8 = Q.C. 9 = Electronic/Mechanical/Hardware 10 = CO-Oximeter codes. The ones (second) number, called the “digit code,” provides more information on the failure.
pHOx Service Manual PCO2 Status Codes Code 11 12 13 14 15 96 97
Status Screen Text PCO2 Slope PCO2 Instability PCO2 Overload PCO2 Drift PCO2 Dependency Hardware Software
PO2 Status Codes Code 21 22 23 24 96 97
Status Screen Text PO2 Slope PO2 Instability PO2 Overload PO2 Drift Hardware Software
SO2 LED 1 Status Codes Code 31 32 33 34 35 96 97
Status Screen Text SO2 % Slope SO2 % Instability SO2 % Overload SO2 % Drift SO2 % Dependency Hardware Software
SO2 LED 2 Status Codes Code 31 32 33 34 96 97
Status Screen Text SO2 % Slope SO2 % Instability SO2 % Overload SO2 % Drift Hardware Software
Hb Status Codes Code Status Screen Text 41 Hb Slope 45 Hb Dependency 97 Software
4-14
Error Log Text PCO2 Slope PCO2 Instability PCO2 Overload PCO2 Drift PCO2 Dependency PCO2 ADC PCO2 Math Error Log Text PO2 Slope PO2 Instability PO2 Overload PO2 Drift PO2 ADC PO2 Math Error Log Text
SO2 % LED1 SO2 % LED1 SO2 % LED1 SO2 % LED1 SO2 % Dependency SO2 % LED1 ADC SO2 % LED1 Math Error Log Text
SO2 % LED2 SO2 % LED2 SO2 % LED2 SO2 % LED2 SO2 % LED2 ADC SO2 % LED2 Math Error Log Text Hb Slope Hb Dependency Hb Math
4 Troubleshooting Error Log Text Hct Hct Hct Hct Hct Hct ADC Hct Math
Na+ Status Codes Code Status Screen Text 61 Na Slope 62 Na Instability 63 Na Overload 64 Na Drift 96 Hardware 97 Software
Error Log Text Na Slope Na Instability Na Overload Na Drift Na ADC Na Math
AD1 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software
Error Log Text AD1 AD1 AD1 AD1 ADC AD1 Math
AD2 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software
Error Log Text AD2 AD2 AD2 AD2 ADC AD2 Math
AD3 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software
Error Log Text AD3 AD3 AD3 AD3 ADC AD3 Math
4. Trouble.
Hct Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 54 Hct/Air Det Drift 55 Hct/Air Det Dependency 96 Hardware 97 Software
4-15
pHOx Service Manual
4-16
AD4 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software
Error Log Text AD4 AD4 AD4 AD4 ADC AD4 Math
Flow Status Codes Code Status Screen Text 71 StdA Flow 72 StdB Flow 73 StdC Flow 74 StdD Flow 75 Flowtime 76 Sample Flow 77 Air Flow 78 Back Flow
Error Log Text StdA Flow StdB Flow StdC Flow StdD Flow Flowtime Sample Flow Air Flow Back Flow
Control Status Codes Code Status Screen Text 81 Ctrl 1 Flow 82 Ctrl 2 Flow 83 Ctrl 3 Flow
Error Log Text Ctrl 1 Flow Ctrl 2 Flow Ctrl 3 Flow
System Status Codes Code Status Screen Text 91 Barometer 92 Printer 93 Temperature 94 Communication 95 Hardware 95 Hardware 95 Hardware 96 Hardware 97 Software 99 Scheduled QC Not Run
Error Log Text Barometer Printer Temperature Communication PEROM Write PEROM Read PEROM CRC Hardware Software Scheduled QC Not Run
RMS Status Codes Code Status Screen Text 688 N/A 690 N/A 691 N/A 693 N/A 694 N/A
Error Log Text Software (Pack Not Present) Software (Pack Empty) Software (Pack Lot Date Expired) Software (Pack Use Life Expired) Communication Error
4 Troubleshooting CMS Status Codes Code Status Screen Text 696 N/A 698 N/A 699 N/A 701 N/A 702 N/A
Error Log Text Software (Pack Not Present) Software (Pack Empty) Software (Pack Lot Date Expired) Software (Pack Use Life Expired) Communication Error
Standards Information: Sensor
Hct& ADET’s
A&D
4. Trouble.
Standards Used Delta mV Calibration To calibrate (between standards) _______________________________________________________ pH C&D 33 + Na (pHOx) A&D 17 (pHOx+/C/L) A&C 16 PO2 Electronic 0.00 / Room Air, Std A used for 1 pt Ca++ PCO2 A&B 17 Insert sheet shows HCO3- value. HCO3 x 1.5=CO2 in mmHg (all ADET’s) > 500 Hct ideal > 800
(Hb) Based on SO2 & Hct (mV not shown) SO2 External Stds 1 & 2 n/a + K A&C 24 iCa A&D 12 Cl A&C 15 Gluc A&C 38-55 Lact A&C 35-50 _______________________________________________________
4-17
pHOx Service Manual 4.3.1 Analyte Troubleshooting ( Series code 1-6 and Digit codes 1-4) Multiple Channel failures: Slope • Ensure the Reagent Pack is fully plugged in. • The pack box edge closest to you should lie inside the small lip of the chassis. • Check the flow times displayed on the service screens. • Inspect the Reference tubing to assure flow. • Inspect the waste line for bubbles. • Prime the system, check each fluid for air. • pH and Na+ require a working reference electrode. Figure 4-1 Reagent Pack Placement • • • • •
Note if the channels that fail use the same standard from the fluid pack. Loss of the +5 Vdc on the analog board Loss of the Analog board A/D/ converter Loss of +5 Vdc to the Analog Board Analog Board over heated (PTC resister open) - If unit is allowed to cool off by shutting it off for 5 minutes. Investigate why the unit or Analog board is overheating.
Dependency Rules • Hct requires calibrated Na+ for calibration and results. • PCO2 requires calibrated pH for results (pH checked on standards to assure bicarbonate is correct). • SO2 requires calibrated Hct and PO2 for results. Calculated SO2% is reported if measured SO2 is not available. • Hb requires Hct and SO2 for reporting. Priority: 1. If linked with COOX, the COOX HGB is reported. 2. If no COOX link, the measured result is reported, if available. 3. If no COOX or measured result, the calculated result from the Hct reading (Hct/3) is reported. 4. If no COOX, no measured, and no Hct, the default value (Hb) is reported.
4-18
4 Troubleshooting • • • • • • • •
Qsp/Qt requires 2 separate blood samples for determination: mixed venous and arterial. RI utilizes input or default FIO2% for calculation. Calculated Results require calibrated and proper results by measured tests used in the calculation. Suppression rules - If measured test is suppressed in the Setup, the calculated results based on test also will not be shown or printed. Password (One) is used for Setup, Service Menu, and QC. (No user passwords are available.) Interfaces are NOVA COOX, PDM, and ASTM. Standby Mode: There is no time dependency. Analyzer will prime and calibrate after coming out of standby. Idle pumping is active, but no fluidic or gas calibrations are performed. Features Auto-Wake-Up, Auto-Cal, and Auto-QC Micro Sample Analysis: Micro-Mode Tests: pH, PCO2, and PO2 Stepped Analysis: Stage 1 = Tip of probe to ADT1 (40 µL) Stage 2 = Up to ADT3 Stage 3 = Between ADT3 (up to ADT4): Other Tests Reading (including contact with reference electrode)
Instability • Inspect sensor module for bubbles during the measurement cycle • Ensure that the waste valve stops the fluid flow during analysis. Fill the flow path with any standard, raise the probe to the syringe position Look for droplets forming at the sampler tip. • Ensure that the waste and Reference solenoid valve are functioning. • Fill the flow path with any standard and view the system test screen. The signals should be steady and there is no flow of fluid there are three Possible conditions, a loose sensor module, an analog board failure, or a poor connection between a set of electrodes and an interconnect board. • Inspect the electrode washers to assure proper alignment. Nonalignment may cause a flow problem. • In rare cases the tubing harness may have become contaminated. Using the cleaning adapter, flush the system with cleaning agent followed by distilled water.
4-19
4. Trouble.
•
P50 is measured and is reported only when PO2 is between 30-75 mmHg. P50c is reported when PO2 is outside those limits. P50 also requires calibrated SO2% and Hb/Hct for reporting.
pHOx Service Manual 4.3.2 Single Channel Failures Slope View the Sensor Subsystem screens. If the change in millivolts between standards is 0 or less than that mentioned in Section 4.3.1, there are several possibilities: • Membrane or sensor has reached the end of its use life. • Bubble of Air is covering the sensing material at the electrode tip. • The electrode is not making good contact with the Interconnect board contacts. • Loss of the Analog board multiplexer will cause a zero input to the A/D converter. • Ensure calibration on required channels: Sodium is required for Hct, Hct is required for SO2%. • Viewing the Sensor subsystem screens you note an overload condition (999.9 mv). • Check for NO connection to the Interconnect board. • Check the PO2 /PCO2 for loss of the Internal Fill solution (membrane rupture). • Loss of the multiplexer or individual channel circuit on the Analog board. • The pH electrode may need conditioning. • Na+ sensor over 30 days of use. The Na+ sensor will be destroyed by conditioning. Instability Single channel instability is seldom a flow problem unless a standard is unique to a particular channel. • De-bubble the electrodes. • Ensure a good connection between the electrode and the Interconnect Board. • Ensure a good connection between the sensor module and Chassis. • Ensure the analyzer's door is closed and the ground strap is making good connection.
4-20
4 Troubleshooting 4.3.3 Temperature Problems • • • • •
Ensure that the room ambient temperature is within specification ( 18o - 32o C). Ensure that the sensor module is plugged in correctly. View the Analog Input screen channel 20. This channel will be a 0 if the module is at 37o C and change at a rate of 22.5 mv/1oC, + if above and - if below 37o C. The power for the preheater comes from the digital control board. The heater control circuitry is on the analog board.
4.3.4 Air Detector (ADT) Problems
• • •
NOTE: •
•
Air Detectors are only calibrated during a 2-point calibration. Typically the Fluid /Air millivolt difference will be about 700. Check the sample probe for air being introduced below the sensor module. Check the sensor module for air being introduced at the electrode/sensor module seal. Ensure that the electrode sealing washers are centered and seated correctly. The Air detector signals are processed by the analog board. They are only displayed on the System Test screen. Alternately aspirate air and then fluid to see the change in the signal. Remember to turn on the air oscillator to obtain an ADT reading. Flow Test During a flow test, the rotary valve rotates to each fluid position. The pump turns on to fill the lines from the rotary valve S-line port to the ADT1. The rotary valve rotates to an air position. The pump turns on and the time in milliseconds until the ADT senses air is recorded. The time is normally 1200 msec. The acceptable range is 100 to 1700 milliseconds. Run a Flow Test; press the soft key on the Service screen. Typical result for Standard A, B, C, D is 1300. If all are high (3000-5000), there is a flow path obstruction. If only one is high, 1. Check for correct fluid pack installation. 2. Check that the valve is in the correct position. 3. Connect a syringe to the S-line port of the valve. Remove the reagent pack and move the valve to each position. Verify that the link is clear by purging air through the S-port.
4-21
4. Trouble.
• • •
pHOx Service Manual
4-22
4 Troubleshooting
4. Trouble. 4-23
pHOx Service Manual
Inspect for air leak. Loss of IFS, Replace or use plasClear the plug by flushbad washer, ruptured membrane, Replace or or interconnect tubing seal. backflush the tic cleaning line to ing plastic line or reclear the plug. place the module. 4-24 air detector.
4 Troubleshooting
4. Trouble. 4-25
pHOx Service Manual
������������������
�������������������������������������������������
�
��� �������������������������������������������� �
� ��������������������������������������������������������������� � ������������������������������������������� ������������������������������������������������ � ��� ���������������������������������������������� � ���� ��������������������
� ��� ��������������������������������������������������������
���� ������������������������������������������������������ �������������������������������������������� � ��� ������������������������������������������������ ���� � �������������������������������������������������������� �� � ������������������� � ���� ���������������������
��� ������������������������������������������������������������� �
� �������������������������������������������������������������� � ����������������������������������������������������������� ����������������������������������������������������������������� � ��� ��������������������� � ���� ��������������������������������������������������
� ��� ����������������������������������������������
��� ������������������������������������������������ ��� ����������������������������� �� ����������������������������������������������������� ��� ����������������������������������������������� ��� ��������������������������������������������������� ������� ����������������������������������������� ��� �������������������������������������������������� � ������������������������������������������������������������ � ����������������������� ������������������������������������������������������������� �� ����������������������������������������������������������������� � ������������������������������������������������������������� � ���������������������������������������������������������
�
� � 4-26
4 Troubleshooting ���������������������������������� ��� � ������������������������������������������������ � �����������������������������������������������
�
��� � ������������������������������������������� � ���������������������������������������������������������������� � ���������������������������������
� ������������������������������������������������������� ��� ��� ������������������������ � ������������������������������������������������ ���� ��������������������������������������������������������� ���������������������������������
� ������������������������������������������������������� ���� ��������������������������������������������������������� �������������������������������������������������������� �����������������������������
� ��
4. Trouble.
��
�� ���� ���������������������������������������
� ������������������������������������������������������� � �������������������������������� ������������������������������������������������������������ ����������������������������������������������������������
�� ��� �������������������������������������������������������� �� ��������
�� ���� �������������������������������������������������������� � � �
������������������������������������������ ��� ����������������������� ����������������������������
���� ���������������������������������������������� ��
� ��������������������������������� ������������������������������������������������ �������������������������������������������������������� �������������������������������������������������������� �����������������������������������������������
��
�� ��������������������������������������������������������������
��
�������������������������������������������������������� ������������������������������������������������������������ �������������������������������������������������������� �������������������������������������������������� ��������������������������������������������������������
�� ����������������������������������������������������������
���������������������������������������������������� ������������������������������������������������������������� ����������������������������������������������������������� �������������������������������������������������������� ��������� �������������������������������������������� ������������������������������������������������
�� 4-27
pHOx Service Manual
4-28
Appendix A
Appendix
A.1
Quick Reference Guide Measured Tests: pH, PO2, PCO2, SO2%, Hb/Hct, (Na+ is not reported)
Alveolar Oxygen Concentration Arterial Alveolar Oxygen Tension Ratio Arterial Alveolar Oxygen Base Excess, Extracellular Fluid Base Excess, Blood Actual Bicarbonate Concentration Oxygen Content P50 Calculated (when PO2 is 75) Physiologic Shunt Respiratory Index Standard Bicarbonate Oxygen Saturation % (Calculated) Total Carbon Dioxide pH/PCO2/PO2 (Temperature Patient Corrected, if selected)
Slope Limits: pH 9.1 to 11.6 (1,7) PO2 -15.0 to -1.6 (1,2,4,8) Na+ 8.8 to 11.5 (1,7,8,9) Electrode Notes: 1 = Long Life Sensor 2 = Int’l Ref Electrode 3 = Membrane Cap; Push-On 4 = Membrane Cap; Screw-On 5 = Reflectance
PCO2 7.9 to 12.6 SO2% 6.9 to 18.2 Hct:
12.0 to 50.
Appendix
Calculated Tests: • A • a/A • A-aDO2 • BEecf • BEblood • HCO3• O2Ctc • P50c • Qsp/Qt • RI • SBC • SO2%c • TCO2 • Temp
(1,2,4,8) (1,5,8) (1,6,7)
6 = Impedance 7 = No Washer 8 = Internal Washer 9 = Non-Reported Test 10= Obtained from SO2% and Hct
A-1
pHOx Service Manual Flow Limits: Flow Test (Service Menu): Typical/Ideal Value = 1300 (milliseconds) ** High Values (ie: > 3000) or “ERROR” note on Std Channel: Indicates major flow blockage. Offset Limits (Slope/Intercept): Test Y-Intercept Range pH +/- 100 mpH H+ NONE PCO2 +/- 10 mm Hg PO2 +/- 20 mm Hg SO2 % NONE Hct +/- 10.0 % Hb +/- 2.0 g/dl
Slope Range NONE 0.8 to 1.2 0.8 to 1.2 0.9 to 1.1 0.9 to 1.1 0.8 to 1.2 0.9 to 1.1
Flow Cell Positions: Top Adet 1 Adet 2 Adet 3 Adet 4
Reference Biosensor Located above Sample Probe Na+ Used to detect 40 uL MicroSample In Sensor Mod/Flow Cell pCO2 Used to detect 70 uL sample Reads Hct, then blood pulled up and air is seen telling stop for electrode sample sO2 reading; sample position Positions top of sample for electrode readings
Reference ADET 4
pH
pO2 ADET 3 ADET 2
Calibration Details (Analysis Time/Throughput): Two Point Calibration: User Selects (not auto control), 2,4,6 hour intervals (turned off in Standby Mode). One Point Cal: User selects for 30 or 45 minutes (same as STP/UL), One point cal for ALL (gases plus pH/Hct), One point SO2 done with each full 2-point calibration. Standard One Point Cal: Done with blood samples: Single Mode (each sample); NOT available with product release; will be for low volume users only.
A-2
Appendix Throughput specs subject to change without notice due to software sequence changes.
Throughput Details:
Mode A = 44 sec.; Mode B = 64 sec (time to view results) Mode A = 62 sec; Mode B = 74 sec (full cycle) Approx 50/Hour (varies by mode) Analysis (view) = 43 sec; Run Time = 65 sec.
pHOx+C: Analysis Cycle: Run Time: Throughput: Micro Mode:
Mode A = 47 sec.; Mode B = 100 sec (time to view results) Mode A = 65 sec; Mode B= 109 sec (full cycle) Approx 33/Hour (varies by mode) Analysis (view) = 45 sec; Run Time = 65 sec.
pHOx+-L: Analysis Cycle: Run Time: Throughput: Micro Mode:
Mode A = 60 sec.; Mode B = 126 sec (time to view results) Mode A = 89 sec; Mode B= 134 sec (full cycle) Goal: 25-30/Hour Analysis (view) = 71 sec; Run Time = 95 sec.
Default Values: Hb Temperature FIO2 %
14.5 g/dL 37 ° C 20.9 % (room air)
Sample Volume: Regular Mode
Micro Mode
70 uL 115 uL 120 uL 40 uL 55 uL
(Range 10.0 to 20.0 g/dL) (Range 10.0 to 40.0°C) (Range 20 to 100.0%)
pHOx/B pHOx-Plus pHOx-Plus L pHOx/B pHOx-Plus- Plus L
(all tests) (all tests) (all tests) (gases) (gases)
Software Version: 4.xx Sample Type: Arterial, venous, capillary, mixed venous, Quality Controls, and Proficiency
A-3
Appendix
pHOx/B: Analysis Cycle: Run Time: Throughput: Micro Mode:
pHOx Service Manual One-Point Cal Modes: Mode A One point fluid cal performed with each Gas Cal I (30-45 min intervals). (Best for medium-high volume customers) Mode B One point fluid and Gas Cal performed with EACH analysis. (Best for LOW volume customers only - No separate Gas Cal I) Rotary Valve Positions:
Closed ("Air Std D")
AIR ("Air Cntrl 1")
Std D
Std C
Cntrl 1
AIR ("Air Std C")
Closed ("Air Cntrl 2")
Std B
Cntrl 2
Sample
AIR ("Air Cntrl 3")
Closed ("Air Std B")
Cntrl 3
Std A AIR ("Air Std A") Home: Closed
"Dummy"/Closed (not used)
Figure A-2 Valve Manifold Assembly (Tubing Connections and Rotation Positions)
Control/Position Valve Manually and View Positions: Service Menu, System Test
A-4
AIR Positions: Closed Positions:
Air Std A, Air Std C, Air Cntrl 1, Air Cntrl 3 HOME, Air Std B, Air Std D, Air Cntrl 2, Dummy
Fluid Positions: Connections on Valve:
Std A (A)
Control (Fluid) Positions: Connections on Valve:
Cntrl 1 (E)
Not Used Position:
Dummy
Std B (B) Cntrl 2 (F)
Std C (C)
Std D (D)
Cntrl 3 (G)
Appendix Linear (Measurement) Ranges: pH 6.50 to 8.00 pH units PCO2 3.0 to 200 mmHg (0.4 to 26.7 kPa) PO2 0 to 800 mmHg (0.0 to 106.7 kPa) SO2% 0.0 to 100% Hct 12% to 70% Hb 4 - 23.3 g/dL - based on Hct Na+ not reported - 1.0 to 200.0 mmol/L)
Memory Storage Capability: Patient Storage 18 Samples (FIFO) QC Storage 6 QC Files (FIFO) Stores up to 10 daily points/level - condensed to 1 monthly point Monthly Data - 32 points for each level Error Codes 96 total codes stored (FIFO) 6 shown per screen (16 potential screens) Appendix
QC Files Details: 6 QC Files plus “File 0” (proficiency) 3 On-Board QC files (combined gases/Hct/SO2 controls; 3 levels) 3 External QC files (3 levels for BG; 2 levels for Hct/SO2) Features Auto QC (on-board QC) w/ preset times Control reminder/lockout; Setting daily QC analysis times. Tracking Daily, Monthly, Cumulative, Levey-Jennings. Fluids Details: Std A - It is used to calibrate PCO2, Na+, Hct, and PO2 (1-point check) and sits in flow cell during idle. Std B - It is used to calibrate PCO2, contains surfactant, and is used to clean flow path during cycles. Std C - It is used to calibrate pH. Std D - It is used to calibrate pH, Na+, and Hct. External SO2 Calibrators - There are 2 levels used to calibrate SO2 (approx 50 and 95 %; final values TBD). Calibrate monthly (or as needed).
A-5
pHOx Service Manual Dependency Rule and Setup Information: HCT Requires Na+ calibrated for calibration/results. PCO2 Requires pH calibrated for results. (pH checked to assure bicarbonate is correct.) SO2% Requires Hct calibrated, uses SO2%c if measured SO2 is not available. Hb Requires Hct and SO2 for reporting. Hb method priority: (1) If linked with COOx, then COOx Hb will be reported. (2) If no COOx link, measured results are reported (if available). (3) If no COOx or measured result, calculated result from Hct reading (Hct/3) is reported. (4) If no Coox, no measured, and no Hct: Reports ‘Default’ Value (Hb). P50 Measured Reported only with PO2 between 30 and 75 mmHg. Also requires calibrated SO2% and Hb/Hct for reporting. Qsp/Qt Requires 2 separate blood samples for determination: Mixed Venous and Arterial. RI Utilizes input or default FIO2 % for calculation. nCa pHOx-Plus Only. Requires properly installed/calibrated iCa and pH. Anion Gap pHOx-Plus Only. Requires properly calibrated/installed Na+,K+,Cl-, PCO2. Calculated Results require calibrated and ‘proper results’ by measured tests used in calculation. Suppression (Hb): Only Hb can be suppressed: Option to turn Hb from measured to calculated (one or the other will always appear). Passwords: • One password (System Password) for Set Up • 200 user unique passwords available - 3 levels of priviledge: 1. Privilege Level 1: Operators have access to all areas of the analyzer except those protected by the existing System Password. Level 1 operators do not require PDM review and may override this feature. Level 1 operators can override QC lockout. 2. Privilege Level 2: Operators have access to all areas of the analyzer except those protected by the existing System Password. Level 2 operators DO require PDM review, but may override this feature. Level 2 operators cannot override QC lockout. 3. Privilege Level 3: Operators have access to analysis only. Level 3 operators require PDM review and cannot override this. Level 3 operators cannot override. Test Suppression: If measured test is suppressed, any calculated/associated tests will also be suppressed. Tests still will be calibrated w/errors displayed on suppressed channels.
A-6
Appendix Interfaces: NOVA COOX, PDM, ASTM interfaces. Standby Mode: No time dependency will prime and calibrate after coming out of Standby, In Standby only does idle pumping, no fluidic/gas calibration, Auto-Wake-Up, Auto-Cal, and Auto-QC Micro-Sample Analysis: Micro-Mode Tests: pH, PCO2, PO2 Stepped Analysis: Stage 1 = Tip of Probe to Adet 1 (45/55 uL Stage 2 = Up to Adet 3. Stage 3 = Between Adet 3 (up to Adet 4): Other Tests Reading (including contact with Reference Electrode)
A-7
Appendix
QC Lockout: Mode A Lockout: If lockout enabled, and tests fail, unit can be ‘unlocked’ by running any control level and passing all tests. Mode B Lockout: If lockout enabled, and tests fail, unit can only be ‘unlocked’ by running specific control level(s) that failed and passing all tests. QC Storage Options: Mode A QC Storage: If all QC results ‘pass’ range limits, QC results are automatically stored in memory. If any results ‘fail’ range limits, all results will remain on unit display until either user accepts/rejects results or another QC or sample result is run/displayed (at which time past [non-released] QC results) will be deleted. Mode B QC Storage: All QC results, whether passing or failing range limits are automatically stored in memory at end of control analysis. Glucose (Overrange Testing) pHOx-Plus/C/L only: Gluc Overrange Solution included in Reagent Pack: Done Automatically Daily; Elimination of manual - ampule running of Performance Check Solution Automatic Glucose Performance done at operator-programmed time daily. (Nova recommends to set time at 5:00 AM for optimum performance.) PASS - FAIL is reporting. Change Gluc Membrane if test lists as FAILED and repeat. Glucose Auto Polishing Cycle: Automatic polishing of Gluc sensor occurs each day at midnight, followed by 2-point cal, and additional 2-pt calibrations at 2 and 4 AM.
pHOx Service Manual Error Code Overview: Code Groups Summary Group: 1 - 6 Group: 7 - 13 Group: 14 - 19 Group: 20 - 32 Group: 33 - 35 Group: 36 - 42 Group: 43 - 48 Group: 49 - 53 Group: 54 - 58 Group: 59 - 63 Group: 64 - 68 Group: 69 - 75 Group: 76 - 78 Group: 79 - 104 Group: 602 - 684 Group: 686 - 703
pH PCO2 PO2 SO2% Hb Hct Na+ ADET 1 ADET 2 ADET 3 ADET 4 Flow-Related QC-Related Hardware-Software Related QC Related RMS Reagent Pack Related
(Refer to the reference manual for specific code details.)
A.2
General Maintenance/QC 1.
Flowpath Conditioning (AS NEEDED) (Use serum/plasma/whole blood - Automated Cycle.) (Sales/Customer Typical Information: Customers as below; Sales- Do once at start of demo/evaluation.) a. Done daily for low-mid volume users (as needed). b. Not required for higher volume users.
2.
Control(s) Analysis (DAILY) (Sales Information: SetUp QC Files fully before demo/evaluation; Customers: Run QC as they wish.) NOTE: Internal & External Controls are unique for all pHOx models. Controls for pHOx cannot be used on pHOx-Plus or “L”, and visa-versa. a. b.
A-8
Internal Controls - 3 levels (All Tests including gases and SO2) External Controls - 3 levels in Multi-Pack Only (pHOx/B =All Tests inc. SO2 & Gases; pHOx+/L = All Tests, inc. electrolytes/Glu/Lac)
Appendix Glucose Overrange Solution (Analysis) pHOx+/L/C Only (DAILY) (Sales Information: Verify Performance Chk passes- AutoRun at start of demo/evaluation & at each followup) Software Versions 2.0.x or Higher (Elimination of manual-ampule running of Performance Check Solution.) Automatic Glucose Performance done at operator-programmed time daily. PASS - FAIL reporting: Change Gluc Membrane if test lists as FAILED then repeat test. For Failed Test: Gluc is BACKLIT on Ready screen indicating failure; tests are unavailable for use until resolved.
4.
Flowpath Cleaning (MONTHLY) a. For product release: Do monthly. b. Automated aspiration cycle, use Deproteinizing Solution. c. For pHOx only, remove Na+ electrode (non-glass sensor) and use a blank electrode.
5.
Calibrate SO2 Channel pHOx/+/L/C (MONTHLY) a. Calibrate using 2 external Calibrators (latex suspension). b. Do monthly or more frequently as needed.
6.
Clean SO2 Sensor (Optics) Surface: (pHOx/+/L/C) (QUARTERLY) a. Wipe optics surface of SO2 Sensor with lint-free tissue soaked with Cleaning Solution or 10% Bleach. Rinse with H2O.
7.
Replace Na+ Electrode pHOx (AS NEEDED) a. Required to be done based on slope or results performance. b. Correct Hct QC is NOT reliable indicator of Na+ channel performance. Condition Na+ Electrode pHOx+/L/C (AS NEEDED) a. Required to be done based on slope or results performance. b. Use Na+ Conditioning Solution (PN 36856). Remove Na+ and soak in Conditioning Solution for 2 minutes.
8.
Electrode Maintenance Details pH Electrode • Condition (manually) troubleshooting w/ pH Cond Soln (PN 06856); 10 min soak. • NO need to condition at start of use (Cond Soln in electrode pouch: pre-soaked) • Long-Life Sensor (similar to Stat Profile M) • Expected Lifetime: 1-2 Years; Warranty = 6 months from installation
A-9
Appendix
3.
pHOx Service Manual PO2 Electrode • •
• • • • •
Screw-On Membrane Cap (w/ Washer attached) Polish electrode (manually) (10 sec. circular on paper) w/ each membrane change. Electrode = Long Life; Membrane = Short Life Electrode estimated life = 1-3 Years; Membrane estimated life = 1-5 months Warranty: Electrode: 6 months; Membrane: Free of defects at installation Blood Condition after new electrode is installed (membrane changed as needed). Prefilled Membrane Caps (no IFS addition required or to be added)
PCO2 Electrode • • • • • • • •
Screw-on Membrane Cap (NEW Version; Different than STP/UL); washer attached. Membrane Life (new membranes/sensors = 2-12 weeks NO Conditioning (PCO2 Condition Solution) at StartUp (Caps presoaked in package) Electrode = Long Life; Membrane = Short Life Electrode estimated life = 1-3 Years; Membrane estimated life = 1-5 months Warranty: Electrode: 6 months; Membrane: Free of defects at installation Blood Condition after new electrode install, membrane change and as needed. Pre-filled Membrane Caps (no IFS addition required or to be added)
Na+ Electrode pHOx: • Short- Life Disposable Electrode. NO maintenance. • Expected Lifetime: 2-4 months. Warranty = 1 month (from install). • Do NOT condition with pH Cond Solution (will damage electrode). • Must remove Na+ when doing Flowpath Cleaning (use ‘blank’). pHOx-Plus/L/C: • Long-Life Sensor (similar to SP-M) • Expected Lifetime: 1-3 Years. Warranty = 6 months (from install). • Condition (as needed): Requires manual soak in Na+ Conditioning Solution. • NO need to remove Na+ (from flowpath) when doing Flowpath Cleaning.
A-10
Appendix SO2 Channel • •
•
pHOx/+/L/C Only Clean flow-channel with cotton-tipped swab. Wipe optics surface of SO2 Sensor with lint-free tissue soaked w/ Cleaning Solution or 10% Bleach. Rinse with H2O. (Quarterly) Expected Lifetime: 6-12 Months Warranty = 3 months from installation
K+ Electrode pHOx-Plus/L/C Only • Mid-life (age) Sensor (similar to SP-M); Max Use Life = 5 months • Blood/protein conditioning for low-volume customers (as needed) iCa Electrode pHOx-Plus/L/C Only • Short-life (age) Sensor (similar to STP-M) • Expected Lifetime: 1-3 Months; Warranty = 7 days or 300 samples from installation • Blood conditioning (optional) can improve sensor performance.
Glucose Electrode pHOx+/L/C Only • Electrode Short-life (age) Sensor (similar to STP-M) • Electrode = Long Life; Membrane = Short Life • Electrode estimated life = 1-3 Years; Membrane estimated life = 1-4 weeks (replace membrane if Glucose Performance Check fails or slope drops out) • Warranty: Electrode: 6 months; Membrane: 3 days from installation • Polish using new YELLOW (1 u) Polishing Paper only. • NO CONDITIONING (soaking) of Sensor required routinely. • Polish sensor ONLY FOR TROUBLESHOOTING (not with each membrane change). (Polish AS NEEDED.) • Polish 10 seconds. (Polish DRY on paper: no water added.) • Rinse with water; blot dry. • Replace new membrane directly. • Condition sensor in Soaking Solution (PN 13409) for 5 minutes with polishing.
A-11
Appendix
Cl- Electrode pHOx-Plus/L/C Only • Short-life (age) Sensor (similar to STP-M) • Expected Lifetime: 1-3 Months; Warranty = 7 days or 300 samples from installation • Blood conditioning (optional) can improve sensor performance • No membrane cap; pre-filled/sealed sensor (similar to STP-M)
pHOx Service Manual Lactate pHOx-Plus-L Only • Electrode = Long Life; Membrane = Short Life • Electrode estimated life = 1-3 Years; Membrane estimated life = 1-3 weeks (Replace membrane if Lact Std D Hi or Lo errors occur or slope drops out.) • Warranty: Electrode: 6 months; Membrane: 3 days from installation • Polish using new YELLOW (1 u) Polishing Paper only. • NO CONDITIONING (soaking) of Sensor required routinely. • Polish sensor ONLY FOR TROUBLESHOOTING (not with each membrane change). (Polish AS NEEDED.) • Polish 10 seconds. (Polish DRY on paper: no water added.) • Rinse with water; blot dry. • Condition sensor in Soaking Solution (PN 13409) for 5 minutes with polishing.
A.2.1 Other Analyzer Maintenance 1. a. b. c. d.
2.
Analyzer Shutdown a. Short-Term ( 48 Hours) • Requires water and air purge of flowpath and all reagent lines • Utilize special Purge Manifold to attach to reagent needle ports and place • Put all lines (except waste) in distilled water. Process 2-3 fluid primes with water. • Then remove lines from water and leave open to air; process 2 fluid primes with air. • After air prime, unit may be shut-off and remove electrodes as needed. • Take pump tubing off pump roller to relieve tubing tension.
Appendix A.3
pHOx Electrode/Standards Electrode
Standards Used in Approximate Delta mV Calibration/Slope (between standards) _________________________________________________________________ Stds C & D
33
Na+
Stds A & D
17
PO2
Stds Room Air, Std A and 0 (Ratio Air to Fluid (A) for 2 pt Cal; Std A used for 1 pt Cal)
Not Shown (Air to 0) 1 (typical ratio Air/A) (should be almost same)
PCO2
Stds A and B (HCO3- for dissolved gas>>monitor pH)
17
Hct and ADET’s
Stds A and D (all ADET’s)
> 500 (Adet 3 = Hct) (ideal > 800)
Hb
Based on SO2 and Hct
TBD
Appendix
pH
SO2%
External Stds 1 and 2 n/a _________________________________________________________________
A.4
Electronics and Mechanicals Overview Electronics Circuit Boards and Main Assemblies NOTE:
• • •
Almost all circuit boards on the pHOx utilize surfacemount technology, therefore board component repair is not recommended except for socketed chips. Analog Board/Sensor Board Digital/Controller Board Display Assembly
A-13
pHOx Service Manual • • • • • • •
A.5
Distribution Board Electrode (Left and Right) Interconnect Boards Keypad Power Entry Module Power Supply Assembly Printer Printhead and Printer Interface Board SO2 Sensor Board
Shutdown Procedure If the analyzer is to be turned off for more than 24 hours, flush the tubing harness first with distilled water then with air. Use the Flush Fixture (PN 24327) to perform this procedure. 1. Remove the Reagent Pack from the analyzer. 2. Install the Flush Fixture into the analyzer in the same way as the Reagent Pack. 3. Place the W-line of the fixture into an empty container. 4. Place the other tubing ends into a beaker of distilled water. 5. From the Ready/Not Ready screen, press Menu (soft key). 6. Select Flowpath Maintenance and press Enter. 7. Press Purge (soft key). 8. When the cycle ends, take all the tubings out of the distilled water. Leave the W-line in its container. 9. Repeat the purge with air. 10. When this cycle is completed, remove the flush fixture. 11. The pHOx Analyzer is now ready to be powered off for extended time.
A.6
Start Up Procedure 1. 2. 3. 4. 5. 6.
A-14
Insert the Reagent Pack and the Quality Control Pack. Reinstall the pump tubing. Power up the analyzer. Perform normal fluid priming and biosensor maintenance. Calibrate the analyzer. Run QC per your normal procedures.
Stat Profile pHOx Installation Check List
DONE
DONE
Model _____________________ Serial # ____________________
Instrument Check Out
Set Up Check Out
W/R Pump Tubing Harness Reagent Cartridge Control Cartridge (when available) Tubing Harness Electrode / Membranes SO2 Sensor Module Software Card and Version Probe/Microsample Adapter W and R Lines
Inspect Unit for Shipping Damage Remove Cover, Check all Cable Connections Fuses Correct Line Voltage N/A
YES
SMR(s) ____________________
Barometric Pressure
N/A
YES
Sensor Module Temperature
Troubleshooting Training Subjects Solenoid Valve Tubing Incorrectly Seated Water Test Flow Blockage Other __________________________ Other __________________________ Other __________________________
Operator Training Subjects System Identificaion Running an Analysis Operational Status Calibrating the Analyzer Replacing Reagents and Paper Troubleshooting Flow Test and Flushing
DONE
Using and Handling Controls
Fill out and review PACE forms Review customer warranty or service financing coverage
Maintenance Set Up
The above checks have been performed on your NOVA analyzer as part of your INSTALLATION. We at NOVA BIOMEDICAL believe that you, our valued customer, are entitled to the highest quality support available. As part of this service, we are leaving a copy of this check list for your records. If any questions arise, feel free to contact us TOLL FREE at 1-800-545-NOVA.
Nova Support Representative
Date
Customer Signature
Date A-15
A-16
Nova Biomedical pHOx Preventive Maintenance Check List SMR _____________________
CLEAN
ADJUST
REPLACE
Mechanical Checkout
INSPECT
CLEAN
ADJUST
REPLACE
INSPECT
Model _____________________ Serial # _____________________
Electronic Checkout LCD Screen Electronics Bay Electrical Grounds RS232 Connections Software Version ______
Pump Assy Sampler Assy W & R Valve Rotary Valve
CLEAN
ADJUST
REPLACE
INSPECT
Printer Assy
Fluidic Checkout Flowcell SO2 Sensor Sensors (replace membrane, if applicable) DONE
External Tubing Internal Tubing Fluids Bay Fluid Fountain Sample Probe Capillary Adapter
The above checks have been performed on your NOVA analyzer as part of your scheduled PREVENTIVE MAINTENANCE service call. We at NOVA BIOMEDICAL believe that you, our valued customer, are entitled to the highest quality support available. As part of this service, we are leaving a copy of this check list for your records. If any questions arise, feel free to contact us TOLL FREE at 1-800-545-NOVA. The PM sticker placed on the upper, front corner of the left side cover indicates today’s date, the month your next scheduled PM is due, the initials of the Nova Support Representative and the service report number.
Nova Support Representative
Date
Customer Signature
Date A-17
A-18
A-19
�
�
��
�
�
�
�
����� ���������������� ����� �������������� ����� ��������������� ����� �������������������������� ����� ����������������������������� � ����� ���������������� ��� � ���������������������������� ��� � ������������������������������ ��� � ��������������������������������� ��� � ���������������� ��� � �������������� ���� ��������������������������������������������� ������������������������������������������������� ������������������������������������������������������ ������������������������������������������������� ������������������������������������������������������ �������������������������������������������������� � � ������������������������������ ��������������������������������������������������� � � ������������������������������ ������������������������������������ ��������������������������������������� ���� �������������������������� ���� ����������������������
�
�
�� ���������
��
��
��
��
��
��
��
��
��
��
��
��
��
��
� �
��
��
��
��
��
���� ��������������������������������� ���� ������������������������������������������ ���� ����������������������������������������������� ���� ����������������������������� ���������������������������������������������� ������������������������������������ ������������������������������������ ���������������������������� �������������������������������������������������������� ������������������������������� ������������������������������������������������� ������������������������������� �������������������������������� ����������������������������� ���� ��������������������������������� ���� ������������������������ ���� ������������������������������������ ���� ������������������������������� ������������������������������������������������ ���� ��������������������������
������� ���������
��
��
��
�������������
�
�
�
�
A-20
PCMCIA
COM3
COM2
COM1
BARCODE
POWER SUPPLY J7
TB2
AC INPUT
TB1
J8 DIGITAL CONTROL BOARD J9 CN3
P1
P5
PUMP
J11
SAMPLER
J10 SEIKO PRINTER
J102
J104
J103
J6
J3
FLUID PACK
FP
P4
FRONT PANEL
LEFT ELECTRODE CLIP
TEST
J7
CN1 CCFL
KEYPAD
J2
ANALOG BOARD
CN2 OPTREX LCD PANEL
P2
J1
DISTRIBUTION BOARD
SENSOR MODULE
RIGHT ELECTRODE CLIP
J101
ROTARY VALVE
CP
Title:
Page S-1
Size
CONTROL PACK
WIRING DOCUMENT FOR PRODUCT: NOVA PHOX Drawing Number
Rev
D Do not scale drawing
Sheet
1
of
1
DWG #10-1087A
AC Mains Entry Module To Power Supply TB1 AC Line Brown
1
AC Neutral Earth Ground
3 5
Blue Green/Yellow
Analog Board J3 to Sensor Module CN1 ADT1-E Yellow
1
ADT1-M
2
THERMISTOR FPACK AGND
Blue Red
3
White
4
Black
5
---
6
HEATER+
Red
7
HEATER-
Brown
8
N.C.
N.C.
---
9
N.C.
---
10
Digital Control Board J9 to Seiko Printer CN3 PV1 ---
1
GND PV2 PV2
-------
2 3 4
PGND PGND
-----
5 6
P STB# P DR1 P DR2
-------
11 12 13
P DR3 P DR4 P DR5
-------
14 15 16
P DR6 P DR7 P DR8
-------
17 18 19
P BUSY P ACK# P ERR#
-------
20 21 22
P PE GND P RESET#
-------
23 24 25
P ONLINO P ONLINI P FEED#
-------
26 27 28
P BFEED# P NMI#
-----
29 30
PV2 PV2 GND
-------
31 32 33
PV1
---
Digital Control Board J11 to Distribution Board J101 GND 1 --2 D LP ---
GND DM GND D FLM GND D CP GND D DD0 GND D DD1 GND D DD2 GND D DD3 GND D SPARE GND D ENAVDD D VEEN D ENABKL D YELLOW# D GREEN# D VCC D VCC D VCC D VCC KBY5# KBY4# KBY3# KBY2# KBY1# KBY0# KBX3# KBX1# KBX2# KBX0# SPCLK DACLK DADAT SPKRSPKR+
Digital Control Board J8 to Main Wire Harness PGND P9 Black
ANA V+ +5VANA -5VANA ANAGND ANAGND ANAGND ANARXD# ANATXD# VCCPROT GND GND INT7 HTRCTL# FAIL# $1N3313 PGND
1
---
2
PGND
P6
Black
3
-----
3 4
PMD1 PMD3
P1 P1
Red Yellow
-----
5 6
PMD0 PMD2
P1 P1
Black Grey
---
10
-----
12 13
-------
-----
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29
-------
30 31 32
-----
33 34
-------
35 36 37
-----
38 39
Main Wire Harness J8 to Rot Val P9 N.C. ---
1
-------
41 42 43
N.C. ROVHOME# ROVLED
--White Red
2 3 4
---
44
GND GND
Black Black
5 6
---------------------------------------------------
34
Main Wire Harness J8 to Control Pack CP TOUCH Orange
1
Main Wire Harness J8 to Lamp P7 LAMPRed
1
LAMP+
2
N.C. N.C. GND
2 3 4
Brown
1
Digital Control Board J10 to Analog Board J1 PGND ---
----Black
5 6
-------
2 3 4
7 8
KBY1# KBY0#
-----
5 6
9 10 11
KBX3# KBX2# KBX1#
-------
7 8 9
12 13 14
KBX0# G AN G KA
-------
10 11 12
Y AN Y KA
-----
13 14
VALVECOM TOUCH
P4 FP
Black Orange
P5
Red
14 15 16
DOORSW#
P6
White
SMD2 PGND
P5 P10
Grey Black
-----
18 19
SMD1
P5
Yellow
15
---
20
GND SMD10
P10 P5
Black Black
16 17
-----
22 24
PGND RVD3 SAMPLED
FP P2 P10
Black Red Red
18 19 20
RVD2 ROVLED HEATER+
P2 P9 P3
Grey Red Red
21 22 23
HOME# DOORLED SAMHOME#
P9 --P10
White N.C. White
24 25 26
HEATERLAMP-
P3 P7
Brown Brown
27 28
GND
---
N.C.
29
LAMP+ VALVE1#
P7 P4
Red White
30 31
P2 P4 P2
Black Violet Yellow
32 33 34
P9 ---
Black N.C.
2 4
D CP DM VADJ VCCS GND VEE D DD3 D DD2 D DD1 D DD0 DISPOFF#
Power Supply TB2 to Digital Control Board J7 PGND Black
1
Main Wire Harness J8 to Rotary Valve P2 RVD1 Yellow
1
+24VDC POWER-FAIL +5VDC
Violet White Red
2 3 4
RVD3 RVD0 RVD2
Red Black Grey
2 3 4
GND +12VDC
Black Orange
5 6
N.C. N.C.
-----
5 6
GND -12VDC
Green Yellow
7 8
Main Wire Harness J8 to Sampler P10 N.C. ---
1
Main Wire Harness J8 to Fluid Pack FP TOUCH Orange
1
N.C. N.C. GND
2 3 4
N.C. SAMHOME# SAMPLED
--White Red
2 3 4
GND GND
Black Black
5 6
1
VALVE0# VALVECOM N.C.
Violet Black ---
2 3 4
ADT1-E ADT1-M
Yellow Blue
5 6
Main Wire Harness J8 to Pump P1 PDM1 Red
1
PDM3 PDM0 PDM2
Yellow Black Grey
2 3 4
Main Wire Harness J8 to Sampler P5 SMD1 Yellow
1
RVD0 VALVE0# RVD1
2 3
PGND PGND
N.C. N.C.
1
KBY4# KBY3# KBY2#
SMD3
Main Wire Harness J8 to Fluid Deck P4 VALVE1# White
SMD3 SMD0 SMD2
Distribution Board J102 to Membrane Keypad KBY5# ---
Red Black Grey
4
-----
5 6
Main Wire Harness J8 to Door Switch P6 DOORSW# White N.C. ---
1 2
PGND
3
Black
----Black
Distribution Board J103 to Optrex Backlight CN1 COM White 1 N.C. N.C. HOT
----Black
2 3 4
Distribution Board J104 to Optrex LCD Panel CN2 D FLM --1 D LP --2 -----
3 4
-----
5 6
---
7 8 9 10 11 12 13
-------------
7
6
5
4
3
H
P2
DB0
1
16
DB1
2
15
DB2
3
14
DB3
4
13
GND
DB8
DB5
6
11
DB6
7
DB7
8
ADB16
15
1
3
14
ADB3
4
13
DB12
ADB4
5
12
DB13
ADB5
6
11
DB14
ADB6
7
10
DB15
ADB7
8
9
ADB17
2
15
ADB18
3
14
ADB10 ADB11 ADB19
4
13
ADB20
5
12
VCC
ADB25
BADS#
2
ADB21
6
11
ADB22
7
10
ADB23
8
RAMADD0
1
16
RAMADD8
RAMADD1
2
15
RAMADD9
RAMADD2
3
14
RAMADD10
RAMADD3
4
13
GND
RAMADD4
5
12
RAS0#
RAMADD5
6
11
RAS1#
RAMADD6
7
10
CASH#
RAMADD7
8
9
CASL#
CLKOUT 3
14
BRSTSYNC 4
13
LATERD#
5
12
CHOKEWR#
6
11
BD/C#
READY#
7
10
DREQ0
BM/IO#
GND
8
9
DACK0#
BW/RD#
ADB15
15
BWR#
BBLE#
ADB14
P6
16
BCLK2
BRD#
BBHE#
ADB13
9
ADB24
1
GND
ADB12
10
TP2
P4 16
ADB9
ADB2
DB11 12
2
ADB1
DB10 5
P3
16
ADB8
DB9
DB4
1
PMPCLK
9
1.8MCLK BUSY
D8
R137 221
R138 221
R123 221
GREEN
GREEN
GREEN
D9
LOGIC ANALYZER TEST POINTS
TP1
1
DIAGNOSTIC LEDS
VCC
P1
2
H
D7
U21 DB15
1
1
3
4
OC
CLK100
DB14
7406D
CLOCK GENERATOR 1 OHM 1W
C75
C64
47 16V
G
C63
0.1
0.1
9
R136 10
X1
14 6 11 7 1
X2
24MHZ
F0
14.3MHZ
F1
12MHZ
F2
1.8MHZ
50.113 MHz
11
CLK2
R150 33 R133 33
MR#
1.8MCLK
RTC / NVRAM
SYSTEM BUS BUFFERS
D10 8 7 5
RST RST#
C96
4
PFI
PFO#
CLK2 RST#
RST# RSTSYNC RSTSYNC
U32
W/RD#
POWER ON RESET
R114 4.75K
D/C# M/IO# RD# WR#
R134
A25
R145 4.75K
PWRFAIL#
221
C65
BARCODE
A24 A23
VCC
A22
100PF
A21 A20
CLK2
A19
install BARCODE pullup OR pulldown only
KU386EXTC
so software sees the right levels when nothing is connected
J1 DSR RXD# RTS TXD# CTS DTR RI
E
SGND
U24
5 6
4
1
BCRXD
R170 4.75K
4.75K
74HCT14
2 8
BCTXD
VCC
9
U23
BCVCC
TMRGATE0
LAMPON#
74HCT14
C7
100PF
NMI
R135
U23
BCODE
C28
R175 4.75K
74HC02D
221
R46
1 6 2 7 3 8 4 9 5
BS8#
221
R49
DCD
VCC
4.75K
4.75K
AUDIOENB DOOR
100PF
R148
110 33 41 90 73 120
A18
RESET
CLK2
BS8#
ADS#
NA#
W/R#
NMI
D/C#
SMI#
M/IO#
SMIACT#
READY# LBA#
101 105 129 102 131 132 104 106
P1.0/DCD0#
RD#
P1.3/DSR0#
WR#
DRIVENB# SHELL
MECHENB#
4.75K
NMIENB#
12 OD13
CS3
R71
CS3#
ACT05
SMD050
CS2# CS1#
U14
111 108 107 126 125 124 123 122
CS0#
P1.1/RTS0#
A25
P2.6/TXD0
A24
P2.7/CTS0#
A23
P1.2/DTR0#
A22
P1.4/RI0#
A21
P1.7/HLDA
A19
P1.6/HOLD
A18
P1.5/LOCK#
A17
P2.4/CS4#
A16
P2.3/CS3#
A15
P2.2/CS2#
A14
P2.1/CS1#
A13
P2.0/CS0#
A12 A11
VCC
R47
1.8MCLK
4.75K
U23
R51
DOORSW#
3
4
INT2 INT1
1.00K
C8
INT3
74HCT14
CLK100
22 40V C29
100PF
TMROUT1 TMROUT0
R164
93 94 95 96
INT4
INT6
5
INT7
6
3
74LS14D
U19
4
OD
92 89 91 PMPCLK
10
RXD3 TXD3
IPCMRX
74HCT14
IPCM PORT
DREQ0
12 13
ANATXD#
P3.6/PWRDWN
A9
P3.5/INT3
A8
P3.4/INT2
A7
P3.3/INT1
A6
P3.2/INT0
A5
P3.1/TMROUT1
A4
P3.0/TMROUT0
A3
TMRCLK0/INT4
A1
TMRGATE0/INT5
BHE#
TMRCLK1/INT6
BLE#
DACK0#
IPCMTX
XCDONE
4.75K XCRLD#
R183 4.75K
BUSY#/TMRGATE2
D14
PEREQ/TMRCLK2
D13
ERROR#/TMROUT2
D12 D11
118 112 117 128 113
DRQ1/RXD1
D10
DACK1#/TXD1
D9
DRQ0/DCD1#
D8
DACK0#/CS5#
D7
EOP#/CTS1#
D6
R147
74HCT14
U23
XCDATA
XINIT
D5
98 77 79 78
STXCLK/DSR1#
D4
SRXCLK/DTR1#
D3
SSIOTX/RTS1#
D2
SSIORX/RI1#
D1 D0
C
76 25 24 26 119
J5
BRST#
VCC
TDI TDO TMS
4.75K
R182
10 8 6 4 2
UCS# CS6#/REFRESH# WDTOUT
ADS#
R176
W/RD# D/C#
221
M/IO#
R189
72 70 68 67 66 65 63 62 61 59 58 57 56 55 54 53 52 51 50 49 48 45 44 43 42 39 37
2 3 5 6 8 9 10 12 13 14 15 16 17 19 20 21 23 24
1B2
1A2
1B3
1A3
1B4
1A4 1A5 1A6
1B7
1A7
1B8
1A8
1B9
1A9
2B1
2A1
2B2
2A2
2B3
2A3
2B4
2A4
2B5
2A5 2A6
2B6 2B7
2A7
2B8
WR#
2A8
2B9
2A9
1DIR
TDI
2DIR
TDO
1OE#
TCK
2OE#
TMS
55 54 52 51 49 48 47 45 44 43 42 41 40 38 37 36 34 33
1 2 114
A16 A25 A15 A24 A14 A23 A13 A22 A12 A21 A11 A20 A10 A19 A9 A18 A8 A17 A7 A16 A6 A15 A5 A14 A4 A13 A3 A12 A2 A11 A1 A10 BHE# A9 BLE#
2 3 5 6 8 9 10 12 13 14 15 16 17 19 20 21 23 24
A8
R174
A6
A16
A13
PFO
A12
VCC
A11
WE
31
BCLK2
ADB12
30
BRST#
ADB11
29
BRSTSYNC
ADB10
28
BADS#
ADB9
27 A9
ADB8
BW/RD#
A10
BD/C#
ADB7
25 A7
BM/IO#
ADB6
24 A6
BRD#
ADB5
23 A5
BWR# ADB25
ADB4
22 A4
ADB3
21 A3
ADB24
ADB2
20 A2
ADB23
ADB1
19 A1
ADB22
BBLE#
18 A0
ADB21
ADB15
3
ADB16
ADB15 ADB14 ADB13
4
CE
6
CHOKEWR#
7
BRD#
8
CS2#
DQ6 DQ5 DQ4 DQ3
10
DB6
11
DB5
12
DB4
13
DB3
DQ1 DQ0
ADB10 ADB9 ADB8 ADB7 ADB6 ADB5 ADB4 ADB3
DB2
14
DQ2
ADB11
DB7
9
DQ7
ADB12
VCC
5
OE
26 A8
2
15
DB1
16
DB0
ADB2 ADB1 BBLE#
17
GND
DS1644P
ADB20
CS1# BRD#
ADB19
U35
ADB18
CHOKEWR#
30 27 29 28
1 26 56 31
1B1
1A1
1B2
1A2
1B3
1A3
1B4
1A4
1B5
1A5
1B6
1A6
1B7
1A7
1B8
1A8
1B9
1A9
2B1
2A1
2B2
2A2
2B3
2A3
2B4
2A4
2B5
2A5
2B6
2A6
2B7
2A7
2B8
2A8
2B9
2A9
1DIR
TDI
2DIR
TDO
ADB14
XTDO
6 75 64 17
ADB13 ADB12
ADB5
71
CS3B#
ADB4
CS4A#
72
CS4A#
CS4B#
69
CS4B#
M/IO#
10
IO#
CS5A#
66
CS5A#
ADS#
11
ADS#
CS5B#
65
CS5B#
RD#
27
RD#
W/RD#
26
W/R#
CHWR#
80
D/C#
95
D/C#
LTRD#
81
BRST#
91
/RESET
SYSTEM DRAM
ADB1
U28
ADB9
UCS# BRD#
CHOKEWR#
ADB7
BWR#
LATERD#
ADB6 ADB5
33 33
44 34
MECHENB#
R196 R198 R195 R197
33 33 33 33
40 43 41 42
R154
33
47
79
RAS0#
CLK2
89
CLK2
RAS1#
78
RAS1#
CLKOUT
92
CLKOUT CASL#
82
CASL#
CASH#
86
CASH#
ADB3 ADB2 ADB1 BBHE# WDTOUT
96
WDTOUT
BBLE# MECHENB#
30 27 29 28
8
A3
RA2
54
RAMADD2
A4
RA3
67
RAMADD3
BHE#
ADB5
22
A5
RA4
74
RAMADD4
BLE#
ADB6
33
A6
RA5
73
RAMADD5
ADB7
100
A7
RA6
57
RAMADD6
ADB8
98
A8
RA7
62
RAMADD7
ADB9
99
A9
RA8
60
RAMADD8
A10
RA9
58
RAMADD9
59
RAMADD10
D3 D2 D1 D0 VCC UCS# CS6#
CLKOUT
4.75K R166 R165 4.75K
30 27 29 28
1A1
1B1
1A2
1B2
1A3
1B3
1A4
1B4
1A5
1B5
1A6
1B6
1A7
TMS
1A8
1B8
1A9
1B9
2A1 2A2
2B2
2A3
2B3
2A4
2B4
2A5
2B5
2A6
2B6
2A7
2B7
2A8
2B8
2A9
2B9
TDI
1DIR
TDO
2DIR
TCK
1OE#
TMS
2OE#
U27
DB15
1MEG = No jumper
DB14
4MEG = No jumper
DB13
ADB10
DB11 DB10
ADB11
3
ADB12
1
ADB13
23
ADB14
18
DB8
ADB15
37
DB7
ADB16
34
DB6
ADB17
24
DB5
ADB18
25
DB4
R213
DB3
4.75K
DB2 DB1
4.75K
R162 4.75K
DB0
R209
VCC
E7
1
BWR#
ADB19
7
ADB20
9
ADB21
16
ADB23
30
A13
DRVEN#
77
A14
RSTSYN
56
A15
POEH
A17
CAS0#
DQ3
CAS1#
DQ4
CAS2#
DQ5
CAS3#
DQ6 DQ7
WE#
DQ8
A0
DQ10
A1
DQ11
A2
DQ12
A3
DQ13
A4
DQ14
A5
DQ15
A6
DQ16
A7
DQ17
A8
DQ18
A9
DQ19
A10
DQ20
55
DRIVENB#
POEL BUSY
A19
DQ23
POEH#
A20 A21
ACK
44
PDEN
A23
DACK0
MEMSIZ0#
15
SIZ0
MEMSIZ1#
12
SIZ1
46 47 48
DB00
BWR
DQ25
4.75K
READY# POEL#
DQ26 DQ27 DQ28
BUSY 43
DREQ0
DQ24
R200
42
PSTB#
A22
19
52
DQ22
RSTSYNC
51
RDYIN
A18
21
VCC DQ29
PSTB# ACK# PDEN
DQ30
MEMSIZ0#
DREQ0 DACK0#
67 68 69 70
DQ31
AT0
35
DB1
PP38
38
DB2
PP49
49
DB3
TDO
TMS
TCLK
TCLK
CARD
DB1 DB2
AT2
NC0
AT3
NC1
ACCESS
8-BIT
16-BIT
DB4 DB5
E6
R211
4.75K
DB6
R215 221
DB7 DB8
4.75K
R212
DB9
BRSTSYNC
DB10
R218
C119
DB11 DB12 DB13
4.75K
R144
DB14
4.75K
47
D0
DB15
R156
4.75K
R140
4.75K
R129
4.75K
R125
4.75K
R142
4.75K
DB0
16V
D1
DB1
D2
DB2 DB3
D3
DB4
D4
DB5 DB6
D5
4.75K
R127
DB7
D
GND4
D6 DB8
R157
4.75K
R139
4.75K
R124
4.75K
R141
4.75K
R126
4.75K
DB9 DB10
FUJITSU FCN565PO68G0231E4
D7 D8
DB11 DB12 DB13 DB14
VCC
DB15
D9
FRAME OF PCMCIA
D10 D11
4.75K
D12
33 45 R173 R155
4.75K
R167 4.75K 4.75K
R168
4.75K
R169
4.75K
CONNECTS TO CHASSIS GND D13 D14
C
D15
256K X 32, 1M X 32, OR 4M X 32 SIMM DRAM R17
PP63
63
SUICI
J7
JUMPER SELECTS SIZE
+24VDCIN
1.0 OHM 1W
1 PGND
R18 2 1.0 OHM 1W
J10
+12VDC
3 PWRFAIL#
R1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
R16
ANA_V+
ALTERA Epm7128SQC100-7
+24VDC +5VANA
T
-5VANA
4 VCC
5 1 6 C33 22 40V
GND 4
7
B
8
-12VDC ANAGND
R146 4.75K
C2 R15
ANARXD#
C41 47 16V
+24VDC
VCC
MC78M05CDT 3 OUT IN U10
+5VANA
1.0 OHM 1W
SMD050
ANAGND
R98 100
R70 10 OHM 1W
1.0 OHM 1W
R44
B
E
16 BSY# 33 WP 36 CD1 42 CE2# 43 RFSH 44 RFU1 45 RFU2 57 RFU3 58 RST 59 WT# 60 RFU4 61 REG# 62 BVD1 63 BVD2 67 CD2 18 VPP1 52 VPP2 17 VCC1 51 VCC2 1 GND1 34 GND2 35 GND3 68
2
BS8#
DB3
AT1
VCC
TMS
DB0
R12833
DB0
PP35
386BIST
R208 4.75K
TDI
DQ1 DQ2
12 13 14 15 16 17 18 28 31 32 19
A16
87
BIST#
RAS1#
F
WE#
2 4 6 8 20 22 24 26 49 51 53 55 57 61 63 65 3 5 7 9 21 23 25 27 50 52 54 56 58 60 62 64
DQ21
RD#
R143 1.00K
DQ0
A12
14
ADB22
RA10
A11
VCC
DB9
R163 4.75K 26 56 31
4
16MEG = Yes jumper
DB12
10 12 13 14 15 16 17 19 20 21 23 24
2B1
DRAM SELECTION
RAS0#
DQ9
TCLK
2 3 5 6 8 9
1B7
BWR#
11 A9 12 A8 22 A7 23 A6 24 A5 25 A4 26 A3 27 A2 28 A1 29 A0 7 CE1# 9 OE# 15
MT8D132M-6 R187 R199
RAS0#
ADB4
U33
55 54 52 51 49 48 47 45 44 43 42 41 40 38 37 36 34 33
BBLE#
ADB8
39
TMS
ADB2
ADB10
2
2OE#
TCK
ADB3
ADB11
ADB4
D4
ADB10
CS3B#
ADB3
D5
ADB11
TDI TDO TCK TMS
RAMADD1
D6
ADB12
CS3A#
83
D7
ADB13
70
RA1
D8
ADB14
CS3A#
A2
D9
ADB15
CS0#
29
D10
ADB16
ADB6 94
CS0#
ADB2
1OE#
A3
D11
ADB17
ADB15
RAMADD0
D12
ADB18
ADB16
85
D13
ADB19
ADB7
55 54 52 51 49 48 47 45 44 43 42 41 40 38 37 36 34 33
RA0
D14
ADB20
U38
A1
D15
ADB21
ADB8
31
R188 221
ADB22
ADB9
ADB1
A1
ADB23
56 A25 55 A24 54 A23 53 A22 50 A21 49 A20 48 A19 47 A18 46 A17 19 A16 20 A15 14 A14 13 A13 21 A12 10 A11 8
TMS
BLEN#
A2
ADB24
DRAM CONTROLLER
TCLK
32
A4
ADB25
U37
BBLE#
221
J6
A10
BHEN#
A5
AT29C010A-90JC 2 A16 21 DB7 D7 3 A15 20 DB6 D6 29 A14 19 DB5 D5 28 A13 18 DB4 D4 4 A12 17 DB3 D3 25 A11 15 DB2 D2 23 A10 14 DB1 D1 26 A9 13 DB0 D0 27 A8 5 A7 6 A6 ADB18 1 7 A5 A18 8 A4 30 ADB17 9 A3 A17 10 A2 11 A1 12 A0 128K x 8 22 CE# 24 RD# 31 WR#
ADB17
50
103
U29
A14
ADB16
1
BBHE#
WDTOUT
FLT#
A15
ADB13
74ABT18245ADL 23 22 21 20 19 18 16 14 13 12 11 10 8 7 6 5
NC2
32
DB1
G
D0
NC1
74ABT18245ADL
A7
DB2
U34
RD#
4.75K
9 7 5 3 1
1A1
1B6
1 26 56 31
221
1B1
1B5
READY#
TRST# CLKOUT
99
A17
TCK
R158
JTAG PORT
DB3
DB0
NC3
ADB14
TMRGATE1/INT7 D15
U11
11
A10
BIST#
ACT05
U23 ANARXD#
P3.7/COMCLK
A2
4.75K
D
87 86 85 84 82 80 75 74
115 40 30 29 27 32 4 34 35
P2.5/RXD0
A20
T
34
74ABT18245ADL
ADS#
100PF
DB4
STORAGE
1.8474 MHz
MAX707CSA 1
R254
PATIENT DATA
33
SUICIDE
R50
DB5
VIDCLK
U25
VCC
F
10
7406D
14.318 MHz
OE
R177 4.75K E5
DB8
DB6
OC
DRIVENB#
VCC
MANUAL RESET
DB9
DB7
VCC
14.318MHZ
DB10
8
U21
MBRS140T
Y1
DB12 DB11
OC
BADS#
VCC2
50MHZ
1
DB13
VCC1
2 16 15 10 9
4
"DRV"
7406D
4 13 3
VCC
"ADS"
U21
AV9154-27CS16
R149
"RST"
41 D15 40 D14 39 D13 38 D12 37 D11 66 D10 65 D9 64 D8 6 D7 5 D6 4 D5 3 D4 2 D3 32 D2 31 D1 30
PCMCIA CARD CONNECTOR
8
R45
4.75K SUICIDE
ANATXD#
E501
1000
Ripple I => 1.5AS12
35V
105c / 3000 hr
1 GND
C34 47 16V
ESR =< 0.1 ohms
3
-5VANA
U9 OUT
IN
MC79M05CDT
SUICI
4
C27 22 40V
R48 10 OHM 1W
VCCPROT
T INT7
SMD050
HTRCTL# FAIL#
6
U23
12 13 74LS14D
A
RIGHT NEXT TO J7
5 74HCT14
POWER SUPPLY CONNECTOR
LOCATE THESE PARTS
NET MUST CARRY 5A CURRENT
DRIVENB#
NOVA Biomedical
ANARST
DRAWING TITLE
Schematic, Cortex Digital Board Processor & Memory
U44
PGND
SIZE
Page S-3
ANALOG BOARD CONNECTOR
DRAWING NUMBER
E DO NOT SCALE DRAWING
8
200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141
7
6
5
4
3
2
REV.
24292-01 LAST MODIFIED ON: 10-5-1998_13:54 SHEET
A
1
OF
3
1
DWG #10-1071B
8
7
6
5
4
3
2
1
U15 MAX211ECAI 12
H
C1-
C55
DCD DSR RXD# RTS TXD# CTS DTR RI SGND
1 6 2 7 3 8 4 9 5
16
0.1
17
R99
25
1.0 OHM 1W
4.75k
R3D
R3Q
R2D
R2Q
R4D
R4Q
T3Q
T3D
T1Q
T1D
R1D
R1Q
T2Q
T2D
R5D
R5Q
T4Q
T4D
SHDN
26 5 22 20 7 8 6 19 21
DCD1# DSR1# RXD1 RTS1# TXD1 CTS1# DTR1# RI1#
1.00K
1.00K MAX211ECAI
C66
IPCM DEBUG PORT
G
2
J3 1 6 2 7 3 8 4 9 5
DSR RXD# RTS TXD# CTS DTR RI SGND
16
R100
17
C70
R3D
T4D
26 5 22 20 7 8 6 19 21
EN#
24
R3Q
R2D
R2Q
R4D
R4Q
T3Q
T3D
T1Q
T1D R1Q
R1D T2Q
T2D
R5D
R5Q
T4Q
SHDN
U20
0.1
IPCMRX
1.00K
2 5 6 9
R29
3 4 7 8
T SMD050
EN
D0
TRI
2
RXD0
D5 D4
RTS0# TXD0
D3
DCD1#
A2
DSR1#
A1
RXD1
A0
CS0#
CTS1#
CS1#
DTR1#
CSP#
PD1 CLK PD2 RESET# PD3 BDO PD4 INTSEL# PD5 INT0 PD6 INT1 PD7 INTP
SLCT
AFD#
BUSY
STB#
PE
SLIN#
ACK# BIDEN
DB12 DB13 DB14 DB15
T R31
221 221 221 221
R6 R32 R7 R33 R8 R34 R9
2 5 6 9
221 221 221 221
EN
R10
221
R36
221
R37
221
R39
221
R11
4 39 44 43 45 60 59 63 65 66 67 68 1
1.8MCLK BRST# R159
1.00K INT2 INT3 PERR_INT BUSY PE_INT
DB0 DB1 DB2 DB3
VCC
R203
4.75k
R153
DB4 DB5 DB6 DB7
4.75k
R35
4.75k
R38
OE
4.75k
1 4.75k
12
C1+
V+
4.75k 221
CLKIN
MD6 MD5
4.75k 1.00k
LDEV#
MD4
LRDY#
MD3
BCFG2
MD1
BCFG1
MD0
VMD11 VMD10 VMD9 VMD8 VMD7 VMD6 VMD5 VMD4 VMD3 VMD2 VMD1 VMD0
6
ADDHI
108 107 106 105 104 103 102 101 100
MA7 MA6
5 4 3 2 1 153 152 151 150 149 148 147 146 145 144 143 141 140 138 137 136 135 134 133
MA5
A23 A22
MA4
A21
MA3
A20
MA2
A19
MA1
A18
MA0
A16
VMA7 VMA6 VMA5 VMA4 VMA3 VMA2 VMA1 VMA0
A15
HCAS#
A14
LCAS#
A13
R225
110 112 113 111
RAS#
WE#
TMROUT0
4.75k
VCC
CLICK
8 9
10
2 3
2
R223 R222 R224
R30
D_VCC
T SMD100
R214 221
R233 4.75K
A10 A9
R231 4.75K
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43
A7 A6 A5 A4 A3 A2 A1 A0
D_VEEN D_YELLOW#
RAS# HCAS#
D_VCC
LCAS#
KBY5#
WE#
KBY3# KBY1#
OE#
KBX3#
U40
KBX2# SPCLK DACLK
A8 A7
27 4 23 1 2 9 3 18 28
DCD3 DSR3 RXD3# RTS3 TXD3# CTS3 DTR3 RI3 R171 R160
1.00K
C97 0.1 C86
C2+
R2D
R2Q
R4D
R4Q
T3Q
T3D
T1Q
T1D
R1D
R1Q
T2Q
T2D
R5D
R5Q
T4Q
25
SHDN
DCD3# DSR3# RXD3 RTS3# TXD3 CTS3# DTR3#
29 31 41 24 26 28 25 30
IPCMRX IPCMTX
8 5 62 12 10 13 11 6
PERR_INT PE_INT RXD4 PGMEN# TXD4 PROGPULSE
221
R179
SYSTEM STATUS LEDS
D12
D13
D_DD2 D_DD3 D_SPARE
D_ENAVDD D_ENABKL D_GREEN# D_VCC KBY4# KBY2# KBY0# KBX1# KBX0#
F
DADAT
SPKR+
A3 A2
N74F244D
A1
2 4 6 8
BLE#
D15
M
D14
FLM LP CP M
53 54 56 55
SHFCLK
D12 D11
ENAVDD
D10
ENAVEE
D9
ENABKL
48 47 76
PSTB#
PNL7
D6
PNL6
D5
PNL5
D4
PNL4
U43
D3
66 65 64 63 61 60 58 57
PNL3
D2
PNL2
D1
PNL1
D0
PNL0
R4
VO2
BYPASS
GND
+IN
VDD
-IN
VO1
BPSTB#
19
R210
E
221 N74F244D
DD3 DD2 DD1 DD0
2 4 6 8
18 16 14 12
17 15 13 11
221 R255
SHUTDOWN
D_ENAVDD D_VEEN D_ENABKL
U41
YELLOW# GREEN#
1 2 3 4
D_FLM D_LP D_CP D_M
3 5 7 9 1
U39
D8 D7
18 16 14 12
17 15 13 11
D13
D_DD3 D_DD2 D_DD1 D_DD0
3 5 7 9 1
D_YELLOW# D_GREEN# D_SPARE
19 R217 221
8 7 6 5
VCC
LM4861M
20.5K
C5
D
+15VDC
C142 C141
0.1
20.5K
R102
3
C44
R3
49.9
0.1
10K
74HC02D U24
4.75k R401
C400
560
1000PF
U16 MIC6211
.1UF
DSR0#
D6 D5
RTS0#
D4 D3
TXD0 CTS0#
D2
DTR0#
D1
RI0#
D0
21 20 19 18 17 16 15 14
DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0
1
4
2
OC
3
5 V+
D7
RXD0
3
U21 PROGPULSE
DCD0#
DCD1#
A2
DSR1#
A1
RXD1
A0
1 R104
R121 20K
3
5 V+
4
C56
11.8VDC
1
D5
10PF
Q5
2 E
4.75K
MIC6211 +2.5VREF
B
1
V-
2
7406D
C 22657
C
V-
2
R122
C58 100PF
2
R120
4.75K
R501
10K +15VDC
VCC
24.9K
CS0#
CTS1#
CS1#
YELLOW# GREEN#
GREEN
D14
XCCLK
R207 R206
DTR1#
CSP#
IORD#
57 56 55 58
32 3 38 36 37
PD0 PD1 CLK PD2 RESET# PD3 BDO PD4 INTSEL# PD5 INT0 PD6 INT1 PD7 INTP ERR#
221 221 221 221
ADB2 ADB1 BBLE#
D500
R119
MBRS140T
1.00K
CS4A# CS4B# CS5B#
R502 324K
RI1# IOWR#
53 52 51 50 49 48 47 46
33 34 35
R118 3.32K
RTS1# TXD1
220
GREEN
D_DD1
24
EN#
U30
GREEN
D_CP D_DD0
SPKR-
U500
T4D
D_FLM
A4
DUART
TRI
1.0 OHM 1W
R172
0.1
17
26 5 22 20 7 8 6 19 21
D_M
A5
TL16C552FN R3Q
D_LP
VCC
C2-
R3D
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44
A6
+15VDC 13
G
J11
A8
14 28 29 13 27
33
A11
R28
CW
1
33 33 33
A12
13 14 15 16 17 18 20 21 23 24 25 26 27 28 30 31
OPTION
1
26 25 24 23 22 19 18 17 16
VMA8
A17
R504
OPTION
R249
BCFG0 MA8
.22uF
BPSTB#
C500
V-
16
0.1
D11
MD7
VMD12
VGA VIDEO CONTROLLER
C1-
C87
GREEN
CPURST
VMD13
VCC
1.5uF
COM3 J4
B
34 35
40 39 38
C143
MAX211ECAI
14 15
VCC
MD9
MT4C16257-70 39 D15 38 D14 37 D13 36 D12 34 D11 33 D10 32 D9 31 D8 10 D7 9 D6 8 D5 7 D4 5 D3 4 D2 3 D1 2 D0
U24
0.1
SHELL
MD10
RD#
DOORBOARD FUSE
16V VMD14
U45
AUDIOENB
C85
1 6 2 7 3 8 4 9 5
MD11
M/IO#
36 37
DB15 DB14 DB13 DB12 DB11 DB10 DB9 DB8 DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0
1
221 221 221
R42
ADS#
FLM
stb used for b_feed output
PRINTER CONNECTOR
SGND
CHOKEWR# LATERD#
R152 1.00K
74HC02D
RI
36 37
221
R41 R12
MD12
4.75k
ADB23 ADB22 ADB21 ADB20 ADB19 ADB18 ADB17 ADB16 ADB15 ADB14 ADB13 ADB12 ADB11 ADB10 ADB9 ADB8 ADB7 ADB6 ADB5 ADB4 ADB3 ADB2 ADB1 BBLE#
R503
R40
BHE#
CS3
CS3A# CS3B# CS5A#
ACK#
221
R13
DTR
32 3 38
221
R5
MD14
H
C90 0.1
VMD15
1.5uF
P_RESET# P_ONLINO P_ONLINI P_FEED# P_BFEED# P_NMI#
CTS
R244
PDEN
25 26 27 28 29 30
RTS
R204
POEL#
P_BUSY P_ACK# P_ERR# P_PE
TXD#
R202
R161
20 21 22 23
RXD#
ADB2 ADB1 BBLE#
TP3
13 14 17 18 11
DSR
R205
33 34 35
U26
U31 3 4 7 8
12 15 16 19
P_STB# P_DR1 P_DR2 P_DR3 P_DR4 P_DR5 P_DR6 P_DR7 P_DR8
MD15
14.3MCLK
LP
N74F373D
R43 SMD100
PV2
RESET
131 130 129 128 127 126 125 124 123 122 121 120 119 117 116 114
MD2
POEH#
2 5 6 24 33
DCD
BCLK2
1
PV1
7 8 9 10 11 12 13 14 15 16 17 18 19
C
R201
READY#
PD0
INIT#
CVCC2
157 158
MD8
RI1#
IORD#
8 9 10 33
BBHE# ADS# M/IO# W/RD#
DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0
RTS1# TXD1
BE3#
OE
11
D
57 56 55 58
DB8 DB9 DB10 DB11
13 14 17 18
12 15 16 19
VCC
E
RI0#
D6
ERR#
P_INIT#
U2
N74F373D
3 4 31 32
53 52 51 50 49 48 47 46
FSELA FSELB FSELC FSELD
RS232 DRIVER/RECEIVERS
1 34
8 5 62 12 10 13 11 6
DCD2# DSR2# RXD2 RTS2# TXD2 CTS2# DTR2# RI2#
R117 25
1.0 OHM 1W
J9
D1
DSR0#
IOWR#
1.00K
F
D2
DTR0#
D7
0.1
C2-
27 4 23 1 2 9 3 18 28
SHELL
CTS0#
21 20 19 18 17 16 15 14
DCD0#
VCC
C2+
CVCC1
MD13
C1-
V-
DCD2 DSR2 RXD2# RTS2 TXD2# CTS2 DTR2 RI2 R151
13
V+
C67
14 15 C68 0.1
DCD
C1+
0.1
E8
COM2
12
159 155
TL16C552FN 29 31 41 24 26 28 25 30
STANDBY#
C71 0.1
R220 10 C115 47
BRSTSYNC VIDCLK
IPCMTX
R72
24
EN#
DUART/PRINTER PORT
C120 0.1
F65535A 11 7
R219 4.75k
C116 47 16V
C121 0.1
R232 4.75K
R230
C2-
27 4 23 1 2 9 3 18 28
DCD1 DSR1 RXD1# RTS1 TXD1# CTS1 DTR1 RI1 R73
SHELL
R216 4.75K
C54 0.1
C2+ V-
J2
VCC 0.1
14 15
COM1
13
V+
C1+
0.1
10
VCC
MBRS140T3
C42
R221
C43
INIT#
SLCT
AFD#
BUSY
STB#
PE
SLIN#
ACK# BIDEN
R192
4 39 44 43 45 60 59 63 65 66 67 68 1
R400
RXD4 R101
6.81K
MBRS140T
U22
CHOKEWR# LATERD#
8 9
TXD4
1.8MCLK BRST#
U21 10
74HC02D
5
OC
D6
1.0 OHM 1W
6
DS2505
R190
U46
option for either package
221
install only one ds2505
TOUCH MEMORY INTERFACE
INT4 INT6 VCC R194 R193 R186 R185 R184
221 221 221 221 221
R181
221
B
TOUCH
7406D
R103
NC
4.75K
DATA GND
3 2
U501
1
1
6 GND
NC
DATA
NC
NC
NC
2 3
5 4
DS2505P
U36
R191
NOVA Biomedical
200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141
DRAWING TITLE
Schematic, Cortex Digital Board Comm, Printer, Display Touch Memory
TP4
A
1
Page S-4
SIZE
DRAWING NUMBER
E DO NOT SCALE DRAWING
8
7
6
5
4
3
2
A
REV.
24292-02 LAST MODIFIED ON: 10-5-1998_16:11 SHEET
2
OF
3
1
DWG #10-1072B
8
7
6
5
4
3
2
1
FUNCTION SELECT H
SYSTEM CLOCK DIVIDER
VCC
KEYBOARD SCANNER MOTOR CONTROLLER
R132
4.75K
R130
4.75K
R131
4.75K
R110
R109
R108
4.75K
VCC
4.75K
4.75K
XC5202-6PQ100C
PGND E1
FSELD
4.75K
E2
FSELC 1 85 22 13 36 11 93 87
DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0
M2
DB6
M1
DB5
M0
SP7
DB1
SP8
AB2
PUMP1
AB1
PUMP2
BBLE
PUMP3
CE#
38 10
81 82 83 84
BRSTSYNC
DB2
PUMP4 PUMPHOLD
RD# PMPCMP12
AB4
SAMP2 CCLK SAMP3 DIN SAMP4 PROG# SAMPHOLD SMPCMP12
79 7 8 9
5 7 11
DENNA
OUT1
IN1
BOOT1
IN2 ENABLE
XXSP7
R253
XXSP8
9 6
C9 0.1
4.75K
R252
4.75K
BOOT2
GND
SENSE
ROV1 TCK ROV2
ROV4 INIT# ROVHOLD HDC ROVGOOSE# LDC# ROVCMP12 DOUT
57 56
PGMEN# PROGPULSE
+24VDC
PUMP1 PUMP2
5 7 11
PUMP3 PUMP4
DENNA
PUMPHOLD
1 C14
SAMP2
5
SAMP3
U11
9 6
VREF
BOOT2
GND
SENSE
1.0 OHM 1W R75 24.9K
ACT05
1
SMPCMP12
U12
COMCLK
SP2
CLK100
SP3
SMPCMP34
9
ROV2
R74 24.9K
ROV3 ACT05
VCC
ROVHOLD
2
U12
ROVCMP12
2
ROV2IN
73
CLICK
SPCLK
+24VDC
5 7 11
DENNA DACLK eepot increment
C15 0.1
U19
XOUT1 XOUT2
16
221
ROVHOME YIN0
74LS14D YIN1
3
ROVHOM2#
YIN2
4
YIN3
R506
221
U19
+24VDC
1
SMD2
OUT2
VREF
BOOT2
GND
SENSE
8 10
L6203
2 5 7 11
VALEN0#
DENNA XOUT0
4 IN1
0.01
XOUT1
C18 0.1
XOUT2
9 6
XOUT3
C19
BOOT2
0.01
GND
SENSE
8 10
74LS14D
YIN4
13
46 59 37 48 49 60
YIN0
U11
YIN1
SMD0
YIN2
14
VCC
YIN3 YIN4
U13
U11 11
9
R83 24.9K U13
4.75K
R93
R228
R229
R238
KBY5#
R239
4.75K
R240
D
R241
4.75K
U42
4.75K
KBY4#
4.75K KBY3#
4.75K 4.75K
KBY2#
+24VDC
R226 R227
221
R234
221
UNUSED GATES
13
221
R237
C21
9 6
9
+24VDC
14 8
U8
U12
5 7 11
2 DENNA
R245
100
KBX2#
R247
100
KBX1#
R246
100
KBX0#
10 11
6
100
R248
LM339AD
C24 13
8 5
9
U12
1 U14
3
9
74LS14D
9 6
0.1
10
74LS14D
4
74LS14D
100PF
100PF
100PF
100PF
100PF
100PF
100PF
100PF
100PF
100PF
C
U44
IN2
R106
U14
1000
U44
BOOT2
GND
SENSE
L6203
LAMP-
8 10
OUT1
IN1
BOOT1
4 IN2
C23
BOOT2 SENSE
ACT05 OD ACT05
13
RVD0
MBRS140T
C124
C125
C129
C130
C131
C132
C135
C137
C136
C138
3
U14
11
ACT05 OD
KEYBOARD CONNECTOR VALEN1#
2 3
DRIVENB#
+15VDC
ROTARY VALVE MOTOR: RUN: 1700 MA
21.5K
U13
C49
11
C74
C107
C39
C139
C35
C51
C57
C140
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
D
18.2K
R84
21.5K
Q3
2 R59
U13
4.75K
IN
OUT3
R92 10K
C3
C1 22
4 0.1
1
C47 220PF
Local +15VDC
5
+2.5V REFERENCE
R86 10K
1 G 3 S
R26
MTD3055EL
R88
+2.5VREF C46 0.001
100K
2 8 4
VCC VCC
R251
VCC
VIN COMP GND
VO TR
1000
C45
MAX872CSA
0.1
0.1
0.1
0.1
C69 0.1
C77 0.1
C37 0.1
C91 0.1
C126 0.1
C122 0.1
C89 MBRS140T
0.1
C80
0.1
VP0610T
0.1
C81
Q1 VALVE0#
6 5 3
0.1 DRIVENB#
1
OD
0.1
C88 0.1
C40 0.1
C38 0.1
C62 0.1
C50 0.1
C60 0.1
C79 0.1
C78 0.1
D4
0.1
C103 0.1
C104 0.1
C95 0.1
C93 0.1
C101 0.1
C102 0.1
C83 0.1
C82 0.1
C72 0.1
C76 0.1
C73 0.1
4
0.1 VALEN0#
C117
+24VDC
C84
C36
DRIVENB#
5 6
D
C6
Q4
U22 4
C134
C123
C111
C110
C114
C106
C105
C118
C112
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
D2 MBRS140T
2 1
40V
1 G
74HC02D
R27 221
0.1
3
S
VALVECOM
R21 324K
MTD3055EL
T
OC
NOVA Biomedical
R22
12 0.1
Schematic, Cortex Digital Board Motor & Valve Drivers
Page S-5
VALVE BOOST
SIZE
DRAWING NUMBER
E DO NOT SCALE DRAWING
8
7
6
5
200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141
DRAWING TITLE
U21
VALBOOST
VALVE DRIVERS
SMD050
C4 7406D
C108
B
U11
22
13
C127
MBRS140T
DENNA
D
0.1
C128
G
0.1
C53
S
0.1
C59
D1
2
TEMP
3
+24VDC C52
C144
2.49K
ACT05
C100
C
220
GND
4.75K
U17 C99
R19
R20
B C98
OUT2
R68
LM339AD
U22
+12VDC
A
+24VDC
2 3 6 7
OUT4
INSTALL ONE DIODE ONLY
C92
OUT1
1
220PF
221 C109
LM317LM
100K
R82 10
VCC 4
74HC02D
C113
U1
4 D3
C94
ROV Motor
40V
ACT05
+24VDC
LAMP IS T3/4 BULB 28V .17A
R69
0.01
4
OD
VALVE1#
1
LAMP DRIVER
C26
1.5uF
U14
MTD3055EL
18.2K
4.75K
R64
S
G R128
4.75K
8 R94
3
Q6
1
74HC02D
DRIVENB#
1.0 OHM 1W R96
D
221
1
GND
13
R66
OUT2
VREF
11 12
LAMPON#
RVD1 C25
4 U22
R67
0.01
0.01
8 10
D
RVD2
3
VS
LAMP+
0.01 1
VREF
+24VDC 1.0 OHM 1W
2
OD
VCC
U44
TP5
C133
MTD3055EL
Jumper 2-3: inverted sense (requires new sw)
C22
OUT2
R90
C61
221
R243 RVD3
LM339AD
VCC
G
HEATER CONTROL POLARITY
3
10 U44
1
74HC02D
R87 20K
BOOT1
ENABLE
11
74LS14D
13
+2.5VREF
4 IN1
2
74LS14D
KBX3#
S
E
DRIVENB#
1.0 OHM 1W
74LS14D
1
74LS14D
3
Q2
R250
11 12
Jumper 1-2: normal sense (original state)
LM339AD
10
12
5 7 11
0.1
74LS14D
11
D
HTRCTL#
OUT1
ENABLE
6
8 74LS14D
221
R236
KBY0#
9
2
DENNA
1000 5
74LS14D
4 U24
220PF
7
U7
VCC
4
221
R235
KBY1#
3
U14
R85 21.5K
L6203 VS
R58 221
Sampler Motor
C48
221
U42
HEATER DRIVER
6
OD ACT05
4.75K
20K
R89 100K
U42
5
R105
LM339AD
221
U42
HEATER-
E500
6
1
R60
U42
4.75K
220PF
R95 21.5K
100K
10
OD
VCC
U42
4.75K
C32
R91
ACT05
ROVLED2
R500
LM339AD
YIN5
VCC
R505
R65
0.01
8
R63 4.75K
HOME DETECTORS R2 221 ROVLED
R14
C20
1.0 OHM 1W R97 24.9K
ACT05
1000 DOORLED
4.75K
12
OD
R25
R61
OUT2
VREF
SMD100
VCC
1
ENABLE
U18
R23
T SMD1
BOOT1
IN2
HEATER+
R62
OUT1
U6
VALEN1#
R24
+24VDC
C17
3
VS
YIN5
SAMLED
0.01
IN2
2
R115
74LS14D
71
DADAT eepot direction
74LS14D
42 43 47 45
BOOT1
XOUT3
SAMHOME
ROVHOME#
9 6
SMD3 C16
1.0 OHM 1W
VALBOOST
VALVE0#
4 IN1
ENABLE
12
8
99
R113 20.5K
F
OUT1
U5
9
VBOOST
XOUT0
E
+2.5VREF 3
VS
CLICK
6
10
100K
L6203
2
U19
74
21.5K
R76 20K
74LS14D
VALVE1#
11
R78
LM339AD
ROVCMP34
U19
GCK2
2
220PF
5
ROVGOOSE#
R54 4.75K
C31
4
ROV4
G
PUMP and SAMPLER MOTORS: RUN: 1300 MA
4.75K
R79 20K
8
OD
PMD1 PMD0 VALVECOM SMD3 SMD2 SMD1 SMD0 RVD3 RVD2 HEATER+ DOORLED HEATERGND VALVE1# VALVE0#
J8
R56
100K
R81
ROVHOM2#
220PF
R77 21.5K
LM339AD
SP4
30
1
7
R55 4.75K
Pump Motor
C30
6
ROV1
23 24 51 72
SAMLED ROVLED ROVHOME# SAMHOME# LAMPLAMP+ RVD0 RVD1
R57
6
OD
VCC
SAMPHOLD
AB3
SAMHOME#
8 10
SAMP4
1
5
R112 221
PMD0
OUT2
SAMP1
74LS14D
U19
GND
0.01 C12 0.1
PMPCMP34
VCC
R116 20.5K
PMD1
4.75K IN2
R80
SP1
PMD3 PMD2 TOUCH DOORSW#
BOOT1
PMPCMP12
90 88 70 89 65 17 69 68
ROVLED2
R53
4 IN1
ENABLE
13
R242
3 C13
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
R52
OUT1
U44
PMPCLK
20.5K
PMD2
0.01
U4
PGMEN#
100 2 50
ADB3
PMD3
1 C11 0.01 8 10
L6203
2 VS
PROGPLS
1.8MCLK CLK100 PMPCLK
3 C10 0.01 4
1.0 OHM 1W
ROVCMP34
F
OUT2
VREF
U11
TDI
39 31 33 76
4.75K
U3
VS
SMPCMP34
ROV3
VCC
L6203
2
TDO
TMS
R111
BRSTSYNC
32 19 35 18 12 95 34
SAMP1
DONE
XINIT
+24VDC
KYBDINT
77 75 55 53
XTDO TDI TCLK TMS
POWER GROUND
PMPCMP34
21 44
XCCLK XCDATA XCRLD# XCDONE
E4
D C B A
ANARST
63 64 58 62 20 86 61
WR#
INT1 ADB4
E3
FSELB FSELA
ANARST
DB3
96 97 94 98
BRD# CHOKEWR#
29 25 27
DB4
DB0
ADB2 ADB1 BBLE# CS6#
G
DB7
H
R107
4
3
2
A
REV.
24292-03 LAST MODIFIED ON: 10-5-1998_13:57
SHEET
3
OF
3
1
DWG #10-1073B
Page S-6
PN 24292
DWG #10-1074B
P10
Page S-7
PN 21642 DWG #10-1075A
8
7
6
5
4
3
2
1
VCC C9 .1UF R16 8.45K
U6 X9C103
DADAT
2
D
8
U/D
7 DACLK
D
VCC
3 CS
VH
INC
VW
VSS
VL
1
R18 1K
5
4
2 E 1
Q2 MMBT4403LT1
B 6
C 3 U8 74ACT05 11 OD 10
VADJ D_ENAVDD
R17 20.5K
R2 1K
R4 1K
G
R21 182K
J104
Q3 VP0610T
2
3
1
D
S
VEE
VCC
1 2 3 4 5 6 7 8
D_FLM D_LP D_CP D_M VADJ VCCS VEE
C16 47 16V
note:
C2 0.1
J101 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43
C
D_VEEN D_YELLOW VCCU KBY5# KBY3# KBY1# KBX3# KBX2# SPCLK DACLK
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44
D_LP D_M D_FLM D_CP D_DD0 D_DD1 D_DD2 D_DD3
R5 1K
R37 1K
SPAREPIN D_ENAVDD D_ENABKL D_GREEN VCCU
5 3 4 2
KBY4# KBY2# KBY0# KBX1# KBX0# SPDAT
CS
SHDN
EXT
REF
OUT
FB
GND
6 7
3
G 1
C
N.C.
Q6 MTD2955E
S
12 11 10 9 13 14
LCD DISPLAY
1 D2 MBRS140T3
4 D
8
R9 324K
VCC
VCCU
C4 47 16V
L1 47UH
C1 0.1
C3 0.1
C17 22 35V
DADAT
TP1
SPKR+
D_DD0 D_DD1 D_DD2 D_DD3
DISPOFF#
VCC
R10 20.5K
L2 .15PH
-23.5V @ 10mA
U5 MAX775
U8 74ACT05 9 OD 8
D_VEEN
Y2 .2W SPKR-
R6 1K
pixels swizzeled here
R11 0.47
VCC
B
J102
R8 1K
R33 1K U8 74ACT05 13 OD 12
VCC SPCLK
R3 1K
R1 1K
R29
U4 E1636C
1K
R35 1K
1
4 HOT
5
ADJ COM GND
D_ENABKL
R19 1K
3 C
Q1 MMBT4401LT1
B 1
C15 47 16V
3
R36 1K
R15 4.75K
R14 4.75K
R12 221
HOT
VIN
U8 74ACT05 1 OD 2
VCC
COM
4 3 2 1
U8 74ACT05 5 OD 6
D_GREEN
R13 221
2 J103
BACKLIGHT
U8 74ACT05 3 OD 4
D_YELLOW
E
KBY5#
1
KBY4#
2
KBY3#
3
KBY2#
4
KBY1#
5
KBY0#
6
KBX3#
7
KBX2#
8
KBX1#
9
KBX0#
10
G_AN
11
G_KA
12
Y_AN
13
Y_KA
14
B
2
KEYPAD
VCC
A
C10 .1UF
C11 .1UF
C12 .1UF
C7 .1UF
NOVA Biomedical
C8 .1UF
A 200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141
DRAWING TITLE
Page S-8
SCHEMATIC, CORTEX DISTRIBITION BOARD
BYPASSING NOTES: Place power supply filter as close to J101 as possible. Bypass each chip with one 0.1uF from Vcc to Gnd unless bypass capacitors are explicitly shown.
SIZE
DRAWING NUMBER
D DO NOT SCALE DRAWING
8
7
6
5
4
3
2
REV.
24727 LAST MODIFIED ON: 7-15-1998_12:16
SHEET
1
OF
2
1 DWG #10-1076B
PN 24291
Page S-9
DWG #10-1077B
12.4K
U4
3 R65
5V REF.
1
LM324
2
BAROMETER 3
4 NC 5 NC U9 6 NC MPX4115AP 2
24.9K
COMMON MODE VOLTAGE CLAMP 2
CLAMP2 3
1
+5V R69 2.5V
1
24.9K
24.9K
R500
R197
1
C34
A
24.9K
100PF
A
100PF
BAR
A
3
D3
C26
100PF
2
SSTPAD5 D1 2 3 1
1000PF C3
C27
R11 BAV99LT1
SSTPAD5 D2 2 1 3
C2
R51
4.75K
A
A 1000PF
SSTPAD5 D5
4.75K
-5V R10
2
1 CLAMP1
D4
2
J6
3
SEE SHEET 3 ISE 5 = PCO2+
SSTPAD5
12
14
200MEG
6
J6
R3 1K
3 5
J6 J6
LMC6484 U1
R167
4
J6
ISE5
13
LOW RES ANALOG BUS
2.5V C12
A
1000PF 7
J6
LMC6484
R166
9
J6 J6
GUARD
3
~CLEAN0,~CLEAN1,PEAK0,PEAK1
REF2
CLEAN/PEAK CTRL BUS
1
200MEG
R2
11
2
U1
1K J6
10 CLAMP2
J6
REF2
R164
14
LMC6484 U1
10 R1
R49
ISE4 24
J6
1
J6
GNDCHS
J6
LMC6484 U1
5
22
J7
7 200MEG
6 REF1
6
U21 LMC6484
10.0K R98
5
J8
A A 2.49K
1
J7
+5V GNDCHS
A 32
J8
J7
200MEG
6
J7
ISE 3 = NA
R171
4
LMC6484 U2
12
14
R9
36
J8
ISE3
J7
10
200MEG
21
J7
ISE 2
R169
19
J7
LMC6484 U2
9
8 R6
GNDCHS
ISE2
9
7
J8
R7
A
J8 J8
11
J8 J8
ISE 1
24
R174
26
200MEG
2 10
3
8
RG
RG
I/V16
AD622
+
REF
5
U23
A
R97A R47
10.0K
R177 24.9MEG
R55
20K
25
J8
1 -
10.0K
ISE1
6
1K J8
R79
C129 0.01
7 R54
8
LMC6484 U2
5
R97
R59 10.0K
1000PF
23
51.1K
14 13
100K
C13
22
J8
A
LMC6484 U21
12 R60
9
J8
1K
10.0K
U21 4 LT1004CD-1 LMC6484 D19 8
C55 .011.235V LT1004
1K
J8
20
cw R180
5
R48
1K
J7
6 7
6
U21 LMC6484
18
J7
1.235V
R83
8
1
J8
.1
R83A LT1004CD-1
LT1004 4 44.2K
40
J8
8
D21
10
1000PF
A
C56
1K 6
R82
39
R94
3
BAS21LT1
D
DG444DY
24.9K
C14
17
2 S
38
J8
5
3
R81
37
J8
3
J7
J8
1
1
1000PF
I/V 1
35
J8
13
1K J7
33
J8
34
J8
D20
U16
C54 21.5K
2
J7
R98A
51.1K
21
J6
I/V1+
3
R50 1K
24
J7
CLAMP1
18
5 8
RG
RG
CLEAN1
REF1 J6
1
U22
3
J8
23
J7
2-
1
6
REF
AD622
2
100K
J8
1000PF
R62
4
J8
C8
A
3+
REF1 20
A
10.0K
2
J8
15
R165
20K
R61
0.01 R80
13
J6
10.0K
C130
9
1K J6
23
J6
8
200MEG
16
J6
24.9MEG
ISE 4 = PH
12
J6
Not used in CORTEX
R178
22
J6 A J6
HIGH RES ANALOG BUS
A/D+,A/D-,F+,F-,E+,E-,D+,D-,C-,ISE[0:5],BAR,2.27V,HX,CAIR,HCTD,I/V[0:2]+,I/V[0:2]-,PK[0:2],TMPR,REF[1:2],VMN,HTR-,BTMP,MH
8
A 1K
C24
A
DG444DY
1000PF
11 S 9 27 29 31
J8
LMC6484 U2
3
1
R56
51.1K
ISE0
R89 100K
2
1K 28
J8
12
R57
LMC6484 U17
S
10
D
LMC6484 U17
PEAK1
8 9 PK1
13 1K
.22
C57
C71
C25 4700pf
1000PF
3
2
1 TP1
+5V
U5
2N7002
14
30
J8
D 10
G Q4
R196
10.0K
D13
J8
ISE 0
R175 200MEG
MMBD914LT1
J8
R90
1 100K
A
R86
A R84 100K
C123 .22UF CTHRU2
R170 DG444DY
24.9MEG
6 S
D 7
+5V
8 100K
C124
PO2+
G Q5 A
2200PF
A
U5
2N7002
PEAK0
R17
R19 100K R8
9
J7
11
J7
5 2RG 8 RG1
R16
2 1
R87
U14
6
S
51.1K 51.1K
3
1K
U13 LMC6484
3
10.0K R74
1
10.0K
2
D10
10
5
LMC6484 U17 PK0
7 6
CLEAN0
C72 4700PF
.22
C58
10.0K 3
100K R88
+5V 1000PF
R91
1
2.49K
A
I/V 0
R92 100K
2
2 3
J7
LMC6484 U17
D
1
BAV99LT1
R18
8
J7
I/V0+ REF
AD622
D14
3+
MMBD914LT1
7
J7
R52 U16
C18
A
16
21.5K
14
R34
R42A
D 6 D11
DG444DY
8
1K C20 .1
LT1004CD-1
R33
LT1004 4 44.2K
9
1.235V
cw R172
6 7
10.0K
8
5 R42
10 U13 LMC6484
C23 .01
C127 .22UF CTHRU2
15 S
24.9K
R85 100K
A
U13 4 LT1004CD-1 LMC6484 D9 6 8
A
LT1004
HIGH RES ANALOG BUS
1.235V
PO251.1K
R5 J7 J7 J7
12
1K R15
14
100K
12
LMC6484 U13
APC CTRL BUS R75
14 13
16
R14 J7 J7
I/V0-
10.0K
13
R20
R4
C118
15
2200PF
A
2 100K
10.0K
3
1 -
8
RG
RG
6
AD622
+
REF
5
R168
U15
NOVA BIOMEDICAL
24.9MEG A
NEW GEN. SENSOR BD. SIZE
REV.
E
24292-01 SHEET 1 OF 4
DWG #10-1078B
Page S-10
DG444DY 11 S 9
D 10
+5V
G Q6 U16
2N7002 R124
12
S
D
LMC6484 U30
10
100K
LMC6484 U30
R125
8 9
14
PEAK2 MMBD914LT1
13
C95
R96A 10.0K
2
C94 4700pf
.22
3
D37
51.1K
R127 10.0K R126
1
A
100K
A
V+ R123
C126
100K
-5V
.22UF CTHRU2
R146 6.81K
PK2 A R158 R162
2
4.99K
10.0K 1 3
100K A-
R159 100K
13 14
D40
MMBD914LT1
R163
A+
R160 100K
12
ACH6
U33
SEE SHEET 3
LMC6484
+5V R161 100K VMN
A/D+,A/D-,F+,F-,E+,E-,D+,D-,C-,ISE[0:5],BAR,2.27V,HX,CAIR,HCTD,I/V[0:2]+,I/V[0:2]-,PK[0:2],TMPR,REF[1:2],VMN,BTMP,MH 2.5V
HI RES ANALOG BUS
~CLEAN2,PEAK2
CLEAN/PEAK CTRL BUS
LOW RES ANALOG BUS
2.5V,AIR[1:4],PHTR,ACH6,MHTRTEMP
GEN CTRL BUS CAIR MHTRTEMP
500mV/C
R130 1K 2 A
IND11
R108
C61
JUMPER J3
16
AD2M
L2
IND11
PAKPRESENT
PHTR
10PF
500mV/C
R115
C65
.22
TMPR
10.0
R116
C69 .15UH
2
74HC14
A
39
J2
1K
MH
R110 AIR2
R111
FLOW CELL TEMP
200
J4
475K
R138 11
2.5V
D7 2
9
3
R31
J4
18.2K 12
C105
8
1
2.5V
1000PF
10
475K
U10 LMC6484
BAV99LT1 A
A
10PF
9.53MEG
HEATER TEMP
HCT/AIR DETECTOR #3
J4 18
1 3
U25 1
12 U10 LMC6484
R143
U33 LMC6484
100
5
2
4.75K
R105
4
J3
38
J2
14
A
R106
R100
13
AIR DETECTOR #2
R144 475K
200
10.0
J4
1 BAV99LT1
+5V
475K
1K
9.53MEG
3
30.1K
T
.22
.15UH
C63 R107
3
J3
AIR1
R133
FLUID_TEMP.
BTMP R101
L7
AD1M
2 J3
1000PF
D26
10PF
AIR DETECTOR #1
C104
+5V
2.5V
J4 C17
R32
HCT-M
J4
7 1K
R145 475K
HCTD
6
HCT/AIR DETECTOR
R137
13
5
R102
6
14
R142
4700pf
HCT1
U6
18.2K
R139
R136
475K
18.2K
18.2K
R141
7 5 U33 LMC6484
LMC6484 10.0K 10PF
C59
A
A
R109
C101 R140
THERMISTOR RETURN C66
AIR DETECTOR #4
R112
2
475K
3
1 AD4M 19
J6
L1
C70
R117
1000PF
10.0K
9.53MEG
AIR3
U10
LMC6484
IND11
.15UH
.22
1K
+5V
6 7
D38
5
2 3
U10 LMC6484
AIR4
1 BAV99LT1
R114 1.25V
D39 2
10.0K
3
A C67
R113 10.0K
1 BAV99LT1
4700pf A
D8
HCT EXCITATION J4
17 HCT-E
[PASS THRU SIGNALS ONLY] 2
+5V
3 1
2.5V
-5V
7 J3
BAV99LT1 1
J3
AD1E
C38 .22
J4
15 AD2E
8 C62
AIR EXCITATION MONITOR
4700pf
9
17
AD4E
2.49K
C39
J3
DG444DY AIR EXCITATION
3 S 1
D 2
10 U5
C64A 1.0UF
J3
J4
24
J4
25
J4
26
LAMPHI
PWM0BUF
R103
.22
23
HEATERLO
HX
C37 .22
J6
J3
PWM0BUF,HSO.3
J4 HEATERHI
LAMPLO
HSO.3
J3
6
R104 49.9
C64B 1.0UF
A
C64C 1.0UF C64D 1.0UF
A
C64E 1.0UF
NOVA BIOMEDICAL NEW GEN. SENSOR BD. SIZE
E
REV.
24292-02 SHEET 2 OF 4
DWG #10-1079B
Page S-11
1 TP2
J2
J2
J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5
29
30 A
J2
27
GND
J2
R181
40
1
AGND
3
ISE0
5
ISE2
REF1
2
ISE1
4
ISE3
6
ISE5
8
11 PK1
PK0
10
13 I/V0+
PK2
12
15 I/V1+
I/V0-
14
17 I/V2+
I/V1-
16
19 MH
I/V2-
18
7
ISE4
9
REF2
J5 J5
TO SHEETS 1 & 2
PEAK[0:2],~CLEAN[0:2]
J5 J5 J5 J5
J5
J5A
J5
23 TMPR
22
25 BTMP
VMN
24
27 AIR1
CAIR
26
29 AIR2
HX
28
31 A+
PHTR
30
33 AGND
A-
32
PGND
34
J5
J2
J2
9
PHTR,ACH[0:2],ACH6,2.5V,AIR[1:4],MHTRTEMP
2.5V
100
A/D+
-5V
1
PMUX0
2
PMUX1
A2
10.0K
PMUX2
4
PMUX3
C90
5
NMUX0
A+ 6 .01
6
NMUX1
5 U30 LMC6484
7
C85
+V 1 F E NC 2 D -V 27 C NC 3 B GND 12 A NC 13 9 U26 8 7 OUT 28 6 5 4 EN 18 3 A3 14 2 A2 15 1 A1 16 0 A0 17
4 5 6 7 8 9 10 11 26 25 24 23 22 21 20 19
EDCBTMP MH VMN REF2 REF1 I/V2I/V1I/V0HCTD CAIR HX
.01
3 U30 LMC6484
3
+5V
.1
+5V
.1
R118 4 5 6 7 8 9 10 11 26 25 24 23 22 21 20 19
E+ D+ SO2DET2 SO2DET1 TMPR PK2 PK1 PK0 I/V2+ I/V1+ I/V0+ HCTD CAIR HX
.1
C84
A NMUXENA0 NMUX3 NMUX2 NMUX1 NMUX0
F E D C B A 9 8 7 6 5 4 3 2 1 0
C106
+V 1 NC 2 -V 27 NC 3 GND 12 NC 13
.1
C103
U32 PMUXENA1
OUT 28
EN A3 A2 A1 A0
A PMUXENA0 PMUX3 PMUX2 PMUX1 PMUX0
18 14 15 16 17
8
14
10
R119
DG406DN
GND
J2
J2
12
31
32
NMUXENA1
+5V
PMUXENA2
PMUXENA3
VDD U31 1 AD7710 24 SCLK DGND 2 23 MCLK IN DVdd 3 22 SDATA MCLK OUT 4 21 A0 /DRDY 5 20 /SYNC /RFS 6 19 /TFS MODE 7 18 AIN1(+) AGND 8 17 Iout AIN1(-) 9 16 AIN2(+) REF OUT 10 15 AIN2(-) REFIN(+) 11 14 Vss REFIN(-) 12 13 AVdd Vbias
MCLKOUT
A/D+
A/D-
D38A 8
R198 2.49k
.1
C97
A R131 121K
R132
4 A
C100 .1
R134
8 10
0.0
U33 LMC6484
A
A
9 10 11 12 7 6 5 4
BAR ISE5 ISE4 ISE3 ISE2 ISE1 ISE0
9
7 6 5 4 3 2 1 0
C109
22PF WRL
FPX-20MHZ
-5V
J2
GND
E+
VDD
SYNC DRDY MODCTRL HSO.5
U34-1
.15UH
LEDSEL0
A J2
J2
J2
16
17
PMUX0
PMUX1
PMUX2
PMUX3
R148
R147
R149
R150
4.75K
4.75K
4.75K
4.75K
23
J1
17
J1
OC CLK 19 1D 1Q 18 2D 2Q 17 3D 3Q 16 4D 4Q 15 5D 5Q 14 6D 6Q 13 7D 7Q 12 8D 8Q 74HC574
9
J1
7
J1
PEAK0 PEAK1 PEAK2 CLEAN0 CLEAN1 CLEAN2 A/DSYNC BLKHTR
R188 SIGOUT 18
U24 1
2
2
3 D33
1 BAV99LT1
LOGIC SUPPLY L3
D28 2
GND
MOD0 MOD1 MOD2 MOD3 PMUXENA1 PMUXENA2 PMUXENA3 NMUXENA1
C134
C138
47UF R128
4.75K
4.75K
MOD0
MOD1
R120
.1
R122
4.75K
J1
IND11 .15UH
1 MMBZ5V6A
R129
14
3
16
J1
15
J1
4.75K
MOD2
MOD3
P2.3 PAKPRESENT SPARE1
GND
PWM0 SPARE2
L6
ANALOG SUPPLIES +5V
VDD
IND11
C137
D27
C135
47UF
2
47UF
C133
.1
3
J1
4
J1
.15UH
L5
HSO.3 C136 IND11 .15UH .1
3 1
5
J1
6
J1
10
J1
MMBZ5V6A A
MODON
18 LEDSEL3
J2
J1
100
74HC14
VDD
MIX/VAL RFS TFS PMUXENA0 NMUXENA0
GND
LEDSEL2
1 11 2 3 4 5 6 7 8 9
PMUX0 PMUX1 PMUX2 PMUX3 NMUX0 NMUX1 NMUX2 NMUX3
PMUX0 PMUX1 PMUX2 PMUX3 NMUX0 NMUX1 NMUX2 NMUX3
U34-14,36,68
LEDSEL1
NMUX3
U27
HSO.1 P1.6 P2.6 HSO.0 P1.5 P1.7 HSO.2 28 29 30 31 32 33 34
C109B .01
C109A .1
15
NMUX2
A PMUXENA1 PMUX2 PMUX1 PMUX0
-5V
+5V IND11 J2
15 16 17 18 19 20 21 22 23 24 25 26 27
PREHTR
+5VF
NMUX1
VDD
PWM2
L8
3 13 14 8 2 15 16 1
A
26
J2
67 66 65 64 63 62 XTAL1 XTAL2 BUS INST ADV RD 61 CLKOUT P3.0 60 /EA NMI P3.1 59 ACH3 P3.2 58 P3.3 57 ACH1 ACH0 P3.4 56 P3.5 55 ACH2 ACH6 P3.6 54 ACH7 P3.7 53 P4.0 52 ACH5 P4.1 51 ACH4 ANGND P4.2 50 VREF P4.3 49 U34 P4.4 48 P4.5 47 EXTINT P4.6 46 /RESET RXD P4.7 45 TXD P2.3 44 READY 43 P1.0 P1.1 P2.4 42 P1.2 WRH 41 8XC196KC PWM1 WRL 40 PWM2 PWM0 39 P2.7 38 HSI.0 VPP 37 HSI.1 HSO.4 HSO.5 HSO.3 35
2 3 4 5 6 7 8 9 10 11 12 13
FAIL PHTR AIR3 AIR4 MHTRTEMP ACH6 2.5V AIR2 AIR1
ALE
28
J2
R154
VDD GND
X1
D+
4.75K
NMUX0
C108
22PF
A
33
4.75K
A
C99
C-
34
R153
4.75K
NOT INSTALLED IN CORTEX
.1
J2
VV+ GND OUT EN A2 A1 A0
DG508ADY
2.5V
SMDXTAL J2
R152
4.75K
.1 U28
2.27V +5V MONITOR
R151 C96
+5V
100K
C98 37
4.75K
-5V
100K
GND
D-
35
R157
4.75K
.1
+5V
5
6
.1 J2
R156
4.75K
DG406DN
R199
+5V
E-
36
R155
4.75K
VDD
A J2
R121
C93
NMUX3
R135 0.0
J2
PMUXENA3 NMUXENA1
PMUXENA2
C91
NMUX2
J2 13
J5A
J5
GND J2
J5A 10
LOW RES ANALOG BUS
J5
J2 11
J5A 8
J5
LT1009S8
J2
J5A
J5A
10.0K J2
6
J5
7 J2
4
SO2DET1
J5
A J2
2
HCTD
J5
1
J2
AIR3
J5A
ANALOG MUX BUS
-5V
J2
HX
9
PMUX[0:3],NMUX[0:3],NMUXENA[0:1],PMUXENA[0:3]
J2
AIR4
5
7 SO2DET2
J5A
A/D-
J2
AGND
3 J5A
20
ACH6
1 J5A
HI RES ANALOG BUS
J5
MHTRTEMP
21 BAR
CLEAN/PEAK CTRL BUS
J5
19 LEDSEL6
MOD CTRL BUS J2
J2
J2
/LEDSEL
20
LEDSEL0 LEDSEL1 LEDSEL2 LEDSEL3 LEDSEL4 LEDSEL5 LEDSEL6 LEDSEL7
LEDSEL7 23 HSO.5 24
P2.3
U29 15 Y0 14 1 Y1 A 13 2 Y2 B 12 3 Y3 C 11 Y4 10 6 Y5 G1 9 4 Y6 G2A 7 5 Y7 G2B 74HC138
MOD0 MOD1 MOD2
21
TO SHEETS 1 & 2
D31 3
4.75K
1 GND
BAV99LT1
LEDDRV
GND
GND
R38
10.0
R193
2
VDD
R64 J2
GEN CTRL BUS MOD2
MODON
U24
VDD
10.0
PREHTR
3
J2
C33
.1 R195 475
9
J4
10
R37
475
150
Q2
Q1
Q3
MMBT4403LT1
MMBT4403LT1
MMBT4403LT1
R35 J4
R27
1K
LED2-A
7
TXD
74HC14
RXD
U24 6 5
2.5V
2
RESET
U24 8 9
74HC14
EXTINT
74HC14
U24 12 13
U24 10 11
74HC14
A
V+ 2 3
1K
5
FAIL
1
PWM0
VP-
7
1K R36
U5 14
LED1
C22
3
12
J1
13
J1
22
J1
11
J1
20
J1
VDD
3 1
2 3 1
BAV99LT1 D29 2 3 1
R189
R182
R184
R191
4.75K
4.75K
4.75K
4.75K
A
8
BAV99LT1 D30
9.09K 22
FAILIN
D34 2
2
R67
J2
INTIN
R190 100
D
DG444DY
RESETIN
R183 100
BAV99LT1 D32
.01UF
100K
15
SRXD
R194 100
1
6 R70
16
200K
LED1-A
5 LM324
J1
R185 100
D35
VDD
U4 R72
S J4
74HC14
J1
21
STXD
R187 100
6
19
SYNCIN
R192 100
U25 5
74HC14
PWM0BUF
R28 24.9 LED1
74HC14
VP+
4 U3 MIC6211B
D6 J4
SYNC
R200 681 1
R21
LED2
U25 4 3
HTRCTL
R186 100
4
R63 10.0
25
LEDCOM R22 10.0
R23
R24
10.0
10.0
A
3
C36
C29
R73 470PF
2 1 GND
49.9K
BAV99LT1
BAV99LT1 D36 2 3 1 BAV99LT1 GND
4700pf R68 1K
NOVA BIOMEDICAL
A
NEW GEN. SENSOR BD. SIZE
REV.
24292-03
E
SHEET 3 OF 4
DWG #10-1080B
Page S-12
VDD
V+ C28
U25 C68
C92
C30
C44
C87
C86
C83
11
.1 .1
.1
.1
.1
.1
.1
U25 10
13
74HC14
.1
12
74HC14
P GND
+5V
C5A
C45
C40
C19
C52
C89
C16
C48
C102
C107
.1
.1
.1
.1
.1
.1
.1
.1
.1
.1
C21
C53
C88
C49
C60
.1
.1
.1
.1
.1
C35
C46
.1
.1
A
C41 .1
C73
C5
.1
.1
C74
C10
.1
.1
C77
C81
C31
C42
C75
C79
.1
.1
.1
.1
.1
.1
C78
C82
C32
C43
C76
C80
.1
.1
.1
.1
.1
.1
C7 .1
C1 .1
-5V
MAKE CONNECTION TO ALL SPARES WITH ETCH SO THAT ECCH CUTS AND JUMPERS CAN USE THESE DEVICES FOR ECO'S J4
1
J4
4
J4
9
5
U4
10
A J4
LM324
U4
12
8 13
14
LM324
+5V
6
-5V
R300
U25 9
MCLKOUT
CLKOUT
8
U12
MOD0 MOD2 MOD3
1 A 2 B 3 C
MODCTRL 6 G1 4 G2A 5 G2B 74HC138
15 14 13 12 11 10 9 7
Y0 Y1 Y2 Y3 Y4 Y5 Y6 Y7
U11
SAM1 B1 SAM1 SAM2 MEAS
1
IN1
3
74HC08
J1
BKGND_IN
9 8 10
74HC08
U11
SAM2 B2
B1 B2
U11
B1 B2
2
8
100
74HC14
4
IN2
6 5
U11
74HC08
SCOPETRIGGER
12 11 13
74HC08
A
DG444DY
R41 2
D
DG444DY
S
14 S 16
3
1
R40
12
LMC6484 U6 14
475K
10.0K U7
SO2-1
D15
13
U7
MEAS
SO2DET1
SAM1
C114 DG444DY 3 S 1
D 2
C119
U8
DG444DY
SAM2
6 S 8
.033UF CTHRU
.22UF CTHRU2
C122 .033UF CTHRU
DG444DY
D 7
11 S 9
D 10
U8
U8 A
A
C121
IN1
C120 .033UF CTHRU
.033UF CTHRU
DG444DY 14 S 16
DG444DY 6 S 8
C115
.22UF CTHRU2
R66
D 15
3
U8
DG444DY 11 S 9
SO2DET2
CTHRU2
CTHRU2
IN2
D 10
U7 A
1
.22UF
.22UF BKGND_IN
LMC6484 U6
2
C116 C117
U7
SO2-2
475K
D 7
MEAS A
MEAS
CO2INPUT
A R30 200K J4
R29 2
9 8
J4
10.0K R26
3
U6 LMC6484
10.0K
C113
.22UF CTHRU2
SO2DET-C
10
R25 36.5K
NOVA BIOMEDICAL A
NEW GEN. SENSOR BD. SIZE
REV.
24292-04
E
SHEET 4 OF 4
DWG #10-1081B
Page S-13
Page S-14
PN 23897 DWG #10-1082B
ITEM 65 5, 40, 58, 38 31 8 52
P/N 22745 24628 24622 24621 20861
DESCRIPTION Flowcell Lamp Kit Sampler Gears Optosensor Motor, Sampler Fluid Fountain
PN 23899
Page S-15
DWG #10-1083B
ITEM 2 3 4 5
P/N 24627 24626 24613 24624
DESCRIPTION Motor Rotary Valve Kit Rotary Valve Gear Rep. Opto Sensor Rotary Valve Rep. Kit Ceramic Rotary Valve
Page S-16
PN 23898 DWG #10-1084B
SPECIFICATION SHEET
TITLE
DWG.NO.
SIZE
PRINTER INTERFACE IF3001 SEIKO
REV
A P/N
24614
SHT 1 OF 1
To Print Head CN01: Head flexible printed circuit connector
CN01
CN02: Motor control connector To Printer Motor
CN03: Input/Output control connector CN02
SW01: Function switch
SW01
a, b:
Position selection notches for mounting to the LTP3245
c:
Holes for mounting to the LTP3245
d:
Two holes for mounting to the outer case (or frame)
d c NOTE: CN01: Lift white plastic part to release print head flat cable
CN03
PT300 P01
To Control Board J9
PT300 GA1
d
Page S-17
REV
ECO#
APPROVED DRP/CO-DRP/DATE
b
EFFECTIVE ECO MGR/DATE
REV
ECO#
APPROVED DRP/CO-DRP/DATE
EFFECTIVE ECO MGR/DATE
For Doc.Cntrl Use Only Disk No.
a
DWG #10-1085B
TITLE
SPECIFICATION SHEET INPUT:
SIGNALS:
OUTPUT:
J1 AMP P.C.B. HEADER P/N 640445-5 PIN 1) AC LINE TB1 PIN 2) N/C PIN 1) PIN 3) AC NEUTRAL PIN 2) PIN 4) N/C PIN 3) PIN 5) AC GROUND J2 AMP P.C.B. HEADER P/N 640456-4 PIN 1) N/C PIN 3) PIN 2) +SENSE PIN 4) J3 AMP P.C.B. HEADER P/N 1-640445-6 PIN 1) OUTPUT #4 PIN 14) PIN 2) OUTPUT #4 PIN 15) PIN 3) OUTPUT #3 PIN 16) PIN 4) OUTPUT #3 PIN 5) OUTPUT #2 TB2 PIN 6) OUTPUT #2 PIN 1) PIN 7) COMMON PIN 2) PIN 8) COMMON PIN 3) PIN 9) COMMON PIN 4) PIN 10) COMMON PIN 5) PIN 11) COMMON PIN 6) PIN 12) COMMON PIN 7) PIN 13) COMMON PIN 8)
SIZE
POWER SUPPLY UNIVERSAL INPUT 5+/-12 24Vo
DWG.NO.
A P/N
24720
SHT 1 OF 1
1
AC LINE AC NEUTRAL AC GROUND
REV
1
TB1 3
F1 T3.5A/250V~
TB2
-SENSE POWER FAIL
OUTPUT #1 OUTPUT #1 OUTPUT #1
V1 ADJ. V2 ADJ.
OUTPUT #4 OUTPUT #3 OUTPUT #2 COMMON COMMON COMMON OUTPUT #1 OUTPUT #1
NC +S -S PF
J2
8.50" (215.9mm) 8.000" (203.2mm)
NOTE: 5A MAXIMUM RECOMMENDED CURRENT PER CONNECTOR PIN
Page S-18
Model (Note A)
Output
Output Voltage
Minimum
Output Current Maximum (Note C)
Maximum (Note D)
Peak
Noise P-P
Total Regulation (Note B)
GPx130D
1 2 3 4
+5V +24V -12V +12V
3A 0A 0A 0A
16A 3.5A 1.2A 1.2A
20A 3.5A 1.2A 1.2A
25A 5A 1.5A 1.5A
50mV 240mV 120mV 120mV
2% 2% 3% 3%
REV
ECO#
APPROVED DRP/CO-DRP/DATE
EFFECTIVE ECO MGR/DATE
REV
ECO#
APPROVED DRP/CO-DRP/DATE
EFFECTIVE ECO MGR/DATE
For Doc.Cntrl Use Only Disk No.
WEIGHT: 2.3 LBS. MAX.
DWG #10-1086A