Tabel 1. Program Quality Control Manual CT Scan (Standar ACR) PROCEDURE Water CT number and standard deviation Artefact evaluation
BATAS TOLERANSI
ALAT UKUR
Nilai HU air 0 ± 5 HU, but must be 0 ± 7 HU. >3 HU diantara artefak berati CT number (dalam ring atau lingkaran)
ACR CT Phantom
Wet laser printer quality control -
Visual cheklist
Tidak ada toleransi, apabila terjadi malfungsi harus segera dilaporkan untuk dibenarkan
DURA SI TES (min) 5
RESPONSIBLE
FREQUENCY MINIMUM
Radiografer
Daily
ACR CT phantom
5 or less
Radiografer
Daily
Densitomet er Laser film QC chart SMPTE test pattern Meliputi :
10
Radiografer
Weekly (jika digunakan untuk alat interpretasi utama)
5
Radiografer
Monthly
Table height indicator functioning Table position indicator functioning Angulation indicator functioning Laser localization light functioning High-voltage cable/other cables safely attached (and not frayed) X-ray on indicator functioning Exposure switch functionin Display window width/level
Dry laser printer quality control -
Display monitor quality control
-
Densitomet er Laser film QC chart SMPTE test pattern SMPTE test pattern or equivalent
10
Radiografer
Monthly (jika digunakan untuk alat interpretasi utama)
5
Radiografer
Monthly
Tabel 2. Program Quality Control CT Scan PARAMETER OR PROCEDURE CT alignment light (IAEA)
CT alignment light
±5 mm
Phantom CT radiografer standar
CT number accuray, image noise, uniformity, artefact (IAEA)
CT number accuracy, image noise, uniformity, artefact
±5 mm from baseline value (ct number accuracy) ±25% of baseline value (image noise) ±10 (uniformity)
water filled test object (or phantom containing a uniform material)
Radiografer + Annually fisikawan medis (phantom head
Dosimeter Phantom CT dose
Fisikawan medis
Radiation dose (IAEA)
PARAMETER
Radiation dose
BATAS TOLERANSI
No artefacts that have the potential to compromise diagnostic confidence (artefact) Dose Calculation in Air : Output at every kVp ±5% from baseline CTDIvol for clinical protocols : Less than Dose Reference Level Verify scanner CTDIvol :
ALAT UKUR
Calibrated electromet er and CTDI pencil ionization chamber (10- or 15cm long) 16-cm (head) CTDI
RESPONSIBLE
FREQUENCY MINIMUM Monthly
and body)
Uji penerimaan Anually Setelah perbaikan
±10%
dosimetry phantom 32-cm (adult body) CTDI dosimetry phantom
kVp and HVL (IAEA)
kVp and HVL
Image Noise
Image Noise
kVp ±5 Detektor nominal HVL >= spesifikasi sesuai regulasi batas toleransi ±25 of baseline phantom
X-ray Beam Width (IAEA)
X-ray Beam Width
Spesifikasi manufacture Within 3mm or 30% of Applied Colimator
Image Slice Width (IAEA)
Image Slice Width
High contrast Spatial resolusi
High Contrast Spatial Resolution
Low contrast resolution (IAEA)
Low Contrast resolution
±0,5 mm for = 5 mm variasi ±1 mm.